You are on page 1of 18

J Neurosurg 97:370–387, 2002

Implantation of deep brain stimulators into the subthalamic


nucleus: technical approach and magnetic resonance
imaging–verified lead locations

PHILIP A. STARR, M.D., PH.D., CHADWICK W. CHRISTINE, M.D.,


PHILIP V. THEODOSOPOULOS, M.D., NADJA LINDSEY, R.N., DEBORAH BYRD, R.N.,
ANTHONY MOSLEY, M.D., AND WILLIAM J. MARKS, JR., M.D.
Departments of Neurological Surgery, Neurology, and Nursing, University of California, San Francisco;
and Division of Neurosurgery and Departments of Neurology and Nursing, San Francisco Veterans
Affairs Medical Center, San Francisco, California

Object. Chronic deep brain stimulation (DBS) of the subthalamic nucleus (STN) is a procedure that is rapidly gain-
ing acceptance for the treatment of symptoms in patients with Parkinson disease (PD), but there are few detailed de-
scriptions of the surgical procedure itself. The authors present the technical approach used to implant 76 stimulators into
the STNs of patients with PD and the lead locations, which were verified on postoperative magnetic resonance (MR)
images.
Methods. Implantation procedures were performed with the aid of stereotactic MR imaging, microelectrode record-
ing (MER) in the region of the stereotactic target to define the motor area of the STN, and intraoperative test stimula-
tion to assess the thresholds for stimulation-induced adverse effects. All patients underwent postoperative MR imaging,
which was performed using volumetric gradient-echo and T2-weighted fast–spin echo techniques, computational refor-
matting of the MR image into standard anatomical planes, and quantitative measurements of lead location with respect
to the midcommissural point and the red nucleus. Lead locations were statistically correlated with physiological data
obtained during MER and intraoperative test stimulation.
Conclusions. The authors’ approach to implantation of DBS leads into the STN was associated with consistent lead
placement in the dorsolateral STN, a low rate of morbidity, efficient use of operating room time, and robust improvement
in motor function. The mean coordinates of the middle of the electrode array, measured on postoperative MR images, were
11.6 mm lateral, 2.9 mm posterior, and 4.7 mm inferior to the midcommissural point, and 6.5 mm lateral and 3.5 mm ante-
rior to the center of the red nucleus. Voltage thresholds for several types of stimulation-induced adverse effects were pre-
dictive of lead location. Technical nuances of the surgery are described in detail.

KEY WORDS • subthalamic nucleus • deep brain stimulation • Parkinson disease •


microelectrode recording • magnetic resonance imaging • surgical approach

EEP brain stimulation of the STN is a promising the complications associated with various approaches. Few

D technique for the relief of symptoms associated with


PD.11,29 This procedure is based on experimental da-
ta that have shown that inactivation of the motor territory of
publications contain quantitative documentation of STN
lead locations by intraoperative6,51 or postoperative2 imag-
ing, and there is no information on the correlation of intra-
the STN alleviates parkinsonism in a nonhuman primate operative observations with imaging-documented lead lo-
model of PD.9 Reports on methods of implanting stimula- cations.
tors into the STN are beginning to emerge.5,6,22,37,45,57 Nev- In this article, we describe the technical approach we
ertheless, there is no consensus regarding the best techni- used for our initial 76 implantations of stimulators into the
cal approach to this surgery, the proper lead locations, or STN, the locations of the leads determined by postoperative
MR imaging, and the clinical outcomes in a subset of pa-
Abbreviations used in this paper: A = anterior; AC = anterior tients. The goal of surgery was to place the lead through the
commissure; AP = anteroposterior; DBS = deep brain stimulation; sensorimotor (dorsolateral) area of the STN. Target local-
FSE = fast–spin echo; IC = internal capsule; IPG = implanted pulse ization was achieved through a combination of stereotactic
generator; L = lateral; M = medial; MER = microelectrode record-
ing; ML = medial lemniscus; MR = magnetic resonance; P = poste- MR imaging, MER in the region of the stereotactic target,
rior; PC = posterior commissure; PD = Parkinson disease; RN = red and intraoperative test stimulation to assess the thresholds
nucleus; SN = substantia nigra; SNr = SN pars reticulata; STN = for stimulation-induced adverse effects. The lead locations
subthalamic nucleus; UPDRS = Unified PD Rating Scale; 3D = verified by MR imaging were correlated with intraoperative
three-dimensional. observations to provide information that may be used to

370 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

predict the anatomical location of a lead based on intraoper- TABLE 1


ative physiological findings. Parameters used for preoperative and postoperative imaging
on two MR systems*
MR Protocol Siemens Symphony General Electric Horizon
Clinical Material and Methods
T2-weighted FSE
TR (msec) 2500 3000
Patient Population TEeff (msec) 110 87
matrix 252  256 384  256
We reviewed and analyzed our initial 76 consecutive cas- NEX 4 4
es of MR imaging–guided implantation of stimulators in- slice thickness (mm) 2.0 2.0
interleaved imaging
to the STN for the treatment of idiopathic PD. Approval time (min:sec) 12:04 10:48
was obtained from our Institutional Review Board. All pa- volumetric gradient echo
tients had experienced a definite improvement in motor type of acquisition MPRAGE SPGR
symptoms in response to administration of levodopa. The TR (msec) 15 36
patients’ mean scores on Part III (motor subscale) of the TE (msec) 7.0 8.0
UPDRS were 48 during the off-medication period (all anti- matrix 512  512 256  192
flip angle (˚) 30 35
parkinsonian medications withheld overnight) and 22 dur- bandwidth (kHz) 15 15.6
ing the on-medication period. The mean age of the patients NEX 1.0 0.75
at the time of surgery was 60.5 years (range 46–79 years). slice thickness (mm) 1.5 1.5
There were 77 attempted and 76 completed stimulator im- imaging time (min:sec) 9:26 11:00
plantations in 44 patients. In one case the lead was not im- * NEX = number of excitations; MPRAGE = magnetization-prepared rap-
planted because an intraoperative hemorrhage occurred. In id acquisition gradient echo; SPGR = spoiled gradient recalled acquisition;
six patients tremor was the predominant symptom, whereas TEeff = effective echo time.
in 38 patients, the symptoms were predominantly akinesia,
rigidity, gait disorder, “on–off” fluctuations, and levodo-
pa-induced dyskinesias or dystonias. Sixty-two stimulator
implantations were performed as unilateral procedures or Magnetic Resonance Imaging and Target
portions of staged bilateral procedures. In seven patients, 14 Point Determination
implants were placed during simultaneous bilateral surgery. Before we began to use MR imaging as the sole modali-
We analyzed the data for each lead separately. ty for anatomical localization, we performed a study with a
phantom head inside the stereotactic frame.53 We found that
Overview of the Surgical Procedure the center coordinates of the phantom, as determined using
Antiparkinsonian medications were withheld from pa- our MR imagers with the fiducial markers, were identical
tients for 12 hours before surgery to avoid medication- (within a single pixel size) to the known coordinates of the
induced dyskinesias during the surgical procedure. The center, based on the actual geometry of the phantom.
patient was given a local anesthetic agent, after which a ste- The MR imaging sequences that we used for preoper-
reotactic frame was applied, taking care to align the frame ative targeting are provided in Table 1. Two different MR
with the midsagittal plane of the brain and with the AC–PC imaging units were used in this series and the sequence pa-
line. Stereotactic MR imaging was performed to define the rameters differed slightly between the two magnets (Table
anatomical target. After we drilled a 15-mm burr hole, we 1). We obtained a 3D volumetric gradient-echo image set
opened the dura mater sharply. Microelectrode mapping covering the entire brain in 1.5-mm-thick slices with a zero
was then performed sequentially along multiple parallel tra- interslice distance, followed by a coronal T2-weighted FSE
jectories to define the motor territory of the STN. The DBS pulse sequence, with a 2-mm slice thickness, acquiring only
lead (intercontact distance 3 mm from center to center) was 18 slices through the targeted region. The FSE images were
inserted. Test stimulation was performed to document the acquired as two interleaved sets, each with a 2-mm gap be-
voltage thresholds for various adverse effects. Propofol was tween slices. The two sets were combined to provide a sin-
administered intravenously during the opening and closing gle FSE image set with a zero interslice distance. This inter-
portions of the surgery, but no sedative agents were given leaving technique doubled the time required for imaging,
during MER and test stimulation. The lead was cemented in but avoided the signal degradation that would otherwise oc-
place with the aid of methylmethacrylate and fixed posteri- cur in FSE imaging when there is a zero slice separation and
or to the burr hole with a titanium miniplate.15 In the last two images are acquired as a single set.50
cases that we treated, a lead-anchoring device was used. Af- To define the initial anatomical target on the basis of MR
ter the scalp wound had been closed, the headframe was re- imaging, we used a hybrid of two methods: indirect target-
moved, general anesthesia was induced, and the pulse gen- ing in which we used fixed distances from the midcommis-
erator was placed. sural point and direct targeting by visualization of the STN
In one case the lead was externalized for 2 days so that itself. The gradient-echo images were used primarily for
we could perform functional MR imaging postoperatively. indirect targeting and for trajectory planning. The FSE im-
In five cases the pulse generators were put in place 2 to 14 ages were used for direct visualization of the targeted struc-
days after lead placement; in other cases they were placed tures because of the nuclear detail they revealed.
the same day. Postoperative MR imaging was performed in Indirect Targeting. The gradient-echo sequence was re-
all patients within 24 hours after lead implantation. formatted parallel to the AC–PC line and to the midsagittal

J. Neurosurg. / Volume 97 / August, 2002 371


P. A. Starr, et al.

STN on FSE images. We did not attempt to adjust the AP


coordinate because the determination of the spatial position
of the STN in the AP domain on the basis of coronal images
would be perpendicular to the imaging plane and, thus, rel-
atively inaccurate.

Entry Point Determination


In our first 40 implantation procedures, we determined
the entry point by setting the stereotactic arc and rings for a
standardized approach angle of 60˚ from the AC–PC line in
the sagittal projection and to 5 to 10˚ lateral from the verti-
cal line in the coronal projection. In the subsequent 37 cas-
es, we used surgical planning software to examine the tra-
jectory and make slight adjustments so that we could avoid
traversing sulci or the lateral ventricle. A typical trajectory
is shown in Fig. 1D.
FIG. 1. Stereotactic targeting with reference to the commissures.
A volumetrically acquired gradient-echo sequence is depicted (pa-
rameters are shown in Table 1), computationally reformatted to Microelectrode Mapping
align with the intercommissural line and the midsagittal plane. The Single-unit extracellular action potentials were record-
cross indicates the target with respect to the midcommissural point
(lateral 12 mm, AP 3 mm, and vertical 3 mm). A: Axial
ed using one of two types of microelectrodes: custom-made
plane. B: Coronal plane. The SNr is seen slightly inferior and tungsten microelectrodes plated with gold and platinum
medial to the target. C: Sagittal plane. D: Navigational view in or platinum–iridium microelectrodes, which were obtained
the plane of the actual trajectory, showing the approach, which was commercially. Both types were coated with Parylene C, had
planned to avoid traversing sulci or the lateral ventricle. impedances of 0.1 M to 1 M at 1000 Hz, and had 15-
to 25-m tip diameters. Action potentials were amplified,
filtered (0.3–10 kHz), displayed on an oscilloscope, and
plane by using surgical planning software. For the initial 20 played on an audio monitor. Microelectrodes and the subse-
implantation procedures, the target was 12 mm lateral, 4 quent DBS lead were advanced into the brain with the aid
mm posterior, and 4 mm inferior to the midcommissural of a micropositioner (a motorized microdrive) or by using a
point. Subsequently, the AP coordinate was changed to 3 custom-built micropositioner with hydraulic microdrive.
mm posterior because microelectrode mapping in the initial The goal of MER was to identify the characteristic spon-
cases usually dictated a more anterior move. An example of taneous discharge patterns of the STN and surrounding nu-
the target determined by this formula is shown in Fig. 1. clei and the locations of movement-related cells within the
Direct Targeting. To account for potential interindivid- STN. While recording in STN, we paused every 0.3 to 0.5
ual variability in STN spatial position, we determined the mm to assess movement-related activity. Cells were consid-
AC/PC–based coordinates of the STN, as visualized on T2- ered to be movement-related if they exhibited modulation
weighted coronal FSE images. Useful landmarks in the re- of cell firing during passive movement of the contralater-
gion of the STN that are visible on MR images include the al shoulder, elbow, or wrist (arm-related cells) or hip, knee,
red nucleus, the interpeduncular cistern, and the SN. The or ankle (leg-related cells). Examination of the face was not
STN was identified as a hypointense almond-shaped focus performed systematically. “Modulation” consisted of au-
located lateral to the anterior portion of the red nucleus and dible alteration in the discharge frequency that was repro-
8 to 12 mm anterior to the PC, as shown in Fig. 2. In most ducible and synchronous with passive movement. Single
cases, the STN was visualized on at least one and up to units located in the STN and SNr with good signal/noise
three FSE coronal images. In cases in which the STN was ratios were digitized at 20 kHz for at least 30 seconds.
seen on three coronal images, the middle image was used to Spikes were discriminated off line by using a waveform–
measure the STN location and, in cases in which the struc- matching algorithm and the spontaneous discharge rate
ture was seen on two images, the more posterior image was was determined using subroutines written in commercially
used. On the selected image, we identified the center of the available software. During each microelectrode penetration,
signal hypointensity corresponding to the STN and mea- physiological data were recorded on graph paper in a scaled
sured its AC–PC coordinates. reconstruction of the microelectrode track.
Synthesis of Indirect and Direct Anatomical Targeting. In Microstimulation up to 40 A peak-to-peak (300 Hz,
cases in which the STN was visualized at a position within 200 sec pulse width, 1–2 second train duration, biphasic
1 mm of the indirect coordinates, the indirect coordinates pulses) was performed on microelectrode penetrations in
were used for the initial anatomical target. In some cases which the STN was missed entirely or in which the STN
the STN was visualized at a position more than 1 mm from was detected but the SNr was not recorded within 2 mm
that dictated by the indirect AC/PC–based coordinates de- of the base of the STN. During microstimulation, we ex-
termined by the aforementioned method. In those cases amined the patient for tonic contraction of the tongue,
the initial lateral or vertical target coordinates determined lips, face, or contralateral arm or for tonic eye deviation.
by the indirect method were adjusted by 0.5 to 1 mm to Although microstimulation-induced effects were helpful
compensate partially for the observed spatial location of the when present, their absence did not necessarily imply a

372 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

safe distance from clinically important microexcitable fiber


pathways, we used low-current stimulation to avoid damag-
ing the microelectrode tip.
The initial penetration was always directed at the ste-
reotactic target. Additional tracks were made sequentially
rather than simultaneously because the location of each
subsequent penetration was individualized based on the in-
formation already obtained in the initial tracks. Tracks were
separated by 2 mm or 3 mm. Typically, the second penetra-
tion was performed in the same parasagittal plane as the ini-
tial one, and a third was performed 2 to 3 mm lateral or me-
dial to the initial plane of penetration. In cases in which
earlier penetrations identified motor-responsive cells within
the STN, relatively fewer tracks were performed; if early
tracks did not identify motor responsive cells or missed the
STN entirely (which was rare), relatively more microelec-
trode penetrations were performed.

Deep Brain Stimulation Lead Placement


Using the microelectrode map to determine DBS lead
placement, the goal was to place the lead into the motor
territory of the STN, but to avoid a location closer than 2
mm from the lateral, anterior, or posterior borders. If the
initial penetration showed a segment of STN cellular activ-
ity greater than 4 mm in length and revealed movement-re-
lated activity, the lead was typically placed in or very near
to the initial microelectrode penetration. Factors dictating
significant deviation ( 1 mm) from the initial penetration
including the following: 1) greater concentration of move-
ment-related cells or a longer trajectory through the STN
during penetrations other than the initial one; and 2) prox-
imity to a border of the STN. For example, if additional
tracks located 2 mm anterior and 2 mm lateral to the initial
one showed no STN activity, the lead was placed more pos-
teriorly and more medially than the initial penetration, even
if the initial penetration represented a long segment with
clear movement-related cells.
FIG. 2. Stereotactic targeting by reference to the STN. A coro-
Test Stimulation nally oriented T2-weighted FSE sequence is displayed (parameters
After a baseline neurological examination had been con- are shown in Table 1) at points 10 mm (A), 12 mm (B), and 14 mm
ducted by the surgeon or a neurologist, the lead was placed (C) anterior to the PC. The arrows indicate the STN. The image in
so that the middle two contacts (in the first 60 implanta- panel B was used for actual target selection.
tions) or the bottom two contacts (in the last 16 implanta-
tions) spanned the STN. Intraoperative test stimulation was axial rather than coronal plane. Postoperative MR images
performed between the two contacts spanning the STN by were transferred to a computer for analysis, which was per-
using a 60-sec pulse width, 185 Hz, and 2 to 10 V. The fol- formed using appropriate software (Framelink). All image
lowing were monitored as the voltage was increased and sets were reformatted to be parallel to the AC–PC line and
decreased: the development of dysarthria; tonic contraction orthogonal to the midsagittal plane. This ensured that lead
of face, arm, or leg; dyskinesia; diplopia or blurred vision; location measurements in different patients were made in
paresthesia; or any other subjective, but reproducible sensa- standardized anatomical planes. Lead locations were mea-
tion that the patient could report. If resting tremor was pres- sured in two ways: 1) as a continuous variable in three
ent, we monitored the patient for tremor suppression, but dimensions with respect to the midpoint of the intercom-
we did not systematically monitor for improvements in oth- missural line; and 2) as a continuous variable in two di-
er parkinsonian signs. The lead was moved to a new trajec- mensions with respect to the center of the red nucleus, as
tory only if the threshold for dysarthria, muscle contraction, measured on the axial plane 4 mm inferior to the intercom-
visual blurring, or diplopia was 2 V or less. missural line.
Lead Coordinates With Respect to the Midcommissural
Measurement of Lead Location
Point. On the gradient-echo images, the AC and PC were
The pulse sequences selected for postoperative MR im- identified by and registered on the Framelink software.
aging were the same as those for preoperative MR imaging Next, the distal tip of the stimulator was identified. It ap-
(Table 1), except that the FSE sequence was acquired in the peared as a relatively discrete round signal void, approxi-

J. Neurosurg. / Volume 97 / August, 2002 373


P. A. Starr, et al.

poned up to 1 month in patients in whom there was extreme


early sensitivity to stimulation-induced dyskinesias or in
whom there was transient postoperative disorientation. Pro-
gramming was usually performed while the patient was
in the on-medication state. In most patients programming
was conducted in a monopolar configuration by using a
single contact that had been placed in the dorsal STN, as
confirmed by intraoperative observations. In cases in which
mild stimulation induced dysarthria or facial contraction oc-
curred in the therapeutic voltage ranges, bipolar stimulation
was used, with the two contiguous contacts spanning the
STN. The more dorsal contact of the pair was programmed
to be the negative electrode. The frequency and pulse width
were initially set at 185 Hz and 60 sec. Voltage was titrat-
ed for clinical benefit over the initial months, usually to 2 to
3 V. We did not systematically assess the effects of stimula-
tion through contacts outside the STN. However, in cases
in which little or no benefit was reported by the patient after
a few weeks in one configuration, more ventral or more
dorsal contacts were used in an attempt to optimize clinical
benefit.
Neurological assessment was performed preoperatively
FIG. 3. Representative MERs in various nuclei along a trajectory by using the UPDRS and timed motor tests. The UPDRS
(dotted line) through the STN. The drawing is a parasagittal section
adapted from the Schaltenbrand and Wahren human brain atlas. motor subscale (Part III) was performed on separate occa-
Each trace is 1 second long. Upper Left: Dorsal thalamic burst- sions when the patient was in the off- and on-medication
ing cell. Right: Three recordings made in the STN. Lower Left: states. The off-medication state was produced by withhold-
Recording made in the SNr. ing antiparkinsonian medication for 12 hours. In a subset of
patients postoperative UPDRS testing was performed dur-
ing periods of off- and on-medication and off- and on-stim-
mately 3 mm in diameter, that was larger than the actual di- ulation.
ameter of the lead. The center of the round signal void was
considered to represent the true lead position. Precise iden- Data Analysis and Statistical Methods
tification of the lead tip was facilitated by viewing the lead All data related to the surgical procedure, intraoperative
in three orthogonal planes on the Framelink software. The physiological findings, lead locations, and programming
AC/PC–based coordinates of the distal tip were computed. were entered into a customized database. The hypothesis
The entry point of the lead into the brain was also identified was that lead or microelectrode track locations that were as-
(this was also a relatively discrete signal void) and its coor- sociated with a particular physiological observation would
dinates with respect to the AC–PC line were calculated. be located at different coordinates than leads or microelec-
The AC/PC–based coordinates of the tip and entry point trode tracks that were not associated with that physiologi-
were entered into a database. The coordinates of the active cal observation. Hypothesis testing was performed using
contact, with respect to the midpoint of the AC–PC line, multivariate logistic, with the aid of commercially available
were calculated from the choice of contact(s) made at the software.
most recent programming session, from the known contact
geometry (1.5 mm length, spaced 3 mm center–center) and Sources of Supplies and Equipment
from the coordinates of the tip and entry point. The angula-
tion of the electrode array with respect to the vertical line to The Leksell series G stereotactic frame and the phantom
the AC–PC line, in both sagittal and coronal projections, head were purchased from Elekta (Norcross, GA). Med-
was calculated from the coordinates of the tip and entry. tronic (Minneapolis, MN) manufactured the model 3397
The formulas for these calculations are provided in the Ap- DBS lead and the Itrell II and Soleta pulse generators. Im-
pendix. age Guided Neurologics (Melbourne, FL) manufactured
the Navigus lead-anchoring device. Access software (Mi-
Lead Coordinates by Reference to Surrounding Nuclei. To crosoft Corp., Redmond, WA) served as the database for
account for anatomical variability in spatial positions of tar- the AC/PC–based coordinates. Two 1.5-tesla MR imag-
geted nuclei with respect to the AC and PC, we also mea- ing units were used in this study: Symphony, provided by
sured lead location with reference to the red nucleus, the Siemens (Erlangen, Germany), and Horizon, provided by
structure nearest to the STN that can be reliably located on General Electric Medical Systems (Milwaukee, WI). The
MR imaging5,51 in the axial plane 4 mm inferior to the AC– Stealth workstation and Framelink surgical planning soft-
PC line. ware were obtained from Medtronic–Sofamor Danek. Two
types of microelectrodes were used for recordings of extra-
Programming of Stimulation Parameters and
cellular action potentials: gold and platinum–plated tung-
Clinical Outcomes
sten microelectrodes, which were custom-made for us at
Programming of stimulation parameters was usually per- Toronto Western Hospital,22 and platinum–iridium micro-
formed within 1 to 2 days after surgery, but it was post- electrodes, which were obtained from Microprobe, Inc.

374 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

(Gaithersburg, MD). The action potentials were visualized was present during MER, cells with discharges grouped at
and given sound by using equipment obtained from Axon tremor frequency were recorded.
Instruments (model GS 3000; Foster City, CA). Axon In-
struments also produced the Clinical Micropositioner and Movement-Related Neuronal Activity
David Kopf Instruments (Tujunga, CA) manufactured the Movement-related activity was found on 131 of the MER
hydraulic microdrive. tracks (53%). Movement-related activity was concentrated
The Multi-Spike Detector, produced by Alpha Omega in the dorsal 3 mm of the nucleus, 10 to 14 mm from the
Engineering (Nazareth, Israel) was used to discriminate midline. A total of 348 movement-related cells (mean 4.6
spikes during MER, and Labview was used to make sub- per procedure) were encountered. Of these cells, 35% were
routines. Multivariate logistic regression analysis was per- leg related, 59% were arm related, and 6% were mixed. Re-
formed using statistical software from SPSS (Chicago, IL). sponses to proximal joint motions were more frequent than
those to wrist, finger, or ankle motion. In eight procedures
Results no movement-related cells were recorded. This was usually
due to a pulsation artifact that interfered with the identifica-
tion of movement-related activity. In two procedures, how-
Targeting Based on MR Imaging ever, the lack of movement-related activity was probably
The STN was directly visualized on T2-weighted coronal due to the fact that all microelectrode penetrations (inter-
FSE images for 70 (92%) of 76 procedures. In those cases preted retrospectively after viewing the postoperative MR
in which the STN was not clearly visualized, the indirect image) were located too anteromedial in nonmotor regions
target method (12 mm lateral, 3 mm posterior, and 4 mm in- of the STN.
ferior to the midcommissural point) was used as the sole Using the lead location from postoperative axial FSE MR
method of anatomical targeting. images as a marker, we retrospectively extrapolated the AP
For the 70 procedures in which the STN was visualized, and lateral coordinates of all 62 microelectrode tracks along
the mean and range of the AC/PC–based coordinates of the which at least three movement-related cells were encoun-
center of the STN were as follows: lateral 12.9 mm (range tered. For example, if a particular MER track was made 1
11–15 mm), AP 3.5 mm (range 1 to 7 mm), and ver- mm medial and 1 mm posterior to the location of the DBS
tical 5.3 mm (range 2 to 8.5 mm). lead and if the DBS lead was found to have a lateral coordi-
In 24 (34%) of 70 procedures, the STN was visualized at nate of 12.5 mm with respect to midline and an AP coordi-
a position that was greater than 1 mm more lateral than the nate of 3 mm with respect to the center of the red nucleus
AC/PC–based coordinate of 12 mm from the midline, re- (defined at a vertical position of 4 mm), the MER track
sulting in a slight increase in the lateral coordinate. In 31 was assigned a lateral coordinate of 11.5 mm and an AP co-
(40%) of 70 procedures the STN was seen greater than 1 ordinate of 2 mm anterior to the red nucleus. Tracks asso-
mm more inferior than the AC/PC–based coordinate of 4 ciated only with leg-related cells had a mean lateral coor-
mm from the midcommissural point, resulting in a slight dinate of 11.27 mm, whereas those associated only with
inferior adjustment in the vertical coordinate. After adjust- arm-related activity had a mean lateral coordinate of 12.09
ments were made on the basis of direct visualization of the mm. Leg-related activity on an MER track was more often
STN, the means and ranges of the initial anatomical tar- seen in medial tracks, whereas tracks with cells exclusively
get coordinates were the following: lateral 12.3 mm (range responding to arm movements were seen more laterally.
11.5–14 mm), AP 3.3 mm (range 5.5 to 2 mm), and This, however, did not reach statistical significance (p 
vertical 4.9 mm (range 2.5 to 7 mm). 0.07). There was no apparent difference between mean AP
locations for arm- and leg-related activities. We noted a ten-
Spontaneous Neuronal Activity dency for leg-related activity to occur in a more concentrat-
ed area than arm-related activity. The leg-related area was
Recordings encountered during a typical microelectrode not only relatively medial, but along the medial parasagittal
penetration are shown in Fig. 3. We usually recorded cells planes, it was concentrated centrally, away from the anteri-
in the caudate nucleus and the anterior thalamus before or and posterior poles, compared with arm-related activity.
moving into the STN. There was little difficulty in recog-
nizing the STN, based on its characteristic irregular dis- Microstimulation Study
charge at 20 to 50 Hz and dense cellularity, resulting in a
noisy baseline and many multiunit recordings. After exiting Microstimulation was typically performed only on tra-
the STN, we explored another 2 mm in an attempt to iden- jectories along which the SNr had not been encountered
tify the SNr by its characteristic rapid, regular discharge. within 2 mm below the base of the STN, or where the STN
The total number of MER tracks made during the 76 pro- and SNr were both missed. Stimulation-induced effects
cedures was 245, for a mean of 3.2 per side (range 1–6). at less than 40 A peak-to-peak were observed in six
The STN was identified physiologically during the first (16%) of the 37 procedures in which microstimulation was
penetration in 73 (96%) of 76 procedures, and in 58 proce- performed (biphasic square wave, pulse width 200 sec,
dures (76%) the first penetration traversed 3 or more mm of frequency 300 Hz). The following responses were noted:
the STN. On average, a 3.8-mm segment of STN was re- tongue contraction (four procedures), ipsilateral eye abduc-
corded on the initial trajectory (range 0–5.9 mm). The SNr tion (one procedure), and tremor arrest (one procedure).
was recorded below the STN on 129 of MER tracks (53%).
Use of MER and Microstimulation Data to Modify the
The mean ( standard deviation) spontaneous firing
Anatomical Target
rates were 34 14 Hz for 102 STN cells and 86 16 Hz
for six SNr cells. In five patients in whom visible tremor Figure 4 illustrates the most important ways in which

J. Neurosurg. / Volume 97 / August, 2002 375


P. A. Starr, et al.

FIG. 4. Modification of the initial anatomical target by MER in four illustrative cases. The column containing drawings
shows data derived from MER. The microelectrode tracks (dotted lines) are superimposed on a drawing of the STN de-
rived from the closest available parasagittal plane of the Schaltenbrand and Wahren atlas. Tracks are labeled T1 through
T4 according to the order in which they were made; the numbers in parentheses represent the AP and lateral positions in
millimeters, respectively, with regard to the initial anatomical target (positive direction defined as anterior and lateral). Seg-
ments are shaded according to the corresponding cellular activity that was recorded: gray represents the STN, black the
SNr, and unshaded the red nucleus. Movement-related cells are shown as triangles (leg related) or circles (arm related).
Microstimulation thresholds (stim) are indicated where a positive result was obtained and labeled according to the current
threshold in microamperes for the observed effect as well as the body part whose motion was observed. The column con-
taining grids shows a schematic in the axial plane of the locations of each microelectrode track (labeled T1 through T4)
and of the final lead placement (L). The grid lines are drawn every 2 mm. Ant, Post, Med, and Lat indicate the anterior,
posterior, medial, and lateral directions. The last column contains postoperative axial FSE MR images (parameters shown
in Table 1) obtained 4 mm inferior to the commissures. The lead corresponding to the case described is indicated by an ar-
row. Note that in some cases, there is a contralateral lead that had been placed during a previous operation.

MER was used to modify the initial anatomical target. Four regular discharge pattern, and occurrence in a more ven-
illustrative cases are shown, including the MER map in the tral location along the MER track than would be expected
region of the STN and immediate vicinity, the modification for STN.
of the DBS lead position in the AP and lateral directions In Case 2, only a short segment (1–2 mm) of STN was
that resulted from the MER map, and the postoperative MR encountered on the first MER track, indicating a peripheral
image. In all examples except for Case 3 the lead appeared localization in the nucleus. The clues that indicated that the
to be correctly located (in the dorsolateral STN) on the post- first MER track was too lateral were provided by micro-
operative MR image. The arrow on the MR image indicates stimulation-induced corticobulbar activation just below the
the lead that corresponds to the MER map. STN, and by the absence of the SNr below the STN (be-
In Case 1, the initial MER track missed the target com- cause the SNr does not extend as laterally as the STN).
pletely, passing posteromedial to the STN to enter the red Cases 3 shows an “outlier” lead location from early in
nucleus. The red nucleus can potentially be confused with our series. Both the initial MER penetration and the final
the STN based on their similar firing rates and the pres- lead location were anteromedial to the desired target, pass-
ence of movement-related activity,40,43 but it may be dis- ing through nonmotor regions of the STN. In retrospect this
tinguished from the STN by its lower cell density, more was evident from the MER map, based on the absence of

376 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

movement-related activity in tracks nearest the final lead TABLE 2


location. Subsequent to this case, we refined our mapping Intraoperative stimulation-induced adverse effects
strategy. If the first penetration demonstrated STN sponta- observed in 70 cases*
neous activity but no movement-related activity, we contin-
ued mapping in an area 2 to 3 mm more lateral and 2 to 3 No. of Procedures (%)
mm more posterior to the initial target. Additional evidence Threshold Lowest
of an excessively anteromedial location is sometimes af- Threshold
forded by microstimulation of the mesencephalic tract of Effect 5 V 10 V Effect
the third cranial nerve (inducing eyelid opening and adduc- dysarthria or facial contraction 10 (14) 46 (66) 31 (44)
tion of the ipsilateral eye), deep to the anteromedial STN. arm, leg, or truncal muscle contraction 2 (3) 10 (14) 6 (9)
Case 4 illustrates implantation of the DBS lead after a paresthesia 15 (21) 29 (41) 26 (37)
single MER track has been performed. We considered lead diplopia 2 (3) 9 (13) 1 (1.4)
implantation after only one MER track had been obtained blurred vision 1 (1.4) 2 (3) 2 (3)
conjugate eye deviation 0 1 (1.4) 1 (1.4)
when the initial track was long ( 4-mm segment of the vague sense of head discomfort 3 (4) 8 (11) 1 (1.4)
STN) and showed robust movement-related activity with at dyskinesia 5 (7) 8 (11) 1 (1.4)
least some leg-related activity. The finding of leg-related vertigo 0 0 1 (1.4)
activity is suggestive of a localization that is both relatively
* Tests were performed in bipolar mode between the two contacts span-
medial in the motor territory41,47 and away from the rostral ning the STN at a pulse width of 60 sec, a frequency of 185 Hz, and a
and caudal poles.47 Thus, the identification of the leg terri- voltage of 0 to 10.
tory is of high localizing value. We usually place the lead 1
mm lateral to a track where extensive leg-related activity
has been recorded. less than 2 V and to a more posterolateral location in one
case in which diplopia was induced at less than 2 V.
Intraoperative Stimulation-Induced Adverse Effects While testing only contacts in the STN, no patient ex-
Adverse effects elicited by intraoperative test stimulation hibited stimulation-induced mood changes. Although we
are shown in Table 2 (bipolar mode using the two contacts did not systematically test deep contacts, we did observe
spanning the STN with the settings at 185 Hz, 60-sec stimulation-induced mood depression intraoperatively in
pulse width,
10 V). In six procedures, complete eval- one patient, with Contacts 0 (negative pole) and Contact 3
uation of stimulation-induced effects was not possible be- (positive pole), 60 sec, 185 Hz, 4 V. In this case, Contact
cause there was a lack of cooperation on the part of the 0 was in the dorsal SNr, based both on intraoperative phys-
patients; thus, the data in Table 2 represent 70 rather than iological findings and on postoperative MR images.
76 implantation procedures. Stimulation-induced adverse
effects were always rapidly reversible with cessation of Intraoperative Stimulation-Induced Therapeutic Effects
stimulation. Stimulation-induced dysarthria or contralater- During the 14 procedures in which there was resting
al facial contraction occurred at less than 10 V in 46 cases tremor during test stimulation, there was a stimulation-in-
(66%) and contralateral paresthesias in 29 cases (41%). duced reduction in tremor in 10 (71%) at 0 to 5 V and in 13
When present, the body part in which paresthesias were (93%) at 0 to 10 V (pulse width 60 sec, frequency 185
perceived at the lowest voltage threshold was variable: up- Hz). In contrast to what is observed during thalamic stimu-
per extremity (10 procedures), lower extremity (nine proce- lation,7 however, tremor was rarely totally eliminated with
dures), hemibody (six procedures), face or head (three pro- acute stimulation and, frequently, the reduction was mod-
cedures), and torso alone (one procedure). All other effects, est. The degree of intraoperative tremor control did not
at the parameters described earlier, were infrequent (Table seem to predict long-term control, because even those who
2). Diplopia occurred in patients during nine procedures displayed only modest or absent relief from tremor intraop-
(12%) and was usually subjective, without an obvious mon- eratively experienced more substantial tremor relief with
ocular deviation. Two patients (3% of procedures) report- chronic stimulation. We did not systematically search for
ed stimulation-induced visual blurring that was not associ- improvements in parkinsonian symptoms other than tremor,
ated with diplopia. Only eight patients (11% of procedures) although we noted that contralateral rigidity usually im-
experienced acute stimulation-induced dyskinesias during proved in response to microelectrode mapping or DBS lead
intraoperative testing. Eight patients (11% of procedures) insertion.
reported a vague, but unpleasant sensation in the head, typ-
ically described as head pressure or a sense of the head ex- Time in the Operating Room
panding. In five patients no objective or subjective stimu-
lation-induced adverse effects were noted at the standard The mean operating room time per lead implantation
testing parameters given earlier; however, in all of these (from incision to closure of the scalp, not including pulse
cases, dysarthria, paresthesias, or diplopia was inducible generator insertion) shortened from 4.5 hours/lead in the
by higher pulse widths (up to 250 sec) and voltages of 5 first ten procedures to 2.5 hours in the last 10 procedures.
to 10 V.
During three procedures, intraoperative test stimulation Magnetic Resonance Imaging–Verified Lead Locations
thresholds for adverse effects were considered unacceptable Examples of gradient-echo MR images demonstrating
( 2 V), resulting in withdrawal and replacement of the the lead tip and entry are shown in Fig. 5. The mean loca-
DBS lead. The lead was moved to a more medial location tion of the contact arrays with respect to the midcommis-
in two procedures in which bulbar activation occurred at sural point are provided in Table 3, and their distribution

J. Neurosurg. / Volume 97 / August, 2002 377


P. A. Starr, et al.

FIG. 5. Postoperative gradient-echo MR images demonstrating the locations of the lead tip and entry point. The images
have been computationally reformatted to align with the intercommissural line and midsagittal plane. A–C: Lead tip
(arrow) in axial, coronal, and sagittal planes, respectively. Note that the tip was placed in the ventral STN; the active con-
tact in this case was Contact 1, centered 3.75 mm rostral to the lead tip. D–F: A point on the lead near its entry into the
brain (arrow) in axial, coronal, and sagittal planes, respectively.

is displayed graphically in Fig. 6. The mean AC/PC– a systematic tendency to target too posteriorly with respect
based coordinates of the midpoint of the quadripolar array to the physiologically mapped STN. When the AP coor-
were the following: lateral 11.6 mm, AP 2.9 mm, and ver- dinate was adjusted to 3 mm after the first 20 proce-
tical 4.7 mm. Following programming of the devices, at dures, the remaining 56 procedures had a mean AP adjust-
a mean follow-up duration of 9 months, the mean coordi- ment of 0.03 mm, thus validating the choice of 3 mm
nates of the active contacts were the following: lateral 11.8 over 4 mm as a starting coordinate.
mm, AP 2.4 mm, vertical 3.8 mm. The mean angula- The mean absolute change in coordinates (Table 4, right
tion of the leads in the sagittal projection, from the vertical column) provides a measure of accuracy, that is, the differ-
line to the AC–PC line, was 29.2˚ (range 14.1˚–41.2˚). The ence between the positions of the initial and final intraoper-
mean angulation of the leads in the coronal projection, from ative targets, without consideration of the direction of the
the midsagittal plane, was 10.2˚ (range 0˚–24.3˚). difference. The mean absolute change was near 1 mm for
Fast–spin echo imaging in the axial plane demonstrated the vertical and AP directions, but only 0.5 mm for the lat-
the position of the lead with respect to surrounding nuclei eral direction. In 19 procedures (25%), however, a change
(several sample FSE MR images are shown in Fig. 4). The of 2 mm or greater was made in the lateral or AP direction.
mean distance of the contact array with respect to the cen- Those cases did not appear to be associated with any partic-
ter of the red nucleus, measured on FSE images at an axial ular age group, brain size, or degree of brain atrophy; thus,
level of 4 mm below the intercommissural line were 3.5 microelectrode mapping substantially altered the position
mm anterior (range 1.2 to 8.1) and 6.5 mm lateral (range of the DBS lead in a significant minority of cases.
3.5–9.3). On axial imaging, the “average” lead was lateral The modification of the initial anatomical target accord-
to the anterior border of the red nucleus. ing to intraoperative physiological findings provides one
method of assessing the accuracy of anatomical targeting.
Differences Among the Anatomical Target, the An additional, less subjective method is to compare the pre-
Physiological Target, and the MR Imaging– dicted AC/PC–based coordinates of the lead center in ste-
Verified Lead Location reotactic space—after adjustment for intraoperative phys-
Table 4 summarizes the degree of modification of the ini- iological findings—with the actual coordinates of the lead
tial anatomical target that occurred as a result of intraoper- center as measured on the postoperative MR image. This
ative physiological mapping. The data were analyzed both comparison is provided in Table 5. As in Table 4, the mid-
for bias (middle column) and for accuracy (right column). dle column represents bias (vector difference), whereas the
The middle column of the table provides the mean vec- right column represents accuracy (absolute difference). The
tor change in each coordinate subsequent to physiological discrepancy between the predicted coordinates of the lead
mapping. This provides a measure of the degree of bias in and the actual coordinates is somewhat greater than the dis-
the initial MR imaging–based targeting. The fact that the crepancy between anatomical and physiological targets de-
means were near zero indicates that, on average, there was tailed in Table 4.
no systematic tendency to move the lead in a particular di-
Lead Locations Related to Intraoperative Test Stimulation
rection based on physiological mapping. Of note, in the first
20 procedures, the AP coordinate used in the indirect target- Figure 7 displays aggregate lead locations for all proce-
ing formula was 4 mm. In those procedures, the mean dures in the axial plane, indicating the lowest-threshold
change in the AP direction was in fact 0.5 mm, indicating adverse effect observed for each lead during intraopera-

378 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

TABLE 3
Lead locations with respect to the midcommissural point
determined using gradient-echo MR imaging*
Coordinate Middle of Contact Array (mm) Active Contact (mm)

AP 2.9 0.16 (7.63 to 0.74) 2.4 0.17 (7.6 to 0.74)


lat 11.6 0.16 (8.39–14.44) 11.8 0.17 (8.4 –14.7)
vertical 4.7 0.16 (8.76 to 1.92) 3.8 0.19 (7.5 to 0.1)
* Values are expressed as the means standard errors of the mean, with
ranges in parentheses.

tive test stimulation. The nuclear outlines are adapted from


the Schaltenbrand and Wahren atlas42 at a vertical position
of 4 mm with respect to the AC. This axial map was
selected because it is closest to the vertical position of the
center of the active contacts. Based on the 3D lead locations
in Table 3 and Fig. 6, the active contact array in most cases
passes through the plane 4 mm inferior to the AC–PC line.
Retrospective analysis by logistic regression showed that FIG. 6. Graphs of lead locations with respect to the AC–PC mid-
the occurrence of visual, sensory, and presumed corticobul- point. Upper and Lower Left: Each symbol represents the mid-
bar (dysarthria or facial contraction) effects did correlate point of the quadripolar contact array. Upper and Lower Right:
with the MR imaging–determined lead locations. In the Each symbol represents the midpoint of the active contact or con-
26 procedures in which stimulation-induced paresthesias tacts after programming for optimal benefit.
occurred as the lowest-threshold adverse effect, the corre-
sponding lead locations were more medial, posterior, and
superior than others in which paresthesias were not the low- nucleus, and 4.8 mm anterior to the center of the red nucle-
est-threshold adverse effect. The difference in the lateral di- us. The left lead alone resulted in an improvement in the
mension was statistically significant (p = 0.028). Leads in off-medication UPDRS Part III score of 39% for unilater-
patients in whom paresthesias were the lowest-threshold al (left) DBS. The inclusion of the right-sided, suboptimal-
effect were located at a mean lateral position of 11.26 mm ly located lead essentially added no incremental benefit,
from the midline compared with 11.91 mm for leads when with an improvement in the off-medication UPDRS Part III
paresthesias were not the lowest-threshold effect. The 31 score of 40% for bilateral DBS.
leads associated with corticobulbar activation as the lowest- The second outlier lead location was excessively lateral:
threshold adverse effect were significantly (p = 0.011) more 14 mm from the midline and 9 mm lateral to the center of
lateral (12.12 mm) than those leads associated with other the red nucleus according to structures on the postoperative
adverse effects (11.23 mm). In the two instances in which MR image. Although not the most lateral lead in this series,
visual blurring occurred at lower thresholds than any other it was placed in a patient whose STN was more medial than
stimulation-induced side effects, the leads were significant- normal, centered 11 mm from the midline. Its AP location
ly more medial at a mean of 9.06 mm from the midline was deemed “average,” at 3.3 mm anterior to the center of
(p = 0.047). the red nucleus. The contact within the dorsal STN could
not be programmed over 1 V without producing facial con-
Clinical Outcomes and Complications traction. The lead was replaced with a new one, during a
Clinical outcomes for a subset of patients are shown in second stereotactic surgery (under fluoroscopic guidance
Table 6. The follow-up period for unilateral DBS of the without MER). Although the final lateral coordinate of 12.5
STN was 3 months and that for bilateral DBS of the STN mm from midline (and 7.5 mm lateral to the center of the
was 12 months. In the off-medication state, the mean im- red nucleus) was only 1.5 mm more medial than the first, it
provement in the score on the motor subscale (Part III) of allowed programming of the contact within the dorsal STN
the UPDRS was 23.3% for unilateral DBS and 45% for bi- at more typical parameters (2.5 V monopolar, 60 sec, 185
lateral DBS. Hz) without adverse effects.
We do not yet have sufficient outcomes data for a statis- Complications are listed in Table 7. There were two hem-
tical correlation of lead location with clinical benefit. There orrhages visible on MR images in this series: one asymp-
were, however, two outlier leads in this series that were in- tomatic 0.7-ml hematoma in the caudate nucleus next to
effective or unusable, and they are suggestive of the bound- the DBS lead and one symptomatic 0.6-ml hemorrhage in
aries of clinically effective lead location. The first was the the STN, genu, and posterior limb of the internal capsule.
most anteromedially placed lead in this study, shown in Fig. The latter occurred during withdrawal of the microelectrode
4, Case 3 (right lead) and in Fig. 6 (most anteromedial sym- after the initial penetration, producing facial and arm weak-
bol). On the postoperative MR image, at an axial level of 4 ness. Lead implantation was not performed, and motor
mm below the midcommissural plane, this right-sided lead strength returned to normal at 1 month. No hemiballism
appeared to be 9.3 mm off the midline, 4.3 mm lateral to the or increased dyskinesias were observed. There was one
center of the red nucleus, and 9.3 mm anterior to the center wound infection that required intravenously administered
of the red nucleus, whereas the contralateral (left) implant antibiotic medication but no infections that required hard-
was 13.4 mm from the midline, 6.8 mm lateral to the red ware removal.

J. Neurosurg. / Volume 97 / August, 2002 379


P. A. Starr, et al.

TABLE 4 TABLE 5
Intraoperative target adjustment after physiological mapping* Differences between final intraoperative target coordinates and
MR imaging–verified lead coordinates*
Vector Change in Target Absolute Change in Target
Coordinate Coordinates (mm) Coordinates (mm) Vector Difference Absolute Difference
Coordinate (mm) (mm)
AP 0.07 (3 to 2.5) 0.96 (0–3)
lat 0.26 (1.5 to 2) 0.45 (0–2) AP 0.76 (3.6 to 8.7) 1.62 (0–5.1)
vertical 0.39 (5 to 2) 1.07 (0–5) lat 0.75 (4.7 to 2.5) 1.40 (0–4.7)
vertical 0.69 (6.4 to 8.8) 1.72 (0–8.8)
* Values are expressed as the means, with ranges in parentheses. Positive
direction is anterior, lateral, and superior. * Values are expressed as the means, with ranges in parentheses. Final
intraoperative target coordinates were determined after the physiological
mapping. Positive direction is anterior, lateral, and inferior.
There were 10 procedures (13.1%) following which
hardware-related problems required a return to the operat-
ing room (mean follow-up time 9 months). Problems in- methods for the STN. Our targeting method, like that re-
cluded stereotactic repositioning of a poorly positioned lead ported by others,6,51,57 was a hybrid of direct and indirect ste-
(one procedure, discussed previously), stereotactic replace- reotactic targeting techniques. We used visualization of the
ment of a broken lead associated with a connector in the STN on T2-weighted FSE images to measure directly the
cervical position (one procedure), elective repositioning of spatial position of the STN with respect to the AC and PC,
cervical connectors to the cranial position (two procedures), using this to make small adjustments in the indirect coor-
repair of component disconnection (two procedures), ero- dinates (AP 3 mm, lateral 12 mm, vertical  4 mm). The
sion of the lead extension over the clavicle (one procedure), relatively high degree of accuracy of our targeting method
suspected IPG malfunction (one procedure), lead exten- was demonstrated by microelectrode mapping, document-
sion malfunction (one procedure), and hematoma around ing that the anatomical target was within the boundaries of
the IPG (one procedure). the physiologically defined STN in almost all cases.
Behavioral changes lasting longer than 24 hours were ini- Others have reported that the STN and its neighboring
tially common in patients undergoing simultaneous bilater- nuclei (the SNr and the red nucleus) can be visualized on
al lead placement (the first five of our seven simultaneous T2-weighted FSE images.5,6,44,51 This is probably due to the
bilateral procedures). These changes consisted of disinhi- high iron content of these nuclei, which decreases their T2
bition, confusion, or flattened affect, and resolved by 1 to signal.14 Bejjani, et al.,5 reported on a series of 24 implan-
2 weeks. Once our surgical protocol was modified so that tation procedures in which visualization of the STN was op-
we would place the lowest contact in the ventral STN rather timized, allowing anatomical targeting that was based ex-
than into the SNr, the subsequent patients who underwent clusively on direct visualization of the nucleus, independent
simultaneous bilateral implantation procedures displayed of the classic AC/PC–based indirect methods. To achieve
no postoperative behavioral changes. Transient postoper- optimal visualization of the STN, they reduced the field
ative behavioral changes were infrequent following uni- of view to the basal ganglia. This excluded visualization of
lateral placement (four of 62 implantation procedures). One the stereotactic fiducial markers and required computation-
patient attempted suicide 1 month following unilateral al fusion of images with another MR imaging data set with
placement of DBS leads into the STN. He had a history of a larger field of view. Our imaging protocol, in contrast,
depression prior to surgery, but no record of a previous sui- maintains a large field of view for the FSE images, which
cide attempt. This patient did not believe that his mood was includes the fiducial markers. This sacrifices some tissue
altered by activation or inactivation of the stimulator. contrast, but avoids the need to fuse image sets computa-
tionally.
Discussion Applying the protocols described here to a 1.5-tesla MR
imaging system, it is rare that all borders of the STN are vi-
Following the pioneering clinical work of Benabid and sualized with perfect precision, whereas the commissures
colleagues,29 stimulation of the STN is a procedure that is are always clearly identified. We have therefore been hes-
rapidly gaining acceptance for the symptomatic treatment itant to rely solely on direct targeting of the STN, with
of PD. Despite its growing popularity there are relatively no reference to AC–PC coordinates. Slice thickness of the
few detailed descriptions of the technical aspects of this images with optimal nuclear visualization (the FSE imag-
surgery.5,6,23,37,45,57 The goal of this report was to present and es) is greater than that for the volumetric set used for AC/
to evaluate critically our technical approach to stimulator PC–based targeting, thus introducing greater error related to
placement into the STN in a way that is useful to other volume-averaging effects. Use of a hybrid of direct and in-
groups performing this procedure. Central to this goal is direct methods allows the surgeon to substitute indirect
the MR imaging–based analysis of DBS lead locations and methods in cases or in certain dimensions of a case (for
their correlation with intraoperative physiological findings. example, the AP dimension in our protocol), in which the
These correlations should assist other practitioners in in- coordinate of the target depicted on MR images is difficult
terpreting the significance of intraoperative findings for the to establish by direct visualization. Zonenshayn and asso-
prediction of lead locations. ciates,57 in an analysis of targeting accuracy in 30 implan-
tation procedures, showed that better anatomical target ac-
Magnetic Resonance Imaging–Based
curacy is achieved using a hybrid of indirect and direct
Anatomical Localization
techniques rather than direct techniques alone. Although we
Several groups have reported their anatomical targeting did not systematically analyze the predicted accuracy that

380 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

TABLE 6
Effect of unilateral and bilateral DBS on UPDRS Part III scores
Motor Score*

Unilat DBS Bilat DBS


Condition (20 procedures) (10 procedures)

preop
off-medication state 48.1 51.3
on-medication state 22.1 23.9
postop w/ DBS turned off*
off-medication state 44.3 50.6
on-medication state 25.1 30.1
postop w/ DBS turned on*
off-medication state 33.4 28.4
on-medication state 20.1 19.2
% improvement†
off-medication state 23.3 45.0
on-medication state 14.6 35.8
* Test was administered at a single time point at 3 months (patients with
unilateral DBS) or 1 year (patients with bilateral DBS) postoperatively.
† Between DBS on and DBS off periods.

In considering the coordinates of the center of the STN as


seen on T2-weighted coronal imaging, individual variability
in the AC–PC coordinates of STN was definitely observed.
FIG. 7. Lead locations plotted with respect to a drawing adapted This variability, however, was somewhat less marked than
from the Schaltenbrand and Wahren atlas depicting the area 4 mm is true for the globus pallidus,17,18,44 for which the impor-
inferior to the AC. Locations were plotted based on the distance of tance of direct targeting in adjusting or supplanting indirect
the contact array from the midline and the distance anterior to the coordinates may be relatively greater than for STN.
center of the red nucleus, as determined on postoperative axial FSE
MR images obtained 4 mm inferior to the commissures. The sym-
bols indicate the type of adverse effect that occurred at the lowest Role of Physiological Confirmation
voltage threshold during intraoperative test stimulation, which was Despite the relative accuracy of anatomical targeting, our
performed in bipolar mode by using the two contacts spanning the study contained a significant minority of cases in which the
STN at 185 Hz. The arrow indicates an outlier lead that was docu- initial anatomical target was either outside the STN, near a
mented to produce no improvement in the postoperative UPDRS
motor score. border of the STN, or within nonmotor regions of the nucle-
us. After refining the final DBS lead position with the aid of
microelectrode mapping, there were 19 procedures (25%)
would have resulted from the use of different targeting in which the final lead position was 2 mm or more away
methods, we found a hybrid method to be the most prac- from the anatomical target in the AP or lateral direction. In
tical. such cases it is clear that intraoperative physiological find-
For groups that perform targeting solely with indirect ings resulted in a final lead location that was significantly
(AC/PC–based) methods, as is necessary when ventriculog- closer to the motor area of the STN. Our data are consis-
raphy or computerized tomography scanning are the only tent with those of several other groups that have quantitat-
imaging modalities, it appears that the formula used in our ed target refinement by using intraoperative MER as the
last 56 cases (lateral 12 mm, AP 3 mm, and vertical 4
mm) is reasonable in that it provides a low degree of sys-
tematic bias. Several lines of evidence support this. The TABLE 7
mean coordinates of the MR imaging–visualized STN on Complications of treatment in 76 procedures
the preoperative images were lateral 12.9 mm, AP 3.5
mm, and vertical 5.3 mm, which are close to the indirect No. of
Complication Cases (%)
formula we used. Second, the mean AC–PC coordinates of
active contact(s) from postoperative MR imaging (which hemorrhagic stroke 1 (1.3)
takes into account adjustments made for intraoperative asymptomatic hemorrhage 0.5 cm3 seen on postop MR image 1 (1.3)
physiological findings and programming for optimal clin- lead fracture 1 (1.3)
ical effect) were lateral 11.8 mm, AP 2.4 mm, and ver- lead repositioning for poor initial position 1 (1.3)
return to op room for exploration/repair of lead extender 8 (10.4)
tical 3.8 mm, values that also are close to the indirect or IPG
target formula. Other groups have reported indirect target wound infection requiring intravenous antibiotics w/o hardware 1 (1.3)
formulas for STN that are similar to ours,3,39,41,51,57 whereas removal
some published formulas appear to place the initial anatom- intraop focal seizure 1 (1.3)
ical target slightly too anterior (AP coordinate 0 mm)35 or postop generalized seizure 1 (1.3)
slightly too medial (8–10 mm from midline)56 for the STN postop behavioral change lasting 24 hrs 9 (11.8)
suicide attempt 1 (1.3)
of the average brain.

J. Neurosurg. / Volume 97 / August, 2002 381


P. A. Starr, et al.

gold standard. Bejjani, et al.,5 demonstrated that five (21%) consistent with other physiological studies of the somato-
of 24 final stimulator locations differed by greater than 2 topical organization of the STN in humans41,47 and nonhu-
mm from the anatomical target. Rodriguez and colleagues39 man primates.13,54 In these studies, the leg territory is locat-
found that in 53% of cases, the initial anatomical target was ed more medially than the arm territory. In the AP domain,
not within the motor territory of the STN according to MER leg area is central, away from the rostral and caudal poles.47
and, thus, presumably was significantly modified. Zonen- Thus, if primarily leg-related cells are encountered on an
shayn and associates57 did not give a percentage of cases MER track, it can be inferred that the microelectrode is rel-
in which MER refinement was considered significant, but atively medial within the motor territory of the STN.
they reported a mean absolute adjustment of 1.27 mm in the Failure to record motor-responsive cells on MER, de-
initial coordinates as a result of MER, which is similar to spite recording a long segment of the STN, can occur with
ours. The role of “microelectrode refinement” in movement relatively anteromedial microelectrode penetrations. If no
disorders surgery is frequently debated. Some groups re- motor-responsive cells are found during the initial micro-
porting excellent outcomes for DBS of the STN do not use electrode exploration, our current practice is to continue ex-
MER.51 Without more studies that correlate DBS lead or le- ploration 2 to 3 mm posterolaterally.
sion location with outcome, it is not known how close is
close enough. Our work and several other reports indicate Intraoperative Stimulation-Induced Effects
that extremely meticulous anatomical targeting with histor- The major stimulation-induced adverse effects we ob-
ical (preoperative) MR imaging can place a lead within 1 served were similar to those reported by other researchers in
mm of the physiologically confirmed target in many, but the context of intraoperative stimulation or postoperative
not all cases. programming,2,5,29,37 and include paresthesias, dysarthria, fa-
Factors that limit the accuracy of image-guided stereo- cial contraction, oculomotor effects, and visual blurring. We
taxis include the following: 1) the application accuracy of observed stimulation-induced dyskinesias intraoperatively
frame-based stereotactic systems;30 2) image distortion ef- in only a minority of cases, whereas some authors have re-
fects (for MR imaging methods);46 3) brain shifts that occur ported that dyskinesias are a frequent observation.5 This dis-
after preoperative imaging is complete; 4) imperfect visu- crepancy may relate to the duration of intraoperative test
alization of the target structure that precludes the ability stimulation. With long-term stimulation (following postop-
to compensate completely for anatomical variability; and 5) erative programming of the IPG), increased dyskinesias of-
the fact that a specific physiological function may not al- ten develop after a lag time of a few minutes or hours.25 The
ways occur in the same anatomical location. The effect of relative infrequency of intraoperative stimulation-induced
microelectrode refinement on intraoperative target coordi- dyskinesias in our series may relate to the fact that our pe-
nates provides a summary measure of all these factors, but riod of intraoperative test stimulation is brief, focusing on
probably underestimates targeting errors because it rarely those effects that occur immediately at a given parameter
provides perfect correction of such errors. In this study, in choice.
addition to quantifying intraoperative target adjustments In one case, in which we used bipolar stimulation where
resulting from physiological mapping, we compared the fi- the negative contact was located in the SNr, we did observe
nal intraoperative target coordinates (after adjustment for acute stimulation-induced mood depression. This was phe-
physiological mapping) with MR imaging–verified lead co- nomenologically identical to the case described by Bejjani,
ordinates. This provided an objective measure of errors in et al.,4 in which the negative contact was also in the SNr.
stereotaxis that related specifically to application accuracy, When testing contacts within the STN itself, we did not
image distortion, and brain shift. We found the mean differ- observe acute stimulation-induced mood changes, although
ence between predicted and actual lead coordinates to be this has been reported to occur while using the same con-
1.4 to 2 mm in all dimensions, with much higher deviations tacts that are clinically effective for PD.24 Given the small
in individual cases. This discrepancy was greater than the size of the STN, the part of the nucleus participating in the
degree of correction achieved by physiological mapping, limbic (rather than motor) basal ganglia–thalamocortical
suggesting that physiological mapping somewhat undercor- circuit1 should be close to the motor circuit; thus, it is per-
rects the actual errors inherent in image-guided stereotaxis. haps surprising that mood-related effects are not observed
Some of these errors could be overcome by the use of real- more frequently.
time rather than preoperative MR imaging, as has recently The exact role of intraoperative test stimulation in guid-
been described for brain biopsy.19,20,34 ing STN lead placement has not been established. We fo-
cus on determining voltage thresholds for adverse effects.
Single-Cell Physiology and Somatotopy
We also assess stimulation-induced reduction in tremor
if present. Our primary criterion for correct lead place-
The mean cell discharge rate in the STN in patients with ment is microelectrode verification of the motor territory of
PD has been reported to be 41 Hz,31 37 Hz,23 39 Hz,5 33 the STN and not intraoperative stimulation-induced relief
Hz,41 46 Hz,28 and 47 Hz.45 In the present study, it was symptoms. Other groups may systematically test for intra-
34 Hz. These determinations are fairly consistent and sup- operative stimulation-induced improvements in rigidity or
port the view that the mean STN discharge rate in humans bradykinesia.5,29,35,51 We do not systematically assess these
with parkinsonian symptoms is greater than that measured benefits intraoperatively for several reasons. Frequently,
in healthy nonprimates, as is predicted in the current model there is an improvement in rigidity and bradykinesia associ-
of basal ganglia circuitry in PD.12 ated with microelectrode mapping or lead insertion, which
We found that microelectrode penetrations during which can confound the ability to assess any additional stimula-
leg-related activity was recorded were more medial than tion-induced improvements. In addition, the expected time
tracks for which arm-related activity was recorded. This is course of stimulation-induced improvements in rigidity and

382 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

bradykinesia is not well described and may take more than to activation of fibers of the cerebellothalamic pathways,
a few minutes. Therefore, the absence of immediate stim- which are intimately associated with the lateral border of
ulation-induced improvements in rigidity or bradykinesia the red nucleus, close to the medial border of STN. Testing
is not necessarily predictive of a poor long-term result. for these effects required electromyographic recording and
Even with respect to tremor control, the acute improve- signal averaging, whereas our study demonstrates that ad-
ment in tremor with intraoperative test stimulation may verse effects that are directly observable or reported by pa-
be less robust and require higher voltages than is the case tients may also be correlated with DBS lead location.
for acute thalamic stimulation,7 even though tremor control
with long-term STN stimulation is excellent.29,39 Use of MR Imaging to Document Lead Locations
Several other reports, one about DBS in the STN2 and
Correlation of Test Stimulation Findings With two about DBS in the globus pallidus internus,8,55 also pro-
Lead Location vide descriptions of the use of postoperative MR imaging to
Several types of stimulation-induced adverse effects cor- document lead locations systematically in a series of pa-
related with the lead location measured on postoperative tients. A number of pitfalls encountered using this tech-
MR images. We performed intraoperative test stimulation nique must be acknowledged. First, the manufacturer of the
in a standardized way at a pulse width of 60 sec and a fre- Medtronic DBS system cautions against MR imaging of
quency of 185 Hz. When paresthesias were the lowest- their device outside a limited range of imaging parameters
threshold adverse effect, the corresponding lead locations and field strengths. No group that uses MR imaging to mea-
proved to be slightly more medial, posterior, and superior sure lead locations has reported adverse effects during the
than leads not associated with paresthesias as the lowest imaging procedure.2,55 Two studies that formally addressed
threshold effect. Paresthesias are likely to be due to an ef- the safety of performing MR imaging in patients who re-
fect on the medial lemniscus, because this pathway courses ceive DBS found no indication of harmful effects (such as
posteromedial to the STN. As it ascends dorsally, it curves significant heating or torque on the lead) when a 1.5-tesla
laterally and approaches closer to the medial border of the magnet and a headcoil were used.38,48 Nevertheless, it is not
STN, which probably accounts for the somewhat sur- known whether the use of higher field strength magnets,
prising observation that contact locations associated with other types of radiofrequency receiving coils, or other pulse
paresthesias were slightly more dorsal than those not asso- sequences would result in clinically dangerous effects.
ciated with paresthesias. Lateral leads were more likely to As noted by Yelnik, et al.,55 the DBS lead produces an ar-
produce dysarthria or muscular contraction as the lowest- tifact that is larger than the actual lead. In our work, as in
threshold adverse effect. This is reasonable because the cor- theirs, we assumed that the actual contact location is at the
ticobulbar and corticospinal tracts travel immediately later- center of the observed artifact. This has not been proven.
al to the STN. When measured near the lead tip, gradient-echo and FSE
The occurrence of visual blurring as the lowest thresh- images contain a relatively small, round, and discrete lead
old-adverse effect was predictive of medial lead locations. artifact. If, however, the actual lead location is not in the
Medial and ventral to the STN lies the Edinger–Westphal center of the observed artifact, this would decrease the ac-
nucleus and its exiting fibers in the superficial layers of the curacy of MR imaging–based lead location measurements.
tract of the third cranial nerve. Stimulation-induced visual Finally, any method of documenting lead locations across
blurring could be due to the spread to these structures or to multiple patients is hampered by interindividual variability
the supranuclear pathways innervating them. Stimulation- in the shape and size of the target nucleus, as well as in its
induced diplopia was also observed in some cases but was exact position with respect to surrounding structures. Mag-
not as predictive of lead location as visual blurring because netic resonance images do not perfectly depict the borders
leads associated with diplopia at the standard test stimula- of the target (particularly with a superimposed lead artifact),
tion range were variably located. There are, however, many thus forcing us to infer lead location by reference to other
ways in which stimulation in the STN region could induce more reliable landmarks. We have attempted to deal with
oculomotor effects, in addition to the most obvious mecha- this issue by reporting lead locations with respect to the
nism of medial spread of current to the nucleus or tract of closest nuclear structure that is reliably visible on MR im-
the oculomotor nerve. Both the STN and the SNr partici- ages (the red nucleus), in addition to reporting AC/PC–
pate in circuits subserving oculomotor control,21,54 and mod- based lead coordinates. Yelnik, et al.,55 deformed a brain
ulation of these oculomotor subcircuits could affect eye atlas linearly in three dimensions so that it could be super-
movements. Alternatively, activation of corticobulbar fibers imposed on the MR image by using as landmarks those
could cause unilateral contraction of periorbital muscles, deep structures that were reliably visualized on both the MR
which would produce diplopia. image and the atlas. They acknowledge that linear transfor-
In one other study of STN stimulation the authors ana- mation of a brain atlas is not necessarily accurate, because
lyzed physiological–radiographic correlations. Ashby and it makes the questionable assumption that in any given di-
coworkers2 studied the effect of STN stimulation on volun- mension, all structures can be stretched or shrunk by the
tary electromyographic recordings of intrinsic hand mus- same factor to compensate for individual variability. Com-
cles. Production of a short-latency electromyographic effect putational mapping of different individual brains onto a sin-
was associated with more lateral leads and was thought to gle standardized brain map is a complex and important
be related to corticospinal tract activation. Production of a problem that has not yet been solved.
longer-latency electromyographic effect, which correlated
Location of the Active Contact in the Dorsal STN
closely with tremor suppression, was associated with more
medially placed leads. This effect was postulated to relate In the nonhuman primate that displays parkinsonian

J. Neurosurg. / Volume 97 / August, 2002 383


P. A. Starr, et al.

symptoms, the antiparkinsonian benefits of STN ablation fiber pathways are most readily activated during intraoper-
appear to be due to an effect on cells within the dorsolateral ative test stimulation. The approach angle of leads within
STN.9 Given the likelihood that ablation and high-frequen- our series, however, did not vary sufficiently to test this hy-
cy stimulation are mechanistically dissimilar,32 we cannot pothesis.
assume a priori that the therapeutic effect of DBS is also We prefer the relatively vertical trajectory because it usu-
due to an effect on the dorsolateral STN. Voges and co- ally grazes the anterior thalamus. The length of the thalam-
workers51 suggest that the contact providing the best anti- ic segment encountered becomes a useful landmark during
parkinsonian effect at the lowest stimulation parameters is microelectrode mapping. For example, if the STN is missed
located dorsal to the STN. Ashby and colleauges2 suggest on the initial microelectrode penetration, the extent of tha-
that the antitremor effect of STN stimulation may be due to lamic activity superior to the target area may be used to
stimulation of surrounding fiber pathways and not due to an determine whether the second MER track should be per-
effect on STN cell bodies. formed anterior or posterior to the first one.
In this study, the mean location of the active contacts of
the DBS leads, after programming for optimal clinical ben- Clinical Outcomes
efit, did correspond to the dorsolateral STN. This was true In this series, unilateral DBS of the STN produced a 23%
with respect to standard anatomical atlases33,42 and to intra- improvement in UPDRS Part III motor scores in the off-
operative single-cell physiological findings. Because we medication state at 3 months postimplantation. Bilateral
did not systematically evaluate all contacts for clinical effi- DBS of the STN produced a 45% improvement at 1 year
cacy, we cannot prove that a contact outside the STN would following the second implantation procedure. This is con-
not have similar efficacy. Nevertheless, it is clear that ben- sistent with outcomes reported by others for unilateral27 and
eficial effects on parkinsonian symptoms that are obtained bilateral10,11,26,29,35,52 DBS of the STN. In nearly all published
at moderate stimulation parameters, which are only expect- series of patients treated with bilateral DBS of the STN a 40
ed to affect tissue in a radius of several millimeters,32 can to 65% improvement in UPDRS motor scores have been re-
occur with the active contact within the dorsal STN. ported, despite major differences in the technical approach-
Our active contact locations (Figs. 6 and 7) do reflect es to the surgery. Thus, currently available outcomes data
some variability. This may relate to an imperfect measure- do not mandate the adoption of one particular set of implan-
ment of leads by MR imaging or to the inherent difficulty tation techniques.
in plotting contacts with respect to a single brain atlas (as in
Fig. 7), which cannot reflect the variability among patients’ Lead Locations in the STN: Comparison With
brains. It is likely, however, that some of this variability re- Other Groups
flects the learning curve in performing the procedure. Some
of the leads that we implanted are not located in the center Several groups have published STN locations of DBS
of the motor territory of the STN (Fig. 4, Case 3). As we leads, as verified by intraoperative ventriculography6,51 or
gained experience by correlating intraoperative physiolog- postoperative MR imaging.2 Benabid, et al.,6 reported loca-
ical findings with postoperative images, we became more tions of 95 STN active contacts in a graphic format (display
adept in the use of intraoperative physiological findings to of contacts on a Guiot diagram) but did not provide numer-
direct appropriate intraoperative adjustments in lead loca- ical mean contact locations. Their active contacts appear to
tion, resulting in more consistent lead locations. have been located 10 to 14 mm from the midline, 0 to 5 mm
Some of our more medially placed leads were located posterior to the midcommisural point, and 3 to 5 mm below
very close to the anterolateral edge of the red nucleus, prob- the intercommissural line. Our active contact locations are
ably close enough for part of the red nucleus to be affected clustered in a region very similar to theirs. This is notewor-
during long-term stimulation. In humans, lesions in the re- thy in light of the fact that their primary endpoint for deter-
gion of the red nucleus may cause coarse tremor49 or gait mining final lead position was intraoperative observation of
ataxia.16 We have not, however, observed stimulation-in- a clinical benefit,29 whereas ours was placement within the
duced tremor or ataxia, even in those patients with more microelectrode-defined motor territory of STN.
medially placed leads. Lead locations reported by Voges, et al.,51 and by Ashby
and coworkers2 differ somewhat from those reported here.
Voges and associates report the mean coordinates of the dis-
Approach Angle tal contact in 29 DBS implants in the STN as lateral 9.7
Stimulation-induced effects in neural tissue are sensitive mm, AP 2.3 mm, and vertical 3.9 mm for right-sid-
to the orientation of the lead with respect to fiber path- ed leads, and as lateral 10.6 mm, AP 2.1 mm, and verti-
ways.32,36 Approach angles used to implant DBS leads in the cal 3.9 mm for left-sided leads. Their electrode arrays
STN vary significantly between surgeons. Bejjani, et al.,5 were located slightly more medial than those in our study.
report an approach to the STN that involves the use of a rel- In addition, many of their active contacts were located more
atively lateral-to-medial trajectory, 20 to 30˚ from the verti- proximal along the quadripolar array than the distal contact
cal axis in the coronal projection compared with our mean and were located anterodorsally with respect to the STN.
lateral-to-medial angle of 10.2˚ from the vertical axis. With Therefore, their mean active contact locations were proba-
respect to the sagittal projection, Rodriguez, et al.,39 report bly also more anterior and superior than those in our study.
a relatively horizontal approach of 45˚ from the vertical axis Ashby and coworkers described locations for 19 STN leads
compared with the 29.2˚ from the vertical axis used in our based on postoperative MR images. They quantified only
study. Voges and associates51 also report a relatively hori- the lateral coordinate of their leads, which on average was
zontal approach, with the entry point 2 to 3 cm anterior to 9.7 mm from midline and, thus, 2 mm more medial than our
the coronal suture. Such differences could influence which average location.

384 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

Lead Location and Clinical Outcome testing. Microelectrode mapping resulted in a major change
Groups that have reported imaging-verified electrode lo- in the final location of the DBS lead in 25% of cases. Use
cations2,6,51 have all reported stimulation-induced improve- of postoperative MR imaging to measure lead location
ments in parkinsonian motor signs that are robust and quantitatively is technically feasible and provides valuable
similar to those reported here,26,27,29,51,52 despite apparent dif- correlations with intraoperative physiological mapping.
ferences in average lead location. Analysis of our own out-
lier lead locations suggests that leads in the anteromedial
Appendix
STN may be clinically ineffective, whereas leads near the
lateral margin of the STN may be unprogrammable due to Formulas for Calculating AC–PC Coordinates of Active Contact(s)
low-threshold corticobulbar activation. We have not yet ac- From Entry Point and Tip Coordinates
cumulated sufficient outcomes data, however, to allow for- 1) Let x = (lateral entry  lateral tip);
mal statistical analysis of the relationship of lead location to 2) Let y = (AP entry  AP tip); and
clinical outcome. More detailed correlations of lead loca- 3) Let z = (vertical entry  vertical tip)
tion with outcome will be needed to establish more clearly where lateral entry, AP entry, and vertical entry refer to the later-
the range of active contact locations that are be associated al, AP, and vertical distances, respectively, between the DBS entry
point and the midcommissural point; and where lateral tip, AP tip,
with optimal improvement in motor function. and vertical tip refer to the lateral, AP, and vertical distances,
respectively, between the DBS tip point and the midcommissural
Complication Avoidance point.
4) Let L = length of lead in brain = square root (x2 y2 z2);
Our rate of serious complications was low, with two 5) Let D = distance between midpoints of adjacent contacts (3
hemorrhages visible on neuroimaging, both of which were mm for the Medtronic model 3387 lead),
smaller than 1 ml and one of which was symptomatic. Post- 6) Let Q = distance from tip to midpoint of active contact array;
operative changes in mental status were initially common 7) then Q = 0.75 D (mean of active contact numbers)
following simultaneous bilateral implantation. During this Examples: If the system is programmed as monopolar Contact 2
series, we altered our protocol to place the distal contact in only, then the mean active contact number = 2; if the system is pro-
grammed as bipolar or monopolar with Contacts 2 and 3, then the
the ventral STN rather than in the SNr, thus reducing the mean active contact number = 2.5.
degree to which we overshot the target. This decreased the Then the lateral, AP, and vertical coordinates of the center of the
incidence of postoperative changes in mental status, sug- active contact array are the following:
gesting that such changes may be related in part to midbrain 8) Lateral center = (Q/L)x lateral tip,
edema associated with deep lead placement. 9) AP center = (Q/L)y AP tip, and
The essential steps that we adopted to minimize compli- 10) vertical center = (Q/L)z vertical tip;
cations while optimizing accuracy were as follows: 1) me- and the angles of approach in the coronal and sagittal projections
are the following:
ticulous MR imaging–based anatomical targeting, with an 11) approach angle in sagittal projection = arctan (y/z) and
attempt to use direct visualization of the STN to make fine 12) approach angle in coronal projection = arctan (x/z).
adjustments in the AC/PC–based targeting formula; 2) use
of surgical planning software for 3D trajectory planning
to avoid traversing deep sulci and, where possible, avoid Acknowledgments
the ventricles; 3) judicious use of sequential single-channel
MER in which the maximum degree of localizing informa- We thank Drs. Thomas Wichmann, William Dillon, Nicholas Bar-
baro, Paul Larson, and Robert Turner for their critical reading of the
tion was inferred from the minimum number of penetra- manuscript. We thank Kathleen Lamborn for assistance with statis-
tions; 4) maintenance of intraoperative systolic blood pres- tical analysis, and Tammy Damalcherevu for administrative assis-
sure at a level lower than 140 mm Hg; 5) placement of tance. We thank Heidi Clay, R.N., and Susan Heath, R.N. for their
the DBS lead tip no deeper than the ventral border of the work in DBS programming.
STN; and 6) postoperative high-resolution MR imaging in
every case, which provided immediate feedback regard-
ing the correlation of intraoperative physiological findings
with anatomical lead location. The last step facilitated rap- References
id ascension of the learning curve for performing DBS of 1. Alexander GE, Crutcher MD, DeLong MR: Basal ganglia-
the STN. thalamocortical circuits: parallel substrates for motor, oculomo-
tor, “prefrontal” and “limbic” functions. Prog Brain Res 85:
119–146, 1990
Conclusions 2. Ashby P, Kim YJ, Kumar R, et al: Neurophysiological effects of
stimulation through electrodes in the human subthalamic nucleus.
Our technical approach to stimulator implantation into Brain 122:1919–1931, 1999
the STN involved the use of MR imaging–based target- 3. Barlas O, Hanagasi HA, Imer M, et al: Do unilateral ablative le-
ing, identification of the motor territory and confirmation of sions of the subthalamic nucleus in parkinsonian patients lead to
nuclear boundaries by using MER, placement of the lead hemiballism? Mov Disord 16:306–310, 2001
4. Bejjani B, Damier P, Arnulf I, et al: Transient acute depression in-
within the motor territory of the STN, and brief intraopera- duced by high-frequency deep-brain stimulation. N Engl J Med
tive test stimulation to confirm the absence of low-voltage 340:1476–1480, 1999
stimulation-induced adverse effects. This approach is rela- 5. Bejjani BP, Dormont D, Pidoux B, et al: Bilateral subthalamic
tively safe and time efficient and resulted in lead locations stimulation for Parkinson’s disease by using three-dimensional
similar to those reported by groups that determine optimal stereotactic magnetic resonance imaging and electrophysiological
lead placement by extensive intraoperative neurological guidance. J Neurosurg 92:615–625, 2000

J. Neurosurg. / Volume 97 / August, 2002 385


P. A. Starr, et al.

6. Benabid AL, Benazzouz A, Gao D, et al: Chronic electrical stim- 28. Levy R, Hutchison WD, Lozano AM, et al: High-frequency
ulation of the ventralis intermedius nucleus of the thalamus and of synchronization of neuronal activity in the subthalamic nucleus
other nuclei as a treatment for Parkinson’s disease. Tech Neuro- of parkinsonian patients with limb tremor. J Neurosci 20:
surg 5:5–30, 1999 7766–7775, 2000
7. Benabid AL, Pollak P, Gao D, et al: Chronic electrical stimulation 29. Limousin P, Krack P, Pollak P, et al: Electrical stimulation of the
of the ventralis intermedius nucleus of the thalamus as a treatment subthalamic nucleus in advanced Parkinson’s disease. N Engl J
of movement disorders. J Neurosurg 84:203–214, 1996 Med 339:1105–1111, 1998
8. Bereznai B, Steude U, Seelos K, et al: Chronic high-frequency 30. Maciunas RJ, Galloway RL Jr, Latimer JW: The application accu-
globus pallidus internus stimulation in different types of dystonia: racy of stereotactic frames. Neurosurgery 35:682–695, 1994
a clinical, video, and MR report of six patients presenting with 31. Magnin M, Morel A, Jeanmonod D: Single-unit analysis of the
segmental, cervical, and generalized dystonia. Mov Disord 17: pallidum, thalamus and subthalamic nucleus in Parkinsonian pa-
138–144, 2002 tients. Neuroscience 96:549–564, 2000
9. Bergman H, Wichmann T, DeLong MR: Reversal of experimen- 32. Montgomery EB Jr, Baker KB: Mechanisms of deep brain stim-
tal parkinsonism by lesions of the subthalamic nucleus. Science ulation and future technical developments. Neurol Res 22:
249:1436–1438, 1990 259–266, 2000
10. Burchiel KJ, Anderson VC, Favre J, et al: Comparison of palli- 33. Morel A, Magnin M, Jeanmonod D: Multiarchitectonic and ste-
dal and subthalamic nucleus deep brain stimulation for advanced reotactic atlas of the human thalamus. J Comp Neurol 387:
Parkinson’s disease: results of a randomized, blinded pilot study. 588–630, 1997
Neurosurgery 45:1375–1384, 1999 34. Moriarty TM, Quinones-Hinojosa A, Larson PS, et al: Frameless
11. Deep Brain Stimulation for Parkinson’s Disease Study Group: stereotactic neurosurgery using intraoperative magnetic resonance
Deep-brain stimulation of the subthalamic nucleus or the pars imaging: stereotactic brain biopsy. Neurosurgery 47:1138–1146,
interna of the globus pallidus in Parkinson’s diseasee. N Engl J 2000
Med 345:956–963, 2001 35. Moro E, Scerrati M, Romito LM, et al: Chronic subthalamic nucle-
12. DeLong MR: Primate models of movement disorders of basal gan- us stimulation reduces medication requirements in Parkinson’s
glia origin. Trends Neurosci 13:281–285, 1990 disease. Neurology 53:85–90, 1999
13. DeLong MR, Crutcher MD, Georgopoulos AP: Primate globus 36. Ranck JB Jr: Which elements are excited in electrical stimulation
pallidus and subthalamic nucleus: functional organization. J Neu- of mammalian central nervous system: a review. Brain Res 98:
rophysiol 53:530–543, 1985 417–440, 1975
14. Drayer B, Burger P, Darwin R, et al: MRI of brain iron. AJR 147: 37. Rezai AR, Hutchison W, Lozano AM: Chronic subthalamic nu-
103–110, 1986 cleus stimulation for Parkinson’s disease, in Rengachary SS, Wil-
15. Favre J, Taha JM, Steel T, et al: Anchoring of deep brain stimu- kins RH (eds): Neurosurgical Operative Atlas. Park Ridge, IL:
lation electrodes using a microplate. Technical note. J Neurosurg American Association of Neurological Surgeons, 1999, Vol 8, pp
85:1181–1183, 1996 195–207
16. Felice KJ, Keilson GR, Schwartz WJ: “Rubral” gait ataxia. Neu- 38. Rezai AR, Lozano AM, Crawley AP, et al: Thalamic stimulation
rology 40:1004–1005, 1990 and functional magnetic resonance imaging: localization of cor-
17. Giller CA, Dewey RB, Ginsburg MI, et al: Stereotactic pallidot- tical and subcortical activation with implanted electrodes. Tech-
omy and thalamotomy using individual variations of anatomic nical note. J Neurosurg 90:583–590, 1999
landmarks for localization. Neurosurgery 42:56–65, 1998 39. Rodriguez MC, Guridi OJ, Alvarez L, et al: The subthalamic nu-
18. Guridi J, Gorospe A, Ramos E, et al: Stereotactic targeting of the cleus and tremor in Parkinson’s disease. Mov Disord 13 (Suppl
globus pallidus internus in Parkinson’s disease: imaging versus 3):111–118, 1998
electrophysiological mapping. Neurosurgery 45:278–289, 1999 40. Rodriguez-Oroz MC, Rodriguez M, Gonzalez S, et al: Neuronal
19. Hall WA, Liu H, Martin AJ, et al: Brain biopsy sampling by using activity in the red nucleus in Parkinson’s disease. Mov Disord 15
prospective stereotaxis and a trajectory guide. J Neurosurg 94: (Suppl 3):65, 2000 (Abstract)
67–71, 2001 41. Rodriguez-Oroz MC, Rodriguez M, Guridi J, et al: The subtha-
20. Hall WA, Martin AJ, Liu H, et al: Brain biopsy using high-field lamic nucleus in Parkinson’s disease: somatotopic organization
strength interventional magnetic resonance imaging. Neurosur- and physiological characteristics. Brain 124:1777–1790, 2001
gery 44:807–814, 1999 42. Schaltenbrand G, Wahren W: Atlas for Stereotaxy of the Hu-
21. Hikosaka O, Wurtz RH: Visual and oculomotor functions of mon- man Brain, ed 2. Stuttgart: Georg Thieme, 1977
key substantia nigra pars reticulata. I. Relation of visual and audi- 43. Starr PA, Subramanian T, Bakay RAE, et al: Electrophysiological
tory responses to saccades. J Neurophysiol 49:1230–1253, 1983 localization of the substantia nigra in the parkinsonian nonhuman
22. Hutchison WD: Microelectrode techniques and findings of glo- primate. J Neurosurg 93:704–710, 2000
bus pallidus, in Krauss JK, Grossman RG, Jankovic J (eds): Pal- 44. Starr PA, Vitek JL, DeLong M, et al: Magnetic resonance imag-
lidal Surgery for the Treatment of Parkinson’s Disease and ing-based stereotactic localization of the globus pallidus and sub-
Movement Disorders. Philadelphia: Lippincott-Raven, 1998, pp thalamic nucleus. Neurosurgery 44:303–314, 1999
135–152 45. Sterio D, Zonenshayn M, Mogilner AY, et al: Neurophysiological
23. Hutchison WD, Allan RJ, Opitz H, et al: Neurophysiological iden- refinement of subthalamic nucleus targeting. Neurosurgery 50:
tification of the subthalamic nucleus in surgery for Parkinson’s 58–69, 2002
disease. Ann Neurol 44:622–628, 1998 46. Sumanaweera TS, Adler JR Jr, Napel S, et al: Characterization of
24. Krack P, Kumar R, Ardouin C, et al: Mirthful laughter induced by spatial distortion in magnetic resonance imaging and its implica-
subthalamic nucleus stimulation. Mov Disord 16:867–875, 2001 tions for stereotactic surgery. Neurosurgery 35:696–704, 1994
25. Krack P, Pollak P, Limousin P, et al: From off-period dystonia to 47. Theodosopoulos PV, Marks W Jr, Christine C, et al: Physiology
peak-dose chorea: the clinical spectrum of varying subthalamic and functional organization of the human subthalamic nucleus.
nucleus activity. Brain 122:1133–1146, 1999 Mov Disord 15 (Suppl 3):43, 2000 (Abstract)
26. Kumar R, Lozano AM, Kim YJ, et al: Double-blind evaluation of 48. Tronnier VM, Staubert A, Hahnel S, et al: Magnetic resonance im-
subthalamic nucleus deep brain stimulation in advanced Parkin- aging with implanted neurostimulators: an in vitro and in vivo
son’s disease. Neurology 51:850–855, 1998 study. Neurosurgery 44:118–126, 1999
27. Kumar R, Lozano AM, Sime E, et al: Comparative effects of uni- 49. Vidailhet M, Jedynak CP, Pollak P, et al: Pathology of symptom-
lateral and bilateral subthalamic nucleus deep brain stimulation. atic tremors. Mov Disord 13 (Suppl 3):49–54, 1998
Neurology 53:561–566, 1999 50. Vlaardingerbroek MT, den Boer JA: Magnetic Resonance Im-

386 J. Neurosurg. / Volume 97 / August, 2002


Deep brain stimulation of the subthalamic nucleus

aging, Theory and Practice. Berlin: Springer-Verlag, 1996, pp 56. Yokoyama T, Sugiyama K, Nishizawa S, et al: Subthalamic nucle-
107–113 us stimulation for gait disturbance in Parkinson’s disease. Neuro-
51. Voges J, Volkmann J, Allert N, et al: Bilateral high-frequency surgery 45:41–49, 1999
stimulation in the subthalamic nucleus for the treatment of Parkin- 57. Zonenshayn M, Rezai AR, Mogilner AY, et al: Comparison of
son disease: correlation of therapeutic effect with anatomical elec- anatomic and neurophysiological methods for subthalamic nucle-
trode position. J Neurosurg 96:269–279, 2002 us targeting. Neurosurgery 47:282–294, 2000
52. Volkmann J, Allert N, Voges J, et al: Safety and efficacy of palli-
dal or subthalamic nucleus stimulation in advanced PD. Neurol-
ogy 56:548–551, 2001
53. Walton L, Hampshire A, Forster DMC, et al: A phantom study to
assess the accuracy of stereotactic localization, using T1-weight- Manuscript received July 19, 2001.
ed magnetic resonance imaging with the Leksell stereotactic sys- Accepted in final form April 29, 2002.
tem. Neurosurgery 38:170–178, 1996 This work was supported in part by a grant from Medtronic Neu-
54. Wichmann T, Bergman H, Delong MR: The primate subthalamic rological (Minneapolis, MN) to Philip A. Starr and by the Parkin-
nucleus. I. Functional properties in intact animals. J Neurophys- son’s Disease Research, Education, and Care Center (PDRECC) at
iol 72:494–506, 1994 the San Francisco Veteran’s Affairs Hospital.
55. Yelnik J, Damier P, Bejjani BP, et al: Functional mapping of the Address reprint requests to: Philip A. Starr, M.D., Ph.D., Depart-
human globus pallidus: contrasting effect of stimulation in the in- ment of Neurological Surgery, University of California at San Fran-
ternal and external pallidum in Parkinson’s disease. Neuroscience cisco, 779 Moffitt Hospital, 505 Parnassus Avenue, San Francisco,
101:77–87, 2000 California 94143. email: starrp@itsa.ucsf.edu.

J. Neurosurg. / Volume 97 / August, 2002 387

You might also like