Dräger Evita® V300 and Evita® Infinity® V500

Disassembly/Assembly

D-12360-2016
Disassembly/Assembly Flow Sensor
91 03 106 | 16.12-1 | PR | LE | Subject to modifications | © 2016 Drägerwerk AG & Co. KGaA

Open the flap by lifting the lower Push the flow sensor as far as it Remove the flow sensor
edge upwards. will go to the left. from the socket.

Push the flow sensor to Fit the flow sensor


Close the flap. the right into the socket. into the socket.

Disassembly/Assembly Expiratory Valve*


Turn the locking ring as far as Remove the expiratory valve Pull the sensor sleeve from
possible to the left. from the fitting. the expiratory valve.

Turn the locking ring as far Turn the locking ring as far as
Fit the flow sensor sleeve
as possible to the right possible to the left, push the
on the expiratory valve.
until it clicks into place. expiratory valve into the fitting.

* The Neonatal Expiratory Valve is


Remove and empty the water
Remove the diaphragm. also working with Evita® V300
trap container.
and Evita Infinity® V500. For
manual reprocessing of the
Neonatal Expiratory Valve please
refer to guideline: Babylog®
VN500 Reprocessing 9103108

Fit the diaphragm onto the edge


Fit the water trap container.
of the expiratory valve housing.

This guideline does not replace or amend the instructions for use.
The instructions for use and the hospital guidelines and policies always have to be observed.
Dräger Evita® V300 and Evita® Infinity® V500
Flow Sensor: Reprocessing instructions
Manual cleaning and disinfection

D-12360-2016
At the beginning and at the Rinse the flow sensor in water
Prepare cleaning solution. end vigorously swirl the flow (minimum drinking-water quality)
sensor in the cleaning solution and wash off all cleaning agent
at least 3 times. residues.

Immerse flow sensor bubble-free Ensure that the cleaning


in the cleaning solution for the solution flushes all surfaces
agent specific exposure time. and interior spaces.

Sufficiently rinse the flow sensor Thoroughly shake off remaining


Prepare the disinfectant solution. in water (minimum drinking-water water.
quality) until no disinfectant
residues can be recognized.

Repeatedly swirl the flow sensor


in the disinfection solution for
the agent specific exposure time.
Ensure that the solution flushes Inspect flow sensor for visible Allow the flow sensor to air
all surfaces and interior spaces. soiling and damage. dry for at least 30 minutes.

In case of a combined cleaning-disinfection agent please also follow the guideline as described (two times immersion).

This guideline does not replace or amend the instructions for use.
The instructions for use and the hospital guidelines and policies always have to be observed.
Dräger Evita® V300 and Evita® Infinity® V500
Reprocessing instructions
Manual cleaning and disinfection

D-10710-2016
Reprocessing instructions overview:

Manual cleaning and Machine cleaning and Sterilization


disinfection disinfection

Machine cleaning and disinfection: Sterilization:


Securely position the product in Check the product for visible
the basket. damage and replace if necessary.
Ensure the following: For these semi critical products steri-
– All surfaces and interior spaces lization is possible but not mandatory.
can be flushed completely. For further information please
– The water can drain off freely. refer to Instruction for Use.

This guideline does not replace or amend the instructions for use.