Breathing System
Disassembly/Assembly
D-39523-2015
Disassembly
91 02 536 | 16.08-1 | HQ | PP | Subject to modifications | © 2016 Drägerwerk AG & Co. KGaA
Loosen the three screws of the Pull the breathing system up Remove the upper ventilator
ventilator with a hexagon key. and out by the handle. diaphragm.
90°
90°
Remove the CLIC CO2 absorber. Insert the cleaning disk into Loosen the inspiration and
Unscrew the CLIC adapter the adapter. Close the adapter expiration nozzles and pull
(optional). until it clicks into place. them off.
Screw on the CLIC adapter Remove cleaning disk from Put on the inspiration nozzle
(optional). Insert the CLIC CO2 the adapter. Close the adapter and expiration nozzle and
absorber. so that it can click into place. turn to tighten.
90°
90°
Place the cover on the breathing Place the middle part of the
Insert both flow sensors. system and tighten the five breathing system on the bottom
screws with a screwdriver. part of the housing.
* Primus Infinity Empowered, Primus, Pallas, Apollo
Assembly
Primus® Family*
Breathing System
Reprocessing Instructions
D-39523-2015
Manual cleaning Machine cleaning Sterilization
and disinfection and disinfection
Visual check
Immersion Disinfection:
Please use the disinfectants listed in the operating instructions
for Dräger flow sensors.
Sterilization:
SpiroLife® flow sensors are not suitable for plasma or
radiation sterilisation.