Professional Documents
Culture Documents
Needle trap device samplers were used for rapid (60 s) INTRODUCTION
quantitative sampling of short-term exposure limit (STEL) and
peak exposure standard concentrations using a manually oper-
ated pump to collect small volume (10 mL) gas phase samples
containing methylene chloride, benzene, toluene, and tetra-
S everal types of exposure standards for air contaminants
have been promulgated by the U.S. Department of La-
bor, Occupational Safety and Health Administration (OSHA)
chloroethylene vapors. Solventless introduction of chemical
samples for gas chromatography analysis with flame ionization besides the frequently used eight-hour time-weighted average
detection yielded linear results (R2 > 0.99) for vapor standard (TWA) permissible exposure limit (PEL) values. These include
mixtures of the four target analytes ranging from 10% to 200% the short-term exposure limit (STEL), ceiling, and combina-
of their respective nominal STEL or peak exposure standard
tion of ceiling/peak exposure standards that exist for a number
concentrations. Needle trap samplers showed ≥86% recovery
(as GC-FID peak area responses) following 14-day storage of air contaminants. As described in 29 CFR 1910.1000, “an
at room temperature compared to the same samplers analyzed employee’s exposure to any substance in Table Z-1, the expo-
immediately, with better recovery values observed with shorter sure limit of which is preceded by a ‘C,’ shall at no time exceed
storage (≥95% at room temperature for seven days, except the exposure limit given for that substance. If instantaneous
for methylene chloride) or with storage at 4◦ C. Calibration
monitoring is not feasible, then the ceiling shall be assessed as
for quantitation of concentrations of benzene, toluene, and
tetrachloroethylene was shown to be possible with the use of a 15-minute time weighted average exposure which shall not
an internal standard to account for injector discrimination be exceeded at any time during the working day.”(1)
between the solventless NTD approach and injections of tar- For a number of stressors listed in Table Z-2 of the OSHA
get analytes in carbon disulfide. Due to the simple sampling air contaminants standard, ceiling exposure standards are given
method (no field calibration and battery-free pumping) and
that should not be exceeded except in some cases for a specified
the avoidance of solvent dilution, a needle trap sampling
approach could simplify sample collection and analysis to duration and not to exceed a higher stated contaminant concen-
chromatographically determine nearly instantaneous (1 min) tration (i.e., the “peak concentration”). For airborne stressors
exposure concentrations. with peak type exposure standards, the peak concentration
should never be exceeded, while exposure greater than a STEL
Keywords needle trap device, gas chromatography, flame ioniza- standard averaged over 15 min is also unacceptable.
tion detector, air sampling, ceiling, short term exposure
limit
Commonly Used Methods to Measure STEL
and Ceiling/Peak Exposures
Address correspondence to: Philip A. Smith, U.S. Department of
Labor, OSHA, Health Response Team, Salt Lake Technical Center, Current methods for evaluating both STEL and ceiling/peak
8660 S. Sandy Parkway, Sandy, Utah 84070; e-mail: smith.philip.a@ exposure compliance typically rely on sorbent tube sampling.
dol.gov This is followed by solvent desorption and analysis by gas
Applies to industry segments exempt from 1 ppm 8-h TWA PEL value found in 29 CFR 1910.1028(7)
E15-min STEL, 29 CFR 1910.1052(8)
FCeiling standard is 200 ppm, 300 ppm acceptable maximum peak above the acceptable ceiling concentration for an 8-h shift (5 min in any 3 h), 29 CFR
Table Z-2
H15-min sample period (50 mL/min) at the STEL standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(2)
I15-min sample period (50 mL/min) at the STEL standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(3)
J1-min sample period (50 mL/min) at the peak standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(4)
K1-min sample period (50 mL/min) at the peak standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(5)
chromatography (GC). Benzene, methylene chloride, tetra- being sampled is not appreciably changed due to the small
chloroethylene, and toluene samples for STEL or ceiling/peak scale of extraction.
exposure compliance are usually collected on coconut shell Qualitative screening is an important practical use for SPME
charcoal (CSC) tubes to determine short-duration interval con- due to the disadvantages associated with quantitative SPME
centrations following analysis by GC with flame ionization sampling: the need for gas phase calibration curves created
detection (GC-FID).(2–5) Solvent desorption requires the col- from sampling target analyte vapors under the same envi-
lection of sufficient analyte to allow quantitative GC-FID ronmental conditions as for field samples(11) and the poten-
analysis. The respective masses collected over a 15-min pe- tial for competitive displacement of relatively low affinity
riod at the STEL or over a 1 min period for peak exposure volatile organic compounds from the sorbent surface when
measurement as applicable for each of the four analytes de- adsorptive fiber coatings are used.(12) However, Augusto et al.
scribed above are shown in Table I. Vapor hazard ratio values showed, for brief sampling periods, that SPME can be used
calculated per Popendorf (6) are also provided, along with the as a rapid quantitative field sampler without the need to create
relevant OSHA exposure standards for these analytes.(1,7,8) vapor standards if a constant flow of air is passed across a
SPME fiber with an adsorptive coating, (13) and this makes it
Possible Methods to Measure STEL a candidate of interest for STEL and ceiling/peak exposure
and Ceiling/Peak Exposures Without Solvent sampling.
Dilution Sampling with a needle trap device (NTD) sampler is not
The pace of sampling method research using small sorbent a microextraction technique, but may be compared to a typi-
volumes has increased since the early 1990s when solid phase cal sorbent sampling approach—although on a much smaller
microextraction (SPME) was first described.(9) “Microextrac- scale. If a few milligrams of sorbent material are immobilized
tion is defined as an extraction technique where the volume within a thin needle, analytes may be exhaustively extracted
of the extracting phase is very small in relation to the volume from a finite (relatively small) volume of air passed through
of the sample, and extraction of analytes is not exhaustive. the needle. The use of very small quantities of sorbent packed
In many cases only a small fraction of the initial analyte is in a needle allows introduction of an NTD sampler into a
extracted for analysis.”(10) The SPME technique is usually standard GC injection port for thermal desorption of trapped
employed for passive sampling, where analytes in the system analytes, eliminating the need for solvent desorption. The
being sampled diffuse into a thin layer of polymer material history and theory of NTD sampler development and exam-
coated over a small fused silica fiber. Eventually, equilibrium ple applications have been reviewed recently by Lord and
may be attained between the coating material and the system Pawliszyn.(14)
being sampled where no further net analyte uptake occurs. Pawliszyn’s research group has demonstrated that NTD
With SPME, the overall analyte concentration in the system samplers may be used for passive quantitative air sampling,(15)
The fronting observed for the toluene and tetrachloroethy- for all four target analytes, with each of the triplicate analysis
lene peaks indicates that the column capacity was exceeded RSD values below 4%. Results from laboratory analysis of
for these analytes, and this is due to the relatively high con- CSC tube samples from Tedlar bags showed excellent linearity
centrations needed to match the respective allowable peak when plotted against expected analyte concentrations (R2 <
concentration values for these two analytes. The use of an NTD 0.99). These samples for the Tedlar bag analyte concentration
sampler to measure the wide range of concentrations studied estimates ranged from 18.0 to 25.5% lower than expected
(5 to 500 ppm, Table II) in a workplace would either require values for unknown reasons. It is unlikely that this deviation
predetermined analyte-specific sample volumes, or different from theoretical in the measured concentration was due to
injector split ratios and/or a GC column with thicker stationary incomplete analyte desorption during laboratory analysis, as
phase film to improve the chromatography. This is important the analytical results reported consider the known CS2 /CSC
to avoid co-elution of potential interferents, and these details desorption efficiencies which are updated on a regular basis
should be optimized during robust method development. by the laboratory.
TABLE III. Results of Quantitative NTD Sampling with Internal Standard Correction for Calibration Using
Standards Dissolved in CS2 (see Figure 8)
Concentration Range R2 Slope
Benzene 5–100 ppm 0.998 0.94
Toluene 50–1100 ppm 0.997 1.07
Tetrachloroethylene 30–750 ppm 0.997 0.99