100 mm Temperature Controlled Gas Cell

Part Number 162-20XX

OPERATION
The temperature controller is factory preset to work with the
heating module. Turn the unit on and adjust it to the desired
temperature. For additional functionality and detailed setup
information please refer to the Watlow User’s Manual
supplied with the control unit. Operation of the computer
controlled system and its software is explained in a separate
PIKE TempPRO Manual provided with this particular
configuration.
The 100 mm long stainless steel cell is designed for collecting
gas spectra at elevated temperatures, decomposition studies SAFETY
and measurements of gas evolving from heated samples. The
cell is 100 mm long, has a diameter of 38 mm (15 mm aperture) The maximum temperature of the accessory can reach
and is sealed with two 38 mm IR windows (purchased 300 °C. The cell components are insulated and the
separately). High temperature O-rings are used to seal the temperature of the outer shell should not exceed 80 °C.
windows (upper temperature range 300 °C). Gas is introduced Nevertheless, extreme caution should be used when operating
through a high quality stainless steel compression fitting and the cell, as some components, e.g. retaining rings, fittings and
the cell can withstand pressures up to 100 psi. The cell features windows cannot be insulated. These components may be
a stainless steel body resistant to most gaseous samples. Heat is extremely hot. They may cause severe burns when touched.
applied by a heating mantle surrounding the gas cylinder and Please allow enough time for the metal parts to cool down
monitored by a single thermocouple. The cell mounts directly in and check the cell temperature on the control unit display
the sample compartment of the spectrometer via a dedicated before working with the gas cell or removing it from the
baseplate. Optional high temperature purge tubes may be spectrometer. Power cord selected must comply with
ordered separately. IEC60320 C13 type connector. No poisonous or injurious gases
should be analyzed without a proper venting hood.
INSTALLATION
A hot surface (>80 °C) is indicated
Using a 3/32” hex wrench remove two stainless steel retaining by the following symbol on the cell:
rings and glass windows provided for cell protection during
shipment. Replace the glass windows with appropriate IR ENVIRONMENTAL
windows (must be ordered separately). Reattach and screw
down the stainless steel retaining rings. Please make sure that Altitude: Up to 2000 m
the windows and the retaining rings are positioned in the same Ambient Temp: 5 °C to 40 °C
order as when shipped. Humidity Range: Max rel 80% RH
Mains Supply: +/- 10% fluctuation max
Remove the original baseplate from the spectrometer’s sample Applicable Pollution Degree: Category 2
compartment. Position the Heated Cell in the spectrometer. Transient Voltages on Mains: <1500 V typ.
Connect one end of the provided cable (d-shaped connector)
to the Gas Cell and the other end (three prong plug) to the back
panel of the temperature control unit. Attach gas transfer lines
(if used). PACKING LIST
Description Qty
SPECIFICATIONS 100 mm Temperature Controlled Gas Cell 1
Temperature Range: Ambient to 300 °C Cell Cable 1
Accuracy: +/- 0.2 °C to 2.0 °C;
PLEASE NOTE: Two IR windows and temperature controller
depending on control unit used
must be ordered separately for complete assembly.
Probe Type: Single RTD
Heater Voltage: 24 VAC, provided by the
The Gas Cell features two compression fittings for 1/8”
Electronic Control Unit
transfer line. Transfer line is not included.
Power Consumption: 8 A max.
ECU Voltage: 115/230 V, selectable
Pressure Rating: 100 psi max.

6125 Cottonwood Drive, Madison, WI 53719 · Phone: (608) 274-2721 · Fax: (608) 274-0103
E-Mail: info@piketech.com · Web: www.piketech.com 350-162230-03