Professional Documents
Culture Documents
For
Radiography Testing
Note:- This Procedure provides general information. This procedure for a project shall be valid only
when reviewed and modified as per specific project requirements and approved by authorized
person.
Job
Document No Revision Number
Number
2 Q M 0 0 J E P M E 0 0 3 00 01 02
Department/
Quality Management Reviewed By DSK RKG PKD / RKG
Division
Location: Jamnagar Approved By SCG AJV AJV
RELIANCE ENGINEERING ASSOCIATES (P) LIMITED
CONTENTS
1. INRODUCTION
2. PURPOSE
3. SCOPE
6. REFERENCE DOCUMENTS
8. PROCEDURE
9. RECORD
10. ATTACHMENTS
Wire type IQI designation and wire diameters, T-233.2 (single page)
• Radiography film length coverage for various sizes of pipes (one page)
2. PURPOSE
This procedure provides step by step instructions on requirements, performance and
documentation of radiography testing applicable to materials & welded joints in fabricated
vessels, piping and structurals.
3. SCOPE
This procedure covers the requirements for radiography testing and the scope is limited to
radiography of welded joints and material, employing gamma ray & X-ray.
RT Radiography Testing
8. PROCEDURE
All personnel involved in radiographic testing and evaluation of exposed films shall be
qualified in accordance with a recognized National / International Standard or as desired by the
referencing code section. For those code sections that directly reference Article 1 of ASME Sec.
V, the qualification of personnel shall be in accordance with SNT-TC-1A.
The assistance radiographer shall have TLD badges, issued by BARC registered in their own
name.
2
Development technician shall be ASNT II and minimum 3 years experience
All equipment shall be of a recognized manufacturer and shall meet the statutory requirement
and specification. It shall be ensured that all the safety precautions as stipulated by BARC are
followed while carrying out radiographic examination
The intensity of gamma rays / X-rays shall be selected in accordance with the thickness of the
material to be radiographed.
All Contractors are to submit their RT working procedure if any, to QMD for approval prior to
the qualification of their radiography technique / exposing production joints.
All the contractor shall qualify the radiography technique in line with their RT working
procedure and in the absence of RT working procedure the contractor shall qualify the
radiography technique in line with clause No. 8.4 & 8.5 of this procedure as a minimum, prior
to the production welding to avoid the accumulation of joints for exposure. Sample format for
recording the “ Qualification of Radiographic Technique” is enclosed as attachment 5.
An adequate number of exposure shall be made to demonstrate that the required coverage has
been obtained. When complete coverage is required for circumferential welds (materials), a
minimum of three exposures taken 120 deg. to each other shall be made.
For welded pipes having a nominal diameter of 3½ inch. or less, a technique may be used in
which the radiation passes through two walls and the weld (material) in both walls is viewed for
acceptance on the same radiograph. For double-wall viewing, only a source side IQI (Image
Quality Indicator) shall be used. Care should be exercised to ensure that the required geometric
sharpness is achieved.
For welds , the radiation beam may be offset from the plane of the weld at an angle sufficient to
separate the images of the source side and film side portions of the weld so that there is no
overlap of the areas to be interpreted. When complete coverage is required, a minimum of two
exposures taken 90 deg. to each other shall be made for each joint.
As an alternative, the weld may be radiographed with the radiation beam positioned so that the
images of both walls are superimposed. When complete coverage is required a minimum of
three exposures taken at 120 deg. apart shall be made for each joint.
When more than one film is used to cover single weld, then the films will have at least ¾”
overlap to ensure full coverage of the weld area.
Ug = Fd/D Where,
Ug = geometric unsharpness
F = source size
d = distance from source side of the weld or object being radiographed to the film.
The value of geometric unsharpness shall not exceed the following unless stringent requirement
is specified.
Under 2 0.020
2 through 3 0.030
The Contractor shall demonstrate / qualify the methods and techniques meeting the specified
parameters of his work procedure for radiography.
Type of film, screens and process controls are to be specified in the Contractors working
procedure. Films used shall be of Industrial Radiography, KODAK, AGFA, INDU or
equivalent brand. The quality / grade of film within the brand is with reference to the qualified
radiography technique, where the film is meeting the requirement of quality / sensitivity.
Front and back Lead Screen thickness shall be 0.125 mm and above, the exact thickness and the
brand to be used is as specified in the qualified radiography technique document.
Films shall be washed in running water and shall be dried completely prior to interpretation.
Developer solution shall be replenished/replaced on a regular basis as the developer retards its
activity due to change in concentration.
Quality Management Dept. (QMD) shall arrange for the regular audit of the Contractors film
processing environment, to ensure the compliance of the requirement.
A lead symbol “B” with minimum dimensions of ½” (13mm) in height and 1/16” (1.6 mm) in
thickness shall be attached to the back of each film holder during each exposure to determine if
back scatter radiation is exposing the film.. If a light image of the “B” appears on a darker
background of the radiograph, protection from back scatter is insufficient and the radiograph
shall be considered unacceptable.
Where back scatter is suspected, lead sheet should be placed below the test piece so that
radiation is absorbed and not scattered back on to the film.
Desired IQI’s shall be used to ensure that the required sensitivity is achieved. Selection of
penetrameters shall be as per T-276 of ASME Sec. – V, unless stringent requirement is
specified. Thickness 1/4" (Approx.6 mm) and below ,DIN ISO wire type IQI may be used. 2
As far as possible IQI’s shall be located on the source side of the weld. Where this is not
practical the IQI’s may be placed on the film side for single wall viewing. In this case, a lead
marker “F” shall be placed adjacent to the penetrameter. For use of IQI / penetrameter, T-277
of ASME Section V and/or any other stipulated specification shall be followed.
Film shall be adequately identified to maintain full traceability of the weld being radiographed.
The following information shall appear on each radiographic film
• Date
• Welder No.
• Weld No.
• Package No.
st
To identify the 1 repair of welds, the letter “R1” shall be used on the film and “R2” for second
repair.
This identification shall be of a permanent nature by using lead letter (6mm height and 1.5 mm
thk.). Lead marker during radiography shall be positioned so that they do not interfere with
film interpretation.
Films shall be examined under subdued background light, using a high intensity variable type
of film viewer capable of viewing density within the range of 1.8 to 4.0. Dimmed diffused
lighting, rather than total darkness, is preferable in the viewing room. Room illumination / film
interpretation shall be arranged such that there is no reflections of light from the surface of the
film under examination.
All the radiographs shall be free from mechanical, chemical or other blemishes like - fogging,
processing defects (streaks, water marks, chemical stains), scratches, finger/nail marks, crimps,
dirtiness, static marks, smudges, tears and screen marks etc, which can mask and can be
confused with the image of any discontinuity
The transmitted film density through the radiographic image of the body of the appropriate IQI
and the area of interest (weld + HAZ on either side) shall be 2.0 to 3.5. The acceptable limit of
variation in density with in the area of interest shall be as per ASME Section V (T-282.2)
and/or any other applicable specification.
8.5.4 Sensitivity requirements
Table T- 276 and Table T- 233.2 for wire IQI designation of ASME SEC V is enclosed as
attachment 2 & 3 respectively for ready reference. The attachments and the requirement of this
radiography procedure need to be revised with reference to the revision in governing
specification.
8.5.6 Documentation
The evaluation/interpretation finding of the radiographed films and all details of radiography
work shall be recorded on the Radiography Inspection Report Form 2QM00 FMT ME 027.
The inspection report shall include the information like,- client, identification of site, W.O. No./
Report No., date of test, project identification, RT technique, IQI detail, film type and screen,
density, radiation source and strength, exposure time, S F D, Sensitivity, welding details, weld
id, welder id, pipe dia & thk, RT result, radiographers and contractors name, signature with date
and stamp.
Considering the standard size (12”x15”) of the radiography film available in the market, the
coverage length of Radiography film for a specific pipe size and number of exposures are
established for Double Wall Double Image and Double wall single Image techniques of pipe
weld joints and incorporated as an attachment-6 to this procedure. In case of panoramic
exposure the number of segments and film length shall be viewed case to case basis. This shall
be taken as guideline for carrying out radiography on pipe circumferential weld joints.
9. RECORD
Nil
10. ATTACHMENT
3. ASTM wire type IQI designation and wire diameters (single page)
6. Radiography film length coverage for various sizes of pipes (one page)
ATTACHMENT 2 – (SINGLE PAGE) TABLE T-276
Nominal Penetrameter
Single wall
Material thickness Source Side Film Side
Range in inches Hole Type Essential Wire dia. Hole Type Essential Wire dia.
ASTM Wire type IQI designation, and wire diameters, in inches and mm
0.0032
0.004 0.08
0.1
A 0.005 0.13
0.0063 0.16
0.008 0.2
0.010 0.25
0.013 0.33
B 0.016 0.4
0.020 0.51
0.025 0.64
0.032 0.81
0.040 1.02
C 0.050 1.27
0.063 1.6
0.080 2.03
0.100 2.5
0.126 3.2
0.160 4.06
D 0.200 5.1
0.250 6.4
0.320 8.0
RADIOGRAPHY TECHNIQUES
Contractor :
Radiography Agency :
Project :
Client :
Procedure No. :
Rev. No. & Date :
------------------------------------------------------------------------------------
1) Date of qualification test :
2) Material specification :
3) Dia &Thickness :
4) Reinforcement :
6) Groove geometry :
7) Welding process :
8) Radiography Equipment/ :
Radiography source
9) Activity :
b) ASTM-Hole Type.
(As per ASME Sec V Article 2, T-276)
2 5"x3" 3 15
3 6" x 3" 3 18