Disassembly/Assembly
D-12360-2016
Disassembly/Assembly Flow Sensor
91 03 106 | 16.12-1 | PR | LE | Subject to modifications | © 2016 Drägerwerk AG & Co. KGaA
Open the flap by lifting the lower Push the flow sensor as far as it Remove the flow sensor
edge upwards. will go to the left. from the socket.
Turn the locking ring as far Turn the locking ring as far as
Fit the flow sensor sleeve
as possible to the right possible to the left, push the
on the expiratory valve.
until it clicks into place. expiratory valve into the fitting.
This guideline does not replace or amend the instructions for use.
The instructions for use and the hospital guidelines and policies always have to be observed.
Dräger Evita® V300 and Evita® Infinity® V500
Flow Sensor: Reprocessing instructions
Manual cleaning and disinfection
D-12360-2016
At the beginning and at the Rinse the flow sensor in water
Prepare cleaning solution. end vigorously swirl the flow (minimum drinking-water quality)
sensor in the cleaning solution and wash off all cleaning agent
at least 3 times. residues.
In case of a combined cleaning-disinfection agent please also follow the guideline as described (two times immersion).
This guideline does not replace or amend the instructions for use.
The instructions for use and the hospital guidelines and policies always have to be observed.
Dräger Evita® V300 and Evita® Infinity® V500
Reprocessing instructions
Manual cleaning and disinfection
D-10710-2016
Reprocessing instructions overview:
This guideline does not replace or amend the instructions for use.
The instructions for use and the hospital guidelines and policies always have to be observed. Manufacturer: