INSTRUCTIONS
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 1 of 31
Table of Contents
Table of Contents ................................................................................................................................................................................2
Service disclaimers ............................................................................................................................................................................3
Overview .............................................................................................................................................................................................. 12
Product Lines ................................................................................................................................................................................ 12
Kit Contents ................................................................................................................................................................................... 12
Manpower Requirements ....................................................................................................................................................... 13
Tools and Test Equipment ...................................................................................................................................................... 13
Reference Documents.............................................................................................................................................................. 13
Safety ................................................................................................................................................................................................ 13
Hazard Alerts ................................................................................................................................................................................ 13
Installation Instructions ................................................................................................................................................................ 14
Pre-install ........................................................................................................................................................................................ 14
Mainframe Cabling .................................................................................................................................................................... 16
Power Supply Mounting........................................................................................................................................................... 18
Reference Monitor Video Cable Installation .................................................................................................................. 19
Reference Monitor Assembly ................................................................................................................................................ 21
Workstation Video Cable Installation................................................................................................................................ 27
Test ......................................................................................................................................................................................................... 29
Labels .................................................................................................................................................................................................... 30
Finish ...................................................................................................................................................................................................... 31
Revision History ................................................................................................................................................................................ 31
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 2 of 31
Service disclaimers
WARNING This Service Manual is available in English only.
(EN)
If a customer's service provider requires a language other than English, it is
the customer's responsibility to provide translation services.
Do not attempt to service the equipment unless this service manual has
been consulted and is understood.
Failure to heed this warning may result in injury to the service provider,
operator, or patient from electric shock, or from mechanical or other
hazards.
ƴƵƪƩƸƴƵƪƫƩƪƲƭƪ ƷǓLJDžǖNJǕLJǍnjǒǓǕǟǏǓLJǓljǖǗLJǓNJǒDžǐǍǜǒǓǖDžǑǓǒDžDžǒLjǐǍǎǖǏǍNJnjǍǏ
(BG)
ƥǏǓljǓǖǗDžLJǜǍǏǒDžǘǖǐǘLjǍǒDžǏǐǍNJǒǗDžǍnjǍǖǏLJDžNJnjǍǏǕDžnjǐǍǜNJǒǓǗDžǒLjǐǍǎǖǏǍ
ǏǐǍNJǒǗǟǗǓǗLjǓLJDžǕǤnjDžǔǕNJljǓǖǗDžLJǤǒNJǗǓǒDžǘǖǐǘLjǍǗNJnjDžǔǕNJLJǓlj
ƲNJǖNJǓǔǍǗLJDžǎǗNJljDžǕNJǑǓǒǗǍǕDžǗNJǓdžǓǕǘljLJDžǒNJǗǓDžǏǓǒNJǖǗNJǖNJ
ǏǓǒǖǘǐǗǍǕDžǐǍǖǟǖǍǕDžnjdžǕDžǐǍǗǓLJDžǖNJǕLJǍnjǒǓǕǟǏǓLJǓljǖǗLJǓ
ƲNJǖǔDžnjLJDžǒNJǗǓǒDžǗǓLJDžǔǕNJljǘǔǕNJNjljNJǒǍNJǑǓNjNJljDžljǓLJNJljNJljǓǒDžǕDžǒǤLJDžǒNJ
ǒDžljǓǖǗDžLJǜǍǏDžǒDžǘǖǐǘLjǍǓǔNJǕDžǗǓǕDžǍǐǍǔDžǛǍNJǒǗDžǓǗǗǓǏǓLJǘljDžǕ
ǑNJǚDžǒǍǜǒǍǍǐǍljǕǘLjǍǓǔDžǖǒǓǖǗǍ
㆙࿌ ᮏ亜ᾖㇳℴṭ㍸ὃ劙㔯䇰㛔ࠋ
(ZH-CN)
ዴᯝᐈ㇟䘬亜ᾖ㚵≉㍸ὃ⪅㟂せ㠀ⱥᩥ∧ᮏ㸪ᐈ㇟暨冒埴㍸ὃ侣孹㚵≉ˤ
ᮍ宎乮断宣⬴ℐ䎮妋㛔亜ᾖㇳℴᷳ⇵炻ᶵ⼿徃埴亜ᾖˤ
㑂Ᏺ宍㆙࿌ྍ⬟⮤农亜ᾖᥦ౪⪅ࠊ᧯స␀ㆾか侭ᅉゐ䓝ˣ㛢㡘ᡈ
༴ᐖ㐀ᡂⓗ㌇Ọࠋ
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 3 of 31
㆙࿌ ᮏ⥔ಟᡭഹ᭷ⱥᩥ∧ᮏࠋ
(ZH-TW)
ዴ㢳ᐈⓗ⥔ಟ᭹ົᥦ౪⪅せồⱥᩥ௨እⓗㄒゝ㸪㢳ᐈ᭷㈐௵ᥦ౪⩻㆞ⓗ᭹
ົࠋ
㝖㠀ᕬ㞍教᷎䝕妋㛔䵕ᾖㇳℲ炻⏎⇯婳⊧▿娎䵕ᾖ姕⁁ˤ
ᮍὀពᮏ㆙࿌ྍ⬟ᑟ⮴⥔ಟ᭹ົᥦ౪⪅ࠊ᧯సேဨᡈᝈ㐼ཷ฿゙
㟁ࠊ⏤ᶵᲔᡈᅉ⣲㐀ᡂⓗ༴ᐖࠋ
UPOZORENJE 2YDMVHUYLVQLSULUXĆQLNGRVWXSDQMHVDPRQDHQJOHVNRPMH]LNX
(HR)
$NRGDYDWHOMXVOXJHNOLMHQWDWUDåLQHNLGUXJLMH]LNNOLMHQWMHGXåDQRVLJXUDWL
uslugu prijevoda.
1HSRNXãDYDMWHVHUYLVLUDWLXUHĊDMDNRQLVWHSURXĆLOLLVKYDWLOLRYDMVHUYLVQL
SULUXĆQLN
1HSULGUåDYDQMHPRYRJXSR]RUHQMDPRåHGRĄLGRWMHOHVQHR]OMHGHGDYDWHOMD
XVOXJHUXNRYDWHOMDLOLSDFLMHQWDGMHORYDQMHPHOHNWULĆQRJudara ili zbog
PHKDQLĆNLKLOLGUXJLKRSDVQRVWL
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 4 of 31
WAARSCHUWING DEZE ONDERHOUDSHANDLEIDING IS ALLEEN VERKRIJGBAAR IN HET ENGELS.
(NL)
ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL NODIG HEEFT, DAN IS
DE KLANT VERANTWOORDELIJK VOOR DE VERTALING VAN DEZE
HANDLEIDING.
PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING IS GERAADPLEEGD EN BEGREPEN.
INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, KAN HET
ONDERHOUDSPERSONEEL, DE BEDIENER OF EEN PATIËNT LETSEL OPLOPEN
ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK OF DOOR MECHANISCHE OF
ANDERE RISICO'S.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 5 of 31
Warnung Diese Serviceanleitung existiert nur in englischer Sprache.
(DE)
Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es Aufgabe
des Kunden, für eine entsprechende Übersetzung zu sorgen.
Versuchen Sie nicht, diese Anlage zu warten, ohne dass die Serviceanleitung
gelesen und verstanden wurde.
Bei Nichtbeachtung dieser Warnung kann es zu Verletzungen des
Kundendiensttechnikers, des Bedieners oder des Patienten durch
Stromschläge, mechanische oder sonstige Gefahren kommen.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 6 of 31
AVVERTENZA Il presente Manuale di assistenza è disponibile soltanto in inglese.
(IT)
Qualora il tecnico del servizio di assistenza del cliente richieda una lingua
diversa dall’inglese, è responsabilità del cliente fornirne la traduzione.
L’intervento sull’apparecchiatura è consentito solo dopo aver consultato e
compreso questo manuale di assistenza.
La mancata osservanza di questa avvertenza può provocare lesioni al
tecnico dell’assistenza o al paziente, causate da scosse elettriche oppure da
pericoli di natura meccanica o altro.
㆙࿌ ࡇࡢࢧ࣮ࣅࢫ࣐ࢽࣗࣝࡣⱥㄒ࡛ࡢࡳᥦ౪ࡉࢀ࡚࠸ࡲࡍࠋ
(JA)
࠾ᐈᵝࡢࢧ࣮ࣅࢫᢸᙜ⪅ࡀⱥㄒ௨እࡢゝㄒࢆᚲせࡍࡿሙྜࠊ⩻ヂࢧ࣮
ࣅࢫࡢᥦ౪ࡣ࠾ᐈᵝࡢ㈐௵࡞ࡾࡲࡍࠋ
ࡇࡢࢧ࣮ࣅࢫ࣐ࢽࣗࣝࢆ⇍ㄞࡋ⌮ゎࡋ࡚ࡽࠊ⨨ࡢࢧ࣮ࣅࢫࢆ⾜࡞
ࡗ࡚ࡃࡔࡉ࠸ࠋ
ࡇࡢ㆙࿌ᚑࢃ࡞࠸ሙྜࠊឤ㟁ࡲࡓࡣᶵᲔⓗࡶࡋࡃࡣࡑࡢࡢ༴㝤せᅉ
ࡼࡾࠊࢧ࣮ࣅࢫᢸᙜ⪅ࠊ᧯సဨࠊࡲࡓࡣᝈ⪅ࡀ㈇യࡍࡿཎᅉ࡞ࡾࡲ
ࡍࠋ
%5ę',1ý-806 âę$3.$/3(652.$6*5ý0$7$,53,((-$0$7,.$,$1*Ģ89$/2'ý
(LV)
-$./,(17$$3.$/3(661,('=ċ-$01(3,(&,(â$0$,1)250ý&,-$&,7ý
9$/2'ý1(9,6$1*Ģ8./,(17$3,(1ý.806,512'52â,1ý778/.2â$18
1(9(,&,(7$35ę.2-80$$3.$/3,%(=$3.$/3(652.$6*5ý0$7$6
,=/$6ęâ$1$6816APRAŠANAS.
âę%5ę',1ý-80$1(,(9ċ52â$1$9$55$'ę7(/(.75,6.ý6675ý9$6
75,(&,(1$0(+ý1,6.89$,&,785,6.8,=5$,6ę7875$808$3.$/3(6
61,('=ċ-$023(5$725$09$,3$&,(17$0
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 7 of 31
ě63ď-,0$6 â,6(.63/2$7$9,029$'29$63$7(,.,$0$67,.$1*/ņ.$/%$
(LT)
-(,./,(1723$6/$8*ņ7,(.ď-$65(,.$/$8-$9$'292.,7$1(,$1*/ņ
.$/%$3$6,5ŀ3,17,-27(.6729(57,08<5$./,(172$76$.20<%ď
1(0ď*,1.,7($7/,.7,ě5$1*267(&+1,1ď635,(ä,ŀ5261(%(176.$,7ď7(âě
(.63/2$7$9,029$'29ā,5-ě6835$727(
-(,1($7.5(,36,7('ď0(6,2ěâě3(563ď-,0ā3$6/$8*ņ7(,.ď-$6
23(5$725,86$53$&,(17$6*$/,%ŀ7,68ä$/27,'ď/(/(.752660ŀ*,2
MECHANINIO AR KITOKIO PAVOJAUS.
2675=(Ŋ(1,( 1,1,(-6=<32'5đ&=1,.6(5:,62:<'267đ31<-(67-('<1,(:-đ=<.8
(PL) ANGIELSKIM.
-(ĸ/,'267$:&$86à8*./,(17$:<0$*$-đ=<.$,11(*21,Ŋ$1*,(/6.,
=$3(:1,(1,(86à8*,7à80$&=(1,$-(672%2:,ā=.,(0./,(17$
1,(35Ï%2:$ă6(5:,62:$ă:<326$Ŋ(1,$%(==$32=1$1,$6,đ,
=52=80,(1,$1,1,(-6=(*232'5đ&=1,.$6(5:,62:(*2
1,(=$67262:$1,(6,đ'27(*22675=(Ŋ(1,$02Ŋ(632:2'2:$ă
2%5$Ŋ(1,$8'267$:&<86à8*23(5$725$/8%3$&-(17$::<1,.8
325$Ŋ(1,$(/(.75<&=1(*2=$*52Ŋ(1,$0(&+$1,&=1(*2%ā'ň,11(*2
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 8 of 31
$7(1ļ,( $&(670$18$/'(6(59,&((67(',6321,%,/180$,Ì1/,0%$(1*/(=ÿ
(RO)
'$&ÿ81)851,=25'(6(59,&,,$/&/,(178/8,1(&(6,7ÿ2$/7ÿ/,0%ÿ
'(&Æ7(1*/(=$(67('('$725,$&/,(178/8,6ÿ)851,=(=(275$'8&(5(
18Ì1&(5&$ļ,6ÿ5(3$5$ļ,(&+,3$0(178/'(&Æ7'$&ÿ$ļ,&2168/7$7ĺ,
$ļ,Ì1ļ(/(6 ACEST MANUAL DE SERVICE.
IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUC(/$5ÿ1,5($
DEPANATORULUI, A OPERATORULUI SAU A PACIENTULUI ÎN URMA
(/(&752&87ÿ5,,6$8$3(5,&2/(/250(&$1,&(6$8'($/7ÿ1$785ÿ
ƳƶƷƳƵƳƫƲƳ ƩDžǒǒǓNJǕǘǏǓLJǓljǖǗLJǓǔǓǓdžǖǐǘNjǍLJDžǒǍǣǔǕNJljǐDžLjDžNJǗǖǤǗǓǐǡǏǓǒDž
(RU) DžǒLjǐǍǎǖǏǓǑǤnjǠǏNJ
ƪǖǐǍǖNJǕLJǍǖǒǓǑǘǔNJǕǖǓǒDžǐǘnjDžǏDžnjǜǍǏDžǒNJǓdžǚǓljǍǑǓǕǘǏǓLJǓljǖǗLJǓǒDž
ljǕǘLjǓǑǤnjǠǏNJnjDžǏDžnjǜǍǏǘǖǐNJljǘNJǗǖDžǑǓǖǗǓǤǗNJǐǡǒǓLJǠǔǓǐǒǍǗǡǔNJǕNJLJǓlj
ƴNJǕNJljǓdžǖǐǘNjǍLJDžǒǍNJǑǓdžǓǕǘljǓLJDžǒǍǤǓdžǤnjDžǗNJǐǡǒǓǔǕǓǜǍǗDžǎǗNJljDžǒǒǓNJ
ǕǘǏǓLJǓljǖǗLJǓǍǔǓǎǑǍǗNJǍnjǐǓNjNJǒǒǠNJLJǒNJǑǖLJNJljNJǒǍǤ
ƲNJǖǓdžǐǣljNJǒǍNJǍǒǙǓǕǑDžǛǍǍǍnjljDžǒǒǓLjǓǔǕNJljǘǔǕNJNjljNJǒǍǤǑǓNjNJǗǖǗDžǗǡ
ǔǕǍǜǍǒǓǎǔǓǕDžNjNJǒǍǤǖǔNJǛǍDžǐǍǖǗDžǔǓǓdžǖǐǘNjǍLJDžǒǍǣǓǔNJǕDžǗǓǕDžǍǐǍ
ǔDžǛǍNJǒǗDžǢǐNJǏǗǕǍǜNJǖǏǍǑǗǓǏǓǑǔǕǍǜǍǒNJǒǍǤǍǑǑNJǚDžǒǍǜNJǖǏǍǚǗǕDžLJǑǍǐǍ
ljǕǘLjǍǚǔǓLJǕNJNjljNJǒǍǎ
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 9 of 31
Opozorilo 7DVHUYLVQLSULURĆQLNMHQDYROMROHYDQJOHãĆLQL
(SL)
ąHVWUDQNLQSRQXGQLNVWRULWYHSRWUHEXMHGUXJMH]LNUD]HQDQJOHãĆLQHSRWHP
je stranka odgovorna, da zagotovi prevajalske storitve.
1HSRVNXVLWHRSUDYOMDWLVHUYLVLUDQMDĆHQLVWHSUHEUDOLLQUD]XPHOLWHJD
VHUYLVQHJDSULURĆQLND
ąHQHXSRãWHYDWHWHJDRSR]RULODODKNRSULGHGRSRãNRdb ponudnika storitve,
RSHUDWHUMDDOLEROQLND]DUDGLHOHNWULĆQHJDXGDUDDOLPHKDQVNLKLQGUXJLK
nevarnosti.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 10 of 31
'ĝ..$7 %86(59ĝ6.,/$98=81816$'(&(ĝ1*ĝ/ĝ=&(6ĝ0(9&87785
(TR)
0hĺ7(5ĝ1ĝ1+ĝ=0(76$ĕ/$<,&,6,ĝ1*ĝ/ĝ=&(',ĺ,1'$%ĝ5%$ĺ.$'ĝ/'(
.,/$98=7$/(3('(56(7(5&h0(+ĝ=0(7/(5ĝ1ĝ16$ĕ/$10$6,0hĺ7(5ĝ1ĝ1
62580/8/8ĕ8'85
6(59ĝ6.,/$98=8182.8<83$1/$0$'$1(.ĝ30$1/$5,1%$.,0,1,
<$30$<,1
%88<$5,<$'ĝ..$7('ĝ/0(0(6ĝ(/(.75ĝ.0(.$1ĝ.9(<$'ĝĕ(5
7(+/ĝ.(/(5'(1'2/$<,+ĝ=0(76$ĕ/$<,&,1,123(5$7g5h19(<$
+$67$1,1<$5$/$10$6,1$<2/$d$%ĝ/ĝ5
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 11 of 31
Overview
This instruction describes the installation of the reference monitor on the C-Arm control panel
housing. The reference monitor replicates the left monitor image at the control panel. The reference
monitor Kit is available for field system upgrades and new system installations.
Product Lines
x GE OEC 9900 Elite C-Arms excluding motorized systems.
Kit Contents
The kit content is for field upgradable systems, and may differ for new system installations.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 12 of 31
Manpower Requirements
x 1 person, 1.5 hours
Reference Documents
x 9900 Service Manual
x Reference Monitor User Manual
Safety
Do not attempt to service or maintain this equipment before reading this entire publication .
Note: Safety glasses and safety shoes are required when performing any installation, servicing, or
maintenance of equipment. See the hazard matrix in the following Energized Electrical Work
Procedure for specific PPE relative to energized electrical work.
Note: Use proper PPE per the 9900 Service manual.
CAUTION Follow all LockOut/TagOut (LOTO) procedures as described in the service manual.
Hazard Alerts
This document uses three kinds of hazard alerts:
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 13 of 31
Installation Instructions
Note: The following instructions below are intended for field installations.
Pre-install
Functional Check
1. Start up the system.
2. Check system for functionality.
3. If any issues are found, resolve them before proceeding.
C-Arm Covers
1. Raise the vertical column 4 to 8 inches to make it easier to remove the C-Arm cover.
2. Power down the system.
3. Unplug the system and disconnect the interconnect cable.
4. Remove front top cover, right side cover, left side cover, rear cover, and front leg cover.
5. Remove the cross arm brake handle and control panel cover.
6. Disconnect the three (3) internal cables.
7. Remove the Up, Down, and X-ray switches from the control panel cover.
8. If there is a non-(QJOLVK;-ray warning label on the existing cover, select the label
5476335-;;IURPWKHNLWPDWFKLQJWKHODQJXDJHRIWKH;-ray warning label on the
original old cover. Find the 5476335-1EN and 00-881074-04 labels. Keep the switches
and the labels in a safe place for later installation.
9. Dispose of the cover.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 14 of 31
Note: Dispose of the cover according to local procedure.
Workstation Covers
Remove the rear and left side cover from the Workstation.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 15 of 31
Mainframe Cabling
1. Open the cover panel by removing the two (2) screws to access the Power Signal
Interface Board. Retain the screws for later use.
2. Disconnect and remove connector J8 from the Power Signal Interface Board.
3. Remove both leads (DS2-A and DS2-B) from the AC indicator lamp connector, using
needle nose pliers and pulling firmly.
4. Locate the Power Motor Relay Board, and then disconnect J7 at the bottom of the board
and remove the cable (PN 00-879456-;;IURPWKHV\VWHP
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 16 of 31
5. From the kit, obtain the power cable (PN 00-885230-01) and then connect J7 to P7 on the
Power Motor Relay Board.
Note: Check the other connectors on the board to ensure they did not come loose.
6. Attach the ground wire from the new cable to the existing ground screw at the bottom
left chassis.
7. Route the portion of the cable with J8 through the side of the power panel, toward the
Power Signal Interface Board, and then connect J8 to P8 on the Power Signal Interface
Board.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 17 of 31
8. Connect DS2-A and DS2-B to the AC indicator lamp. The leads are not polarized and can
be connected either way.
9. Close the panel and secure it with the two (2) screws that were previously removed.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 18 of 31
Note: The power supply mounts to the inside left wall of the toe so that the power cord elbow points
down.
2. From the kit, obtain the power supply retention clip (PN 20478-001) from the kit.
3. On the power supply, remove the release liners from the double-sided tape.
4. Slide the metal retainer clip over the power supply about halfway up the body of the
power supply. Do not let the clip stick to the tape.
5. Mount the power supply flush with the inside left wall of the C-Arm toe (as viewed from
the rear of the mainframe) with the clip over the outside of the toe wall. Press the power
supply firmly to bond the adhesive tape to the wall.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 19 of 31
2. From the kit, obtain the new LEMO adapter cable (PN 5477291) and insert the two (2)
LEMO pins into the P1 LEMO socket (center white) lead to P1 pin 21 and ground
(white/black stripe) lead to P1 pin 23.
3. From the kit, obtain a tie wrap and install it to the P1 LEMO wires that were removed
earlier, including the new wires.
4. Route the BNC connector end of the video cable down through the side hole in the
chassis, taking care to keep the cable away from the high voltage lines, and down behind
the round bar into the toe space.
Video
cable
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 20 of 31
5. Use a tie wrap from the kit to re-bundle the cables.
6. Connect the LEMO Adapter cable BNC connector to the balun BNC connector.
7. From the kit, obtain the balun retainer clip (PN 20476-001) and remove the release liner
from the double-sided tape.
8. Slide the balun retainer clip over the balun. Do not let the clip stick to the tape.
9. With the BNC connector towards the front of the C-Arm, mount the balun to the inside
right wall of the C-Arm toe (as viewed from the rear of the mainframe) with the clip over
the outside of the toe wall. Press firmly to bond the adhesive tape to the wall.
Power
Balun
Supply
Note: Avoid dropping the screws into the circuit area below the control board.
2. Install the cover mounting bracket (PN 5477294).
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 21 of 31
3. Using Loctite 222 with the four (4) screws previously removed; re-install the cross arm
brake assembly.
3. Route the monitor cables through the top cut out of the control panel cover.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 22 of 31
Reference monitor
assembly
4. Make sure the reference monitor is oriented such that it can be fully rotated when facing
WKH;-ray switch, and then attach the reference monitor assembly to the control panel
housing using the four (4) 10-14 pan head screws (PN 20271-017) from the kit.
10-14 Pan
head screws
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 23 of 31
Cover
plate
Monitor
channels
7. Install the coiled cord strain relief to the slot on the cover and secure to the two (2) cable
anchors using 6-inch tie wraps.
Tie
wraps
8. Slide the entire reference monitor cover assembly into the mounting grooves on the left
side (as viewed from the rear of the mainframe).
Caution: Be very careful to secure the cover in the grooves so the assembly does not fall during
installation and get damaged.
9. Remove the four (4) screws holding the right control panel (as viewed from the rear of the
mainframe).
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 24 of 31
10. Lower the right control panel to access the control panel chassis. Connect the ground
wire from the reference monitor assembly to the control panel chassis at the lower right
corner with a 10-32 pan head screw (PN 20268-017) and star washer (PN 20292-005)
provided within the kit.
Ground
11. Reinstall the right control panel with the four (4) screws previously removed.
12. Connect the three (3) control panel cover internal cables (disconnected earlier).
13. Lower the reference monitor assembly onto the control panel chassis. Carefully dress all
cables inside the cover to avoid crimping, cutting, or short circuit.
Caution: The cables inside the cover are a tight fit. Dress them carefully.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 25 of 31
14. Secure the reference monitor assembly to the inside cover mounting bracket with four (4)
10-32 head cap screws (PN 20320-010) and #10 flat washers (PN 20296-005) provided in
the kit. 7LJKWHQWKHVFUHZVòWXUQSDVWVQXJ
Note: Check to make sure all balun connections and wires are tight and avoid high voltage leads.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 26 of 31
RJ45
Receptacle
J7
2. From the kit, obtain the Workstation reference monitor video cable (PN 5477288) and
connect it to J11 at the back of the Workstation.
3. Route the other end of the cable through the opening towards the right of the Ebox.
4. Route the cable along the top side of the Workstation to the opening at the top (just
below the keyboard) and down towards the bottom of the Workstation.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 27 of 31
5. From the kit, obtain shrink tubing (PN 20462-001).
6. Slide it over the cable and connect P7 to the new Workstation cable. Slide the heat shrink
tubing over the BNC connection.
7. Tie wrap each end of the heat shrink tubing as shown below. Route the cable neatly
through the cable clamps.
Cable
clamps
Connect P7 to new
video cable. Tie wrap
heat shrink tubing at
both ends of the
connection.
Tie
wraps
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 28 of 31
Test
WARNING Workstation covers are removed. Dangerous voltages are present when the covers
are removed. Take appropriate safety precautions.
1. With the covers still off, connect the interconnect cable and start up the system.
2. Verify that the system starts up with no errors, and that the reference monitor turns on
and displays the same image as the left monitor displays.
3. Verify that the control panel console keys, buttons, and control panel display function
properly.
4. Verify that the reference monitor swivels and pivots up and down smoothly.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 29 of 31
5. If needed, the swivel can be adjusted by loosening or tightening the set screw located at
the lower rear of the curved arm. Tilt can be adjusted by loosening or tightening the two
(2) socket head cap screws located at the upper rear of the curved arm. If the tilt screws
are adjusted, both screws should be adjusted the same amount.
Labels
1. Apply the cross-arm brake handle lock/unlock label.
2. On all systems, apply the English ;-ray warning label (PN 5476335-1EN) WRWKHULJKWRI;-
ray activation button as shown below.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 30 of 31
3. In addition, for non-(QJOLVKODQJXDJHV\VWHPVDSSO\WKH;-ray warning label (PN
5476335-;; in the language appropriate for the specific country on the curved arch
VXUIDFHEHORZWKH;-ray activation button as shown below. Match the appropriate
language to WKHODQJXDJHRIWKH;-ray warning label on the original cover removed from
the system.
Finish
1. Properly shut down the system and unplug it.
2. Re-install all covers.
Revision History
Rev # Date Description of Change
1 April 2013 Initial release
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5442124-1EN-01 1 31 of 31