Professional Documents
Culture Documents
CEC205 Lab Manual Part1
CEC205 Lab Manual Part1
Laboratory Manual
Part-1
METHOD USED
Apparatus
The pH scale runs from 0.0 to 14, with 7.0 being neutral. Most of the
surface water pH ranges from 6.0 to 8.5. The pH can be measured either
by the indicator method or by the electrometric method using a pH
meter.
Calibration of pH meter
PROCEDURE
2. Once calibrated, the probe should be rinsed and dried with a soft
tissue paper before every sample analysis.
5. Rinse the probe when you finish the measurement and leave the
probe immersed in pH 7 buffer solution for storage.
Caution: Never leave the probe in dry condition, always dip in buffer
solution.
OBSERVATION AND CALCULATION
RESULT
Introduction:
Turbidity is the technical term referring to the cloudiness of a solution, and it
is a qualitative characteristic that is imparted by solid particles obstructing the
transmittance of light through a water sample. Turbidity often indicates the
presence of dispersed and suspended solids like clay, organic matter, silt,
algae, and other microorganisms. So, in short, Turbidity is an expression of the
optical property that causes light to be scattered and absorbed rather than
transmitted in straight lines through the sample.
Environmental significance:
When the turbid water in a small, transparent container such as a drinking
glass is held up to the light, an aesthetically displeasing opaqueness or milky
coloration is apparent. The colloidal material which exerts Turbidity provides
adsorption sites for chemicals and for biological organisms that may not be
harmful. They may be harmful or cause undesirable tastes and odors.
Disinfection of turbid water is difficult because of the adsorptive characteristics
of some colloids and because the solids may partially shield organisms from
disinfectant. In natural water bodies, Turbidity may impart a brown or other
colors to water and may interfere with light penetration and photosynthetic
reaction in streams and lakes. Turbidity increases the load on slow sand filters.
The filter may go out of operation if excess Turbidity exists. Knowledge of the
turbidity variation in raw water supplies is useful to determine whether a
supply requires special treatment by chemical coagulation and filtration before
it may be used for a public water supply. Turbidity measurements are used to
determine the effectiveness of treatment produced with different chemicals and
the dosages needed. Turbidity measurements help to gauge the amount of
chemicals needed from the day-to-day operation of water treatment works.
Measurement of Turbidity in settled water prior to filtration is useful in
controlling chemical dosages so as to prevent excessive loading of rapid sand
filters. Turbidity measurements of the filtered water are needed to check on
faulty filter operation. Turbidity measurements are useful to determine the
optimum dosage of coagulants to treat domestic and industrial wastewaters.
Turbidity determination is used to evaluate the performance of water treatment
Plants. Turbidity in water may be caused by a wide variety of suspended
matter, such as clay, silt, and finely divided organic and inorganic matter,
soluble colored organic compounds, and other organisms. Under flood
conditions, great amounts of topsoil are washed to receiving streams. As the
rivers pass through urban areas, the domestic and industrial wastewaters may
be added.
Guideline:
According to WHO standard, 5 NTU is suggested as the turbidity limit for
drinking water, while 1 NTU is recommended to achieve the adequate
disinfecting safety. According to Bangladesh Environment Conservation Rules
(1997), drinking Water standard for Turbidity is 10 NTU (Nephelometric
turbidity unit).
Principle:
Turbidity is based on the comparison of the intensity of light scattered by the
sample under defined conditions with the intensity of the light scattered by a
standard reference suspension under the same conditions. The Turbidity of the
sample is thus measured from the amount of light scattered by the sample
taking a reference with standard turbidity suspension. The higher the intensity
of scattered light, the higher is the Turbidity. Formazin polymer is used as the
primary standard reference suspension.
Because of the wide variety of materials that cause Turbidity in natural waters,
it has been necessary to use an arbitrary standard. The original standard
chosen was; 1 mg SiO2/l =1 unit of Turbidity. The Silica used had to meet
certain specifications as to particle size. The Jackson candle turbidimeter has
been replaced by more reliable, sensitive, and easier to use instruments that
depend upon the principle of nephelometry. As a standard reference material,
Silica has been replaced by Formazin polymer. The Formazin suspensions were
first calibrated against the Jackson candle turbidimeter. The standard
nephelometry procedure is now reported in nephelometric turbidity units
(NTU). Because of the basic principles difference for Jackson candle
turbidimeter and nephelometric methods, results from the two methods can
vary widely. To avoid any confusion this may cause, turbidity measurements by
the standard nephelometry procedure are now reported in nephelometric
turbidity units (NTU), and the other one is reported in Jackson candle
turbidimeter units (JTU).
40 NTU is about equivalent to 40 JTU.
The applicable range of this method is 0-40 nephelometric turbidity units
(NTU). Higher values may be obtained with the dilution of the sample.
Sample handling and preservation:
Water samples should be collected in plastic cans or glass bottles. All bottles
must be cleaned thoroughly and should be rinsed with Turbidity free water.
The volume collected should be sufficient to ensure a representative sample,
allow for replicate analysis (if required), and minimize waste disposal. No
chemical preservation is needed. Do not allow samples to freeze. The analysis
should begin as soon as possible after the collection. If storage is required,
samples maintained at 4°C may be held for up to 48 hours.
Precautions:
The following precautions should be observed while performing the experiment:
The presence of colored solutes causes a reduction in the measured
turbidity values. Precipitation of dissolved constituents (for example,
Fe) causes measured turbidity values to be high.
Light absorbing materials such as activated carbon in significant
concentrations can cause low readings.
The presence of floating debris and coarse sediments that settle out
rapidly will give low readings. Finely divided air bubbles can cause
high readings.
Apparatus:
1. Turbidity Meter
Chemicals
1. Formazin Polymer standards
Procedure:
1. The turbidity meter is required to be calibrated.
2. To the sample cells, add sample water up to the horizontal mark, wipe
gently with soft tissue and place it in the turbidity meter. Cover the
sample cell with the light shield.
3. Check for the reading in the turbidity meter. Wait until you get a stable
reading.
Result:
Aim: To determine the amount of sulphate present in the given sample by the
gravimetric method.
Principle:
The sulphate in water is precipitated as Barium Sulphate by adding Barium
Chloride in hydrochloric acid medium. The precipitated matter is filtered,
washed free of Chloride, ignited, and weighed as barium sulphate. The reaction
taking place is provided below.
Na2SO4+BaC1---------------->BaSO4+2NaCl
The hydrochloric acid is added first to remove any carbonate ions that might be
present they would also produce a white precipitate, giving a false-positive
result. The reaction of Barium Sulphate with carbon in the water is as follows.
BaSO4+2C--------------------- >BaS+2C0 2
BaS+2O------------------------- >BaSO 4
Apparatus Required:
Crucible, Oven, weighing balance, pipette, beaker, water bath,
desiccators.
Reagents Required:
Dilute HCL
Solid Ammonium Chloride
1O%Barium Chloride solution
Reagents Preparation:
Barium Chloride Solution:
Dissolve l0g of Barium Chloride in 100 ml of distilled water.
Procedure:
1. Pipette out 5Oml of the sample into a clean 250m1 beaker.
2. Add 10 ml of dilute HCL & 1gm of solid ammonium chloride.
3. Heat to boiling & add l0ml of 10% Barium Chloride solution drop by drop
with constant stirring. Continue boiling for another 2 to 3 minutes.
4. Allow the precipitate to settle and test for completion of precipitation by
adding a small amount of Barium Chloride solution through the sides of
the beaker.
5. If any turbidity is noticed, add a sufficient quantity of barium chloride to
precipitate all the sulphate.
6. Transfer-the contents to a sand bath & digest for half an hour to promote
granulation of the precipitate.
7. Filter through Waliman no 42 filter paper and wash with boiling water till
the filtrate runs free of chlorine.
8. Transfer the filter paper along with the precipitate to a weighed silica
crucible and dry it in an air oven.
9. Ignite over a low flame initially, taking care to ash the filter paper
completely, then ignite strongly over a rose head flame to constant
weight.
10. From the weight of Barium Sulphate obtained, calculate the
Sulphate content of the sample.
RESULT
SIGNIFICANCE OF TEST/PARAMETER
PROCEDURE
Steps
(2).
3b. Determination of Alkalinity of given water sample
SIGNIFICANCE OF TEST/PARAMETER
Alkalinity Relationships
Apparatus
Reagents
Phenolphthalein Indicator
PROCEDURE
VOLUME
S.NO
OF METHYL
PHENOLPHTHALEI
SAMPLE INITIAL ORANGE
N POINT (P)
POINT (M)
H2SO4*50*1000)/Vol. of sample.
50*1000)/Vol. of sample
Observation:
1. Titrant
2. Indicators: Eriochrome Black T Indicator
3. End Point: Wine Red to Blue
4b. Determination of Chloride of water sample.
Aim: To determine the chloride content of the given water sample by Mohr's
method
Theory: Chloride in the form of Chlorine ion is one of the major inorganic
anions. The salty taste produced by Chloride concentration is variable and
dependent on the chemical composition of water. The Chloride is higher in
wastewater than in raw water. Along the sea coastal, Chloride may be
present in high concentrations because of saltwater intrusion into the water
and sewage system. In this method, a slightly alkaline solution is used.
Potassium chromate can indicate the endpoint of Silver nitrate titration of
Chloride as quantitatively precipitated before red Silver chromate is formed.
As the silver nitrate solution is slowly added, a precipitate of silver chloride
forms as per the following reaction:
Ag+(aq) + Cl–(aq) → AgCl(s)
The endpoint of the titration occurs when all the chloride ions are
precipitated. Then additional silver ions react with the indicator's chromate
ions, potassium chromate, to form a red-brown precipitate of silver
chromate as per the following reaction.
2 Ag+(aq) + CrO 4 2– (aq) → Ag2 CrO4(s)
Environmental Significance:
Chlorides associated with sodium (Sodium Chloride) exert salty taste
when its concentration is more than 250 mg/L. These impact a salty
taste to water. Chlorides are generally limited to 250 mg/L in water
supplies intended for public water supply.
In many arrears of the world where water supplies are scarce, sources
containing as much as 2000 mg/L are used for domestic purposes
without adverse effects once the human system becomes adapted to
the water.
It can also corrode concrete. Magnesium chloride in water generates
hydrochloric acid after heating, which is highly corrosive and creates
boilers problem.
Chloride determinations in natural waters are useful in the selection
of water supplies for human use.
Chloride determination is used to determine the type of desalting
apparatus to be used.
Chloride determination is used to control pumping of ground water
from locations where seawater intrusion is a problem.
Procedure:
1. Take a well-mixed suitable volume of sample (25, 50, or 100 mL) in a
conical flask (V).
2. Add 1ml of potassium Chromate indicator. The color of the solution turns
to yellow color.
3. Titrate against Standard silver Nitrate solution till color change from
yellow to brick red.
4. Note the Burette volume of silver Nitrate added (A)
5. For better accuracy, titrate distilled water (blank) in the same manner.
Note the burette volume of silver Nitrate added for distilled water (B).
6. Calculate the volume of EDTA required by sample; C = (A-B)
Observation:
1. Titrant
2. Indicators: Potassium Chromate Solution
3. End Point: Yellow to Brick Red
Introduction:
Chemical coagulation is a treatment method widely used for the removal of
small-sized and colloidal impurities from water. Surface water generally
contains a wide variety of colloidal impurities that may cause the water to
appear turbid and may impart color to the water. Colloidal particles that
cause color and Turbidity are difficult to separate from water because they
will not settle by gravity and are so small that they pass through the most
common filtration media pores. To be removed, the individual colloids must
aggregate and grow in size to settle by gravity. Chemical agents are used to
promoting colloid aggregation by destroying the forces that stabilize colloidal
particles.
The process of destroying the stabilizing forces and causing aggregation of
colloids is referred to as chemical coagulation. Coagulation involves reducing
electrical forces of repulsion and promoting "chemical type" interaction
between colloids, which eventually results in settling of the colloids and
accomplishes removal of Turbidity and color. At higher coagulant doses,
"charge reversal" is possible, resulting in resuspension of the colloids. Hence
optimum coagulant doses are determined through laboratory model tests
where the water to be treated subjected to a range of coagulant doses, and
the removal efficiencies are observed. Many authors used the term
"coagulation" to describe the process by which the charge on particles is
destroyed, and the term "flocculation" to describe the aggregation of
particles into larger units. The chemical used for this purpose is called
coagulants. The most common coagulants used in water and wastewater
treatment are aluminium and ferric salts such as Alum, Ferric Chloride, and
Ferric Sulfate. The common metal salt Alum (Aluminium Sulphate) is a good
coagulant for water containing appreciable organic matter.
Theoretically, each mg/L of alum consumes approximately 0.50 mg/L (as
CaCO3) of Alkalinity for water with low Alkalinity. This may result in a
significant reduction in pH that may interfere with aluminum hydroxide
flocs' formation. If the Alkalinity is insufficient, coagulant aids such as lime
[Ca (OH)2], soda ash (Na2 CO3 ), activated Silica, and polyelectrolytes are
used to provide the necessary Alkalinity. Iron coagulants can be operated
over a wider pH range and generally effective in removing Turbidity and
watercolor. However, they are usually more costly.
𝐴𝑙 (𝑆𝑂 ) . 18𝐻 𝑂 + 3𝐶𝑎(𝐻𝐶𝑂) → 3CaS𝑂 + 2Al (OH) + 6C𝑂
Environmental Significance:
Besides efficient removal of Turbidity and color, coagulation with alum and
ferric Chloride or ferric sulfate is also widely used for the removal of heavy
metal ions (e.g., lead, arsenic) from water. In this process, heavy metal ions
are primarily removed by adsorption (and Subsequent precipitation) onto
coagulated flocs of metal (either aluminum or iron) hydroxides. Coagulation
with alum and ferric chloride/sulfate is successfully used for the removal of
arsenic from water
Hard water can cause various problems. The problems created by hard
waters are as follows;
They cause excessive soap consumption.
They cause skin irritation.
They cause lime accumulation in boilers, hot water pipes,
and heaters.
They cause discoloration in porcelain. Especially in homes,
the white color of the sinks and bathtubs are discolored.
They reduce the life of fabrics and cause them to wear out.
They cause problems in the canned food industry.
Hardness removal (chemical softening) is a process that removes all or part of
the hardness by adding various chemical substances into the water. The
processes used for water softening are chemical sedimentation and ion
exchange methods. Chemical sedimentation can be carried out in 3 ways.
1-Lime-soda process.
2-Caustic-soda process
3-Sodium phosphate process
Lime soda process: In the lime-soda process, hard water is treated with
lime (CaO or Ca (OH)2) firstly, after that with soda. In this process, the
hardness is removed by sedimentation as calcium carbonate or magnesium
hydroxide. Lime is added either as calcium hydroxide or calcium oxide, and
soda is added as sodium carbonate. The substances form hardness in water,
and the reactions of lime and soda can be written as follows.
The usual method for lime soda processing involves treatment with excessive
lime. The pH should be greater than 9 for sedimentation of the magnesium
hydroxide. Usually, pH is between 10-10.5. A small amount of alum or sodium
aluminate is added to ensure good sedimentation. If the water contains
excessive Alkalinity after the softening process is completed, there is a
possibility of calcium carbonate precipitation in the pipes. The softened water
is treated with H2SO4 or CO2 to remove the Calcium Carbonate from the
softened water. This reaction is called Recarbonization.
If it is requested to retain permanent hardness in water, soda should not be
used after lime dosing. This application is defined as a partial softening
operation.
MATERIALS USED
- Ca(OH)2 and Na2CO3
- NH4Cl
- Eriochrome Black T and EDTA
- Phenolphthalein and Methyl Orange
- H2SO4
- Beaker
- Jar test setup
PROCEDURE
1. Determine the pH of the raw water sample.
2. Place 1 liter of raw water in each of the three beakers of the laboratory
stirrer. Immerse blades and stir the raw water samples at about 100
rpm.
3. Add lime-soda solutions into each of the beaker to obtain the desired
concentrations in the raw water samples.
4. Let the samples mix at approx. 100 rpm for 1minute, then decrease the
speed to approx. 30 rpm. Allow the sample to mix for 10 minutes.
Observe any changes in the suspended matter in the sample.
5. At the end of the mixing period, turn off the stirrer, let the flocs settle (at
least 20 minutes). Determine the pH of each treated water sample.
6. Measure the hardness of the samples and find the minimum lime dose to
remove hardness.
RESULTS, CALCULATIONS, and DISCUSSION
-
The amount of lime and soda required to soften hard water is calculated.
Graphically associate the lime scale with total hardness and Alkalinity.