PRODUCT .DATA
SCALABLE .AND. AFFORDABLE. FLOOR-MOUNTED .RADIOGRAPHY. SYSTEM.
Feature
Thanks to its ergonomic
design and operator friendly
controls, the system is very
easy to use. Also, HF-525
PLUS designed for total cost
saving, user friendly,
minimized training efforts
and easy maintenance. A
high-frequency generator
with Automatic Exposure
Control and an Anatomical
Program enables optimized
X-ray techniques without
failure.
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr
HF-525 PLUS.
P R O D U C T ..D A T A.
SYSTEM CONFIGURATION
SYSTEM 1 SYSTEM 2 SYSTEM 3 SYSTEM 4 SYSTEM 5 SYSTEM 6
HFG 32kW 40kW 50kW 50kW 65kW 80kW
POWER OUTPUT 125 kV 150 kV
Max. mA 400 mA 500 mA 630 mA 800 mA 800 mA 1000 mA
220V-230VAC, 50-60Hz,
LINE VOLTAGE 380VAC, 50-60Hz, Three Phase
Single Phase
CONTROLLER Invertor Type
TUBE 140kHU 300kHU 400kHU
TUBE STAND Floor Mounted Type
BUCKY TABLE 4 Way Floating Top Bucky Table
BUCKY STAND Wall Bucky Stand
O/P CONSOLE FnD console LCD console
OPTION
ITEMS SYSTEM 1 SYSTEM 2 SYSTEM 3 SYSTEM 4 SYSTEM 5 SYSTEM 6
TUBE**
Toshiba 300kHU ● ●
Toshiba 400kHU ● ● ● ● ●
Toshiba 600kHU ● ● ● ● ● ●
HFG
32 kW / 3 phase ● ●
40 kW / 3 phase ● ●
50 kW ● ●
65 kW ● ● ● ●
80kW ● ● ● ● ● ●
TUBE STAND
Ceiling Suspended
● ● ● ● ● ●
Tube Stand
BUCKY TABLE
6 Way
● ● ● ● ● ●
Elevating Table
PSU : Power Supply Unit
Battery Bank ● ●
Capacitors ● ●
AEC : Auto Exposure Control
AEC board ● ● ● ● ● ●
3 Field Ion Chamber ● ● ● ● ● ●
DAP : Dose Area Product
DAP chamber ● ● ● ● ● ●
OP CONSOLE
LCD console ● ● ●
Touch LCD console ● ● ●
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr
HF-525 PLUS.
P R O D U C T ..D A T A.
Radiographic 10, 25, 32, 40, 10, 25, 32, 40, 10, 25, 32, 40, 10, 25, 32, 40, 10, 11, 12.5 10, 11, 12.5
mA range 50, 63, 80, 50, 63, 80, 50, 63, 80, 50, 63, 80, 14, 16, 20, 22, 14, 16, 20, 22,
100, 125, 160, 100, 125, 160, 100, 125, 160, 100, 125, 160, 25, 28, 32, 36, 25, 28, 32, 36,
200, 250, 320, 200, 250, 320, 200, 250, 320, 200, 250, 320, 40, 45, 50, 56, 40, 45, 50, 56,
400 400, 500 400, 400, 500, 400, 400, 500, 63, 71, 80, 90, 63, 71, 80, 90,
630 630, 800 100, 110, 125, 100, 110, 125,
140, 160, 180, 140, 160, 180,
200, 220, 250, 200, 220, 250,
280, 320, 360, 280, 320, 360,
400, 450, 500, 400, 450, 500,
630, 710, 800 630, 710, 800,
900, 1000
0.1 ~
mAs Range 0.1 ~ 400mAs 0.1 ~ 500mAs 0.1 ~ 630mAs 0.1 ~ 800mAs 1,000mAs
High Voltage
<1kV@100kV
Ripple(TYP)
Automatic Exposure
Control Optional
(up to 2 chambers)
Power Supply Unit Optional (battery type
-
(PSU) / capacitor type)
Anatomical
288 APR (Standard) 1,290 APR
Programming
220-230VAC, 50/60Hz
Line Voltage (phase) 380VAC, 50/60Hz, Three phase
, Single Phase
Automatic Line
± 10%
compensation
*For the proper performance of generator, the X-ray room must fulfill with 130A main circuit breaker.
3
Page
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr
HF-525 PLUS.
P R O D U C T ..D A T A.
Vertical tube travel: Approx. 1,500mm (400 ~ 1,900mm from the ground level)
Longitudinal travel: Approx. 3,000mm or more (depends on the ceiling rail length)
Transverse travel: Approx. 2,000mm or more (depends on the ceiling rail length)
Vertical tube travel: Approx. 1,500mm (900 ~ 2,400mm from the ceiling to Focal Spot)
X-ray tube rotation: +/- 180° (detents at +90° ~ -90°)
Tube stand column rotation: +/- 180° (detents at +90° ~ -90°)
Locks: Electromagnetic brakes
Tube arm controls: Longitudinal, transverse, vertical, and rotational lock control buttons
COLLIMATOR
CATEGORY SPECIFICATION
Collimator type: Manual square field X-ray collimator
Field lamp: Halogen lamp 24V AC 150W
Lighting timer : Less than 30 sec
Average luminosity: Over 160lux at SID 100Cm
Inherent filtration: 1.2mm aluminum
Added Filtration Added Filtration
Option Tape measure
4
Page
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr
HF-525 PLUS.
P R O D U C T ..D A T A.
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr
OPTIONAL ITEMS
AEC, Auto Exposure Control
SSMC 601 (Claymount)
The SSMC 601 is placed between the patient and the imaging device (film, detector) in order to detect the actual dose
value for the automatic exposure control (AEC) an error free image is guaranteed even at low kV. Its attenuation factor is
very low. X-ray scattering has been brought back to the bare minimum, to improve the brightness of the image.
The KermaX® plus indicates the product of dose and radiation field area in air irrespective of patient distance from the X
ray tube as shown in the figure below.
Page 6
**The real time display of DAP values are shown on OP console immediately after X-ray exposure.
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr
12 : 1 GRID
COMPRESSION BAND
LATERAL CASSETTE HOLDER
HAND GRIP FOR TABLE
OPERATING CONDITIONS
Storage Conditions
Ambient Temperature : -10°C to 50°C
Relative Humidity : 30% to 80% (no condensation)
Atmospheric Pressure : 700 hPa to 1060 hPa
Operating Conditions
Ambient Temperature : 18°C to 30°C
Relative Humidity : 30% to 75% (no condensation)
Atmospheric Pressure : 700 hPa to 1060 hPa
Atmosphere : Do NOT use the system in an atmosphere in an potentially
explosive or corrosive gases may be present
ECORAY ┃ 3F, Urbanlight B/D, Eonju-ro, Gangnam-gu, Seoul, Korea 135-822 / www.ecoray.kr