Real EC [Real Exposure Control]

Auto mA adjustment for each region

Dose reduction
 Easy set of mA
   High quality, quantitative images
Real EC [Real Exposure Control]
● Automatic control about Tube current for every
rotation with Helical scanning, depending on the
x-ray absorption of Body.
・  150mA
● Calculation of X-ray absorption is used only ・  120mA
1-direction Scanogram. ・  100mA
・  120mA
・  150mA
● In order to uniform the Imaging Noise (SD) at all ・  200mA
slice position, tube current is controlled.

● The optimal scan conditions with Real-EC which


suppressed useless DOSE, and provide uniform
image quality.

● 3 type scan mode


   High Quality / Low Dose / Manual
Real EC [Real Exposure Control]
Acquire Scanogram (only 1-direction)

Scan range Setting

Mode selection

Auto calculation of tube current


with Scanogram

Confirm tube current

Helical scanning with Real-EC

Image reconstruction
Real EC [Real Exposure Control]
Real EC [Real Exposure Control]

250 50

200 40

管 電 流 (m A )

画 像 ノイズ
150 30
管電流
100 20 画像ノイズ

50 10

0 0
1 6 11 16 21
ヘリカルスキャン位置

Body phantom scanogram Relationship of Tube Current and SD value of


image. (Body phantom from lung to liver.)
About 40% total dose reduction is possible.