Standard Calibration Procedure

Doc. No. Call/SCP/020 Rev. 00 May 01, 2015


Temperature Sensor (Data Tracer)

STANDARD CALIBRATION PROCEDURE

TEMPERATURE SENSOR (DATA TRACER)

Prepared by

Muhammad Ajlan Signature Date : 01.05.2015

Reviewed by

Muhammad Sarwar Signature Date : 01.05.2015

Approved by

Qaswar Mahmood Signature Date : 01.05.2015

Controlled Copy No. Uncontrolled Copy

Issued to :

Page 1 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/020 Rev. 00 May 01, 2015
Temperature Sensor (Data Tracer)

This document is the property of Qstar Engineering, and is issued only for persons or entity as
written on this document. It is not to be retransmitted, printed or copied without prior written
permission of company.

Table of Contents

Sr. No. Title Page No(s).


1.0 Scope 03
2.0 Standard Equipment 03
3.0 Preliminary Operations 03
4.0 Calibration Procedure 03
5.0 Related Documents 03

Page 2 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/020 Rev. 00 May 01, 2015
Temperature Sensor (Data Tracer)

1.0 Scope

This procedure provides guidelines for calibration of Temperature Sensor (Data tracer).

2.0 Standard Equipment

2.1 Standard Process Calibrator with a valid certificate of calibration traceable to


international standard in accordance with an industrial code or a regulation shall be used
as a standard for calibration.
2.2 The reference standard shall be kept under controlled environmental conditions e.g. dust
free environment and Lab. temperature 22°C ± 3°C. Lab Humidity 60%RH ± 20%RH.

3.0 Preliminary Operations

3.1 The details of Temperature Sensor (Data tracer) received for calibration shall be entered
in the Calibration Job (Form: Call/001).
3.2 Inspection of the Temperature Sensor (Data tracer) for general condition shall be carried
out.
3.3 The Temperature Sensor (Data tracer) shall be inspected and marked () in relevant
block of equipment’s primary inspection section of Calibration Data Sheet (Form:
Call/013).

4.0 Calibration Procedure

4.1 The Temperature Sensor (Data tracer) to be calibrated shall be maintained at same
place of Standard (Process Calibrator) for 01 hours minimum.
4.2 For the measurement of temperature Sensor (Data tracer) the Process Calibrator shall
be selected in temperature measuring mode
4.3 The sensor of standard process calibrator and temperature sensor (Data tracer) shall be
placed in the temperature cell.
4.4 The Temperature Sensor (Data tracer) shall be checked at approximately 20%, 40%,
60% and 80% of full range.
4.5 The relevant section of Calibration Data Sheet (Form: Call/013) shall be marked

Page 3 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/020 Rev. 00 May 01, 2015
Temperature Sensor (Data Tracer)

4.6 The readings of Temperature Sensor (Data tracer) shall be recorded in measured value
column against the standard readings in set value column of Calibration Data Sheet
(Form: Call/013).
4.7 Calibration Certificate (Form: Call/002) shall be issued for calibrated Temperature
Sensor (Data tracer) as per Calibration Data Sheet Record (Form: Call/013).
4.8 A Calibration Sticker (Form: Call/003) containing following information shall be affixed
on the calibrated Temperature Sensor (Data tracer).
a) Item code
b) Calibrated on
c) Next calibration date

5.0 Related Documents

All the latest formats are documented in Master List of Document / Record and placed in CQF
file for reference.
All the latest Standard Calibration Procedures (SCP) are controlled and issued to concerned
Calibration Engineers.