You are on page 1of 2

DPM4

Parameter Tester

Technical Data

The versatile DPM4 tests and calibrates flow and


pressure generators used in many medical devices.
With several measurements combined in a single,
handheld device, the DPM4 provides a cost-effective
solution, eliminating the need for multiple test meters.
The DPM4 features a menu-driven interface
for simple operation and an easy-to-read screen
that displays multiple parameter measurements
simultaneously.
Built-to-order, the palm-sized device comes in
four models to meet the varied needs of biomedical
engineers and technicians – the 1G, 1H, 2G, and 2H.
All models measure differential pressure, vacuum,
and temperature. Models 2G and 2H feature a built-
in barometer and measure barometric pressure, flow,
and humidity. The G and H models vary according to
pressure measurement ranges. The 1G and 2G models
measure pressure in the -700 mmHg to 5000 mmHg
range, and the 1H and 2H models measure pressure in
the -350 mmHg to 350 mmHg range. All the devices
provide highly accurate test results.

Key Features
All models Model 1H
• Palm size • Pressure measurements in -350 mmHg
• High accuracy to 350 mmHg range
• Differential pressure, vacuum, and temperature Model 2G
measurements
• Multiple user-selectable units of measurement • Barometric pressure, flow, and humidity
• Simultaneous display of multiple parameter measurements
measurements • Pressure measurements in -700 mmHg
• Leak-detection/leak-rate calculation to 5000 mmHg range
• RS-232 for computer control Model 2H
Model 1G • Barometric pressure, flow, and humidity
• Pressure measurements in -700 mmHg measurements
to 5000 mmHg range • Pressure measurements in -350 mmHg
to 350 mmHg range
Technical Temperature Probe PT-100 Relative humidity Model 2G or 2H
and PT-1000 Note: An integrated sensor in the
specifications instrument determines relative humidity
PT-100 operating range: measurements.
Pressure measurement -200 ˚C to 750 ˚C (-328 ˚F to 1382 ˚F)
Accuracy: Operating range: 12 % RH
Model 1H or 2H
± 0.13 ˚C @ 100 ˚C (0.23 ˚F at 212 ˚F); to 95 % RH
Operating range: -350 mmHg
± 0.1 ˚C @ 0˚C (0.18 ˚F @ 32 ˚F); Accuracy: ± 3.5 % of reading
to 350 mmHg
± 0.2 ˚C @ 100 ˚C (0.36 ˚F @ 212 ˚F) ± 2 % @ 25 ºC (77 ºF)
Accuracy: ± 0.3 % of range
Resolution: 0.1 % RH
Resolution: 0.1 mmHg PT-1000 operating range: Gas compatibility: Air
Units of measure: mmHg, mBar, -200 ˚C to 750 ˚C (328 ˚F to 1382 ˚F) Units of measure: % RH
cmH2O, psi, InHg, InH2O, kgcm2, Accuracy: 0.5 ˚F (0.3 ˚C)
and kPa barometric pressure Controls
LCD graphic display,
Model 1G or 2G Model 2G or 2H 128 pixels x 32 pixels
Operating range: - 700 mmHg Note: It is possible to compensate for the Data input/outputs: 1; bidirectional
to + 5000 mmHg sea level and calibrate for offsets.
RS-232 for computer control
Accuracy: ± 0.3 % of range for
Operating range: 380 mmHg
temperatures from 21 °C to 25 °C Power
to 900 mmHg
and relative humidity from 9 V alkaline battery RG9 or
Accuracy: ± 2 % of reading
30 % to 70 % ± 0.3 % of range; battery eliminator
Resolution: 1 mmHg
± 0.02 % of range per degree
Units of measure: mmHg, mBar, InHg,
C for temperatures < 21 °C or > Power consumption
and hPa
25 °C with relative humidity from < 70 mA
30 % to 70 %
Gas Flow Model 2G or 2H Battery life
Resolution: 0.5 mmHg
Note: Gas flow measures with an > 7 hours
Units of measure: mmHg, mBar, embedded sensor with 11 calibration
cmH2O, psi, InHg, InH2O, kgcm2, points to compensate non-linearity:
Case
and kPa calibration constants are stored in
firmware. ABS plastic case
Temperature measurement
Operating range: -750 ml/min Dimensions (LxWxH)
Operating range: -40 ˚C to 200 ˚C
to 750 ml/min 156 mm x 94 mm x 34 mm
(-40 ˚F to 392 ˚F )
Accuracy: ± 1 % of range or 6.1 in x 3.7 in x 1.3 in
Accuracy: ± 2 % of reading to
± 5 % of reading
± 0.5 % from -40 ˚C to 200 ˚C Weight
Resolution: 0.1 ml/min
Resolution: 0.1 ˚F and ˚C 0.4 kg with battery (0.9 lb)
Compatibility: Gas: Air, N2, O2, CO2,H2,
Units of measure: ˚F and ˚C
and He
Use standard external temperature Temperature
Units of measure: ml/min
probe type PT-100 or PT-1000 Operating: 15 ºC to 35 ºC
(or SCCM – standard cubic
(DIN/IEC 751 Class A) for temperature (59 ºF to 95 ºF)
centimeters per minute)
measurements in ˚F or ˚C. Storage: 0 ºC to 50 ºC
(32 ºF to 122 ºF)

Fluke Biomedical.
Ordering information Better products.
2583121: DPM4 Parameter Tester Model 1H More choices.
2631330: DPM4 Parameter Tester Model 1G One company.
2637760: DPM4 Parameter Tester Model 2H
2637772: DPM4 Parameter Tester Model 2G
Fluke Biomedical
PO Box 9090, Everett, WA 98206-9090 U.S.A.
Standard accessories
Fluke Biomedical Europe AS
2572323: User/service manual Vegamot 8, N-7048 Trondheim, Norway
2461946: Tubing kit For more information, contact us:
In the U.S.A. (800) 648-7952 or
Optional accessories Fax (425) 446-5629
2462177: Soft-sided carrying case In Europe/M-East/Africa +47 73954700 or
2647372: BE-UNVSL-IEC20C14 battery eliminator Fax +47 73954701
From other countries +1 (425) 347-6100 or
2461910: PT-100 temperature probe Fax +1 (425) 446-5629
2461922: PT-1000 temperature probe Email: sales@flukebiomedical.com
2461905: Expansion chamber Web access: www.flukebiomedical.com
2461946: Tubing kit w/Inflation bulb ©2007 Fluke Biomedical.
2462335: RS-232 cable Specifications subject to change without notice.
Printed in U.S.A. 9/2007 2817810A D-EN - Rev D