Professional Documents
Culture Documents
CR To DR - Optimizing Image Quality and Dose
CR To DR - Optimizing Image Quality and Dose
Floor mounted
Ceiling mounted
R/F rooms
Why the Move from CR to DR?
Outcomes:
• Streamlined workflow & time savings
• 8.16 seconds less time per exam (portable chest, bone
surveys)
• More patients per day / year
• 100% FTE gain
• Rapid ROI
Making the Move from CR to DR…
ASRT HCIAC:
ASRT White Paper
Best Practices, ASRT
Advisory Council 2012:
• Background
• Dose / ALARA
• Social Marketing and
Safety Initiatives
• ACR Guidelines before
getting to the scope of
the white paper
Making the Move from CR to DR…
Electronic control
Triggers the switching diodes 1. Absorption of X-rays in the phosphor screen
2. X-ray is converted into visible light
3. Photo diodes read light signal and generates charges
4. Signals are amplified, digitized, processed and archived
5. Soft Copy display or hardcopy optional
in
Switching diodes
Connects each pixel to readout device
out
Analog-to digital
Multiplexer conversion
Readout the electronic signal
*Lança L., Silva A. (2013) Digital Imaging Systems for Plain Radiography, New York: Springer
Image Quality Improvement
Requires Improved Phosphor Technology
Traditional Powder Phosphors High Efficiency Needle Phosphors
The thickness of CR BaFlBr powder phosphor layer is limited to less than 300 µm, because of light
scattering. This thickness limits the X-ray absorption.
Due to the low light scattering of CsI a thicker phosphor layer can be used without jeopardizing the
sharpness of the imaging system. Higher X-ray absorption is possible with DR detectors using CsI
needle crystalline phosphors. This results in lower dose and better image quality.
Image Quality Improvement
CR DR
70 kVp 10 mAs
70 kVp 10 mAs 70 kVp 10 mAs
Image Quality Improvement and Lower Dose
CR DR
10% variation In
* With Agfa Grid Line Suppression Software - GLS
grid specification
Incorrect Grid Alignment = Poor Image Quality
Angle of tube collimator is not the same as the grid/panel
Tube/collimator is not parallel to the grid
Grid
Grid
Tube Collimator and Grid ARE Parallel (6:1)
Low Scatter Improved Lung and Spine Detail
Tube Collimator and Grid are not parallel
High Scatter Poor Lung and Spine Detail
“Decubitus” grid orientation
Angled positioning but Good Lung and Spine Detail
Incorrect Alignment = Poor Image Quality
Angle of tube collimator is not the same as the grid/panel
Tube collimator is not parallel to Grid
Grid
Grid
• Product Types
• Agfa MUSICA Chest +
• Fuji Virtual Grid
• Philips Skyflow
• Other ?
• How does it work ?
• Reduces the need for a grid using advanced image processing
• Scatter radiation is low frequency in an image
• Scatter Suppression Software reduces the visualization of low
frequencies while enhancing medium to high frequencies
Scatter Suppression Software extracts low frequency scatter
information while enhancing relevant clinic information
Non Grid Chest with Standard Processing Low Frequency Scatter Contribution
Scatter Suppression Software extracts low frequency scatter
information while enhancing relevant clinic information
Non Grid Chest with Scatter Suppression Low Frequency Scatter Contribution
Software
Scatter Suppression Software can improve lung field detail without
the use of an anti-scatter grid
Chest Phantom with 6:1 Grid Chest Phantom with 6:1 Grid
MUSICA 2 GenRad Soft Chest Processing MUSICA 3 Chest Processing flipped for comparison
100 kVp 3.2 mAs 100 kVp 3.2 mAs
* When compared to non grid exposures , grid exposures may require up to 50% higher exposure depending on conditions
Clinical Examples
CR DR
DI =10 * Log [ ]
EI_
TEI DI =10 * Log [ ]=0
500_
500
Target Exposure Index and Deviation Index
Actual Target
Exposure Exposure Deviation Exposure %
Index Index Index Factor Change
1300 500 4 2.6 160%
1000 500 3 2 100%
800 500 2 1.6 60%
630 500 1 1.26 26%
500 500 0 1 0%
400 500 -1 0.8 -20%
300 500 -2 0.6 -40%
250 500 -3 0.5 -50%
200 500 -4 0.4 -60%
. Exposure Monitoring Software
1. Color Coded Exposure Bar
2. Exposure Index Performance Table
60 kVp 1 mAs
Green
within range
The bar gives a relative indication of the exposure to the plate and it is a good
measure of the variation of exposure to the plate within a given exam type, but it
is not an absolute dose measurement value.
Yellow Slightly
Overexposed
60 kVp 4.75 mAs
Overexposed far
outside of range.
Greater than 4X exposure
Greater than 6.0 DI
Red Significantly
Overexposed
60 kVp 0.23 mAs
Underexposed far
outside of range.
Less than 1/4X exposure
Less than - 6.0 DI
Red Significantly
Underexposed
Region of Interest ROI Selection Affects Exposure Index
EI 375
DI 1.0
TEI 300
Region of Interest ROI Selection Affects Exposure Index
EI 64
DI -6.7
TEI 300
Region of Interest ROI Selection Affects Exposure Index
EI 1924
DI 8.1
TEI 300
Dose Monitoring
2. Proper Image Processing - Can significantly reduce dose and improve image
quality
3. Proper Grid selection and position – Will greatly influence image quality
a. Grid lines per inch
b. Grid position and distance
c. Non Grid Scatter Suppression software - Is an option to improve workflow and dose
6. X-ray Collimation
a. Influences scatter, image processing and overall image quality
b. Improper Electronic collimation or masking reduces image quality and increases
dose