Sandia National Laboratories Building 1090


Laboratory Modifications
PRESENTED BY: Mark B. Allen
SAND REPORT: SAN I 201 9-XXXX
Co-Author: Joe Zigmond
Radiation Protection Sample Diagnostics

Sandia National Labccatories is a multimission laboratory managedand


operated by National Technology and Engineering Solutions of Sandia LLC, a
wholly owned subsidiary of 1-loneryell International inc. for the U.S.
)A P RRMC 2019
Sandia National Laboratories is a multimission laboratory managed and operated by National Technology & Engineering Solutions of Sandia, LLC, a wholly owned
Departmentof Energys National Nuclear Security Administration under
subsidiary of Honeywell International Inc., for the U.S. Department of Energy’s National
contract DE-NA0003525. NuclearSERVICES
ANALYTICAL Security Administration under contract DE-NA0003525.
PROGRAM
Radiation Protection Sample Diagnostics

Mission: Laboratory Analysis Services in Support of


Worker Safety and Health
Liquid Scintillation Counting ICP-MS, ICP-OES
Gas Proportional Counting Flame Atomic Absorption
Gamma spectroscopy Sample preparation facilities
Whole Body Counting
I Alpha spectroscopy

C)Sandia National Laboratories


0
Ea i NMENT
SAFETY & HEALTH
i
2
Building Design Modification History

U Rush to re-locate to new facility (2004)


U Design and build—same contractor
U No prior experience with lab construction ®
o Undersized, stainless steel ductwork.
o Positive air pressure differential from
1090 Evacuation Plan
lab spaces to office spaces. PRIMARY ASSEM1SILY POINT — South end of the east
athing lot for Building 1090.
SECONDARY ASSEMBLY POINT — F re hydrant on the
north side of East Ordinance road, north of Bukfrig 1090.
MGM ;

o Emergency evac plan—traffic flow


Fre Alarm Pull

toward the hazard. [4] Fre Extinguisher


k Post Inclicalthig Valve

U Energy audit(2009)—High power


consumption due to undersized ductwork
o New variable frequency drive installed
to serve exhaust fans.
o New stainless steel ductwork installed
(correct size).
o Energy use significary reduced.

Sandia National Laboratories .livAmo&NHMEgTi 3


Building Design & Modification History

CI Extended power outages (2012)


o Ventilation shuts down while acid
digestions in process.
o No emergency lighting in lab spaces.
CI Stainless steel ductwork incompatible with acid
digestions
o Condensate collects in hoods.
• Insufficient air flow.
o Sub-optimal hood design.
o Corrosion of control valves—leakage into
ceiling spaces.

Sandia National Laboratories or,=Tram 4


Laborator acility Upgrades
Goals:
o Replace corroded ductwork with
PermaShield Pipe (PSP) fluoropolymer
coated ductwork.
o Install new acid-compatible hoods.
o Provide backup power and lighting.
o Improve lab work flow/traffic patterns and
evac routes.
o Re-purpose some under-utilized lab
spaces.
L

11111iiiiirAn

Sandia National Laboratories or% milizrzn 5


Upgraded Facility
Do not use these exits during
an oxygen depleting event

Covered -
Storage

izge -

178
186 Sample Recelyln!
RPSD 172
stable
Radiochemistry CI: 174
la e
Sample
Management
Office
182 180
RPSD Lab
Count Lab Storage

184
RPSD
190A Sample
188 Prep Lab
179 1771 w172
190 RPSD Particulate
Whole Body 168 Fitier Prep
Counting LataL_,
_(1,

•11111

U Consolidated 7 fume hoods (spread out


over 5 rooms)to 3 fume hoods in 1 room
LI Dedicated exhaust fan for acid digestions
U Added additional emergency evacuation
exit(away from potential lab hazards)

Project Cost:$845K
Construction Timeline:4 months

0Sandia National Laboratories E NMENT


SAFETY & HEALTH 6
onstruction Issue: Inadequate Air Flow
Do not use these exits dunng
an oxygen depleting eve t

o Identified from newly installed Covered


Storage

194

differential air pressure monitors. Slorage alb -

ta
86
SD
178
Sample Receiving.' _ HI 172
So hne
e 174 Sample
Rad. : ifif Portable Management
Model SRPM Secure-Sense@ Lab Office
182 180
RPSD Lab
Count Lab Storage 170
Normal Dis.dbled SMO
Sample
0.0015 "4/C Receiving

190A
Alarm 179 172
190 R SD
7:1' EMI ,
I r .1 Whole.,,y
186
Parficulate
Fifier Prep

101
Room Pressure Monitor

1
The pressure reading 1301e drlierentral
Pressure between the hallway and room

7
: 1
002
267 g
1111

111111

o Positive pressure from lab


space to office space during
windy conditions.

\I0 ENNARONMENT
0Sandia National Laboratories —oak SAFETY 6 HEALTH 7
New Project to Extend Walls to Roof Deck

Eli Sandia National Laboratories ‘r aF


lVI ir&NZI H"'
ET 8
Wall Extension Project In Progress

Custom
scaffolding
over hoods

Finished wall
Old ceiling
extended to
line
roof deck

0Sandia National Laboratories Or% =0,14,rn 9


Wall Extension Finished Product

Before After

Project Cost:$ I 54K


Construction Timeline: 3 months

(3Sandia National Laboratories Tek


\F
1V
ET
I IVTH
Ir I -Hi l
Big Improvement to Differential Air Pressure

Before After
Model SRPM Secure-Sense®
Model SRPM Secure-Sense®
lb MEW
Normal Disabled Normal Disabled
—0.0015 "WC !WM —0.0797 "WC

I• 410 0.100 0.100


-0.100 I 0.100 Alarm
Alarm Reset
Reset l * „41

Room Pressure Monitor

Room Pressure Monitor The pressure reading is the differential


1
lowpressure between the hallway and room
The pressure reading is the differential 184.

pressure between the hallway and room 184.

0Sandia National Laboratories ►ENVIROTY HENMENT


gio SAFEALTH
Et
Issues E ounte During Recent Building Mods

UUnderestimated personnel resources needed to support the project


o One lab POC to interface with designers and construction contractors.
o Many competing routine and non-routine laboratory operations and projects.
OUnforeseen hazards/issues
o Mice in the ceilings
o Removal of flooring—increased background in adjacent lab count room
UHow to work around existing hoods
o Special scaffolding required

ENVIRONMENT
Sandia National Laboratories I2
p SAFETY & HEALTH
11 11 essons Learned

O Insist on a design firm with experience in designing analytical lab spaces.


O Consider independent review of design plans by consultant.
O Match hoods and ducting materials to be compatible with chemical and
radiological needs.
O Consider emergency lighting & backup power options.
O Involve Radiation Protection and Industrial Hygiene professionals early in
design and work planning process.
O Need frequent update meetings during construction phase to plan around
lab operations.

Sandia National Laboratories .1


.miTr&NHmEln
11
13
essons Learned

O Consider emergency evacuation routes and options.


O Plan for how doorways are going to used
o Large openings for equipment movement
O Evaluate need for windows on doors and door type (i.e. crash bars).
O Evaluate the need for differential pressure monitors in critical areas.
O Consider an independent post-construction evaluation.
O Need plan for storing construction designs, documents, plans, photos for
future reference.

\
I0ENVVRONMENT
Sandia National Laboratories —oak SAFETY & HEALTH I4
L ASP
• ANALYTICAL SERVICES PROGRAM