You are on page 1of 5

Physica Medica 30 (2014) 708e712

Contents lists available at ScienceDirect

Physica Medica
journal homepage: http://www.physicamedica.com

Technical notes

Analysis of the EBT3 Gafchromic film irradiated with 6 MV photons


and 6 MeV electrons using reflective mode scanners
Nicolas Farah a, b, Ziad Francis a, c, *, Marie Abboud a
a
Physics Department, Faculty of Science, Saint Joseph University, B.P. 11-514 Riad El Solh, Beirut 1107 2050, Lebanon
b
Department of Radiation-Oncology, Hôtel-Dieu de France Hospital, PO Box HDF 166830, Beirut, Lebanon
c
The Open University, Faculty of Science, Department of Physical Sciences, Walton Hall, MK7 6AA Milton Keynes, United Kingdom

a r t i c l e i n f o a b s t r a c t

Article history: We explore in our study the effects of electrons and X-rays irradiations on the newest version of the
Received 30 January 2014 Gafchromic EBT3 film. Experiments are performed using the Varian “TrueBeam 1.6” medical accelerator
Received in revised form delivering 6 MV X-ray photons and 6 MeV electron beams as desired. The main interest is to compare the
4 April 2014
responses of EBT3 films exposed to two separate beams of electrons and photons, for radiation doses
Accepted 25 April 2014
Available online 29 May 2014
ranging up to 500 cGy. The analysis is done on a flatbed EPSON 10000 XL scanner and cross checked on a
HP Scanjet 4850 scanner. Both scanners are used in reflection mode taking into account landscape and
portrait scanning positions. After thorough verifications, the reflective scanning method can be used on
Keywords:
Gafchromic EBT3
EBT3 as an economic alternative to the transmission method which was also one of the goals of this
Film dosimetry study. A comparison is also done between single scan configuration including all samples in a single A4
Radiotherapy (HP) or A3 (EPSON) format area and multiple scan procedure where each sample is scanned separately
on its own. The images analyses are done using the ImageJ software. Results show significant influence of
the scanning configuration but no significant differences between electron and photon irradiations for
both single and multiple scan configurations. In conclusion, the film provides a reliable relative dose
measurement method for electrons and photons irradiations in the medical field applications.
! 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Introduction thus care should be taken during film handling and storage.
Another study by Ref. [2] revealed a 9% post-irradiation coloration
In the past years, using radiochromic films for dose assessment increase of MD-55-2 films up to 20 h after the irradiation, sug-
went through noticeable progress, and studies were emphasized on gesting that films should be scanned for read out at least 6 h after
this type of measurements in the medical field for radiation exposure. The properties of EBT Gafchromic films exposed to X-rays
dosimetry and radiotherapy control. Gafchromic film irradiation is radiation were also investigated by other authors e.g. Ref. [3] and
seen today as a simple and quick method for dose profiling in the energy response of the EBT2 films was found to be independent
different irradiation situations. However some properties should be of the incident photon energy over a wide range from 15 keV to
considered for more reliable results taking into account the sensi- w2.5 MeV for doses ranging between 25 cGy and 500 cGy. The
tivity of such type of films to several parameters e.g. radiation type absorption spectra variations during read out scanning of the EBT
and energy. Studies on the effect of read out light sources and film showed a clear increase in net optical density for frequencies
ambient light on previous versions of Gafchromic films, e.g. MD-55- around 636 nm in comparison with older versions of Gafchromic
1 and MD-55-2 [1], showed an interesting color modification after films [4]. The improvements applied on the EBT made this type of
few hours of exposure especially for fluorescent light sources, and films a powerful candidate for several applications like measuring
solar UV radiation [5], and as a potential dosimeter for ion beams
therapy [6], underlining the precision that can be achieved in
practical cases like medical tomotherapy [7]. A detailed review by
* Corresponding author. Physics Department, Faculty of Science, Saint Joseph Ref. [8] summarizes the improvement of Gafchromic films through
University, B.P. 11-514 Riad El Solh, Beirut 1107 2050, Lebanon. Tel.: þ961 1 421
000ext3450; fax: þ961 4 532 657.
the recent years and underlines the importance of precision in
E-mail addresses: ziad.francis@gmail.com (Z. Francis), marie.abboud@usj.edu.lb dosimetry for medical applications. However, some performance
(M. Abboud). issues were still reported for the EBT2 version, e.g., the

http://dx.doi.org/10.1016/j.ejmp.2014.04.010
1120-1797/! 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
N. Farah et al. / Physica Medica 30 (2014) 708e712 709

inhomogeneity of the film response [9], showing the importance of between low and high field sizes ranging from 3 " 3 cm2 up to
the repetitive and complementary verification studies that are re- 25 " 25 cm2 [14] on dose response so only the mentioned
ported in the literature for the different types of Gafchromic films. 10 " 10 cm2 configuration is adopted. The irradiation procedure is
In fact, comparisons between different versions of the EBT are of performed for doses ranging from 0 to 500 cGy with a step of
extreme importance, especially when it comes to assessing the 50 cGy for both electron and photon beams. The relative uncer-
improvements that are made for the medical applications dosim- tainty of the delivered dose is around 2%; this is the standard
etry. Moreover, the analysis protocols that are reported in the conventional value that is clinically used when continuous quality
literature can differ easily one from another leading to differences control procedures are implemented using the ionization chamber
in the obtained results. Here comes the importance of reporting as methods.
much details as possible when it comes to analysis methods.
In our study we use the most recent version of the EBT series
Scan procedure
namely the EBT3 film. The EBT3 is an improved version of the EBT2
film; it is made by laminating an active layer between two identical
The irradiated films were folded in dark envelopes at least for
polyester layers [10]. Studies published in the literature showed
24 h to ensure the darkening effect stabilization. The scanning is
some of the characteristics of the EBT3 e.g., energy dependence of
performed in landscape and portrait positions using two different
the film reaction after X-rays irradiation [11], particle type depen-
scanners: HP scanjet 4850 and EPSON 10000 XL with a resolution of
dence [12], specific IMRT applications [10] and comparisons with its
300 dots per inch (dpi) producing 48 bit RGB colored images in tiff
predecessor the EBT2 in medical irradiation applications [13]. One
format. When HP scanner is used, five blank scans are performed in
of the advantages of the EBT3 is that the analysis is independent
order to warm up the lamp of the scanner and the sixth scan is
from the film scan orientation which was not the case for the EBT2.
saved for analysis. In the case of Epson 10000 XL scanner, an
Here in this work, our interest lies mostly in comparing the
average of 2e3 preview scans was considered sufficient at the
response of the films to doses delivered by X-ray and electron
beginning of the scanning procedure to take into account the
beams, and using different analysis configurations with two flatbed
warm-up effect. According to Ref. [14] the impact of these effects on
reflective scanners. In fact, transmission scanners are more costly
the results does not exceed 0.1%. We consider two approaches of
than reflective scanners and might be more difficult to acquire in
scanning: (i) a single scan configuration where all the samples are
some cases. One of the issues we also wanted to verify here is the
placed according to increasing, decreasing or random dose order
reliability of the reflective analysis method comparing a relatively
from left to right in order to ensure that the position on the scanner
expensive scanner to the results obtained by a cheaper scanner
bed does not affect the results and then scanned at once fitting in an
model.
A4 or A3 format area for the HP and the EPSON scanners respec-
tively and (ii) a ‘multiple scan’ configuration where every sample is
Materials and methods
placed in the center of the scanner bed and scanned on its own. The
comparison between single and multiple scans approaches would
Gafchromic film EBT3
show the effect of the scanning phase parameter on the results. As
advised by the manufacturer, for the scanning, no filters or color
The Gafchromic EBT3 (lot number A 01231202, Ashland Spe-
correction functions are applied to raw pixel value results.
cialty Ingredients, NJ, USA) radiochromic film is available in sheets
of 20.32 " 25.4 cm2. It consists of an active layer of w28 mm of
thickness coated on both sides by a protective polyester layer of Scan analysis
100 mm. Samples are cut into pieces of 5 " 6 cm2 before the irra-
diation. This size order is acceptable and common to several studies Image measurement and analysis are performed on a region of
in the literature [7,12]; it covers an area that is wide enough to limit interest (ROI) of 480 " 480 pixels2 around the film samples center.
statistical uncertainties during irradiation and analysis. Handling Three values are taken to obtain the mean pixel value with its
the samples from cutting phase to the scanning and folding is done standard deviation in the ROI. The scanned images are then
using latex gloves to avoid stains and marks that can affect the analyzed using ImageJ [15] software. The RGB and the red
scanning results. The samples are folded with care in a dark en- component of the color are taken into account [5,16].
velope whenever not in use in order to avoid ambient light and The net reflective optical density “net ROD” is calculated using
other sources effects during storage. the pixels values from the scanned images according to the
following expression
Irradiation procedure ! "
Iu
Net ROD ¼ log (1)
The irradiations are accomplished using a Varian “TrueBeam Ii
1.6” accelerator providing electron beams of 6 MeV average energy
where Iu and Ii are the average pixel values of the reflected in-
and photon beams of 6 MV acceleration potential. Relatively low
tensities through non irradiated and irradiated films respectively
energies are chosen and phantoms are used in order to achieve the
[17].
best dose absorption rate and to minimize the irradiation time,
without altering the quality of the results. The considered config-
uration is thus compatible with our study since the dose response Results and discussion
of the film is independent of the depth in the phantom between 2
and 25 cm [14]. The irradiation field is considered uniform and As already mentioned, we consider in our study several analysis
limited within 10 " 10 cm2. The films are positioned between two configurations. The combination of different setups and approaches
water equivalent phantom layers of polystyrene; the build-up would lead to a high number of similar results that would be
phantom (1.5 cm thickness as in Ref. [4]) is placed between the cumbersome to show at once, thus, for the sake of clarity only the
source field and the irradiated sample and the support phantom interesting selection is displayed in this section. Furthermore, the
(10 cm thickness) laying behind the sample in the beam propaga- error bars on experimental values represent less than 0.1% and they
tion direction. Additionally, no significant influence is observed do not appear on the figures as they are smaller than the symbols.
710 N. Farah et al. / Physica Medica 30 (2014) 708e712

Scan orientation comparison

Figures 1 and 2 show the landscape and portrait scan orienta-


tion comparison for 6 MV photon and 6 MeV electron irradiations
respectively, for doses between 0 and 500 cGy. The Epson 10000 XL
scanner is used in reflective mode and the analysis is made in 2
different ways taking into account the complete RGB scanned im-
age for one analysis and only the RED channel for the other. The
landscape and portrait configurations lead to the same results for
the RGB and the Red channel analyses; the differences of the ROD
values are below 0.03% for the whole spectrum with an exception
reaching 0.07% for doses above 400 cGy for electrons irradiations.
Differences of around 0.45%e1.1% were reported in the literature
for a dose range reaching up to 10 Gy [12]. In the previous EBT
versions (EBT2) the scanning results showed noticeable differences
between landscape and portrait scan positions, which is not the
case for the EBT3 where the difference is negligible as mentioned
by the manufacturer and verified by Ref. [13]. Furthermore, in
previous studies on previous EBT films, the red channel showed a
higher sensitivity than the RGB [18] and this tendency is still valid
for the particular conditions applied in our study.
Figure 2. Dose response of EBT3 Gafchromic film irradiated with 6 MeV electrons. The
analysis details are the same as for Fig. 1.
Particle type and energy dependence

In order to verify the effect of the particle type on the EBT3 dependence on the particle type was also reported to be non-
coloring, samples are irradiated using 2 different particles, photons significant by Refs. [12] and [19] for the EBT3 predecessors (The
and electrons. The scan of the films is done in a landscape config- EBT2 and the EBT versions).
uration and a comparison between RGB and red channel analyses is
conducted. Figure 3 shows a comparison of the net ROD for electron Single versus multiple scan comparison
and photon irradiations. The obtained results show very close re-
sults for both radiation types and for the same doses. Discrepancies As the results can vary according to the scanning details e.g. scan
remain less than 0.07% for doses below 350 cGy and increase lamp temperature reference [20] and other parameters, we
slightly up to 0.1% for doses around 500 cGy. These findings are in compare two scan configurations. In a multiple scan approach, each
agreement with previous results from the literature where the irradiated sample is scanned together with the non-irradiated one
EBT3 film was used for IMRT energy range e.g., 6 MVe15 MV and consecutively from the lower dose to the higher one, then in a
the energy dependence was proven to be negligible [10]. For high decreasing order of dose and finally in a random order. In all three
energy photon beams generally used in radiotherapy energy cases, the analyses of the scanned images yield to the same result.
dependence of the EBT3 coloring is also weak [11] and the In a single scan approach, all the samples are placed in an A3 format
area and scanned at once using the EPSON 10000 XL scanner.

Figure 1. Dose response of EBT3 Gafchromic film irradiated with 6 MV photon beam
and scanned in landscape and portrait configurations using an Epson 10000 XL Figure 3. Dose response curve for EBT3 Gafchromic film irradiated with 6 MV photons
scanner. The analyses are shown using separately the RGB and the Red channel and 6 MeV electrons, scanned in landscape orientation using an Epson 10000 XL
filtering. scanner and analyzed using RGB and red color filtering.
N. Farah et al. / Physica Medica 30 (2014) 708e712 711

Figure 4. Superposition of red channel calibration curves for EBT3 Gafchromic film Figure 5. Dose response curve for EBT3 Gafchromic film irradiated with 6 MV photons
irradiated with 6 MV photons and 6 MeV electrons, scanned in landscape orientation and 6 MeV electrons, scanned in landscape orientation using 2 different scanners the
in single and multiple scans procedure using an Epson 10000 XL scanner. EPSON 10000 XL and the HP Scanjet 4850. The analysis is done using a red channel
filtering.

Results of multiple scan and single scan approaches are reported in


Fig. 4 for the red channel analysis. Differences between electron in this case as previously confirmed by Ref. [10]. Although the
and photon irradiations are less than 0.01%. Although these dis- multiple and the single scan procedures did not show any signifi-
crepancies are insignificant, we recommend a single scan approach cant difference in the results, we recommend to proceed in a single
for the whole measurements set, which allows for faster analysis scan approach as long as the size and the number of the samples
and avoids any potential doubts regarding the scan procedure as still fit within the scanning area.
discussed in Ref. [21] e.g. scan to scan differences, temperature While comparing both scanners, the difference is relatively large
stability and other environmental issues. between the 2 scanners analysis. This may depend on many pa-
rameters, mainly related to the scanner state and its design.
Therefore it is impossible to compare results from different scan-
Scanner type effect
ners. Despite the observed significant variation of sensitivity, the
two scanners exhibit similar precision for Gafchromic film
In our study, two different scanners are used, the Epson 10000
measurements.
XL which is commonly dedicated to this kind of analysis in the
In conclusion, the EBT3 film is suitable for use as a dose control
medical field and a desktop flatbed HP scanner (Scanjet 4850). We
method in the clinical field for different radiation types and
illustrate in Fig. 5 a comparison between both scanners. The dis-
different doses. The analyses in reflective mode are reliable, and the
played results are obtained for a landscape orientation and using a
use of the red channel filtering leads to more ROD sensitivity
red channel filtered analysis. Notwithstanding the significant var-
therefore to better dose measurement results. For a given system,
iations in sensitivity, both considered scanners exhibit good pre-
pertinent methods should be developed as part of the dosimetry
cision for Gafchromic film based dose measurements.
protocol during the calibration procedure.
Although results show a higher sensitivity of the Epson scanner
reaching nearly 2 times the HP sensitivity for doses above 400 cGy,
one can use relatively old and cost saving scanners as long as a Acknowledgments
suitable calibration is performed before use.
Authors are grateful for the Research Council of the Saint Joseph
University for funding this project (Funding FS45), and would like
Conclusion
to thank the Department of Radiation-Oncology of the HDF for the
technical support.
Based on our comparisons for the Gafchromic film dosimetry in
the range 0e500 cGy, and supported by the results from Ref. [11]
and those obtained by Ref. [12], this study suggests that the EBT3 References
film is suitable for reflective scanning analysis. Furthermore the
[1] Butson MJ, Yu PKN, Metcalfe PE. Effects of read-out light sources and ambient
results showed no difference between landscape and portrait scan light on radiochromic film. Phys Med Biol 1998;43:2407e12.
orientation. In fact, the EBT3 properties should be investigated [2] Cheung T, Butson MJ, Yu PKN. Post-irradiation colouration of Gafchromic EBT
more thoroughly in future studies in order to confirm the reliability radiochromic film. Phys Med Biol 2005;50:N281e5.
[3] Butson MJ, Yu PKN, Cheung T, Alnawaf H. Energy response of the new EBT2
of the method described in this paper. Similarly to its predecessor
radiochromic film to X-ray radiation. Radiat Meas 2010;45:836e9.
the EBT2, the EBT3 film reaches larger values of optical density for [4] Butson MJ, Cheung T, Yu PKN. Absorption spectra variations of EBT radio-
the red channel, thus showing more sensitivity than with the RGB chromic film from radiation exposure. Phys Med Biol 2005;50:N135e40.
spectrum analysis as revealed previously by Fig. 3. The differences [5] Butson ET, Cheung T, Yu PKN, Butson MJ. Measuring solar UV radiation with
EBT radiochromic film. Phys Med Biol 2010;55:N487e93.
between photon and electron irradiation results can be neglected [6] Martisíková M, Jakel O. Study of Gafchromic EBT film response over a large
showing that the EBT3 response is independent of the particle type dose range. Phys Med Biol 2010;55:N281e90.
712 N. Farah et al. / Physica Medica 30 (2014) 708e712

[7] Avanzo M, Drigo A, Ren Kaiser S, Roggio A, Sartor G, Chiovati P, et al. Dose to [15] ImageJ. http://imagej.nih.gov/ij/.
the skin in helical tomotherapy: results of in vivo measurements with radi- [16] Richley L, John A, Coomber H, Fletcher S. Evaluation and optimization of the
ochromic films. Phys Med 2013;29:304e11. new EBT2 radiochromic film dosimetry system for patient dose verification in
[8] Devic S. Radiochromic film dosimetry: past, present, and future. Phys Med radiotherapy. Phys Med Biol 2010;55:2601e17.
2011;27:122e34. [17] Ohuchi H. High sensitivity radiochromic film dosimetry using an optical
[9] Hartmann B, Martisíková M, Jäkel O. Technical note: homogeneity of gaf- common-mode rejection and a reflective-mode flatbed color scanner. Med
chromic EBT2 film. Med Phys 2010;37:1753. Phys 2007;34:4207e12.
[10] Casanova BV, Pasquino M, Russo G, Grosso P, Cante D, Sciacero P, et al. [18] Alva H, Mercado-Uribe H, Rodriguez-Villafuerte M, Brandan ME. The use of a
Dosimetric characterization and use of Gafchromic EBT3 film for IMRT dose reflective scanner to study radiochromic film response. Phys Med Biol
verification. J Appl Clin Med Phys 2013;14:4111. 2002;47:2925e33.
[11] Massillon-JL G, Chiu-Tsao S-T, Domingo-Muñoz I, Chan MF. Energy depen- [19] Arjomandy B, Tailor R, Anand A, Sahoo N, Gillin M, Prado K, et al. Energy
dence of the new Gafchromic EBT3 film: dose response curves for 50 kV, 6 and dependence and dose response of Gafchromic EBT2 film over a wide
15 MV X-ray beams. Int J Med Phys Clin Eng Radiat Oncol 2012;1:60e5. range of photon, electron, and proton beam energies. Med Phys 2010;37:
[12] Sorriaux J, Kacperek A, Rossomme S, Lee J, Bertrand D, Vynckier S, et al. 1942e7.
Evaluation of Gafchromic EBT3 films characteristics in therapy photon, elec- [20] Sorriaux J, Kacperek A, Rossomme S, Lee J, Bertrand D, Vynckier S, et al.
tron and proton beams. Phys Med 2013;29:599e606. Evaluation of Gafchromic EBT3 films characteristics in therapy photon, elec-
[13] Reinhardt S, Hillbrand M, Wilkens JJ, Assmann W. Comparison of Gafchromic tron and proton beams. Phys Med 2012;12:S1120e797.
EBT2 and EBT3 films for clinical photon and proton beams. Med Phys [21] Lewis D, Micke A, Yu X, Chan MF. An efficient protocol for radiochromic film
2012;39:5257e62. dosimetry combining calibration and measurement in a single scan. Med Phys
[14] Fuss M, Sturtewagen E, De Wagter C, Georg D. Dosimetric characterization of 2012;39:6339e50.
GafChromic EBT film and its implication on film dosimetry quality assurance.
Phys Med Biol 2007;52:4211e25.

You might also like