470-SU-e-9811
CONTENTS
View of Weighing Mechanism : Square Pan Type (KERN 470-38, KERN 470-44, KERN 470-46) 4
ELECTRONIC CONSTRUCTION 5
Block Diagram 5
Whole Wiring 6
TROUBLESHOOTING 7
Troubleshooting Procedure 7
Troubleshooting Table 8
Primary Checks 9
PERFOMANCE TESTS 11
Repeatability Test 11
PARTS REPLACEMENTS 13
ADJUSTING 16
Built-in Battery Unit Installation Manual for KERN 470 Series Balances 18
2 470-SU-e-9811
View of Weighing Mechanism
470-SU-e-9811 3
View of Weighing Mechanism : Square Pan Type (KERN 470-38, KERN 470-44,
KERN 470-46)
4 470-SU-e-9811
Electronic Construction
Block Diagram
470-SU-e-9811 5
Whole Wiring
6 470-SU-e-9811
Troubleshooting
Troubleshooting Procedure
470-SU-e-9811 7
Troubleshooting Table
8 470-SU-e-9811
Symptoms Causes & Remedy
Linearity is out of • Tuning-fork sensor or mechanism unit is defective.
specified range.
• Coefficient memories have changed by noises or static
electricity.
Corner error is too • Mechanism unit is defective, such Roberval plate
much. (spring) has been bent or twisted.
• Pan base touches other parts.
Display suddenly • Built-in battery (option) has been consumed.
disappeared in
• HGD board is defective.
operation.
• AC adapter is defective.
b-Err appears. • HGD board is defective.
• Coefficient memories have changed by noises or static
electricity. Adjust linearity.
Primary Checks
470-SU-e-9811 9
Removing / Fitting of Upper Case
10 470-SU-e-9811
Perfomance Tests
Repeatability Test
(1) Referring to operation manual, set AUTO-ZERO at off (3 A.0 0). Also set the least
readability at the finest (72 dA 1, 74 db 1).
(2) Press the T key to clear display to "0". Place a full balance weight just at the
centre of the weighing pan and read the value. Repeat this for 2 or 3 times to
verify that the errors at zero and full capacity are within the least readability.
(1) Referring to operation manual, set AUTO-ZERO at off (3 A.0 0). Also set the least
readability at the finest (72 dA 1, 74 db 1).
(2) Place a full balance weight just on the centre of the weighing pan and press the T
key to clear display to "0".
(3) Place the weight in order from A, to B, C, D, E referring to the above illustration,
reading the difference between "0" at the point A. If the errors are over the
tolerance, that is 3 divisions, adjust the corner errors referring to page 14.
470-SU-e-9811 11
Corner Error Adjustment
Remove the upper case referring to page 11. Place the pan base, then place the
weighing pan on the mechanism. Adjust the errors by adjusting bolts referring to the
following illustration.
12 470-SU-e-9811
PARTS REPLACEMENTS
(4) Unplug Output cord from CN5 on HGD circuit board, and pull out of Base Block.
(5) Remove three Mechanism Fixing Nuts on the bottom of the balance by a nut
driver of app. 7mm.
(6) Remove the mechanism unit by holding at base block*.
470-SU-e-9811 13
*Caution: Do not hold other parts than the base block.
(1) Place the mechanism on the lower case. Fix it by three nuts on the bottom of the
balance. Insert output cord into the base block referring to the following
illustration.
14 470-SU-e-9811
(2) Plug the output cord to CN5 on HGD circuit board, passing inside of the
supporting metal.
(3) Plug connection cord of the balance on CH1 of VIBA-2A, oscillation circuit board.
(5) Cover the balance with upper case referring to page 11.
470-SU-e-9811 15
Adjusting
(2) While pressing T key, press F key and release both together.
(3) unit ο appears showing the weight unit which currently set for Group "A" at
function "71 uA". If any adjusting weights for that unit is unavailable, change the
parameter to that of available weight, referring to next page, "Parameters and
Corresponding Adjusting Weights". Pressing T key changes the parameter in
following sequence.
If the parameter for 73_ub is set at "0", unit ο will not appear. In this case, you
should use adjusting weights either of the unit currently set in function 71 uA ο, or
of any available unit after changing the parameter at 71 uA ο.
(5) When automatic zero adjustment is completed, display changes to on F.S. Load
proper full capacity weight in the centre of the weighing pan, referring to the table
in next page. Make sure that the weight is of the unit selected in above (3). Load
the weight very gently.
(6) The span adjustment will be achieved automatically. Display will return to
weighing mode to indicate the adjustment is completed.
Notes
1. To stop operation in the middle, press S key.
2. 0-Err signs that the reference weight is over the full capacity.
3. Adjusting is available even with 50% weight of the capacity, still we recommend to
adjust with full capacity weights. 1-Err will appear if the weight is less than 50% of
the capacity.
4. 2-Err will appear if the error exceeds 1% of the capacity, or any other object than a
adjusting weight is loaded. Or, perhaps the balance is still defective.
16 470-SU-e-9811
For preparation of reference weight for adjusting:
3 (kg) 2 5 10 15
6 (lb) 4.4 11 22 33
470-SU-e-9811 17
Built-in Battery Unit Installation Manual for KERN 470 Series Balances
(1) Remove the weighing pan and the pan base. Then remove two fixing screws by a
Philips screw driver.
(2) Lift the case upward and unplug connector CN3 on HGD circuit board.
2. Remove the fixing screw on the battery hold once, slide the battery holder into the
Hold Support at the cut of the holder. Then fix the holder by the fixing screw
removed before.
3. Plug the HGBT-1 connection cord to CN2 on HGD circuit board.
18 470-SU-e-9811
4. Plug the connector of the case to CN3 on HGD circuit board, preventing from
holding cables by the cases.
5. Secure the case with two fixing screws.
6. Apply the pan base and the weighing pan. Over.
470-SU-e-9811 19
Parts List for KERN 470 Series Balances
20 470-SU-e-9811
470-SU-e-9811 21