You are on page 1of 11

Journal of Occupational and Environmental Hygiene, 10: 674–684

ISSN: 1545-9624 print / 1545-9632 online


DOI: 10.1080/15459624.2013.831982

Short Duration Needle Trap Sampling with Gas


Chromatography Analysis to Determine Nearly
Instantaneous Concentrations of Selected Organic Vapor
Contaminants
Simon J. Strating,1 Theodore J. Juarez,1 Michael E. Stevens Jr.,1
Duvel W. White,1 and Philip A. Smith1,2
1
Department of Preventive Medicine and Biometrics, Uniformed Services University of the Health
Sciences, Bethesda, Maryland
2
Health Response Team, Occupational Safety and Health Administration, U.S. Department of Labor, Salt
Lake Technical Center, Sandy, Utah

Needle trap device samplers were used for rapid (60 s) INTRODUCTION
quantitative sampling of short-term exposure limit (STEL) and
peak exposure standard concentrations using a manually oper-
ated pump to collect small volume (10 mL) gas phase samples
containing methylene chloride, benzene, toluene, and tetra-
S everal types of exposure standards for air contaminants
have been promulgated by the U.S. Department of La-
bor, Occupational Safety and Health Administration (OSHA)
chloroethylene vapors. Solventless introduction of chemical
samples for gas chromatography analysis with flame ionization besides the frequently used eight-hour time-weighted average
detection yielded linear results (R2 > 0.99) for vapor standard (TWA) permissible exposure limit (PEL) values. These include
mixtures of the four target analytes ranging from 10% to 200% the short-term exposure limit (STEL), ceiling, and combina-
of their respective nominal STEL or peak exposure standard
tion of ceiling/peak exposure standards that exist for a number
concentrations. Needle trap samplers showed ≥86% recovery
(as GC-FID peak area responses) following 14-day storage of air contaminants. As described in 29 CFR 1910.1000, “an
at room temperature compared to the same samplers analyzed employee’s exposure to any substance in Table Z-1, the expo-
immediately, with better recovery values observed with shorter sure limit of which is preceded by a ‘C,’ shall at no time exceed
storage (≥95% at room temperature for seven days, except the exposure limit given for that substance. If instantaneous
for methylene chloride) or with storage at 4◦ C. Calibration
monitoring is not feasible, then the ceiling shall be assessed as
for quantitation of concentrations of benzene, toluene, and
tetrachloroethylene was shown to be possible with the use of a 15-minute time weighted average exposure which shall not
an internal standard to account for injector discrimination be exceeded at any time during the working day.”(1)
between the solventless NTD approach and injections of tar- For a number of stressors listed in Table Z-2 of the OSHA
get analytes in carbon disulfide. Due to the simple sampling air contaminants standard, ceiling exposure standards are given
method (no field calibration and battery-free pumping) and
that should not be exceeded except in some cases for a specified
the avoidance of solvent dilution, a needle trap sampling
approach could simplify sample collection and analysis to duration and not to exceed a higher stated contaminant concen-
chromatographically determine nearly instantaneous (1 min) tration (i.e., the “peak concentration”). For airborne stressors
exposure concentrations. with peak type exposure standards, the peak concentration
should never be exceeded, while exposure greater than a STEL
Keywords needle trap device, gas chromatography, flame ioniza- standard averaged over 15 min is also unacceptable.
tion detector, air sampling, ceiling, short term exposure
limit
Commonly Used Methods to Measure STEL
and Ceiling/Peak Exposures
Address correspondence to: Philip A. Smith, U.S. Department of
Labor, OSHA, Health Response Team, Salt Lake Technical Center, Current methods for evaluating both STEL and ceiling/peak
8660 S. Sandy Parkway, Sandy, Utah 84070; e-mail: smith.philip.a@ exposure compliance typically rely on sorbent tube sampling.
dol.gov This is followed by solvent desorption and analysis by gas

674 Journal of Occupational and Environmental Hygiene December 2013


TABLE I. Analytes Studied
Allowable STEL Allowable Peak Vapor Hazard Mass of analyte
Analyte Exposure (ppm) Exposure (ppm) Ratio (VHR) A injected (ng)B
Benzene 5C 50D 125,263 12H
Methylene Chloride 125E —– 22,895 326 I
Tetrachloroethylene —– 300F 245 101 J
Toluene —– 500G 187 94 K
AVapor hazard ratio (VHR) is defined per Popendorf (6) as VHR = Pvapor × 106/TLV × 760; relevant 8-hour OSHA permissible exposure limit standards used in
place of TLVR , Pvapor = vapor pressure, mm Hg at 25 ◦ C
BCalculated by considering sampling duration, volumetric sampling rate, and sample dilution from solvent desorption following current methods for STEL and

peak exposure determination using charcole tube sampling


C15-min STEL, 29 CFR 1910.1028(7)
DCeiling standard is 25 ppm, 50 ppm acceptable maximum peak above the acceptable ceiling concentration for an 8-h shift, 29 CFR 1910.1000(1) Table Z-2.

Applies to industry segments exempt from 1 ppm 8-h TWA PEL value found in 29 CFR 1910.1028(7)
E15-min STEL, 29 CFR 1910.1052(8)
FCeiling standard is 200 ppm, 300 ppm acceptable maximum peak above the acceptable ceiling concentration for an 8-h shift (5 min in any 3 h), 29 CFR

1910.1000(1) Table Z-2


GCeiling standard is 300 ppm, 500 ppm acceptable maximum peak above the acceptable ceiling concentration for an 8-h shift (10 min), 29 CFR 1910.1000(1)

Table Z-2
H15-min sample period (50 mL/min) at the STEL standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(2)
I15-min sample period (50 mL/min) at the STEL standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(3)
J1-min sample period (50 mL/min) at the peak standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(4)
K1-min sample period (50 mL/min) at the peak standard concentration, 1.0 mL solvent extraction and injection of 1.0 μL solvent extract for GC-FID analysis(5)

chromatography (GC). Benzene, methylene chloride, tetra- being sampled is not appreciably changed due to the small
chloroethylene, and toluene samples for STEL or ceiling/peak scale of extraction.
exposure compliance are usually collected on coconut shell Qualitative screening is an important practical use for SPME
charcoal (CSC) tubes to determine short-duration interval con- due to the disadvantages associated with quantitative SPME
centrations following analysis by GC with flame ionization sampling: the need for gas phase calibration curves created
detection (GC-FID).(2–5) Solvent desorption requires the col- from sampling target analyte vapors under the same envi-
lection of sufficient analyte to allow quantitative GC-FID ronmental conditions as for field samples(11) and the poten-
analysis. The respective masses collected over a 15-min pe- tial for competitive displacement of relatively low affinity
riod at the STEL or over a 1 min period for peak exposure volatile organic compounds from the sorbent surface when
measurement as applicable for each of the four analytes de- adsorptive fiber coatings are used.(12) However, Augusto et al.
scribed above are shown in Table I. Vapor hazard ratio values showed, for brief sampling periods, that SPME can be used
calculated per Popendorf (6) are also provided, along with the as a rapid quantitative field sampler without the need to create
relevant OSHA exposure standards for these analytes.(1,7,8) vapor standards if a constant flow of air is passed across a
SPME fiber with an adsorptive coating, (13) and this makes it
Possible Methods to Measure STEL a candidate of interest for STEL and ceiling/peak exposure
and Ceiling/Peak Exposures Without Solvent sampling.
Dilution Sampling with a needle trap device (NTD) sampler is not
The pace of sampling method research using small sorbent a microextraction technique, but may be compared to a typi-
volumes has increased since the early 1990s when solid phase cal sorbent sampling approach—although on a much smaller
microextraction (SPME) was first described.(9) “Microextrac- scale. If a few milligrams of sorbent material are immobilized
tion is defined as an extraction technique where the volume within a thin needle, analytes may be exhaustively extracted
of the extracting phase is very small in relation to the volume from a finite (relatively small) volume of air passed through
of the sample, and extraction of analytes is not exhaustive. the needle. The use of very small quantities of sorbent packed
In many cases only a small fraction of the initial analyte is in a needle allows introduction of an NTD sampler into a
extracted for analysis.”(10) The SPME technique is usually standard GC injection port for thermal desorption of trapped
employed for passive sampling, where analytes in the system analytes, eliminating the need for solvent desorption. The
being sampled diffuse into a thin layer of polymer material history and theory of NTD sampler development and exam-
coated over a small fused silica fiber. Eventually, equilibrium ple applications have been reviewed recently by Lord and
may be attained between the coating material and the system Pawliszyn.(14)
being sampled where no further net analyte uptake occurs. Pawliszyn’s research group has demonstrated that NTD
With SPME, the overall analyte concentration in the system samplers may be used for passive quantitative air sampling,(15)

Journal of Occupational and Environmental Hygiene December 2013 675


even over periods as long as 95 h for benzene, toluene, ethyl-
benzene, and xylenes (BTEX).(16) A diffusive sampling rate
of up to 16 μL air/min was described for NTD samplers used
this way to passively sample BTEX analytes.(16) In the case
of benzene at the STEL concentration of 5 ppm, a 1 mL
air sample actively drawn through an NTD sampler packed FIGURE 1. Kitagawa AP-20N Gas Sample Pump shown with
with suitable sorbent would result in trapping of 16 ng of Shinwa NeedlEx NTD sampler. The pump handle (1) can be pulled
benzene (assuming no breakthrough), easily detectable by and locked to provide 10, 50, or 100 mL sample volumes. The NTD
sampler (2 ) restricts the flow rate to approximately 10 mL per min.
thermal desorption followed by GC-FID analysis. Passively A red “pop-out” indicator (3 ) releases for visual confirmation that
sampling at the rate described by Gong et al.(16) for 15 min sampling is complete. The overall length with needle trap sampler
would result in trapping < 1 ng of benzene. The mass of attached is 34 cm. For flow-assisted backflushed desorption the
benzene that would be trapped with only a 1 mL active NTD NTD sampler (2) is detached from the pump and is connected
sample is similar to the mass of benzene injected for GC-FID at its Luer fitting to a flexible gas line with carrier gas flow
diverted through the NTD sampler in the heated injector of the
analysis using current 15 min CSC tube sampling and solvent GC instrument used.
desorption methods for STEL exposure assessment (Table
I).(2) Thus, for short duration quantitative sampling of the
kind needed to determine STEL compliance for benzene (with
the lowest short interval exposure standard values among the Japan, and Shinwa, Japan. The Kitagawa AP-20N hand pump
analytes listed in Table I), active NTD sampling will potentially is designed to sample 10, 50, or 100 mL volumes of air through
trap sufficient analyte for analysis, even with a much smaller a Shinwa NeedlEx NTD sampler at a rate of approximately
air sampling volume compared to current methods. 10 mL per min. For the proposed measurement of air con-
A volumetric field sample collected over a very brief dura- taminant concentrations relevant to STEL or peak exposure
tion is relatively easy to obtain rapidly using an NTD sampler standards, the AP-20N pump was operated using a 10-mL draw
and a manually operated pump, and a breathing zone sampling volume through an organic vapor type NeedlEx NTD sampler.
approach capable of sampling over a relevant period (15 min The sorbent provided in the NeedlEx sampler is a copolymer
in the case of the benzene and methylene chloride STEL of methacrylic acid and ethylene glycol dimethacrylate (6 μL
standards) could be developed if NTD-based sampling and volume).(17) Figure 1 shows the Kitagawa pump with NeedlEx
analysis methods show promise for collection of short duration NTD sampler and Figure 2 details the arrangement of the Leur
occupational exposure information. Laboratory analysis of fitting connection for pump and backflush desorption flow,
NTD samples by thermal desorption in the heated inlet of a sorbent bed, and sample inlet for this NTD sampler.
calibrated gas chromatograph would provide for analysis with
no sample preparation while also avoiding the use of solvents. Quantitative Standards
This latter point eliminates the expense and environmental Vapor Standards for Determination of Breakthrough,
risks related to collection and disposal of contaminated solvent, NTD Sampler Variability, Optimal Inlet Liner, Analysis
while also eliminating solvent background that may obscure Linearity, and Sample Stability
early eluting analytes. Tedlar bags with a single polypropylene fitting (5 L volume,
The purpose of this study was to examine the usefulness SKC, Eighty Four, Pa.) were prepared with a mixture of the
of active NTD sampling for GC-FID analysis to determine
airborne contaminant concentrations over short durations. The
analytes listed in Table I were selected for study as these are
amenable to trapping by a simple single-bed sorbent system
for analysis on a non-polar liquid film capillary GC
column.

MATERIALS AND METHODS


Chemicals
Methylene chloride (99.9%), benzene (99%), toluene
(99.8%), and ethylbenzene (99%) were purchased from Fisher
Scientific (Pittsburgh, Pa.). Tetrachloroethylene (99%) and FIGURE 2. Photo of Shinwa NeedlEx NTD sampler (Figure 2a)
carbon disulfide (CS2 , 99.9%) were purchased from Acros and diagram of the NTD sampler components (Figure 2b, not to
Organics (Geel, Belgium). scale): (1) Luer-Lok fitting for connection to pump or desorption gas
flow, (2 ) narrow needle shaft measuring 0.7 mm O.D. by 85 mm
Needle Trap Device and Pump length, (3 ) sorbent bed containing 6 μL of copolymer, and (4)
side-hole for sample entry and analyte desorption when the NTD
Pumps and NTD samplers were purchased from Kitagawa sampler is inserted into a heated GC injector.
America (Pompton Lakes, N.J.—a distributor for Kitagawa,

676 Journal of Occupational and Environmental Hygiene December 2013


TABLE II. Quantitative Vapor Standards DB-5 column assembly (Agilent Technologies, Wilmington,
Del.). The column had an internal diameter of 0.25 mm, with a
Nominal Percent of STEL or Peak Air Contaminant length of 30 m and 0.25 μm df . The GC injector was operated at
ConcentrationA 200◦ C, and used a 4-mm gooseneck liner and a 20:1 split ratio
10 50 100 150 200 for most analyses (exceptions described below). The column
was operated at a constant pressure of 21.9 psi providing an
Benzene (ppm) 0.5 2.5 5 7.5 10 average linear velocity of 38 cm/s at 50◦ C column temperature.
Methylene Chloride 12.5 62.5 125 187.5 250 At the start of each run for the initial experiments, the LTM
(ppm) column module was held at 50◦ C for 1.0 min followed by
Tetrachloroethylene 30 150 300 450 600 a 30◦ C/min ramp to 110◦ C, at which point all four analytes
(ppm) were observed to have eluted from the column. For calibration
Toluene (ppm) 50 250 500 750 1000 using standards in solvent the initial 1.0 min hold at 50◦ C was
Stock solution 2 μL 9 μL 19 μL 28 μL 38 μL followed by a 30◦ C/min ramp to 120◦ C to ensure elution of the
volume injected ethylbenzene internal standard. In both cases the final portion
into 5.0 L bagB of a run involved a 300◦ C/min ramp to a final temperature of
AAs applicable to each analyte 250◦ C with a 0.5 min hold time.
BStock solution mixture ratio by volume was 5 parts benzene; 87 parts After collection of an NTD sample the device was dis-
methylene chloride; 333 parts tetrachloroethylene; and 576 parts toluene. connected from the hand-operated pump shown in Figure 1.
It was then attached at its Luer type fitting (see Figures 1
four target analytes having nominal concentrations expressed and 2) to a flexible small-diameter gas line connected to one
as a percentage of their respective STEL or peak exposure of two automated external valves that were operated by the
standard concentrations as shown in Table II. Sweet et al. GC instrument control software. Carrier gas flow could be
previously demonstrated that tetrachloroethylene was found diverted through the valve connected to the Luer fitting with
to be stable in Tedlar bags in concentrations between 5 and the second valve (normally supplying the GC injector) closed.
100 ppm for at least 12 h.(18) One day prior to each experiment, Upon introduction of an NTD sample into the heated injector,
bags to be used were flushed with dry nitrogen gas three times, carrier gas was diverted through the NTD sampler for 60 s.
and were then filled with 5.0 L of dry nitrogen using a large This arrangement provided for backflushed desorption with a
gastight syringe (1 L volume, Hamilton, Reno, Nev.). Bags that total desorption gas volume of 21 mL (1.0 mL/min through
were to contain non-humid vapor standards were not handled the GC column, and 20 mL/min split flow) by maintaining the
further, while bags for humid standards each received 85 μL of constant carrier gas pressure that provided injector carrier gas
deionized water, with the expectation that a portion of the water flow through the NTD sampler during the 60 s period of carrier
would adhere to the inside wall of the bag. The bags were then gas diversion.
left to equilibrate overnight at 23◦ C yielding 0% and 40–50% At the conclusion of this desorption time, the GC instrument
relative humidity (0 mg/L and 9–10 mg/L absolute humidity, control software cycled both external valves so that carrier gas
respectively), as measured by length of stain indicator tubes flow to the NTD sampler was stopped and reopened to the
for water vapor (1–40 mg/L, Draäger, Luebeck, Germany). carrier gas supply line found on the injector for the remainder
One hour prior to sampling events, analyte mixture was of a GC run. The 1.0 min thermal desorption eliminated
injected into each bag from a stock solution containing the four carryover and negated the need for NTD sampler conditioning
target analytes. For multi-hour experiments where more than between samples as confirmed by a lack of response from
one bag would be sampled, the addition of stock solution was blank analyses run at regular intervals with NTD samplers
staggered in order to maintain between 1 and 2 h of contact desorbed using this approach.
time between vapor standards and bag walls. No noticeable
decreases in analyte concentrations were observed using this Needle Trap Device Performance
approach. Breakthrough
Each bag concentration was validated with triplicate CSC To study breakthrough, two NTD samplers were connected
tube samples analyzed offsite by an American Industrial Hy- in series, and a 10 mL sample was collected and analyzed for
giene Association (AIHA)-accredited laboratory using OSHA each of the five STEL or peak exposure standard concentra-
methods that included correction for extraction efficiencies.(2–5) tions listed in Table II. Increasingly larger volumes were sam-
Unless otherwise stated, experiments utilized 50% nominal pled in 10 mL increments from a 100% nominal STEL/peak
relative humidity with 100% nominal STEL and peak vapor standard concentration bag. As the Kitagawa AP-20N pump
standard concentrations, and 10 mL of air was sampled from only samples with 10, 50, or 100 mL volumes, repeated 10 mL
each bag through an NTD sampler. draws through the front NTD sampler were taken without
analysis while the back NTD sampler was analyzed after each
Gas Chromatography Analysis 10 mL draw. This allowed separate breakthrough analyses of
An Agilent 6890N GC-FID instrument was used in this the back NTD sampler at cumulative volumes through the front
study, employing a resistively heated low thermal mass (LTM) NTD sampler of 10, 20, 30, 40, and 50 mL.

Journal of Occupational and Environmental Hygiene December 2013 677


NTD Sampler Variability stored at room temperature (23◦ C). Nine NTD samplers were
Variability was tested using nine NTD samplers with trip- used for each time-course, with all nine samplers analyzed
licate analyses for each from a 100% nominal concentration on day 0 followed immediately by a second sample drawn
standard bag (Table II). Run-to-run variability on a single nee- for time-course storage. At day 1 following sampling: NTD
dle was calculated individually for each NTD sampler using samplers 1, 2, and 3 were analyzed for comparison to the values
the triplicate relative standard deviation (RSD) values for GC- obtained from analysis of the same NTD samplers on day 0.
FID peak areas. Needle-to-needle variability was calculated This was repeated for NTD samplers 4, 5, and 6 analyzed again
from the overall RSD of the triplicate analyses for all nine on the 7th day of storage, and for NTD samplers 7, 8, and 9
NTD samplers (n = 27 analyses). analyzed on the 14th day of storage.
Additionally, variability was compared between NTD sam-
plers conditioned 10 days prior to use and freshly conditioned Calibration Using Standards in Solvent
NTD samplers. Nine NTD samplers were conditioned and After examining breakthrough, NTD sampler variability,
capped for 10 days with room temperature storage (23◦ C). Af- optimal inlet liner, analysis linearity, and sample stability for
ter the 10-day storage, the nine NTD samplers were uncapped the four analytes listed in Table II, work was completed to
and used to collect 10 mL samples from a 100% nominal calibrate the observed GC-FID responses for NTD sampling
STEL/peak standard concentration bag. Analysis/conditioning of benzene, tetrachloroethylene, and toluene. Several variables
of each NTD sampler was immediately followed by a sec- affected the slope of peak areas plotted against the mass of
ond 10 mL sample collection from the same 100% nominal analyte trapped on an NTD sampler relative to injection of
STEL/peak standard concentration bag. The second analysis standards dissolved in CS2 , including injector temperature and
provided data for freshly conditioned NTD samplers to com- the depth of NTD insertion in the heated injector.
pare to the samples taken after storing a sampler for 10 days Assuming that injector discrimination was the same for all
following conditioning. Since the data were paired for each of analytes and an internal standard (ethylbenzene), whether from
the nine NTD samplers, two-tailed paired t-tests were used to an NTD sampler or a solvent injection, quantitation of NTD
compare accuracy. Precision was compared using the resulting analyses was possible without creating gas phase standards.
RSD values. To accomplish this, injections were completed with the target
analytes and internal standard in CS2 at nine concentrations
GC Liner and Sample Humidity Effect on to exceed the range of expected analyte recovery masses, and
Chromatography the results were used to determine a response ratio (R) for
Using a single NTD sampler, four commercially available each analyte relative to the internal standard. These solvent
GC injector liner designs were evaluated for quantitative run- injection calibration curves were also used to determine the
to-run variability, qualitative chromatography aspects, and hu- concentration of analytes within dry Tedlar bag samples by on-
midity effects of water vapor expansion in the heated injector. site analysis of desorption solvent obtained from CSC sorbent
The evaluated liners included 2 mm straight, 2 mm gooseneck, tube samples.(2,4,5) Gas from the same dry Tedlar bag samples
4 mm straight, and 4 mm gooseneck types. After each liner was that also contained a known internal standard concentration
installed, two sets of triplicate samples were collected from (1.73 μg of this compound collected with a 10 mL NTD sample
100% nominal STEL/peak standard concentration bags—one volume) was pulled through an NTD sampler, and the values
triplicate sample set from a non-humid vapor standard bag and of R for each analyte were used with Equation 1 to determine
the other triplicate sample set from a humid vapor standard bag. the amount of each analyte trapped on an NTD sampler.
Quantitative run-to-run variability was evaluated using RSD AX AI S
comparisons. Chromatography quality and humidity effects =R× (1)
MX MI S
were evaluated visually with overlaid chromatograms.
In Equation 1, AX is the integrated peak area for analyte X,
Linearity AIS is the integrated peak area for the internal standard, MIS is
The linearity for NTD sampling was evaluated for the five the known mass of internal standard desorbed from the NTD
vapor standard concentrations of each analyte listed in Table sampler, and MX is the unknown mass of analyte X desorbed
II. Triplicate NTD samples were drawn at each concentration from the NTD sampler.
relative to the STEL and peak exposure standards. This yielded
15 data points for each target analyte. Each target analyte set RESULTS
was then plotted separately and evaluated for linear regression
fit using least squares R2 values. Breakthrough
No breakthrough of detectable analyte on the backup NTD
Stability Time-Course sampler was observed in any of the 10 mL samples collected
Two 14-day stability time-course studies were conducted and analyzed from 10% to 200% nominal STEL/peak standard
on NTD samples drawn from a 100% nominal STEL/peak concentration bags. In sample volumes greater than 10 mL
standard concentration bag. One set of NTD samples was from the 100% nominal STEL/peak standard concentration
stored refrigerated (4◦ C) and the other set of samples was bag, breakthrough was observed for methylene chloride at

678 Journal of Occupational and Environmental Hygiene December 2013


RSD values of GC-FID peak areas) on a single needle ranged
from 1 to 5%, while needle-to-needle variability was 5% for
three of the analytes and 10% for methylene chloride. Figure 4
shows a graphical representation for these results with run-
to-run variability of each needle represented by error bars
showing the RSD, and needle-to-needle variability represented
by the y-axis showing differences for each NTD sampler
from the overall mean. The bar chart is useful for quick
visual evaluation of individual NTD sampler performance in
comparison to the others but the RSD values mentioned above
provide the best overall needle-to-needle statistical evaluation.
For example, comparing methylene chloride results from
triplicate samples on NTD sampler #2 to triplicate samples on
NTD sampler #8 yields the largest difference for needle-to-
needle comparison (+16% to −11% compared to the overall
mean) but the overall methylene chloride needle-to-needle
RSD was 10%. Since run-to-run variability on a single nee-
dle was lower than needle-to-needle variability, it may be
prudent to pull a very small volume (e.g., 1 mL) from a
stable, concentrated vapor of an analyte that would serve as
FIGURE 3. Overlaid GC-FID chromatograms resulting from
analysis of a second NTD sampler placed in series to capture an internal standard prior to sampling. Field samples were
analyte breaking through a front NTD sampler used to sample not analyzed in the scope of this preliminary study, but in
a humid 100% nominal STEL/peak standard concentration bag: future studies post-analysis correction of some type would
125 ppm methylene chloride (peak 1); 5 ppm benzene (peak 2, be warranted due in part to the observed needle-to-needle
peak not present until 50 mL sample volume -not shown here);
variability.
500 ppm toluene (peak 3 ); and 300 ppm tetrachloroethylene
(peak 4, peak not present until 50 mL sample volume - not Addressing the question related to post-conditioning stor-
shown here). All chromatograms are magnified 10x for peak 3 . age prior to sample collection, results for experiments to detect
The lightly dotted 10 mL sample volume chromatogram shows changes in 10-day stored vs. fresh NTD samplers are shown
no FID response for any of the analytes. The solid 20 mL in Figure 5. The t-tests failed to show a statistical difference
sample volume chromatogram shows a slight FID response for
between the means of GC-FID peak areas for 10-day stored and
methylene chloride, indicating the beginning of breakthrough.
The dashed 30 mL sample volume chromatogram shows more fresh NTD samplers at an alpha value of 0.05. Additionally,
pronounced breakthrough of methylene chloride and slight toluene the variabilities were similar for both conditioning approaches
breakthrough. The dotted 40 mL sample volume chromatogram (4–6% RSD for three analytes and 9–11% RSD for methylene
shows increased breakthrough of methylene chloride and toluene. chloride) as shown by the error bars in Figure 5.

20 mL sample volume, then for toluene at 30 mL sample


volume, and finally for all four analytes at 50 mL sample GC Inlet Liner and Sample Humidity Effects on
volume. Although toluene was sampled at its 500 ppm peak Chromatography
concentration and methylene chloride was sampled at its lower During the early stages of the work described in this article,
125 ppm STEL, it was not surprising to observe methylene a 2-mm straight liner was used in the GC-FID instrument
chloride as the first analyte to break through as other estab- because the volume of a 0.75-mm liner designed for SPME
lished sampling methods acknowledge difficulties in capturing was thought to be too small to allow for analyte thermal
this analyte. The documentation for OSHA Method 59 calls for expansion from the relatively high capacity NTD sampler,
collection of methylene chloride through three CSC sections and a 4-mm liner designed for liquid injection was thought to
(350 mg in each section) in series to avoid breakthrough be too large for rapid analyte transfer to the column during
loss. The documentation for this method also states that high thermal desorption. NeedlEx NTD sampler instructions call
(80%) humidity increases methylene chloride breakthrough in for a 10 s delay between insertion of the NTD sampler into
charcoal tubes.(3) Figure 3 chromatograms show breakthrough the heated GC injector and pushing an inert gas through the
study results for sample volumes ranging from 10 to 40 mL. sampler to achieve flow assisted thermal desorption.(19)
However, the four target analytes all exhibited peak dou-
NTD Sampler Variability bling with this approach, which was thought to be caused
In preliminary studies, the use of a single NeedlEx NTD by an initial thermal expansion peak followed by a second
sampler produced highly repeatable results, whereas sampling flow-assisted peak. To eliminate the extra peak, the carrier gas
conducted with a different NeedlEx NTD for each subsequent bypass valve described previously was used to allow simulta-
sample collection and analysis was found to produce results neous NTD sampler insertion and gas flow across the heated
that were slightly less reproducible. Run-to-run variability (as NTD sorbent bed. Chromatography of the four target analytes

Journal of Occupational and Environmental Hygiene December 2013 679


FIGURE 4. Results from the use of 9 separate NTD samplers with triplicate samples for each drawn from a humid 100% STEL/peak standard
concentration bag: methylene chloride (white bars); benzene (light gray bars); toluene (gray bars); and tetrachloroethylene (dark gray bars).
The X-axis shows the 4 separate analytes on each of the 9 individual NTD samplers. The Y-axis represents the needle-to-needle GC-FID peak
area variability by showing the difference of each triplicate NTD sampler average for each analyte from the overall average for all 27 analysis
events. The error bars show run-to-run GC-FID peak area variability (triplicate RSD) for a single NTD sampler.

improved so that the peak doubling was only observed at the


apex of peaks. In order to eliminate the double apex, the GC
injector split ratio was increased from 1:10 to 1:20 and column
flow was decreased from 2.0 mL per min to 1.0 mL per min
in order to maintain a nearly identical flow rate through the
NTD sampler during the period of carrier gas flow through
the heated NTD sampler. This change lessened the apex peak
doubling, leaving only a fronting peak shoulder as shown in
Figure 6a.
However, analysis of humid samples resulted in a slight
increase in the fronting shoulder (e.g., the toluene peak in
Figure 6a). The use of a 4-mm straight liner eliminated fronting
peak shoulders but GC peak widths increased, apparently due
to slower analyte transfer caused by the larger volume of this
liner. Figure 6b shows that the 2-mm gooseneck liner improved
the chromatography peak sharpness and reduced the fronting
peak shoulder associated with inefficient analyte transfer that
was observed with the 2-mm straight liner. Figure 6c shows
that the 4-mm gooseneck liner completely eliminated fronting
peak shoulders due to slow analyte transfer from the NTD FIGURE 5. Effect of conditioning NTD samplers immediately
sorbent, with minimal loss of peak sharpness. The RSD for prior to use, and use following ten-day post-conditioning storage.
replicate samples of the four target analytes decreased as well GC-FID peak areas for NTD samplers used immediately following
conditioning were compared with peak areas for the same
from 4–7% RSD for the 2 mm liners to 1–2% RSD using the
samplers used following ten days of post-conditioning storage.
4 mm gooseneck liner.

680 Journal of Occupational and Environmental Hygiene December 2013


FIGURE 6. GC injector liner and humidity effects. Overlaid GC-FID chromatograms resulting from 10 mL NTD samples drawn from both
humid and non-humid 100% STEL/peak standard concentration bags: methylene chloride (peak 1); benzene (peak 2 ); toluene (peak 3 ); and
tetrachloroethylene (peak 4 ). A 6x magnified view for peaks 1 and 2 is provided for each set of overlaid chromatograms. Figure 6a shows a
slight humidity effect (larger fronting peak shoulders) in a chromatogram comparison of humid and dry samples introduced into a 2-mm straight
liner (RSD values ranging from 3.5 to 5.9%). Figure 6b shows reduced fronting peak shoulder size and humidity effect in a comparison of humid
and dry samples introduced into a 2-mm gooseneck liner (RSD values ranging from 5.4 to 6.7%). Figure 6c shows elimination of fronting peak
shoulders and humidity effect in a comparison of humid and dry samples introduced into a 4-mm gooseneck liner (RSD values ranging from 0.3
to 2.2%). The GC peak fronting observed for toluene and tetrachloroethylene are related to the relatively large concentrations of these analytes
(e.g., the toluene peak exposure standard concentration is two orders of magnitude higher than the benzene STEL concentration).

The fronting observed for the toluene and tetrachloroethy- for all four target analytes, with each of the triplicate analysis
lene peaks indicates that the column capacity was exceeded RSD values below 4%. Results from laboratory analysis of
for these analytes, and this is due to the relatively high con- CSC tube samples from Tedlar bags showed excellent linearity
centrations needed to match the respective allowable peak when plotted against expected analyte concentrations (R2 <
concentration values for these two analytes. The use of an NTD 0.99). These samples for the Tedlar bag analyte concentration
sampler to measure the wide range of concentrations studied estimates ranged from 18.0 to 25.5% lower than expected
(5 to 500 ppm, Table II) in a workplace would either require values for unknown reasons. It is unlikely that this deviation
predetermined analyte-specific sample volumes, or different from theoretical in the measured concentration was due to
injector split ratios and/or a GC column with thicker stationary incomplete analyte desorption during laboratory analysis, as
phase film to improve the chromatography. This is important the analytical results reported consider the known CS2 /CSC
to avoid co-elution of potential interferents, and these details desorption efficiencies which are updated on a regular basis
should be optimized during robust method development. by the laboratory.

Linearity Stability Time-Course


The use of a single NTD sampler with the five vapor stan- No statistically significant difference was observed between
dard bag concentrations yielded R2 values greater than 0.99 samples stored at 4◦ C and those stored at 23◦ C over a period

Journal of Occupational and Environmental Hygiene December 2013 681


FIGURE 8. Results of quantitative NTD sampling with internal
standard correction for calibration using standards dissolved in
CS2 (See Table III).

Calibration Using Liquid Standards


The ability to calibrate the GC-FID instrument response for
NTD samples using standards dissolved in CS2 and internal
standard correction is demonstrated by Figure 8 where NTD
results are plotted against CSC results for the same bags. With-
out internal standard correction, plots for the same analytes
produced regression lines with good linearity, but with slope
FIGURE 7. Storage stability of the four target analytes over values far from unity. This was due to injector discrimination
14 days at 4◦ C and 23◦ C yielded the following respective average
between the two types of sample introduction approaches, and
recoveries (as GC-FID peak area responses) at 14 days: 80.1
and 86.0% for methylene chloride (Figure 8a); 97.9 and 89.9% for a possible explanation is expansion of the CS2 solvent delivery
benzene (Figure 8b); 96.1 and 88.1% for toluene (Figure 8c); and vehicle in the heated injector while this does not occur to the
97.3 and 87.5% for tetrachloroethylene (Figure 8d). At 7 days, same degree for the NTD samples. The slopes of the regression
recoveries were ≥95.0% for all analytes with the exception of lines for the three analytes with internal standard correction
methylene chloride, which was recovered at 91.7% compared to
shown in Figure 8 are close to 1, with high correlation between
samples analyzed on day 0. Two-tailed paired t-tests (alpha =
0.05) failed to show a statistical difference between 4◦ C and 23◦ C the NTD and CSC results (see Table III).
storage for the four analytes.
DISCUSSION

of 14 days. At room temperature, GC-FID peak areas for NTD


samplers were 92.9% or greater for the sampled target analytes
T he primary advantage to the use of an NTD sampler for
collection of a nearly instantaneous sample is the ability
to potentially desorb all of the trapped analyte into a GC instru-
with storage for 7 days but fell to 86.0% or greater with storage ment, while traditional methods that rely on solvent desorption
of 14 days. Results are shown graphically in Figure 7. require longer sampling times that would not be considered to

TABLE III. Results of Quantitative NTD Sampling with Internal Standard Correction for Calibration Using
Standards Dissolved in CS2 (see Figure 8)
Concentration Range R2 Slope
Benzene 5–100 ppm 0.998 0.94
Toluene 50–1100 ppm 0.997 1.07
Tetrachloroethylene 30–750 ppm 0.997 0.99

682 Journal of Occupational and Environmental Hygiene December 2013


be nearly instantaneous. For example, OSHA Method 111 for resulting in faster than normal septum degradation and occa-
toluene recommends a sampling time >10 min for diffusive sionally clogging of the NTD sampler entry/exit port. Addi-
sampling to detect exposures above the 300 ppm ceiling value. tionally, the recommended injector temperature limit of 200◦ C
While the use of diffusive samplers for periods as short as 1 min for thermal desorption of the copolymer sorbent bed could
was shown to be possible for determination of the peak toluene limit the usefulness of the method for larger, less volatile
concentration (500 ppm) under carefully controlled conditions, compounds.
as this type of device begins to sample immediately upon However, in preliminary work we observed that tributyl
removal from its sealed container the potential for significant phosphate (a relatively much less volatile analyte compared
sampling error exists for such short duration sampling periods to those listed in Table I) was successfully desorbed at 200◦ C
and the “shortest recommended sampling time for toluene without observable carryover from the copolymer sorbent bed
using diffusive samplers was determined to be 10 minutes.”(5) using 60 s NTD flow assisted injection. No evidence was seen
The volumetric nature of NTD sampling, its speed, and for any degradation of the sorbent in the NTD samplers used
the elimination of solvent use for desorption deal with these (e.g., spurious GC peaks, loss of sampling capacity, increased
problems. While rapid volumetric sampling across sorbent GC baseline) during this study with injections carried out at the
packed within a traditional thermal desorption tube would pro- recommended temperature of 200◦ C. If needed, other sorbent
vide the same benefits, dedicated thermal desorption hardware bed materials that may be heated to temperatures greater than
would be required for analysis. The relatively high flow from 300◦ C are available, and further work to validate the approach
the desorption of a large (slowly heated) tube with internal discussed here for very rapid sampling of STEL or ceiling/peak
diameter of for example, 4 mm, must be concentrated onto concentrations should occur after considering the wide range
a smaller secondary trap that may be rapidly heated, with of sorbents that were not available in commercially available
a much lower volumetric flow to closely match that of a NTD samplers at the time this work was completed.
modern capillary GC column typically having a diameter The need for internal standard application to the NTD
≤0.32 mm.(20) sampler introduces additional problems, although this could
The use of a hand pump or vacuum reservoir and a constant be solved by pulling a very small precisely metered volume
restricted flow for NTD sampling could simplify short duration of a constant-concentration vapor standard through the NTD
air contaminant collection in the field by eliminating the need sampler in the field prior to sampling. With knowledge of the
for calibration of a battery operated sampling pump. In addition values of R for target analytes relative to the internal standard
to the potential use of NTD sampling for target analytes obtained from the analysis of standards dissolved in solvent,
with OSHA STEL or ceiling/peak exposure standards, the the calculation of target analyte air concentration could be
approach may also be of interest to those wishing to compare completed.
worker exposures to TLV values published by the American
R
Lastly, a substantial limitation of this NTD sampling and
Conference of Governmental Industrial Hygienists (ACGIH). R
analysis method is the need for a carrier gas bypass valve
Ceiling or STEL TLV values have been promulgated for more
R
to be retrofitted to a GC instrument to avoid split GC peaks
than 150 chemical substances.(21) observed for the light analytes studied here when the bypass
Although rapid collection of ceiling or peak exposure sam- valve was not used. This is a minor modification compared to
ples is advantageous, a problem exists with the method de- the addition of an entire thermal desorption module that would
scribed here for measurement of STEL values, as a sample be necessary to use traditional thermal desorption tubes, and
averaging time of 15 min is needed for these measurements. the issue of split peaks would be a lesser problem (if at all)
An approach to remedy this while avoiding breakthrough could for semivolatile analytes trapped on an NTD sampler. For
entail the use of a critical orifice to allow for 15 min sample analytes of this type, nearly instantaneous thermal desorption
collection at a constant flow rate, with a total air sample volume is less important compared to light solvent analytes as the less
of 10 mL or less. Also, the ceiling/peak and STEL exposure volatile analytes tend to focus at the proximal GC column
standards apply to personal breathing zone exposures, and end where the column interfaces with a heated injector, as
modifications to the sampling system would be needed to temperature programmed GC can employ a relatively cool
allow for mounting of a combined NTD sampler and vacuum initial column temperature (e.g., 40◦ C).
reservoir for worker breathing zone sampling, with subsequent Another approach that has shown promise in simplifying
activation of sampling at a time when a maximum exposure is NTD injection in a standard GC instrument injector is the use
expected. of a side-hole located above the sorbent packing in a NTD
Another limitation of the methods described here is NTD sampler. Sampling is through a hole in the tip of the NTD,
sampler cost relative to traditional sampling media and the identical to the NeedlEx NTD samplers used in this work. This
current lack of validated methods. However, reusability of is accomplished with vacuum applied to the side-hole above
NTD samplers (estimated 25–30 uses)(19) along with solvent- the sorbent. For analysis, the tip of the NTD bottoms out in an
less laboratory sample desorption are likely to mitigate the injector liner designed to seal the tip, and the pressurized and
cost issue to some degree. Another limitation is related to heated injector gas in the liner then pushes down through the
the side-hole sample entry/exit port (Figure 2) that tends to side-hole, through the sorbent, and out the tip of the NTD for
damage a typical injector septum during insertion or removal, backflushed thermal desorption of trapped analytes.(14)

Journal of Occupational and Environmental Hygiene December 2013 683


CONCLUSION 3. “Methylene Chloride: Method 59: OSHA Sampling and Analytical
Methods.” Available at http://www.osha.gov/dts/sltc/methods/organic/
org059/org059.html (accessed May 1, 2012).
A t the time of this preliminary study, the NTD samplers
used were obtainable from a commercial source, but
NTD samplers with the wide range of well-known sorbents
4. “Tetrachloroethylene Trichloroethylene: Method 1001. OSHA Sam-
pling and Analytical Methods.” Available at http://www.osha.gov/dts/sltc/
methods/mdt/mdt1001/1001.html (accessed May 1, 2012).
used for thermal desorption tube sampling were not. Further 5. “Toluene: Method 111: OSHA Sampling and Analytical
validation studies are needed for NTD sampling, and these Methods.” Available at http://www.osha.gov/dts/sltc/methods/organic/
should examine carefully selected sorbents that are already org111/org111.html (accessed May 1, 2012).
6. Popendorf, W.: Source control via substitution. In Industrial Hygiene
available for use in traditional thermal desorption tubes, -
Control of Airborne Chemical Hazards. Boca Raton, Fla: CRC/Taylor &
possibly using a multi-bed design. Robust testing with op- Francis, 2006. pp. 185–214.
timized sorbent types should include dynamic standard gen- 7. “Benzene,” Code of Federal Regulations Title 29, Part 1910.1028, 2013.
eration with relative humidity levels ≥ 80%. This preliminary 8. “Methylene Chloride,” Code of Federal Regulations Title 29, Part
study demonstrated the fundamental ability to vary rapidly 1910.1052, 2013.
9. Arthur, C.L., and J. Pawliszyn: Solid-phase microextraction with
sample target chemical vapors using an NTD sampler for
thermal-desorption using fused-silica optical fibers. Anal. Chem.
analytes with both STEL and ceiling/peak type exposure stan- 62:2145–2148 (1990).
dards. The GC-FID results obtained were highly repeatable 10. Lord, H., and J. Pawliszyn: Microextraction of drugs. J. Chromatogr. A
and linear from 10% to 200% of the nominal respective OSHA 902:17–63 (2000).
STEL or peak exposure standard concentrations for benzene, 11. Bocchini, P., D. Dello Monaco, R. Pozzi, et al.: Solid-phase microex-
traction coupled to gas chromatography with flame ionization detection
methylene chloride, toluene, and tetrachloroethylene. Analytes
for monitoring of organic solvents in working areas. Microchim. Acta
remained stable after collection on the NTD sampler for at 165:271–278 (2009).
least 7 days at room temperature. A short duration expo- 12. Mocho, P., V. Larroque, and V. Desauziers: Modelling of toluene solid-
sure assessment approach where nearly instantaneous analyte phase microextraction for indoor air sampling. Anal. Bioanal. Chem.
concentration information is desired could potentially benefit 388:147–156 (2007).
13. Augusto, F., J. Koziel, and J. Pawliszyn: Design and validation of
from NTD sampling and its associated low-effort, solventless
portable SPME devices for rapid field air sampling and diffusion-based
sample preparation/introduction method without the purchase calibration. Anal. Chem. 73:481–486 (2001).
and maintenance of dedicated thermal desorption equipment. 14. Lord, H.L., W. Zhan, and J. Pawliszyn: Fundamentals and applications
of needle trap devices: A critical review. Anal. Chim. Acta 677:3–18
(2010).
ACKNOWLEDGMENTS
15. Cai, J., G. Ouyang, Y. Gong, and J. Pawliszyn: Simultaneous sampling
and analysis for vapor mercury in ambient air using needle trap

T his work received financial support from the U.S. Army


Center for Environmental Health Research and material
support from the U.S. Department of Labor, OSHA Salt Lake
coupled with gas chromatography-mass spectrometry. J. Chromatogr. A
1213:19–24 (2008).
16. Gong, Y., I.-Y. Eom, D.-W. Lou, D. Hein, and J. Pawliszyn:
Technical Center. The views presented are those of the authors, Development and application of a needle trap device for time-
weighted average diffusive sampling. Anal. Chem. 80:7275–7282
and do not represent those of their respective organizations. (2008).
This document is not a standard or regulation. It creates no 17. Saito, Y., I. Ueta, K. Kotera, et al.: In-needle extraction device designed
new legal obligations and alters no existing obligations created for gas chromatographic analysis of volatile organic compounds. J.
by OSHA standards or the Occupational Safety and Health Chromatogr. A 1106:190–195 (2006).
Act. It may contain recommendations that are advisory in 18. Sweet, N.D., G.E. Burroughs, L. Ewers, and G. Talaska:
A field method for near real-time analysis of perchloroethy-
nature, informational in content, and are intended to support lene in end-exhaled breath. J. Occup. Environ. Hyg. 1:515–520
the provision of safe and healthful workplaces. This article is (2004).
not subject to US copyright law. 19. “NeedlEx Instruction Manual.” Available at http://www.airmet.com.au/
admin/Products/DownloadProductFile.aspx?FileTitle=Manual&uniquen
ame=NeedlEx manual.pdf (accessed May 4, 2012).
20. Woolfenden, E.: Sorbent-based methods for volatile and semi-volatile
REFERENCES organic compounds in air. Part 1: Sorbent-based air monitoring options.
J. Chromatogr. A 1217:2674–2684 (2010).
1. “Air Contaminants,” Code of Federal Regulations Title 29, Part
21. American Conference of Governmental Industrial Hygienists
1910.1000. 2013.
(ACGIHR ): 2012 Threshold Limit Values for Chemical Substances
2. “Benzene: Method 1005: OSHA Sampling and Analytical Meth-
and Physical Agents and Biological Exposure Limits. Cincinnati, Ohio:
ods.” Available at http://www.osha.gov/dts/sltc/methods/validated/
ACGIH Worldwide, 2012.
1005/1005.html (accessed May 1, 2012).

684 Journal of Occupational and Environmental Hygiene December 2013

You might also like