Professional Documents
Culture Documents
CE and CEC
1 Introduction using fluorescently labeled primers and a multiplex ampli-
fication with 3±16 different sets of primers. This process
In the past few years there has been an expansion in the conserves sample and reduces analysis time. The loca-
application of capillary gel electrophoresis (CE) to DNA tion at which each primer anneals to the template DNA is
analysis. Methods have been developed and validated for selected to ensure that there is no overlap between differ-
DNA quantitation [1, 2], genotyping [3, 4], mutation analy- ent coamplified loci. As there is a linear relationship be-
sis [5, 6], and sequencing [7]. The reliability and stability tween DNA size and migration time from approximately
of the technique has progressed to the point that the use 75±500 bases in size [12], up to four different STR loci
of many of these applications is becoming routine in can be analyzed with a single dye label by simply adjust-
assays for product analysis and quality control. CE sys- ing the location at which the primers anneal to the sam-
tems using multicapillary, multichannel fluorescence ple. Additional loci can be analyzed simply by changing
detection are playing an increasing role in sequencing the the dye label.
human genome due to their high speed and easy automa-
tion [8]. CE is also playing an increasingly important role In DNA typing, identity is determined by calculating the
in forensic human identification [9±11]. In this application, probability that a given individual will have a specific set
CE using entangled polymer buffers is used to analyze of alleles at an STR locus. In the United States, a set of
sets of highly polymorphic microsatellite DNA known as 13 STR loci have been selected for use in a database
short tandem repeats (STRs). The STRs used in human known as CODIS (combined DNA identification system)
identity testing consist of four base motifs such as AATG [13]. Convicted sex offenders and certain other classes of
that repeat a certain number of times at particular loca- felons are compelled to submit samples for DNA analysis,
tions on a given individual©s DNA. The STRs are targeted the results of which are stored in this database. By com-
bining the frequencies of all thirteen loci, extremely high
probabilities of identity are reached, ensuring that repeat
Correspondence: Dr. Bruce McCord, Department of Chemistry
offenders can be quickly detained through comparison of
and Biochemistry, 136 Clippinger Laboratories, Athens,
OH 45701, USA crime scene samples with results in the CODIS database.
E-mail: mccord@helios.phy.ohiou.edu
Fax: +740-593-0148 In our research, we are interested in the application of the
multiplex STR systems as a mechanism to explore elec-
Abbreviation: HEC, hydroxyethylcellulose trophoretic principles. The repetitive nature and regular
spacing of the STR systems provides a useful tool for the In order to investigate the plausibility of reducing the tem-
characterization of the effect of different parameters on perature and concentration of urea in these assays, we
the separation. Another issue in STR analysis is the pres- have characterized the effect of temperature and urea
ence of variant alleles that contain one and two base concentration on the analysis of a particular multiplexed
deletions from the four base repeat motif. To detect these set of 9 STRs, the Profiler+ system. These 9 STRs are an
variants, methods must be developed with optimal resolu- important subset of the 13 STRs recommended for use in
tion and precision. In addition, the judicial aspects of the the CODIS database of convicted sexual offenders and
forensic application require a high degree of confidence in other felons. The analyses in this study are performed
the reliability of the results. Thus, the development of sys- using the PE/ABI 310, the CE system utilized in the vast
tems for DNA typing by CE represents a unique challenge majority of DNA profiling assays.
for the analyst.
To perform this work, we have utilized an experimental
sieving buffer consisting of 3% hydroxyethylcellulose
To meet these requirements, analytical systems have
(HEC) as well as a commercial buffer POP4 consisting of
been developed with multiple controls to maintain preci-
4% poly(dimethylacrylamide), at varying temperatures
sion [10]. For each set of analyses, both internal and
and/or concentrations of urea. We have also examined
external standards are utilized. To each amplified sample
the application of elevated pH as an alternative to high
an internal standard is added which consists of a mixture
temperatures and urea in an effort to minimize DNA sec-
of DNA fragments labeled with a red (ROX) fluorescent
ondary structure. The elevated pH work was performed
dye. The internal standard is then used to estimate the
using fluorocarbon-coated capillaries. Our results illus-
size of each allele in the amplified samples. An external
trate the effect of urea, pH and temperature on the preci-
standard is also part of the analytical protocol. It consists
sion and accuracy of DNA genotyping.
of a mixture of the most common alleles present in each
STR locus in the multiplex combined with the ROX-
2 Materials and methods
labeled internal standard. To perform a determination, the
size of an unknown allele is determined with reference to 2.1 Materials
the internal standard and the identity of the allele is deter-
mined by comparing the calculated size of the unknown Fluorescently labeled allelic ladders and GS500 ROX
with the known allele sizes determined in the external standards were acquired from Perkin Elmer (Foster City,
standard. CA, USA). Trizma base, boric acid, CAPS, EDTA, HEC
(#30,863-3; Aldrich Chemical, Milwaukee, WI, USA), and
Amberlite MB-150 Resins were obtained by Sigma Chem-
The multiple comparisons between internal and external
ical (St. Louis, MO, USA). Urea was acquired from Spec-
standards are necessary in order to maintain analytical
trum (Gardena, CA, USA). High-purity formamide was ob-
precision, however, additional measures are also
tained from Amresco (Solon, OH, USA).
required. Secondary structure differences can occur be-
tween the amplified sample and the internal standard and
will result in a loss of precision. These effects are caused 2.2 Sample preparation
by conformational and sequence differences and will Allelic ladder samples (AmpFISTR Profiler+; Perkin
result in poor sizing accuracy. To minimize DNA secon- Elmer) were prepared by adding 11 mL of formamide to
dary structure effects, analyses are performed using ele- 0.5 mL of allelic ladder and 0.5 mL of the ROX standard.
vated temperatures (60oC) and high concentrations of To denature the samples, the sample tubes were placed
urea (7 M). in a PE/ABI 9700 Thermal cycler, heated to 95oC for
2 min, and snap-cooled to 4oC in an ice bath. Blood stains
The use of high temperature and urea in these systems on FTA paper were extracted according to the manufac-
necessitates careful control of instrumental parameters. turer©s protocols (part #10786-010; Gibco BRL, Grand
High urea concentrations limit the shelf life of the buffers Island, NY, USA), and amplified using the protocols de-
and can result in sublimation and crystal formation. The scribed in the AmpFISTR Profiler+ Users Manual (part
elevated temperatures required can be difficult to main- #4303812; Perkin-Elmer).
tain in conditions where high humidity or rapid room tem-
perature shifts occur. Although these issues can be easily
2.3 Capillary electrophoresis
mitigated by proper validation of the analytical systems, a
less cumbersome approach would be to perform separa- An ABI Prism 310 Genetic Analyzer (Perkin Elmer) was
tions at room temperature and with denaturants other used for all experiments and calibrated using matrix dye
than urea. standards labeled with ROX, JOE, NED, and FAM fluo-
Electrophoresis 2001, 22, 755±762 Temperature and pH studies of STR systems 757
rescent dyes. J&W Scientific (Folsom, CA, USA) mSil-FC modified to permit measurement using peak width at half
capillary columns with 50 mm inner diameter were used. height:
The sieving medium was 3% w/v HEC and contained
20 mM CAPS, pH 11, and varying concentrations of urea; Resolution = 2(T1 - T2)/(W1 + W2) = (2ln2)1/2(T1 - T2)/
7, 5, 3.5, 2, and 0 M. The HEC sieving polymer solution [(W(1/2)1 + W(1/2)2)]
was prepared by adding various amount of urea to 25 mL
of CAPS buffer and shaking until dissolved, to obtain the where T is the peak retention time, W is the baseline peak
proper concentration of urea. 1.5 g of HEC was stirred width and W(1/2) is the peak width at half height [15].
into the solution, the solution was allowed to mix overnight Dividing the resolution into the DNA fragment size differ-
and approximately 0.5 g of Amberlite mixed-bed ion- ence between any two peaks produces an estimate of the
exchange resin were added and stirred for about 1 h resolution in units of bases.
using procedures similar to those described by Baskin et
al. [14]. This solution was centrifuged at 3000 rpm for
3 Results and discussion
10 min and the polymer solution removed from the ion-
exchange resin. The solution was then filtered through a In a paper by Rosenblum and co-workers in 1997 [16], a
5 mm syringe filter (Millipore, Bedford, MA, USA). So- technique was described for the analysis and mitigation of
lutions were stored in the refrigerator. Buffers containing anomalously migrating DNA fragments. In their paper,
HEC in Tris-borate/EDTA were prepared as described they identified specific fragments in a TAMRA dye labeled
previously [9], and buffers containing POP4 (Perkin Genescan 500 standard whose migration times were tem-
Elmer) were prepared using the manufacturer©s sug- perature sensitive. The effect of this temperature sensitiv-
gested protocols. Samples were electrophoretically sep- ity was to produce size estimates for these fragments that
arated using 50 mm ID capillaries with 41 cm length varied with temperature. The size estimates were ob-
(30 cm to the detector) and a field strength of 350 V/cm. tained through comparison with DNA fragments which
The capillaries were prepared by rinsing with buffer for were insensitive to temperature. Separations performed
5 min prior to running the samples. Injections were made under stronger denaturation conditions produced size
at 15 kV for 5 s. Electrophoresis was conducted at 30± estimates closer to the true (sequenced) size.
70oC and data were analyzed using GeneScan software
(Version 2.0.1; Perkin Elmer). Regression analysis on the In a previous work in our laboratory we have detected a
slope and standard deviation of the slope was performed concomitant improvement in precision with increasing
using the LINEST function present in Microsoft Excel temperature which we believe is also associated with the
(Version 5.0). The resolution between two peaks was cal- DNA secondary structure. It is possible that elevated tem-
culated using the standard chromatographic definition, peratures and strong denaturing conditions limit the varia-
Table 1. Slope of the relationship between temperature (45±70oC) and estimated size for individual alleles in the Profiler+
multiplex
STR Allele# Sizea) Run 1 SDb) Run 2 SD Run 3 SD Run 4 SD Average SDc)
Slope slope slope slope slope
D3S1358 12 111.2 ±0.12 0.01 ±0.091 0.003 ±0.098 0.006 ±0.097 0.005 ±0.10 0.01
VWA 21 194.9 ±0.083 0.009 ±0.085 0.006 ±0.058 0.005 ±0.055 0.008 ±0.07 0.02
FGA 30 264.7 ±0.144 0.005 ±0.163 0.01 ±0.129 0.006 ±0.131 0.003 ±0.14 0.02
Amel. x 103.5 ±0.147 0.01 ±0.12 0.004 ±0.118 0.007 ±0.117 0.004 ±0.13 0.01
D8S1179 19 170.4 ±0.16 0.016 ±0.193 0.014 ±0.148 0.004 ±0.15 0.006 ±0.16 0.02
D21S11 36 232.4 ±0.037 0.009 ±0.042 0.01 ±0.019 0.004 ±0.02 0.004 ±0.03 0.01
D18S51 26 341.9 ±0.186 0.006 ±0.191 0.009 ±0.167 0.011 ±0.161 0.011 ±0.18 0.01
D5S818 7 131.2 ±0.107 0.009 ±0.082 0.003 ±0.084 0.002 ±0.078 0.003 ±0.09 0.01
D13S317 8 205.0 ±0.121 0.005 ±0.127 0.003 ±0.109 0.006 ±0.108 0.005 ±0.12 0.01
D7S820 15 292.8 ±0.094 0.008 ±0.095 0.005 ±0.082 0.003 ±0.078 0.004 ±0.09 0.01
Results are given in units of bases/C
a) Estimated size at 61oC
b) Standard deviation of the calculated slope
c) Standard deviation of the average of the four estimated slopes
Slope and standard deviations are calculated using Microsoft Excel. POP4 buffer; 43 cm, 50 mm uncoated capillary;
350 V/cm; 5 s 15 kV injection. See also Fig. 1 for an example dataset.
758 T. Nock et al. Electrophoresis 2001, 22, 755±762
Table 2A. Slope of the relationship between temperature Table 2B. Slope of the relationship between temperature
(45±70oC) and estimated size for individual (45±70oC) and estimated size for individual
alleles in the Profiler+ multiplex, 0 M urea alleles in the Profiler+ multiplex, 7 M urea
STR Allele Size 61 Run 1 SD Run 2 SD Average STR Allele Size 61 Run 1 SD Run 2 SD Average
(oC) slope slope slope (oC) slope slope slope
D3S1358 12 114.10 ±0.016 0.004 ±0.014 0.003 ±0.015 D3S1358 12 115.11 ±0.024 0.003 ±0.017 0.005 ±0.021
vWA 21 196.96 ±0.032 0.004 ±0.031 0.003 ±0.032 VWA 21 197.75 ±0.040 0.006 ±0.041 0.006 ±0.039
FGA 30 254.34 0.022 0.003 0.026 0.004 0.024 FGA 30 258.97 0.039 0.005 0.042 0.006 0.040
Amel. X 106.12 ±0.032 0.003 ±0.028 0.003 ±0.030 Amel. X 108.59 ±0.063 0.009 ±0.067 0.008 ±0.062
D8S1179 19 167.17 ±0.039 0.005 ±0.033 0.004 ±0.039 D8S1179 19 171.12 ±0.066 0.012 ±0.057 0.009 ±0.060
D21S11 36 238.86 ±0.036 0.009 ±0.030 0.003 ±0.033 D21S11 36 239.25 0.027 0.004 0.017 0.008 0.021
D18S51 26 328.60 0.046 0.004 0.052 0.008 0.048 D18S51 26 333.42 0.041 0.007 0.015 0.006 0.026
D5S818 7 136.76 ±0.030 0.005 ±0.026 0.002 ±0.028 D5S818 7 137.31 ±0.021 0.004 ±0.023 0.005 ±0.022
D13S317 8 203.14 ±0.033 0.003 ±0.027 0.003 ±0.030 D13S317 8 207.80 ±0.050 0.005 ±0.030 0.007 ±0.039
D7S820 15 293.74 ±0.011 0.005 ±0.020 0.005 ±0.016 D7S820 15 294.60 0.008 0.004 0.023 0.005 0.015
20 mM CAPS buffer, pH 11; 3% HEC; 43 cm, 50 mm J&W 20 mM CAPS buffer, pH 11; 3% HEC; 43 cm, 50 mm J&W
msil FC capillary msil FC capillary
fragments and not to the size standard. The overall lower pH using urea as a denaturant. However, DNA res-
results clearly illustrate the importance of temperature olution of only 10±11 bases was illustrated and little infor-
control on the size estimates of alleles in the multiplex. mation was given on the reproducibility of the results. In
one study, separation efficiency was maximized at pH 11
In controlled studies from these and other laboratories, although complete denaturation of DNA was expected at
the precision (standard deviation) of allele size estimates pH 11.5 [24].
at 60oC ranges from 0.24±0.07 bases, with 0.17±0.12
bases being fairly typical [19±23]. An issue in forensic A goal of the present work was to determine if it was
DNA typing of STRs is the existence of variant alleles. possible to produce high-resolution separations under
These alleles contain deletions of one to two bases from alkaline conditions using fluorescently labeled DNA. Pre-
the standard 4-base motif. The level of precision, required vious studies in the literature utilized UV detection tech-
for reproducibly determining the difference between two niques and peak resolution was limited. Our expectation
alleles differing in size by one base, is 0.15 bp [21]. Thus, was that resolution would improve using higher concen-
present CE technology is just adequate to perform this trations of polymer and that the fluorescent dye linkers
task. The appearance of these variant alleles can some- would remain stable at high pH. We were also interested
times lead to an inconclusive result at one locus in a multi- in how buffers at elevated pH would affect the tempera-
plex. While the absence of one allele from a total of 13 ture response, precision and resolution of the separation.
sets of STR loci has minimal effect on the statistics of the We expected that the use of high pH buffers could obviate
analysis (in fact the rarity of these types often results in a the need for elevated temperatures and high concentra-
unique profile), improved precision would be useful in the tions of urea. Initially, we examined buffers at pH 11 and
identification of these variant alleles. Our findings on the 12 using a 50 mm fluorocarbon-coated capillary (msil FC)
temperature dependence of sizing indicate that improved filled with 3% HEC. This specially coated capillary was
control of temperature should result in increased preci- necessary because of the solubility of silica at pH above
sion. 8.5. However, the bulk of this work was performed in a
20 mM CAPS buffer at pH 11, as the fluorocarbon coating
was found to have limited stability at pH 12. The HEC
3.2 Electrophoresis at alkaline pH
polymer was chosen as a result of our previous work
An alternate method for affecting the conformational demonstrating that high precision DNA typing can be per-
effects of ssDNA is to adjust buffer pH. Recent reports formed with this system [9] and due to its demonstrated
have demonstrated the applicability of CE for DNA analy- stability at high pH. It should be noted that these results
sis at elevated pH [24, 25]. The denaturing effect of alka- should be fairly independent of the type of polymer sys-
line pH is believed to be due to the development of neg- tem as long as the polymer remains chemically stable. In
ative charges on guanine and thymine bases. In both order to characterize the effect of alkaline pH, the temper-
reports, CE with UV detection was used to analyze DNA ature dependence of the DNA size estimates using the
at pH 11 and above. The authors found the separation HEC/CAPS buffer was compared to the results previously
mechanism in alkaline buffers to be very similar to that at obtained with the POP4 system (Fig. 1).
760 T. Nock et al. Electrophoresis 2001, 22, 755±762
Table 3. Resolution and size estimates for the analysis of selected alleles in the profiler+
multiplex at three different concentrations of urea
0 M Urea 3.5 M Urea 7 M Urea
Locus Allele# Resolution Allele size Resolution Allele size Resolution Allele size
Amel. X 2.04 +/± 0.12 111.09 +/± 0.24 1.09 +/± 0.02 108.61 +/± 0.04 0.63 +/± 0.01 107.14 +/± 0.06
D21S11 26 2.71 +/± 0.09 199.10 +/± 0.19 1.94 +/± 0.06 199.24 +/± 0.03 1.26 +/± 0.01 197.99 +/± 0.05
D18S51 20 3.26 +/± 0.09 316.62 +/± 0.05 2.88 +/± 0.17 314.66 +/± 0.08 2.26 +/± 0.27 312.54 +/± 0.10
D13S317 14 2.64 +/± 0.17 237.47 +/± 0.11 2.07 +/± 0.07 230.17 +/± 0.09 1.38 +/± 0.12 229.47 +/± 0.08
VWA 20 2.91 +/± 0.16 194.87 +/± 0.10 1.82 +/± 0.13 194.53 +/± 0.05 1.26 +/± 0.07 193.88 +/± 0.06
Experimental error is measured as the SD of ten replicates. Conditions: 20 mM CAPS buff-
er, pH 11; 3% HEC; 43 cm, 50 mm J&W msil FC capillary; 5 s, 15 kV injection, 30oC. Results
are the average of ten runs.
Electrophoresis 2001, 22, 755±762 Temperature and pH studies of STR systems 761
wise, the results are roughly equivalent to those reported analysis. Reduction of the temperature dependence of
for the 3% HEC and POP4 polymers at 60oC with urea allele-size estimates will improve the long-term precision
but at far lower temperatures [26]. The reproducibility of of the analysis. Future work will explore the effects of buff-
the size estimates are also shown in Table 3 and range of er additives other than urea to improve DNA resolution at
standard deviations from 0.06 to 0.012 for a set of ten elevated pH.
analyses using 7 M urea. In a final series of experiments,
a set of 10 samples was extracted from FTA blood stain The authors would like to thank John Pedersen for con-
cards and the results were amplified and analyzed using tributing samples utilized in this study, and J&W Scientific
POP4 at 60oC, 3% HEC in 100 mM Tris-borate-EDTA for the donation of certain capillaries used in this study.
(TBE) buffer with 7 M urea at 60oC [9] and in 20 nM CAPS. Federica Crivellente, Sharon Williams, and Alice Isenberg
The results were typed using Genotyper software and are acknowledged for helpful discussions. Major parts of
shown to be in complete concordance with each other. this work were funded through grant # 1999-IJ-XC-K014
Figure 3 illustrates the analysis of an amplified sample at from the National Institute of Justice and through start-up
30oC in a pH 11 CAPS buffer at 7 M urea. funds provided by Ohio University.
[8] Huang, X. C., Quesada, M. A., Mathies, R. A., Anal. Chem. [18] Shewale, S. G., Richey, S. L., Sinha, S. K., Retarded Migra-
1992, 64, 2149±2154. tion of GS500 ROX and Sample Alleles During STR Analy-
[9] Isenberg, A. R., Allen, R. O., Keys, K. M., Smerick, J. B., sis on 310 Genetic Analyzer, Reliagene Technologies, New
Budowle, B., McCord, B. R., Electrophoresis 1998, 19, Orleans, LA 2000.
94±100. [19] Van Hoofstat, D., Deforce, D. L., Millecamps, R. E.,
[10] Lazaruk, K., Walsh, P. S., Oaks, F., Gilbert, D., Rosenblum, Brochez, R. E., Van Geldre, E. G. L., Vergauwe, A., Sijs, A.,
B. B., Menchen, S., Scheibler, D., Wenz, H. M., Holt, C., Van den Eeckhout, E. G., Electrophoresis 1998, 19,
Wallin, J., Electrophoresis 1998, 19, 86±93. 719±722.
[11] Mansfield, E. S., Robertson, J. M., Vainer, M., Isenberg, A. [20] Buel, E., Schwartz, M. B., LaFountain, M. J., J. Forens. Sci.
R., Frazier, R. R., Ferguson, K., Chow, S., Harris, D. W., 1998, 43, 164±170.
Barker,D. L., Gill, P. D., Budowle, B., McCord, B. R., [21] Wenz, H., Robertson, J. M., Menchen, S., Oaks, F., Demor-
Electrophoresis 1998, 19, 101±107. est, D. M., Scheibler, D., Rosenblum, B. B., Wike, C., Gil-
[12] Butler, J. M., McCord, B. R., Jung, J. M., Allen, R. O., bert, D. A., Efcavitch, J. W., Genome Res. 1998, 8, 69±80.
BioTechniques 1994, 17, 1062±1070. [22] Isenberg, A. R., Allen, R. O., Keys, K. M., Smerick, J. B.,
[13] Niezgoda, S. J., Brown, B., The FBI Laboratory©s Combined Budowle, B., McCord, B. R., Electrophoresis 1998, 19,
DNA Index System Program; Promega, Madison, WI 1999. 94±100.
[14] Bashkin, J., Marsh, M., Barker, D., Johnston, R., Appl. [23] Vainer, M., Enad, S., Dolnik, V., Xu, D., Bashkin, J., Marsh,
Theor. Electrophor. 1996, 6, 23±28. M., Tu, O., Harris, D. W., Barker, D. L., Mansfield, E. S.,
Genomics 1997, 41, 1±9.
[15] Luckey, J. A., Norris, T. B., Smith, L. M., J. Phys. Chem.
1993, 97, 3067±3075. [24] Liu, Y., Kuhr, W. G., Anal. Chem. 1999, 71, 1668±1673.
[16] Rosenblum, B. B., Oaks, F., Menchen, S., Johnson, B., [25] Mala, Z., Kleparnik, K., BocÆek, P., J. Chromatogr. A 1999,
Nucleic Acids Res. 1997, 25, 3925±3929. 853, 371±379.
[17] Isenberg, A. R., Allen, R. O., Keys, K. M., Smerick, J. B., [26] Isenberg, A., PhD Thesis, University of Virginia, Charlottes-
Budowle, B., McCord, B. R., Electrophoresis 1998, 19, ville, VA 1998.
94±100.