You are on page 1of 1

BUVECO ST700-IR

Fault codes that could appear on internal LED-display


----------------------------------------------------------------------------------------------------------------------------------------
F2 = sensor negative drift                          try to rezero in clean air or nitrogen.
F4 = zero factor outside limits                     try to rezero in clean air or nitrogen, otherwise sensor
change 
F8 = span factor outside limits                    try to recalibrate with gas, check gasconcentration,
otherwise sensor change 
F16 = raw reference sensor signal to low      change sensor  
F32 = raw reference sensor signal to high    change sensor
F64 = raw active sensor signal to high         change sensor
F99 = configuration data corrupt                   change pcb
-----------------------------------------------------------------------------------------------------------------

You might also like