Professional Documents
Culture Documents
Features
Due to its wide measurement range and excellent long-term stability, Vaisala
DRYCAPâ Dew Point Transmitter DMT143L is an ideal choice for low dew point
industrial applications, such as compressed air dryers, plastic dryers, and other OEM
applications.
continuous measurement
Mechanical Specifications
Process temperature (°C) Mechanical connection G1/2" ISO228-1 with bonded seal ring
(U-seal) or NPT1/2" thread
Housing material Stainless steel (AISI 316L)
Dew Point Accuracy vs. Measurement Conditions
Weight
Response time 63 % [90 %] at +20 °C (+68 °F) gas temperature and 1 bar
pressure and 1 liter/min flow rate: G thread model 90 g (3.2 oz)
NPT thread model 100 g (3.5 oz)
-60 → -20 °C Td (-76 → -4 °F Td) 5 s [10 s] (typical)
-20 → -60 °C Td (-4 → -76 °F Td) 45 s [10 min] (typical) AW24 G1/2" ISO228/1 mm
[in]
Accuracy with DRYCAP 180S â
±2 °C (±3.6 °F) 1)
ø28 [1.10]
1) When the dew point is below 0 °C (32 °F), the transmitter outputs frost point.
ø12 [0.47]
41 [1.61]
Dew point temperature (°C)
70
60 57 [2.24] 51 [2.01]
50
ne
40 ration li 120 [4.72]
30 te r v apor satu
W a
20
10 DMT143L with G1/2" Thread
0
-10 Accuracy < ± 2°C Td
-20 AW30 NPT1/2" mm
Accuracy not specified
-30 [in]
-40
lower limit
ø28 [1.10]
0 10 20 30 40 50 60 41 [1.61]
Process temperature (°C)
49.1 [1.93] 51 [2.01]
120 [4.72]
Compliance
DMT143L with NPT1/2" Thread
IP rating IP66
EMC compliance EN61326-1, industrial environment
All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.
www.vaisala.com Any reproduction, transfer, distribution or storage of information contained in this document is strictly
prohibited. All specifications — technical included — are subject to change without notice.