Professional Documents
Culture Documents
EC 350 User Manual: Release R100.1
EC 350 User Manual: Release R100.1
Release R100.1
Table of Contents
Chapter 8: Maintenance..............................................................................................................133
Temperature Probe Measurement Kits.............................................................................................. 134
Transducer Replacement Kits............................................................................................................ 134
Redundant Uncorrected Switch.......................................................................................................... 136
Installing Measurement Canada (MC) Sealing Cover........................................................................136
Removing and Re-Installing Human Factor (HF) Cover.................................................................... 137
Honeywell Process Solutions 1860 W. Rose Garden Lane Phoenix, AZ 85027 USA 1-800 822-7673
Chapter
1
About this document
Topics: This document is for operators and technicians working in the
natural gas industry. This document provides an overview of EC
• Terms and definitions 350 and instructions to install EC 350 on any meter equipped with
an instrument drive capability (common to diaphragm and turbine
meters) using a Universal Mounting Bracket (UMB) or a rotary mount.
To be able to perform the tasks in this document, you must have
knowledge about the physical properties of gas (volume, pressure,
temperature, and supercompressibility).
About this document
Term Definition
Meter gasket It is used as an environmental seal for mounting EC
350 on a meter.
Opto-isolators Devices that prevent unwanted current flow or
possible damage from high voltage or/and from
external devices connected to the instrument.
Rivets Used to install the slides on the index.
Slide A small piece of plastic used for covering the digits
on the index.
Spline wrench. Used to loosen and tighten set screw in order to
move the Thimble gear assembly UP for clock- wise
(CW) direction or DOWN for counter clock-wise
(CCW) direction.
Thimble gear assembly Rotates the digits on the index in CW direction when
it is set UP or CCW when it is set DOWN.
2
About EC 350
Topics: This section includes information on the main interfaces of EC 350.
These are LCD, Keypad, and External connections. It also lists the
• Safety instructions safety instructions that must be remembered while installing and
• Main interfaces of EC 350 commissioning EC 350.
About EC 350
Safety instructions
EC 350 complies with the general safety standards and regulations. However, failure to operate EC 350 as per
the safety instructions available in this document may lead to hazards.
EC 350 is approved for use in hazardous areas (Class I Division 1 or Class I Division 2). Different versions of
EC 350 are available depending on the operating conditions. The permitted operating conditions are marked
on each EC 350. Ensure to check the operating conditions of EC 350 before installing. You must be familiar
with the currently applicable electrical installation standards and regulations before installing and operating EC
350 in hazardous areas.
Warning: The warning denotes an explosion hazard. Ensure to follow the instructions described in
the warning.
To avoid explosion hazard, ensure to note the following:
• Install EC 350 depending on the operating conditions permitted for that particular EC 350 unit.
• EC 350 device for Class I Division 1 is intrinsically safe and must be connected to other circuits as per the
installation drawing specified on each EC 350.
• EC 350 device for Class I Division 2 must be connected to other circuits as per the installation drawing
specified on each EC 350.
• Substitution of components may impair suitability for use in a hazardous location.
Caution: The caution warns you of possible damage to property and provides instructions to avoid
damage to EC 350.
Honeywell recommends you to observe the warning information described in this document and other generally
applicable safety rules.
No warranty claims can be asserted if there is an unauthorized interference with the device.
Things to remember
You are allowed to use EC 350 in hazardous areas, under some permitted operating conditions. Ensure to
comply with the applicable laws and regulations, and company policies for the usage of EC 350.
EC 350 must be installed and commissioned only by specially trained and qualified staff. The device is designed
in accordance with the IP 65 degree of protection as per EN 60529. The installation of the intrinsically safe
circuits must comply with the applicable local laws or regulations. Operate EC 350 only if the instrument is
completely intact.
The service, maintenance and troubleshooting of EC 350 device operating in hazardous areas must be
performed only by specially trained and qualified staff.
To maintain acceptability of use in hazardous locations, you must use replacement batteries supplied by
Honeywell. For further information about changing the EC 350 battery pack, refer to the section, “Installing or
replacing battery”
• External connections
LCD
EC 350 provides a ten character, configurable, alphanumeric LCD display with icons to display the status
information and alarm conditions. The LCD display can be configured to on or off at different times of day. By
default it is ON all the time. During normal operation (Correction mode), the corrected volume is displayed on
the LCD. To conserve battery power, after each processing cycle the EC 350 keypad is locked, if there is no
input received within the time out period (1 to 10 minutes). Following is an LCD display illustration, showing
all segments ON.
Icon Description
Heart Beat : flashes every 3 seconds to indicate
normal operation (Correction mode).
Icon Description
Pulse output channel indicator : indicates the
pulse output channel that is active.
Keypad
The keypad is used for scrolling through the menu options. The following table lists the keys.
Option Description
Use these keys for unlocking the keypad and activating the
display. To unlock the keypad, press and hold ESC and UP arrow
at the same time until display text appears.
Use these keys for scrolling up and down in a menu. You can also
use to enter input by increasing and decreasing a value at the
current position.
Use these keys for navigating to the next and/or previous digit.
Use these keys to move the cursor one character at a time, in
forward or backward directions.
Use this key to access the main menu or display the submenu of
the current menu. It is also used for accepting an input value.
Use this key to display the Home screen, cancel an entry, or go
back to the previous menu.
External connections
The external connections from EC 350 are:
• Pulse outputs
• Serial port
• Alarm outputs
• External supply voltage (TB1)
• Battery pack (P5)
3
EC 350 installation
Topics: This chapter deals with the installation of EC 350 on conventional gas
meters (such as rotary, diaphragm, and turbine). This chapter also
• EC 350 contents verification provides procedures for installing and replacing the battery and output
• Overview of installation wire connections to obtain pulse output from EC 350.
• Model number interpretation
• Prerequisites
• Installing EC 350 on
conventional diaphragm, rotary
or turbine gas meters
• Installing EC 350 on a rotary
mount
• Power Supply Options
• General Wiring connections
• Installation Drawings
EC 350 installation
Overview of installation
The EC 350 installation consists of mounting and wiring EC 350 according to the instructions given in this
section. Before proceeding, read the installation information provided in this section and refer to the section
“Model number interpretation ”, to familiarize yourself with the EC 350 model you have selected.
Based on the mounting selected, refer to one of the following two methods for installing EC 350 on a meter.
• Installing EC 350 on conventional diaphragm, rotary or turbine gas meters. The following image illustrates
with a UMB.
• Installing EC 350 on a rotary mount. The following image illustrates EC 350 configured for a rotary mount.
(Adapter plate, which varies by meter model, not shown.)
Prerequisites
Ensure that the following components are installed and connected inside EC 350, before installing EC 350
on meters.
• Battery pack
• Pressure transducer (if ordered)
• Temperature probe (if ordered)
The following image illustrates the labels applied to the UMB, slides, and rivets.
Next steps
Verify if the digital index reading rotates in the proper direction. If not, perform the tasks described in the
following section, “ Changing the drive rotation ”, to position the bevel-gear thimble of EC 350 to ensure the
reading rotates in the proper direction.
Use the base plate to rotate EC 350 about 360 degrees in any four directions. To rotate EC 350, remove all
four screws, which attach the base plate to the bracket housing.
2. Use the temperature probe as a measuring device, and insert the probe until it bottoms out from the
thermowell of the meter as illustrated in the following image.
3. Mark the temperature probe cable at the point where the cable comes out from the meter body as
illustrated in the following image.
4. The temperature probe is held in place with a rubber gland in the mounting plate. Push the probe into EC
350 until the mark is one inch from the mounting plate, as illustrated in the following image.
The one inch additional length ensures that the probe bottoms out at the end of the thermowell.
5. Insert the temperature probe into the thermowell as illustrated in the following image.
The temperature probe is mounted into the meter base.
LMMA 7m to 22-2090
16m mounting kit
with snap seals.
LMMA 7m to 22-2090-1
16m mounting kit
with cross-drilled
screws
LMMA 7m to 22-2090-2
16m mounting
kit with McGARD
screw, PG and E
key
B3 or TQM 22-2108-1
mounting kit with
cross- drilled
screws.
B3 or TQM 22-2108-2
mounting kit with
McGARD screw,
PG and E key
Installations requiring CLASS 1 - DIV 1 approval, only the Battery Pack 40-6064 is approved. Approval
Pending.
When operating from an externally supplied DC power source, configure the External Supply Low Alarm Limit
(Item 795) to be a value greater than 5.0 and generally less than 7.0 V. The default value of Item 795 is -1.0
to effectively disable the Alarm Item 796. The External Supply voltage is measured on a 10 minute interval.
Three consecutively low readings are required to the trip the Low Voltage Alarm (Item 796) – this is done to
help guard against falsely setting alarm due to a power supply glitch.
Battery Powered
Connections for the Battery pack are made at the P5 connector on the EC 350 IO Board. There are three
Battery choices for operating the EC 350 from battery power:
• 40-6050 (4-cell Alkaline) – 5 year operating life under specified conditions
• 40-6048 (2-cell Lithium)
• Dual set of 40-6048 (2-cell Lithium) – extended life or heavy usage / comms applications.
Use the Battery Type (Item 1061) to select the type of Battery pack that is being used to power the
instrument.
• 0= 4 Cell Alkaline (default type)
• 1= 2 Cell Lithium
• 2= 4 Cell Lithium (dual set of p/n 40-6048).
The Battery voltage is measured on a 10 minute interval. Three consecutively low readings are required to
the trip the Low Voltage Alarm (Item 99) – this is done to help guard against falsely setting alarm due to a
supply glitch.
Note: The Low Battery Voltage Alarm (Item 99) will not set if an External Supply voltage is present
at TB1 with a voltage level greater than the Battery voltage by over 1.0 VDC. When a Low Battery
Voltage Alarm is tripped (set), an Alarm Log record entry is generated along with updates to timestamp
Items 462 and 463.
condition in which volume continues to be accumulated and event logging and alarms are still active, but all
other functions are stopped to ensure very low power drain. This allows a low power battery (40-6054) to be
plugged in while the main battery pack is changed, and then removed after the new battery is plugged in.
To replace the battery in a hazardous DIV-1/ZONE-0 environment, perform the following steps:
1. Enter HMI and set the EC 350 to Battery Change mode. You can use the L2.11 or L3.13 HMI menus.
2. The BATT CHNG message is displayed on the LCD screen. Press the Enter key.
3. Select Yes and press Enter. The unit is now ready for a battery change out. CHNG BATT THEN ESC
is dispalyed on the LCD screen.
4. Plug the special battery change battery (part no. 40-6054) into the open connector on the battery Y cable.
5. Unplug the old battery.
6. Plug in the fresh battery where the old battery was plugged in.
7. Remove the special battery change battery.
8. Press the ESCkey twice to exit the battery change mode. When the ESCkey is pressed, all the battery
life items are automatically set to 100% new values and the battery usage cycle is set to zero.
While in low power shutdown mode, the LCD displays the following (except if user enters HMI mode):
__________
“REPLACE”
“BATTERY”
Toexit Low Power Shutdown, the Battery voltage must be above the Item 50 limit value and the user must enter
HMI mode. User enters HMI and menus down arrow to the screen showing: SHUTDOWN - HMI menus: L2.10
or L3.12. Next, the User presses Enter key and display will show: ‘ENTER MODE’ as first selection. Arrow
down to ‘EXIT MODE’ selection. User again presses the Enter key to confirm – the unit leave (exit) Shutdown/
Shelf mode. When the User exits the HMI menus – the LCD will then return to normal default screen.
Refer to the table below to see which functionality is limited or blocked in low battery shutdown mode.
Refer to the table below to see which functionality is not blocked in shutdown mode.
1. Unlock the keypad and type the PASSCODE to enter level 2 mode or level 3 mode Refer to the section
"Accessing level 2 mode” for accessing level 2 mode or section “Access level 3 mode ” for accessing
level 3 mode.
2. By default, the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
SHUTDOWN appears.
4. Press OK. The following appears on the display.
ENTER MODE
5. Press OK to confirm.
EC 350 enters shutdown/shelf mode.
1. Establish a serial communication between EC 350 and MasterLinkSQL. Refer to the MasterLinkSQL
User’s Guide for information about establishing a serial connection between EC 350 and MasterLinkSQL.
2. In the MasterLinkSQL window, click Instrument > Shutdown.
3. Click Yes.
Refer to the table below to see which functionality is not blocked in User Shelf mode.
An alarm pulse (on channel C or D) is generated when EC 350 enters an alarm condition. Use channel C (NO
or Normally Open) or channel D (NC or Normally Closed) depending on the AMR device accepting the alarm
pulse. Use of channel D (NC) allows for a cut wire to be interpreted as an alarm condition to the AMR device.
EC 350 provides three Form-A pulse outputs (channels A, B, and C) for volume output pulses.
Volume pulses can correspond to either uncorrected volume (as counted in item 2) or corrected volume (as
counted in item 0). The type of volume pulse (and whether the channel is enabled) is configured through items
93-95.
The "weight" of each pulse (e.g. whether a pulse corresponds to 10 cubic feet or 1 cubic meter) is configured
via items 1193-1195. (Note: in prior Mercury products the pulse weight was configured differently through items
93-95. These are now read-only values provided for reference.)
Pulse output timing is configurable via items 1014, 1015, and 1024. Various pulse ON and OFF times can be
chosen. Default is a pulse ON time of 62.5 milliseconds and a pulse OFF time of at least 62.5 milliseconds,
for a minimum pulse period of 125 milliseconds.
Pulse queue = 2
x the number of
Channel A output
pulse remaining
from the previous
wakeup period.
Pulse queue = 2
x the number of
Channel B output
pulse remaining
from the previous
wakeup period.
Pulse queue = 2
x the number of
Channel C output
pulse remaining
from the previous
wakeup period.
Note: *If channel C is used as an alarm output, the pulse width is fixed at 100 milliseconds.
Installation Drawings
4
Securing the device
Topics: This chapter describes the different safety and security features of a
EC 350 device.
• Case
• Metrological protection modes
• Defining access privileges
• Metrological configuration
mode
• Validating setup configuration
Securing the device
Case
The case can be locked or tamper sealed with a padlock or seal on the door hasp.
Item classifications
Each item has a fixed classification – one of those below. Contact Honeywell for classification of each item.
These classifications pertain to the access restriction modes below.
Item 139 Type Item Change Event Logging Firmware Lock Icon
Restrictions Upgradable
2 Metrological OPEN items Only changes No Closed
Event Protection may be to EVENT items
modified. are logged.
SEALED items Changes to
may not be OPEN items are
modified. NOT logged.
EVENT
items may be
modified only
if the event
log is not full
of unread
records. (Only
log records
that have been
read can be
overwritten.)
(This mode
corresponds to
a Measurement
Canada Type B
Event Log.)
3 Metrological Only OPEN All item No Closed
Sealed items may be changes are
modified. event logged.
Item 139 Type Item Change Event Logging Firmware Lock Icon
Restrictions Upgradable
4 Metrological OPEN items Only changes No Closed
Event Locked may be to EVENT items
modified. are logged.
SEALED items Changes to
may not be OPEN items are
modified. NOT logged.
EVENT
items may be
modified only
if the event log
is not full of
changes since
put in protected
mode. Reading
Event Log has
NO affect on
allowing write
access. Once
event log is
full of changes
since it was put
in protected
mode (with item
139 change) no
more changes
to protected
items are
allowed until
item 139 is
set back to
zero (see
notes below).
(This mode
corresponds to
a Measurement
Canada Type A
Event Log.)
Attention: Firmware can be upgraded if either the override jumper is off OR if item 139 is set to
unrestricted.
The METROLOGICAL ACCESS JUMPER is located in the lower right corner of the IO Board at the back of
the case (labeled “METR JMPR”). If a sealing plate is in place that must first be removed.
1. Establish a serial communication between EC 350 and MasterLinkSQL (4.41 or above). Refer to the
MasterLinkSQL User’s Guide for information about establishing a serial connection between EC 350 and
MasterLinkSQL.
2. In the MasterLinkSQL window, click Find Item by Number icon.
5. Click Change.
The Change Item dialog box appears.
6. From the list, select Unrestricted and then click Save.
Event log records reference this user ID, to identify who made configuration changes.
Full use of multi-user features requires host software supporting the SS protocol command. For backwards
compatibility with host software that only supports the older SN sign on command (which does not support
multiple users), the user 0 passcode is required for sign on via the SN command. In that case all configuration
changes made through an SN sign on are attributed to user 0 in the event log records. The legacy item 196
(event log user ID) is not supported.
The screenshot below shows a typical User Table.
Valid user IDs are 0 through 99 (decimal numeric). Valid passcodes are 0 through 99999 (decimal numeric).
The following privileges can be granted or denied users:
• MODIFY OPEN ITEMS
• MODIFY EVENT ITEMS
• MODIFY SEALED ITEMS
• READ EVENT LOG
• USER TABLE DOWNLOAD
• HMI LEVEL 2 ACCESS
• HMI LEVEL 3 ACCESS
OPEN, EVENT, and SEALED are three classifications of items (the other classification is READONLY, which
cannot be modified with any privilege).
A user granted level 3 HMI access will enter level 3 HMI menus after entering their user ID and password into
the HMI (regardless of whether they are granted level 2 access). To allow a user to enter level 2, they must not
be granted level 3 access. A single user can either have level 2 HMI access or level 3 HMI access or neither,
but may not access both level 2 and 3.
5. Click Apply to save the privileges for the newly created user.
6. Click Save.
The Save User Table File As dialog box appears.
7. Enter a filename and click OK.
The Choose Password dialog box appears.
8. Enter a password and click OK. This password is required when you want to send the user table file.
The EC 350 ships in unsecure mode with default passwords. Change the default passcodes at least to
secure the device.
Note that, for convenient device access by users choosing not to secure their devices, MasterLinkSQL
will attempt to access user 0 with the default passcode and will only prompt the user if that fails.
5
Key features
Topics: This chapter describes the key features of a EC 350 device.
• Volume measurement
• Meter proving
• Supercompressibility Factor
• Item Description for
Supercompressibility factors
• Alarms
• Logging
• Display ON/OFF
Key features
Volume measurement
EC 350 generally operates in the Corrector mode. In this mode, EC 350 receives and processes uncorrected
volume, pressure, and temperature inputs to produce corrected volume information. While in Corrector mode,
totalized corrected volume information is continuously displayed on the display. EC 350 always returns to
Corrector mode after completing operations in any other mode. To enter the Corrector mode, from any other
mode, press the ESC key several times until the Corrector mode display appears. From every other mode, the
instrument returns to Corrector mode if no key is pressed for 1 to 10 minutes depending on the mode.
The following image illustrates a typical Corrector mode display.
Corrected volume
EC 350 accurately measures and maintains the total uncorrected volume passing through the meter on which
it is installed. In addition to the uncorrected volume, EC 350 also computes the corrected volume.
The corrected volume is computed by multiplying a total correction factor by the uncorrected volume input from
the meter.
EC 350 maintains items called Hi-Res (High Resolution) volumes for indicating a fractional part of the corrected
and uncorrected volumes that are not visible in the primary items (0 and 2).
Refer to the section “Items reference” for information about Corrected volume (item 000) and its corresponding
High resolution corrected volume (item 113) and Uncorrected volume (item 002) and its High resolution
uncorrected volume (item 892).
Total Correction Factor = Temperature Factor X Pressure Factor X Auxiliary Factor (normally 1.0) X Squared
Supercompressibility Factor.
Refer to the section “Items reference” for information about Total Correction Factor (item 043), Temperature
Correction Factor (item 045), Pressure Factor (item 044), Auxiliary Factor (item 046), and Squared
Supercompressibility Factor (item 116).
EC 350 supports a large range of units for measuring volume; some of them are as follows:
CF, CFx10, CFx100, CCF, CFx1000, MCF, CFx10000, m3x0.1, m3, m3x10, m3x100, and m3x1000.
You can independently configure corrected and uncorrected volume items.
Refer to the section “Items reference” for information about Corrected volume units (item 090) and Uncorrected
volume units (item 092) .
Energy
In addition to the corrected and uncorrected volume, EC 350 also calculates the energy (in item 140)
corresponding to the corrected volume, based on a fixed user-configurable conversion factor (item 142).
EC 350 supports a large range of units for measuring energy; some of them are as follows: Therms,
Decatherms, Mega Joules, Giga Joules, Kilo Calories, and Kilo Watts.
Refer to the section “Items reference” for information about Energy (item 140), Energy units (item 141), and
Gas energy value (item 142) .
High resolution
For direct-mount rotary configurations (not UMB), EC 350 supports an option called high resolution mode for
obtaining a higher degree of gas volume measurement resolution. Use the Inputvolume mode (item 433) for
selecting the high resolution mode option.
In high resolution mode, EC 350 measures the gas volume using a very low divider ratio for the input signals
supplied by the two volume input sensors. Generally, the divider ratio is much higher for making a sensible
trade-off between the measurement resolution and the instrument battery life.
The high resolution mode consumes more battery power while achieving a greater degree of volume
measurement resolution. You can use this mode as a test and verification tool for obtaining accurate volume
measurement within a shorter period of time (less gas passing through the meter).
Refer to the section “Items reference” for information about Input volume mode (item 433) .
Attention: Field applications do not use the high resolution mode for day-to-day volume
measurement, as the gain in volume resolution is not worth the inherent battery life drain.
Volume statistics
The EC 350 supports various gas volume statistical items. Following are the supported gas volume statistical
items.
• Daily Cor - Vol (item 223)
• Previous Day Cor - Vol (item 183)
• Daily Unc- Vol (item 224)
• Previous Day Unc- Vol (item 184)
• Max Day Cor - Vol (item 253)
• Max Day Cor - Vol Date (item 254)
Refer to the section “Items reference” for information about the gas volume statistical Items.
Temperature readings
EC 350 reads the gas temperature probe on a time-based approach, as opposed to reading temperature only
when it receives volume input.
Regardless of flow rate, the temperature of the gas is read, and the correction factors are computed, every 30
seconds. Hence, high and low temperature alarms are detected even under low and no-flow conditions.
While measuring a temperature, EC 350 samples the Gas temperature probe a large number of times
to produce a very accurate final reading. With each gas temperature measurement, both the temperature
correction and the total correction factors are updated to reflect the new reading.
Refer to the section “Items reference” for information about the following temperature items.
• Gas temperature (item 026)
• Gas temperature units (item 089)
Temperature statistics
EC 350 supports various gas temperature statistical items. Following are the supported gas temperature
statistical items:
• Interval high gas temperature (item 216)
• Interval low gas temperature (item 216)
• Daily average gas temperature (item 257)
• Previous day gas temperature (item 186)
• Max gas temperature (item 293)
• Max gas temperature date/time (item 294/295)
• Min gas temperature (item 297)
• Min gas temperature date/time (item 298/299)
Refer to the section “Items reference” for information about the gas temperature statistical items.
Pressure statistics
EC 350 supports various pressure statistical items. Following are the supported pressure statistical items:
• P2 Interval avgerage pressure (item 421)
• P2 Interval high pressure (item 422)
• P2 Interval low pressure (item 423)
• P2 Daily average pressure (item 424)
• P2 Previous daily average pressure (item 425)
• P2 Max pressure (item 426)
• P2 Max pressure time (item 427)
• P2 Max pressure date (item 428)
• P2 Min pressure (item 429)
• P2 Min pressure time (item 430)
• P2 Min pressure date (item 431)
• P1 Max pressure (item 285)
• P1 Max pressure date (item 287)
• P1 Max pressure time (item 286)
• P1 Min pressure (item 289)
• P1 Min pressure date (item 291)
• P1 Min pressure time (item 290)
• Prev day average P1 pressure (item 185)
• P1 Interval average pressure (item 206)
• P1 Interval high pressure (item 214)
• P1 Interval low pressure (item 215)
• P1 Daily average pressure (item 256)
Refer to the section “Items reference” for information about the gas temperature statistical items.
Fixed temperature
The EC 350 can be configured to function as a fixed temperature device by setting item 111 to fixed temperature.
In that mode, the temperature value configured in item 1162–Fixed temperature value (instead of the live
temperature at item 26) is used for temperature correction.
Pressure measurement
The pressure of the gas is read, and the correction factors are computed, every 30 seconds, regardless of flow
rate. Hence high and low pressure alarms are detected even under low and no-flow conditions.
When measuring gas pressure, the EC 350 uses an ultra-high-resolution analog-to-digital converter system
along with advanced characterization/linearization techniques, including temperature compensation, to
produce a very accurate final reading. The EC 350 Plug and Play Pressure Transducer stores unique
characterization information within each transducer, which is transferred automatically if the pressure
transducer is changed.
Fixed pressure
The EC 350 can be configured to function as a fixed pressure device by setting item 109 to fixed pressure.
In this mode, the pressure value configured in item 1161–Fixed pressure value (instead of the live pressure
at item 8) is used for pressure correction.
Meter proving
The public utility commissions or various governmental regulations may require you to check the accuracy of
the actual volume of gas flowing through a gas meter. Prover devices such as the Dresser Model 5 Prover
and the Elster/American Meter SNAP Sonic Nozzle Prover are used for verifying the accuracy of the volume
of gas flowing through a gas meter. Prover devices pass a known volume and temperature of gas through
the meter being tested. They compare their own standard volume to the volume reported by EC 350, which is
mechanically connected to the rotary gas meter.
EC 350 uses switch closure pulse counts for providing volume data to the Prover device. The meter bodies
and EC 350 are tested at various rates such as 10%, 50%, and 100% of rated capacity. You can connect EC
350 to a prover to test both uncorrected and temperature corrected volumes.
EC 350 has a special operating mode that makes meter proving process very easy and fast. This mode is called
Pushbutton Proving . No special configuration settings are required to use the Pushbutton Proving feature of
EC 350.
The proving kit includes the following components for connecting the EC 350 to a Dresser Model 5 Prover:
• EC 350 prover dongle
• Cable assembly (USB and Model 5 connectors)
• Magnetic coupler
The cable adapter assembly is used with an Elster/American Meter SNAP prover system. This cable adapter
is included in the 40-4913-1 prover kit along with all of the components in the 40-4913 base prover kit. The
following image illustrates the cable adapter for the SNAP prover system.
Proving dongle
The infrared prover dongle is a device that can be connected to the IrDA data port on EC 350. It receives
IR LED pulses from EC 350 and converts the appropriate pulses into electronic switch contacts. This allows
a rotary meter with the EC 350 to be tested as a unit without breaking the meter seal. The following image
illustrates an IR proving dongle.
• PULSE: A red light appears to indicate that the electronic switch to the Model-5 prover is closed.
• LINKED: A green light appears to indicate that EC 350 and the prover dongle are synchronized. This must
be ON when you are in the Pushbutton Proving mode.
• CPU OK: A yellow light appears to indicate that the prover dongle is powered and is operational. Ensure
that the yellow light is always ON during operation. If the light goes off, it indicates that the prover dongle
does not receive the 5V power supplied by the computer.
Attention: When EC 350 is in Pushbutton Proving mode, the IrDA Serial communications
are disabled. Before you enter the Pushbutton Proving mode, ensure that the prover
dongle is connected to EC 350.
Supercompressibility Factor
Gases actually behave in a manner slightly different from what the ideal gas laws indicate. This deviation
depends on the molecular composition of the gas and the specific gravity as well as the pressure and
temperature. Natural gas, for instance, compresses by a greater amount than that computed by Boyle’s law and
hence the term “supercompressibility” is used for this deviation. It is small at very low pressure, but becomes
substantial as the pressure increases. The EC 350 automatically applies the supercompressibility factor and
therefore the equation for total volume correction that EC 350 applies to metered volume is expressed as:
Vb = Vm * Fp * Ft * (Fpv)²
Where:
• Vb = gas volume corrected to base conditions
• Vm = measured volume from meter
• Fp = pressure factor
• Ft = temperature factor
• Fpv = supercompressibility factor determined from NX-19 or AGA-8
The EC 350 can calculate the supercompressibility factor according to the following standards:
• AGA-8
• Gross method 1
• Gross method 2
• Detail method
• NX-19
Note: * Items 8 and 26 are live readings (not user configurable). They are used for
supercompressibility calculations when pressure and/or temperature calculations are “live” per items
109 and 111, respectively. If pressure and/or temperature is fixed, items 1161 and/or 1162 (user
configurable fixed pressure and fixed temperature values, respectively) are used instead of items 8
and/or 26 in supercompressibility calculations.
Alarms
EC 350 monitors a list of parameters and compares their measured values to configurable limits to determine
if a fault condition has occurred. When any of these measured parameters has exceeded their respective limit,
EC 350 will create an Alarm.
EC 350 currently supports monitoring eight (8) parameters and supplies an Alarm Item for each. Shown below
are the eight Alarm Items.
Alarm Item Alarm Item No. Alarm Limit Item Alarm Limit Basic Operational
Item No. Description
Battery Low 99 Battery Low Limit 49 Battery life
remaining < set limit
Battery Cycles 100 Battery Months 1001 Battery life months
Alarm Remaining remaining < set limit
Volume Sensor 1 102 n/a n/a Internal tests
Volume Sensor 2 103 n/a n/a Internal tests
System Alarm 104 n/a n/a Internal tests
Pressure Low 143 Pressure low limit 11 Gas pressure < set
low limit
Alarm Item Alarm Item No. Alarm Limit Item Alarm Limit Basic Operational
Item No. Description
Pressure High 145 Pressure high limit 10 Gas pressure > set
high limit
Temperature Low 144 Temperature low 027 Gas temperature <
limit set low limit
Temperature High 146 Temperature high 028 Gas temperature >
limit set high limit
Flow Rate high 163 Flow rate high limit 164 Flow rate > set high
limit
Daily Cor volume 222 Daily cor volume 221 Daily CorVol > set
limit limit
P2 Press High 451 P2 Press High 455 Alarm is raised
Alarm Alarm Limit when P2 goes
above its high limit
P2 Press Low 452 P2 Press Low 456 Alarm is raised
Alarm Alarm Limit when P2 goes
below its low limit
Extern Supply Low 796 External Supply 795 Alarm is raised
Alarm Low Limit when external
supply voltage goes
below its low limit
Super Compress 1025 n/a - internal error n/a - internal error Any one of several
Alarm errors in super
comp calculation
process
Metro Cnfg Chng 1035 n/a - internal error n/a - internal error An attempt was
Alarm made to change
an item not
appropriate for
current metrological
state / jumper
P1 Transducer 1058 n/a - internal error n/a - internal error Any one of several
Alarm internal failures of
pressure transducer
P2 Transducer 1059 n/a - internal error n/a - internal error Any one of several
Alarm internal failures of
pressure transducer
Temp Probe Alarm 1335 n/a - internal error n/a - internal error Any of several
internal failures of
probe itself
Comms Login 1388 n/a - internal error n/a - internal error Attempt to login
Alarm serially with invalid
credentials
HMI Login Alarm 1392 n/a - internal error n/a - internal error Attempt to login to
HMI with invalid
credentials
Alarm Item Alarm Item No. Alarm Limit Item Alarm Limit Basic Operational
Item No. Description
Alarm Log Record 1409 n/a - internal error n/a - internal error if > 0, some number
CRC Alarm of corrupt records
were encountered
during record
search/read
Event Log Record 1410 n/a - internal error n/a - internal error if > 0, some number
CRC Alarm of corrupt records
were encountered
during record
search/read
Audit Log Record 1411 n/a - internal error n/a - internal error if > 0, some number
CRC Alarm of corrupt records
were encountered
during record
search/read
Vol Sensor-3 Alarm 1424 n/a - internal error n/a - internal error malfunction
indicated by volume
sensor 3
Vol Sensor-4 Alarm 1425 n/a - internal error n/a - internal error malfunction
indicated by volume
sensor 4
RBX (Report By Exception) is an alarm mode that is enabled via item 165. RBX controls the behavior of the
following alarms.
• Pressure 1 High
• Pressure 1 Low
• Temperature High
• Temperature Low
• Pressure 2 High
• Pressure 2 Low
• Flow Rate High
When enabled, the RBX function automatically clears an instrument alarm after an alarm parameter has
returned to its normal operating range, buffered by a user specified deadband (hysteresis). The deadband
values are specified in items 166 (Pressure 1), 167 (Temperature), 169 (Flow Rate), and 459 (Pressure 2).
When disabled, the alarms remain in their current state until the user manually clears it.
Logging
To configure logs:
1. Ensure that MasterLink is linked to the EC 350.
3. Drag and drop the desired items from the Select Items list over to the log configuration tree.
In order to remove an item from the log configuration tree, highlight it and press the ‘Delete’ key.
4. The order of the items in the Log Configuration list determines the order in which the values are stored
and thus the order in which they will later appear in a report. Items in the list can be ‘grabbed’ and moved
up and down to change the order.
5. You can save a configuration for future reference to a configuration file (*.cnfg) with the Save setup to
file button. It can later be retrieved with the Read setup form file button. Only *.cnfg files can be read;
item files (*.ie3) are not compatible.
6. You can right click on the time interval displayed in the Log Configuration tree, to configure the period
at which log records are recorded.
7. In the Allocate log memory tab, configure the percentage of available memory allocated to each log.
For each log, you can select the percentage of memory allocated, if fewer than 5 logs are desired the
percent of the unused ones must be set to zero.
Note: The sum total of memory allocated for all logs must be equal to 100%.
The smaller a log’s allocation, the less data it will be able to store. When the limit is reached, the oldest
records will be overwritten with the newer ones. The consequence of a particular allocation setting is
reflected in the number of Days worth of data as well as the Number of records that can be written before
overwriting will occur. Note: A log’s interval setting will impact Days, but not Numbers.
8. The Enable checkboxes may be used to suspend data collection for one or more logs. A log’s existing
records will not be affected when the log is disabled. The fact that a log is disabled will be reflected in the
Log Configuration window by the presence of a red ‘x’ at the top of that log’s item tree. In addition, the
area for that log in the bar graphic will be grayed and marked as ‘Disabled’
9. Finally, click Send to Instrument to push the log configuration to the EC 350 Device. A few seconds will
be required for the transfer. A window should appear indicating ‘Successfully Configured’.
the host. The process can be initiated from the Transfer menu or icon using the Read Audit Trail Data
window.
You must specify the number of logs to read and the amount of data to transfer. All configured and enabled
logs can be checked in the Log Selection frame. If a log is not configured it cannot be checked. The amount
of data is specified by date/time ranges. The simplest option is Since Last Download. The host will examine
its database for the most recent record for each log for this instrument and construct a command to the EC 350
for all data it has collected since then. Using this option each time will ensure that the host database contains
a complete set of audit trail records for each log in each instrument. There are three other options if a specific
data range of data is desired.
Note: Reading of audit trail records does not remove them from EC 350’s memory. They can be
read any number of times until they are overwritten, as explained above.
Once Masterlink finishes reading all data for all logs the user is given the chance to view all the downloaded
records. This is a useful practice to ensure that the transfer succeeded and that all the expected records were
read.
1. In the MasterLinkSQL window, click Reports and select one of the following options:
• Audit Trail Report (With Summary)
• Audit Trail Report (Without Summary)
• Audit Trail Report (Summary Only)
• Audit Trail Report (Daily Summary)
• Audit Trail Report (Monthly Summary)
Based on the option selected, a corresponding dialog appears allowing the user to specify the Audit Trail
data to view.
2. Click the Select Sites button to browse to the desired EC 350 unit by SITE ID/ Site name.
3. Click on the Date Range tab to bring up a control window to select date range display options for the
log report.
4. Click Preview to view the Audit Trail report, or click the Print to send the report to a specified printer.
Event logger
The function of the Event Logger is to record configuration changes and significant system events. Each record
includes:
• Date
• Time
• User ID
• Sequence Number (starts at 1, wraps from 65535 to 0)
• Before and After for changes made to instrument parameters.
Event Log capacity is 1024 records
See section Metrological Protection Modes for details of Event Log behavior in different item 139 modes.
Write a value of 19230429 to item 264 using Setup > Advanced>Raw Instrument Access >Raw Item Access
in MastgerLinkSQL.
First view the log records uploaded with MasterLinkSQL. Audit Trail record CRCs can be viewed through the
Audit Trail CSD Conversion feature in the File menu. That outputs a Comma Separated Value file that includes
the record CRC as transmitted by the device and also as computed by MasterLinkSQL of the record as stored
in the MasterLinkSQL database. Event Log record CRCs can be viewed through the Event Log Report. In the
report configuration, select the option to view the CRC values.
To view records in the HMI, select the corresponding menu item, select the desired log (Audit Trail 1 through 5,
or Event Log), enter a Sequence Number for a record (as seen in MasterLinkSQL). The corresponding record
will be shown with the sequence number, the CRC value (in decimal), and the date and time of the record.
(Other record values cannot be displayed via the HMI.
Compare the CRC value on the device display with the one in the Audit Trail CSD file or the Event Log report.
The UP and DOWN arrow keys can be used to show the previous or next record.
Display ON/OFF
The EC 350 display can be configured to turn on and off at specific times during the day.
You can configure this in the MasterLink software using the following itemcodes:
6
User access
Topics: This chapter provides an overiew on using a EC 350 device. It
includes information on how to work with HMI displays and how the
• Getting started with the keypad
connections must be done between EC 350 and MasterLink.
• Working with HMI
• Connecting to EC 350 via
MasterLinkSQL
• Working with MasterLinkSQL
• Working with PowerSpring
User access
2. Press the DOWN arrow to enter Scroll List/ Meter Reader mode or press OK to access any one of the
HMI keypad modes.
• Level 0 mode- It is the scroll list mode where the customer-configured list of items appear on the display.
No passkey is required to access this mode. After the keypad is active, use the UP arrow and DOWN arrow
to scroll through the items in this mode. The items in this mode are read-only.
• Level 1 read only mode- It is used for viewing the instrument settings and live parameters. You do not
require any passkey to access this mode. In this mode, the instrument functions are categorized into
11 submenus. By scrolling through the 11 submenus, you can view over 60 factory-defined instrument
parameters. The items in this mode are read-only.
• Level 2 mode- It is used for accessing the less sensitive parameters and configuration options of EC 350.
This mode provides six submenus. You can view and configure the items in this mode by scrolling through
the six submenus. A passkey is required to access this mode.
• Level 3 mode- It is used for accessing all the parameters, configuration, and functions of EC 350. This
mode provides nine submenus. A passkey is required to access this mode.
Level 0 mode
The level 0 mode is also known as the Scroll List Mode (or Meter Reader Mode). The following table lists the
factory-default items that appear on the display when you enter the level 0 mode (in the order in which they
appear as you press the DOWN arrow). This list is configurable. The Configurable item column identifies the
item number at which each list item is programmed.
Scroll list Displayed on LCD Item description Item code Configurable item
Scroll list Displayed on LCD Item description Item code Configurable item
Using the UP arrow and DOWN arrow you can scroll through this list of item readings. The list (12 items) is
configured using items 130 through 135 and items 075 through 086. All values that are displayed in this mode
are static readings, which means that they are sampled when you enter the level 0 mode and the displayed
values will not change while you remain in the scroll list mode. Scroll List mode has a 60-second inactivity
time-out. If you do not press any key for 60 consecutive seconds, EC 350 exits the level 0 mode and returns
to the normal Corrector mode.
This unlocks the EC 350 keypad and EC 350 enters the HMI keypad mode.
2. Press the DOWN arrow.
EC 350 enters the level 0 mode.
3. Use the UP arrow and DOWN arrow to scroll through the available menus.
Attention
When one of the arrows on the display disappears, it indicates that you have reached either the bottom
or top of the scroll list, and therefore you must scroll the opposite direction..
4. Press ESC to exit level 0 mode and return to the Corrector mode.
Refer to the section “Level 0 mode” for information about level 0 mode.
Level 1 mode
Level 1 mode provides read-only information on the following:
The level 1 mode is read-only mode where you can view alarms, firmware, and configuration information.
Perform the following steps to enter the level 1 mode.
The following table lists the factory-defined list for level 1 mode.
The individual
alarm items are
not displayed
in the L1.2
menu, unless
that particular
alarm is active.
This unlocks the EC 350 keypad and EC 350 enters the HMI keypad mode.
2. Press OK to enter level 1 mode.
Refer to the section “Level 1 read only mode” for information about level 1 mode.
Level 2 mode
Level 2 mode is the limited-access configuration mode and provides read-only information on the following:
• Passcode
• Alarms
• Clear Alarms
• Configuration
• Verify Pressure
• Verify Temperature
• Meter Prover
• Diagnostics
The following flowchart illustrates the main menus and submenus in the level 2 mode.
After you enter level 2 mode, the items listed in the table appear on the display; by default, ALARMS appear.
Use the UP arrow and DOWN arrow to scroll through the items in level 2 mode.
L2.1 PASSCODE
The active
alarms appear
on the display.
L2.3 CLR ALARMS Allows you to clear the active alarms on the display.
If the condition that caused the alarm is still present,
a new alarm is activated on the next measurement.
L2.4.1 ALARMS
L2.4.1.1 LIMITS
L2.4.1.1.1 P HI LIMIT
L2.4.1.1.2 P LO LIMIT
L2.4.1.1.3 T LO LIMIT
L2.4.1.1.4 T HI LIMIT
L2.4.1.1.8 D CV LIMIT
L2.4.2 INSTRUMENT
L2.4.2.1.2 FW VERSION
L2.4.2.1.3 SITE ID ●1
L2.4.2.1.4 SITE ID ●2
L2.4.2.1.6 DATE
L2.4.2.1.7 TIME
L2.4.2.2 BATTERY
L2.4.3.1 DRESSER
L2.4.3.1.1 I-D B3
L2.4.3.1.3 ROT B3
L2.4.3.3 ROMET ®
L2.4.3.3.1 ROT RM
L2.4.3.4 SENSUS
L2.4.3.5 OTHER
L2.4.3.5.1 I-D/PULSE
L2.4.4 VOLUME
L2.4.4.2 CORVOL
L2.4.4.2.1 CORVOL
L2.4.4.2.2 CV UNITS
L2.4.4.2.3 CV DIGITS
L2.4.4.3 UNCVOL
L2.4.4.3.1 UNCVOL
L2.4.4.3.2 UV UNITS
L2.4.4.3.3 UV DIGITS
L2.4.5 ENERGY
L2.4.5.1 ENERGY
L2.4.6 PRESSURE
L2.4.6.1.3 P-DECIMALS
L2.4.6.1.9 FIXED P
L2.4.7 TEMPERATUR
L2.4.8.2.1 SQ SU FACT
L2.4.8.3 PARAMETERS
L2.4.9.2 COM-1
L2.4.9.2.2 COMM
FORMAT
L2.4.9.2.4 HANDSHAKE
L2.4.9.2.5 COMMWAKEUP
L2.4.10.1 CH-A
L2.4.10.2 CH-B
L2.4.10.3 CH-C
L2.5 VERIFY P/T Allows you to validate the accuracy of the pressure/
temperature measured by EC 350.
The EC 350
calculates the %
ERROR.
L2.6 MTR PROVER Allows you to check the accuracy of the actual
volume of gas flowing through a gas meter. Press
. The following items appears.
UC IDIAL RATE
- Uncorrected
Input Dial Rate.
TC IDIAL RATE
- stands for
Temperature
Corrected Input
Dial Rate.
I FLOW RT -
stands for Input
Flow Rate.
L2.8 VIEW LOGS Allows you to view a specified Log Record’s Date,
Time, and CRC value (in decimal).
L2.8.1 AT-LOG1 Actual Log Item Values are not available. This
function is mainly to serve the needs of Compliance
verification. Specify a Log Record number to begin
and then press the Enter Key to view the actual
record’s parameters (Date, Time, and CRC).
L2.8.2 AT-LOG2
L2.8.3 AT-LOG3
L2.8.4 AT-LOG4
L2.8.5 AT-LOG5
L2.10 SHUTDOWN Special mode to take the unit out of service – i.e. to
put in long term storage (shelf mode)
Limited
functionality
is enabled in
this mode –just
time update.
No volume or
measurements
are performed.
HMI is active.
L2.11 BATT CHNG Enter this mode when replacing the battery –
includes resetting battery gauge.
This unlocks the EC 350 keypad and EC 350 enters the HMI keypad mode.
2. Press OK to enter level 2 mode.
3. To access the passkey screen, in level 2 mode press OK. The following appears on the display.
0000000
PASSCODE
For firmware versions before 1.0068, you must enter a five digit security code followed by a two digit user
ID code to enter level 2 mode. The default value for the level 2 passcode is 20000. The default value of the
last two digit user ID is 00. You can enter any value for the event log user ID. If the unit is configured to log
events in an event log, the user ID identifies the user who made the changes in the event log.
For firmware version 1.0068 and later, of the 7 digits entered, the first two are the user id and the last five
are the passcode. By default only user 02 is permitted access to Level 2. The default passcode is 20000, so
the value entered would be 0220000. Other users can be granted access to level 2 menus, but a user can
only access level 2 if they are not granted access to level 3. (A single user can access level 2, level 3, or
neither, but cannot access both level 2 and level 3. Level 2 functionality is a subset of level 3 functionality.)
4. To enter passcode, press the UP arrow or the DOWN arrow at the current position to change the displayed
digit to the required numeral.
5. For example, press the UP arrow twice at the current position, the value 0000000 changes to 2000000.
6. Press the RIGHT arrow to move one position to the right. Using the UP arrow or the DOWN arrow change
the displayed digit to the required numeral.
7. For example, press the RIGHT arrow until you reach the sixth digit. At the sixth digit press the UP arrow
twice. The value 2000000 changes to 2000020.
8. Repeat this step until you enter the passcode of the instrument.
9. To accept changes after entering the last digit (all seven digits of passcode), press the RIGHT arrow.
Attention: You can also press OK for entering the passkey value.
The OK key enters the displayed digit at the current position and advances to the next position.
If the log in was successful, the display indicates that you are in level 2 mode. If the user ID or
passcode were invalid, you will be returned to the main level 1 menu.
10.Use the UP arrow and DOWN arrow to scroll through the main menu items (L2.2 through L2.7) in level
2 mode.
11.Press OK to select the necessary main menu item.
12.Use the UP arrow and DOWN arrow to scroll through the submenu items in level 2 mode.
13.Press OK to select the necessary submenu.
14.Press ESC to return to the main menu item.
15.From the main menu item, press ESC to exit level 2 mode.
Level 3 mode
Level 3 mode is the limited-access configuration mode and provides read-only information on the following:
• Passcode
• Alarms
• Clear Alarms
• Configuration
• Advance Configuration
• Verify Pressure
• Verify Temperature
• Calibrate Pressure
• Calibrate Temperature
• Meter Prover
• Diagnostics
• Shutdown
After you enter the level 3 mode, the items listed in the table appear on the display; by default, ALARMS
appear. Use UP arrow and DOWN arrow to scroll through the items in the level 3 mode.
L3.3 CLR ALARMS Clears the alarms. Press and then press YES to
clear the alarms.
L3.4.1 ALARMS
L3.4.1.1 LIMITS
L3.4.2 INSTRUMENT
L3.4.2.2 BATTERY
L3.4.3.1 DRESSER
L3.4.3.1.1 I-D B3
L3.4.3.1.3 ROT B3
L3.4.3.3 ROMET ®
L3.4.3.3.1 ROT RM
L3.4.3.4 SENSUS
L3.4.3.5 OTHER
L3.4.3.5.1 I-D/PULSE
L3.4.4 VOLUME
L3.4.4.2 CORVOL
L3.4.4.3 UNCVOL
L3.4.5 ENERGY
L3.4.5.1 ENERGY
L3.4.6 PRESSURE
L3.4.7 TEMPERATUR
L3.4.8.3 PARAMETERS
L3.4.10.1 CH-A
L3.4.10.2 CH-B
L3.4.10.3 CH-C
L3.5 ADV CONFIG Allows you to enter the item number. Press . The
following appears.
ENTER
ITEM NUMBR
Using the
keypad, you
can enter the
required item
number. After
you enter the
item number,
the respective
item appears
on the display.
You can modify/
set the value
of the item.
For example, if
you enter the
item number
0000008, GAS
PRESS appears
on the display.
L3.6 VERIFY P/T Allows you to validate the accuracy of the pressure
or temperature measured by EC 350.
L3.7 CALIB P/T Allows you to calibrate the pressure and temperature
measured by EC 350.
L3.8 MTR PROVER Allows you to check the accuracy of the actual
volume of gas flowing through a gas meter. Press
. The following items appear.
• UC IDIAL
RATE
• TC IDIAL
RATE
• I FLOW RT
UC IDIAL
RATE, stands
for Uncorrected
Input Dial Rate,
TC IDIAL RATE
stands for
Temperature
Corrected Input
Dial Rate, and
I FLOW RT
stands for Input
Flow Rate.
L3.10 VIEW LOGS Allows you to view a specified Log Record’s Date,
Time, and CRC value (in decimal).
L3.10.1 AT-LOG1
L3.10.2 AT-LOG2
L3.10.3 AT-LOG3
L3.10.4 AT-LOG4
L3.10.5 AT-LOG5
L3.12 SHUTDOWN Special mode to take unit out of service – i.e. to put
in long term storage (shelf mode).
Limited
functionality
is enabled in
this mode –just
time update.
No volume or
measurements
are performed.
HMI is active.
L3.13 BATT CHNG Enter this mode when replacing the battery –
includes resetting battery gauge.
This unlocks the EC 350 keypad and EC 350 enters the HMI keypad mode.
2. Press OK to enter level 3 mode.
3. To access the passkey screen, in level 3 mode press OK. The following appears on the display.
0000000
PASSCODE appears.
For firmware versions before 1.0068, you must enter a five digit security code followed by a two digit user
ID code to enter level 3 mode. The default value for the level 3 passcode is 30000. The default value of the
last two digit user ID is 00. You can enter any value for the event log user ID. If the unit is configured to log
events in an event log, the user ID identifies the user who made the changes in the event log.
For firmware version 1.0068 and later, of the 7 digits entered, the first two are the user id and the last five
are the passcode. By default only user 03 is permitted access to Level 3. The default passcode for user 03
is 30000, so the value entered would be 0330000. Other users can be granted access to level 3 menus,
but a user can only access level 2 if they are not granted access to level 3. (A single user can access
level 2, level 3, or neither, but cannot access both level 2 and level 3. Level 2 functionality is a subset of
level 3 functionality.)
4. To enter the passcode, press the UP arrow or the DOWN arrow at the current position to change the
displayed digit to the required numeral.
5. For example, press the UP arrow thrice at the current position, the value 0000000 changes to 3000000.
6. Press the RIGHT arrow to move one position to the right. Using the UP arrow or the DOWN arrow change
the displayed digit to the required numeral.
7. For example, press the RIGHT arrow until you reach the sixth digit. At the sixth digit, press the UP arrow
thrice. The value 3000000 is changed to 3000030.
8. Repeat this step until you enter the passcode of the instrument.
9. To accept the changes after entering the last digit (all seven digits of passcode), press the RIGHT arrow.
Attention: You can also press OK for entering the passkey value.
The OK key enters the displayed digit at the current position and advances to the next position.
If the log in was successful, the display indicates that you are in level 3 mode. If the user ID or
passcode were invalid, you will be returned to the main level 1 menu.
10.Use the UP arrow and DOWN arrow to scroll through the main menu items (L3.2 through L3.10) in level
3 mode.
11.Press OK to select the necessary main menu item.
12.Use the UP arrow and DOWN arrow to scroll through the submenu items in level 3 mode.
13.Press OK to select the required submenu.
14.Press ESC to return to the main menu item.
15.From the main menu item, press ESC to exit level 3 mode.
1. Unlock the keypad and type the PASSCODE to enter level 2 mode or level 3 mode. Refer to the section
"Access level 2 mode" for accessing level 2 mode or section “ Access level 3 mode ” for accessing level
3 mode.
2. By default, in level 2 mode or level 3 mode the following appears on the display.
MAIN MENU
ALARMS
Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
CONFIG appears.
3. Press OK. The following appears on the display.
CONFIG
ALARMS
4. Using the UP arrow and DOWN arrow scroll through the options until METER TYPE appears.
5. Press OK.
The following default meter maker appears on the display.
METER TYPE
DRESSER
The following preconfigured types of makers of meters appear. Using the UP arrow and DOWN arrow
scroll through the options.
• ELSTER AMC
• ROMET
• SENSUS
• OTHER
6. Press OK to select the required type of meter maker.
7. Using the UP arrow and DOWN arrow scroll through the options and press OK to select the required
type of meter.
The following image illustrates the preconfigured types of meter makers and the meters available under each
maker.
Verifying pressure
To verify the pressure:
1. Unlock the keypad and type the PASSCODE to enter level 2 mode or level 3 mode Refer to the section
“ Accessing level 2 mode ” for accessing level 2 mode or section “ Access level 3 mode ” for accessing
level 3 mode.
2. By default, the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
VERITY P/T appears.
4. Press OK. For example, the following appears on the display.
VERIFY P/T
VALIDATE P
5. Press OK. The live pressure measurement made by the instrument appears on the display. For example:
51.00
PSI
LIVE PRESS
The pressure value that appears in this step is the value that EC 350 reads.
6. Press OK. For example, the following appears on the display.
50.00
PSI
INPUT REFP
7. Enter the reference pressure (the actual/exact pressure being piped to EC 350) using the keypad and
press
OK.
The percentage difference between the pressure that EC 350 reads and the reference pressure appears
on the display. For example:
2.00
% ERROR LIVE
PRESS
8. Press OK. The control returns back to the following:
VERIFY P/T
VALIDATE P
Verifying temperature
To verify the temperature:
1. Unlock the keypad and type the PASSCODE to enter level 2 mode or level 3 mode. Refer to the section “
Accessing level 2 mode ” for accessing level 2 mode or section “ Accessing level 3 mode ” for accessing
level 3 mode.
2. By default, the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
VERITY P/T appears.
4. Press OK. The following appears on the display.
VERIFY P/T
VALIDATE P
VALIDATE T
ERR CNTS
SENSOR TST
Ensure that the value of ERR remains zero +/- one count, whereas the value of CNTS must increase with
each volume input.
1. Unlock the keypad and type the PASSCODE to enter level 2 mode or level 3 mode. Refer to the section
“ Accessing level 2 mode ” for accessing level 2 mode or section “ Accessing level 3 mode ” for accessing
level 3 mode.
2. By default, in level 2 mode or level 3 mode the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
CONFIG appears.
4. Press OK. The following appears on the display.
CONFIG
ALARMS
5. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
INSTRUMENT appears.
6. Press OK.
The following appears on the display.
INSTRUMENT
SITE INFO
7. Press OK, the following appears on the display.
UNIT S/N
8. Using the UP arrow and DOWN arrow scroll through the options until SITE ID #1 appears.
9. To change the SITE ID #1, press OK and then using the keypad enter the eight-digit SITE ID.
10.SITE ID is in addition to the UNIT S/N. Follow the site-specific rules for configuring the site ID number.
11.Once you enter all the eight digits, press OK.
INSTRUMENT
SITE INFO
7. Press OK, the following appears on the display.
UNIT S/N
8. Using the UP arrow and DOWN arrow scroll through the options until DATE appears.
9. Press OK.
10.Using the keypad set the DATE and press OK.
Code Units
Code 0 CF
Code 1 CFX10
Code 2 CFX100
Code 3 CFX1K
Code 4 CFX10K
Code 5 CCF
Code 6 MCF
Code 7 M3X0.1
Code 8 M3
Code 9 M3X10
Code 10 M3X100
Code 11 M3X1K
11.Press OK to select the required unit.
For example, if you press OK at Code 0, CF is selected as the unit of corrected volume.
12.Repeat step 6 through step 10 for selecting the unit for uncorrected volume. Selecting the unit of
measurement for energy
Code Units
Code 0 THERMS
Code 1 DECATHERMS
Code 2 MEGAJOULES
Code 3 GIGAJOULES
Code 4 KILOCALS
Code 5 K-WATT HRS
9. Press OK to select the required unit.
For example, if you press OK at Code 0, THERMS is selected as the unit of ENERG Y .
2. By default, in level 2 mode or level 3 mode the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
CONFIG appears.
4. Press OK. The following appears on the display.
CONFIG
ALARMS
5. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
PRESSURE appears.
6. Press OK.
By default, the following appears on the display.
PRESSURE
GAS PRESS
7. Press OK.
By default, the following appears on the display.
0
PSI
GAS PRESS
8. Press the DOWN arrow key.
The following appears on the display.
0
PSI
PRESS UNIT
9. Press OK and then using the UP arrow and DOWN arrow increase or decrease the value of Code.
The corresponding unit associated with the code appears. For example, Code 0 represents PSI.
Following are the codes and the corresponding units of measurement for PRESSURE.
Code Units
Code 0 PSI
Code 1 kPa
Code 2 mPa
Code 3 Bar
Code 4 bar
Code 5 KGcm2
Code 6 In WC
Code 7 In HG
Code 8 mm HG
10.Press OK to select the required unit.
For example, if you press OK at Code 0, PSI is selected as the unit of PRESSURE.
1. Unlock the keypad and type the PASSCODE to enter level 2 mode or level 3 mode. Refer to the section
“ Accessing level 2 mode ” for accessing level 2 mode or section “ Accessing level 3 mode ” for accessing
level 3 mode.
2. By default, in level 2 mode or level 3 mode the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 2 mode or level 3 mode until
CONFIG appears.
4. Press OK. The following appears on the display.
CONFIG
ALARMS
5. Using the UP arrow and DOWN arrow scroll through the options until TEMPERATURE appears.
6. Press OK.
By default, the following appears on the display.
TEMPERATURE
GAS TEMP
7. Press OK.
By default, the following appears on the display.
60.00
F
GAS TEMP
8. Press the DOWN arrow key.
The following appears on the display.
F
TEMP UNITS
9. Press OK and then using the UP arrow and DOWN arrow increase or decrease the value of Code.
The corresponding unit associated with the code appears. For example, Code 0 represents F.
Following are the codes and the corresponding units of measurement for PRESSURE.
Code Units
Code 0 F
Code 1 C
Code 2 R
Code 3 K
10.Press OK to select the required unit.
For example, if you press OK at Code 0, F is selected as the unit of TEMPERATURE .
Calibrating temperature
To calibrate temperature:
1. Unlock the keypad and type the PASSCODE to enter level 3 mode. Refer to the section "Accessing level
3 mode ” for accessing level 3 mode.
2. By default, the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 3 mode until CALIB P/T appears.
4. Press OK. The following appears on the display.
CALIB P/T
CALIB PRES
5. Press the DOWN arrow, the following appears on the display.
CALIB P/T
CALIB TEMP
Press OK, the temperature read by EC 350 appears on the display.
For example:
75.20
F
LIVE TEMP
Here, 75.20 is the value that EC 350 reads and F stands for the unit of temperature that is Fahrenheit.
6. Press OK. The following appears on the display.
75.20
F
INPUT REFT
Here, REFT stands for reference temperature.
7. Enter the reference temperature (the temperature that EC 350 must read) by using the keypad and then
press OK.
The percentage difference between the temperature that EC 350 reads and the reference temperature
appears on the display. For example, the following appears.
0.04
% ERROR LIVE TEMP
8. Press OK, the control returns back to the following:
CALIB P/T
CALIB TEMP
Calibrating pressure
To calibrate pressure:
1. Unlock the keypad and type the PASSCODE to enter level 3 mode. Refer to the section "Accessing level
3 mode ” for accessing level 3 mode.
2. By default, the following appears on the display.
MAIN MENU
ALARMS
3. Using the UP arrow and DOWN arrow scroll through the options in level 3 mode until CALIB P/T appears.
4. Press OK. The following appears on the display.
CALIB P/T
CALIB PRES
5. Press OK, the pressure read by EC 350 appears on the display.
51.00
PSI
LIVE PRESS
The pressure value that appears in this step is the value that EC 350 reads.
6. Press OK. The following appears on the display.
50.00
PSI
INPUT REFP
7. Enter the reference pressure (the actual/exact pressure being piped to EC 350 ) using the keypad and
press OK.
The percentage difference between the pressure that EC 350 reads and the reference pressure appears
on the display. For example:
2.00
% ERROR
LIVE PRESS
8. Press OK, the control returns back to the following:
CALIB P/T
CALIB TEMP
Default users 2 and 3 are primarily intended for HMI access (levels 2 and 3 respectively) but serial access with
MasterLinkSQL through those users is also possible with those passcodes.
The firmware upgrade is performed through the IrDA serial interface that is used by the MasterLinkSQL
software. The IrDA serial interface provides an electrically isolated interface and requires no additional
hardware or software beyond what is already needed to utilize the MasterLinkSQL software. (Firmware
upgrades can also be performed over the RS-232 port but not the RS-485 port.)
Caution: It is important that you read all items and logs (audit trail logs, event log, alarm log) before
performing a firmware upgrade because most items will be set to default values and the audit trail
logs will be erased.
2. To select the firmware file, click Browse and navigate to the path where the EC 350 firmware file is saved.
3. From the Serial Port list, select USB IrDA.
Ensure to select the proper serial port for communicating with EC 350.
4. Select the Set instrument’s date and time check box.
5. Click Start Upgrade.
The program begins to read the current item file from EC 350. The WARNING dialog box appears.
6. Click OK only if you have saved ALL (Calibration and non-Calibration) items in EC 350.
The program begins to load the new firmware and then validates its integrity. Once the firmware is
validated, the firmware is sent to the MasterLinkSQL software, as illustrated in the following image.
7. Click OK only if you have saved ALL (Calibration and non-Calibration) items in EC 350.
The program begins to load the new firmware and then validates its integrity. Once the firmware is
validated, the firmware is sent to the MasterLinkSQL software, as illustrated in the following image.
8. After the upgrade is complete, Upgrade was successful (in green) appears at the lower left corner of
the main Firmware Upgrader dialog box.
You can select and configure all the items that are displayed in the Display/Change Items By Function
dialog box.
For example, if you want to configure Volume & Energy, perform the following:
1. Select Volume & Energy and click OK. The Volume & Energy dialog box appears.
2. Select Corr Volume Units and click Change. The Change Item dialog box appears.
3. From the list, select the required unit for volume and then click Save.
Similarly you can select and configure all the items in the Volume & Energy dialog box.
Attention: Follow the same procedure to change other functions.
1. Establish a serial communication between EC 350 and MasterLinkSQL. Refer to the MasterLinkSQL
User’s Guide for information about establishing a serial connection between EC 350 and MasterLinkSQL.
2. In the MasterLinkSQL window, click on Instrument > Set Instrument Date and Time via Computer to
update EC 350 with the current PC’s date and time.
A dialog appears allowing the user to accept the date and time update.
3. Click the Yes button to proceed with the Date/Time update. MasterLinkSQL will always re-read the PC’s
clock date and time just before sending the data packet over to EC 350.
Item files
Calibrating pressure
Attention: Check the sensors before calibrating pressure.
To calibrate pressure:
1. In the MasterLinkSQL window, click Calibrate > Pressure .
The Pressure 1 dialog box appears.
2. Apply a zero reference pressure (0.00 PSI) to the P1 pressure transducer of EC 350.
Wait for the pressure reading displayed on the Pressure 1 dialog box to stabilize.
3. After the reading stabilizes, click Average Pressure Now .
MasterLinkSQL averages the reading and displays the average pressure.
4. Click Change.
The Change Pressure Calibration Value dialog box appears.
5. Change the value to accurately match the zero reference pressure applied to the pressure transducer.
6. Click OK .
7. Click Span Calib.
The Pressure 1 Span Calibration page appears.
8. Apply a high pressure such that the pressure applied is at least 50% more than the zero reference
pressure.
Wait for the pressure reading displayed on the Pressure 1 dialog box to exceed the zero reference
pressure by 50%. For example, if the zero reference pressure on a 100 PSI transducer equals 0.00 PSI,
then the span reference pressure must be between 50.00 and 100.00 PSI. If the zero reference pressure
on a 600 PSI transducer equals 0.00 PSI, then the span reference pressure must be between 300.00
and 600.00 PSI.
9. After the pressure stabilizes, click Average Pressure Now.
MasterLinkSQL averages the reading and displays the average pressure.
10.Click Change.
The Change Pressure Calibration Value dialog box appears.
11.Change the value to accurately match the high reference pressure applied to the pressure transducer.
12.Click OK .
The Restore Line Pressure dialog box appears.
13.Restore the normal pressure line to EC 350 and click Done to exit the calibration.
3. Apply a zero reference pressure (0.00 PSI) to the pressure transducer of EC 350. Wait for the pressure
reading displayed on the Pressure 2 dialog box to stabilize.
4. After the reading stabilizes, click Average Pressure Now.
9. Apply a high pressure such that the pressure applied is at least 50% more than the zero reference
pressure.
Wait for the pressure reading displayed on the Pressure2 dialog box to exceed the zero reference
pressure by 50%. For example, if the zero reference pressure on a 100 PSI transducer equals 0.00 PSI,
then the span reference pressure must be between 50.00 and 100.00 PSI. If the zero reference pressure
on a 600 PSI transducer equals 0.00 PSI, then the span reference pressure must be between 300.00
and 600.00 PSI.
10.After the pressure stabilizes, click Average Pressure Now.
MasterLinkSQL averages the reading and displays the average pressure.
11.Click Change.
The Change Pressure Calibration Value dialog box appears.
12.Change the value to accurately match the high reference pressure applied to the pressure transducer.
13.Click OK.
The Restore Line Pressure dialog box appears.
14.Restore the normal pressure line to EC 350 and click Done to exit the calibration.
Calibrating temperature
Attention: Check the sensors before calibrating temperature.
To calibrate temperature:
1. In the MasterLinkSQL window, click Calibrate > Temperature .
The Temperature Calibration dialog box appears.
2. Insert the temperature probe of EC 350, along with an accurately calibrated thermometer calibrated
thermometer into a low temperature bath or dry well, which provides the temperature zero reference.
Wait for the temperature reading displayed on the Temperature Calibration dialog box to be stabilized.
3. After the reading stabilizes, click Average Temperature Now .
MasterLinkSQL averages the reading and displays the average temperature.
4. Click Change.
The Change Temperature Calibration Value dialog box appears.
5. Change the value to accurately match the temperature measured by the calibrated thermometer placed
in the low temperature bath.
6. Click OK.
7. Click Span Calib.
The Temperature Span Calibration page appears.
8. Insert the temperature probe of EC 350 and the calibrated thermometer into a high temperature bath,
which provides the temperature span reference.
The temperature reading displayed on the Temperature Calibration dialog box must exceed the zero
reference temperature by at least 15% to proceed with span calibration.
9. After the reading stabilizes, click Average Temperature Now .
MasterLinkSQL averages the reading and displays the average temperature.
10.Click Change.
The Change Temperature Calibration Value dialog box appears.
11.Change the value to accurately match the temperature measured by the calibrated thermometer placed
in the high calibration bath.
12.Click OK .
4. Click Change.
The Change Item dialog box appears.
5. Type the item number of your choice that must be listed first in the Meter reader list and then click Save.
For example, if you want Battery usage alarm (item 100) to appear first in the Meter reader list, then type
item number 100 and then click Save.
The Scroll List dialog box appears and the item 130 (first list in Meter reader list) is configured to Battery
usage alarm (item 100).
6. Perform step 3 through step 5 sequentially to change items 130 through 135 and 75 through 86, if you
want to change all the 12 default items.
3. The site ID must be identical to the one configured in PowerSpring during the EC 350 product addition. It
is a six digit, hexadecimal number.
4. Click OK to close the Site Information window.
5. Double click the Audit Trail Log Configuration option from the Display/Change Items By Function dialog
box. The EC 350 audit log appears.
6. Specify the log interval size (this should match the interval size specified in the Input field definition in
PowerSpring). Configure the Audit Trail items of EC 350 from the left pane. The sequence and number of
Audit Trail items must match the inputs provided for RUID Inputs on Definiton page in PowerSpring.
7. Click OK to close the Configuration window.
8. Double click the Call-In Config option from the Display/Change Items By Function dialog box. The Call-
In Config dialog box appears.
9. Type the information as shown, items: 1230, 333 (you may choose some other item numbers as well,
491, 492, 494,495, 496, upto last). Critical items are, 339 & 493, where the Host machine's IP Address
for MERCURYSERVER and TCP Port (50467) should be correctly specified. The remaining fields can be
defined by the user.
Note: The Corrected Volume # of digits is defined for item-90, 97 and the Uncorrected Volume
# of digits is defined for item-92, 97 These values must match the No. of Dials field in the Unit
Configuration tab in PowerSpring.
7
Remote Communications
Topics: This chapter describes the different communication mechanisms
featured in a EC 350 device.
• Use of RS232/ 485
• Call in and call out
• Modbus Host Communication
Remote Communications
For both RS232 and RS484 TXD (Transmit Data), RXD (Receive Data) and GND (Ground) are required in all
cases. In some applications TXD of the EC 350 may connect to RXD of the external device; other times they
may be crossed. Also, RTS (Request To Send) and CTS (Clear To Send) may or not be used and may or may
not be crossed. Consult the documentation of the external device for these details.
The RS232 and RS485 standard specifications should be consulted for the proper wire type, maximum cable
length, etc.
Jumpers
In addition to wiring, jumpers on JP1 and JP2 must be properly positioned. For RS232, both jumpers must be
set to B and C as shown in the following diagram. They must be placed on A and B for RS485.
Configuration Parameters
Several EC 350 items must be set to accommodate a particular external device. Modbus configuration items
are found in the ‘Modbus Protocol’ function group of MasterLink:
If the external device communicates via Modbus protocol, item 985 must be set to 1. Item 994 will need to
be set to the proper Modbus variation, either RTU or ASCII. The correct setting for all of these items will be
determined by the external device used.
Others items, specifically 272, 995 and 1220, are found in the ‘Communications’ group:
Items 170 and 172 are primarily associated with the behavior of MI Protocol. These should normally left as
default (shown here). The remainder will vary depending on the external devices.
Call in
When connected to either a cellular or landline modem the EC 350 can be configured to autonomously dial and
connect to a host computer. It can do this according to a preset schedule or in the event of an alarm condition.
Previous Mercury Instruments products have incorporated similar features, but the EC 350 is unique in its
support of four phone numbers (or IP addresses); two for scheduled calls and two for alarms. This provides
the ability to communicate with four different hosts. In turn, this feature leads to some differences in the retry
algorithm. Call in configuration items are:
Modem items are shared between the scheduled and alarm call in features. If the modem requires an
initialization string it must be set into item 491 along with an Init Delay (Item 821). The initialization delay is
applied so that the modem, having just been powered up, has a chance to stabilize before being sent the
initialization string. A dial prefix must be configured in item 492 and a hangup string in 494. The Dialing Delay will
take place following the issuance of the initialization string. Call in sequence, Trigger Type and retry parameters
are also shared between scheduled and alarm calls. Those will be discussed later.
Scheduled Call In
The EC 350 can initiate a call at a preset time. The time may have been manually specified, but on an ongoing
basis the host(s) is responsible for the schedule. At the end of each call the host is must set item values that
determine when the next call is to take place. That time will always be contained in items 334 and 335. Items
338 and 337 indicate the result of the latest call ( or call attempt ). If only one host exists, its phone number
is set in item 339. If a second host exists, its number is set in item 1030. To enable scheduled call in, at
least phone number must be configured and item 333 must be set to either Alarm and Scheduled Call In or
Scheduled Call In Only.
Alarm Call In
If alarm call in is configured and enabled, a call will be initiated whenever a new alarm occurs or, if in RBX
mode, an alarm condition clears. This requires a phone number in item 493 ( and optionally in 785 ) and item
333 must be set to either Alarm and Scheduled Call In or Alarm Call In Only.
Retry Timing
Not all call failures are resolved with ‘retries’. ‘Retries’ is defined as the termination of the current attempt and
the scheduling of another attempt some minutes or hours hence. In some cases the EC 350 may try to resolve
the failure immediately. If, for instance, an initialization string is sent to the modem but no positive response
is received within several seconds, it will simply send it again. If several initialization attempts fail, then it will
be deemed hopeless (for now) and a future retry will be scheduled. The same is true if a dial string goes
unanswered. But after a certain point another immediate attempt would be futile and a future retry is scheduled.
There are three retry timing parameters – two delays ( A and B ) and a counter. For any given trigger (schedule
or alarm ) the first retry will be scheduled to take place Retry A minutes in the future. Retry A will be used
for subsequent retries until Modem Retry A count is exhausted. The next ( and subsequent ) failures will be
scheduled at intervals of Retry B minutes. Once the call finally succeeds ( or another trigger occurs ) the
counter, and thus the retry sequence, will be reset.
LCD indications
When a call is triggered, the information on the 2nd and 3rd lines of the LCD will be replaced with
messages indicating the progression of the call, such as ‘INIT MODEM’, ‘DIALING ●’, ‘DIAL FAILED’,
‘CONNECTED’, ‘WAIT FOR CONN RESPONSE’, ‘CONNECTED’, ‘WAITING +-+CLRALMS’, and ‘HOST
SYSTEM CONNECTED’.
Call out
The Call Out window (COW) allows EC 350 to accept remote communications during one or two blocks of
time during the day.
When used with a CNI2 cellular modem, the CNI2 will be set to “server mode” during the call out window(s) so
it can accept calls from a host system during those times. The CNI2 consumes more power in server mode,
regardless of whether communication occurs.
EC 350 can also be kept “awake” during the call out window(s) such that even the initial characters of
a communication session are accepted. Otherwise, the initial characters “wake up” the EC 350. The MI
Protocol includes such wake up characters in its protocol so that this feature is not required for MI Protocol
communication. But the Modbus protocol includes no wake-up characters, so this feature can be used so that
the EC 350 processes the first Modbus packet. Without it, the first packet is not processed and the Modbus host
system must send a “retry” packet. More power is consumed during the call out window in this mode, regardless
of whether communication occurs or not. To use this mode, in addition to specifying a call out window, set
item 1236 (Port Active During COW) to enabled. This mode consumes more battery power but reduces retries
on Modbus activity.
There are items for specifying the start and stop times for two call out windows. Set the start and stop times
equal to disable a call out window. (Do so for both if they have no call out window.) Times are specified in 24-
hour format. The maximum value is “23 59 00”. Seconds other than “00” are ignored. The resolution for the
times is one minute. No single call out window can be longer than 18 hours, but greater than 18 hours can be
achieved by scheduling two call out windows “back to back” (e.g. 00 00 00 - 12 00 00, 12 00 00 - 23 59 00
would effectively give you a call out window for the entire day except one minute before midnight).
The modem will stay active until a stop time is reached or when the current call (if after the stop time) has
ended. EC 350 goes back to sleep if there is no call out.
4. Click Change.
The Change Item dialog box appears.
5. Enter the desired start time and click Save.
6. Repeat steps 2 to 5 if you want to change item 1233.
4. Click Change.
Pin Connection
8
Maintenance
Topics: This chapter lists out the temperatue and transducer related kits. It
also includes information on how to remove the HF cover.
• Temperature Probe
Measurement Kits
• Transducer Replacement Kits
• Redundant Uncorrected
Switch
• Installing Measurement
Canada (MC) Sealing Cover
• Removing and Re-Installing
Human Factor (HF) Cover
Maintenance
1. The following figure shows MC Cover installed with 3 Cross-Drilled security screws at locations indicated
by arrows.
2. Using sealing wire and lead seal, seal cover at these locations. Run wire through the screw head, and
then through the MC cover sealing features.
3. Re-install Human Factor cover tightening the screws to 9 +/- 1 in-lbs of torque at these locations. Ensure
power cable is routed in the orientation shown for battery connection.
9
Item Reference
Topics: This chapter lists out all the item references, along with the item
number, the item name, a brief description of the item functionality,
• Item reference table and the default value of the items.
Item Reference
007 Ch-C Pulses Waiting The channel C waiting pulse equals twice 0
the number of channel C output pulses
remaining from the last wakeup period,
waiting to be sent to remote device.
008 P1 Gas Pressure Most recently measured P1 gas pressure.
Usedin calculating the pressure correction
factor (Item 044). The pressure value is
scaled to the unit of measure selected at
Item 087
010 P1 Press High Alarm High limit for P1 pressure alarm. Initiates 099999.99
Limit an alarm (Item 145) if exceeded.
011 P1 Press Low Alarm Low limit for P1 pressure alarm. Initiates -1.0
Limit an alarm (Item 143) if exceeded.
012 Atmospheric Barometric pressure optionally entered by 0.0000
Calibration Ref. the technician during the most recent
P1 calibration of an absolute pressure
transducer.
013 Base Pressure Base pressure per the contract, used 14.7300 (PSIA)
in calculatingpressure correction factor
at item 044 and ultimately the total
correction factor at item 043
016 P1 Span Calibration The pressure value used during the most 0.00
Pressure recent P1 Pressure Span calibration.
017 P1 Press Calibration The most recent offset for the pressure 0.0000
Zero transducer, obtained during most recent
P1 Pressure Zero calibration. The value
is normally very close to 0.0000 and is
recalculated during each Pressure Zero
Calibration.
018 P1 Prev-1 Pressure The previous value of Item 017 0.0000
Cal. Zero (Calibration Zero), updated on each P1
Pressure Zero Calibration.
019 P1 Prev-2 Pressure The previous value of Item 018 0.0000
Cal. Zero (Calibration Zero), updated on each P1
Pressure Zero Calibration.
020 P1 Press Calibration The most recent span value for 1.0000
Span the pressure transducer, obtained
during most recent P1 Pressure Span
calibration. The value is normally very
close to 1.0000 and is recalculated during
each Pressure Span Calibration.
021 P1 Prev-1 Pressure The previous value of Item 1.0000
Cal. Span 020(Calibration Span), updated on each
P1 Pressure Span Calibration.
022 P1 Prev-2 Pressure The previous value of Item 1.0000
Cal. Span 021(Calibration Span), updated on each
P1 Pressure Span Calibration.
023 Minimum PCal Point The configurable spread or difference 50.0000
Diff % required (expressed in percent of full-
scale of pressure range) between
PressureCal Zero and Pressure Cal Span
calibration points.
024 Excess PCal The configurable, maximum allowable 2.0000
Change % change (expressed in percent of full scale
of pressure range) to Item 017 during
Pressure Zero calibration and to Item 020
during Pressure Span calibration.
025 P1 Transducer The upper range limit of the installed P1 100
Range (PSI) pressuretransducer, always expressed in
PSI.
027 Gas Temp Low Low limit for gas temperature alarm. -35.00
Alarm Limit Initiates an alarm (Item 144) if exceeded.
028 Gas Temp High High limit for gas temperature alarm. 165.00
Alarm Limit Initiates an alarm (Item 146) if exceeded
029 Zero Calibration The temperature value used during 0.0000
Temperature themost recent temperature zero
calibration.
030 Span Calibration The temperature value used during 1.0000
Temperature themost recent temperature Span
calibration.
031 Case Temperature Temperature measured inside the 0.0000
instrument enclosure.
032 Case Temperature Highest case temperature measured by -50
Maximum the unit since item was reset.
033 Case Temperature Lowest case temperature measured by 170.00
Minimum unit since item was reset.
034 Base Temperature Base temperature used to calculate the 60.00
temperature correction factor (item 045).
Deg.F
035 Temperature The most recent offset value for the 0.0000
Calibration Zero temperature transducer, obtained during
a Temperature Zero calibration. The
valueis normally very close to 0.0000 and
is recalculated during each Temperature
Zero Calibration.
036 Previous-1 The previous value of item 035, updated 0.0000
Temperature on each Temperature Zero Calibration.
Calibration Zero
037 Previous-2 The previous value of item 036, updated 0.0000
Temperature on each Temperature Zero Calibration.
Calibration Zero
057 Ch-B Pulse-out Scaling factor that determines the volume 2.000
Scaling per pulse for channel-B pulses when Item
093 is set to CorVol or UncVol.
Historically this has been a configurable
item but in EC 350, it is read-only (for
reference) and is now calculated based on
the pulse weight selection at item1194.
058 Ch-C Pulse-out Scaling factor that determines the volume 2.000
Scaling per pulse for channel-C pulses when Item
093 is set to CorVol or UncVol.
Historically this has been a configurable
item but in EC 350, it is read-only (for
reference) and is now calculated based on
the pulse weight selection at item1195.
062 Unit serial number Factory assigned, 8-digit instrument serial FirmwareDefault =
number. 00000000, actual
number is Factory
The first 2 digits do not represent the year set.
of manufacture.
075 HMI Scroll List - Item The 7th of 18 user configurable, instrument 045
7 parameters that will appear on the LCD
(Temperature Factor)
display when HMI Level-0 is accessed.
(HMI Level - 0)
See Scroll List parameter description at
item 130.
076 HMI Scroll List - Item The 8th of 18 user configurable, 043
8 instrument parameters that will appear
on the LCD display when HMI Level-0 is (Total Factor)
(HMI Level - 0) accessed.
080 HMI Scroll List - Item The 12th of 18 user configurable, 122
12 instrument parameters that will appear
on the LCD display when HMI Level-0 is (Firmware Version)
(HMI Level - 0) accessed.
081 HMI Scroll List - Item The 13th of 18 user configurable, 255
13 instrument parameters that will appear
on the LCD display when HMI Level-0 is (End of List)
(HMI Level - 0) accessed.
082 HMI Scroll List - Item The 14th of 18 user configurable, 255
14 instrument parameters that will appear
on the LCD display when HMI Level-0 is (End of List)
(HMI Level - 0) accessed.
083 HMI Scroll List - Item The 15th of 18 user configurable, 255
15 instrument parameters that will appear
on the LCD display when HMI Level-0 is (End of List)
(HMI Level - 0) accessed
084 HMI Scroll List - Item The 16th of 18 user configurable, 255
16 instrument parameters that will appear
on the LCD display when HMI Level-0 is (End of List)
(HMI Level - 0) accessed
085 HMI Scroll List - Item The 17th of 18 user configurable, 255
17 instrument parameters that will appear
on the LCD display when HMI Level-0 is (End of List)
(HMI Level - 0) accessed.
086 HMI Scroll List - Item The 18th of 18 user configurable, 255
18 instrument parameters that will appear
on the LCD display when HMI Level-0 is (End of List)
(HMI Level - 0) accessed.
089 Temperature Units Unit of measure for Gas Temperature 0 Deg (F)
(026) and other related temperature items.
The following arethe available temperature
units:
• 0= F (Fahrenheit)
• 1= C (Celsius)
• 2= R (Rankine)
• 3= K (Kelvin)
090 Corrected Volume Unitof measure for Corrected Volume and 5 (CCF)
Units other corrected volume related items. The
following are the available volume units:
• 0= CF
• 1 = CFx10
• 2 = CFx100
• 3 = CFx1K (x1,000)
• 4 = CFx10K (x10,000)
• 5 = CCF
• 6 = MCF
• 7 = m3x0.1
• 8 = m3
• 9 = m3x10
• 10 = m3x100
• 11 = m3x1K (x1,000)
099 Battery Low Alarm When active, indicates the battery(s) has 00000000
reached or exceeded the low limit set by
item 049 (Battery Low Limit). Item 099
displays:
11111111 (HMI) & Yes (Software)
when in an alarm state
00000000 (HMI) & No (Software)
when there is no low battery alarm
The alarm remains active until cleared
manually or automatically when the
batteries are replaced.
113 High Resolution Display of the fractional portion (to the 000.0000
4th decimal) of item 000 (Totalized
Corrected Volume Corrected Volume) with the last three
CorVol integers provided for reference.
For example, if the value at item 000
equals "12345678 MCF", item 113 may
appear as 678.2477 MCF". The four
decimal places are useful during field
calibration checks when using a small
number of meter revolutions.
114 Meter Scaling Scaling value (multiplier) applied to the 1.0000
value selected at item 098 (Meter Index
Rate), ie., the scaling applied the volume
per each meter output shaft rotation or
volume pulse input.
Note: If Item 433 is set to Rotary mode, the
value at item 114 is automatically inserted
when the meter model is selected at item
432.
130 HMI Scroll List - Item The 1st of 18 user configurable, instrument 002
1 parameters that will appear on the LCD
(UncVol)
display when HMI Level-0 is accessed.
(HMI Level - 0)
Placing the value 255 at any Scroll List
parameter is interrupted as the end of the
list.
See Chapter-6 in this manual for
information about the Scroll List feature.
See items 075 – 086 for Scroll List items
7 – 18.
131 HMI Scroll List - Item The 2nd of 18 user configurable, 000
2 instrument parameters that will appear
on the LCD display when HMI Level-0 is (CorVol)
(HMI Level - 0) accessed.
132 HMI Scroll List - Item The 3rd of 18 user configurable, 008
3 instrument parameters that will appear
on the LCD display when HMI Level-0 is (Gas Pressure)
(HMI Level - 0) accessed.
133 HMI Scroll List - Item The 4th of 18 user configurable, 026
4 instrument parameters that will appear
on the LCD display when HMI Level-0 is (Gas Temperature)
(HMI Level - 0) accessed.
135 HMI Scroll List - Item The 6th of 18 user configurable, 044
6 instrument parameters that will appear
on the LCD display when HMI Level-0 is (Pressure Factor)
(HMI Level - 0) accessed.
137 Pressure The upper pressure range limit, 100
Transducer Range automatically scaled to the pressure units
selected at Item 087.
Examples for a 100 PSI transducer:
If Item 087 is set PSI, Item 137 = 100.00
If Item 087 is set kPa, Item 137 = 689.476
If Item 087 is set KGcm2, Item 137 = 7.031
142 Gas Energy Value The amount of Energy containedin each 1000.0
volumetric unit of corrected volume. This
value is user configurable and is a
multiplier for calculating Energy at Item
140. The nominal Gas Energy Values for
the Energy Units (item 141) are:
THERMS = 1000.0 BTU per CuFt
DKTHERMS = 1000.0 BTU per CuFt
MJOULES = 37259.00 Kilojoules per m3
GJOULES = 37259.00 Kilojoules per m3
KILOCAL = 8905.00 Kilocalories per m3
KILOWATTHR = 10349.73 Watt/Hr per m3
191 Previous day energy Yesterday's Daily Energy. That is, the daily 00000000
energy that is one `Gas Day' old at
the time the instrument was accessed. This
information is updated with the value from
item 190 at the beginning of every new
gas day.
The gas day begins at the time entered at
Item 205.
10 = 10 minutes
15 = 15 minutes
24 = Daily (24 hours)
30 = 30 minutes
31 = Monthly (31 days)
60 = Hourly (60 mins)
210 Peak flow rate This item indicates the highest flow rate of
the gas.
211 Peak hour cor vol The largest 1 hour of corrected volume
during the gas day.
214 Interval high Highest Gas Pressure measured during 0.00
pressure the time interval selected at item 202.
The value at item 214 is initialized (re-
zeroed) at the beginning of each Gas Day
(per item 205). If you select the Display
items by function in the MasterLinkSQL
software, it is the 3rd item in group 15. It
has 87 units.
254 Max day date The date the max day CorVol (253) 01-01-01
occurred. Format: Determined by item
262.
256 Daily average The daily average of pressure of gas. 0.00
pressure Resets at the start of each gas day.
257 Daily average The daily average of gas temperature 0.00
temperature (026). Resets at the start of each gas day.
258 Audit Trail - item 1 Used only as a reference to indicate 000 (Corrected
the 1st item value logged in the Backup Volume)
Memory module.
259 Audit Trail - item 2 Indicates the 2nd item value logged in the 002 (Uncorrected
Backup Memory module. Volume)
260 Audit Trail - item 3 Indicates the 3rd item value logged in the 043 (Total Correction
backup memory module. Factor)
261 Audit Trail - item 4 Indicates the 4th item value logged in the 257 (Daily Avg.
backup memory module. Temperature)
262 Date format Code (0-2) selects the format in which MM-DD-YY
the Date is entered and displayed at item
204. The format for all date related items (Month/Day/Year)
are also governed by the selection of item
262. The date formats are MM DD YY,
DD MM YY, and YY MM DD.
264 Miscellaneous action For Honeywell use only. 0
265 Memory Address For Honeywell use only.
266 Memory Data For Honeywell use only.
267 Miscellaneous config For Honeywell use only. 0
284 Maximum dial rate The pressure at which Max Dial Rate
press occurred.
285 Maximum pressure The highest value of Gas pressure since
(P1) last manually reset.
286 Maximum pressure The time when the highest value of Gas 00:00:00
time pressure is obtained.
287 Maximum Pressure The date on which the highest value of 01-01-01
Date (P1) Gas pressure is obtained.
288 Maximum press flow The maximum flow rate of pressure.
rate (P1)
289 Minimum pressure The lowest value of Gas pressure since 0.00
(P1) last manually reset.
295 Maximum gas The date when the Max Gas Temp (item 01-01-01. Format set
temperature date
293) occurred. by item 262
296 Max gas The Instantaneous Flow Rate (item 209) 0.00
temperature flow when the Max Gas Temperature (item
rate
293) occurred.
300 Minimum gas The instantaneous flow rate (item 209) 0.00
temperature flow when the min gas temperature (item 297)
rate occurred.
333 Call-in trigger type 0 = No Call-In 0 (No Call-In)
1 = Absolute
2 = None
Note: This item is read-only.
It is read automatically from
the actual pressure transducer
installed.
0 = 4 Cell Alkaline
1 = 2 Cell Lithium
2 = 4 Cell Lithium
1062 Door Tamper State State of the door tamper switch input. 0 = 0
no tamper (switch closed)
1 = tamper (switch open)
1063 Ext Power Alarm External power low voltage alarm time 00:00:00
Time
1064 Ext Power Alarm External power low voltage alarm date 01-01-01
Date
1065 Log 2 Interval User configurable Log-2 interval options: Hourly
10 = 10 minutes
15 = 15 minutes
24 = Daily (24 hours)
30 = 30 minutes
31 = Monthly (31 days)
60 = Hourly (60 mins)
10 = 10 minutes
15 = 15 minutes
24 = Daily (24 hours)
30 = 30 minutes
31 = Monthly (31 days)
60 = Hourly (60 mins)
10 = 10 min.
15 = 15 min.
24 = Daily (24 Hrs)
30 = 30 min.
31 = Monthly (31 days)
60 = Hourly (60 min)
1190 Meter S/N User can enter the serial number of the 00000000
meter into the corrector for traceability
1191 Battery usage alarm Time stamp when battery usage alarm 00:00:00
time occurred.
1192 Battery usage alarm Date stamp when battery usage alarm 01-01-01
date occurred.
1193 Ch-A pulse output Amount of volume associated with one 100 CF per pulse
value Channel-A Output Pulse.
The following options are available for the
Ch-A pulse output value:
0 = 1 CF
1 = 10 CF
2 = 100 CF
3 = 1000 CF
4 = 10000 CF
5 = 0.1 m3
6 = 1 m3
7 = 10 m3
8 = 100 m3
9 = 1000 m3
1195 Ch-C pulse output Amount of volume associated with one 100 CF per pulse
value Channel-C Output Pulse.
The following options are available for the
Ch-C pulse output value:
0 = 1 CF
1 = 10 CF
2 = 100 CF
3 = 1000 CF
4 = 10000 CF
5 = 0.1 m3
6 = 1 m3
7 = 10 m3
8 = 100 m3
9 = 1000 m3
1196 Log 1 name User has the ability to give each audit
trail log a descriptive name of up to 20
characters. This is used as reference for
assisting the user to determine the correct
information before downloading the data or
to makes it easier to sort the data once it is
on their computer.
For example, a 20 character log name
such as “Daily Billing” or “Hourly Volume”
or “Technical Checks” or “Daily Status”
or “Log for Joe” or any short description
helps the user to determine the correct
information before downloading the data or
to make it easier to sort the data once it is
on their computer
1237 Super Comp Alarm Alarm time when a Super Compress 00:00:00
Time calculation type error (Alarm) first occurs.
10
EC 350 specifications
Topics: This section provides the specifications and certifications for EC 350.
• Certifications
• Power
• Temperature measurement
system
• Pressure measurement
system
EC 350 specifications
Certifications
• Electrostatic Discharge Immunity Test (EN61000-4-2)
• Radiated, Radio-Frequency Electromagnetic Field Immunity Test (EN61000-4-3)
• Electrical Fast Transient/Burst Immunity Test (EN61000-4-4)
• Conducted Radio-Frequency Electromagnetic Field Immunity Test (EN61000-4-6)
• Radiated Emissions (EN 61000-6-4)
• FCC Part 15
• Measurement Canada (pending)
• CSA C/US--- Class I, Div 1/Zone 0 (pending)
• CSA C/US Class I, Division 2, Groups C and D, T3C
• IECEx --- Class I, Zone 0 (pending)
• IECEx --- Class I, Zone 2 (pending)
Power
• 7 Volt 2 cell lithium battery pack
• 6 Volt 4 cell alkaline battery pack
• External DC supply 5 - 15 Volts DC