Professional Documents
Culture Documents
Hoda S. Seddeq
Reprinted from
JOURNAL OF
BUILDING ACOUSTICS
Volume 17 · Number 1 · 2010
ABSTRACT
The standard ISO/IEC 17025:2005 “General requirements for the competence of testing and
calibration laboratories” introduces the necessity of evaluation of measurement uncertainties for
making them reliable and comparable. In order to fulfill acoustic comfort conditions indoors,
building acoustic have to establishes a set of criteria for the acoustic performance of the
building. These criteria may be expressed in terms of permissible minimum values for
normalized noise reduction NNR, noise reduction NR or apparent transmission loss ATL. This
paper discusses the uncertainty sources and presents a methodology for the calculation of
uncertainty of the airborne sound insulation between rooms carried out according to the of
standards ASTME 336 and examples of uncertainty budget, evaluated for the measurement
results of the sound insulation. The range of values obtained for the measurement results
uncertainty is 0.68 to 1.77.
1. INTRODUCTION
For noise control in buildings a set of criteria for the acoustic performance of building
has to be established in order to fulfill acoustic comfort conditions indoors. In the case
of a field evaluation of airborne sound insulation of a separating element between two
rooms, the uncertainty must be evaluated and added to the final result to be compared
to the prescribed limit. The housing and building national research centre in Egypt are
encouraged that tests in the field of building physics should have accreditation
according to the standard ISO/IEC 17025 [1]. Therefore the building acoustics
department in the housing and building national research centre should perform a
complete evaluation of uncertainty of in-situ sound insulation measurements.
70 Evaluated Uncertainties for Measurements of Airborne Sound Attenuation
between Rooms in Buildings
The standard ASTM E 336 uses a procedure originally developed for laboratory
measurements of the transmission loss of partitions. These procedures assume that the
rooms in which the measurement is made have a sound field that reasonably
approximates a diffuse field. Sound pressure levels in such rooms are reasonably
uniform throughout the room and average levels vary inversely with the logarithm of
the room sound absorption. Unfortunately not all rooms not specifically designed as
laboratory testing rooms will satisfy these conditions and conditions in a field test
should not be modified by any temporary means to improve performance of testing.
Practical experience and studies have shown that the test method is applicable to
smaller spaces with volume less than 150 cubic meters. The repeatability for this test
method may be due to variation in microphone and loudspeaker positions [2].
Higginson, R. F. estimated the repeatability standard deviations of the same
measurement team between the same two rooms using the same equipments as
follows: 2 dB for the frequency bands 100 to 200 Hz and 1 dB for the bands above
that. The corresponding 95% repeatability limits are 5.6 dB and 2.8 dB [3]. ASTM E
90 prescribes procedure for determining where the sound fields in the rooms and
microphone systems used to sample them give measurements uncertainties that are
low enough for the purpose of this test method. This is decided by calculating
confidence intervals from measurements of sound pressure level and reverberation
time [4].
The strict mathematical way for estimating uncertainties is described extensively in
the GUM [5], The uncertainty of measurement associated with the input estimates is
evaluated according to either a ‘Type A’ or a ‘Type B’ method of evaluation. The Type
A evaluation of standard uncertainty is the method of evaluating the uncertainty by
the statistical analysis of a series of observations. In this case the standard uncertainty
is the experimental standard deviation of the mean that follows from an averaging
procedure or an appropriate regression analysis. The Type B evaluation of standard
uncertainty is the method of evaluating the uncertainty by means other than the
statistical analysis of a series of observations. In this case the evaluation of the standard
uncertainty is based on some other scientific knowledge [5].
u( x i ) = s ( x i ) (1)
∂f
where the partial derivative, is termed a sensitivity coefficient.
∂x i
The interval in which the value of the quantity subject to measurement can be asserted
to lie with a specified level of confidence is referred to as the, expanded uncertainty U
(it means that one can with the relevant level of confidence believe that the value of Y,
estimated by y, lies within the limits y – U ≤ Y ≤ y + U ). The expanded uncertainty is
obtained by multiplying a combined standard uncertainty by a coverage factor k
depending on the desired level of confidence and the type of statistical distribution.
(a− − a+ ) / 2
u( x i ) = (4)
3
T
NNR = LS − LR + 10 log (5)
0.5
where
LS = the average sound pressure level in the source room, dB, and
LR = the average sound pressure level in the receiving room, dB,
T = the average reverberation time in the receiving room,
72 Evaluated Uncertainties for Measurements of Airborne Sound Attenuation
between Rooms in Buildings
For each frequency band the averaged sound pressure L in each room: is calculated
according to:
1 n Li
L = 10 log ∑ 10 10 (7)
n i =1
Each individual sound pressure level measurement, Li, will contribute to the type A
uncertainty. The standard uncertainty is calculated from the standard deviation, s(Li) for
N different measurements as follows:
u( L ) = s( Li ) / N (8)
Standard Uncertainty
Quantity Uncertainty Source of Probability uncertainty contribution
Xi symbol uncertainty distribution u(xi) ui(y) = ciu(xi)
where
LS = the average sound pressure level in the source room, dB, and
LR = the average sound pressure level in the receiving room, dB, and
S = the area of the tested sample, m2, and
A2 = the sound absorption in the receiving room m2
The sound absorption of the receiving room, A2 is determined, by [8]:
where:
c = speed of sound in air, m/s,
V = volume of room, m3, and
d = rate of decay of sound pressure level in the room, dB/s.
where:
d = 60/T,
T = average reverberation time in the receiving room, s.
The speed of sound changes with temperature, and is calculated for the conditions
existing at the time of test from the equation [8]:
1
c = 20.047 ( 273.15 + t ) 2 m/s (11)
where:
t = receiving room temperature, °C.
For very large halls with highly reflecting surfaces, air absorption at high frequencies
can be the dominant phenomenon which is strongly influenced by the relative humidity.
For frequencies below 2 kHz, absorption due to the air is not significant and therefore
ignored [9].
For each frequency band equation (9) can be expanded as :
Standard Uncertainty
Quantity Uncertainty Source of Probability Uncertainty contribution
Xi symbol uncertainty Distribution u(xi) ui(y) = ci u(xi)
NR = LS − LR (13)
where
LS = the average sound pressure level in the source room, dB, and
LR = the average sound pressure level in the receiving room, dB.
where the sensitivity coefficients, ci are equal to:
∂NR ∂NR
=1, = −1
∂LS ∂LR
Standard Uncertainty
Quantity Uncertainty Source of Probability Uncertainty contribution
Xi symbol uncertainty Distribution u(xi) ui(y) = ci u(xi)
70
Transmission loss and level difference (dB)
60
50
40
30
ATL
NR
20 NNR
10
100 125 160 200 250 315 400 500 630 800 1000 1250 1600 2000 2500 3150 4000
Third octave frequencies Hz
Figure 1. Mean values of the apparent transmission loss ATL, normalized noise
reduction NNR and noise reduction NR.
with a 1000 Hz sound level calibrator type 4231 (B&K). The reverberation time of the
receiving room was measured with two source positions and five microphone positions,
two measurements each. The area of the separating wall was 13 m2.
Figure 1 presents the mean values of the apparent transmission loss ATL, normalized
noise reduction, NNR and noise reduction, NR.
3.1. Results
Figures (2), (3), (4) show the contribution for sound insulation measurement results at
1000 Hz. These figures show that the sound pressure level measurements make the
most contribution to the total uncertainty. The measurement of reverberation time also
has a significant effect on the total uncertainty. Calibration errors – as may be expected
– made less contribution to the total uncertainty. Figure (4) shows the contribution of
measuring area and volume is low. The contribution of temperature measurement is
very low and may be neglected.
Table (4) gives the values of the expanded uncertainty, with 95% confidence
interval (k = 2) for the measured ATL, NNR and NR at third octave frequencies from
100 to 4000 Hz. These values obtained for the expanded uncertainty range between
BUILDING ACOUSTICS · Volume 17 · Number 1 · 2010 79
1
0.9
0.8
Uncertainty contribution value
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
)
T)
p)
l)
uc
LS
LR
ic
na
es
ca
m
u(
Tr
a
u(
u(
Lδ
ra
δL
δL
u(
p
u(
u(
u(
δL
u(
0.9
0.8
Uncertainty contribution value
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
u(LS) u(LR) u(δLpramp) u(δLmic) u(δLana) u(δLcal) uc
0.9
0.8
Uncertainty contribution value
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
)
v)
s)
t)
T)
p)
)
.)
u( l)
)
uc
LS
LR
al
ic
na
es
al
ca
u(
al
u(
u(
am
u(
m
c
Vc
Sc
Tr
a
u(
u(
δt
Lδ
δL
δL
pr
u(
u(
u(
u(
u(
u(
δL
u(
0.68 and 1.77. These fall within the limits published in ASTM E336 and Portuguese
legislation [10].
4. CONCLUSION
This study presents a methodology for the standard uncertainty calculation of the field
airborne sound insulation measured according to ASTM E336. The analysis of
uncertainties was carried out using the same measurement team making measurements
between the same two rooms using the same equipment. The rooms were small spaces
with volumes less than 150 cubic meters. The final combined uncertainty for airborne
sound attenuation was found to depend mainly on the spread of values of sound level
in the source room, sound level in the receiving room and the reverberation time. To
reduce the magnitude of the combined total standard uncertainty effort should be
concentrated on reducing the variation of sound pressure level in rooms and variation
of reverberation time by selecting suitable positions of the sound source and
microphone. Whilst it is clear that the magnitude of the final combined total standard
uncertainty effort is affected by calibration errors of the instrumentation this was minor
in comparison.
The area and volume measurement made a quite low contribution to the total
uncertainty. The temperature uncertainty had a negligible effect on the total uncertainty
so it can be neglected.
REFERENCES
[1] ISO/IEC 17025, General requirements for the competence of testing and
calibration laboratories, Second edition, 2005-05-15
[2] ASTM E336, Standard test method for measurement of airborne sound
attenuation between rooms in buildings, 2005
[3] Higginson, R. F., A Study of Measuring Techniques for Airborne Sound Insulation
in Building” Journal of Sound and Vibration, Vol. 21, 1972 p. 405
[4] ASTM E90-04, Standard test method for laboratory measurement of airborne
sound transmission loss of building partition and elements
[5] ISO Guide 98, Guide to the Expression of Uncertainty in Measurement.
International Organization for Standardization, 1995
[6] EA-4/02, Expression of the Uncertainty of Measurement in Calibration, European
cooperation for Accreditation, December, 1999.
[7] EUROLAB Technical Report 1/2002 , Measurement Uncertainty in testing, 2002
[8] ASTM E2235, Test Method For The Measurement of Decay Rates for Use In
Sound Insulation Test Methods
[9] Dennis A. Bohn, Environmental Effects on the Speed of Sound, J. Audio Eng.
Soc. Vol. 36, No. 4, 1988
[10] Regulamento dos Requisitos Acústicos de Edifícios. Building Acoustics Noise
Code- in Portuguese., Law nº 129/2002 of 11 de May 2002