Professional Documents
Culture Documents
Babyleo TN500
WARNING IncuWarmer
To properly use this medical device, Software 1.0n
read and comply with these
instructions for use.
This page has been left blank intentionally.
Contents
1 Information about this document ........................................................... 6
1.1 Typographical conventions ............................................................. 6
1.2 Use of terms ................................................................................... 6
1.3 Illustrations...................................................................................... 6
1.4 Trademarks..................................................................................... 7
6 Operation .................................................................................................. 63
6.1 Checking for operational readiness ................................................ 63
6.2 Switching on the device .................................................................. 64
6.3 Switching off the device .................................................................. 67
6.4 Operating modes ............................................................................ 68
6.5 Incubator operation......................................................................... 69
6.6 Radiant warmer operation .............................................................. 71
6.7 Transition ........................................................................................ 74
6.8 Skin temperature mode .................................................................. 77
6.9 Air temperature mode ..................................................................... 85
6.10 Manual mode .................................................................................. 91
6.11 Humidity.......................................................................................... 94
6.12 Oxygen (option) .............................................................................. 96
6.13 Heated mattress (option) ................................................................ 100
6.14 Kangaroo mode .............................................................................. 104
6.15 Tolerate cooling (option) ................................................................. 109
6.16 Warm-up (option)............................................................................ 111
6.17 Weaning (option) ............................................................................ 113
6.18 Scale (option).................................................................................. 116
6.19 Taking x-rays .................................................................................. 122
6.20 Developmental care (option)........................................................... 124
7 Alarms....................................................................................................... 127
7.1 Alarm priorities and alarm signals................................................... 127
7.2 Suppressing and delaying the alarm signal .................................... 127
7.3 Dismissing the alarm message....................................................... 130
7.4 Viewing the alarm history................................................................ 130
9 Transfer..................................................................................................... 150
9.1 Intrahospital patient transfer with switched off warming therapy .... 150
10 Troubleshooting....................................................................................... 153
10.1 Alarm – Cause – Remedy............................................................... 153
10.2 Power failure alarm......................................................................... 161
10.3 Adjustment of device height not possible ....................................... 162
10.4 Adjustment of mattress tray not possible........................................ 162
10.5 Fault during weighing...................................................................... 162
10.6 Fault with the oxygen supply .......................................................... 163
10.7 Removing liquid from the heating plate .......................................... 164
Index.......................................................................................................... 215
1.3 Illustrations
Depending on the configuration, the products and screen content shown in this
document may differ from the actual on-site products.
1.4 Trademarks
Trademarks owned by Dräger
The trademarks are registered in the following countries:
Trademark Country
Germany, Australia, Switzerland, China, EU, Japan,
Babyleo® Korea, Mexico, New Zealand, Russia, Singapore, USA
Infinity® EU, USA, China, Brazil, Australia, India
IncuWarmer® EU
DrägerService® Germany, USA
MEDIBUS.X® EU, USA
® EU, USA
Oxycell
® EU, USA
KangarooMode
ThermoMonitoring® Germany, France, USA
2 Safety-related information
2.1 Intended use
The IncuWarmer Babyleo TN500 is intended for use with premature babies and
neonates and can be used as both an incubator and a radiant warmer. When the
product is switched between incubator and radiant warmer operation, patients
continue to be kept warm during the transition. The device provides a thermally
regulated environment for patients with a body weight of up to 5 kg (11 lbs) and a
height of up to 55 cm (22 in). The device can be operated as either a closed care
unit or an open care unit. As a closed care unit, Babyleo TN500 is an incubator.
Neonates are kept warm in the patient compartment with humidifiable air, which can
be enriched with oxygen (option). As an open care unit, Babyleo TN500 is a radiant
warmer.
2.2 Indications
The IncuWarmer Babyleo TN500 provides controlled ambient conditions for
premature babies and neonates. The following parameters are regulated, according
to the intended use:
– Temperature
– Humidity
– Oxygen (option)
2.3 Contraindications
Do not use Babyleo TN500 outside the specified environments of use.
2.6.1.1 Users
Users are clinical personnel such as doctors and nursing staff trained in
neonatology.
Activity Requirement
Use of the product in accordance with Specialist medical knowledge in neonatol-
the intended use ogy
Knowledge of currently known risks and
benefits of warming therapy
Activity Requirement
Reprocessing Specialist knowledge in the reprocessing of
medical devices
Activity Requirement
Installation Specialist knowledge in electrical engineer-
Basic and complex service work ing and mechanics
(inspection, maintenance, repair) Experience in the servicing of medical
devices
Experience in complex service work on this
product
The instructions for use do not contain any information on the following points:
– Risks that are obvious to users
– Consequences of obvious improper use of the product
– Potentially negative effects on patients with different underlying diseases
2.8.7 Alarms
Failure to notice alarm signals may put the patient at risk.
► Check that 2 acoustic signals sound and the alarm light shows yellow and red
when the device is switched on.
► Set the alarm volume so that alarm signals can be heard.
► Be aware that only high- and medium-priority alarms trigger the nurse call.
► Take note of alarms directly on the device. Be aware that using the nurse call
does not unconditionally guarantee that alarms will be transmitted.
► Make sure that the alarm system has not been rendered ineffective by setting
the alarm limits to extreme values or deactivated by switching off the alarms.
► Adjust the basic alarm settings on devices to the needs of the patient.
Electromagnetic fields may disrupt the function of the device, putting the patient at
risk. Electromagnetic fields are generated by, for example:
– Cellular phones
– Radio-frequency surgical equipment
– Defibrillators
– Shortwave therapy equipment
► Observe the separation distances. For further information, see the following
chapter: ''Recommended safety clearances for portable and mobile high-
frequency communication equipment'' (page 199).
28227
1
3
10 4
5
9
7
8
30230
Fig. 2 Radiant warmer protective grid
The radiant warmer may heat up objects in the beam path close to the radiant
warmer to such an extent that touching the objects causes burns. In addition, the
objects may ignite and cause a fire.
► Do not permanently locate objects (e.g., examination lights) in the beam path of
the radiant warmer.
► Allow the objects to cool down before touching, to avoid burns.
► Before pivoting accessories (e.g., examination lights) out of the beam path of the
radiant warmer, take precautions to prevent burns.
The warm air duct passes through the air ducts of the X-ray flap. The convective
heater directs the warm air through the air ducts of the X-ray flap into the patient
compartment.
30231
1
2.15.3 Service
This product must be serviced at regular intervals and in the appropriate manner.
Failure to perform service measures correctly may result in personal injury and
property damage.
► Perform service work as described in ''Service''.
► Service work must only be performed by the target groups authorized to perform
the work in question.
► Only carry out service measures when there is no patient in the device.
► Perform calibration, adjustment, and verification of the scale (option) at the
specified intervals.
2.16.2 Ventilation
The temperature measurement for breathing gas humidification may be distorted by
the radiant warmer or the incubator temperature. There is a risk of hypothermia if
the breathing gas is too cold and of suffocation due to condensate formation.
► Follow the instructions for use of the humidifier in conjunction with the
instructions for use of the breathing hose.
3 Product description
3.1 Structure and function
This chapter contains descriptions of the device components in table format.
Some components and functions are options. The actual device may differ in
appearance and functions from the descriptions below. For further information,
see the following chapter: ''Range of functions'' (page 41).
30273
Fig. 4 Incubator operation (hood closed) and radiant warmer operation (hood
open)
The user's operating location can be on the right side, on the left side, or at the
front of the device.
28227
1 2
15
4
14
5
13
12
11
7
8
9
10
27221
15 16 1
14 2
3
13
12
6
7
8
11
9
10
27220
1 2 3
5 4
6
7
11 8
10
27216
1
9
8
7
27215
1
8 7 6 5 4
The actual screen views may differ in appearance or configuration. The screen
views in these instructions for use show the maximum range of functions of the
device. The set values shown in the screen views are for illustration purposes
only. For further information, see the following chapter: ''Range of
functions'' (page 41)
28500
1
2
6 5
28477
1
3.2.2.1 Timer
The APGAR timer or stopwatch can be activated using the Timer... button in the
main menu bar.
When the APGAR timer has been started, a signal tone is emitted after one
minute, after 5 minutes, and after 10 minutes. The stopwatch counts up to
99:59 minutes.
Procedure:
1 Touch the Timer... button in the main menu bar.
The Timer dialog window is displayed.
2 Touch the APGAR or Stopwatch button.
The selected timer is displayed in the header bar.
3 To use a timer, proceed as follows:
a Touch the button next to the timer.
b Select Start, Stop, or Reset.
c To deactivate the timer, touch the Timer... button in the main menu bar.
d In the Timer dialog window, touch the timer that should be deactivated.
✓ The button turns light green. The timer disappears from the header bar.
28482
Fig. 12 Quick setup dialog window for the air temperature (example)
The quick setup dialog window for the alarm limits can be opened by touching the
following parameter fields in the monitoring area:
– Skin temperature
– Air temperature
– Oxygen (option)
– Noise level (option)
– Light level (option)
27240
1
5
6
7
8
13 12 11 10 9
Fig. 13 Standard view in incubator operation with air temperature mode activated
27253
1
4
5
6
8 7
Fig. 14 Standard view in radiant warmer operation with manual mode activated
3.2.6 Thermomonitoring
Thermomonitoring is the continuous measurement and display of the central and
peripheral skin temperature, even when skin temperature mode is not active.
27264
1
27247
1
4 3
3.2.8 Family
If the patient's parents are present at the device, the Family screen view can be
selected. Only the most important therapy settings are displayed in this screen
view.
The therapy values and settings cannot be changed in this screen view. To
change the screen view, press the rotary knob.
27250
3 2 1
3.3.1.1 Options
– Scale (standard or NAWI/OIML certified)
– Heated, x-ray transparent mattress
– Drawer
– Oxygen regulation in the patient compartment
– Up to 2 gas cylinder holders
– AutoThermo package for advanced therapy applications
– Tolerate cooling
– Warm-up
– Weaning
– Developmental care package
– Audio stimulation
– Measurement of light level and noise level in the patient compartment
The device is intended for use with the accessories indicated in the list of
accessories.
3.3.1.2 Monitoring
The device can monitor the following parameters and events (some are only
available if the corresponding option was selected):
– Thermomonitoring: skin temperature
– Air temperature
– Mattress core temperature (option)
– Relative humidity
– O2 concentration in the patient compartment (option)
– Noise level in the patient compartment (option)
– Light level in the patient compartment (option)
– Patient weight (option)
– Convection heater power
– Radiant warmer power
– Occurrence of Hood open and Access panel open events
– Kangaroo mode
3.3.1.4 Protocoling
The device can save the following data, among other things, in a logbook:
– Measured values
– Set values and changes to values
– Patient data
– Events (e.g., alarms, confirmed alarms)
– Test results
USB port
The USB port enables the following actions to be carried out once a USB mass
storage device (e.g., USB flash drive from Dräger) has been connected:
– Saving and loading device configurations
– Saving and loading patient configurations
Service port
Specialized service personnel can connect the device to a DrägerService computer.
The following are used:
– Device status queries
– Support of software installation
Alarms
Setting alarm limits for modes and therapy controls Page 63
Suppressing the acoustic alarm signal Page 127
Dismissing the alarm message Page 130
4 Operating concept
4.1 Display unit
This chapter describes how to change the therapy settings using the display unit.
The chapter also explains how the device supports the user in operating the device
by means of visual and textual indicators (color signals, information texts, alarm
texts).
30607
1
30608
1
2
3
30607
4 3 2 1
WARNING
Risk of electric shock
If height adjustment is actuated and the power cable is trapped or wrapped around
a castor, the power cable may tear.
► When positioning the device, ensure that the power cable is not trapped.
► Be aware of the length of the power cable when operating the height
adjustment.
► During intrahospital transport, do not push the device over the power cable.
WARNING
Risk of injury by collision with compressed gas cylinders
When height adjustment is actuated, accessories on the duct rail may collide with
the gas cylinders.
► Actuate height adjustment with caution and stop it promptly.
WARNING
Risk to patient if hoses and cables are too short
Hoses and cables may be subjected to tensile loading when opening the hood,
moving the device, or actuating the device height adjustment. This risk applies
especially if the hoses and cables are routed into the patient compartment across
the hood.
► Choose hoses and cables that are long enough to perform the above
movements safely.
► Do not route hoses and cables across the hood; always pass them into the
patient compartment via the hose grommets.
► Pay attention to the hoses and cables when carrying out the above movements.
WARNING
Unintentional disconnection of hose and cable connections
Possible interruption to patient treatment due to uncontrolled movement of the
device via height adjustment or the trolley.
► When using the height adjustment mechanism and the trolley, always ensure
that no hoses and cables become disconnected.
► Lock the double castors on the trolley with the locking brakes.
CAUTION
Risk of injury due to collision when using height adjustment
The radiant warmer or accessories may collide with obstacles when the device
height is raised. The patient may be injured by the jolting. The radiant warmer and
the accessories may be damaged.
► Make sure there is sufficient space above the radiant warmer and the installed
accessories.
► Stop the device promptly ahead of any possible obstacle.
Procedure:
1 Adjust the device height using the foot pedals:
– The right-hand foot pedal (1) moves the device up.
– The left-hand foot pedal (2) moves the device down.
28263
2 1
WARNING
Risk of injury when adjusting the mattress tray
If the mattress tray is adjusted without due care, parts of the body may become
trapped. If an external ventilator is being used, the patient may be accidentally
extubated.
► When lowering or tilting the mattress tray, make sure that no part of the patient's
body or user's body is within the hazard areas.
► The height adjustment and bed-tilt mechanism must only be operated by a
single user.
► If the mattress tray is pulled out, operate the device height adjustment with
caution and stop it promptly ahead of an obstacle.
► Ensure that hoses of a sufficient length are used when using an external
ventilator.
CAUTION
Uneven heat distribution
If the mattress tray is tilted, heat may be unevenly distributed on the mattress tray.
► If the mattress tray is tilted, make sure that the patient is adequately warmed.
Procedure:
1 Adjust the mattress tray height with the upper buttons:
– The right-hand key (1) raises the mattress tray.
– The left-hand key (2) lowers the mattress tray.
28262
1
2
28262
3
4
When the maximum mattress tray tilt is reached, the LED in the button (3 or
4) goes out.
✓ The height and tilt angle of the mattress tray are set.
4.3.1 Mattress
The heated mattress or the SoftBed mattress can be placed on the mattress tray.
Procedure:
1 Place the mattress on the mattress tray.
2 Cover the mattress with a sheet.
For further information, see the following chapter: ''Connecting the heated
mattress'' (page 101)
4.4 Hood
This chapter describes how to open and close the hood.
28260
CAUTION
Risk of injury due to unexpected opening or closing of the hood
Opening or closing the hood unexpectedly may cause head or other injuries to
people nearby.
► Make sure that no-one else is using the device at the same time and that there is
no-one else in the vicinity of the hood.
CAUTION
Risk from falling objects
Objects placed on the hood may fall off, injuring the patient.
► Remove all objects before opening the hood.
Procedure:
1 To open the hood, take hold of the handle at the side of the hood and lift it
forwards and up.
The hood moves up.
2 To close the hood, pull the handle at the side of the hood downwards.
The device closes the hood automatically at the end.
28257
1
2
When the hand ports or access panels are open, a warm air curtain remains in
place, preventing the patient compartment from cooling down.
WARNING
Risk of falling out
If the access panels or the hand ports are open or the mattress tray has been pulled
out, there is a risk of the patient falling out of the device.
► Check that the access panels and hand ports are securely locked.
► Monitor the patient when the access panels or the hand ports are open or the
mattress tray is pulled out.
Procedure:
Operating the hand ports
1 To open the hand port, push down the ridged part of the lock (1) with the hand or
elbow.
The hand port opens.
2 To close the hand port, allow the lock (1) to audibly engage.
The hand port is closed.
28257
1
30843
1
30326
1
3 2
5
1 2
9
8 6
3
7
4
Additional information:
The following loading limits apply if only 1 pivot arm is fitted to the device:
Duct rail on the side of the hood arm:
Maximum load Arm length
15 kg (33.1 lbs) max. 300 mm (11.81 in)
Or:
Maximum load Arm length
25 kg (55.1 lbs) max. 250 mm (9.84 in)
The following loading limits apply if a total of 2 pivot arms are fitted to the device:
– Either 15 kg (33.1 lbs) or 12 kg (26.5 lbs) can be fitted to each duct rail.
– The user can decide which weight is fitted to which side of the device.
WARNING
Risk of electric shock
If the interface connectors and the patient are touched at the same time, there is a
risk of an electric shock.
► Do not simultaneously touch the nurse call connector and the patient.
CAUTION
No alarm in event of a nurse call fault
The nurse call may fail in the event of an incorrect device configuration, operating
errors (e.g., nurse call connector incorrectly connected) or a fault in the connection
between the nurse call and the central hospital alarm system.Alarms are not
transmitted.
► Be aware that only high- and medium-priority alarms trigger the nurse call.
► Bear in mind that using the nurse call does not absolutely guarantee that alarms
will be transmitted.
► Take note of alarms directly on the device.
Procedure:
1 Push the nurse call connector into the nurse call socket so that it engages
audibly.
30229
Fig. 29 Correct side of device for oxygen cylinders
WARNING
Risk of explosion and fire
If oxygen cylinders or devices containing oxygen are installed in the vicinity of the
ventilation slot, oxygen may get into the device at the power supply unit. In the
event of sparking at the power supply unit there is an increased risk of fire, which
may lead to personal injury and property damage.
► Position oxygen cylinders and devices containing oxygen on that side of the
device without the ventilation slot.
WARNING
Risk to patient if unauthorized gases are used
The use of compressed gases that are not approved for medical purposes may
adversely affect the function of the device and put the patient's health at risk.
► Only use compressed gases that are approved for medical use.
► Only use compressed gases that are dry and free from dust and oil.
WARNING
Increased risk of fire from unauthorized oxygen cylinder valves
The use of unauthorized or poorly maintained oxygen cylinder valves presents an
increased risk of fire, which may lead to personal injury and property damage.
► Do not use unauthorized oxygen cylinder valves.
► Make sure that oxygen connectors and oxygen seals are kept free from oil and
dust at all times.
► Always open oxygen cylinder valves slowly.
WARNING
Risk of crushing if accessories are mounted at too low a height
If compressed gas cylinders are present and accessories are mounted too low
down on the duct rail, hands or fingers may be crushed between the duct rail and
the compressed gas cylinders.
► If compressed gas cylinders are present, do not mount accessories too low
down.
► When lowering the height adjustment, do not touch the compressed gas
cylinders and the accessory arm.
Before installation, set the lifting column to its lowest position. Mount
accessories with the lifting column in its lowest position. Do not mount
accessories too low down.
Procedure:
1 Adjust the bottom of the gas cylinder holder (3) to the height of the oxygen
cylinder. The height of the gas cylinder holder can be set to 2 levels.
2 Ensure that the pressure reducer is pointing towards the rear and protrudes over
the upper edge of the gas cylinder holder.
3 Adjust the height of the protection bar (2) to the size of the oxygen cylinder and
the position of the pressure reducer.
30227
1
30228
1
The gas cylinder holder can also be used for compressed air cylinders.
Compressed air cylinders can be attached to both sides of the device. Further
information about gas cylinders can be found in the technical data.
WARNING
Risk of electric shock
If the interface connectors and the patient are touched at the same time, there is a
risk of an electric shock.
► Do not simultaneously touch the COM port and the patient.
CAUTION
Risk to patient due to incorrect data transfer
All data transferred via the COM port are for information only and must not be used
as the sole basis for diagnostic or therapeutic decisions. This may put the patient at
risk.
► Do not use data transferred via the COM port as the sole source of information.
Procedure:
1 Connect an external device to the COM port.
2 Configure the COM port.
6 Operation
6.1 Checking for operational readiness
This chapter describes how to check the device for operational readiness. For
further information, see the following chapter: ''Automatic selftest'' (page 202)
Prerequisites:
– The device has been acclimatized in the environment of use for 24 hours prior to
use.
– The device has been reprocessed and assembled ready for operation.
– The mains plug is plugged in.
WARNING
Risk to patient due to undetected damage to the device
Reprocessing, wear and tear, and improper storage may damage parts of the
device such that they no longer function correctly.
► Comply with the ambient conditions for operation and storage of the device.
► Before operating the device, always carry out the test steps listed in the
"Checking for operational readiness" chapter.
► Only operate the device when all test steps and the automatic selftest have been
completed successfully.
Position the device so that the mains plug is readily accessible at all times. The
device can only be safely disconnected from the mains power supply by
unplugging the mains plug.
Procedure:
Always perform the following test steps before operating the device:
1 Check that all parts of the device are free from cracks and sharp edges,
especially the hood, the access panels, and the mattress tray.
2 Check that all cables and hoses are free from damage and are routed correctly.
3 Check that the heated mattress (option) is free from damage.
4 Check that all access panels and hand ports are free from damage and close
securely.
5 Check that 2 acoustic signals sound and the alarm light shows yellow and red
when the device is switched on.
6 After switching the device on, check that the touch screen and rotary knob are
working correctly.
7 Check whether the O2 sensors need to be adjusted. For further information, see
the following chapter: ''Adjusting the O2 sensors'' (page 99)
✓ Only operate the device when all test steps and the automatic selftest have been
completed successfully.
The device carries out an automatic selftest when it is switched on. If the device
detects a fault, a corresponding message or an alarm is displayed. The
automatic self-test checks, e.g., the functional integrity of the primary and
secondary alarm systems and the power supply of the device. For further
information, see the following chapter: ''Automatic selftest'' (page 202)
Procedure:
1 Plug in the mains plug.
2 Switch on the device at the main switch on the rear of the main column (position
I).
The start process commences and lasts about 7 seconds.
3 Switch on the device on the display unit using the Start/standby key (1).
The device carries out an automatic selftest.
4 Check the following points shortly after switching on the device:
– An acoustic signal sounds (power failure alarm).
– The power supply indicator next to the Start/standby key lights up red.
– Another acoustic signal sounds (alarm system).
– The alarm light lights yellow and red alternately.
27215
If the device was switched off for less than 2 hours, the following dialog window is
displayed:
28478
Fig. 33 New patient dialog window
– By touching the New patient New patient button, all existing patient data are
deleted.
– By touching the Current patient button, the device continues operation with all
existing patient data and the most recent settings.
Skin temperature mode is not suitable for warming up the device during
incubator and radiant warmer operation.
WARNING
Risk due to incorrect settings
Devices within an environment of use can be configured differently. User-specific
alarm limits or therapy settings may be set on the devices. If the settings are
unsuitable for the patient, the patient may be put at risk.
► Ensure that the set therapy values are suitable for the new patient.
► Make sure that the alarm system has not been rendered ineffective by setting
the alarm limits to extreme values or deactivated by switching off the alarms.
► Check the basic alarm settings and the patient-specific alarm settings when a
new therapy is started.
CAUTION
Risk of hypothermia
If the device has not been warmed up, the patient may become hypothermic.
► Wait for the warm-up time of the radiant warmer.
► In incubator operation, wait for a warm-up time of 30 minutes.
Procedure:
1 Select the desired mode in the therapy bar.
2 Set the desired temperature with the therapy control.
3 Set additional parameters (e.g., humidity).
✓ The therapy settings have been adjusted.
31326
%DE\OHR71
,QFXEDWRURSHUDWLRQ 5DGLDQWZDUPHURSHUDWLRQ
0RGHV
6NLQWHPSHUDWXUHPRGH 6NLQWHPSHUDWXUHPRGH
.DQJDURRPRGH .DQJDURRPRGH
$LUWHPSHUDWXUHPRGH
0DQXDOPRGH
6HWWLQJV
+HDWHGPDWWUHVV +HDWHGPDWWUHVV
2[\JHQ
+XPLGLW\
6NLQWHPSHUDWXUH 6NLQWHPSHUDWXUH
$LUWHPSHUDWXUH
5DGLDQWZDUPHUSRZHU
$SSOLFDWLRQV
:HLJKLQJ :HLJKLQJ
;UD\ ;UD\
:DUPXS :DUPXS
7ROHUDWHFRROLQJ 7ROHUDWHFRROLQJ
:HDQLQJ
27240
1
3 2
Fig. 35 Standard view in incubator operation with air temperature mode activated
The Standard function or the Eco function can be selected for the transition
between incubator operation and radiant warmer operation. For further
information, see the following chapter: ''Configuring the transition'' (page 76).
The Standard function automatically activates the ClearView on setting. The
ClearView on setting can only be deactivated when the Eco function is
selected.
In incubator operation set the lowest position for the mattress tray so that there
is as much space as possible for the patient. This position also offers good
access to the patient.
27253
1
Fig. 36 Standard view in radiant warmer operation with manual mode activated
The patient's thermal equilibrium may be influenced by the ambient conditions, e.g.,
by drafts.
In radiant warmer operation Dräger recommends setting the highest position for
the mattress tray. This position provides the most effective warming for the
patient. This position also offers good access to the patient.
The radiant warmer starts up with 100 % power. Within a period of 14 minutes and
30 seconds the power is gradually reduced. In pre-warm mode the therapy control
displays either Pre or a value, depending on the radiant warmer power.
28481
1
6.7 Transition
This chapter describes the transition between incubator operation and radiant
warmer operation.
WARNING
Risk to patient due to incorrect temperature settings
When switching between incubator operation and radiant warmer operation, the
temperature setting may subsequently be too high or too low.
► After opening or closing the hood, always check the temperature settings.
► After the transition, adjust the temperature settings to the patient's needs.
Procedure:
1 Touch the System setup... button in the main menu bar.
2 Touch the Therapy tab.
3 Enter the password and confirm with OK.
4 Select a button in the Thermoregulation tab:
Button Meaning
Standard – In incubator operation the radiant warmer operates contin-
uously at low power.
– In radiant warmer operation the convection heater oper-
ates continuously at low power.
This ensures optimum warming during and after the tran-
sition.
Eco – In incubator operation the radiant warmer is switched off.
– In radiant warmer operation the convection heater is
switched off.
This saves energy.
To set the temperature in the extended range, first confirm the extended range
by pressing the rotary knob.
To open the quick setup dialog window, touch a skin temperature value. The
buttons for the alarm limits are displayed.
27245
1 2 3 4 5 6
30327
1 2 3
The following table shows the differences between the skin temperature sensors:
Symbols on Sensor color Temperature measure- Attachment position
the device ment
Yellow Central skin temperature In the liver or kidney
region
White Peripheral skin tempera- Foot or hand
ture
Prerequisites:
– The device is in operation.
– The patient is inside the device.
– The yellow and white skin temperature sensors and the corresponding covers
are ready to use.
– The area of skin where the sensor is to be placed is clean and dry.
WARNING
Risk of electric shock
If the contact pins are touched, the user and the patient will be exposed to the risk
of an electric shock.
► Do not touch the pins in the skin temperature sensor plugs.
Procedure:
1 Push the yellow skin temperature sensor into the yellow socket (1) on the rear
sensor wall.
The Skin mode tab in the therapy bar is available.
2 Push the white skin temperature sensor into the white socket (2) on the rear
sensor wall.
26457
1
3 Pass the sensor cables into the device through a hose grommet.
WARNING
Risk of hypothermia or hyperthermia
If the attachment positions of the skin temperature sensors have been swapped
accidentally or the skin temperature sensors are not attached correctly to the
patient, the patient may become too hot or too cold.
► Place the yellow skin temperature sensor in the liver or kidney region only.
► Place the white skin temperature sensor on the foot or hand only.
► Do not place the skin temperature sensors under the patient.
► Check at regular intervals that the skin temperature sensors are correctly
attached.
WARNING
Risk of hypothermia or hyperthermia
If the skin temperature sensors are wet, the patient may become too hot or too cold.
► Use dry skin temperature sensors only.
WARNING
Risk to patient due to incorrect use of the skin temperature sensors
If the rectal temperature is measured with the skin temperature sensors, the device
will display incorrect values or will regulate the temperature on the basis of incorrect
values.
► Only use the yellow skin temperature sensor for measuring the central skin
temperature.
► Only use the white skin temperature sensor for measuring the peripheral skin
temperature.
Procedure:
1 Attach the yellow skin temperature sensor for measuring the central skin
temperature in one of the following positions on the patient:
– Supine position: using a sensor cover, attach the sensor to the abdomen in
the liver region.
– Prone position: using a sensor cover, attach the sensor to the back in the
kidney region.
2 Attach the white skin temperature sensor for measuring the peripheral skin
temperature to the patient's foot or hand using a sensor cover.
WARNING
Risk of hypothermia or hyperthermia
If skin temperature mode is used for patients with a fever or in a state of shock, the
patient may become too hot or too cold.
► Do not use skin temperature mode in cases of fever or shock.
Procedure:
1 Touch the Skin mode tab (1) in the therapy bar.
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
27245
1 2
6.8.5 Thermomonitoring
If 2 skin temperature sensors are used, the Thermomonitoring view shows the
continuous measurement and display of a central and a peripheral temperature.
Thermomonitoring can be used for the early detection of cold stress, heat stress, or
infection. Thermomonitoring means the continuous monitoring of the central skin
temperature and the peripheral skin temperature. The difference between the two
temperature values is displayed. The difference can indicate a change in the
temperature of the patient. The diagram below shows thermomonitoring in a patient
with cold stress: the peripheral skin temperature (2) drops before the central skin
temperature (1) and can be the first indication of cold stress.
30262
1
No. Designation
1 Central skin temperature
2 Peripheral skin temperature
28483
Fig. 43 Additional skin temperature alarm limits
6 Set and confirm the alarm limits for the difference between the central and the
peripheral skin temperature.
✓ The advanced skin temperature alarm limits are set.
If one of the advanced skin temperature alarm limits is switched off, the following
symbol is displayed on the main screen:
To set the temperature in the extended range, first confirm the extended range
by pressing the rotary knob.
The target air temperature must be at least 3 °C (5.4 °F) above the ambient
temperature in order for the device to reach the desired temperature.
In air temperature mode the patient's skin temperature can be monitored with skin
temperature sensors.
27240
1
2 3 4 5 6
WARNING
Risk of hyperthermia due to external heat sources
The temperature in the patient compartment may rise to dangerous levels due to
direct sunlight or other external heat sources (e.g., phototherapy devices). The
patient may become overheated.
► When additional heat sources (e.g., solar radiation, examination lights, or
phototherapy lights) are active, make sure that the temperature in the patient
compartment remains non-critical.
► Check the patient's body core temperature at regular intervals.
WARNING
Risk to patient due to incorrect settings
If the air temperature and humidity settings are unsuitable for the patient, the patient
may be put at risk.
► Check the suggested air temperature and humidity settings carefully and adjust
them to the patient's needs if necessary.
Procedure:
1 Touch the Air mode tab (1) in the therapy bar.
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
The mode is activated.
27240
2 1
28473
Procedure:
1 Set and confirm the upper and lower alarm limit for the central skin temperature.
2 Set and confirm the upper and lower alarm limit for the peripheral skin
temperature.
3 Set and confirm the upper and lower alarm limit for the difference between the
central and the peripheral skin temperature.
✓ The alarm limits for skin temperature monitoring are set.
28473
2
1
The automatic alarm limits relate in each case to the current skin temperature
measurement or to the current difference.
27253
1
3 2
WARNING
Risk of hyperthermia in manual mode
If the radiant warmer power is set too high or additional heat sources are active
(e.g., phototherapy devices or sunlight), the patient may become overheated.
► In manual mode check the patient's body core temperature at regular intervals.
► Only supply the patient with as much thermal radiation as is necessary.
► Do not leave the patient unattended during radiant warmer operation.
WARNING
Risk of fire underneath the radiant warmer
Objects in the beam path of the radiant warmer may ignite, putting people and
property at risk.
► Do not place combustible materials under the radiant warmer.
► Do not place objects in the beam path of the radiant warmer.
CAUTION
Risk to patient due to increased water loss when exposed to thermal radiation
The thermal radiation may cause increased water loss in the patient.
► Only expose the patient to the necessary amount of thermal radiation and only in
a controlled manner.
Procedure:
1 Touch the Manual mode tab (1).
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
27253
2 1
If the radiant warmer power is set above 30 %, a timer starts. The device
switches the radiant warmer off after 15 minutes.
6.11 Humidity
This chapter describes how to humidify the air in incubator operation. It also
describes how to attach an external water supply to the device.
Manual Automatic
The user sets the relative humidity level The device sets the relative humidity
manually according to the patient's automatically according to the air tem-
needs. perature.
Example: If the air temperature is
increased, the relative humidity is
increased correspondingly.
WARNING
Risk to patient due to germs in the humidification system
If the humidification system is filled with water, germs may form and contaminate
the breathing gas (e.g., with exogenous pyrogens).
► Only fill the water reservoir of the humidification system with Aquadest/distilled
water.
WARNING
Risk to patient due to unsuitable liquids in the humidification system
Undistilled water, water containing additives, and infusion solutions damage the
humidification system and may endanger the patient.
► Only fill the water reservoir of the humidification system with Aquadest/distilled
water.
► Do not introduce infusion solutions into the humidification system.
Procedure:
1 Disinfect your hands.
2 Attach the water bag or water bottle to the water bag holder on the rear of the
main column. The water bag or water bottle must be attached above the Luer
Lock connector. For further information, see the following chapter: ''Connectors
on the rear sensor wall'' (page 30)
3 Close the clamp on the connection hose.
4 Attach the connection hose to the water bag or water bottle.
5 Connect the connection hose to the Luer Lock connector on the device.
6 Open the clamp on the connection hose.
✓ The humidification system is ready for use.
WARNING
Risk to patient due to incorrect settings
If the air temperature and humidity settings are unsuitable for the patient, the patient
may be put at risk.
► Check the suggested air temperature and humidity settings carefully and adjust
them to the patient's needs if necessary.
If a high humidity level is set and the ambient temperature is low, condensation
may form next to the device. This may cause the user or a third-party to slip
over.
Procedure:
1 Touch the Humidity therapy control (1) in the therapy bar.
The therapy control turns yellow.
27240
1
To set the oxygen supply in the extended range, first confirm the extended range
by pressing the rotary knob.
WARNING
Risk due to increased external O2 supply
If O2 is supplied from an external source (e.g., via the central supply unit, an oxygen
cylinder, or an O2 hood), the O2 concentration may exceed 65 %. The health of the
patient may be endangered and there is an increased risk of fire in the device.
► Measure and monitor the oxygen concentration in the patient compartment.
► Make sure that the oxygen concentration in the patient compartment is always
below 65 %.
► Always monitor the oxygen concentration in the patient's blood, e.g., using blood
gas analysis and a pulse oxymeter.
► Always monitor the external O2 supply.
► Be aware that Babyleo TN500 displays no alarms in the case of an external
oxygen supply, e.g., via an O2 hood.
WARNING
Risk of fire if unauthorized electrical devices are used
If the air in the patient compartment is enriched with oxygen, there is an increased
risk of fire if unauthorized electrical devices are used.
► In the patient compartment use only electrical devices that are approved for use
in an oxygen-enriched atmosphere.
CAUTION
Risk to patient due to background noise
The supply of oxygen may cause additional background noise.
► Protect the patient from background noise.
Procedure:
1 Touch the Oxygen therapy control (1) in the therapy bar.
The therapy control turns yellow.
27240
1
32774
2
3
5 If oxygen has been used, air the device thoroughly: Open the hood and the side
access panels for at least 3 minutes. Close them again after the airing.
The Start button changes from gray to green when the following conditions
are met:
– The warm-up time has elapsed.
– The Oxygen therapy control is at Off.
– The device has been adequately aired.
If the Start button is displayed in gray, repeat steps 3 to 5.
After the O2 sensors have been adjusted, the date of the next adjustment is
displayed in the Oxygen dialog window. An alarm message will be displayed if
the O2 sensors have not been adjusted by this date. Additionally, the oxygen
measurement on the main screen is shown in gray. A gray measurement
indicates that the measurement accuracy is no longer guaranteed.
The main screen of the user interface displays the mattress core temperature. The
mattress core temperature is measured inside the mattress. The mattress core
temperature is the mean of 2 temperatures measured at the edges of the heating
element. The mattress core temperature can also be influenced by other heat
sources.
The heated mattress is x-ray transparent. Therefore, x-rays can be taken while
using the heated mattress. In this case, the x-ray cassette is placed in the x-ray
tray.
WARNING
Risk of patient injury if the heated mattress is used as the only source of heat
The heated mattress is not sufficient for use as the only source of heat, especially in
the case of hypothermic patients.
► Provide the patient with the heated mattress and an additional heat source.
► Monitor the body core temperature of the patient with a thermometer.
WARNING
Risk of patient injury due to damaged heated mattress
If the cover of the heated mattress is damaged, fluid may penetrate into the
mattress. This may damage the electronic components in the heated mattress. The
temperature sensors of the heated mattress may be damaged.
► Check the heated mattress for damage before use.
► Replace any damaged heated mattresses.
WARNING
Risk of electric shock
If the heated mattress is connected when radio-frequency surgical equipment,
endocardial catheters, or defibrillators are used, this may endanger the patient.
► If the heated mattress is connected, do not use radio-frequency surgical
equipment, endocardial catheters, or defibrillators.
WARNING
Risk of electric shock
If the contact pins are touched, the user and the patient will be exposed to the risk
of an electric shock.
► Do not touch the pins in the plug of the heated mattress.
Procedure:
1 Place the heated mattress on the mattress tray. Ensure the following:
– The heating side of the heated mattress points towards the patient. The
product label on the heated mattress shows the correct side.
– The heated mattress cable is facing the sensor wall.
2 Pull the cable through the hose grommet in the center of the sensor wall and
push it into the cable guide.
3 Make sure that the cable runs downward between the bed support and the
sensor wall and is not caught.
4 Connect the orange plug of the heated mattress to the orange socket (1) on the
rear of the sensor wall.
28322
1
27240
1
Fig. 55 Therapy control for the heated mattress
2 Adjust the mattress core temperature manually using the rotary knob.
or
3 Turn the rotary knob to the right as far as it will go and set the Auto function.
The device sets the mattress core temperature automatically according to the
air temperature.
4 Press the rotary knob to confirm.
The therapy control turns dark green.
✓ The mattress core temperature is set.
27244
1 2 3
Procedure:
1 Touch the Kangaroo mode tab (1).
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
The duration of kangaroo mode is continuously displayed on the main
screen.
27244
1
6 Close the hood or the side access panel. Make sure that the hoses or cables do
not get caught.
7 Check the alarm limits for skin temperature and adjust them if necessary.
28475
1
27244
1
28476
6.15.2 Starting "Tolerate cooling"
The chapter describes how to start the "Tolerate cooling" function. If a cooling
device is used, interrupt the warming therapy with the "Tolerate cooling" function.
Prerequisites:
– The device is in operation.
– The patient is inside the device.
Procedure:
1 Touch the Special procedures... button (1) in the main menu bar.
The Special procedures dialog window is displayed.
28486
3 2
28485
1
Procedure:
1 Touch the Autoset all limits button.
2 Press the rotary knob to confirm.
✓ The following alarm limits are set automatically for the current skin temperature
value:
28490
6.16.2 Starting warm-up
This chapter describes how to start warm-up.
Prerequisites:
– The device is in operation.
– At least the yellow skin temperature sensor for measuring the central skin
temperature is connected. For further information, see the following chapter:
''Connecting the skin temperature sensors'' (page 79).
Procedure:
1 Touch the Special procedures... button (1) in the main menu bar.
The Special procedures dialog window is displayed.
28489
1
When ending warm-up, the target value defined during warm-up is adopted as
set value for the skin temperature.
28492
6.17.2 Starting weaning
This chapter describes how to start automatic weaning.
Prerequisites:
– The patient is inside the device.
– The device is in incubator operation.
– At least the yellow skin temperature sensor for measuring the central skin
temperature is connected. For further information, see the following chapter:
''Connecting the skin temperature sensors'' (page 79)
Automatic weaning can only be started when the start temperature is less than
38 °C (100.4 °F) and does not lie within the extended temperature range.
Procedure:
1 Touch the Special procedures... button (1) in the main menu bar.
The Special procedures dialog window is displayed.
28491
5 Set and confirm the interval for the selected temperature step with the rotary
knob.
6 Set and confirm the target value for the central skin temperature with the rotary
knob.
7 Set and confirm the value for the tolerated deviation from the central skin
temperature with the rotary knob.
8 Touch the On button.
9 Press the rotary knob to confirm.
The On button and the previously set buttons change to dark green.
✓ The device switches to air temperature mode. When the target value for the air
temperature is reached, a message appears.
28493
1
4 3
Offers access to a table of the last weight measurement with the following
information: date and time, weight and manually adjusted weight. This button allows
to switch between the graphical view and the tabular view.
Only one weight measurement per day is saved in the graphical view and the
tabular view. Only the last measured value is saved.
WARNING
Risk of electric shock
If the contact pins are touched, the user and the patient will be exposed to the risk
of an electric shock.
► Do not touch the pins in the plug of the scale.
Procedure:
1 Make sure that the scale cable faces the sensor wall.
2 Place the bed support with the scale on the T-bars from above.
3 Pull the cable through the hose grommet in the center of the sensor wall and
push it into the cable guide.
4 Make sure that the cable runs downward between the bed support and the
sensor wall and is not caught.
5 Insert the gray plug for the scale into the gray socket on the rear of the sensor
wall.
✓ The scale is connected to the device.
6.18.3 Weighing
This chapter describes how to weigh the patient in the device.
Prerequisites:
– The device is in operation and the scale is connected.
– The bed support is resting securely on the T-bars and is not tilted to the side.
– The locking brakes are engaged.
– A mattress is placed on the mattress tray.
– The patient is lying in the center of the mattress.
WARNING
Risk due to measuring error or measuring inaccuracy
If the mattress tray is not horizontal during weighing or if the weight is falsified by
hoses or cables, measuring errors or inaccuracies may occur.This may result in
incorrect therapy decisions.
► Always position the mattress tray horizontally before weighing.
► Before weighing, make sure that the air bubble in the spirit level is within the
black ring and is not touching the ring.
► Make sure that cables (such as the heated mattress cable) are correctly routed
and do not falsify the weighing result.
► If a weighing result seems implausible, cross check it with a known weight value.
► Double-check therapeutic decisions that are based on the patient's weight by
performing a reference measurement using an external scale.
WARNING
Measuring error due to incorrect taring weight
If a taring weight if not determined before weighing, measuring errors may occur.
► Tare the scale before each weighing process.
Procedure:
1 Touch the Scale button (1) in the main menu bar.
The Scale dialog window is displayed.
28493
1
30295
Fig. 66 Note the spirit level
3 Proceed as follows in order to tare the scale:
– Keep as many hoses and cables that lead to the patient as possible off the
mattress tray.
– Fasten the hoses and cables in the holders.
– Pull the hoses and cables tautly through the hose grommets.
This prevents hoses and cables from jamming the mattress tray or falsifying
the weight measurement.
4 Touch the Weigh patient button (2).
5 To tare the scale, lift the patient up. At least 200 g (0.44 lbs) must be lifted off the
mattress tray for the scale to be tared. Wait until a signal sounds.
The scale is tared.
27266
6.18.4 Reweighing
If the mattress tray is not adjusted or moved for 6 minutes after the previous
weighing, the patient can be reweighed without needing to tare the scale again. The
Scale dialog window must remain open.
Example: The patient's weight needs to be determined with a diaper and then
without a diaper.
Prerequisites:
– The patient was weighed earlier and the Scale dialog window remains open.
– The mattress tray has not been adjusted or moved for 6 minutes.
WARNING
Risk if taring is not performed
When changes are made between measurements, measuring errors may occur if
the device is not tared, e.g., if the mattress tray is moved or adjusted or additional
items are placed on the mattress tray.
► Before carrying out another weight measurement, lift the patient up and tare the
scale.
Procedure:
1 To weigh the patient again, touch the Reweigh (1) button.
The measuring result is displayed.
The weight is also displayed as a trend.
30328
27254
1
For digital x-ray detectors with larger formats, remove the x-ray tray and insert the
digital x-ray detector in the compartment.
6.19.2 X-ray
The chapter describes how to take an x-ray of the patient in the device.
Prior to taking the x-ray, remove all positioning aids from the imaging area. The
positioning aids may absorb x-rays and adversely affect the image quality.
Prerequisites:
– The device is in operation.
– The patient is inside the device.
– A mobile x-ray device and an image receiver (x-ray cassette or x-ray imaging
plate) are available.
– The regulations for taking x-rays in incubators are known.
– The weight limit for the x-ray cassette has been taken into account.
– Bring the mattress tray into the bottom position and ensure it is horizontal.
WARNING
Risk of tipping over if overloaded
If compartments, x-ray trays, drawers, and supports are overloaded, the device may
tip over. This may result in injury to the patient and user. The device and
accessories may also be damaged due to the tip over.
► Do not overload the open x-ray tray.
► Observe the maximum loading of all shelves, drawers and holding devices.
NOTICE
Risk of x-ray detector failure
If the ambient conditions (temperature and relative humidity) in the device are
unsuitable for a digital x-ray detector, the x-ray detector may provide incorrect
results.
► Observe the permitted temperature range for the digital x-ray detector.
Procedure:
1 Open the flap for the x-ray tray on one side of the device. Pull the x-ray tray (1)
out.
30274
1
NOTICE
Risk of impact with x-ray tube assembly
If the clearance above the device is too small, the x-ray tube assembly may hit
the hood of the device.
► Before aligning the x-ray tube assembly, make sure that there is enough
room for maneuver above the device.
4 Position the x-ray tube assembly between the radiant warmer and the patient
compartment.
5 Use the light field of the x-ray device to set the aperture to determine the desired
image section.
6 X-ray the patient.
7 Take the image receiver out of the x-ray tray and read it.
✓ The patient's x-ray is available.
The Developmental care view graphically shows how the noise level and the light
level in the patient compartment change over time.
27247
If the monitoring limits for noise and light are exceeded, the color of the graph
changes in the Developmental care view and in the Trends/data dialog
window as follows:
– If the medium threshold is exceeded, the graph turns yellow.
– If the upper threshold is exceeded, the graph turns red.
28479
1
The monitoring limits for noise and light are only indicated visually. There is no
acoustic signal.
WARNING
Risk of electric shock
If the interface connectors and the patient are touched at the same time, there is a
risk of an electric shock.
► Do not simultaneously touch the audio port and the patient.
Procedure:
1 Connect the audio cable to an MP3 player or a cellular phone.
2 Insert the audio cable into the audio port on the rear of the sensor wall. For
further information, see the following chapter: ''Connectors on the rear sensor
wall'' (page 30)
3 Start playback and adjust the volume on the playback device.
The voices or music are played in the patient compartment.
4 Observe the patient while the audio file is played.
5 Stop the playback after an appropriate amount of time.
The selection and use of audio material is left to the discretion of the clinical
staff.
7 Alarms
7.1 Alarm priorities and alarm signals
This device indicates alarms visually and acoustically. Each alarm is assigned a
specific priority, which indicates the level of urgency. The alarm conditions are
monitored automatically. The alarm presettings for the factory settings, user-
specific settings and patient-specific settings are retained after a power supply
failure, no matter how long it lasts.
If several alarms are triggered at the same time, the alarm message with the
highest priority is displayed first. High-priority alarm messages that are no longer
active are displayed in the background color of the alarm message field.
If the fault triggering the alarm is not eliminated after the suppression time has
elapsed, the alarm signal sounds again.
1 Press the "Audio paused" key.
✓ In the header bar, the symbol and the remaining time for the suppressed alarm
signal are displayed.
Alarm
Air temperature too high
Central skin temp. sensor missing
Check oxygen supply
Data loss
Fan motor faulty, heating off
Mattress faulty
Mattress temperature too high
Oxygen module faulty
Oxygen sensor faulty
Peripheral skin temp. sensor missing
Temperature high in warm air duct
Procedure:
1 Touch the Alarm reset button on the header bar on the main screen.
✓ The alarm message has been dismissed and is no longer displayed.
8 Configuration
8.1 Entering patient data
This chapter describes how to enter patient data. At the start of therapy, the device
suggests an air temperature range based on the patient data and weight entered.*
For further information, see the following section: ''Recommended air
temperature'' (page 133)
Prerequisite:
– The device is in operation.
Procedure:
1 Touch the System setup... button (1) in the main menu bar.
The System setup dialog window and the Patient dialog page are
displayed.
28495
1
2 Touch the pen icon to the right of the relevant input field.
The keyboard is displayed. 27252
Fig. 74 Keyboard
* Sauer PJ, Dane HJ, Visser HK. New standards for neutral thermal environment of healthy very
low birth weight infants in week one of life. 1984
1 Switch off the device on the display unit using the Start/standby key .
The Shut down dialog window is displayed.
2 Touch the Shut down button.
The device shuts down.
3 Switch on the device on the display unit using the Start/standby key .
The start-up screen is displayed.
4 Touch the New patient button.
The existing patient data is deleted.
The therapy settings and the patient-specific alarm settings are reset to the
user-specific settings.
The logbook is not deleted.
✓ Patient data can be entered anew.
28495
1
* Sauer PJ, Dane HJ, Visser HK. New standards for neutral thermal environment of healthy very
low birth weight infants in week one of life.1984
Additional information:
– According to the data from this study, the recommended air temperature for a
patient aged 1 to 7 days does not depend on the patient's weight. Consequently,
the weight range cannot be selected for patients of this age.
– The last measured weight is used for determining the recommended air
temperature range. If there is no scale in the device, the weight range must be
selected manually.
– The range for the recommended air temperature in the System setup dialog
window is automatically adjusted in line with the age. The required air
temperature must be set on the main screen.
– If the ClearView on setting or the Auto setting of the Humidity therapy control
is changed or the patient is clothed or covered, this must be taken into account
in the air temperature settings.
– The recommended air temperature applies for patients up to an age of 36 days
and < 2500 g (5.5 lbs).
For further information, see the following chapter: ''Starting air temperature
mode'' (page 87)
WARNING
Risk of patient injury due to incorrect patient data
Only one data record with configuration data and one data record with patient data
can be stored on the USB mass storage device. If a data record already exists on
the USB mass storage device, the new data record overwrites the existing data.
► Check the configuration data and the current settings on the new device after
the data import.
WARNING
Risk of electric shock on USB port
Connecting mains voltage carrying accessories to the USB port may trigger an
electric shock.
► Only connect USB mass storage devices that are included in the list of approved
accessories to the USB port.
► Do not connect any accessory operating on mains voltage to the USB port.
► Do not simultaneously touch the USB port and the patient.
Procedure:
1 Connect the USB mass storage device to the USB port on the rear of the main
column.
It may take a few seconds for the device to recognize the USB mass storage
device.
2 Touch the System setup... button (3) in the main menu bar.
30325
2
3
7 To import the configuration data, touch the Import button and confirm with the
rotary knob.
The configuration data is imported.
The imported data is displayed with date and time.
Once the import is complete, the device restarts automatically.
Once the import is complete, the patient data is automatically deleted from
the USB mass storage device. This prevents patient data from being
transferred accidentally. The configuration data is not deleted.
8 Once the device is ready for operation again, check the alarm limits and the
therapy values.
9 Remove the USB mass storage device from the USB port.
The configuration data and the patient data have been transferred to the
other device.
✓ The configuration data can be transferred to another device.
This chapter describes the maximum possible display of trends and data.
Depending on how the device is configured, the display on the actual device
may differ.
Prerequisite:
– The device is in operation.
Procedure:
1 Touch the Trends/data... button (1) in the main menu bar.
28487
28488
1
The color of the buttons indicates the overviews in which parameters can still be
selected.
Color Meaning
Dark green The parameter is selected and displayed as a
graphical trend.
Light green The parameter can be selected.
Gray The parameter cannot be selected.
8.3.4 Logbook
All events and alarms are displayed in chronological order in the Logbook tab. The
latest event/latest alarm is at the top. Earlier events and alarms can be displayed
using the arrow keys.
The entries in the logbook cannot be deleted and are preserved even after the
device is switched off and on or after a power supply failure. There is no record kept
of how long the device and alarm system were shut down. When the capacity of the
logbook is reached, the oldest entries are overwritten.
The following events and alarms are displayed:
– All alarms, their confirmation, and muting by the user
– Alarm limit changes
– Operating mode changes
– Therapy control changes
– Start and end of applications
– Events such as hood open, side access panel open
– Data import and export to USB mass storage device
– Device time changes
The description assumes the maximum device configuration. The actual device
may differ in its functions from the descriptions below.
Factory settings
Dräger delivers the device with factory settings that are used when starting the
device for the first time. Users can reset the device to factory settings.
User-specific settings
The user can define specific settings for the device. The settings (e.g., therapy
values, alarm limits, or screen layout) can be made in the System setup dialog
window. User-specific settings are saved automatically once they are confirmed.
User-specific settings can be transferred to other devices via the Import/export
function. For further information, see the following chapter: ''Transferring data to
another device'' (page 134)
User-specific settings are automatically applied in the following situations:
– When the device is restarted after it has been switched off for more than 2
hours.
– If the New patient button was activated when the device was switched on.
Patient-specific settings
The user can adjust the factory settings or the user-specific settings to suit the
current patient. The values for the patient-specific settings are set on the main
screen via the therapy bar (modes and therapy controls). Patient-specific alarms
are set via the main screen in the Alarms dialog window. The patient-specific
settings also include the information in System setup > Patient dialog window.
Patient-specific settings can be transferred to other devices via the Import/export
function. For further information, see the following chapter: ''Transferring data to
another device'' (page 134)
Patient-specific settings are automatically set in the following situations:
– When the device is restarted after it has been switched off for less than 2 hours.
– If the Current patient button was activated when the device was switched on.
The configuration password must be entered for some settings. For further
information, see the following section: ''Configuration password for Babyleo
TN500 Software 1.0n'' (page 213).
The Standard function or the Eco function can be selected for the transition
between incubator operation and radiant warmer operation. For further
information, see the following chapter: ''Configuring the transition'' (page 76).
The Standard function automatically activates the ClearView on setting. The
ClearView on setting can only be deactivated when the Eco function is
selected.
9 Transfer
9.1 Intrahospital patient transfer with switched off warming
therapy
This chapter describes how to move the patient in the device into another room.
28226
Fig. 79 Label on the lifting column
WARNING
Risk of patient injury due to interruption of warming therapy
When the patient is moved to another room in the device, the device is switched off
during the transfer.The warming therapy is interrupted and the skin temperature is
not measured.
► During the transfer, monitor the patient's body temperature with a separate
measuring device.
► Make sure that the patient is sufficiently warmed during the transfer.
WARNING
Risk of patient injury during transfer
Device parts may be damaged during transfer. The device may tip over. The patient
may be injured and device functions may be impaired. Before transfer, perform the
following measures:
► Set the device to transfer height and observe the information in the chapter
"Transferring the patient".
► Make sure that there is always sufficient clearance above the radiant warmer
and that the radiant warmer is not damaged.
► Make sure that accessories, e.g., pressure reducers and cylinder valves, do not
protrude beyond the trolley during the transfer.
► When transferring the patient, the device must always be moved by two people.
► Avoid shocks and vibrations as much as possible, e.g., when crossing door
thresholds and passing through door frames.
Check the functional integrity of the accessories during and after the
intrahospital transport. Observe the chapter "Adjusting the device height".
Procedure:
1 Adjust the height of the device using the foot pedals. The correct transfer height
is reached when the red hatched label on the lifting column is completely
covered.
The correct device height for transfer is set.
26836
Check for operational readiness after the transfer. For further information, see
the following chapter: ''Checking for operational readiness'' (page 63)
10 Troubleshooting
10.1 Alarm – Cause – Remedy
The alarm messages are displayed in the message field of the header bar in
hierarchical order. For further information, see the following chapter: ''Alarm
priorities and alarm signals'' (page 127)
Different background colors indicate the priority levels of the alarms.
In the Current alarms and Alarm history tables, the priority of the alarm
messages is also indicated by exclamation marks.
Alarm priority Exclamation marks Color
High !!! Red
Medium !! Yellow
Low ! Cyan
In order to classify the alarms within an alarm priority, internal priority numbers are
specified in the table below. The most critical alarm is given the number 255. The
priority of the alarm decreases the lower the number is.
In the following table, the alarm messages are listed in alphabetical order. If an
alarm occurs, the table helps to identify causes and remedies. The different causes
and remedies should be worked through in the order listed until the alarm has been
resolved.
Some alarms appear in this table several times with different priorities because their
priority may change under certain conditions.
When the mains power supply is restored, the device continues the warming
therapy with the settings that were activated before the power supply failure.
30800
Fig. 81 Heating plate
WARNING
Risk due to penetrating liquid
If too much liquid is on the heating plate, the liquid may penetrate into the device
and cause damage to the device. If the device fails, the patient is put at risk.
► Make sure that no liquid penetrates into the device.
► If the liquid level on the heating plate reaches the upper edge of the fan impeller
receptacle, stop using the device and contact the specialized service personnel.
Procedure:
1 Switch off the device on the display unit using the Start/standby key .
2 Switch off the device with the main switch on the rear of the main column.
3 Disconnect the mains plug.
4 Remove any liquid on the mattress and the mattress tray.
5 Remove the mattress and mattress tray.
6 Remove any liquid on the bed support.
7 Remove the bed support and the T-bars.
8 Remove any liquid from the warm air duct covers.
9 Remove the warm air duct covers.
10 Check if the fan impeller is in the liquid and remove the fan impeller.
11 Check if there is liquid on the heating plate. The heating plate may be hot.
12 If the liquid level reaches the upper edge of the fan impeller receptacle (1),
contact the specialized service personnel. The specialized service personnel
must check if liquid has penetrated into the device.
30275
1
11 Reprocessing
11.1 Automatic humidifier cleaning
This chapter describes how to perform automatic humidifier cleaning.
WARNING
Risk of patient injury due to overheating
If the automatic humidifier cleaning is started when the patient is in the device, this
may result in the patient overheating of the patient.
► Only start automatic humidifier cleaning when there is no patient in the device.
Procedure:
1 Make sure there is no patient in the device.
2 Open the hood.
3 Touch the Special procedures... button (2) in the main menu bar.
The Special procedures dialog window is displayed.
28494
27243
Fig. 84 Starting humidifier cleaning
7 Wait until the humidifier cleaning has ended. Do not disassemble the device
during cleaning.
11.2 Dismantling
This chapter describes how to disassemble the device before reprocessing.
Only disassemble the device after the patient and all items have been removed
from the device. All additional devices and accessories must be removed before
reprocessing.
30733
1
2
12
10 11
7 5
Fig. 85 Dismantling
No. Designation
1 Heated mattress (option) or SoftBed mattress
2 Mini-walls
3 Hood sealings
4 Side access panels and front access panel
5 Drawer insert (option)
6 Hose grommets
7 Fan impeller
8 Small and large warm air duct cover
9 T-bars
10 Bed support with scale (option)
11 X-ray tray
12 Mattress tray
CAUTION
Risk of injury due to hot surfaces
If the device is dismantled directly after operation or after the automatic humidifier
cleaning, hot surfaces may cause injuries.
► Before dismantling, make sure that the device has cooled down.
► Do not touch the hot surfaces near to the radiant warmer, the convection heater,
and the humidifier.
Procedure:
Surface disinfection
Surface disinfection using the following surface disinfectant is suitable for the
device and all components (non-critical and semi-critical):
Class of active Product name ManufacturerContact Concentra-
ingredient time tion
Oxygen-releasing Dismozon plus BODE Chemie 15 min 1,6 %
agents
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electrical shock when switching on the device
– Device malfunctions
► Make sure that no liquid penetrates into the device.
WARNING
Risk of O2 sensor malfunction
Disinfectants or cleaning agents which contain formaldehyde or ammonia may
damage the O2 sensors. The O2 sensors may provide incorrect measured values.
This may put the patient at risk.
► Only use disinfectants or cleaning agents which do not contain formaldehyde or
ammonia.
If no specific instructions for use are available for the approved device accessory,
then the accessory can be reprocessed by means of surface disinfection using the
listed surface disinfectants.
Procedure:
1 Wipe the inside and outside of the hood.
2 When treating the hood seal, ensure that the disinfectant also comes into
contact with the inner part of the hood seal.
3 Wipe the inside and outside of the side panels. Pay particular attention to the
locking mechanism of the hand ports.
4 Carefully wipe the grip on the X-ray flap.
5 To clean the air ducts of the X-ray flap (1), pull a piece of cloth soaked in
disinfectant through each duct.
30231
1
WARNING
Risk of wear and material fatigue
Cleaning and disinfecting can cause components to wear, resulting in device
malfunctions. There is a risk of injury.
► Check the device and reusable products for signs of wear (e.g., cracks,
deformation, discoloration, or detachment) and replace if necessary.
Procedure:
12 Service
12.1 Overview
This chapter describes the service measures required to maintain the proper
functioning of the product. Technical documentation is available on request.
12.2 Inspection
Measure Interval Target group
Inspection and safety check 1) Every 12 months Service personnel
Metrological check of the skin Every 24 months Service personnel
temperature measurement
system
Visual check of the internal 6 years after commis- Service personnel
compressed gas hose sioning. If replacement
is not required, check
every 24 months.
Calibrating the scale Every 12 months Service personnel
Verifying the scale According to regulatory Officially authorized spe-
requirements for scales cialist personnel
which must be verified
1) Designation applies to the Federal Republic of Germany; corresponds to the “Recurring safety
inspection” in the Federal Republic of Austria
WARNING
Risk due to missing service measures
If the device is not maintained at the prescribed intervals, the function and the
safety of the device cannot be ensured.This may put the health of patients and
users at risk.
► Perform service measures on the device at the prescribed intervals.
► Service measures must only be performed by suitably qualified persons.
WARNING
Risk of patient injury due to service measures performed during therapy
Performing service measures during therapy may endanger the patient.
► Only perform service measures when there is no patient in the device.
Set value 70 %
Accuracy ±10 %
12 Check the operating parameters for the lifting column under Service menu >
Operating Data > Lifting Column. The lifting column must be replaced if one of
the two operating parameters exceeds the maximum value. Contact
DrägerService.
Resistance Temperature
1586 Ω 33.2 °C (91.8 °F)
1412 Ω 36.0 °C (96.8 °F)
1312 Ω 37.8 °C (100.0 °F)
Procedure:
1 Set and connect the 3 simulation resistors.
2 Compare the displayed measured temperature values with the temperature
values in the table.
The displayed measured temperature value must not deviate from the
temperature value simulated by the relevant resistance by more than 0.1 °C
(0.18 °F).
12.3 Maintenance
Component Interval Measure Target group
Skin temperature sen- Weekly, at the lat- Replace Users
sors est when the
patient changes
Sensor covers Weekly, at the lat- Replace Users
est when the
patient changes
Hose grommets As required, when Replace Users
the material is
brittle or when the
ribs are torn out.
Heated mattress (option), As required, when Replace Users
SoftBed mattress the heated mat-
tress or the Soft-
Bed mattress are
damaged.
Filter cloth (in the sensor Every 6 months Replace, see Users
wall) page 185
Oxycell O2 sensors Every 2 weeks Adjustment Users
Every 24 months Replace Service personnel
12.4 Replacing the filter cloth for patient air in the sensor
wall
The filter cloth for patient air is located in the rear of the sensor wall.
WARNING
Risk due to soiled or expired filter cloth
If the filter cloth is soiled or the life span has elapsed, the patient may be put at risk.
► Replace a soiled or worn out filter cloth.
Procedure:
1 Press the flap lock (1) down.
30704
1
12.5 Repair
Dräger recommends that all repairs are performed by DrägerService and that only
authentic Dräger spare parts are used.
12.6.1 Loading the activation code from a USB mass storage device
Prerequisites:
– A USB mass storage device with valid activation codes for this device is
connected to the USB interface.
– The Service dialog window is opened.
1 Open the Applications (Applications) dialog page.
2 Touch the Load from USB (Load from USB) button.
The activation codes found on the storage medium are uploaded and
displayed in the list.
13 Disposal
13.1 Disposing of the device
The disposal of electrical and electronic devices is subject to special guidelines.
This device must be disposed of in accordance with national regulations. In
countries of the European Union, Dräger will organize the return of the device.
Additional information is available at www.draeger.com (search term: WEEE).
14 Technical data
14.1 Ambient conditions
During operation
Temperature 20 to 35 °C (68 to 95 °F)
Ambient pressure 620 to 1100 hPa (9.0 to 16.0 psi)
Relative humidity 20 to 95 %, non-condensing
Air flow in air-conditioned rooms Max. 0.3 m/s (11.8 in/s)
During storage and transportation
Temperature - 20 to 60 °C (- 4 to 140 °F)
Ambient pressure 500 to 1100 hPa (7.3 to 16.0 psi)
Relative humidity 10 to 95 %, non-condensing
14.5 Materials
Bisphenol A (BPA) The patient compartment of the device (without
accessories) is free from bisphenol A
Latex The device (without accessories) contains no
latex-containing components
1 - Common port
2 - Alarm
3 - No alarm
4 - Nurse call
COM port Only connect devices that fulfill the require-
ments of the standard IEC 60950-1 on
unearthed SELV circuits or the requirements of
the standard IEC 60601-1 on accessible sec-
ondary circuits with max. 60 V DC nominal volt-
age.
Type 9-pole D-sub, electrically isolated:
1.5 kV against internal electronics
1.0 kV against housing
Protocol MEDIBUS.X
Alarm delay Typically < 2 s
Baud rate 1200, 2400, 4800, 9600, 19200, 38400 baud
15 Annex
15.1 Description of skin temperature regulation
This chapter describes skin temperature regulation in incubator operation and in
radiant warmer operation.
Alarm priority Standard (in accordance with with IEC 60601-1- Repeated
8)
Yes
No
Low
The acoustic alarm signals described here will be output by a reserve loudspeaker
if the main loudspeaker should fail. The reserve loudspeaker plays the tone
sequences of the Warning and Caution alarm priorities at the same tone frequency
and volume level.
When Signal
Time-out following non-confirmation of
selection
15.6 Interfaces
The following devices can be connected to the interfaces on the device:
Audio interface
15.7 Abbreviations
Abbreviation Explanation
% Radiant warmer power
%, Vol % Percentage gas ratio, related to total volume
A Ampere
Air Medical compressed air
CISPR Comité International Spécial des Perturbations Radioélec-
triques, International Special Committee on Radio Interference
CO2 Carbon dioxide
COM Serial port
CSA Canadian Standards Agency
dB(A) Sound pressure level, A-weighted
EMC Electromagnetic compatibility
ESD Electrostatic Discharge, electrostatic discharge
fc Foot candle
HF High-frequency
hPa Hectopascal
ID Identification
kg Kilogram
L Liter
LAN Local network
lbs Pound; unit of mass
LED Light emitting diode
mbar Millibar
MEDIBUS.X Dräger communications protocol for medical devices with uni-
form data definition for all devices
min Minute
mL Milliliter
MRI Magnetic resonance imaging
oz Ounce
ppm Parts per million
RFID Radio Frequency Identification, radio frequency identification
RI Revision index
SNMP Simple Network Management Protocol
UMDNS Universal Medical Device Nomenclature System, nomenclature
for medical devices
USB Universal Serial Bus, computer interface
15.8 Symbols
Symbol Explanation
Manufacturer
Date of manufacture
WEEE marking
Start/standby
Audio paused
Mains voltage
Serial number
Revision index
Part number
Symbol Explanation
Use by
Storage temperature
Relative humidity
Atmospheric pressure
Do not reuse
Spare part
Nurse call
USB port
2
High-density polyethylene, class 2
PE-HD
Working light
Symbol Explanation
Night light
Air temperature
Radiant warmer
Convection heater
Humidity
Alarm inactive
Locked
Unlocked
Enter key
Scale Explanation
Socket for the scale
Castor Explanation
Green arrow: castor is movable in longi-
tudinal direction (direction lock)
Green lock: castor is not locked (no
direction lock)
Red lock: castor is locked
16 Password
16.1 Configuration password for Babyleo
TN500 Software 1.0n
Cut out from the instructions for use Babyleo TN500 Software 1.0n
To prevent unauthorized alteration, the device settings for Babyleo TN500 are
protected by the following configuration password:
0000
Index
A Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Causes . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . 205
Colors. . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Access panel . . . . . . . . . . . . . . . . . . . . . . . . . 52
Device settings. . . . . . . . . . . . . . . . . . . . 144
Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Logbook . . . . . . . . . . . . . . . . . . . . . . . . . 141
Opening . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Manual mode . . . . . . . . . . . . . . . . . . . . . . 91
Accessories
Remedy . . . . . . . . . . . . . . . . . . . . . . . . . 153
Safety instruction . . . . . . . . . . . . . . . . . . . 23
Safety instruction . . . . . . . . . . . . . . . . . . . 12
Activating
APGAR
Applications . . . . . . . . . . . . . . . . . . . . . . 185
Timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Licensed application . . . . . . . . . . . . . . . . 186
Applications
Activation code
Activating . . . . . . . . . . . . . . . . . . . . . . . . 185
Loading from USB mass storage device 186
Assembling
Manual entry . . . . . . . . . . . . . . . . . . . . . . 186
Device . . . . . . . . . . . . . . . . . . . . . . . . . . 176
Adjusting
Assembly and preparation. . . . . . . . . . . . . . . 56
Device height . . . . . . . . . . . . . . . . . . . . . . 48
Attaching
Mattress tray . . . . . . . . . . . . . . . . . . . . . . . 49
Skin temperature sensor . . . . . . . . . . 79, 81
O2 sensors . . . . . . . . . . . . . . . . . . . . . 15, 99
Audio file
Patient weight . . . . . . . . . . . . . . . . . . . . . 121
Playing . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Advanced skin temperature alarm limit
Auto
Setting . . . . . . . . . . . . . . . . . . . . . . . . 84, 88
Heated mattress. . . . . . . . . . . . . . . . . . . 103
Air temperature alarm limit
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Air temperature mode . . . . . . . . . . . . . . . . . . 85
Ending . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 B
Extended temperature range . . . . . . . . . . 85 Bed-tilt mechanism
Main screen . . . . . . . . . . . . . . . . . . . . . . . 86 Activating . . . . . . . . . . . . . . . . . . . . . . . . . 49
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Temperature range . . . . . . . . . . . . . . . . . . 85 C
Alarm condition
Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . 203 Canceling
Alarm limit Humidifier cleaning. . . . . . . . . . . . . . . . . 167
Advanced . . . . . . . . . . . . . . . . . . . . . . 84, 88 Castor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Air temperature . . . . . . . . . . . . . . . . . . . . . 88 Locking brake. . . . . . . . . . . . . . . . . . . . . . 26
Automatic . . . . . . . . . . . . . . . . . . . . . 89, 110 Checks
Kangaroo mode . . . . . . . . . . . . . . . . . . . 107 Metrological . . . . . . . . . . . . . . . . . . . . . . 183
Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Skin temperature . . . . . . . . . . . . . . . 83, 110 Closing
Tolerate cooling . . . . . . . . . . . . . . . . . . . 110 Access panel . . . . . . . . . . . . . . . . . . . . . . 53
Alarm message . . . . . . . . . . . . . . . . . . . . . . 153 Hand port . . . . . . . . . . . . . . . . . . . . . . . . . 53
Dismissing . . . . . . . . . . . . . . . . . . . . . . . 130 Hood. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Alarm priorities. . . . . . . . . . . . . . . 127, 127, 153 Color
Alarm signal . . . . . . . . . . . . . . . . . . . . . . . . . 127 Control element . . . . . . . . . . . . . . . . . . . . 46
Delaying . . . . . . . . . . . . . . . . . . . . . . . . . 127 Skin temperature sensor . . . . . . . . . . . . . 79
Suppressing . . . . . . . . . . . . . . . . . . 127, 130 Components
Alarm tones Disassembling . . . . . . . . . . . . . . . . . . . . 169
Schematic representation . . . . . . . . . . . . 203 Condensation management . . . . . . . . . . . . . 69
Configuration . . . . . . . . . . . . . . . . . . . . . . . . 131
Hood . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 L
Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Licensed application
Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Activating . . . . . . . . . . . . . . . . . . . . . . . . 186
Direction of movement . . . . . . . . . . . . . . . 51
Lifting column
Opening . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Safety check. . . . . . . . . . . . . . . . . . . . . . 181
Hood cover . . . . . . . . . . . . . . . . . . . . . . . . . 126
Liquid on heating plate
Humidification system
Removing . . . . . . . . . . . . . . . . . . . . . . . . 164
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Load
Humidifier cleaning . . . . . . . . . . . . . . . . . . . 166
Maximum . . . . . . . . . . . . . . . . . . . . . . . . . 56
Canceling . . . . . . . . . . . . . . . . . . . . . . . . 167
Loading from USB mass storage device
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Activation code. . . . . . . . . . . . . . . . . . . . 186
Humidity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Lock. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Auto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Bed-tilt mechanism. . . . . . . . . . . . . . . . . . 55
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . 94
Device height adjustment. . . . . . . . . . . . . 55
Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Locking brake
Setting range . . . . . . . . . . . . . . . . . . . . . . 94
Castor. . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Double castor . . . . . . . . . . . . . . . . . . . . . . 26
I Logbook . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Importing
Configuration data . . . . . . . . . . . . . . . . . 136 M
Patient data . . . . . . . . . . . . . . . . . . . . . . 136
Main column
Incubator operation . . . . . . . . . . . . . . . . . . . . 69
Detailed view . . . . . . . . . . . . . . . . . . . . . . 28
Interrupting . . . . . . . . . . . . . . . . . . . . . . . . 75
Main menu bar . . . . . . . . . . . . . . . . . . . . . . . 33
Main screen . . . . . . . . . . . . . . . . . . . . . . . 78
Main screen
Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Air temperature mode . . . . . . . . . . . . . . . 86
Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Incubator operation . . . . . . . . . . . . . . . . . 78
Transition . . . . . . . . . . . . . . . . . . . . . . . . . 74
Kangaroo mode . . . . . . . . . . . . . . . . . . . 104
Information tones
Manual mode . . . . . . . . . . . . . . . . . . . . . . 91
Schematic representation . . . . . . . . . . . . 203
Radiant warmer operation . . . . . . . . . . . . 79
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 184
Metrological checks . . . . . . . . . . . . . . . . 183
Manual entry
Safety checks . . . . . . . . . . . . . . . . . . . . . 181
Activation code. . . . . . . . . . . . . . . . . . . . 186
Instructions for use
Manual mode . . . . . . . . . . . . . . . . . . . . . . . . 91
Safety instruction . . . . . . . . . . . . . . . . . . . 11
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Ending . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Interrupting
Main screen . . . . . . . . . . . . . . . . . . . . . . . 91
Incubator operation. . . . . . . . . . . . . . . . . . 75
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Intrahospital patient transfer . . . . . . . . . . . . 150
Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
IT networks . . . . . . . . . . . . . . . . . . . . . . . . . 194
Heated mattress. . . . . . . . . . . . . . . . . . . . 51
SoftBed . . . . . . . . . . . . . . . . . . . . . . . . . . 51
K Mattress core temperature
Kangaroo mode . . . . . . . . . . . . . . . . . . . . . . 104 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Alarm limit . . . . . . . . . . . . . . . . . . . . . . . . 107 Mattress tray
Duration . . . . . . . . . . . . . . . . . . . . . . . . . 104 Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . 49
Ending . . . . . . . . . . . . . . . . . . . . . . . . . . 108 Maximum load . . . . . . . . . . . . . . . . . . . . . . . . 56
Main screen . . . . . . . . . . . . . . . . . . . . . . 104 Modifications
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Safety instruction . . . . . . . . . . . . . . . . . . . 12
Transition . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Configuring . . . . . . . . . . . . . . . . . . . . . . . . 76
Temperature adjustment. . . . . . . . . . . . . . 75
Trends
Displaying . . . . . . . . . . . . . . . . . . . . . . . . 137
Graphical . . . . . . . . . . . . . . . . . . . . . . . . 138
Tabular . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 153
Adjusting the height of the device. . . . . . 162
Adjusting the mattress tray . . . . . . . . . . . 162
Fault with the oxygen supply . . . . . . . . . 163
Liquid on heating plate . . . . . . . . . . . . . . 164
Power failure alarm. . . . . . . . . . . . . . . . . 161
Weighing. . . . . . . . . . . . . . . . . . . . . . . . . 162
U
Unlock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Bed-tilt mechanism . . . . . . . . . . . . . . . . . . 55
Device height adjustment . . . . . . . . . . . . . 55
Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
V
Verification
NAWI scale . . . . . . . . . . . . . . . . . . . . . . . 181
W
Warm-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Ending . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Weaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Ending . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Weighing . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
X
X-ray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
X-rays
Taking . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Manufacturer
Á9054863gÈ
9054863 – GA 6016.170 en
© Drägerwerk AG & Co. KGaA
Edition: 2 – 2017-01
(Edition: 1 – 2016-09)
Dräger reserves the right to make modifications
to the device without prior notice.