You are on page 1of 7

Comparison of Immersion Ultrasound,

Partial Coherence Interferometry, and


Low Coherence Reflectometry for Ocular
Biometry in Cataract Patients
Robert Montés-Micó, PhD; Francesco Carones, MD; Antonietta Buttacchio, OD;
Teresa Ferrer-Blasco, PhD; David Madrid-Costa, PhD

A
ccurate biometry estimating ocular dimensions is nec-
ABSTRACT
essary in cataract and refractive surgery. Axial length,
PURPOSE: To compare ocular biometry parameters
anterior chamber depth, and corneal power are needed
measured with immersion ultrasound, partial coherence for theoretical intraocular lens (IOL) power calculation.
interferometry, and low coherence reflectometry in cata- Current ocular biometry techniques include both ultra-
ract patients. sound and optical biometers. Applanation ultrasound is the
most widely used technique for ocular biometry.1 However,
METHODS: Measurements of axial length and anterior direct applanation may compress the cornea and result in
chamber depth were analyzed and compared using im-
mersion ultrasound, partial coherence interferometry, and
an underestimation of axial length by 0.14 to 0.47 mm com-
low coherence reflectometry. Keratometry (K), flattest pared to immersion ultrasound.2-4 Immersion ultrasound is
axis, and white-to-white measurements were compared an accurate method for axial length measurement, generally
between partial coherence interferometry and low considered superior to applanation ultrasound techniques.
coherence reflectometry. Seventy-eight cataract (LOCS Immersion ultrasound is used only by a minority of cataract
II range: 1 to 3) eyes of 45 patients aged between 42
and 90 years were evaluated. A subanalysis as a function
and refractive surgeons (12%)1 despite its reduced range of
of cataract degree was done for axial length and anterior postoperative refractive errors when compared to applana-
chamber depth between techniques. tion ultrasound.
Optical biometry uses laser instead of ultrasound for data
RESULTS: No statistically significant differences were acquisition. Two different types of optical laser systems based
noted for the study cohort or within each cataract on different technology for ocular biometry measurement
degree among the three techniques for axial length and
anterior chamber depth (P⬎.05, ANOVA test). Mea-
are available currently. The IOLMaster (Carl Zeiss Meditec,
surements between techniques were highly correlated Jena, Germany) uses partial coherence interferometry and the
for axial length (R=0.99) and anterior chamber depth Lenstar LS 900/Biograph (Haag-Streit AG, Koeniz, Switzerland/
(R=0.90 to 0.96) for all methods. Keratometry, flattest Alcon Laboratories Inc, Ft Worth, Texas) uses low coher-
axis, and white-to-white measurements were compa- ence reflectometry. Partial coherence interferometry has been
rable (paired t test, P⬎.1) and correlated well between
partial coherence interferometry and low coherence
shown to have the same accuracy as immersion ultrasound
reflectometry (K1 [R=0.95), K2 [R=0.97], flattest axis biometry,5-8 whereas low coherence reflectometry has yet to
[R=0.95], and white-to-white [R=0.92]). be compared to immersion ultrasound biometry to determine
possible differences. Partial coherence interferometry and
CONCLUSIONS: Immersion ultrasound, partial coherence
interferometry, and low coherence reflectometry provided
comparable ocular biometry measurements in cataractous From Optometry Research Group, University of Valencia, Valencia, Spain
eyes. [J Refract Surg. 2011;27(9):665-671.] (Montés-Micó, Ferrer-Blasco, Madrid-Costa); and Carones Ophthalmology
Center, Milan, Italy (Carones, Buttacchio).
doi:10.3928/1081597X-20110202-01
This research was supported in part by Ministerio de Ciencia e Innovación
Research Grants to Robert Montés-Micó (#SAF2008-01114-E/# and #SAF2009-
13342-E#).
The authors have no proprietary interest in the materials presented herein.
Correspondence: Robert Montés-Micó, PhD, Dept of Optics, University of
Valencia, C/ Dr Moliner 50, 46100 Burjassot, Spain. Tel: 34 96 354 4764;
Fax: 34 96 354 4715; E-mail: robert.montes@uv.es
Received: September 12, 2010; Accepted: January 18, 2011
Posted online: February 15, 2011

Journal of Refractive Surgery • Vol. 27, No. 9, 2011 665


Comparison of Three Biometers/Montés-Micó et al

low coherence reflectometry have been compared in rior chamber depth, defined as the measurement from
healthy, cataract, pseudophakic, aphakic, and silicon the corneal endothelium to the anterior lens surface.
oil-filled eyes with comparable results between tech- It also measures crystalline lens thickness and retinal
niques.9-15 thickness. The keratometry readings are calculated by
The present study measures and compares ocular analyzing the anterior corneal curvature at 32 refer-
biometry values with immersion ultrasound, partial ence points orientated in 2 circles at approximately the
coherence interferometry, and low coherence reflec- 2.30-mm and 1.65-mm optical zones. It also measures
tometry in cataractous eyes. white-to-white distance.
Immersion ultrasound was done using the Hansen
PATIENTS AND METHODS Shell immersion technique with the Ocuscan (Alcon
Seventy-eight cataractous eyes of 45 patients (20 Laboratories Inc) ultrasound biometer.
men and 25 women) with a mean age of 65.3⫾12.9
years (range: 42 to 90 years) were included in this non- DATA ANALYSIS
randomized, prospective evaluation study. Cataracts Data analysis was performed using SPSS for Win-
ranged from grade 1 to 3 according to the Lens Opaci- dows version 12.0 (SPSS Inc, Chicago, Illinois). Nor-
ties Classification System (LOCS) II (mean: 1.8⫾0.8). mality was checked by the Shapiro-Wilk test. For axial
The study was conducted in accordance with the length and anterior chamber depth measured with par-
Declaration of Helsinki, and investigational review tial coherence interferometry, low coherence reflec-
board approval was obtained from the Carones Oph- tometry, and immersion ultrasound, analysis of vari-
thalmology Center (Milan, Italy). Patients with cor- ance (ANOVA) was used to compare the outcomes. A
neal pathology, previous refractive surgery, or abnor- subanalysis as a function of cataract degree was also
malities in the anterior segment were excluded from performed. Paired t tests were applied to compare
the study. Patients were fully informed about the pur- keratometry, flattest axis, and white-to-white measure-
pose of the study and provided informed consent. ments between partial coherence interferometry and
low coherence reflectometry. To determine the rela-
BIOMETRY TECHNIQUES tionship between measurements of the two devices,
Ocular biometry was performed using immersion the Pearson correlation coefficient (R) was used. The
ultrasound, partial coherence interferometry, and low distribution of differences between each device was
coherence reflectometry in all patients. Partial coher- plotted by means of Bland-Altman plots. This method
ence interferometry and low coherence reflectometry also computed 95% limits of agreement as ⫾2 stan-
were performed first using block randomization, after dard deviations of the differences between two mea-
which measurement by immersion ultrasound was car- surement techniques. Differences were considered sta-
ried out. Immersion ultrasound was the last technique tistically significant at P⬍.05.
performed to avoid any possible discomfort to the patient
due to the use of the shell. Five consecutive measure- RESULTS
ments were taken per eye with each instrument. Three Table 1 shows the axial length and anterior cham-
examiners (A.B., F.C., T.F.B.), one per technique, par- ber depth measurements with the three techniques as
ticipated and were masked to the values found using well as the difference in measurements between the
the other two instruments. units for all eyes and as a function of cataract degree.
The IOLMaster optical biometer uses partial coher- No statistically significant differences were noted for
ence interferometry with a 780-mm laser diode infra- the entire cohort and cataract degree among the three
red light to measure axial length. The anterior chamber techniques for axial length and anterior chamber depth
depth is measured through a lateral slit-illumination (P⬎.05, ANOVA test). Measurements between tech-
and is defined as the measurement between the corneal niques were highly correlated for axial length (partial
epithelium and anterior lens surface. The keratom- coherence interferometry vs low coherence reflectom-
etry readings are calculated by analyzing the anterior etry, R=0.99; partial coherence interferometry vs im-
corneal curvature at six reference points oriented in a mersion ultrasound, R=0.99; and low coherence reflec-
hexagonal pattern at approximately the 2.3-mm optical tometry vs immersion ultrasound, R=0.99) and anterior
zone. White-to-white distance is also measured. chamber depth (partial coherence interferometry vs low
The Lenstar LS 900/Biograph optical biometer is coherence reflectometry, R=0.90; partial coherence in-
based on low coherence reflectometry, with an 820-mm terferometry vs immersion ultrasound, R=0.96; and
superluminescent diode. In addition to axial length, low coherence reflectometry vs immersion ultrasound,
the unit measures central corneal thickness and ante- R=0.92). Figure 1 shows the distribution of differences

666 Copyright © SLACK Incorporated


Comparison of Three Biometers/Montés-Micó et al

TABLE 1
Comparison of Axial Length and Anterior Chamber Depth Among Instruments in All
Eyes and as a Function of Cataract Degree (LOCS II Classification)
Mean⫾Standard Deviation (Range) (mm)
Difference Difference PCI- Difference LCR-
Parameter PCI LCR Immersion US PCI-LCR Immersion US Immersion US
All eyes AL 24.70⫾2.23 24.70⫾2.26 24.59⫾2.23 ⫺0.004⫾0.14 0.11⫾0.14 0.11⫾0.13
(N=78) (21.00 to 31.15) (20.95 to 31.28) (20.90 to 31.31) (⫺0.43 to 0.42) (⫺0.33 to 0.95) (⫺0.47 to 0.40)
ACD 3.12⫾0.42 3.19⫾0.40 3.30⫾0.39 ⫺0.07⫾0.28 ⫺0.18⫾0.28 ⫺0.11⫾0.15
(2.15 to 3.95) (2.31 to 4.15) (2.44 to 4.11) (⫺1.38 to 0.38) (⫺1.37 to 0.37) (⫺0.49 to 0.48)
LOCS II 1 AL 25.44⫾2.57 25.47⫾2.60 25.33⫾2.58 ⫺0.04⫾0.03 0.10⫾0.13 0.14⫾0.13
(n=32) (21.00 to 31.15) (20.95 to 31.28) (20.90 to 31.31) (⫺0.13 to 0.06) (⫺0.33 to 0.31) (⫺0.31 to 0.32)
ACD 3.23⫾0.40 3.33⫾0.36 3.40⫾0.33 ⫺0.10⫾0.35 ⫺0.17⫾0.34 ⫺0.06⫾0.18
(2.56 to 3.84) (2.36 to 3.97) (2.56 to 3.84) (⫺1.37 to 0.32) (⫺1.37 to 0.37) (⫺0.24 to 0.47)
LOCS II 2 AL 24.14⫾1.66 24.11⫾1.73 24.0⫾1.67 0.02⫾0.21 0.12⫾0.18 0.09⫾0.14
(n=27) (21.25 to 27.33) (21.15 to 27.37) (21.20 to 27.20) (⫺0.11 to 0.35) (⫺0.10 to 0.95) (⫺0.47 to 0.29)
ACD 3.08⫾0.42 3.13⫾0.38 3.26⫾0.41 ⫺0.10⫾0.24 ⫺0.17⫾0.25 ⫺0.12⫾0.07
(2.15 to 3.82) (2.31 to 3.87) (2.44 to 4.03) (⫺0.60 to 0.38) (⫺0.71 to 0.32) (⫺0.24 to 0.06)
LOCS II 3 AL 24.24⫾2.06 24.24⫾2.02 24.15⫾2.02 0.005⫾0.07 0.09⫾0.08 0.08⫾0.11
(n=19) (21.63 to 29.10) (21.62 to 28.87) (21.67 to 28.97) (⫺0.07 to 0.32) (⫺0.05 to 0.33) (⫺0.10 to 0.40)
ACD 2.98⫾0.39 3.03⫾0.40 3.18⫾0.44 ⫺0.05⫾0.15 ⫺0.20±0.19 ⫺0.15±0.17
(2.32 to 3.95) (2.38 to 4.15) (2.50 to 4.11) (⫺0.27 to 0.32) (⫺0.66 to 0.12) (⫺0.48 to 0.23)
PCI = partial coherence interferometry, LCR = low coherence reflectometry, US = ultrasound, AL = axial length, ACD = anterior chamber depth (from epithelium)

by means of six Bland-Altman plots of axial length and DISCUSSION


anterior chamber depth measured by partial coherence Previous reports have shown that both devices (par-
interferometry, low coherence reflectometry, and im- tial coherence interferometry [IOLMaster] and low co-
mersion ultrasound for all eyes analyzed. Results show herence reflectometry [Lenstar LS 900/Biograph], which
that the observed distribution of differences in axial are based on different technologies, correlate well for
length and anterior chamber depth measurements for axial length, anterior chamber depth, and keratometry,
each comparison between techniques was unrelated to providing valid measurements to be used for preopera-
the mean axial length and anterior chamber depth values, tive examination of cataract patients.9-15 However, some
respectively. discrepancies exist among studies although they are not
Table 2 shows the keratometry (K1 [flat] and K2 clinically significant. The purpose of the present study
[steep] meridians), flattest axis, and white-to-white was to analyze and compare both devices to one another
measurements with partial coherence interferom- as well as to immersion ultrasound for axial length and
etry and low coherence reflectometry for all eyes. All anterior chamber depth measurements. A subanalysis
values were comparable between techniques show- as a function of cataract degree was done, considering
ing no statistically significant differences among that cataract opacity might affect the outcomes.
them (paired t test, P⬎.1). Measurements between
techniques were highly correlated for both K1 AXIAL LENGTH AND ANTERIOR CHAMBER DEPTH
(R=0.95), K2 (R=0.97), flattest axis (R=0.95), and MEASUREMENTS
white-to-white (R=0.92). Figures 2 and 3 show the Comparable and highly correlated (R⬎0.9 for all
Bland-Altman plots of white-to-white and keratom- comparisons) measurements of axial length and ante-
etry, respectively, measured by partial coherence rior chamber depth were obtained with the three tech-
interferometry and low coherence reflectometry for niques analyzed. No statistically significant differences
all eyes analyzed. Results show that the observed were noted among the values found for all eyes or as a
differences in white-to-white and keratometry (K1 function of cataract degree (P⬎.05). No previous stud-
and K2) measurements between techniques were un- ies comparing the three techniques in the same eyes
related to the mean white-to-white and keratometry have been published to date and hence direct compari-
values, respectively. son with the present study is not possible. However,

Journal of Refractive Surgery • Vol. 27, No. 9, 2011 667


Comparison of Three Biometers/Montés-Micó et al

0.5 2.0 2.0


0.4 1.5 1.5
0.3
1.0 1.0
0.2
0.1 0.5 0.5

0 0 0
-0.1 -0.5 -0.5
-0.2
-1.0 -1.0
-0.3
-1.5 -1.5
-0.4
-0.5 -2.0 -2.0
20 22 24 26 28 30 32 20 22 24 26 28 30 32 20 22 24 26 28 30 32

1.0 1.0 1.0


0.8 0.8 0.8
0.6 0.6 0.6
0.4 0.4 0.4
0.2 0.2 0.2
0 0 0
-0.2 -0.2 -0.2
-0.4 -0.4 -0.4
-0.6 -0.6 -0.6
-0.8 -0.8 -0.8
-1.0 -1.0 -1.0
2.0 2.5 3.0 3.5 4.0 2.0 2.5 3.0 3.5 4.0 2.0 2.5 3.0 3.5 4.0

Figure 1. Bland-Altman plots showing the distribution of differences as a function of mean values for top) axial length (AL) and bottom) anterior cham-
ber depth (ACD) among the three techniques (partial coherence interferometry [PCI], low coherence reflectometry [LCR], and immersion ultrasound
[IU]). The 95% limits of agreement are represented by dotted lines and the bold horizontal line refers to the mean difference between techniques.

TABLE 2
Comparison of Keratometry, Flattest Axis, and White-to-White Between
Partial Coherence Interferometry and Low Coherence Reflectometry
Mean⫾Standard Deviation (Range)
Parameter PCI LCR Difference PCI-LCR P Value
Keratometry K1 (D) 42.32⫾1.71 42.24⫾1.68 0.08⫾1.41 .310
(36.37 to 46.49) (35.42 to 45.68) (⫺0.01 to 1.65)
Keratometry K2 (D) 43.33⫾1.81 43.12⫾1.72 0.22⫾1.53 .108
(36.72 to 48.01) (36.01 to 46.94) (⫺0.26 to 2.78)
Flattest axis (°) 93.33⫾59.98 93.27⫾60.48 0.77⫾61.55 .495
(1 to 178) (0 to 179) (⫺170 to 169)
White-to-white (mm) 12.08⫾0.47 12.11⫾0.39 ⫺0.04⫾0.50 .269
(11.08 to 12.70) (11.01 to 12.86) (⫺1.16 to 1.40)
PCI = partial coherence interferometry, LCR = low coherence reflectometry, K1 = flat meridian, K2 = steep meridian

reports comparing the IOLMaster and Lenstar are of However, in 112 cataract patients, Buckhurst et al11
interest. showed significant differences between devices for
Rohrer et al9 compared the IOLMaster with the Len- axial length (longer for low coherence reflectometry by
star LS 900 in 144 eyes (83 cataractous, 28 pseudo- 0.01⫾0.15 mm, P⬍.001) and anterior chamber depth
phakic, 5 aphakic, 14 silicon oil-filled, and 14 healthy (longer for low coherence reflectometry by 0.10⫾0.40 mm,
phakic eyes). In all eyes, the results showed a high cor- P⬍.001). Regarding the anterior chamber depth dis-
relation for axial length and anterior chamber depth crepancy in the study by Buckhurst et al, the IOLMas-
measurements between both instruments (R coefficient ter does not use coherent interferometry to measure
close to 1). Holzer et al10 compared 200 healthy eyes anterior chamber depth, using instead image analy-
using both instruments, showing that axial length and sis of the distance between the anterior surface of the
anterior chamber depth were similar and comparable. cornea and crystalline lens when illuminated by an

668 Copyright © SLACK Incorporated


Comparison of Three Biometers/Montés-Micó et al

1.5 2.0

1.5
1.0
1.0
0.5
0.5

0 0

-0.5
-0.5
-1.0
-1.0
-1.5

-1.5 2.0
11 11.5 12 12.5 13 38 40 42 44 46 48

Figure 2. Bland-Altman plot showing the distribution of differences as Figure 3. Bland-Altman plot showing differences against mean values
a function of mean values for white-to-white measurement between for keratometry (K1 and K2) between the partial coherence interfer-
the partial coherence interferometry and low coherence reflectometry ometry and low coherence reflectometry techniques. The 95% limits of
techniques. The 95% limits of agreement are represented by dotted lines agreement are not represented to avoid confusion. The bold and dotted
and the bold horizontal line represents the mean difference between horizontal lines represent the mean difference between techniques for
techniques. Results show that the observed differences in white-to-white K1 and K2, respectively. Results show that the observed differences in
measurements between techniques were unrelated to the mean white- keratometry measurements between techniques were unrelated to the
to-white. mean keratometry for both meridians.

optical section with a 0.7-mm wide slit beam of light found in the present study, discrepancies may exist in
through the anterior segment of the eye at an angle of axial length and anterior chamber depth measurements.
38° to the visual axis.16 The Lenstar detects the ante- Hoffer et al14 suggested that the reason for axial length
rior and posterior corneal and anterior crystalline lens measurement discrepancy is unclear. We believe it is
peaks in the low coherence reflectometry waveform to likely to be related to the fact that measurements from
measure the anterior chamber depth and corneal thick- the partial coherence interferometry unit are recali-
ness, which were combined for comparison with the brated to improve agreement with immersion ultra-
IOLMaster result. sound. The partial coherence interferometry unit mea-
Rabsilber et al12 compared the IOLMaster and Lenstar sures the distance from the anterior corneal vertex to
in 100 cataractous eyes showing comparable outcomes the retinal pigment epithelium, whereas immersion
for axial length (P⬎.1) but significantly higher mea- ultrasound measures the distance up to the internal
surements for anterior chamber depth with the Lenstar limiting membrane.15 Without recalibration, partial
(0.05⫾0.11 mm, P⬍.1). In 145 cataractous eyes, Liampa coherence interferometry would result in higher mea-
et al13 showed a small, systematic statistically, although surements than immersion ultrasound. Similarly, it
not clinically, significant difference in axial length can be speculated that longer mean axial length mea-
(0.03 mm, P⬍.0001) and anterior chamber depth (0.2 mm, sured by low coherence reflectometry depends on a
P⬍.0001) between the devices. Hoffer et al14 evaluated different algorithm used to closely reflect ultrasound
50 cataract and 50 clear lens eyes. They found a good values. Despite recalibration, the clinical significance
correlation between axial length and anterior chamber of 0.01,11 0.02,14 and 0.0313 mm axial length variation
depth in both groups. However, the Lenstar measured between units reported in other IOL power calculation
a slightly longer axial length (mean difference 0.026 mm studies would be a ⬍0.10-diopter difference.17 Regard-
and 0.023 mm for cataract and clear lens eyes, respec- ing anterior chamber depth, as previously discussed,
tively; P⬍.0001) and a deeper anterior chamber depth differences found in some studies might be attributed
(0.128 mm and 0.146 mm for cataract and clear lens to the different technique for measurement used by
eyes, respectively; P⬍.0001). Cruysberg et al15 ana- each instrument. Partial coherence interferometry
lyzed 78 healthy eyes showing that the correlation devices measure anterior chamber depth through a
between these two devices regarding the axial length lateral slit illumination,18 whereas low coherence re-
was very high (R=1.000), yet a significant difference flectometry devices measure anterior chamber depth
was found (P⬍.001). Anterior chamber depth measure- with optical biometry.
ments of the IOLMaster were significantly lower than In addition, differences as a function of cataract
the Lenstar (P=.011). degree were not found. The presence of dense media
Although no differences between partial coherence seems to be a similar problem for both devices. Buck-
interferometry and low coherence reflectometry were hurst et al11 reported that axial length measurement

Journal of Refractive Surgery • Vol. 27, No. 9, 2011 669


Comparison of Three Biometers/Montés-Micó et al

was not possible in 9% to 10% of patients because of in our opinion, is not clinically significant. Caution
this factor. must be taken when using a dioptric representation
of corneal curvature, as differences in the refractive in-
KERATOMETRY, FLATTEST AXIS, AND WHITE-TO-WHITE dex attributed to the cornea between the instruments
MEASUREMENTS (n=1.3375 [IOLMaster] and n=1.332 [Lenstar]) would
The results reported herein revealed no significant result in a clinically significant difference in average
differences between partial coherence interferometry curvature.
and low coherence reflectometry for keratometry, flat- Partial coherence interferometry, low coherence
test axis, or white-to-white measurements. Both units reflectometry, and immersion ultrasound provided
correlated well, achieving R values close to 1. Bland- comparable ocular biometry in cataract patients. Non-
Altman plots in Figures 2 and 3 show that the ob- contact optical biometry using partial coherence in-
served differences in white-to-white and keratometry terferometry or low coherence reflectometry can be
measurements between techniques were unrelated to used in clinical practice to measure those parameters
the actual white-to-white and keratometry values mea- needed for accurate IOL power calculation in cataract
sured, respectively. and refractive surgery. Future studies should include a
Previous studies found comparable outcomes be- large sample of eyes, considering short and long eyes,
tween units for keratometry,9-11 flattest axis,9 and white- to analyze possible differences.
to-white11,15 measurements. However, others reported
small but statistically significant differences between AUTHOR CONTRIBUTIONS
the devices when analyzing keratometry. Rabsilber et Study concept and design (R.M.M., F.C., A.B., D.M.C.); data col-
al12 and Liampa et al13 showed mean differences be- lection (F.C., A.B., T.F.B., D.M.C.); analysis and interpretation of
tween R2 and (R1⫹R2)/2 corneal radii of ⭐0.02 mm data (R.M.M., T.F.B.); drafting of the manuscript (R.M.M., F.C., A.B.,
(P⬍.01) and 0.01 mm (P⬍.0001), respectively. Hoffer et T.F.B., D.M.C.); critical revision of the manuscript (R.M.M.)
al14 reported a flatter keratometry with low coherence
reflectometry compared to partial coherence interfer- REFERENCES
ometry (⫺0.107 D and ⫺0.121 D for cataract and clear 1. Leaming DV. Practice styles and preferences of ASCRS mem-
bers—2003 survey. J Cataract Refract Surg. 2004;30(4):892-900.
lens eyes, respectively; P⬍.0001). Cruysberg et al15 re-
2. Schelenz J, Kammann J. Comparison of contact and immersion
ported larger keratometry readings for low coherence techniques for axial length measurement and implant power
reflectometry versus partial coherence interferometry calculation. J Cataract Refract Surg. 1989;15(4):425-428.
(K1: 7.80 vs 7.79 mm, P=.034; K2: 7.65 vs 7.63 mm, 3. Olsen T, Nielsen PJ. Immersion versus contact technique in the
P⬍.001). For keratometry, both units acquire corneal measurement of axial length by ultrasound. Acta Ophthalmol.
1989;67(1):101-102.
radius measurements in the flat and steep meridian by
analyzing a pattern of light-emitting diodes (LEDs) im- 4. Giers U, Epple C. Comparison of A-scan device accuracy.
J Cataract Refract Surg. 1990;16(2):235-242.
aged by the corneal front surface. With the Lenstar, the
5. Haigis W, Lege B, Miller N, Schneider B. Comparison of immer-
corneal curvature value is calculated by 32 projected sion ultrasound biometry and partial coherence interferometry
light reflections, which are arranged on 2 circles with for intraocular lens calculation according to Haigis. Graefes
16 measuring points each. It has been suggested that Arch Clin Exp Ophthalmol. 2000;238(9):765-773.
compared to an arrangement with fewer LEDs, the 6. Haigis W. Optical coherence biometry. Dev Ophthalmol.
2002;34:119-130.
evaluation of a larger number of LEDs has the poten-
tial to improve the measurement reproducibility and 7. Packer M, Fine IH, Hoffman RS, Coffman PG, Brown LK. Immer-
sion A-scan compared with partial coherence interferometry:
accuracy because of the averaging over a larger number outcomes analysis. J Cataract Refract Surg. 2002;28(2):239-242.
of measurement points.9 The outer circle consisting of 8. Narváez J, Cherwek DH, Stulting RD, Waldron R, Zimmerman
16 LEDs is projected onto the cornea in a ring 2.3 mm GJ, Wessels IF, Waring GO III. Comparing immersion ultrasound
in diameter corresponding to the typical measurement with partial coherence interferometry for intraocular lens power
calculation. Ophthalmic Surg Lasers Imaging. 2008;39(1):30-34.
zone of commercially available keratometers. The in-
9. Rohrer K, Frueh BE, Wälti R, Clemetson IA, Tappeiner C,
ner circle consisting of 16 LEDs is projected onto the Goldblum D. Comparison and evaluation of ocular biometry
cornea in a ring 1.65 mm in diameter. Rohrer et al9 sug- using a new noncontact optical low coherence reflectometer.
gested that the relatively small diameter of the inner Ophthalmology. 2009;116(11):2087-2092.
LEDs may be beneficial for keratometry of corneas that 10. Holzer MP, Mamusa M, Auffarth GU. Accuracy of a new partial
are centrally altered after excimer laser correction. coherence interferometry analyser for biometric measurements.
Br J Ophthalmol. 2009;93(6):807-810.
Discrepancies among literature studies are found for
11. Buckhurst PJ, Wolffsohn JS, Shah S, Naroo SA, Davies LN,
corneal radius estimation.9-12,15 However, keratometry Berrow EJ. A new optical low coherence reflectometry de-
variations reported were approximately 0.12 D, which

670 Copyright © SLACK Incorporated


Comparison of Three Biometers/Montés-Micó et al

vice for ocular biometry in cataract patients. Br J Ophthalmol. bander J, Nuijts RM. Evaluation of the Lenstar LS 900 non-con-
2009;93(7):949-953. tact biometer. Br J Ophthalmol. 2010;94(1):106-110.
12. Rabsilber TM, Jepsen C, Auffarth GU, Holzer MP. Intraocular 16. Santodomingo-Rubido J, Mallen EA, Gilmartin B, Wolffsohn
lens power calculation: clinical comparison of 2 optical biometry JS. A new non-contact optical device for ocular biometry.
devices. J Cataract Refract Surg. 2010;36(2):230-234. Br J Ophthalmol. 2002;86(4):458-462.
13. Liampa Z, Kynigopoulos M, Pallas G, Gerding H. Comparison of 17. Hill W, Angeles R, Otani T. Evaluation of a new IOLMaster
two partial coherence interferometry devices for ocular biometry. algorithm to measure axial length. J Cataract Refract Surg.
Klin Monbl Augenheilkd. 2010;227(4):285-288. 2008;34(6):920-924.
14. Hoffer KJ, Shammas HJ, Savini G. Comparison of 2 laser in- 18. Lara F, Fernández-Sánchez V, López-Gil N, Cerviño A, Montés-
struments for measuring axial length. J Cataract Refract Surg. Micó R. Comparison of partial coherence interferometry and
2010;36(4):644-648. ultrasound for anterior segment biometry. J Cataract Refract Surg.
2009;35(2):324-329.
15. Cruysberg LP, Doors M, Verbakel F, Berendschot TT, De Bra-

Journal of Refractive Surgery • Vol. 27, No. 9, 2011 671

You might also like