CERTIFICATE OF CALIBRATION

CERTIFICATE No.

SACPR/00224/23

Customer : Renata Limited, Rajendrapur Plant,


Noyapara, Bhawal Mirzapur, Rajendrapur, Gazipur, Bangladesh, 1700
Instrument : Pressure Gauge
Manufacturer : N/A
Serial No. : N/A
Tag No. : PG-01
Model No. : N/A
Range : 0~16 bar
Resolution : 0.2 bar
Location : Liquid Transferring Device, Eq-WH-039
As Received Physical Condition : Good
As Return Physical Condition : Good
Date of Issue : 07-02-2023
Date Calibrated : 07-02-2023
Recommended Due Date : 07-02-2024

This instrument has been Calibrated at Customer Place.


The Calibration was performed at an ambient temperature of 20.3°C, and a relative humidity of 54.6% RH

Calibration Method
Procedure number : IESL CP-02-05, Procedure name : IESL CP-02: Analog/Digital Pressure Gauge, pressure Switch & Transmitter both at the
Laboratory & Onsite, International standard used :DKD-R6-1 Calibration of Pressure Gauge, Deutsche Norm DIN 16005 - February 1987, OIML
recommendation -September 1987, JCGM - 100:2008, EA-04/02

Reference Equipments Used


Equipment Name Serial No Traceability Certificate No Due On
1 TIS Digital Pressure Gauge 15-041030 NPL, India through FL/C/M/22092022-C001 24-09-2023
Fare Labs Pvt. Ltd.

Calibration Results / Remarks (if any)


1 Please refer to the attached Calibration certificate
2 The user should decide on the usability of this document for its intended use

Calibrated By Reviewed By Approved Signatory


Musa Kalimulla Md.Aslam Uddin Molla Md.Monwar Hosen
Emp ID : ASC 047 Emp ID : ASC 025 Emp ID : ASC 008
Date : 07-02-2023 Date : 07-02-2023 Date : 07-02-2023

The results reported herein have been performed in accordance with the Laboratories terms of accreditation under the Accreditation Board.
This report shall not be reproduced except in full, without the prior written approval of IESL. Certificate is electronically generated

Page 1 of 2
CERTIFICATE OF CALIBRATION

CERTIFICATE No.

SACPR/00224/23

Calibration Results(As Found)


Table : Calibration of Pressure

Average Acceptance
Pressure on Expanded
Sl. No. Standard Error Limit Remarks
UUC Uncertainty(±)
Pressure (±)

1 0.0 bar 0.000 bar 0.0 bar 0.04 bar OK

2 4.0 bar 3.983 bar 0.017 bar 0.04 bar OK

3 8.0 bar 8.002 bar -0.002 bar 0.04 bar 5% OK

4 12.0 bar 12.004 bar -0.004 bar 0.04 bar OK

5 16.0 bar 16.009 bar -0.009 bar 0.04 bar OK


The result has an expanded uncertainty of approximately 95% confidence level with a coverage factor k=2.00

UUC : “Unit Under Calibration”


Comments: Based on above calibration data the above equipment is declared calibrated.

Calibrated By Reviewed By Approved Signatory


Musa Kalimulla Md.Aslam Uddin Molla Md.Monwar Hosen
Emp ID : ASC 047 Emp ID : ASC 025 Emp ID : ASC 008
Date : 07-02-2023 Date : 07-02-2023 Date : 07-02-2023

The results reported herein have been performed in accordance with the Laboratories terms of accreditation under the Accreditation Board.
This report shall not be reproduced except in full, without the prior written approval of IESL. Certificate is electronically generated

Page 2 of 2
CERTIFICATE OF CALIBRATION

CERTIFICATE No.
SAC/00384/23

Customer : Renata Limited, Rajendrapur Plant,


Noyapara, Bhawal Mirzapur, Rajendrapur, Gazipur, Bangladesh, 1700
Instrument : Temperature Sensor(Gelatin Service Tank-03)
Manufacturer : Autonics
Serial No. : N/A
Tag No. : N/A
Model No. : TZ 4ST
Range : 0~150 °C
Resolution : 1°C
Location : RPGF ANNEX
As Received Physical Condition : Good
As Return Physical Condition : Good
Date of Issue : 29-12-2022
Date Calibrated : 28-12-2022
Recommended Due Date : 28-12-2023

This instrument has been Calibrated at Customer Place.


The Calibration was performed at an ambient temperature of 20.1 °C, and a relative humidity of 50.4% RH

Calibration Method
Procedure number : IESL CP-01-04, Procedure name : IESL CP-01-04: Calibration Procedure for Temperature Sensor with/without Indicator
&Temperature Transmitter with/without Indicator, International standard used :DKD R-5-1:2003, ASTM E 77 2014, ‘Traceable Temperatures’,
second Edition-Wiley

Reference Equipments Used


Equipment Name Serial No Traceability Certificate No Due On
1 Digital Thermometer (Taishio) 97519080139 NPL, India through FL/C/TH/22092022-C003 24-09-2023
Fare Labs Pvt. Ltd.
2 Dry Well Temperature Calibrator 660TH18010068 NPL, India through SAL/00502/22 05-03-2023
IESL
3 Dry Well Temperature Calibrator 660TL18010053 NPL, India through SAL/00503/22 05-03-2023
IESL

Calibration Results / Remarks (if any)


1 Please refer to the attached Calibration certificate
2 The user should decide on the usability of this document for its intended use

Calibrated By Reviewed By Approved Signatory


Musa Kalimulla Md.Aslam Uddin Molla Md.Monwar Hosen
Emp ID : ASC 047 Emp ID : ASC 025 Emp ID : ASC 008
Date : 28-12-2022 Date : 29-12-2022 Date : 29-12-2022

The results reported herein have been performed in accordance with the Laboratories terms of accreditation under the Accreditation Board.
This report shall not be reproduced except in full, without the prior written approval of IESL. Certificate is electronically generated

Page 1 of 2
CERTIFICATE OF CALIBRATION

CERTIFICATE No.
SAC/00384/23

Calibration Results(As Found)


Sl. No. Standard Found Error °C Expanded Acceptance Remarks
Temperature Temperature Uncertainty Limit
°C in UUC °C ±°C ±°C
1 0.000 0 0 0.605 2 OK

2 25.001 25 -0.001 0.605 2 OK

3 50.001 49 -1.001 0.605 2 OK

4 75.000 74 -1 0.608 2 OK

5 99.999 99 -0.999 0.609 2 OK


The result has an expanded uncertainty of approximately 95% confidence level with a coverage factor k=2.00

UUC : “Unit Under Calibration”


Comments: Based on above calibration data the above equipment is declared calibrated.

Calibrated By Reviewed By Approved Signatory


Musa Kalimulla Md.Aslam Uddin Molla Md.Monwar Hosen
Emp ID : ASC 047 Emp ID : ASC 025 Emp ID : ASC 008
Date : 28-12-2022 Date : 29-12-2022 Date : 29-12-2022

The results reported herein have been performed in accordance with the Laboratories terms of accreditation under the Accreditation Board.
This report shall not be reproduced except in full, without the prior written approval of IESL. Certificate is electronically generated

Page 2 of 2
CERTIFICATE OF CALIBRATION

CERTIFICATE No.
SAC/00383/23

Customer : Renata Limited, Rajendrapur Plant,


Noyapara, Bhawal Mirzapur, Rajendrapur, Gazipur, Bangladesh, 1700
Instrument : Temperature Sensor (Medicine Service Tank-03)
Manufacturer : HANYOUNG NUX
Serial No. : N/A
Tag No. : AHI12K027
Model No. : NX4-01
Range : 0~150 °C
Resolution : 1°C
Location : RPGX ANNEX
As Received Physical Condition : Good
As Return Physical Condition : Good
Date of Issue : 29-12-2022
Date Calibrated : 28-12-2022
Recommended Due Date : 28-12-2023

This instrument has been Calibrated at Customer Place.


The Calibration was performed at an ambient temperature of 19.2°C, and a relative humidity of 50.1% RH

Calibration Method
Procedure number : IESL CP-01-04, Procedure name : IESL CP-01-04: Calibration Procedure for Temperature Sensor with/without Indicator
&Temperature Transmitter with/without Indicator, International standard used :DKD R-5-1:2003, ASTM E 77 2014, ‘Traceable Temperatures’,
second Edition-Wiley

Reference Equipments Used


Equipment Name Serial No Traceability Certificate No Due On
1 Digital Thermometer (Taishio) 97519080139 NPL, India through FL/C/TH/22092022-C003 24-09-2023
Fare Labs Pvt. Ltd.
2 Dry Well Temperature Calibrator 660TH18010068 NPL, India through SAL/00502/22 05-03-2023
IESL
3 Dry Well Temperature Calibrator 660TL18010053 NPL, India through SAL/00503/22 05-03-2023
IESL

Calibration Results / Remarks (if any)


1 Please refer to the attached Calibration certificate
2 The user should decide on the usability of this document for its intended use

Calibrated By Reviewed By Approved Signatory


Musa Kalimulla Md.Aslam Uddin Molla Md.Monwar Hosen
Emp ID : ASC 047 Emp ID : ASC 025 Emp ID : ASC 008
Date : 28-12-2022 Date : 29-12-2022 Date : 29-12-2022

The results reported herein have been performed in accordance with the Laboratories terms of accreditation under the Accreditation Board.
This report shall not be reproduced except in full, without the prior written approval of IESL. Certificate is electronically generated

Page 1 of 2
CERTIFICATE OF CALIBRATION

CERTIFICATE No.
SAC/00383/23

Calibration Results(As Found)


Sl. No. Standard Found Error °C Expanded Acceptance Remarks
Temperature Temperature Uncertainty Limit
°C in UUC °C ±°C ±°C
1 0.000 0 0 0.605 2 OK

2 25.001 25 -0.001 0.605 2 OK

3 49.999 49 -0.999 0.605 2 OK

4 75.002 74 -1.002 0.608 2 OK

5 99.999 99 -0.999 0.607 2 OK


The result has an expanded uncertainty of approximately 95% confidence level with a coverage factor k=2.00

UUC : “Unit Under Calibration”


Comments: Based on above calibration data the above equipment is declared calibrated.

Calibrated By Reviewed By Approved Signatory


Musa Kalimulla Md.Aslam Uddin Molla Md.Monwar Hosen
Emp ID : ASC 047 Emp ID : ASC 025 Emp ID : ASC 008
Date : 28-12-2022 Date : 29-12-2022 Date : 29-12-2022

The results reported herein have been performed in accordance with the Laboratories terms of accreditation under the Accreditation Board.
This report shall not be reproduced except in full, without the prior written approval of IESL. Certificate is electronically generated