Professional Documents
Culture Documents
Atomic Force Microscopy - AFM
Atomic Force Microscopy - AFM
MICROSCOPY–AFM
1. Introduction
Kirk-Othmer Encyclopedia of Chemical Technology. Copyright John Wiley & Sons, Inc. All rights reserved.
2 ATOMIC FORCE MICROSCOPY– AFM
Fig. 1. Schematic diagrams showing principles of (a) the mechanical surface profiler and
(b) the scanning tunneling microscope.
sample topography (or shape), and thus a profile of the sample surface is built up.
The principle of operation of a surface profile is shown in Figure 1.
The motion of the surface profiler may be one dimensional (ie, line scans are
made), or two dimensional, such that a kind of imaging occurs. This means that a
surface profiler can be thought of as a kind of microscope. In fact, a surface pro-
filer is somewhat like a simplified version of an atomic force microscope because
it uses a stylus to obtain a map of surface topography. However, the resolution of
this technique is fundamentally limited because the force between the tip and
the sample is uncontrolled. Because of this lack of force control, damage to tip
and the sample can occur. This means that in surface profilers, sharp tips cannot
be used because they would be broken quickly. Another predecessor of the atomic
force microscope was the scanning tunneling microscope, the STM (also shown in
Fig. 1). Unlike the surface profiler, in the STM, the tip does not touch the sample
surface. Instead, the tip is brought close to the surface (which must be conduc-
tive), and a tunneling current is measured between the tip and sample. To mea-
sure the tunneling current, the tip must be kept close to the sample while
scanning over it without touching. Therefore, instead of just moving the tip in
one or two dimensions, it is moved in three dimensions, typically by piezo trans-
lators, which can make precise movements. A high-speed feedback circuit
between the current measured and the z-movement of the piezos keeps the tip
the same distance from the surface at all times while also rastering the tip
over the sample surface in the x and y directions. A major drawback of STM is
that it is limited to conductive samples and works best in vacuum.
AFM is a generalization of the principles of STM, and because it is applic-
able to almost any sample, it overtook STM in popularity quickly and is now used
much more widely than STM. The principles of atomic force microscope operation
are illustrated in Figure 2.
The heart of the atomic force microscope is the probe. The probe in AFM is a
combination of a thin, flexible cantilever with an integrated tip at the end. Typi-
cally, the tip will have pyramidal shape and be very sharp, with a tip diameter of
5–20 nm. This tip interacts physically with the sample. In most imaging experi-
ments, there is a small repulsive force between the tip and the sample. The
motion of the probe, which depends on these forces, can be measured by several
techniques (4), but most commonly an optical lever is used. This setup is very
sensitive to the motions of the cantilever and is simple to implement, so it is
ATOMIC FORCE MICROSCOPY– AFM 3
Fig. 2. AFM principles of operation. The scanning in x and y is continuous. The laser,
probe, and photodetector serve to detect changes in height in the sample as this occurs.
The signal from the photodetector is fed back into the z scanner to maintain the probe-
sample force at a constant level.
used in almost all commercial AFM instruments. The signal from the optical
lever can have sub-Angstrom precision, and this signal is passed via a feedback
loop into the z movement transducer. The movement in most atomic force micro-
scopes is achieved using piezoelectric materials, which allows extremely precise
control of the motion of the probe relative to the sample (again, subnanometer
precision can be achieved). Whereas the x and y axis piezoelectrics scan in a repe-
titive pattern to raster the probe over the area of the sample of interest, the z axis
piezo is involved in the feedback loop to enable the fine control of tip-sample
force. The distance the z piezo has to move up and down to maintain the tip-sam-
ple force at the same set point is equivalent to the height (topography) of the
sample. A plot of this z height versus the x-y position of the probe gives us the
AFM image.
The combination of the following three factors using the feedback loop
explains the greatly improved resolution of atomic force microscopy compared
with profilometers:
1. Probe: the tip must be sharp to allow discrimination of small features. The
cantilever must be flexible (for contact mode) to allow for transduction of
small forces into movement.
2. Optical lever: greatly amplifies the movement of the probe so that small
heights can be measured and gentle forces are applied.
3. Piezoelectric translators: their high precision allows fine control of the
movement of the probe over the sample.
Fig. 3. Probes and probe layout. (a) The generalized layout showing probe substrate
(chip), cantilever, and tip. (b) Scanning electron microscopy (SEM) image of a probe
with rectangular cantilever for oscillating modes such as intermittent contact AFM.
(c) SEM image of a probe with ‘‘v’’-shaped cantilever for contact mode AFM. Often chips
for contact mode have multiple integrated probes to enable the selection of a cantilever
with appropriate spring constant.
ATOMIC FORCE MICROSCOPY– AFM 5
vertical deflection is used. Several other mechanisms are used (4) (notably inter-
ferometric measurements (7)), but the optical lever is employed in the over-
whelming number of atomic force microscope designs. The major advantages of
this design are that it is extremely sensitive, as well as cheap and simple to
implement. Drawbacks include the fact that it is not particularly compact and
that is somewhat more complex to implement in probe-scanning designs. There
can be some variation on the laser used; for example, low coherence lasers reduce
interference effects, whereas nonvisible (eg, infrared) lasers can be used to avoid
interfering with simultaneous optical microscopy.
Feedback Electronics. The signal from the photodetector must go
through a feedback loop into the z piezoelectric element. Typically, this is done
using a proportional, integral, derivative (PID) controller, which moves the z
translator such that the cantilever deflection returns to a setpoint. The setpoint
is programmed either automatically by the software or by the instrument opera-
tor. The P, I, and D terms can be altered in real time by the operator to optimize
imaging, along with other parameters including the setpoint and imaging speed.
Typically, the feedback is handled by an ‘‘electronics box’’ located between the
AFM head and a controlling PC, which also generates the signals for x and y
translators, and handles the digitization of the image signals.
Piezoelectric Translators. A wide variety of translation mechanisms for
use in AFM instruments has been investigated, but in general, the mechanisms
nearly always rely on piezoelectric ceramic materials. As mentioned previously,
this is because these devices can produce extremely precise movements. When
calibrated properly, they can also be capable of extremely accurate scanning;
however, various issues with piezoelectric materials can give rise to significant
nonlinearity in their real-world operation (8,9). Piezoelectric materials expand
or contract on the application of an electrical potential, and the materials used
typically have expansion coefficients of the order of 0.1 nm per applied volt. Thus,
it is simple to use piezoelectrics to scan an AFM probe with the precision
required to image individual atoms with the atomic force microscope. However,
some problems associated with piezoelectrics include nonlinear response, hyster-
esis, and creep. These problems lead to distorted images and difficulties in per-
forming zoom-to-feature, which is an important requirement for any microscope.
To overcome some of these problems, a variety of different piezoelectric device
geometries has been described, as well as the use of piezoelectrics in more com-
plex scanning devices, which are designed to improve linearity of scanning (10).
Ultimately, the best linearity is currently found in instruments that add a sensor
to the scanning mechanism and use the signal from the sensor to control the
scanning. This is usually done in a ‘‘closed-loop’’ configuration for the X-Y scan-
ning. More details about how this is done along with the description of some dif-
ferent sensor designs are given in Reference 4.
2.2. Topographic Scanning Modes. Topographic scanning is the most
important and common experiment in atomic force microscopy. Most AFM
images presented are actually topography images. Often they are shown in
three-dimensional or light-shaded views, but nonetheless these images are
based on topographic height maps. This is perhaps so widely used because,
despite all the useful complementary information that can be obtained using
other modes and signals, normally topographic images are the only ones
6 ATOMIC FORCE MICROSCOPY– AFM
It is a widely held assumption that contact mode is a poor choice for soft
samples because of the normal forces mentioned previously. This is not always
true, and there are many examples where contact mode has been shown to be
suitable for imaging such samples as living cells (12–14), soft polymers (15,16),
and individual molecules (17–19). In fact, the lateral forces mentioned tend to
cause more problems. For example, imaging weekly adhered samples on a
solid substrate can present serious problems because of a ‘‘sweeping’’ action of
the probe associated with these lateral forces. Despite these issues and the
increasing use of other modes, contact mode is often still used today for specific
applications, and because of certain advantages it has as topographic mode. For
example, contact mode is the fastest scanning AFM mode, since the feedback sys-
tem does not have to sum oscillations before it can act. In normal circumstances,
contact mode can often scan at two or three times the speed of intermittent
contact mode. Furthermore, contact mode scanning has been developed with
high-speed scanners into so-called video rate AFM, allowing the user to collect
whole frames in milliseconds (19,20). Another strength of contact mode is high
ATOMIC FORCE MICROSCOPY– AFM 7
resolution. Contact mode has been successful in attaining true atomic resolution
(21) and submolecular resolution on individual molecules in liquid (17,21,22),
and it is used commonly for demonstrations of atomic lattice resolution, which
is simple to achieve (23–25).
Oscillating Probe Modes. It was realized at the start of the development
of AFM that advantages could be gained by oscillating the probe while scanning
(11). The advantages of this are twofold. First, monitoring the oscillation of a can-
tilever is a much more sensitive way of measuring the forces acting on it than
measuring the deflection directly. This allows the implementation of so-called
‘‘noncontact’’ techniques, where feedback can be maintained without the probe
ever coming into repulsive contact with the sample surface. The second reason
is that these modes effectively eliminate the lateral forces that exist in contact
mode AFM. In this way, it can be considerably easier to image weakly adhered
samples and to reduce tip and/or sample damage during scanning. Because of the
simplicity of operation with a wide variety of samples, these modes have become
popular and are now undoubtedly applied much more often than contact mode
for topographic imaging applications.
Oscillating probe methods use an additional piezoelectric device (sometimes
called a ‘‘dither piezo’’), close to the probe substrate, to oscillate the probe verti-
cally, usually close to its natural resonant frequency. The resonant frequencies of
commercial probes for oscillating modes typically vary between approximately 60
and 400 kHz. In most instruments, the oscillation is monitored by feeding the
signal from the photodetector into a lock-in amplifier, which measures the ampli-
tude and phase of oscillation. When the probe approaches the sample, both the
phase and amplitude of oscillation will be affected. In most instruments, the
amplitude is used as the error signal in oscillating modes. The amplitude will
decrease as the tip starts to come into intermittent contact with the sample sur-
face during oscillation. The feedback electronics would then maintain this
decrease in amplitude at the preselected setpoint. The same can be done using
the phase of oscillation as the error signal. If the probe does not actually touch
the sample, then interaction usually occurs by long-range forces damping oscilla-
tion and reducing the frequency of operation. In this case, the amplitude or phase
at a fixed frequency will typically also be altered, but it is also possible to monitor
change in frequency more directly. However, this requires additional electronics
for most instruments (26,27). Intermittent-contact AFM, which is named differ-
ently by each manufacturer of AFM instruments (28) but that typically involves
large oscillations (10–100 nm),and amplitude as the error signal, is probably the
most widely used AFM imaging mode today. It is applied widely, especially in
ambient environments, and it can be used to image almost any sample. It can
also be applied in liquid, although some complication in selecting the most appro-
priate resonant frequency can occur (29). It gives medium resolution, being
typically incapable of resolution less than 1 nm. A common application of inter-
mittent contact AFM is so-called phase imaging, which means using the ampli-
tude as the error signal and recording the phase of oscillation, which can have
sensitivity to material properties such as viscoelasticity (30–32). A related oscil-
lating mode is carried out by using direct frequency detection and very small
oscillations (1–10 nm). Under these circumstances, the technique is usually
called noncontact AFM. Noncontact AFM is often applied in vacuum, where it
8 ATOMIC FORCE MICROSCOPY– AFM
has considerable advantages and can give true atomic resolution (33,34). Non-
contact AFM can also be applied in liquids, and although it is used less often
than intermittent contact AFM, it can give extremely high-resolution imaging
in liquid (35).
2.3. Nontopographic Modes. One of the major strengths of AFM is
that there are many ‘‘secondary modes’’ that measure properties other than topo-
graphy. These modes, like the topographic modes discussed previously but unlike
force spectroscopy (discussed in the next section), are imaging modes, but the sig-
nals they generate give more information than the basic topography of standard
AFM. Many of these modes can be carried out with a standard AFM instrument,
only requiring the mode to be enabled in software or the use of special probes.
Lateral Force Microscopy. Lateral force microscopy (LFM) is a technique
that is extremely simple to carry out, requires no special equipment, and can give
material contrast on many samples. Lateral force microscopy works by measur-
ing the lateral bending (twisting) of the cantilever as it scans the sample. This is
done by comparing the segments on the left and right of the photodetector shown
on Figure 2. Twisting of the cantilever always occurs as the sample is scanned,
and it shows changes in regions of changing slope but is also sensitive to the fric-
tion force between the probe tip and the sample. For this reason, LFM is some-
times called friction force microscopy (FFM). Unfortunately, the topographic
sensitivity of the LFM signal can often complicate the interpretation of possible
frictional differences in regions of the samples, unless the sample is flat. How-
ever, lateral force microscopy has been applied successfully to measure friction
on organic monolayers (36), friction between two monolayers (37), and discrimi-
nation of phases in polymer blends (38). It is used often to generate contrast at
the atomic scale (39). With care, the resulting signal can be calibrated to give
important frictional measurements of sample surfaces (40).
Mechanical Property Modes. In addition to LFM, which can measure
friction, a variety of modes are sensitive to the mechanical properties of the sam-
ple. Probably the most commonly used of these is phase imaging, which as pre-
viously mentioned is carried out by imaging in amplitude-modulation-based
oscillating mode while recording the phase shift. This mode is certainly sensitive
to different materials, and it is thought that the contrast is based on adhesion
and viscoelastic properties of the material under study (31). Like many AFM
techniques, it is also sensitive to topographic changes, so care must be taken
in interpreting phase images. Other techniques that can generate images show-
ing mechanical properties include pulsed-force AFM (41) and jumping-mode
AFM (42). Both of these techniques involve ramping up the z position of the
probe while scanning but at a rate lower than the resonant frequency. By mea-
suring the slope of the cantilever deflection versus distance, a parameter related
to sample stiffness is obtained, which can be displayed in a separate channel to
the topography image. Similarly, the ‘‘pull-off height’’ can be plotted to show
tip-sample adhesion properties. Both of these parameters are explained in the
section on force spectroscopy because they are more often determined by force
spectroscopic (ie, one-dimensional instead of three-dimensional) experiments.
Neither jumping mode nor pulsed force mode is widely available because they
require specific instruments or additional hardware to implement, which
explains why they are used less often than phase imaging.
ATOMIC FORCE MICROSCOPY– AFM 9
Fig. 4. Illustration of lifting modes used in MFM and EFM. Typically, a first scan (left)
measures sample topography, which is then used to enable scanning at a set height above
the surface (right). This second scan records the magnetic (or electrostatic) field informa-
tion, free of interference from short-range forces.
10 ATOMIC FORCE MICROSCOPY– AFM
Fig. 5. Simplified force-distance curve. The measurement of the curve begins with the
probe far from surface (a), where there is no tip-sample interaction force and no deflection.
At (b), the probe is attracted to the surface and ‘‘snap-in’’ occurs as the probe jumps onto
the sample. Here, the force is attractive. As movement continues, the sample applies
repulsive forces to the probe and, thus, the cantilever deflects in the opposite sense, until
movement is reversed at (c). Typically, adhesive tip-sample interactions mean that as the
probe is retracted, the tip remains on the sample until these forces are overcome, when at
(d), ‘‘pull-off’’ occurs. As shown, the slope (sample stiffness) and adhesion (tip-sample
interaction force) data can be measured from such a curve.
can be carried out in a grid pattern over a sample surface to generate a map of
interaction forces. In this mode, it is sometimes called chemical force microscopy
(CFM) (58,59).
Nanoindentation. During the acquisition of force curves, as shown in
Figure 5, there are two phases: the approach and the retract segments. By mon-
itoring what happens during the approach phase while the probe tip is in physi-
cal contact with the sample, we can carry out nanoindentation experiments. This
quantitative technique, in principle, can give useful mechanical parameters of
the sample such as spring constant or Young’s modulus (60,61). It is possible
to carry out nanoindentation experiments and then to image the resulting
indents in the sample at high resolution to gain insight into deformation
mechanisms (62). A major advantage of AFM for nanoindentation experiments
is that it is possible to use the highly precise positioning capabilities of AFM to
carry out such experiments in specific parts of the sample. For example,
researchers made nanoindentation measurements on individual microorganisms
(63) and on nanoparticles (64). Furthermore, multiple curves can be acquired in a
grid pattern to enable the mapping of the hardness of samples, for example, on
heterogeneous polymer films (16,65,66). A disadvantage of AFM-based nanoin-
dentation experiments is that only estimated values of the properties measured
can be obtained unless calibration of both cantilever spring constant and tip-
shape are made. Data analysis can be complicated also (19). The tip-shape pro-
blem is often overcome by using a tipless cantilever with a spherical colloid glued
to it in place of the more normal sharp tip.
12 ATOMIC FORCE MICROSCOPY– AFM
The range of applications of AFM in chemistry is vast. It has been shown to have
many uses not only in the traditional disciplines of physical, organic, inorganic,
and analytical chemistry but also in more applied and specialized areas such as
food science, surface science, nanotechnology, biochemistry, and even in chemi-
cal industry. In this article, therefore, I will not try to cover the range of applica-
tions in which AFM is used, but instead just a few examples are described to
illustrate the capabilities of AFM with direct relevance to chemists.
3.1. Electrochemical AFM. One technique of most direct relevance to
chemistry is that of electrochemical AFM. In reality, this technique makes use
of standard AFM techniques but with a specialized sample cell to allow electro-
chemical modifications and measurements to be made simultaneously with AFM
measurements. Typically, a small. sealed liquid cell is used with connections for
the electrodes to bias the sample versus a reference electrode. These can be con-
structed by the atomic force microscope user and are commonly available for pur-
chase from the major atomic force microscope manufacturers, requiring only the
addition of a potentiostat.
With such a setup, it is possible to carry out in situ imaging of electroche-
mical processes occurring on sample surfaces, typically at metal electrode sur-
face. By ramping up the applied potential during or between AFM scans, it is
possible to observe the results of reduction and oxidation processes directly at
the electrode surface. Such processes tend to give rise to small changes, so it is
useful to use AFM to observe them. Using the AFM probe (typically with a con-
ducting coating), it is also possible to carry out electrical measurements at the
nanoscale, which is referred to as scanning electrochemical AFM. An example
showing images collected during reduction of gold on an electrode surface is
shown in Figure 6.
3.2. High-Resolution Imaging. Most AFM imaging does not actually
use the high-resolution capabilities of AFM, but instead it involves collecting
images with sizes in the range of 2 to 50 mm, meaning resolution does not nor-
mally go beyond 5 nm. One reason for this is that for useful high-resolution ima-
ging, it is necessary to have a well-organized and clean sample. Another reason is
that achieving higher resolution can be difficult, and it requires great care and
persistence. In some cases, specialized AFM techniques are required, meaning
that sometimes these capabilities are beyond the scope of a typical commercial
atomic force microscope scanning in ambient conditions.
However, this is not always the case. For example, an experiment that
is carried out routinely with atomic force microscopes to demonstrate high-
resolution imaging is atomic-lattice imaging. In this type of imaging experiment,
the atomic force microscope enables imaging of the repeating pattern of an
atomic lattice of a well-organized material. The most common material for this
experiment is muscovite mica because it is extremely simple to prepare and
has a rather large lattice periodicity, making it relatively simple to measure.
An example image of this sample showing the atomic lattice is shown in
Figure 7. In addition to mica, other samples that have been imaged in this
way include the Au (111) surface, salt crystals, highly ordered pyrolytic graphite,
and many others (21). These experiments can be carried out in ambient
ATOMIC FORCE MICROSCOPY– AFM 13
Fig. 6. Electrochemical AFM example. Images showing the morphology of a CdTe film
during electrochemical deposition of Au, at various times of applying a potential of
0.35 V. Electrochemical AFM allows the user to follow the progression of electrochemical
processes such as this in situ. Reproduced with permission from Reference 67.
conditions or in liquid at room temperature using contact mode. The only special
care that is required is the use of a sharp tip, removal of all sources of noise,
thermal equilibrium, and typically, extremely fast scanning to overcome residual
thermal expansion. It is important to remember that these images do not in
fact show true atomic resolution but instead show the average arrangement of
atoms on a surface. Thus, this ‘‘atomic lattice resolution’’ AFM imaging, as we
Fig. 7. Examples of high-resolution imaging with the AFM. (a) Example of lattice resolu-
tion on mica measured in liquid in contact mode. Inset shows the FFT pattern. Repro-
duced with permission from Reference 68. (b) True atomic resolution image of (Si, Sn,
and Pb alloy) measured using noncontact AFM in vacuum. Reproduced with permission
from Reference 69.
14 ATOMIC FORCE MICROSCOPY– AFM
shall refer to it here, does not allow imaging of individual adatoms or of atomic
vacancies.
However, true atomic resolution can be achieved with AFM, although it is
usually achieved using so-called FM detection (which allows the precise
control of the frequency of oscillation in an oscillating mode) (33) and in most
cases operation in ultra high vacuum and sometimes at cryogenic temperatures
(69–71). These conditions enable the imaging of atomic vacancies, dislocations,
stacking faults, submonolayers, and so on (2,21). Furthermore, this technique
can even reveal subatomic detail, for example, the spectacular imaging of mole-
cular orbitals (72). Some recent advances allowed atomic resolution imaging in
liquids as well (35). In addition to cases where atomic resolution can be obtained,
submolecular information is obtainable under more relevant conditions (eg, with
hydrated samples) in the case of large molecules such as proteins, protein
complexes (22,73,74), and polymers (75,76).
3.3. Molecular Interactions. One of the most important abilities of
AFM is that it can sensitively measure tip-sample interaction force. The interac-
tion between the material on the AFM probe tip and the material on the sample
surface can give useful information about the nature of the sample. Further-
more, by attaching a species of interest to the AFM tip, the force of interaction
it has with some other species on the sample can be measured. This could be, for
example, molecule–molecule interactions or colloid–colloid interaction forces, or
even the force of interaction between an animal cell and a biomaterial. Because
the atomic force microscope is so sensitive, in principle, even an interaction
between single pairs of molecules can be detected. This technique, which as men-
tioned previously is referred to as force spectroscopy, has a huge range of poten-
tial applications, from surface science, through chemistry to biology. For
chemistry, however, probably the most interesting applications are those of sin-
gle-molecule interaction studies. These studies can be used also in a mapping-
type configuration, for example, to use molecular interactions to locate single
molecules on the surface of living cells (77,78). Probably the most common use
for this type of molecule–molecule interaction is in biological systems such as lec-
tin–carbohydrate interactions or antibody–antigen binding, but small molecules
can also be probed. For example, researchers studied hydrophobic interactions
between hexadecane molecules (34,79) and carbohydrate-carbohydrate interac-
tions (54) with this technique. Typically for this application, the molecule to
interact will be bound to the tip and possibly to a flat substrate (for self-interac-
tions), often being coupled by a flexible linker to increase the chance of the
molecules binding in appropriate configurations.
3.4. Nanolithography. In addition to imaging, the ability of AFM to
control finely the position of the probe close to a surface and to control tip-sample
force can be used to modify samples. Since the first experiments in AFM, it was
noticed that the sample could be changed by interaction with the tip. In the inter-
vening years, many techniques have been developed to modify and construct
nanostructures with the atomic force microscope. Some are unsophisticated tech-
niques, such as nanoscratching, which is a simple way to created patterns in soft
materials, and assembly of nanostructures. The latter technique involves using
the AFM probe to move nanoparticles, nanotubes, and so on around on the
surface, and it can be used for applications such as wire building or moving
ATOMIC FORCE MICROSCOPY– AFM 15
Fig. 9. Friction and phase imaging of examples. (a) Friction image (difference of left to
right and right to left scans) on a filled polymer films showing low friction on additive
articles. Adapted with permission from Reference 92. (b) Topography image of crystalliz-
ing polymer with spherulites, which give low contrast in height. (c) Phase image of the
same area, showing clear contrast on the spherulites. Reproduced with permission from
Reference 94.
ATOMIC FORCE MICROSCOPY– AFM 17
imaging is working well (95). Because it is simple to apply and sensitive, phase
imaging has been applied to a wide range of polymer samples, including block
copolymers (96,97), imaging spherulites in crystallizing polymers (94), polymer
blends (98), and more. However, in general it is difficult to identify features
using phase imaging without some prior information to help, such as the propor-
tions of the various components, and so on (98). Even with prior knowledge of the
response of the bulk materials in phase imaging, it can be problematic to identify
the material using phase imaging definitely because the phase signal is highly
dependent on experimental setup and phase signal reversal can even occur
under certain circumstances (99).
4. Conclusion
BIBLIOGRAPHY
‘‘Atomic Force Microscopy-AFM’’ in ECT 5th ed., Vol. 5, pp. 319–341, by Charles Anthony
Peterson, Intel Corporation and Dror Sarid, Optical Sciences Center, University of
Arizona.
62. Y. Gaillard, C. Tromas, and J. Woirgard, Acta Mater. 51, 1059–1065 (2003).
63. P. Eaton and co-workers, Ultramicroscopy 108, 1128–1134 (2008).
64. H. P. Wampler and A. Ivanisevic, Micron 40, 444–448 (2009).
65. M. S. Bischel and co-workers, J. Mater. Sci. 35, 221–228 (2000).
66. M. R. Vanlandingham and co-workers, J. Adhes. 64, 31–59 (1997).
67. R. Vidu, J.-R. Ku, and P. Stroeve, J. Colloid Interface Sci. 300, 404–412 (2006).
68. Y. Kuwahara, Phys. Chem. Miner. 28, 1–8 (2001).
69. Y. Sugimoto and co-workers, Nature 446, 64–67 (2007).
70. R. Garcia, Wiley VCH Verlag GmbH & Co. KGaA, Berlin, Germany, 2010,Chapt. 1.
71. L. Gross and co-workers, Nat. Chem. 2, 821–825 (2010).
72. F. J. Giessibl and co-workers, Annalen Der Physik 10, 887–910 (2001).
73. J. X. Mou and co-workers, Biophys. J. 71, 2213–2221 (1996).
74. D. J. Muller and Y. F. Dufrene, Nat. Nanotechnol. 3, 261–269 (2008).
75. J. Kumaki, S. Sakurai, and E. Yashima, Chem. Soc. Rev. 38, 737–746 (2009).
76. N. H. Thomson, in B. Bhushan, H. Fuchs, and S. Hosaka, eds., Applied Scanning
Probe Methods, Vol. VI: Characterization, Springer-Verlag, Berlin, Germany,
2006, pp. 127–164.
77. Y. Gilbert and co-workers, Nano Lett. 7, 796–801 (2007).
78. L. A. Chtcheglova and co-workers, Biophys. J. 93, L11–L13 (2007).
79. C. Ray and co-workers, J. Phys. Chem. C 112, 18164–18172 (2008).
80. C. Ho-Yin, X. Ning, Z. Jiangbo, and L. Guangyong, 5th IEEE Conference on Nano-
technology, Vol. 2, 2005, pp. 713–716.
81. T. Junno and co-workers, Appl. Phys. Lett. 72, 548–550 (1998).
82. D. Stievenard and B. Legrand, Prog. Surf. Sci. 81, 112–140 (2006).
83. D. S. Ginger, H. Zhang, and C. A. Mirkin, Angew. Chem., Int. Ed. 43, 30–45 (2004).
84. X. N. Xie and co-workers, Mater. Sci. Eng. R 54, 1–48 (2006).
85. R. V. Martinez and co-workers, Nano Lett. 7, 1846-1850 (2007).
86. A. Noy and co-workers, Nano Lett. 2, 109–112 (2002).
87. R. D. Piner and co-workers, Science 283, 661–663 (1999).
88. K. L. Christman, V. D. Enriquez-Rios, and H. D. Maynard, Soft Matter 2, 928–939
(2006).
89. L. Fu and co-workers, Nano Lett. 3, 757–760 (2003).
90. R. V. Martinez and co-workers, Adv. Mater. 22, 588–591 (2010).
91. C. Ton-That and co-workers, Polymer 42, 1121–1129 (2001).
92. B. D. Beake, G. J. Leggett, and P. H. Shipway, Surf. Interface Anal. 27, 1084–1091
(1999).
93. S. E. Morgan, R. Misra, and P. Jones, Polymer 47, 2865–2873 (2006).
94. J. Xu and co-workers, Macromolecules 37, 4118–4123 (2004).
95. Spm refererences and standards (afmhelp.com) http://bit.ly/spm_standards.
Accessed 25th May 2011.
96. Z. W. Xu and co-workers, Ultramicroscopy 105, 72–78 (2005).
97. L. Peponi and co-workers, Macromol. Mater. Eng. 293, 568–573 (2008).
98. D. Raghavan and co-workers, J. Polym. Sci. Polym. Phys. 39, 1460–1470 (2001).
99. G. Bar and co-workers, Langmuir 13, 3807–3812 (1997).
PETER EATON
Requimte