Professional Documents
Culture Documents
beasercer
Química Analítica II
3º Grado en Química
Facultad de Química
Universidad de Sevilla
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
1- Galvanic cell and electrolytic cell. Fundamentals, differences and applications.
2- Types of electrodes and examples.
3- Electrode processes. What is the double layer, what polarization is and what is its
importance.
4- Ideally non-polarizable electrode and ideally polarizable electrode.
5- Mechanisms of mass transfer.
Potentiometry
a64b0469ff35958ef4ab887a898bd50bdfbbe91a-1921780
Key-points-second-term-2019-Ruth...
beasercer
Química Analítica II
3º Grado en Química
Facultad de Química
Universidad de Sevilla
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
A) Introduction to electroanalytical methods.
1. Galvanic cell and electrolytic cell: fundamentals, differences and applications.
A galvanic cell uses a spontaneous electrochemical reaction to generate electricity. To
accomplish this, one reagent must be oxidized and another must be reduced. The two cannot be
in contact, or electrons would flow directly from the reducing agent to the oxidizing agent.
Instead, the oxidizing and reducing agents are physically separated, and electrons are forced to
flow through an external circuit to go from one reactant to the other. It’s used on potentiometry.
An electrolytic cell is the inverse of the galvanic cell: it uses electricity (potential is applied)
to generated a non-spontaneous an electrochemical reaction. It’s used on voltammetry or
coulometry.
3. Electrode processes. What is double layer, what polarization is and what is its
importance.
a) Electrode processes.
ü Faradaic processes.
They are associated to the electronic transfer in the oxidation reaction that takes
place in one electrode and the reduction reaction occurring in the other electrode. These
processes originate the faradaic current. They follow the Faraday’s law and the amount of
product (Q) of reaction is:
Q = n · F · N; where:
n is number of electrons involved in the reaction.
F is the Faraday’s constant
N the number of moles.
ü Non-faradaic processes.
Under some conditions and due to thermodynamic or kinetic reasons cells show a
potential interval at which the faradaic processes are excluded. The surface of the
electrode behaves like capacitor. These processes originate the non-faradaic current. They
follow the Ohm’s law.
E=I·R
b) Double layer.
a64b0469ff35958ef4ab887a898bd50bdfbbe91a-1921782
Origin of the double layer: a charge on the electrode originates a charge of opposite sign
in the solution close to the electrode. This is called double electric layer. If negative potential
is increased more positive ions flow to the surface of the electrode to neutralize the charge of
it, producing a transient charge current to attain the equilibrium.
Composition of the solution adjacent to the electrode is different to the bulk solution. We
can difference two parts: the compact layer, with an excess of positive charge, in which
potential linearly decreases with the distance to the surface of the electrode, and the diffuse
layer, in which potential is exponentially diminished. The compact layer has two parts. The first
one is the inner Helmholtz layer, that may include solvent and any adsorbed solute molecule as
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
neutral molecules, anions, or cations. The second one is the outer Helmholtz layer with solvated
cations.
c) Polarization.
Polarization is the variation of the linear relationship that exists between the intensity of
current of the cell (I) and the applied potential (E), due to a decrease of the current, by the
decrease in the speed of some stage of the electrical process. Separation of charges occurs.
The degree of polarization is measured with the overpotential, and is the deviation of the
electrode from its theoretical value at equilibrium. It is taken as a negative value.
B) Potentiometry.
1. Characteristics of a potentiometric measurement instrument.
Potentiometry techniques are based on the measurement of the potential of an analyte in
the absence of current to determine the concentration of an analyte in a sample. The
potentiogram is the graph where the potential is represented as a function of the time.
The instrumentation required to perform these measurements is a galvanic cell made by a
reference electrode, and a standard electrode connected by a salt bridge, and a device capable
of measuring the potential.
Reference electrodes are electrodes whose potential (Eref) fixed, so they don’t change with
the temperature, the current that goes through the electrode, or the concentration of the
analyte or any other ion on the matrix of our sample.
The working electrode’s potential changes (Esam) with the analyte’s activity.
The salt bridge is the connection between both electrodes, which doesn’t allow the ions of
the reference electrode and the working electrode to mix by them. In each of the side of the
salt bridge may develop a potential. This potential, if the mobility of the anion and the cation on
the salt bridge are more-or-less the same, the potential may be same, so the difference in
potential through the bridge, the junction potential (Ej) is cancelled.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
The reference electrode is an electrode whose potential is fixed. It provides a stable and
known potential, so that we can attribute any change in Ecell to the analyte’s effect on the
indicator electrode’s potential.
ü Standard Calomel Electrode (SCE).
It consists of an inner tube made with of Hg, Hg2Cl2, and KCl,
situated inside of a second tube that contains a saturated solution of
KCl. A small hole connects the two tubes, and a porous wick serves as a
salt bridge. It is based on the following redox reaction:
Hg2Cl2 (s) + 2e- ® 2Hg (l) + 2Cl-(aq)
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
electrolyte solution, with a Ag/AgCl reference electrode immersed
inside, enclosed in a semipermeable membrane. When CO2 diffuses
through the semipermeable membrane, it lowers the pH in the
electrolyte solution. The response of the glass electrode to the
change in pH is a measure of the CO2 concentration outside the
electrode. Other acidic or basic gases, including NH3, SO2, NOx can
be detected in the same manner.
ü b) Biocatalytic sensors.
They respond biochemically to an important species. The most
common are enzyme electrodes, in which an enzyme is trapped in a gel or immobilized by
adsorption on a porous layer, at the surface of a potentiometric electrode. The analyte’s
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
one, a hygroscopic gel layer, hydrated by the reference
solution; the intermediate one or “dry sheet”, where Na+
cations conduct the electricity; and the outer one, the
same that the inner one, but hydrated by the external
analyte’s solution, and not by the internal reference
solution. So then, the conduction depends on the
movement of H+ ions through the hygroscopic layers.
The potential difference between internal and external Ag/AgCl reference electrodes
depends on the |Cl-| concentration in each electrode compartment, and on the potential
difference across the glass membrane. Due to |Cl-| is fixed in each compartment, and due to |H+|
is fixed on the inside of the glass membrane, the only variable is the pH of analyte solution
outside the glass membrane.
ü Fluoride electrode.
2. Types of voltammetry techniques and what is the main difference between them.
3. Description of the parts that a curve i-E presents. What does happen in each part?
a64b0469ff35958ef4ab887a898bd50bdfbbe91a-1921782
The voltammogram is the i-E sigmoid curve obtained (voltammetric
wave), where the current is represented as a function of the potential,
when it is sweep. Three parts can be considered:
a) Initial, with zero current. Potential is not negative enough to
produce reduction reaction.
b) Reaction takes place, and cathodic current is measured. E1/2: half-
wave potential. It is the potential at which half the maximum current is
reached.
c) At a more negative potential current becomes constant.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
4. Working electrodes in polarography: detailed description (only
DME)
It is formed by a capillary connected to a reservoir containing Hg. Suspend a drop of Hg from
the bottom of the capillary, and then the current and voltage are measured and the drop is
dislodged. Then another drop is suspended and the next measurement is made. The drop acts
like an electrode. Constant dropping occurs while potential is increased. Current increases while
size of the drop increases, but it decays when drop falls. Very quick increasing is produced with
the new drop. As a consequence, oscillation of current is produced. Mechanism of mass transfer
is only diffusion.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
7. Clark electrode: measurement of O2 in solution.
It is based on an amperometric measurement of the current produced between a pair of
-
electrodes that are driving the reduction reaction of O2 to OH , and is proportional to the
-
concentration of OH , in consequence, also to the analyte (O2). The cathode, a Pt working
electrode, is held at -0.6 V with respect to the anode, a Ag/AgCl reference electrode. The cell is
covered by a semi- permeable membrane, across which O2 can diffuse.
9. Types of coulometries.
ü Coulometry at a constant potential.
Potential is applied during enough time to oxidize
or reduce completely the analyte. Because analyte is
consumed, current decreases with the time in an
exponential form. Q is calculated from a graph of i vs
t (area under the curve).
But it is not necessary to complete the curve. Current is zero when 100% of electrolysis
$
is accomplished, and then 𝑄 = %. Constant of the cell, k, can be obtained plotting Ln i vs t.
&
𝑖 = 𝑖( · 𝑒 +&, → 𝐿𝑛 𝑖 = 𝐿𝑛 𝑖( − 𝐿𝑛 𝑒 · 𝑘𝑡 = 𝐿𝑛 𝑖( − 𝑘𝑡
Q = i · t = n · F · N = n · F · V · C; where:
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
ü Coulometry at constant current.
This coulometric method is also called coulometric titration. A reagent, produced
by an electrolytic process, is used in the procedure to react with the analyte. When
the end-point is reached keeping constant the current, the charge is obtained by
multiplying the current by time. An indicator system is necessary to detect the end-
point.
PARTITION
Separation is due to the “distribution” of the solute between the mobile and stationary
phases (relative solubility)
Stationary phases: bonded silica…
ION INTERCHANGE
Separation is due to the electrostatic attraction between the analyte ions and the opposite
charge ions located at the surface of an ion exchanger resin which acts as stationary phase
(polystyrene divinylbenzene)
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
Analyte separation is mainly based on its molecular size (sieving). Also known as gel filtration
chromatography
AFFINITY CHROMATOGRAPHY
Separation is based on highly specific interactions between the analyte and the called
affinity ligands, bounded to the stationary phase.
4. What are the causes of asymmetric band shapes? Could they be eliminated?
Asymmetry in band shapes can be be of two types: overloading and tailing.
The center isotherm is the ideal one, producing to a
symmetric peak.
The upper isotherm arises from an overloaded column in
which too much solute has been applied to the column. It
produces a long front of gradually increasing concentration
a64b0469ff35958ef4ab887a898bd50bdfbbe91a-1921782
3
• Longitudinal diffusion ( ).
45
Molecules tend to diffuse ahead and behind of the
chromatographic band, according to a concentration gradient.
The band slowly broadens as molecules diffuse from the high
concentration zones to lower concentration zones on the edges of
the band. The faster the flow, the less time is spent on the column
and the less longitudinal diffusion occurs. It is called longitudinal
diffusion because it takes place along of the column while the band
is transported by the mobile phase.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
It depends on Dm (diffusion coefficient), that is higher in a
gas than in a liquid, and then, this effect is bigger in gas
chromatography (B = 2 · 𝛾 · Dm).
E) HPLC.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
ADVANTAGES: no reconditioning, shorter time of analysis and reduced solvent consumption.
DISADVANTAGES: low resolution for difficult samples.
5. Common stationary phases for liquid-liquid partition chromatography. How are they
synthetized and their main mechanism of action according to its structure? Most
common types.
The most common support is silica, that is permeable to
solvent and have a huge specific area. Bare silica has a surface of
silanol (Si-OH) polar groups. Above pH 3, silanol groups are
dissociate to negative siloxane groups (Si-O-). This causes that
some polar compounds are retained, producing the presence of
important tails in the peaks of the chromatogram, which increases
the retention time of them, and getting worse its resolution.
To solve this problem, we use a technique called ENDCAPPING.
This technique consists in prepare "bonded silica", which is the
silica that is bound by the free O- of siloxane groups, to the Cl of a ClSiMe3, blocking the free
O- of siloxane groups. The polarity of the bound silica can be modified by bonding it with
different R groups, and the most common is C18.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
absorbance at all wavelengths simultaneously.
8. Draw a scheme of an evaporative light scattering detector for HPLC explaining how it
works.
It responds to any solute that is significantly less volatile than the
mobile phase. Eluate enters the detector at the top. In the nebulizer,
the eluate is mixed with nitrogen and an aerosol (dispersion of droplets)
is produced. Solvent evaporates in the heated drift tube, leaving a fine
mist of solid particles to enter the detection zone at the bottom. The
particles are detected by the light that they scatter from a diode laser
to a photodiode
The evaporative light scattering detector response is related to the
mass of the solute, not to the structure or molecular mass of the solute.
It is universal and is compatible with gradient elution.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
It is an ionization technique at
atmospheric pressure where the solvent-
analyte flow from the LC passes through a
positively charged and very narrow capillary,
and gets nebulized as microscopic, positively
charged solvent-analyte droplets. These
droplets fly towards the negatively-charged
faceplate, with solvent evaporating on the way,
until they disintegrate in a Coulomb explosion,
when the repulsive charge of their ionized
components exceeds their surface tension. The
individual ionized analyte molecules then pass through the faceplate entry hole into the mass
spectrometer.
a64b0469ff35958ef4ab887a898bd50bdfbbe91a-1921782
m/z ratio, being expelled from the trap. Changes in the electromagnetic field, will cause the ions
to be expelled in increasing order of their m/z ratio, giving rise to a mass spectrum.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
notable source of error. To minimize this
time, this system has to employ electric
fields to enclose the ions, and then release
them all at the very same time.
Another important fact would be the distance covered by the fragments. A further distance
would result on a higher resolution (because they take longer to reach the detector). To increase
it without make much bigger the analyzer, we use a reflectron, an optical device through which
the ions pass through and their path is reversed.
The benzene rings are modified to produce a cation-exchange resin, containing sulfonate
groups (-SO3-), or anion-exchange resin, containing ammonium groups (-NR3+).
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
peak is obtained.
In a mixture of weak acids, that of a higher pKa elutes before.
b) Buffer concentration.
When concentration of buffer in mobile phase increases, retention diminishes
c) Temperature.
Temperature decreases retention, because equilibrium is attained more rapidly.
d) Organic solvents.
Organic solvents in the mobile phase may reduce retention but it is difficult to predict.
3. Explain in detail what suppressed ion chromatography and ion chromatography without
suppression are.
No se permite la explotación económica ni la transformación de esta obra. Queda permitida la impresión en su totalidad.
For proteins is important to use an ionic strength high
enough (0.05 M) to eliminate electrostatic adsorption of
solute by occasional charged sites on the gel.