Standard Calibration Procedure
Doc. No. Call/SCP/004 Rev. 00 May 01, 2015
Decade Resistance Box
This document is the property of Qstar Engineering, and is issued only for persons or entity as
written on this document. It is not to be retransmitted, printed or copied without prior written
permission of company.
Table of Contents
1.0 Scope 03
2.0 Standard Equipment 03
3.0 Preliminary Operations 03
4.0 Calibration Procedure 03
5.0 Related Documents 03
Page 2 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/004 Rev. 00 May 01, 2015
Decade Resistance Box
1.0 Scope
This method defines the procedure for calibration of Decade resistance Box.
2.1 Digital Multi Meter with a valid certificate of calibration traceable to international standard
in accordance with an industrial code or a regulation shall be used as a standard for
calibration.
2.2 The master standard shall be kept under controlled environmental conditions e.g. dust
free environment and Lab. temperature 22°C ± 3°C. Lab Humidity 60%RH ± 20%RH.
3.1 The details of Decade Resistance Box received for calibration shall be entered in the
Calibration Job Form (Form: call/001).
3.2 Inspection of the Decade Resistance Box for general condition shall be carried out.
3.3 The Decade Resistance Box shall be inspected and marked () in relevant block of
equipment’s primary inspection section of Calibration Data Sheet (Form: call/013).
4.1 The Decade Resistance Box to be calibrated shall be maintained at same place of
standard Digital Multi Meter for 03 hours minimum.
4.2 Decade Resistance Box shall be connected to appropriate terminal of the Digital
Multimeter
4.3 The range of resistance shall be selected for the Decade Resistance Box being
calibrated.
4.4 Resistance shall be gradually increased to each cardinal point upto full scale of test
instrument *(IUT).
4.5 The readings of Decade Resistance Box shall be recorded in measured value column
against the standard readings in set value column of Calibration Data Sheet (Form:
call/013).
4.6 The relevant section of Calibration Data Sheet (Form: call/013) shall be marked.
4.7 After completion of calibration the power of the standard and Decade Resistance Box
under test shall be switched off and secure all equipment.
Page 3 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/004 Rev. 00 May 01, 2015
Decade Resistance Box
4.8 Calibration Certificate (Form: call/002) shall be issued for calibrated Decade Resistance
Box as per Calibration Data Sheet Record (Form: call/013).
4.9 A Calibration Sticker (Form: call/003) containing / having following information shall be
affixed on the calibrated Decade Resistance Box.
a) Item code
b) Calibrated on
c) Next calibration date
All the latest formats are documented in Master List of Document / Record and placed in CQF
file for reference.
All the latest Standard Calibration Procedures (SCP) are controlled and issued to concerned
Calibration Engineers.
Page 4 of 4