You are on page 1of 8

ARTICLE IN PRESS

Medical Dosimetry ■■ (2017) ■■–■■

Medical Dosimetry
j o u r n a l h o m e p a g e : w w w. m e d d o s . o r g

Dosimetry Contribution:

Fine-tuning the normal tissue objective in eclipse for lung


stereotactic body radiation therapy
James P. Bell, M.S., R.T.T.,* Pretesh Patel, M.D.,† Kristin Higgins, M.D.,†
Mark W. McDonald, M.D.,‡ and Justin Roper, Ph.D., D.A.B.R.‡
*Department of Radiation Oncology, Simmons Comprehensive Cancer Center at the University of Texas Southwestern Medical Center, Dallas, TX;

Winship Cancer Institute of Emory University, 1356 Clifton Road NE, Building C, Atlanta, Georgia 30322; and ‡Hospital Corporation of America,
Sarah Cannon Cancer Center, Department of Radiation Oncology, 2410 Patterson Street, Basement Level, Nashville, TN 37203

A R T I C L E I N F O A B S T R A C T

Article history: The purpose of this study was to characterize the effects of the normal tissue objective (NTO)
Received 2 May 2017 on lung stereotactic body radiation therapy (SBRT) dose distributions. The NTO is a spatial-
Received in revised form 13 ly varying constraint used in Eclipse to limit dose to normal tissues by steepening the dose
November 2017
gradient. However, the multitude of potential NTO setting combinations challenges optimal
Accepted 13 November 2017
NTO tuning. In the present study, a broad range of NTO settings are investigated for lung
SBRT treatment planning with volumetric modulated arc therapy(VMAT). Ten prior lung SBRT
Keywords:
cases were replanned using NTO priorities of 1, 50, 100, 200, 500, and 999 in combination
NTO
with fall-off values of 0.01, 0.05, 0.10, 0.15, 0.20, 0.30, 0.50, 1.00, and 5.00 mm−1 and the au-
Fall-off
tomatic NTO. NTO distances to planning target volume (PTV), start dose, and end dose were
SBRT
1 mm, 100%, and 10%, respectively, for all 600 plans. Prescription dose covered 95% of the
PTV. The following metrics were recorded: conformity index (CI), ratio of the 50% prescrip-
tion isodose volume to PTV (R50%), maximum dose 2 cm away from PTV (D2cm), lung volume
of ≥20 Gy (V20Gy), maximum PTV dose (PTVmax), and monitor units (MUs). Differences between
prior plans and NTO plans were evaluated using the Wilcoxon signed-rank test. Different
combinations of NTO settings resulted in wide-ranging plan quality metrics: CI (1.00 to 1.54),
R50% (3.95 to 7.57), D2cm (33.4% to 67.9%), V20Gy (1.66% to 2.75%), MU (1.81 cGy−1 to 4.69 cGy−1),
and PTVmax (118% to 175%). Although no settings were optimal for all metrics, a fall-off of
0.15 mm−1 and a priority of 500 best satisfied institutional criteria. Compared with prior plans,
NTO plans resulted in significantly lower R50% (4.00 vs 4.35, p = 0.002), lower V20Gy (1.22% vs
1.32%, p = 0.006), and higher PTVmax (138% vs 122%, p = 0.002). All of the prior and well-
tuned NTO plans met Radiation Therapy Oncology Group (RTOG) 0813 guidelines. Lung SBRT
dose distributions were characterized across a range of NTO settings. NTO plans with well-
tuned settings compared favorably with prior plans.
© 2017 American Association of Medical Dosimetrists.

Introduction

Reprint requests to James P. Bell, M.S., R.T.T., Department of Radiation


Oncology, Simmons Comprehensive Cancer Center at the University of
For medically inoperable early-stage lung cancer pa-
Texas Southwestern Medical Center, Dallas, TX, USA. tients, stereotactic body radiation therapy (SBRT) has
E-mail: James.Bell@UTSouthwestern.edu emerged as a viable treatment option, providing an
https://doi.org/10.1016/j.meddos.2017.11.004
0958-3947/Copyright © 2017 American Association of Medical Dosimetrists
ARTICLE IN PRESS
2 J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■

ablative effect on tumors rivaling outcomes found with sur- 2) start dose (f0),
gical resection.1-3 In a recent review of retrospective studies, 3) end dose (f∞),
SBRT was recommended because of substantially better local 4) and fall-off (k).
control and overall survival rates compared with conven-
tional radiation therapy (RT) techniques.3 SBRT mandates high The NTO equationa is as follows:
accuracy and high precision RT; therefore, quality assur-
⎧ x ≥ xstart ⎫
ance standards are tighter than those associated with f ( x ) = ⎨ f0 − k( x − xstart ) + f∞ (1 − e − k( x − xstart ) ) , ⎬
conventionally fractionated RT.4-6 Successful implementa- ⎩ f0,< xstart ⎭
tion of SBRT requires accurate patient positioning, secure
The first parameter xstart is the distance away from a PTV
immobilization, and a distinctive set of planning strategies
border, which defines a plateau region of uniform dose abut-
for optimal delivery of highly biologically effective doses.3,7,8
ting a target volume.b The second parameter, f0, is the start
In contrast with the large treatment volumes and relative-
dose. The start dose may be set as a certain percentage of
ly uniform doses used in conventional RT, relatively small
the prescription dose and defines the dose level in the plateau
targets are treated in SBRT using highly conformal dose dis-
region. The third parameter, f∞, is the end dose. The end dose
tributions with sharp dose gradients. Excellent conformity
is the lowest dose penalized by the NTO. The location of the
and rapid dose gradients are important because healthy
end dose is influenced by the fall-off parameter k. This ex-
tissues that receive intermediate to high-dose spillage may
ponential decay constant may range from 0 to 10 (mm−1) and
become severely damaged, with the potential for long-
determines the steepness of the NTO curve beyond the dose
term toxicity.2
plateau. In addition to these shape parameters, the impor-
To achieve the desired dose distribution for SBRT of the
tance of the NTO is determined by a priority parameter.
lung, several treatment planning strategies may be em-
Priority may range from 0 to 1000 and determines the rel-
ployed clinically. Multiple static beams or arcs are often spread
ative importance of the NTO to any other optimization
over a wide angular range, including noncoplanar
constraints. The NTO has no effect when the priority is set
orientations.4,9 Additionally, treatment planners may adjust
to 0, whereas the highest allowable priority will apply a
beam weights, multileaf collimators, or other modifiers when
maximum penalty.
using 3-dimensional-conformal RT techniques. Inverse plan-
While the NTO has a flexible equation that allows for a
ning, e.g., intensity modulated radiation therapy (IMRT) and
customized dose profile penalty and weighting, there is little
volumetric modulated arc therapy (VMAT), is also used for
guidance as to which settings are optimal for a specific task.
SBRT. Inverse planning techniques may include optimiza-
Here, NTO settings were studied for the task of lung SBRT
tion structures such as concentric rings used to isotropically
treatment planning. A broad range of NTO priorities and fall-
steepen the dose distribution. Additional penalties may be
off values were investigated in an effort to determine settings
applied to preferentially spare sensitive healthy tissues.10 Al-
that achieve highly conformal dose distributions with sharp
though effective, the design and the use of ring structures
dose gradients. Plans were evaluated using dosimetric bench-
increase the planning time. The planner is faced with the
marks from Radiation Therapy Oncology Group (RTOG) 0813,
additional challenge of choosing the number of rings, the
a clinical trial on SBRT for early-stage lung cancer.11
ring width, gaps or overlaps, and the associated penalties.
For users of the Eclipse treatment planning system, the normal
tissue objective (NTO) is a tool available during optimiza- Methods and Materials
tion that requires no additional contouring time to achieve
sharp dose gradients beyond the target volume. Little time Ten consecutive patients who were previously treated at
or effort is needed to specify NTO settings; however, it is dif- our institution using SBRT for lung tumors were selected for
ficult to know which NTO settings are appropriate for a the present study. Cases were anonymized. Plans con-
particular plan. Guidelines on optimal NTO settings are lacking. sisted of 3 to 6 noncoplanar 6X-FFF VMAT arcs that broadly
Clarifying the NTO selection process during optimization could cover the collision-free zone. The collision-free zone was de-
potentially pave the way for NTO-based planning. termined before planning to make full use of dosimetrically
Eclipse has 2 types of NTO: an automatic formulism advantageous safe angles. The arc arrangement from the orig-
defined by the vendor and a spatially varying dose con- inal plan was used as a template when generating 60
straint defined by the user. The treatment planner sets the additional VMAT plans. Across the 60 plans for each of the
priority for both types. The shape of the user-defined NTO 10 cases, NTO priority values of 1, 50, 100, 200, 500, and 999
curve is described by 4 parameters:

a
J. A. Alakuijala and K.M. Pesola (Google Patents, 2008).
1) distance from planning target volume (PTV) border b
Varian Medical Systems, Inc. Eclipse Photon and Electron Algo-
(xstart), rithms Reference Guide. P1015026-001-A; 2015.
ARTICLE IN PRESS
J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■ 3

were investigated for both the automatic and user-defined descending order of importance: conformity index (CI), ratio
NTOs. All user-defined NTO curves had an xstart of 1 mm, an of the 50% prescription isodose volume to PTV (R 50% ),
f0 of 100%, and an f∞ of 10%, whereas k varied from a gradual maximum dose 2 cm away from a PTV (D2cm), V20Gy, and
fall-off of 0.01 mm−1 to a much steeper fall-off of 2.0 mm−1. maximum PTV dose (PTVmax). Monitor units (MUs) were also
Plan optimization was simple and easy to implement. Only recorded. Statistical significance with respect to differ-
the NTO and the PTV were penalized. Besides the NTO pri- ences between prior plans and NTO plans was evaluated
ority and fall-off settings described earlier, other NTO settings using the Wilcoxon signed-rank test.
were held constant. The PTV was penalized using a lower
objective with a priority of 100 set at the prescription dose Results
to cover 100% of the PTV. An upper PTV objective with a pri-
ority of 50 was set to limit the maximum dose to 140% of In the current study, a broad range of NTO settings were
the prescription. No constraints were placed on any of the investigated for 10 lung SBRT cases to characterize the effects
contoured organs at risk. All plans were optimized in auto- on plan quality metrics. PTVs ranged from 5.0 to 15.1 cc and
matic mode with an intermediate dose calculation using the were relatively far from critical structures. Figure 1 shows
Photon Optimization algorithm in Eclipse, version 13.6 NTO curves from the investigated NTO fall-off values. The
(Varian Medical Systems Inc., Palo Alto, CA). No ring struc- dose profile from a representative original plan is also plotted
tures were used for optimization. Plans were neither paused for reference. NTO priority values of 1, 50, 100, 200, 500, and
during optimization nor reoptimized following the final dose 999 were studied for each of these curves.
calculation. The final calculation was performed using the NTO settings were shown to have a substantial effect on
anisotropic analytical algorithm with tissue heterogeneity the dose distribution and corresponding dose metrics. For
corrections. each metric, mean values were calculated across the 10 cases.
Before plan quality analysis, plans were normalized so that Figure 2 shows the mean CI, which ranged from 1.00 to 1.54.
the prescription dose covered 95% of a PTV while also en- CI approached 1.00 using an NTO priority value greater than
suring that 90% of the prescription dose covered at least 99% 1 and a fall-off value of ≥0.1 mm−1. An automatic NTO with
of a PTV. Physicians gave consensus recommendations on priorities of 100 and 200 also resulted in a CI approaching
the rank order importance of plan quality metrics: first to 1.00.
achieve adequate coverage of the PTV and then to mini- Figure 3 shows a plot of mean R50% values. R50% ranged from
mize high-dose spillage, low-dose spillage, and the lung 3.95 to 7.57. The automatic NTO performed poorly at reduc-
volume receiving ≥20 Gy (V20Gy). More specifically each plan ing R50% to meet RTOG 0813 protocol guidelines.11 Only 1
was evaluated based on the following criteria listed in automatic NTO did not result in a minor or major

100

90

80 NTO Fall-off
0.01
70
0.05
0.10
60
0.15
Dose [%]

50 0.20
0.30
40 0.50
1.00
30 5.00
OP
20

10

0
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5
Distance from PTV [cm]

Fig. 1. The NTO function defines the desired dose fall-off away from a PTV. In the present study, SBRT dose distributions are characterized as a func-
tion of NTO fall-off and priority. (Color version of figure is available online.)
ARTICLE IN PRESS
4 J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■

1.6

NTO Priority
1.5
1

Deviation
Minor Deviation
1.4 50
Conformity Index

100

Minor
1.3
200
1.2
500

Deviation
No Deviation
1.1 999

No
1

0.01 0.1 1 Automatic NTO


NTO Fall-off

Fig. 2. Conformity index. Each data point is the mean value from 10 cases (RTOG 0813 deviation thresholds are marked with dashed lines.). (Color
version of figure is available online.)

deviation, and it was outperformed by all user-defined NTOs A plot of mean PTVmax dose values is shown in Fig. 7. The
of the same priority with fall-off values of ≥0.1 mm−1. PTVmax dose ranged from 118% to 175%. In general, the au-
Mean D2cm values can be seen in Fig. 4. D2cm ranged from tomatic NTO performed worse than the user-defined NTO
33.4% to 67.9%. Lowest values of D2cm were obtained with a as demonstrated in Figs. 2 to 6 and yielded lower PTVmax
fall-off set to 0.15 mm−1 and a priority of 500. As with R50%, values as seen in Fig. 7. As NTO priority and fall-off values
the automatic NTO typically performed more poorly than the increased, so did PTVmax. For two of the NTO setting com-
user-defined NTO. binations, the recommended upper limit of 167% was
A plot of mean MU values is shown in Fig. 5. MUs per cGy exceeded.
ranged from 1.81 to 4.69 cGy−1. The total number of MUs from Although no combination of NTO settings were optimal
all fields was divided by the nominal daily dose. A higher for all of the dose metrics, an NTO with a fall-off value of
NTO priority typically reduced the number of MUs. 0.15 mm−1 and a priority of 500 best satisfied the rank order
Mean lung V20Gy values ranged from 1.66% to 2.75%. Results criteria. Median results from cases with well-tuned NTO set-
are plotted in Fig. 6. In general, lung V20Gy decreased as NTO tings (priority of 500, fall-off of 0.15 mm −1 ) were then
priority increased, although the decrements were relative- compared with the median results from prior plans of the
ly small. In all cases, lung V20Gy was well below RTOG 0813 same 10 cases. Statistical significance was assessed with the
minor and major deviation limits of 10% and 15%. Wilcoxon signed-rank test. Comparisons of the median

7.5 NTO Priority


Deviation
Deviation
Major

1
Major

6.5 50

100
6
R50

Deviation
Deviation

200
Minor
Minor

5.5

500
5
999
Deviation

4.5
No

3.5
0.01 0.1 1 Automatic NTO
NTO Fall-off

Fig. 3. R50. Each data point is the mean value from 10 cases (RTOG 0813 thresholds are marked with dashed lines.). (Color version of figure is avail-
able online.)
ARTICLE IN PRESS
J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■ 5

70

Deviation
Deviation
Major
Major
65 NTO Priority
1
60
50

Deviation
Minor
55
100
D2cm [%]

50
200

45
500

Deviation
No
40 999

35

30
0.01 0.1 1 Automatic NTO
NTO Fall-off

Fig. 4. D2cm. Each data point is the mean value from 10 cases. (RTOG 0813 thresholds are marked with dashed lines). (Color version of figure is
available online.)

results from the 10 well-tuned NTO and prior cases were as Discussion
follows: CI of 1.02 vs 1.03 (p = 0.133), R50% of 4.00 vs 4.35
(p = 0.002), D2cm of 34.5% vs 36.5% (p = 0.131), V20Gy of 1.22% In the present study, the dosimetric effects of NTO set-
vs 1.32% (p = 0.006), MUs of 2.76 cGy −1 vs 2.88 cGy −1 tings were characterized for lung SBRT. VMAT optimization
(p = 0.131), PTVmax of 138% vs 122% (p = 0.002), spinal cord in Eclipse was simplified, consisting of the NTO and lower
max of 3.78 Gy vs 3.57 Gy (p = 0.322), and heart max of and upper constraints on a PTV. No rings or other struc-
5.22 Gy vs 6.64 Gy (p = 0.313). In all cases, both the well- tures were penalized in the optimization. Following
tuned NTO and prior plans comfortably satisfied RTOG 0813 institutional criteria for plan evaluation, a well-tuned NTO
guidelines.11 had a fall-off of 0.15 mm−1 and a priority of 500. This com-
Figure 8 depicts the isodose distributions from plans gen- bination of NTO settings achieved high conformity,
erated with varying fall-off values, all with an NTO priority minimized low-dose spillage, and required relatively few
of 500. Note that plans with more isotropic isodose distri- MUs. When compared with median results from the 10 prior
butions have relatively lower levels of low-dose spillage, as cases, well-tuned settings of the NTO outperformed prior
shown by the curve height outside the PTV. A profile of the cases with respect to achieving lower R50% and lung V20Gy. The
PTV is drawn at the prescription dose level for reference. results (R50% of 4.00 vs 4.35, p = 0.002, lung V20Gy of 1.22% vs

4.5 NTO Priority


1
4
50
Monitor Units/cGy

3.5 100

200
3
500
2.5
999

1.5
0.01 0.1 1 Automatic NTO
NTO Fall-off

Fig. 5. Monitor units. Each data point is the mean value from 10 cases (lower monitor units = less complex MLC shaping). (Color version of figure is
available online.)
ARTICLE IN PRESS
6 J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■

16

14 NTO Priority
1
12
50
Lung V20Gy [%]

10
100
8
200

6
500

4 999

0
0.01 0.1 1 Automatic NTO
NTO Fall-off

Fig. 6. V20Gy. Each data point is the mean value from 10 cases (RTOG 0813 thresholds are marked with dashed lines.). (Color version of figure is
available online.)

1.32%, p = 0.006) were statistically significant. Although it is well-tuned NTO settings in this study achieved R50% values
possible that prior plans could be improved by using dif- within the RTOG 0813 protocol that were significantly lower
ferent rings or penalties, NTO plans had the advantage of than R50% values from the prior cases utilizing the same
easier implementation as compared with ring-based plan- noncoplanar arc beam orientations. Similar trends were found
ning; the 4 NTO parameters were defined within seconds. when comparing D2cm values, another metric of low-dose
All plans were well within RTOG 0813 guidelines. spillage, although results were not statistically significant.
Low-dose spillage is an important concern with SBRT. R50% The mean PTVmax was higher for NTO plans compared with
is a common metric for quantifying low-dose spillage in lung prior plans: 138% vs 122%, p = 0.002. The upper objective
SBRT.12 The R50% primarily describes intermediate dose spill- penalty for the target was set at 50, whereas the objective
age that is absorbed by healthy tissue beyond the target itself was 140% of the prescription dose for each of the cal-
border. Patients undergoing SBRT of the lung may have pre- culated NTO plans in the present study. Typically, planning
existing pulmonary limitations; therefore, relatively small strategies that obtain sharper dose gradients often result in
changes in R50% could cause significant negative effects with a higher maximum dose inside of a PTV. A higher PTV dose
respect to pulmonary function.12 Acceptable values of the may be perceived as beneficial as the highest dose in the
R50% are determined by the PTV volume and have been center of the field could coincide with a hypoxic and there-
defined in table 1 of RTOG 0813.11 Plans generated with fore more radioresistant tumor core.13,14 All prior plans

180

170 NTO Priority


1
-60%
90% -60%

160
50
PTV Max Dose [%]

surface: 90%

150 100
isodose surface:

200
140
Prescription isodose

500
130
Prescription

999

120

110
0.01 0.1 1 Automatic NTO
NTO Fall-off

Fig. 7. PTVmax (RTOG 0813 expected range is within dashed lines.). (Color version of figure is available online.)
ARTICLE IN PRESS
J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■ 7

A B

Fig. 8. (A) Dose line profiles of plans with priority set to 500 and fall-off values of 0.01 (red), 0.1 (yellow), 0.15 (green), 0.5 (cyan), 1.0 (purple), and
5.0 (magenta). (B) Axial view of isodose line distributions of plans with priority of 500 and a fall-off, from top left: 0.01, 0.1, 0.15, 0.5, 1.0, and 5.0.
(Color version of figure is available online.)

achieved a PTVmax well below the RTOG 0813 upper limit, additional research is needed to determine optimal NTO set-
although perhaps with the trade-off of relatively worse low- tings across the range of tumor sizes and tumor locations
dose spillage.11 In contrast, certain NTO plans exceed the seen clinically.
upper limit, although most were within the acceptable vari- The NTO has been commercially available for some time;
ation range.11,14 however, the authors do not know of any previously pub-
Results show an interplay effect between PTV dose and lished work on optimal NTO settings for lung SBRT. A
low-dose spillage. By tuning NTO settings and allowing a previous study has investigated SBRT planning with the use
maximum PTV dose of approximately 140%, low-dose of ring structures.14 The current study investigates a broad
spillage—D2cm in particular—was minimized. As shown in range but not an exhaustive set of NTO parameters. The
Fig. 8, a plan with these settings had relatively symmetric current study explores how the calculated dose distribu-
isodose lines forming shells around the PTV. Whereas sym- tion is affected by NTO priority and fall-off, parameters that
metric dose distributions appear to correspond to reduced respectively describe the steepness of the dose gradient and
low-dose spillage and may be desirable in many cases, asym- the importance of achieving that gradient in relation to other
metric dose distributions might be beneficial for sparing a optimization constraints, i.e., PTV coverage. The param-
nearby critical structure. In such cases, the NTO alone may eters were investigated, given their conceptual relationships,
not be adequate for meeting normal tissue constraints. along with goals of achieving steep dose gradients beyond
The cases evaluated in the present study had relatively a PTV and adequately covering a PTV with the prescribed
small PTVs, which were relatively far from any critical struc- dose. A broad array of priority and fall-off combinations were
tures. Tumor size is related to SBRT treatment outcomes.15 studied; however, other NTO parameters were held con-
A recent study comparing T1 and T2 lung tumors found that stant. The parameters xstart and f∞ were set at 100% and 10%
larger tumors were associated with higher local recur- of the prescription dose, whereas xstart was 1 mm. The mo-
rence and reduced survival rates.15 Additionally, tumor tivation for these choices was to deliver full dose to a PTV
location influences SBRT planning techniques and the dose while penalizing much of the low-dose wash. Limiting the
to healthy tissues. RTOG 0813 proposes that dose-volume extent of any high-dose spillage to a small rind around a PTV
limits be used for tissues located in close proximity to the was an additional concern. Although beyond the scope of
PTV.11 Although NTO tuning could be a useful strategy for the present study, these NTO parameters could be tuned in
dose escalation or greater sparing of normal tissues, combination with the fall-off and priority. Furthermore, other
ARTICLE IN PRESS
8 J.P. Bell et al. / Medical Dosimetry ■■ (2017) ■■–■■

plan evaluation criteria could be considered, e.g., the unified 5. Benedict, S.H.; Yenice, K.M.; Followill, D.; et al. Stereotactic body
dosimetry index, which integrates the CI, the coverage index, radiation therapy: The report of AAPM Task Group 101. Med. Phys.
37(8):4078–101; 2010.
the homogeneity index, and the dose gradient index all into
6. Klein, E.E.; Hanley, J.; Bayouth, J.; et al. Task Group 142 report: Quality
1 score.16-18 Optimal NTO settings may also depend on the assurance of medical accelerators. Med. Phys. 36(9):4197–212; 2009.
beam arrangement. The NTO is limited to 1 dimension, 7. Ong, C.L.; Verbakel, W.F.; Cuijpers, J.P.; et al. Stereotactic radiotherapy
whereas the 3-dimensional dose distribution depends for peripheral lung tumors: A comparison of volumetric modulated
arc therapy with 3 other delivery techniques. Radiother. Oncol.
strongly on the beam arrangement. Further investigation is
97(3):437–42; 2010.
needed to determine optimal NTO settings across the many 8. Timmerman, R.; Galvin, J.; Michalski, J.; et al. Accreditation and
possible treatment planning applications. quality assurance for Radiation Therapy Oncology Group: Multicenter
clinical trials using stereotactic body radiation therapy in lung cancer.
Acta Oncol. 45:779–86; 2006.
Conclusions 9. Mulryan, K.; Leech, M.; Forde, E. Effect of stereotactic dosimetric end
points on overall survival for Stage I non-small cell lung cancer: A
Lung SBRT dose metrics were characterized across a critical review. Med. Dosim. 40:340–6; 2015.
10. Sangroh, K.; Tzu-Chi, T. Off-target-isocentric approach in non-
broad array of NTO priority and fall-off settings. The range
coplanar volumetric modulated arc therapy (VMAT) planning for lung
of dosimetric results, some unacceptable and many less SBRT treatments. J. Radiosurg. SBRT 3(3):215–24; 2015.
than optimal, demonstrates the importance of selecting 11. Bezjak, A.; Paulus, R.; Gaspar, L.E.; et al. Efficacy and Toxicity Analysis
appropriate NTO settings and the exponential effects of of NRG Oncology/RTOG 0813 Trial of Stereotactic Body Radiation
the fall-off parameter (k). An automatic NTO is not recom- Therapy (SBRT) for Centrally Located Non-Small Cell Lung Cancer
(NSCLC). Int. J. Radiat. Oncol. Biol. Phys. 96(2):S8; 2016.
mended for lung SBRT planning because of poor performance
12. Lim, D.H.; Yi, B.Y.; Mirmiran, A.; et al. Optimal beam arrangement
at reducing low-dose spillage. Plans with well-tuned NTO for stereotactic body radiation therapy delivery in lung tumors. Acta
settings (priority of 500, fall-off of 0.15 mm−1) compared Oncol. 49(2):219–24; 2010.
favorably with prior plans created with ring structures, 13. Rasey, J.S.; Koh, W.J.; Evans, M.L.; et al. Quantifying regional hypoxia
in human tumors with positron emission tomography of [18F]
achieving significantly lower R50% and lung V20 values.
fluromisonidazole: A pretherapy study of 37 patients. Int. J. Radiat.
Based on the ease of NTO planning and the corresponding Oncol. Biol. Phys. 36(2):417–28; 1996.
results found in the present study, when well-tuned, the 14. Sonier, M.; Chu, W.; Lalani, N.; et al. Implementation of a volumetric
NTO is a viable and powerful treatment planning tech- modulated arc therapy treatment planning solution for kidney and
adrenal stereotactic body radiation therapy. Med. Dosim. 41:323–8;
nique for lung SBRT.
2016.
15. Dunlap, N.E.; Larner, J.M.; Read, P.W.; et al. Size matters: A
References comparison of T1 and T2 peripheral non-small-cell lung cancers
treated with stereotactic body radiation therapy (SBRT). J. Thorac.
1. Timmerman, R.D.; Kavanagh, B.D. Stereotactic body radiation therapy Cardiovasc. Surg. 140(3):583–9; 2010.
for lung cancer, Ch. 17. In: Roth, J.A.; Cox, J.D.; Hong, W.K.; eds. Lung 16. Lindblom, E.; Antonovic, L.; Dasu, A.; et al. Treatment fractionation
Cancer. 3rd ed. Oxford: Blackwell Publishing Ltd.: 256–70; 2008. for stereotactic radiotherapy of lung tumours: A modelling study of
2. Timmerman, R.; Paulus, R.; Galvin, J.; et al. Stereotactic body radiation the influence of chronic and acute hypoxia on tumour control
therapy for inoperable early stage lung cancer. JAMA 303:1070–6; probability. Radiat. Oncol. 9:149; 2014.
2010. 17. Krishnan, J.; Shetty, J.; Rao, S.; et al. Comparison of Rapid Arc and
3. Falkson, C.B. Guideline for radiotherapy with curative intent in intensity-modulated radiotherapy plans using unified dosimetry
patients with early-stage medically inoperable non-small-cell lung index and the impact of conformity index on unified dosimetry index
cancer. Curr. Oncol. 24(1):e44–9; 2017. evaluation. J. Med. Phys. 42:14–7; 2017.
4. Khan, F.M.; Gibbons, J.P. Khan’s the Physics of Radiation Therapy. 5th 18. Akpati, H.; Kim, C.; Kim, B.; et al. Unified dosimetry index (UDI): A
ed. Philadelphia, PA: Lippincott Williams & Wilkins/Wolters Kluwer; figure of merit for ranking treatment plans. J. Appl. Clin. Med. Phys.
2014. 9:2803; 2008.

You might also like