Professional Documents
Culture Documents
MIRD PAMPHLET
No. 16 •
Siegel et al. 37S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
The document has been prepared in a format to address two The time-dependent factor A incorporates characteristics of
audiences: both uptake and retention of activity in the regions of interest
and includes consideration of the physical half-life of the
1. Individuals with a primary interest in designing clinical radionuclide and the biologic half-time of the radiopharmaceu
trials who are not experts in dosimetry; and tical. The time-independent factor S includes consideration of
2. Individuals with extensive experience with dosimetry the types and energies of the radiations emitted, geometrical
based protocols and calculational methodologies. aspects such as the size and shape of the source and target
regions and the distance between them and the composition of
For the first group, the general concepts involved in data the absorbing and intervening media.
acquisition are presented with guidance provided for the num To determine the cumulated activity in the desired source
ber of measurements (data points) required. For those with regions, serial measurements of region activity must be made
expertise in dosimetry, highlighted sections (in boxed areas following administrationof the radiopharmaceutical.A general
below), examples and appendices have been included to provide principle for collecting activity data in vivo is that quantitative
calculational details as well as references for the techniques procedures or calculations must account for all the activity
involved. Appendix A provides a table of all symbols and administered. A minimum number of quantitative measure
definitions used in this document (Table Al). This document is ments must be made for each source region with an appropriate
intended also to serve as a guide for the investigator in choosing temporal sampling frequency. The latter depends upon the
the appropriate methodologies when acquiring and preparing patternof the uptakeand retentionof the activity in the various
product data for review by national regulatory agencies. sourceregions. Serial measurementsof activity in the source
The quantitative biologic data collection, analysis and pro regionscan be performedusingquantitativeimaging (including
cessing schemes presented here will yield results that represent planar scintillation camera, SPECT or PET), external nonimag
an average uniform distribution of activity within the source ing radiation monitoring, tissue sampling (blood or biopsy) and
regions. Nonuniform distribution of the radionuclide is not excreta counting. Total-body retention may be determined
considered. Bolch et al. (4), in the last issue of The Journal of using whole-body quantitative imaging, external probe moni
Nuclear Medicine, provide a more comprehensive treatment of
toring (nonimaging) and/or quantitative total recovery of body
nonuniform activity distributions. The potentials of SPECT and excreta. Activity in blood is readily obtained by direct sam
PET technologiesfor quantitativedata acquisitionand analysis pling. The resulting time—activity curve obtained for each
are discussed in brief; however, the intended emphasis of this source region can be analyzed using several different tech
document is on planar imaging techniques commonly available niques to provide A or T, for use in absorbed dose calculations.
in most nuclear medicine departments and laboratories. Planar This report describes data acquisition protocols that may be
imaging systems, particularly dual-head whole-body scanners, implemented to obtain source region activities and how these
provide both a cost- and time-effective method for acquiring data are analyzed and processed to estimate absorbed radiation
data, as required for sequential dosimetry studies. dose using the MIRD schema (a study design protocol checklist
is given in Appendix B). Three phases are involved:
II. ThE MIRD SCHEMA 1. Data collection: identification of the various source re
An accurate determination of the time-dependent activity in gions containing activity, determination of appropriate
tissuesof the body is required for calculatingabsorbeddoseto temporal sampling and acquisition of radioactivity or
target regions ofthe body using the MIRD schema (1,2). In this count data;
document, the term “region―will be used to designate sources 2. Data analysis: calculation of activity in source regions as
of radioactivity in the body and to designate targets for the a function of time using the count data and calibration
assessment of radiation absorbed dose. Source regions should factors obtained from quantitative measurement tech
always be considered target regions as well; however, the dose niques; and
to other target regions is usually of interest. In many cases, a 3. Data processing: integration ofthe time—activitycurves to
source region will be synonymous with a source organ, such as obtain the sum of all the nuclear transitions, cumulated
the liver, or a distributed organ, such as the bone marrow. In activity A or residence time T in each source region.
some cases, however, the source region may be an organ
subregion such as the caudate nucleus within the brain, or These three phases are described below in detail.
suborgan region of tissue (i.e., voxel).
The absorbed dose is defined as the energy absorbed per unit Ill. DATA COLLECTiON
mass. The meanabsorbed dose to tissue is given in the MIRD A. Introduction
schema by D = A X 5, where D is the mean absorbed dose (Gy To determine the time—activity profile of the radioactivity in
or rad), A is the cumulatedactivity (Bq . secor p@Cihi) and S sourceregions,four questionsmust be answered:
is the mean absorbed dose per unit cumulated activity (Gy/Bq@
1. What regions are source regions?
sec or rad/@.tCi. @@j)
The absorbed dose to the target may also 2. How fast does the radioactivity accumulate in these source
be expressed in terms of absorbed dose per unit administered regions?
activity, A0 (Bq or pCi), and the source region residence time 3. How long does the activity remain in the source regions?
‘1@,
defined as T A/An. Therefore, the mean dose to the target
and
per unit administered activity is given by D/A@ = T X S. The 4. How much activity is in the source regions?
estimation of absorbed dose is, thus, dependent upon two types
of information: The first question concerns identification of the source regions,
whereas the second and third questions relate to the appropriate
1. Time-deyendent (biokinetic) factors: those incorporated number of measurements to be made in the source regions, as
within A or @;and well as the timing of these measurements. The fourth question
2. Time-independent (physical) factors: those represented is addressed through quantitative external counting and/or
within S. sampling of tissues and excreta.
Each source region must be identified, and its uptake and sourceregionscanbe identified by tissueand excretasampling.
retention of activity as a function of time must be determined. All measurements must provide quantitative results in terms of
This providesthe data required to calculatecumulatedactivity absolute activity (Bc@Jor percent or fraction of administered
or residence time in all source regions. Each region exhibiting activity within source regions. Counts or count rates uncor
significant radionuclide uptake should be evaluated directly rected for administeredactivity are inadequatefor the quanti
where possible. The remainder of the body (total body minus tative analysis required in dose estimate calculations.
the source regions) must usually be considered as a potential For determining source region volume, especially for tumors
source as well. Mathematical models that describe the kinetic or regions of abnormal size where standard phantom geometries
processes of a particular agent may be used to predict its may be inadequate, the tomographic capability of SPECT or
behavior in regions where direct measurements are not possible PET may be required. Independent imaging modalities such as
but where sufficient independent knowledge about the physiol x-ray CT or MRI may also be used for source volume
ogy of the region is available to specify its interrelationship determination, provided a one-to-one correspondenceexists
with the regions or tissues in which uptake and retention can be between uptake and anatomic structure.
measured directly. These models can account for the presence
C. TemporalSampling
of metabolic products. Compartmental modeling can also be
After the source regions have been identified, the activity
used to separate the activities in regions that overlap on imaging
retention in these regions must be determined to answer the
studies, such as the renal cortex and renal pelvis or the liver and
following questions: How fast does the activity get to the
right colon. Some compartmental models are complex and
various sourceregions?How long doesthe activity stay there?
include compartments representing many different biologic
The calculation of absorbed dose requires that the region and
processes. Several computer software packages (SAAM, CON
total-body uptake, washoutand long-term retentionbe charac
SAM, BMDP and so on) are available to solve the equations
terized. Four basic kinetic models are considered here (calcu
associated with the different compartmental models (5—7).
lational examples are given in Appendix E):
The statistical foundation of a data acquisition protocol
designed for dosimetry requires that an adequate number of data 1. Instantaneous uptake (wash-in) with no biologic removal;
points be obtained and that the timing of these points be 2. Instantaneous uptake with removal by both physical decay
carefully selected. As the number of measurements increases, and biologic elimination (washout);
the confidence in the fit to the data and in the estimates of 3. Noninstantaneous uptake with no biologic removal; and
unknown parameters in the model improves. As a heuristic or 4. Noninstantaneous uptake with removal by both physical
general rule of thumb, at least as many data points should be decay and biologic elimination.
obtained as the number of initially unknown variables in the
Selection of optimal time points for sampling radiopharma
mathematical curve-fitting function(s) or in the compartmental
ceutical biodistributions in humans with the objective of defin
model applied to the data set. For example, each exponential
ing the uptake and retention pattern is dependent upon the
term in a multiexponential curve-fitting function requires two
biokinetic variables to be measured, e.g., effective half-times.
data points to be adequately characterized. On the other hand, if
Relevant quantitative data for similar radiopharmaceuticals
it is known a priori that the activity retention in a region can be
obtained in animal or human trials may serve as a guide for the
accurately represented by a monoexponential fimction, restric
initial choice ofthese time points. The selection ofthe sampling
tions on sampling times are less stringent as long as enough data
times can have a significant effect on parameter estimation
points are obtained to derive the fitted function. Because of
precision (10). Readers interested in pharmacokineticsmay
problems inherent in the collection of patient data (e.g., patient
refer to Appendix D for a rigorous mathematical treatment of
motion, loss of specimen and so on), the collection of data
the statistical consequence of applying inappropriate temporal
above the necessary minimum is advisable.
sampling.
As a practical guide for those who are designing clinical trials
B. Identification of Source Regions
with new radiopharmaceuticals, consideration should be given
The types of measurements required for identifying source
to which one of the following three general categories most
regions can be categorized into four basic groups: imaging,
closely describes the experimental status of the new compound
discrete probe monitoring (whole-body or region counting),
(11):
tissue sample (blood or biopsy) counting and excreta counting.
Published literature on humans or animal studies along with ex Category 1. Radiopharmaceutical compounds for which nei
vivo or in vitro data may yield valuable information to assist in ther in vivo animal studies nor human pharma
identifying tissues that accumulate a particular radiopharmaceu cokinetic clinical trials have been conducted.
tical. This information, if available, may serve as the basis for Included also in this category are those com
designing the preliminary data acquisition protocol. poundsfor which pertinent information is lim
The distribution of a radiopharmaceutical in the body can be ited to general chemical and biologic class
determined by sequential imaging methods. Quantitative data comparison (e.g., antibody fragments with a
can be obtained from planar scintillation camera images or chelated metal, single-chain haptens undergoing
tomographic SPECT or PET images (example quantitative a standard iodination procedure and so on). The
imaging protocols are given in Appendix C). Because of the labeled compound should be injected into at
time-consuming and more complex nature of quantifying least two animal species, as per Food and Drug
SPECT and PET tomographic image data, conjugate view Administration (FDA) guidelines (12) to deter
quantitative planar imaging, using anterior and posterior views, mine the uptake and retention characteristics in
is the most widely used method (8,9) and will be emphasized in all source regions prior to administration in
this report. The whole-body imaging capability of current human subjects. As a starting point, pharmaco
single- and dual-head scintillation cameras is especially useful kinetic temporal sampling points may be se
for this purpose. Total-body measurements may also be ob lected basedon the physical half-life (T) of the
tamed with an external nonimaging radiation probe. In addition, radionuclide or the effective half-time (Te) of
MIRD PAMPHLET
No. 16 . Siegel et al. 39S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
the pharmaceutical compound in the blood or used to approximate the measured time—activitycurve. Because
source region of interest (ROI). A minimum of many processes in the body are governed by first-order kinetics,
three sampling points should be chosen for each sums of exponentials are often used as these mathematical
clearance exponential, if known or suspected. A functions. As illustrated in Appendix D, the number of mea
general protocol could involve one or two data surements needed is directly related to the number of exponen
points taken at some fraction of Te; one near Te tial terms in the retention curve. Ideally, two to three measure
and one or two other data points taken at —3X ments per exponential term are required. Therefore, if the
Te and 5 x T@. Data points can serve for more total-body clearance curve is a single exponential, three data
than one exponential. If nothing is known about points are needed. A biexponential retention curve would
the retention of the compound in animals and require a minimum of four and preferably six data points and so
humans, multiple sampling times, say a dozen, on. When the distribution of samples is appropriately spread
should provide a reasonable starting point. In over the time interval for which activity is taken up and
sufficientdata pointsmay make it impossibleto removed, then the mathematically simulated curve is an accu
properly characterize a multicompartmental rate representation of the time—activity profile. However, re
model, thereby resulting in patient irradiation to strictions on the timing or number of samples may necessitate
little purpose. As an alternative, an estimate of extrapolation of the simulated curve to time intervals for which
sampling times may be based on the molecular no data exist. These intervals extend from t 0 (the time of
weight and size of the compound as well its radiopharmaceutical administration) to the first sample and
ionizationpotentialor chargeas comparedto from the last sample point to very long times postadministra
similar compounds in the literature. tion, i.e., to the total clearance time of the compound.
Category 2. Radiopharmaceutical compounds for which the For further refinement, both the uptake phase and the
biodistribution has been characterized in at least washout phase must be characterized accurately. The uptake
two animal species and some preliminary phar phase is the period over which the region activity increases
macokinetic data have been acquired in a patient from zero (at the time of administration) to its maximum value.
study (5—20normal volunteers or patients with Answers to the following questions will characterize this phase:
known disease). Extensive literature on this Was the uptake slow or fast (qualitative information)? What
compound or similar analogs should be avail was the uptake half-time (relevant if curve fitting)? What was
able. The investigator should be able to project the fractional uptake in the region? If a compartmental model
a reasonable hypothesis as to whether clearance was used, were there any special parameters necessary to
parameters are mono- or biexponential or more describe the uptake phase? The duration of the uptake phase in
complicatedbecauseof retentionandexcretion a given region may vary widely with different radiopharmaceu
from multiple specific uptake compartments. ticals. It is often assumed that uptake ofthe radiopharmaceutical
Again, at least three sampling points should be is instantaneous (i.e., the peak uptake occurs immediately
taken for each exponential, based on Te in the following administration); however, in many cases this is
blood or source region of interest in either inaccurate. Errors generated by this assumption will result in an
animals or humans. overestimate of the source region residence time and thus an
Category 3. Radiopharmaceutical compounds for which all overestimate of the dose component from that source to various
pertinent literature related to the pharmacokinet targets. The instantaneous uptake assumption may be consid
ics of similar labeled compounds has been ered satisfied when extrapolation of the observed data back to
compiled, and extensive animal model and hu t = 0 results in less than a 10% error in the residence time as
man pharmacokinetic data have been acquired. compared to the residence time estimated from a curve fit that
The compound has most likely completed Phase accurately reflects the noninstantaneous uptake phase.
I and Phase II clinical trials. Uptake should be The washout (clearance) phase is the period over which the
characterized in all the source regions for the activity retained in a region decreases from its maximum value.
animal models and in the human subjects. Re In addition to physical decay, biologic elimination may also
mainder ofthe body and blood retention, as well occur, making the effective washout time more rapid than
as urinary excretion characteristics, should be physical decay alone. This portion of the curve will also affect
known. From these data, a cost-effective, highly the A calculation (area under the curve) independent of the
directed Phase III clinical trial may be designed. uptake phase. For example, if the uptake phase is ignored (i.e.,
by extrapolating the linear portion of the semilog time—activity
Radiopharmaceuticals that show rapid uptake or washout curve back to time 0 and integrating the area under the
may need two or three activity measurements at early times extrapolated curve), the overestimation of area under the curve
postadministration. Radiopharmaceuticals that show very rapid will be greater for faster washout conditions. Therefore, if rapid
uptake may require no wash-in measurements. Similarly, two, washout occurs, as is typical of freely diffusible ions (such as
three or more additional data points may be required during the 18F@and 201TF1),data must be acquired at early times postad
long-term retention phase to adequately characterize the late ministration of the radiopharmaceutical.
time course of the radiolabeled material. Where the presence of To characterize the long-term retention (tissue incorporation
diseasemayalterregionfunction,theacquisitionof additional or trapping of the radiopharmaceutical), measurements at times
datain theaffectedtissuesmaybeadvisable. equal to multiples (2—5)of the effective half-time are needed.
From serial activity measurements, one can determine the Frequently, this may be accomplished within several hours
residence time defined asthe area under the time—activitycurve postadministration; however, measurements are more often
(i.e., cumulated activity A)divided by the administered activity required at much later time points. If the data acquisition ends
A0. Errors in estimates of A and thus in the residence time may too early, the area under the curve would not be representative
be introduced as a result of improper sample timing and an of that under the true retention curve because the long-term
inadequate number of samples when analytical functions are retention phase would not be included. Because the late
retention is usually characterized by the longest effective effects from scattered radiation, statistical noise associated with
half-times, the calculated area under the curve would typically low count densities and attenuation. The spatial resolution in
be an underestimate. Large errors can result in residence time planar images depends upon the intrinsic resolution of the
calculations if the long-term retention phase is ignored. For NaI(Tl) scintillation crystal, the collimator and Compton scat
example, if the ratio of the effective washout half-times of the ter. The collimator also plays the important role of defining
slow-to-rapid clearing components is 4: 1 or greater, the errors geometric sensitivity (the fraction of emitted photons per unit
start at 30% and rapidly increase to 100% (see Appendix D). time which reach the crystal from a predefined location/
The physical half-life of the radionuclide plays an important direction). Parallel-hole collimators are recommended for quan
role in determining the period over which reliable data may be titation of radioactivity because they exhibit less geometric
obtained. After approximately five half-lives, count statistics distortion when compared to other types of collimators (e.g.,
can be extremely poor, and the data may be less reliable. Under converging, pinhole and diverging), with the result that sensi
these conditions, counting times should be increased, if possi tivity will be relatively independent of source-to-collimator
ble, to maintain adequate statistical reliability. distance within a defined region. Spatial resolution of a scm
In summary, improper sampling may introduce error in tillation camera degrades, however, as the distance between the
calculated values of A (and thus residence time) under condi source and the detector increases. Consequently, the detector
tions that neglect uptake, early rapid washout or the long-term should be as close as possible to the subject.
retention component. The first two conditions relate to insuffi Biodistribution (kinetic) data for radiopharmaceuticals within
cient sampling of tissue activity at times just after administra the whole body and specific regions may be obtained from
tion and will result in an overestimation of the area under the planar scintillation camera views. The accuracy of this method
time—activity curve. The loss of activity from the source region will be greatest for radiopharmaceuticals distributed in a single
can occur in one of two ways: physical decay only or physical region or isolated regions that do not overlap (nonsuperim
decay plus biologic elimination. Clearly, if biologic processes posed) in the planar projection. Determination of absolute
are contributing to removal of activity from the source region, radioactivity requires defmition of ROIs corresponding to the
the residence time for that region will be less than if losses are source regions as differentiated from their adjacent or surround
due to physical decay only. When instantaneous uptake is ing background activity. Most scintillation camera systems used
assumed, the peak uptake (at time 0) is determined by extrap in clinical nuclear medicine are computer-based and provide
olation back to t = 0 based on the data measured at later times. software for generation of ROIs and statistical analysis. Com
The extrapolation process results in greater overestimates of the mercial software packages typically provide the means for
peak uptake as the rate ofbiologic removal increases. The major creating ROIs of simple geometry (circles, ellipses and rectan
implication of overestimating peak uptake and not accounting gles) and for interactive drawing of irregularly shaped ROIs.
for noninstantaneous uptake is an overestimate of the residence This latter capability is important for determining activity in
time for that particular source region. The third condition regions that cannot be described by simple geometries such as
(neglect of the long-term retention), resulting from insufficient most source regions and tumors. Automated or semiautomated
sampling at times long after administration, will result in an methods for ROI generation are also available. Statistical
underestimate of the residence time. Error analyses associated analysis packages provided with these commercial systems
with these considerations are discussed in Appendix D. allow determination of the number of pixels contained within
the ROI, the planar area encompassed, the total number of
D. Data AcquisitionConcepts/Overview
counts and the mean and s.d. ofthe counts per pixel. The count
The acquisition of count data for the source regions can be
rate within the ROI is obtained from the total number of counts
performed with a number of measurement techniques. They
divided by the image acquisition time. Sequential imaging as a
include:
function of time postadministration of the radiopharmaceutical
1. Quantitative imaging, with a planar scintillation camera or provides the time dependence of activity (time—activity curve)
tomographic SPECT or PET system; within the source regions. Generally, the computer software
2. External nonimaging radiation monitoring with a Nal packages permit storage of the ROIs, allowing reproducible,
probe or GM survey meter; efficient analysis of serial image data.
3. Tissue sample counting of the blood or biopsy specimen;
and 2. Conjugate View Counting
4. Excreta counting (e.g., urine and feces). An overview of the methodology. The most commonly
employed imaging method for quantitation of radioactivity in
The technique used depends on the nature of the source
vivo uses 1800opposed planar images (known as the conjugate
region. For instance, the total-body activity can be measured by
view approach) in combination with transmission data through
whole-body monitoring with a nonimaging device, whole-body
the subject and a system calibration factor (9,13—16). This
imaging or stationary counting with a scintillation camera or
technique offers an improvement over the single-view proce
quantitative recovery of body excretions. Individual tissue
dures involving comparison with a standard phantom under
activity can be determined by quantitative imaging, blood
sampling and biopsy specimen measurement.
fixedgeometryin thattherigorousmathematical formalismfor
conjugate view quantitation provides correction for source
IV. QUANTITATIVE MEASUREMENT TECHNIQUESI@ thickness, inhomogeneity and attenuation. Ofsignificance is the
IMAGING fact that calculational results are theoretically independent of
the source depth in tissue. The transmission scan, which
A. Scintillation Camera Planar Imaging Techniques involves counting an external source of activity through the
1. Introduction source ROI to correct for attenuation, is generally performed at
A number of factors affect the accuracy of quantitating one time point, either at the beginning of the study before the
radioactivity with a scintillation camera. These include limita administration ofthe radiopharmaceutical or at a later time with
tions on energy resolution, degradation of spatial resolution due appropriate correction made for activity in the body.
to collimator septal penetration by high-energy photons and The conjugate-view image pair is typically an anterior and
MIRD PAMPHLET
No. 16 •
Siegel et al. 41S
@ M@
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
where:
@ A1#@
(@@;/2)
@ f; sinh (@.tjtj/2) Eq. 2
@;-@ A2 p2
and
@ I
‘3 n n
42S THEJOURNAL
OFNUCLEAR
MEDICINE
•
Vol. 40 •
No. 2 (Suppl). February1999
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
: @A :
Ct)
Cl)
@ t@ .@*1_ , ,
@ v--- ‘ ,
LU
0.01
FiGURE 2. A schematic diagram showing a single source region surrounded
by uniformbackgroundactivity.Asdescribed inthe text, the count rate from
an adjacent,equal-@zed ROlQ@)may be usedin a simp@background
subtractionmethodto correcttheconjugate-view countratesforthecontri
bution of the over/unded@ngactMty.
I I I I I I I I I I III
consider the portion of the background equivalent to the source
2 46 8 10 12 14 16 18 20 22 24 26
region volume and may underestimate the source region activity
due to oversubtraction ofbackground (1 7). In addition, defining WATER PHANTOM ThICKNESS (4 cm)
background ROIs is sometimes difficult as care must be taken
to avoid hot or cold areas, such as major blood vessels. FiGURE 3. An experimentally measured transmission curve in a tissue
A simple geometrically based subtraction technique may be equivalentmediumforthetwo photopeaksof 67Gausinga medium-energy
collimator.Thecurveexhibitsa significantshouldercharacteristicof scat
applied to correct for the oversubtraction ofbackground activity tered radiation effeCtSWhich requires use ofthe pSeUdOeXtrapOlatiOnnumber
and to avoid the problems inherent in background ROl defini modificationofthe conjugateviewquantitationequations.Fort 6 cm, @f
=
tion. Under conditions in which the background activity is ne@ andtheactivityisCalculated accordingto Equation10.Adaptedfrom
uniform, the fraction ofthe geometric mean counts (IAIp)@2that a previousreport (20).
originates from the region (or volume) of interest alone is given
as:
The individual f factors are defined according to Equa
@ F = {[l —(1ADJ/IA)(l —t@/t)][l —(IADJ/IP)(l — tion 2, whereas Y2 A2/A3 and y4 A4/A3 represent the
Eq. 6 ratios of the activity in the adjacent regions to that in the
@ where ‘ADJ the count rate through the patient from a region source volume. Calculation of the k('y) constant requires
adjacent to the organ ROI with equal area. The A/P count rates estimates of the factors (f, Y2' ‘Y4)that may be obtained
(IA/IP)' source thickness (9 and patient thickness (t) remain as
through analysis of additional views in combination with
previously defined. Direct patient measurement will provide t, the main conjugate view pair. The k(y) correction may
and ; may be obtained from scaled lateral views or through use become significant when the activity in the adjacent re
of standard organ dimensions as appropriate (18, 19). The gions approaches 10% of that in the source volume (9,16).
lateral view may also be used to document the assumption of
c. Correction for Scattered Photons. As stated, the above
uniform background activity distribution. For the situation in
analytical methods (Eqs. 1—9)have been derived under the
which the source is located near the midline, a simplified
model of narrow-beam geometry in which scatter effects are
expression for this fraction is:
negligible. The investigator must verify that these conditions
@ F [1 —(IADJIIA)(l — Eq. 7 are met for his/her experimental data acquisition configuration
before applying these methods. High-resolution collimators and
In either case F(IAIP)―2would replace (IAIp)@2in Equation I. narrow, asymmetric energy window settings assist in approxi
ii. Analytical Methodology. Figure lc illustrates the mating these conditions. However, in most nuclear medicine
situation in which a well-defined source is surrounded by imaging applications, broad-beam geometry predominates and
background activity uniformly distributed in the adjacent scatter effects are significant. The presence of scattered photons
tissue. For this configuration, the source region activity A3 originating from outside the source organ ROI will artificially
is given by the expression: inflate the ROI count density and introduce error in the
quantitation of radioactivity. The following sections describe a
i-i@j;
f3 number of approaches designed to correct for scatter.
A3= Eq. 8 i. Pseudoextrapolation Number. An illustration of the effect
of scattered radiation is provided in Figure 3, where the curve
where for the relative transmission (@?1)
as a function of depth within a
k(y) = {l + (y2f3/f2)2+ (-y4f3/f4)2+ 2-y2'y4(f@/f2f4)
cosh [(@2t2 scattering medium exhibits a shoulder on the semilog plot rather
+ 2@3t3+ p.@t@)/2]
+ 2y2f@/f2cosh [(@2t2+ @3t3)/2] than the straight line expected from the simple narrow beam
model for transmission of @J= I/Is = e@et. The shoulder is a
+ 2y4f3/f4cosh [(@3t3+ @i.@t@)/2F
1/2•
Eq. 9 manifestation of increased count density within the ROI caused
by acceptance of scattered photons from activity close to the
MIRD PAMPHLET
No. 16 . Siegel et al. 43S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
collimator. A modified expression to account for this situation sponding to the straight-line portion of the transmission curve
has been developed by Thomas et al. (20) in which the intercept (Figure 3). As described previously, the parameter n is taken as
n is thought of as a pseudo-extrapolation number (following the a constant; therefore, Eq. 10 will not be valid for regions near
terminology used with radiobiology survival curves). For the the surface (i.e., in the shoulder of the curve). The DDBF with
straight line portion ofthe semilog plot ofFigure 3, @J= ne_@@t its variable, experimentally determined look-up table will in
@ where is the equivalent narrow-beam geometry linear dude compensation for this effect. The DDBF technique
attenuation coefficient. For a single-source region embedded requires a more complicated protocol of calibration set-up
within a scattering medium with no surrounding background procedures and calculational algorithms than does the transmis
activity, the modified expression for the source activity would sion method of Thomas et al. (9); however, it would have its
be: greatest utility under conditions of significant scatter effects
such as those encountered with wide windows and high
ARJ sensitivity collimators. The investigator may determine the
Eq. 10
@ j— n@?1C necessityof following this procedureby obtaining a transmis
sion curve similar to that in Figure 3 under the appropriate
This is valid for source regions located at depths beyond the instrumental conditions and assessing the extent ofthe shoulder
transmission curve shoulder corresponding to the straight line region. To implement the buildup factor algorithms, a computer
portion of the plot. program may be written requiring only four input parameters
@ ii. Buildup Factor Method. Another approach to the scatter (IA' ‘P' and C) to calculate the activity.
ing problem has been described by Wu and Siegel (21). Their A variation on the buildup factor theme has been introduced
method employs conjugate views and experimentally deter by Siegel et al. (22,23) involving a depth-independent buildup
mined depth-dependent buildup factors (DDBF) which are factor (DIBF). The DIBF method is easier to apply than the
defined as the factor by which the transmission is increased DDBF method in that only the single parameter B(oo) is
under broad-beam (scatter) conditions. No transmission mea required which is independent of source size and depth for a
surement is made across the patient. An iterative calculational given window setting (23). The utility of the DIBF approach
procedure is used involving the solution of the set of simulta has been verified by a number of investigators (1 7,24).
neous equations representing the conjugate-view count rate
pairs. In the DIBF method, the transmission factor @Tas a
function of depth d in tissue equivalent material with linear
‘A CodB(d)e_I@I[e_@tJ2(l —e_@tJ)//.Ltj] attenuation coefficient p. is fit to the function:
= C@jB(d)e@―'{sinh (p@t@/2)/(p@t@/2)] Eq. 11 @J-
= 1 —(1 —e)B(@@), Eq. 14
and where B(oo) is the buildup factor at infmite depth (corre
‘P= C@B(t —d)e_@t_@[e_@LtI2(l —e_1.LtJ)/p,tjl sponding to the straight line portion of Fig. 3 and equiva
lent to the pseudoextrapolation number n previously dis
= C@B(t — d)e_@t_@[sinh (ptj/2)/(p.tj/2)], Eq. 12 cussed). The expressions for the experimentally measured
conjugate view count rate pairs are:
with the depth-dependent buildup factor B(d) defined as
‘A C@[l — (I —e_I@@)B@][sinh (p.tj/2)/(p.tj/2)]
Cd
B(d) = C@je'@―
‘ Eq. 13 Eq. 15
and
where Cd is the count rate measured at a depth d at the
center of the lesion in a phantom, C@ is the count rate ‘P= C@[l —(1 —e_@t_dI))B@][sinh (ptj/2)/(p.t@/2)].
measured in air at the same source-to-gamma camera
Eq. 16
distance, j.Lis the attenuationcoefficient, t is the overall
patient thickness and ; is the source thickness. Equations 15 and 16 may be combined to give the ratio:
The depth-dependent buildup factors B(d) must be deter [1 —(1 —e@@I)B(@@)]
mined from experimental measurements and depend upon the ‘A―P [1 —(1 —e_@t_dI))B@] ‘ Eq. 17
radionuclide, collimator type, energy window, measurement
geometry, source depth, source size and source thickness. The which may be solved numerically for d by varying the
@ B(d) values are stored in the computer memory as specific value of d under the constraint 0 d t. Either Equation
tables to be used in association with the appropriate system 15 or 16 may be used to calculate C@ from which the
conditions. The equation set (Eqs. 11 and 12) are solved activity may be calculated through use of the calibration
iteratively for C@ and d under the initial assumption that the factor. Alternatively, if the anterior depth d is measured,
source is located exactly at the center by using B(tJ2) in both COdmay be calculated from the expression for ‘A (Eq. 15).
expressions. This initial calculated value of d is used to select
buildup factors B(d) and B(t —d) from the look-up table for the iii. Multiple-Energy Window Techniques. As already mdi
second iteration and new values for C@ and d are calculated. cated, use of an asymmetric energy window positioned on the
This iterative procedure continues until a solution for C@ high side ofthe photopeak will decrease the amount of scattered
converges. The activity is then determined by A@= C@/C, radiation accepted from Compton scattering of the primary
where C (count rate per unit activity) is the system calibration (photopeak) photons. However, the number of counts per unit
factor. time will be reduced, thus lowering the signal-to-noise ratio
As pointed out by Wu and Siegel (21 ), the pseudoextrapola (25,26). In addition, scatter from higher-energy photons present
tion number modification method of Thomas et al. (20) is in the specific radionuclide decay scheme will still be included
equivalent by definition to the buildup factor at depths corre within the window.
@ ever, it may vary as a function of position on the camera face) P(364) = —PS(364) —P5(637/723), Eq. 19
and the ratio of the scatter component in the two windows
@ where is the total count recorded within the photopeak
(under the assumption that the relative shape of the scattered
photon energy spectrum remains constant). Calibration is per window. Three scatter multipliers (kl = P5(637/723)!
formed using a two-step process: US(637!723); k2 = LS(637!723)/US(637!723); and, k3 =
PS(364)ILS(364)) will allow counts detected in the upper
1. The fraction of unscattered counts in one window com window (US) to be used to sequentially remove scatter
pared to the other is found for a small source that is events from the lower two windows (PP and LS). Once this
essentially scatter-free; and is done, the 364-keV scattered photons that fall into the
2. A sequence of planar images is acquired for a flat source photopeak (PP) window can be removed. Techniques for
placed at various depths in a nonradioactive water bath. estimating the scatter multipliers empirically bi using the
@ scintillation camera response to 13‘Csand Cr (which
Triple-energy window (TEW) techniques were introduced for when superimposed may be considered to simulate the 1311
SPECT (29) but can be implemented for conjugate planar views spectrum) have been described (30).
to provide a more accurate correction in the situation when iv. Digital Filtering Techniques. Scattered photon ef
scatter from higher-energy photons “spillsdown― into the fects may also be removed through computer processing
primary photopeak window. An example of this is 1311where utilizing digital filter techniques. A number of methodolo
the 637-keV (7.3%) and 723-keV (1.8%) photons scatter into gies have been implemented for both planar and SPECT
the 364-keV (81%) window. One relatively straightforward imaging. The Wiener restoration filter has been shown to
correction procedure involves establishing adjacent windows on compensate for the blurring effects of scattering and
either side of the photopeak window, such that the area of the collimator septal penetration (31—35).As a general rule,
two similar adjacent windows is equal to that of the photopeak the Wiener filter uses minimization of the mean square
(Fig. 4a). The corrected (true) photopeak counts CT are given by error between the restored image and object function as the
the expression: criterion for optimization. The technique exploits a priori
estimates of the object and noise power spectrum deter
mined from the degraded image.
@ CT —CLS Cus, Eq. 18
Energy-weighted acquisition (EWA) represents a dis
tinctly different image filtering approach in which a differ
@ where is the total count recorded within the photopeak
ent spatial filter (weighting function) is applied for each
window, whereas CL5 and C@5are the counts within the lower
energy to remove the effects of Compton scattering (36—
and upper scatter windows, respectively. In this model, subtrac
39). The EWA approach uses all scintillation events and
tion of the adjacent windows is assumed to compensate for the
fractionally weights their contribution to the image forma
high-energy photon scatter tail, upon which the true photopeak
tion, depending on energy. The sign and magnitude of the
events ride.
MIRD PAMPHLET
No. 16 •
Siegel et al. 45S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
collimators have each contributed to the improvement of the
weighting values are related to the probability that a given
counting statistics available for SPECT. The use of filter
event represents scatter or primary radiation. For events
functions with cutoff frequencies matched to the system's
more likely to be scattered photons, negative weights are
spatial resolution has proven to be useful for reducing the
assigned. The weighting function incorporates information
effects of statistical uncertainty (51—53).
about the instrument response (e.g., collimator characteris
Despite the limitations, SPECT has been used to measure
tics and so on), as well as the scattering environment.
activity levels in source tissues with acceptable accuracy and
Although, in principle, the weighting functions should be
precision. The SPECT system must be properly calibrated to
determined individually for each imaging configuration
ensure acceptable reconstructed uniformity, sensitivity, spatial
encountered in practice, functions specific to the given
resolution, linearity and count rate performance. Acquisition
radionuclide and camera system may be stored and used
variables must be selected to provide adequate counts per view,
successfully(36).
appropriate spatial and angular sampling and the desirable
a Single-View
Effectk'e
Point
Source
Methods imaging orbit. Selection takes into consideration the intrinsic
Occasionally source volumes or features are seen on only one and system performance properties, in vivo activity levels at the
scintillation camera view. This may occur for a superficial time of measurement and the determination of imaging times
source region with minimal uptake and/or relatively high based on subject comfort as well as the uptake and clearance
background conditions. For these situations, a simplified single properties of the radiopharmaceutical. In addition, suitable
view effective point source method (35,40) may be used instead software for image reconstruction and image processing is
of the conjugate view method. The activity calculation is based necessary(41—45).Researchinto new methodsfor improving
on the known (previously determined) effective linear attenua SPECT quantitation is under way, and the approaches for
tion coefficient, P@e'and the measured depth of the source activity quantitation in vivo are expected to change as new
region, d, according to the expression: methods are validated for scatter and attenuation correction.
Quantitative SPECT scans can be used for all imaging time
A = IAe'@'F/C, Eq. 20 points if the tomograph has sufficient sensitivity to keep
acquisition time to reasonable intervals so that patients will
@ where ‘A the counts per unit time within the image ROI
tolerate the procedure. Alternatively, the sequential images can
(denoted here as the anterior view), and C is the system
be acquired by planar procedures, with a SPECT image taken at
calibration factor (count rate per unit activity). An orthogonal
one time point during the study. The result from the SPECT
scaled image may be used to assist in depth determination. The
scan can be used to constrain the planar data at that point. Then,
factor F corrects for background activity that is counted within
assuming that the image contrast is time invariant, one can
the ROI (see Eqs. 6 and 7).
apply the correction factor required for that constraint to the
B. SPECT Techniques earlier and later planar-imaging results (54).
Because activity in under- or overlying regions and irregular 1. Phantom Studies
distribution of background may interfere with accurate quanti The use of software-generated phantoms and fabricated
tation, SPECT imaging has the potential for improving the anthropomorphic phantoms is recommended to test and validate
accuracy ofplanar imaging measurements. Quantitative SPECT the performance properties of the SPECT system as well as the
(41—46) enables the determination of actual tissue activity reconstruction and correction methods, and to develop calibra
concentration (e.g., MBq/cm3) and the associated total volume tion factors for converting counts observed in ROIs to activity
(e.g., cm3) over which the activity is distributed. Although levels (42,43). Phantoms that replicate the size, shape, electron
attenuation correction schemes, nonoptimal statistics, collima density, mass density and elemental composition of the source
tor resolution and scatter result in some limitations to the tissues and surrounding structures can be used with activity
accuracy of quantitative SPECT determinations of radioactiv levels comparable to the range expected in human subjects to
ity, SPECT imaging can provide major advantages over planar obtain these calibration factors for SPECT image quantitation.
conjugate pair imaging for selectively detecting and discrimi
nating activity in source tissues from surrounding structures. a Reconstruction
Methods
andAttenuation
The use of SPECT is particularly advantageous for measuring SPECT reconstruction and image restoration may be imple
activity in regions in which activity in contiguous overlying, mented using either an iterative or noniterative algorithm (41).
underlying or adjacent structures limit the accuracy of the A critical parameter affecting the accuracy of SPECT data is
conjugate pair planar view measurements. The use of appropri attenuation. Attenuation compensation is performed either by
ate SPECT acquisition parameters, a suitable reconstruction assuming uniform attenuation throughout the body, or measure
algorithm for image reconstruction and image processing, a ment of the nonuniform attenuation coefficient distribution.
properly calibrated imaging system and a constant distribution Compensation methods for the uniform attenuation assumption
of radiopharmaceutical during the imaging procedure can all used in SPECT can be divided into three classes: preprocessing
provide an accurate assay of radioactivity in vivo (41—46). correction, intrinsic correction and postprocessing correction.
To date, the majority of reports have involved the use of a The most widely used approach in current commercial SPECT
single rotating scintillation camera for SPECT image acquisi systems is Chang's first-order postprocessing method (55),
tion to determine activity in regions and tissues. The relatively wherein the image is initially reconstructed using filtered
low sensitivity of these systems has held back development of backprojection and then multiplied on a voxel-by-voxel basis
correction techniques that would allow improved accuracy for by a mean depth-dependent attenuation correction. The appli
quantitating nonuniform activity distributions in inhomoge cable equation is:
neous and irregularly shaped attenuation media (41—43,46—50). 1
Numerous approaches have been implemented to correct for the CF = Eq. 21
limited counting statistics available from conventional SPECT
imaging. Multiple detector SPECT systems, either multicamera @i@ri
or ring systems, and utilization of fanbeam and conebeam
46S THEJOURNAL
OFNUCLEAR
MEDICINE
•
Vol. 40 •
No. 2 (Suppl)•
February1999
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
A LU
B Scatter DirectWindow
Window Cpp CT+SC FiGURE5. Dual-energywindowtech
niques. (a) The photopeak is dMded
into two equal nonoverlapping win
dows (lower, L upper, U) whk@hencom
passthe standardsymmetricacquisi
tion window. Adapted from a previous
report(28).(b)Thespectrumis dMded
a: intoa scatterwindowwherethecounts
SC/cs are attributedentiraiyto lower-energy
C scatteredphotons(Ce)aiid a direct
0 window over the photopeak where the
C-) observed counts (C@)are composed
of bothscattered(Sd andtrue photo
peak (C1-)counts.The assumptionis
made that the scattered counts S,@are
correlatedwith Cs by the samecon
Energy(keV) Energy(keV) slant k for all imagepixels.Adapted
froma previousreport(26).
@ where CF is the correction factor, is the transmission at angle attenuation coefficient within the voxel and 1 is the gamma-ray
i from the voxel through the body to the skin surface and N is path length through the voxel at the projection angle of interest.
the total number ofprojections. This correction is quantitatively Examples of reconstruction algorithms available are the maxi
rigorous only for uniformly attenuating media. Such uniform mum-likelihood expectation-maximization (ML-EM) algorithm
attenuation is often assumed for abdominal imaging. With (69) and the space-alternating generalized EM (SAGE) algo
commercial systems, the skin surface is either defined by the rithm (70). The utility of nonuniform attenuation correction
user or by a simple algorithm. A typical algorithm uses an with an iterative algorithm has been verified by tests of relative
intensity threshold that is a prechosen percentage of the maxi diagnostic accuracy in coronary heart disease (71).
mum intensity in the image. This image can either be a single
transverse slice, a sum of several slices, or a composite ofall the a Scatter@ci
Photon
Correction
Methods
slices in the reconstruction. Several different approaches have been advocated to reduce
In areas of nonuniform density, such as the thorax, a the detrimental effects of detected scattered photons on activity
nonuniform attenuation map should be used. This map can be quantitation and on image quality (28,41—43,72—83). These
calculated from radioisotope transmission data obtained at the methods include use of a reduced-value attenuation coefficient,
same time the emission data is being acquired. The source of the dual- or multiple-energy window acquisition, convolutionlde
radiation can be a sheet source (56), a translating line source convolution methods (in space or energy), spectral fitting,
with parallel-hole collimation (57), a point source with cone energy-weighted acquisition and computer modeling. Most of
beam collimation (58) or a line source with fan-beam collima these methods are applicable to planar imaging as well.
tion (59—61). Some multihead commercial systems currently The reduced-value attenuation coefficient approach (broad
provide such an option (62,63). An important parameter in beam geometry) compensates for the additional scatter counts
these procedures is the collimation used. For example, fan present by undercorrecting the measured image counts. Dual
beam collimation can lead to truncation of the data at the edge photopeak window methods provide a subtractive method for
of the patient. A compensation for this truncation must then be correction of scattered photons within the lower energy portion
used. The relative merits ofdifferent methods ofcollimation are of the photopeak (28, 78—81). In one technique, the photopeak
still under investigation. region is divided into two nonoverlapping energy windows
Another option in the case of nonuniform attenuation is the (Fig. 5a) from which a regression relationship is obtained
use of an attenuation map obtained from a CT image by a between the ratio of counts within these windows and the
suitable energy extrapolation from the gamma-ray energy of the scatter contribution within the total photopeak window (28).
x-ray scanner (@70 keV) to the energy of the gamma ray of The scatter distribution within the photopeak window is esti
interest (54, 64). This map is useful only ifthe CT image set can mated through calibration to determine the regression coeffi
be fused (i.e., superimposed exactly) into the space of the cients for the specific system and pair of energy windows
SPECT image set. Various methods have been proposed for without the use of arbitrary scaling factors. The more general
image fusion (65—67). ized scheme for dual-window acquisition involves acquiring
The nonuniform attenuation map can be incorporated into the two sets of images, one at the radionuclide photopeak and a
Chang attenuation compensation method (50,68,69) or applied second one suitably positioned to image Compton scattered
through an intrinsic attenuation correction. The latter usually photons. An appropriate fraction of the scatter image is sub
occurs by means of one of many iterative reconstruction tracted from the photopeak image to compensate for the
algorithms. These algorithms rely on finding a consistency Compton scatter photons (80,81).
between the measured data and results from the current estimate Koral et al. (80) have described a method which utilizes an
of the image after forward projection to the detector. The adjacent lower energy scatter window to obtain a scatter
forward projection utilizes a model to specify the process. multiplier “k―
defined as the ratio of the scattered counts in the
Within that model, the attenuation is usually accomplished one photopeakwindow to thosein the lower window (Fig. Sb). The
voxel at a time by the formula true photopeak counts CT are given by the simple relationship:
@ where @;; is the gamma-ray intensity entering the voxel, I' is the where is the total count recorded within the photopeak
@ intensity (count rate) leaving the voxel, is the average window and C@ is the count within the scatter window. The
MIRD PAMPHLET
No. 16 •
Siegel et al. 47S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
multiplier k may be determined empirically from imaging In areas of nonuniform density, such as the thorax, it is
experiments conducted with sources inside scattering medium necessary to incorporate a nonuniform attenuation map in the
phantoms and in air. Chang attenuation compensation method (102). Also, the use of
Deconvolution techniques are either true deconvolution a fixed threshold method to distinguish between source and
methods (83—85) or convolution calculations that produce a background voxels is based on a global maximum and can be
scatter estimate which is subtracted from the usual image to incorrect if the activity in the selected volume of interest
provide correction (74). A scatter function (74,84,85) or an changes rapidly with time. In addition, Mortelmans et al. (103)
energy-response function (83) is required to carry out the reported that a single fixed threshold value may not be accurate
calculation. The scatter function has spatial “tails―
that define enough because the threshold value was found to be dependent
the magnitude and extent of the effect of detecting scattered upon object size and image contrast.
photons. It is usually measured by placing a radioactive source Because SPECT is a three-dimensional imaging modality, it
within a water bath that provides the representative scatter. The can provide information about the distribution ofactivity within
energy-response function simulates the detection process of the a source volume (i.e., the activity in each individual voxel may
scintillation camera; a Gaussian shape with a width depending be obtained). However, in practice, this may be difficult due to
on energy is usually assumed. the large uncertainty associated with quantitating radioactivity
In some spatial convolutions, the calculation begins with a within a very small volume (i.e., the voxel) due to finite system
first approximation to the corrected image (74,86). This esti spatial resolution and partial volume effects (42,99). To mini
mate would ideally be improved upon through an iterative mize this problem, the average activity within a collection of
process. However, only one iteration is actually performed to subvolumes, each containing an appropriate number of voxels,
avoid noise increments and potential instabilities. may have to be integrated (104). A number of investigators
Energy-weighted acquisition is a technique available through have determined the spatial distribution of activity within a
an add-on to a commercial gamma camera system (87,88). It source region at the voxel level (104—108) and used these data
requires the knowledge of predefmed weights, which are to generate patient-specific absorbed dose estimates. In addi
derived from desired criteria for the resulting images. These tion, image fusion (i.e., registration of multimodality images),
criteria are improved resolution without increase in noise and has also been shown to be important for patient-specific
stability against changes in energy gain from point to point over dosimetryin radionuclidetherapy (104,105,107,108). Bolch et
the camera face (39). Spectral-fitting methods (82) are also al. (4), in the last issue of The Journal of Nuclear Medicine,
now available on a commercial system (89). Last, computer provide a more comprehensive treatment ofnonuniform activity
modeling (90) is an elegant technique but potentially requires and absorbed dose distributions.
an accurate Monte Carlo simulation of photon transport for the
particular patient of interest (91).
C.PET Techniques
Simulation studies in heart perfusion, especially in cases of Quantitative PET imaging is generally thought to be more
large liver activity, have led to the conclusion that both accurate than SPECT, largely because of the commercial
attenuation and scatter correction are needed (92,93). However, availability of a convenient attenuation correction (109). Stan
because of the complexity of the problem and due to the dard PET corrections are made for detector nonuniformity,
multiplicity of proposed methods, no one single method of deadtime, random coincidences and attenuation; otherwise, the
scatter correction has been widely accepted, although scattered procedure is similar to that for SPECT. PET is sometimes used
photons may account for as much as 50% ofthe recorded counts for measuring activity of a positron emitter to simulate another
in a SPECT image. radionuclide of the same atomic number (not a positron emit
ter), for which activity measurements and absorbed dose esti
4. Activity Determination
mates are desired. The assumption is made that isotopes have
Reasonably accurate and precise quantitative SPECT imag the same biokinetic behavior. For example, measurements have
ing is clinically feasible, even without sophisticated scatter been made using an ‘241-labeled radiopharmaceutical as a
corrections, at least in uniformly attenuating parts of the body surrogate for the ‘31I-labeledradiopharmaceutical (110). Un
such as the abdomen and pelvis (42). With Chang's first order like most positron emitters, 1241has a sufficiently long half-life
postprocessing method of attenuation compensation (55), the (4.2 days) to permit imaging over several days during the
most commonly used approach in commercial SPECT systems, biologic uptake and washout of the agent. Therefore, when an
investigators have obtained fairly accurate activity estimates analogous positron-emitting isotope (e.g., as a surrogate for a
using a semi-automated processing method employing thresh desirable therapeutic radionuclide) that has a physical half-life
olding (94,95). The attenuation-corrected reconstructed slices long enough relative to the pharmacokinetics is available, PET
are used to obtain source region activity based on a fixed imaging may increase the accuracy of activity measurements.
threshold method. Briefly, all the reconstructed slices spanning
the object of interest typically are analyzed to determine the V. QUANTITATIVEMEASUREMENTTECHNIQUES:NON
maximum voxel count. Only voxels containing more than a IMAGING
predetermined percentage of this maximum count, as derived A. External Nonimaging Radiation Monitoring
from phantom studies, are included for the activity calculation. Whole-body retention can be estimated by placing a shielded
The activity is determined from the sum of the counts in the Nal detector or other suitable survey meter 3—7m from the
included voxels divided by the acquisition time and an mdc patient and recording the number of counts detected (111 ). A
pendently acquired count rate per unit activity system calibra source ofknown activity is used to calibrate the detector. Initial
tion factor. In addition to threshold methods (96,97), image measurements for the total body must be made within a short
segmentation has also been based on edge detection (98—101). time after administration, defined as within the “nopatient
It should be pointed out that any technique that does not excretion―period, so that later measurements may be easily and
specifically compensate for scatter will not be rigorously more accurately compared to initial body content of the admin
correct, especially in the presence of low-energy photons, istered activity and to the reference standard. The use of a
which are known to contribute a significant amount of scatter. reference standard measured at each time corrects for radioac
MIRD PAMPHLET
No. 16 •
Siegel et al. 49S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
where A@(O)is the initial activity value of the jth exponential would probably be divided into several compartments, includ
component, A is the physical decay constant corresponding to ing a gallbladder compartment that receives activity from one or
the physical half-life I of the radionuclide through the relation more of the intrahepatic compartments. The gut, which subdi
ship A = 0.693/T and A@is the biologic elimination constant vides anatomically into a number of compartments, would also
corresponding to the biologic half-time T@(A@ = O.693/T@)of the be important in a liver agent model because it receives activity
jth exponential component. Summing all the nuclear transitions from both the liver and gallbladder. But the gut may also
during the time interval is equivalent to integrating Ah(t) over receive activity directly from the blood through the mesenteric
that time interval. The limits of the integral are usually taken as circulation. If the tracer is partially or completely reabsorbed,
time zero to infinity: the gut sends activity back to the liver and, perhaps, to the
blood. In the study of a liver tracer, the kidney would probably
be treated as a single compartment.
@ Ah j' Ah(t)dt= A@/(A
+ A@)
= 1.443 Aj(Tj)e, Eq. 25 A compartmental model, for dosimetry purposes, describes
the kinetics of the tracer in question as transfers among the
compartments of interest. These compartments may represent
where (Tj)e 0.693/(A + Ai). (T@)@is the effective half-time of
individual organs, in which case the normal assumptions about
the jth exponential component, where the combination of
a compartment, e.g., being a well-mixed space, are not strictly
physical decay and biologic removal results in an exponential
valid. Usually, however, a reasonable representation of the
disappearance rate that is the sum of the physical and biologic
organ's kinetics may be obtained through the analysis and an
decay rates. The quantity A@is negative for uptake components.
accurate residence time may be determined. Alternatively,
Analytically, the A@and A@values can be determined by
compartments may represent pools or spaces in the body, and
fitting the Ah(t) data to Equation 24 by a least squares technique
the residence time obtained must later be apportioned to
(see Example 4 in Appendix E). In some cases, adequate results
individual regions to estimate the radiation dose. In dosimetry
can be obtained by plotting Ah(t) on semilogarithmic graph
studies, as many as possible of these compartments are mea
paper, drawing the best-fit line through the data corresponding
sured directly. All available data are used and known physiol
to the slowest component (which will be a straight line),
ogy furnishes the configuration of the model. Some of the
subtracting its values from the original data, and repeating the
parameters of the model may be taken from the literature,
process to generate the A@and A@values.
especially when observed data are not sufficient to identify the
Alternatively, Ah may be calculated as the area under the
model completely. The more compartments for which valid
curve in an ordinary plot of Ah(t), by use of any of several
data exist to test the model, the more confidence one can place
numerical methods, such as Simpson's rule or the trapezoidal
on the model projections ofthe time—activityfunctions for those
rule. Of these,the simplestto apply is the trapezoidalrule compartments without direct observational data.
(122):
Most models assume linear (first-order) kinetics between
j Ah(t)
dt
=[w1yo
(‘b
a
+(w1
+w2)y1
+(w2
+w3)y2
+... linked compartments. In other words, transfer out ofa compart
ment per unit oftime is a constant fraction ofthe content of that
compartment. Therefore, the time-dependent activity of any
compartment can be described by a differential equation, and
+w@@y,@@J/2,
Eq.26 the entire system is described by a series of coupled differential
equations. Even when a transfer of tracee is nonlinear, the use
where w's are the widths and y's are the heights of the m
of tracers linearizes the function. When the steady state is
rectangles. An example of the use of the trapezoidal rule can be
perturbed during a study (e.g., a large dose of iodine is given
found in Example 4 of Appendix E.
during a radioiodine study or cholecystokinen is given during a
These curve processing methods can be applied to any source
gallbladder study), this perturbation may be modeled by adjust
region time—activity data. The resulting cumulated activities or
ing one or more of the model parameters. Radioactive decay is
residence times can then be used to estimate absorbed radiation
incorporated into the physiological model as an exit rate
dose using the MIRD schema (1,2). However, obtaining the
constant from each compartment.
residence times for the remainder ofbody, excretory organs and
Compartmental models are described by a set of differential
regions that cannot be measured directly is often difficult. Data
equations, the parameters of which are the sizes of the com
sets for whole body and cumulative or differential excretion and
partments and the rate coefficients for transfer between com
other regionsmay require interpretationand manipulation.
partments. Unless the model is quite simple (four compartments
B. Compartmental Modeling or less), the only practical way to solve these models is by
An alternative approach to the direct calculation ofA and r is numerical methods. Analytic solutions for small models which
to fit a compartmental model to the measured data. Often, it is have unique solution sets are known, and software is available
impractical to measure the time—activity curves of all of the for generating the rate constants from the input data points
source regions. When the physiological interactions of these (126, 127). Because the transfer rate coefficients in these mod
regions with the blood or with other directly measurable tissues els are assumed to be linear (i.e., they do not vary with time),
are known, the time—activity curves of unmeasured tissues can extrapolation ofthe results beyond the end ofthe measured data
be inferred by compartmental modeling. This technique also to infinite time, using these coefficients, is often thought to be
makes it feasible to separate the time—activitycurves of over a reasonable approach, assuming the model accurately repre
lapping regions from each other. A compartment in such a sents reality (which may not be the case for short-term bone
model is a unit of the body which is assumed to act kinetically models, for example; see Appendix D). A variety of programs
as a distinct, homogeneous, well-mixed unit (125). For exam are available for solving models and can be used on desktop
pie, if a kidney agent is studied, the renal cortex and renal computers (128). The most useful programs have utilities for
medulla might be treated as separatemeasurable compartments, optimizing the fit of time—activity curves of the model solution
whereas the liver would probably be a single compartment. On to the observed data through the use of iterative, nonlinear least
the other hand, if the agent studied is a hepatic agent, the liver squares routines.
MIRD PAMPHLET
No. 16 . Siegel et al. 51S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
TABLE Al
Table of Symbols and Definitions
Section
(units)II Symbol Definition
Effectivelinearattenuationcoefficient(cm1).
@ Eq.3 IL. (lit) @.t,t,
= @.Lj
+ (lit) (@ —
i=@ 1=1
IVA2.a.i. e@ Transmission factor(@J)
acrossthe patientthicknesst.
t Patientthicknessthroughthe regionof interest(cm).
Effectivelinearattenuationcoefficientacrossthe patientthickness(cm1).
I/to Ratioof countrateswith(I)andwfthout(Ia)the patientin posftion.
@ C Systemcalibrationfactor(countrateperunitactivity)(Bq@
@Ci1 . min1).
IVA2.a.ii.
fi@f2 ActMtyin sourcevolume2 (Bqor MCi).
Eq.4 A2=
52S THEJOURNAL
OFNUCLEAR
MEDICINE
•
Vol. 40 •
No. 2 (Suppl)•
February1999
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
TABLE Al
Continued
DDBF
IVA2.c.ii. Acronymfor depth-dependentbuildupfactor in Eqs.11 and 12 for ‘A
and Ii,.
Depth-dependentbuildup factor.
Eq.13 B(d) = Cd
d Depthat the centerof thesourceOesion)
(cm).
Cd Countratemeasuredat depthd (sec@1 or min1).
C@ Count rate measuredin air at same source to gamma camera distance
(@_1 or min1).
linear attenuationcoefficient(cm@).
OVendpatientthickness(cm).
Sourcethickness(cm).
A1= COcJ/C Activity (buildupfactor method)(Bq or pci).
@ C System calibrationfactor (count rate per unit actMty) (Bq@ . or
@tCi@
. min@).
DIBF Acronym for depth-independentbuildup factor.
Eq.14 T=1—(1—e@)@ Transmission(DIBFmethod).
B(cxD) Buildupfactor at infinitedepth. (Equivalentto the pseudo-extrapolation
number,n.
IVA2.c.iii. CR Acronym for the channelratio method.
TEW Acronym for the triple energywindow tachnk@ue.
@ Eq.18 Corrected(true)photopeakcounts.
Total count recordedwithin the photopeakwindow.
C@,C@ Counts within lower and upper scatter windows.
ki, k2, k3 Scatter multipliers.
IVA2.civ. EWA Acronym for energyweightedacquisition.
IVA3.
Eq.20 A = IAe@F/C Activityseenonlyon oneview(versustheconjugatemethod).
‘A Count rate within the image AOl.
C Systemcalibrationfactor(countrateperunitactMty).
F Correctionfactor for backgroundactivity counted within the ROl.
IV.B.2.
Correctionfactor(SPECTimaging).
Eq.21 CF=
Transmissionat angle i from the pbel through the body to the skin surface.
N Total number of projections.
Eq.22 I'=I@e_@ Gamma ray intensityleavinga voxel.
l@; Gamma ray intensityenteringthe voxel.
Averagelinearattenuationcoefficientwithin the voxel (cm@).
e Gammaraypathlengththroughthevoxelat the projectionangleof interest.
IV.B.3. k Scattermultiplierdefinedastheratioof thescatteredcountsin the
photopeakwindow to those in the lower window.
Eq.23 CT=C@,—kCs Truephotopeakcounts.
CPp Total count recordedwithin the photopeakwindow.
C@ Count within the scatter window.
V.B. @.tCVml Unitsfor blood actMtyconcentration.
VIA Ah(t) Activity in organ h is a function of time t.
Time.
@ Eq.24 Ah(t) = Aj(0)e_@Ai)t
MIRD PAMPHLET
No. 16 •
Siegel et al. 53S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
TABLE Al
Continued
(units)Appendb(
Section SymbolDefinition
patient.‘0Count
C.1. ICount ratewithtransmissionsourceviewedthroughthe
air.‘ptCount rate with the transmissionsource in
(counts/time).!i@i@ rate obtelned with patient alone
Eq.C1
f
region.FBackground
A=F@—@-@Absolute
activity A (MBq or @i)
in each
6).fCorrection correctionfactor(Eq.
2).Appendb for source region attenuationand source thickness (Eq.
D.1. A1,A@Component coefficientfor source regionat t = 0 (fraCtionalactMty,
unitless).TeEffective
hr).ThResidence half-time(sec or
hr).Ainet(t)ActMty time in source regionh (sec or
uptake.‘1instResidence associatedwith instantaneous
hr).kFractional time assuminginstantaneousuptake (secor
calculations.tmaxTime error in residencetime
maximum.Tel at whichtime-aCtMtyequationreachesa
washout.T@
0.693/(A + A1)Eff@ctiV@ half-timefor uptakeor rapid
washout.Appendb(= 0.693/(A+ A@)Effective half-timefor singlecomponent removalor slow
T@.Tk,ngResidence
0.2. A@(t)Activity associatedwiththe effecthiehalf-timeof the slowwashout,
A.@(t).Appendbc time associatedwith
Tei.T@Residence
0.3.A@(t)Activity associatedwiththe effectivehalf-timeof the rapidwashout,
A,@(t).Eq. time assodated with
D19 r = A1/(A1+ A@J@efined ratio of coefficients.
APPENDIX C: QUANTiTATIVE IMAGING PROTOCOLS imaging study. The count rate per unit activity (cpm/MBq)
(EXAMPLE FORMATS) represents the calibration factor. The collimator count-rate
1. Conjugate View Imaging—Planar Views response as a function of the source-to-collimator distance
must be known. For parallel-hole collimators, collimator
The steps involved in activity quantitation based on transmission
conjugate-view planar imaging are as follows: efficiency is invariant; however, for other collimators, such
as diverging, converging and pinhole collimators, the effi
1. Determine the transmission factor, @?J(=e_@t), through each ciency is dependent upon distance;
source ROI. Usually the transmission factor is measured 3. Acquire anterior and posterior views of each source ROI for
before administering the radiopharmaceutical to the patient, a fixed time (or for a fixed number of counts which must be
but it may also be determined after administration. To normalized to unit count time for each of the conjugate
measure the transmission factor, prepare a source of activity views);
whose surface area is greater than that of the source region 4. Determine the anterior ‘A and posterior I@conjugate-view
with the same radionuclide as that to be used for the patient count rates for the region to be quantitated through ROI
imaging study. As an example, for small regions fill the analysis. Choose an appropriate background correction tech
bottom ofa Petri dish (covered and sealed to prevent possible nique ifrequired as described in the Section IV.A.2.b. For the
contamination); for large regions, fill a flood source. Count simple background subtraction method, the adjacent ROIs are
the transmission source for a fixed time (e.g., 5 mm) through usually drawn contiguous to the region ROI and may be
the patient across the area of the specific body regions of normalized to the area of the organ region if a different size
interest and again in air at the same source-to-collimator ROI is used for ‘ADJ
versus ‘A
and I@.Care should be taken
distance. Another method for acquiring transmission data is to avoid high and/or low activity areas for selection of the
through a whole-body transmission scan which may be background ROI; and
performed with a line or flood source. The transmission 5. Determine the absolute activity, A (MBq), in each region. For
factor is calculated as @J= I/Ia, where I is the count rate example, for simple background subtraction, per Eqs. 1 and
(cpm) obtained with the transmission source viewed through 6:
the patient and I@ is the count rate obtained with the
transmission source in air. The count rates are obtained by
drawing appropriate ROIs encompassing the source region. A=F@j—@--X@@ Eq.Cl
The measurement of the transmission factor is thus based on
relative counts with and without the patient. If the transmis where F is the background correction factor and f provides the
sion factor is determined postadministration, a separate mea correction for the source region attenuation and source thickness
surement of activity in the region at that time must be (see Eqs. 2 and 6 or 7). In most cases, f will be approximately
@ obtained. In this case, @1= (I —‘pt)/I@'
where is the count equal to unity. If the ROI is small and the variation in body
rate obtained with the patient alone; thickness within the ROI is also small, this equation can be used
2. Obtain the imaging system calibration factor, C, by preparing to calculate the activity for the entire ROI. On the other hand, if
a standard of known activity, usually 37—74MBq (1—2mCi), the ROI is large (e.g., encompasses the entire liver), a pixel-by
in a suitable container ofthe same radionuclide to be used for pixel calculation may have to be made.
patient administration. Count this standard in air for a fixed A conjugate-view measurement is not necessary at each of the
time (e.g., 5 mm) at a source-to-collimator distance that time points chosen for generating ROIs unless the distribution of
approximates that ofthe patient midline distance used for the the activity is time-dependent. (This occurs, for example, in the
MIRD PAMPHLET
No. 16 •
Siegel et al. 55S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
70
40% —____ 60 -@
E 50
(1) Cl)
30% —“@—-.--
I
0
@ tmax I .443kTe2 ln(k + 1)1(k). Eq. D7 @/(A2Te2) 1.443 [1 + AiTei/(A2Te2)1 Eq. D12
Figure D2 shows the variation of t@ (the time for initial and
sampling that constrains the error to <k%) as a function of
effective washout half-time (Te2) for four different values of Ti0@@/(A2T@2)
= 1.443. Eq. Dl3
percentage error (< 10%) in residence time estimates. To use this As before, the fractional error in the residence time calculations
figure, find the appropriate Te2, decide on the acceptable error, and is defined as the ratio of the difference between r and Tiongto T.
then find the y-intercept (t,,@) that satisfies these conditions. From
This result (Eq. Dl4) is obtained by dividing the difference
Figure D2, for an effective washout half-time of 50 hr, sampling is
between Equations Di2 and Di3 by Equation Di2:
necessary before 10 hr postadministration (t@@)to keep the error
below 5% as a result of neglecting the uptake phase. k = {A1Te1/(A2Te2)}/[l+ A1Te1/(A2Te2)1. Eq. Dl4
@: E
.@
Cl)
C
:D
I
@ I S • @1 25
29
SIow:RapidWashout Halftime Ratio Slow:RapidWashoutHalftimeRatio
—...-A1/A2=0.1 -@--A1/A2=0.2--@@---A1/A2=0.3--.--A1/A2=0.5 —@--r=0.1—*--r=0.2--@'--r=0.3--.--r=0.5
FIGURED3. Error in residence time calculationwhen the rapid washout FIGURE D4. Potentlal errors in source region determination when the
phase is neglected. For source regions exhibiting biphasic clearance,the long-termretentionis neglected.The ratio of activityremovedfrom the
magnitudeof the residencetime underestimate(whenneglectingthe rapid source regionto the total uptake is r, where r = A@/(A@
+ A@j.(1 - r) is equal
washout component)are fUnctionSof the ratio of the slow to rapid washout to thefractionof theactivityretainedwitha longeffectiveremovalhalf-time.
It is clear from this figurethat not determiningthe long-termretention
effective half-times (F@/ro1)and the ratio of the fractional activity removed
rapidlyto that removedslowly(A,/k@J. Whenthe relativeamountof activity componentintroduceslargeerrors(underestimates) in the sourceregion
rapidlyremovedfromtheregionissmallcomparedtothe long-termretention residencetimecalculations.
component,theerrorsdueto neglectingtherapidwashoutareminor.Asthe
long-term retention component effective half-time becomes much larger where
thanthatoftherapidlyremovedcomponent,theerrorsareminorregardless
of how the activity was distributed betweenthe two washout components. r = A1/(A1 + A2) Eq. D19
The foregoingdoes not account for errors in residence time determination
dueto neglectingnoninstantaneous uptake. 1 —r = A2/(A1 + A2). Eq. D20
The fractional error in the residence time calculations, k, can be
Figure D3 shows how the percentage error resulting from defined, asbefore, asthe ratio of the difference betweenTand T@pid
neglect of the rapid washout component varies as a function of the to T. This result (Eq. Dl9) is obtained by dividing the difference
relative magnitude of the slow washout effective half-time com between Equations D17 and Di8 by Equation Dl7:
pared to the rapid washout effective half-time. Two general trends
should be noted. First, the larger the fraction of activity associated k = {[r + (1 —r)Te2/Tei] l}/[r + (1 —r)Te2/Tei].
Eq. D2l
with the initial rapid removal effective half-time, the larger the
percentage error. Also, the smaller the ratio of the slow washout Figure D4 shows how the percentage error from neglecting the
effective removal half-time, Te2, relative to T@1 , the larger the long-term retention of activity phase or the slow washout compo
percentage error. To illustrate, assume a radiopharmaceutical nent varies as a function of the relative magnitude of the slow
exhibits biexponential washout where 10 kBq is removed with an washout effective half-time compared to the rapid effective wash
effective half-time of2 hr and 50 kBq is removed with an effective out half-time. The percentage error is very sensitive to two
half-time of 10 hr. Therefore, A1/A2 0.2 and T@2/T@1= 5. variables, namely r and the ratio of /T1 . The smaller the
According to Figure D3, the error in neglecting the rapid washout fraction of activity associated with the rapid effective half-time, r,
component is —4%. the larger the percentage error. Also, the larger the effective
half-time for slow washout, Te2, relative to , ‘the larger the
3. Assessment
ofErrors
forLong-Term
Retention
of percentage error. For example, assume the radiolabeled compound
Activity
is 30% removed with an effective half-time of 2 hr and is 70%
The actual activity retention function can again be described by removed with an effective half-time of 10 hr. Then r = 0.3 and
Equation D8 and the associated residence time by Equation D9. In
Te2ITei = 5. According to Figure D4, the error in neglecting the
this case assume that samples are not obtained at later times and
long-term washout component is —45%.
only data that describe the early or rapid washout from the tissue
are acquired. The resulting derived retention function is given by APPENDIX E: CALCULATIONAL EXAMPLES
Equation D15. Note that the estimated initial activity is equal to the
I. Instantaneous Uptake: No Biologic Removal
sum of the individual activity fractions for both the rapid and slow
Problem. Consider 10 mCi of a I‘C-labeledcompound that is
washout components:
taken up rapidly and distributes uniformly in the whole body (WB).
A,@@Id(t)
= (A1 + A2)e_@@k1@. Eq. Di5 Assume that no biologic removal occurs. The physical half-life of
1‘Cis 0.34 hr. Calculate the cumulated activity and residence time
The associated residence time is given by: in the whole body.
E Dl6 Solution. The activity in the whole body is removed exponen
Trapid l.443(A1 + A2)Tei.
q. tially with a half-time equal to the physical half-life T. From Eq.
Eqs. D9 and Dl6 can be rewritten as follows: 24, one obtains:
and
where AwB(t) is the activity in the whole body at time t, A is the
Tr@jd{Tei(Ai + A2)} = 1.443, Eq. Dl8 physical decay constant, A@is the biologic disappearance constant,
MIRD PAMPHLET
No. 16 •
Siegel et al. 57S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
TABLE El TABLE E2
Cumulated ACtIVityand ReskienceTime in the Stomach Cumulated Activity in the Kidneys
.513,54667.72.06,95734.82.53,57319.93.01,8359.2
,790/186,058232280141
,64722944094,600/108,790150,000/172,50018258073,69cv84,744125,050
,630/162,874166,650/191
AWB f AWB(t)
dt = A0f e0693@dt = l.443A@T a Noninstantaneous
Uptake:
Removal
ByBoth
Physical
Jo Jo Decay and Biologic Elimination
Problem. A patient is administered 37 MBq (1 mCi) of 99mTc@
= I .443(1 0,000 @Ci)(0.34 hr) = 4,900 @Ci . hr. MAG3. Conjugate planar gamma camera images ofthe kidneys are
acquired for 20 sec at various times post administration. The counts
The residence time is: in the anterior (A) and posterior (P) views of the left and right
TWB AwB/@ —(4,900 @Ci. hr)/(lO,000 MCi) = 0.49 hr.
kidney (K) ROIs are given in Table E2. The physical half-life of
@“Tc
is 6 hr, the transmission factor through the kidneys is 0.75,
2. Instantaneous
Uptake
Removal
byBothPhysical
Decay and the imaging system calibration factor is 5000 cpm/@Ci.
and Biologic Elimination Assume that the kidney correction factor, f, is equal to unity.
Problem. To measure the gastric emptying rate for liquids, a Calculate the cumulated activity in the kidneys (Table E2).
patient is instructed to drink 300 ml ofwater containing 500 @Ci
of Solution. The total activity in the two kidneys can be calculated
I J3mInDTPA Planar images of the stomach (5) are obtained at 30 using Eq. 1 and is given in Table E2. If these time—activitydata
minute intervals for three hours. The net counts per minute (cpm) were plotted on graph paper it would be observed that the retention
in the stomach region after subtracting background, if present, are ofactivity from the kidneys can be described by a three-component
@ given in Table E1. The physical half-life of I3mIn is 1.658 hr. exponential function with a single term for the uptake phase and
Calculate the cumulated activity and residence time in the stomach two terms for the clearance phase. Therefore, the kidney activity
(Table El). can be described by:
Solution. Because all of the activity is in the stomach at t 0
(instantaneousuptake),the activity in the stomachat this time is AK(t) = @:Aj(0)e_+@t = @: Aj(0)e_@@t.
A0, where A0 is the administered activity. At all other times, the J
58S THEJOURNAL
OFNUCLEAR
MEDICINE
•
Vol. 40 •
No. 2 (Suppl)•
February1999
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
TABLE E3 function, which can then be integrated analytically. For the
Whole-Body Retention at Vatious Times example above, curve stripping gives a biexponential function, and
a least squares fit will yield the following equation:
Time(hr)Activity
(j@C@01000.5721352244206151012
A(t) = l8.6e°°3@
+ 81.4e'231.
Note that the sum of the coefficients of the two exponential terms
is equal to the administered activity of 100 pCi. By integrating the
least squares equation to time t 10 hr and comparing the result
with the estimate generated by the trapezoidal method, a reasonable
agreement is achieved:
A = 18.6/0.039
x [1—
e°°39@°]
+ 81.4/1.23x [1 —
e@'23>@0]
Solution. To determine A, we integrate the time—activity curve
shown in Figure El. = 220 @Ci . hr.
MIRD PAMPHLET
No. 16 •
Siegel et al. 59S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
17. Kojima M, Takaki Y, Matsumoto M, et al. A preliminary phantom study on a 54. Koral KY, Zasadny KR, Kessler ML, et al. A method using CT-SPECT fusion plus
proposed model for quantification of renal planar scintigraphy. Med Phys l993;20: conjugate views for dosimetry in ‘31I-MoAb
therapy of lymphoma patients. J Nucl
33—37. Med 1994;35:l7l4—l720.
18. Rohe RC, Thomas SR. Stabin MG, et al. Biokinetics and dosimetry analysis in 55. Chang LT. A method attenuation correction in radionuclide computed tomography.
healthy volunteers for a two-injection (rest-stress) protocol of the myocardial IEEE Trans Nucl Sci 1978;N5-25:638—642.
perfusion imaging agent 9@―Tc-Q3.J Nucl Cardiol l995;2:395—404. 56. Frey EC, Tsui BM, Perry JR. Simultaneous acquisition of emission and transmission
19. International Commission on Radiation Protection. Report of the task group on data for improved Tl-20l cardiac SPECT imaging using a Tc-99m transmission
reftrence man, ICRP Pubi. No. 23. New York: Pergamon Press; 1975:103—140. source. J Nuci Med l992;33:2238—2245.
20. Thomas SR. Gelfand MI, Burns GS, et al. Radiation absorbed-dose estimates for the 57. Tan P. Bailey DL, Meikle SR. Eberl 5, Fulton RR. Hutton BF. A scanning line source
liver, spleen, and metaphyseal growth complexes in children undergoing gallium-67 for simultaneous emission and transmission measurements in SPECT. J Nucl Med
citrate scanning. Radiology l983;146:8l7—820. l993;34: 1752—1760.
21. Wu RK, Siegel JA. Absolute quantitation of radioactivity using the buildup factor. 58. Manglos SH, Bassano DA, Thomas FD, Grossman ZD. hnaging ofthe human torso
MedPhys 1984;ll:l89—l92. using cone-beam transmission CT implemented on a rotating gamma camera. J NucI
22. Siegel JA, Wu RK, Maurer AH. The buildup factor: effect of scatter on absolute Med l992;33:l50—156.
volume determination. J Nucl Med 26:390—394, 1985. 59. Jaszczak Ri, Gilland DR. Hanson MW, Jang S. Greer KL, Coleman RE. Fast
23. Siegel IA, The effect of source size on the buildup factor calculation of absolute transmission CT for detennining attenuation maps using collimated line source,
volume.J NuclMed 26:1319—1322; 1985. rotatable air-copper-lead attenuators and fan-beam collimation. JNuclMed 1993;34:
24. Van Rensburg AJ, Loner MG, Heyns AP, et al. An evaluation of four methods of I577—1586.
@ I ‘Inplanar image quantitation. Med Pkvs l988;l5:853—86I. 60. Tung C-H, Gullberg GL, Zeng GL, Christian PE, Data FL, Morgan HT. Non-uniform
25. Jaszczak Ri, Floyd CE, Coleman RE. Scatter compensation techniques for SPECT. attenuation correction using simultaneous transmission and emission converging
IEEE Trans Nucl Sci l985;32:786—793. tomography. IEEE Trans NucI Sci l992;39:1 134—1143.
26. Koral KF, Clinthome NH, Rogers WL. Improving emission-computed tomography 61. Ficaro EP, Fessler IA, Rogers WL, Schwaiger M. Comparison of Am-241 and
quantification by Compton-scatter rejection through offset windows. Nucl Inst Tc-99m as transmission source for the attenuation correction of T1-20l SPECT
Methods Phys Res l986;A242:610—6l4. imaging ofthe heart. J NucI Med l994;35:652—663.
27. Pretorius PH, Van Rensburg AJ, Van Assswegen, et al. The channel ratio method of 62. Stewart R, Ponto R, Dickinson C, Meakem L, Chava R, Juni I. In-vivo validation of
scatter correction for radionuclide image quantitation. J NucI Med 1993;34:330—335. simultaneous transmission-emission protocol (STEP) for Tc-99m-sestamibi SPECT
28. King MA, Hademenon GJ, Glick Si. A dual-photopeak window method for scatter quantitative comparison with N-l3-ammonia PET [Abstractj. J Am Coll Cardiol
correction. J NucI Med 1992;33:605—612. 1995;25:2 I7A.
29. Ogawa K, Harata Y, Ichihara T, Kubo A, Hashimoto S. A practical method for 63. Miron 5, Conant R, Sodee D, Amini S. Clinical evaluation of simultaneous
position-dependent Compton-scatter correction in single emission CT. IEEE Trans transmission SPECT myocardial perfusion images (STEP) [Abstract]. J NucI Med
Medimag 199l;lO:408—4l2. l995;36: l2P.
30. Macey DA, Grant EF, Bayouth JE, et al. Improved conjugate view quantitation of 64. Nickoloff EL, Perman WH, Esser PD, Bashist B, Alderson P0. Left ventricular
I-I 3 1 by subtraction of scatter and septal penetration events with a triple energy volume: physical basis for attenuation corrections in radionuclide determinations.
window method. Med Phvs 1995;22:l637—l643. Radiologyl98l;l52:5l1—5l5.
31. King MA, Doherty PW, Schwinger RB. A Wiener filter for nuclear medicine images. 65. Pelizzari CA, Chen GTY, Spelbring DR. Weichselbaum RR, Chen CT. Accurate
Med Phvs l983;lO:876—880. three-dimensional registration ofCT, PET, and/or MRI images ofthe brain. J Comput
32. King MA, Schwinger PW, Doherty PW, Penney BC. Two-dimensional filtering of AssistTomogr 1989;14:20—26.
SPECT images using the Mets and Wiener filters. J Nucl Med 1984;25:l234—l240. 66. Weber DA, Ivanovic M. Correlative image registration. Semin Nucl Med l994;24:
33. Penney BC, Glick Si, King MA. Relative importance ofthe error success in Wiener
311—323.
restoration of scintigrams. IEEE Trans Med imag 1990;9:60—70.
67. Parsai El, Ayyangar KM. Dobelbower RR, Siegel JA. Clinical fusion of three
34. Clarke LP, Cullom Si, Shaw R, et al. Bremsstrahlung imaging using the gamma
dimensional images using bremsstrahlung SPECT and CT. J Nucl Med l997;38:3l9—
camera: factors affecting attenuation. J Nucl Med I992;33:l6l—l66.
324.
35. Shen 5, DeNardo GL, DeNardo S. Quantitative bremsstrahlung imaging of yttri
68. Manglos Si, Jaszczak Ri, Floyd CE. Weighted backprojection implemented with a
um-90 using a Wiener filter. Med Pkvs l994;2l :1409
nonuniform attenuation map for improved SPECT quantitation. IEEE Trans NucI Sd
36. Halama, iR, Henkin RE, Friend LE. Gamma camera radionuclide images: improved
I988;NS-35:625—628.
contrast with energy-weighted acquisition. Radiology l988;169:533—538.
69. Tsui BMW, Gullberg GT, Edgerton ER, et al. Correction of nonuniform attenuation
37. DeVito RP, Hamill ii, Treffert I, Staub EW. Energy-weighted acquisition of
in cardiac SPECT imaging. J Nucl Med l989;30:497—507.
scintigraphic images using finite spatial filters. J Nucl Med l989;30:2029—2035.
70. Fessler JA, Hero AO. Space-alternating generalized expectation-maximization algo
38. Hamill, ii, DeVito RP. Scatter reduction with energy-weighted acquisition. IEEE
rithm. IEEE Trans Sig Proc 1994;42:2664—2677.
Trans Nuci Sci l989;NS-36:l334—l339.
71. Ficaro EP, Fessler IA, Shreve PD, Kritzman iN, Rose PA, Corbett JR. Simultaneous
39. DeVito RP, Hamill ii. Determination of weighting functions for energy-weighted
transmission/emission myocardial perfusion tomography. Diagnostic accuracy of
acquisition. J Nuci Med l991;32:343—349.
attenuation-corrected 9@―Tc-sestamibisingle-photon emission computed tomogra
40. Breitz HB, Fisher DR. Weiden PL, et al. Dosimetry of rhenium-l86-labeled
monoclonal antibodies: methods, prediction from technetium-99m-labeled antibodies
phy. Circulation 1996;93:463—473.
72. Jaszczak RI, Greer KL, Floyd CE, et al. Improved SPECT quantification using
and results of Phase I trials, J Nucl Med l993;34:908—9l7.
41 . Rosenthal MS. Cullom I, Hawkins W, Moore SC, Tsui BMW, Yester M. Quantitative compensation for scattered photons. J Nuci Med 1984;25:893—900.
SPECT imaging: a review and recommendations by the Focus Committee of the 73. Gilardi MC, Bettinardi V. Todd-Pokropek A, et al. Assessment and comparison of
Society of Nuclear Medicine Computer and Instrumentation Council. J NucI Med three scatter correction techniques in single photon emission computed tomography.
l995;36: 1489—1513. J NuciMed l988;29:l97l—l979.
42. Zanzonico PB, Bigler RE, Sgouros G, Strauss A. Quantitative SPECT in radiation 74. Axelsson B, Maski P, Israelsson A. Subtraction of Compton-scattered photons in
dosimetry. Semin Nucl Med 1989;19:47—61. single photon emission computed tomography. J Nucl Med l984;25:490—494.
43. Tsui BMW, Zhao X, Frey EC, McCartney WH. Quantitative single-photon computed 75. Frey EC, Tsui BMW. A practical method for incorporating scatter in a projector
tomography: basics and clinical considerations. Semin Nucl Med l994;24:38—65. back-projector for accurate scatter compensation in SPECT. IEEE Trans Nuci Set
44. Tsui BMW, Frey EC, Zhao X, Lalush DS, iohnson RE, McCartney WH. The 1993;NS-40:l 107—1116.
importance and implementation of accurate 3D compensation methods for quantita 76. Smith MF, iaszczak RI. Generalized dual-energy-window scatter compensation in
tive SPECT. Phys Med Biol l994;39:509—530. spatially varying media for SPECT. Phys Med Biol 1994;39:531—546.
45. Schold SC ir, Zalutsky MR. Coleman RE, et al. Distribution and dosimetry of 1-123 77. Liang z. Compensation for attenuation, scatter, and detector response in SPECT
labeled monoclonal antibody 8lC6 in patients with anaplastic glioma. Invest Radiol reconstruction via iterative FBP methods. Phys Med Biol l993;20:l097—l 106.
I993;28:488 —496. 78. Ljungberg M, Msaki P. Strand SE. Comparison of dual-window and convolution
46. Parker JA. Quantitative SPECT: basic theoretical considerations. Semin NucI Med scatter correction techniques using the Monte Carlo method. Phys Med Biol
1989;19:3—12. l990;35:I099—l 110.
47. Budinger TF. Physical attributes of single photon tomography. J Nucl Med l980;2l: 79. Logan KW, McFarland WD. Single photon scatter compensation by photopeak
579—592. energy distribution analysis. IEEE Trans Nucl Sci 199I;38:l178—1l82.
48. Gullberg GT, Budinger TF. The use of filtering methods to compensate for constant 80. Koral KF, Swailem FM, Buchbinder 5, Clinthorne NH, Rogers WL, Tsui BMW.
attenuation in single photon emission computed tomography. IEEE Trans Biomed SPECTdual-energywindowComptoncorrection:scattermultiplierrequiredfor
Eng 198l;28:142—l57. quantifcation. J Nucl Med I990;3 1:893—900.
49. Jaszczak RI, Coleman RE, Whitehead FR. Physical factors affecting quantitative 81. Bayouth JE, Macey Di. Quantitative imaging ofholmium-l66 with an Anger camera.
measurements using camera-based single photon emission computed tomography. Phys Med Bioll994;39:265—297.
IEEE Trans Nuci Sci 198 l;28:69—80. 82. Koral KF, Wang X, Rogers WL, Clinthorne NH, Wang X. SPECT Compton
50. Tanaka E, Toyama H, Murayama H. Convolutional image reconstruction for scattering correction by analysis of energy spectra. J Nucl Med 1988;29:195—202.
quantitative single photon emission computed tomography. Phys Med Biol l984;29: 83. Wang X, Koral KF. A regularized deconvolution-fitting method for Compton-scatter
1489—I500. correction in SPECT. IEEE Trans Med lmag l992;1 1:351—360.
51 . Mas I, Ben-Younes R, Bidet R. Improvement of quantification in SPECT studies by 84. Floyd CE, Jaszczak RI, Greer KL, et al. Deconvolution of Compton scatter in
scatter and attenuation compensation. Eur J Nucl Med l989;l5:351—356. SPECT.J NuclMed l985;26:403—408.
52. King MA, Schwinger RB, Doherty PW, Penney BC. Variation of the count 85. Msaki P, Axelsson B, DaM CM, Larsson SA. Generalized scatter correction method
dependent Mets filter with imaging system modulation transfer function. Med Phys in SPECT using point scatter distribution functions. J Nucl Med 1987;28:1861—l869.
l986;25: 139—140. 86. LjungbergM,StrandS-E.Scatterandattenuationcorrectionin SPECTusingdensity
53. Gilland DR. Tsui BMW, Perry iR, et al. Optimum filter function for SPECT imaging. maps and Monte Carlo simulated scatter functions. J NucI Med 1990;3l:1560—1567.
J NucI Med l988;29:643—650. 87. Floyd J, Mann R, Shaw A. Changes in quantitative SPECT thallium-20l results
associated with the use of energy-weighted acquisition. J NucI Med 1991;32:805— 108. Kolbert KS, Sgouros G, Scott AM, et al. Implementation and evaluation of
807. patient-specific three-dimensional internal dosimetry. J NucI Med 1997;38:30l—308.
88. Staff RT, Gemmell HG, Sharp PF. Assessment of energy-weighted acquisition in 109. Hoffman El, Phelps ME. Positron emission tomography: principles and quantitation.
SPECT using ROC analysis. J Nucl Med 1995;36:2352—2355. In: Phelps ME, Mazziotta J, Schelbert H, eds. Positron emission tomography and
89. O'Connor MK, Caiti C, Christian TF, Gibbons Ri. Effects ofscatter correction on the autoradiography: principles and applications for the brain and heart. New York:
measurement of infarct size from SPECT cardiac phantom studies. J Nucl Med Raven Press; 1986:237—286.
1995;36:2080—2086. 110. Larson SM, Pentlow KS, Volkow ND, et al. PET scanning of iodine-l24-3F9 as an
90. Floyd CE, iaszczak Ri, Greer KL, Coleman RE. Inverse Monte Carlo as a unified approach to tumor dosimetry during treatment planning for radioimmunotherapy in a
reconstruction algorithm for ECT. J Nucl Med 1986;27:l577—1585. child with neuroblastoma. J NucI Med 1992;33:2020—2023.
91 . Koral KF. Editorial: correction for patient Compton-scattering—current status. I I 1. Leichner PT, Klein IL, Garrison LB. et al. Dosimetry of ‘31l-labeledanti-ferritin in
JNuclMed 199l;32:1291—I293. hepatoma: a model for radioimmunoglobulin dosimetry. Int J Radiat Oncol Phys
92. Frey E, Li I, Tsui B. The importance of combined scatter and attenuation compen l981;7:323—333.
sation in T1-20l cardiac SPECT [Abstract]. J Nuci Med 1995;36:60P. 112. Cohn SH. Whole body counters. In: Editors: Wagner HN, Szabo Z, Buchanan JW.
93. King MA, Xia W, DeVries DI, et al. A Monte Carlo investigation ofartifacts caused Principles ofnuclear medicine, 2nd ed. Philadelphia: W.B. Saunders; 1995:298 —305.
by liver uptake in single-photon emission computed tomography perfusion imaging 113. Lathrop KA, Harper PV, Charleston DB, Atkins FB, Mock BH. Acquisition of
with technetium 99m-labeled agents. J NucI Cardiol l996;3: 18—29. quantitative biologic data in humans for radiation absorbed dose estimates. In:
94. Iosilevsky G, Israel 0, Frenkel A, et al. A practical SPECT technique for quantitation Cloutier Ri, Coffey IL, Snyder WS, Watson EE, eds. Radiopharmaceutical dosimetry
ofdrug delivery to human tumors and organ absorbed radiation dose. Semin Nucl Med symposium. Proceedings ofa conference held at Oak Ridge, TN. April 26—29.1976.
l989;l9:33—46. Washington: U.S. Government Printing Office 1976:164—173.
95. Pretorius PH, van Aswegen Al, Herbst CP, Lotter MG. The effects of different 114. Sgouros G. Bone marrow dosimetry for radioimmunotherapy: theoretical consider
correction techniques on absolute volume determination with SPECT using a ations. J NucI Med l993;34:689—694.
threshold edge detection method. Med Phys 199I;I8:390—393. 115. Lee RE, Siegel IA, Horowitz JA, et al. The use of SPECT in monoclonal antibody
96. ErdiYE,wessclsBW,LoewMH,ErdiAK.Thresholdestimationin singlephoton imaging to determine marrow-to-blood activity concentration ratio [Abstract]. J Nuci
emission computed tomography and planar imaging for clinical radioimmunotherapy.
Med Technol1990;18:143.
Cancer Res 1995;55(suppl):5823s—5826s.
116. Fisher DR. Radiation dosimetry for radioimmunotherapy. Cancer Suppl 1994;73:
97. Pretorius PH, van Aswegen A, Latter MG, et al. Verification of a varying threshold
905—911.
edge detection SPECT technique for spleen volume: a comparison with computed
I 17. Cloutier Ri, Smith EA, Watson EE, et al. Dose to the fetus from radionuclides in the
tomography volumes. J Nucl Med 1993;34:963—967.
bladder. Health Phys l973;25:147—l6l.
98. Long DT, King MA, Penney BC. 2D vs. 3D edge detection as a basis for volume
118. Thomas SR, Stabin MG, Chen C, Samaratunga RC. MIRD Pamphlet No. 14 Revised:
quantitation in SPECT. In: Ortendahl DA, Liacer I, eds. Information processing in
a dynamic urinary bladder model for radiation dose calculations. J Nucl Med
medical imaging. New York: Wiley Liss; 1991:457—471.
99. King MA, Long DT, Brill AB. SPECT volume quantitation: influence of spatial I999;40:Forthcoming April.
resolution, source size and shape, and voxel size. Med Phys l991;l8:lOl6—l024. 119. Limits for intakes of radionuclides by workers. Annals of the ICRP, 2, ICRP
100. Long DT, King MA, Sheehan I. Comparative evaluation of image segmentation Publication 30 (Part 1). Oxford: Pergamon Press; l979;30—34.
method for volume quantitation in SPECT. Med Phys 1992;l9:483—490. 120. Stubbs lB. Results from a new mathematical model of gastrointestinal transit that
101. Glick SI, Penney BC, King MA, Byrne CL. Noniterative compensation for the incorporates age and gender-dependent physiological parameters. Radiat Protein
distance dependent detector response and photon attenuation in SPECT imaging. Dosim l992;4l :63—69.
IEEE Trans Med Imag 1994;l3:363—374. 121. Stubbs JB, Marshall BJ. Radiation dose estimates for the carbon-l4-labeled urea
102. Gilland DR, Jaszczak RI, Turkington TG, Greer KL, Coleman RE. Volume and breath test. J Nuci Med l993;34:82l—825.
activity quantitation with iodine-123 SPECT. J NucI Med 1994;35:I707—l713. 122. Bers L. Calculus. New York: Holt, Rineholt, and Winston, Inc; 1969:413—416.
103. Mortelmans L, Nuyts J, Van Pamel G, Van den Maegdenbergh V, Dc Roo M, Suetens 123. Macon N. Numerical analysis. New York: Wiley; 1963.
P. A new thresholding method for volume determination by SPECT. Eur J Nuci Med 124. Riggs DS. The mathematical approach to physiological problems. Cambridge, MA:
1986;12:284—290. The M.I.T. Press; 1976.
104. Sgouros G, Chiu 5, Pentlow KS, Ct al. Three-dimensional dosimetry for radioimmu 125. iacquez IA. Compartmental analysis in biology and medicine. Ann Arbor: The
notherapy treatment planning. J Nucl Med 1993;34:1595—l6Ol. University of Michigan Press; 1985.
105. Sgouros G, Barest G, Thekkumthala I, et al. Treatment planning for internal 126. Feldman HA. A numerical method for fitting compartmental models directly to tracer
radionuclide therapy: three-dimensional dosimetry for nonuniformly distributed data. Am J Physiol l977;233:Rl—R7.
radionuclides. J Nucl Med l990;3 1:I 884—1891. 127. D'Argenio Dz, Schumitzky A. A program package for simulation and parameter
106. Giap HB, Macey DJ, Podoloff DA. Development of a SPECT-based three-dimen estimation in pharmacokinetic systems. Comput Prog Biomed l979;9: I 15—134.
sional treatment planning system for radioimmunotherapy. J Nucl Med l995;36: 128. Karayanakis NM. Advanced system modelling and simulation with block diagram
1885—1894. languages. Boca Raton, FL: CRC Press; 1995.
107. Furhang EE, Chui C, Sgouros G. A Monte Carlo approach to patient-specific 129. Hays MT. Radiation dosimetry of radioiodinated thyroid hormones. J Nucl Med
dosimetry. Med Phys 1996;23:1523—l529. 1985;25: 1068—1074.
MIRD PAMPHLET
No. 16 â€S̃iegel et al. 61S
Downloaded from jnm.snmjournals.org by on October 22, 2020. For personal use only.
Information about reproducing figures, tables, or other portions of this article can be found online at:
http://jnm.snmjournals.org/site/misc/permission.xhtml