Professional Documents
Culture Documents
®
OWNER’S MANUAL
Multiload
Ethylene Oxide
Sterilization
SERIES 3
sterilization redefined
PN4004 050306
Andersen Sterilizers, Inc. Health Science Park Haw River, NC 27258 USA
AN ISO 9001 - EN46001 certified company
EOGas™ Series Plus Owner’s Manual
Andersen Sterilizers, Inc. reserves the right to correct typographical errors. Specifications are
subject to change. Nothing in this manual may be reproduced in any manner, either
wholly or in part, for any use without prior written permission from Andersen Sterilizers,
Inc.
®
is a registered AirScan Enviromental Technologies, Inc.
2
EOGas™ Series Plus Owner’s Manual
Phone: 336-376-8622
Fax: 336-376-3088
The user should be aware that if the EOGas sterilizer is installed and operated in a
manner not specified by the company the protection provided by the equipment may
be impaired.
The specifications contained in this manual are subject to change without notice.
3
EOGas™ Series Plus Owner’s Manual
INTRODUCTION
This manual provides an overview of the installation, operation, and maintenance require-
ments of EOGas Series 3 Plus Sterilizers, and the information necessary for troubleshoot-
ing, testing, and repairing the sterilizer cabinets and heat sealer.
EOGas Series 3 Plus is a sterilization system manufactured by Andersen Sterilizers, Inc. Its
components include:
The EOGas
sterilizer cabinet
EOGas cartridges
Dosimeters®
Humidichip® relative
humidity stabilizing chips
4
EOGas™ Series Plus Owner’s Manual
The EOGas system uses 100% Ethylene oxide gas in sealed cartridges. Items to be
sterilized are prepared, wrapped, and placed inside a sterilization bag along with the
appropriate size EOGas cartridge, a Humidichip relative humidity stabilizer chip and an
EOGas Dosimeter. After the sterilization bag is sealed, the gas is released inside the
sterilization bag by activating the cartridge through the bag wall. The standard sterilization
cycle takes 16-hours.
EOGas Series 3 Plus sterilizers have three different operating modes - STANDARD,
BATCH, and INDUSTRIAL - enabling operators to choose the most efficient mode for their
sterilization needs.
5
EOGas™ Series Plus Owner’s Manual
CONTENTS
INSTALLATION……………………………………………………………………..8
Unpacking…………………………………………………………………………...9
Initial Setup………………………………………………………………………….8
Installation Requirements………………………………………………………….10
Power Failure……………………………………………………………………….11
Label Printer Connection………………………………………………………….12
PROCESSING REVIEW……………………………………………………..……13
Disassembly………………………………………………………………………...13
Cleaning……………………………………………………………………………..13
Humidifying………………………………………………………………………….14
Drying………………………………………………………………………………..15
Wrapping…………………………………………………………………………….15
Description of an EOGas Cycle…………………………………………………..16
6
EOGas™ Series Plus Owner’s Manual
APPENDIX
Diagrams and Drawings
Figure 1 EOGas Sterilizer Operation Reference Chart
Figure 2 Sterilizer Dimensions
Figure 3 Heat Sealer
Figure 4 Vacuum Sealer
Figure 5 Venting Through Outside Wall
Figure 6 Venting to the Outside within an Interior Wall
Figure 7 Exhaust Calculations
Figure 8 Component Module Layout
Figure 9 EOGas Extractor Unit
Figure 10 Gas Disposal System
Figure 11 Equipment Ratings Summary
Figure 12 Controls Layout
Figure 13 Ventilation Alarm Schematic
Figure 14 Electrical Schematic 120 V
Figure 15 Electrical Schematic 240 V
Figure 16 Replacement Parts List
Figure 17 IEC Warning Symbols
Figure 18 Routine Maintenance Schedule
7
EOGas™ Series Plus Owner’s Manual
EOGas
Sterilizer
Installation
8
EOGas™ Series Plus Owner’s Manual
INSTALLATION INSTRUCTIONS
Unpacking
Remove the sterilizer from the shipping crate. Look for the following items:
Exhaust Blower
Exhaust Hose
Uninterruptible Power Supply (UPS) for units without an abator
Label Printer
Power Cable
Spare Fuses
Cabinet Positioning
When mounted on a base unit, EOGas Sterilizer Cabinets are freestanding on casters with
brakes. After positioning the cabinet, the two front casters can be used to lock the cabinet
in place. Alternatively, the cabinet may be mounted on a bench, without a base unit.
Position the cabinet and lock in place if desired. If an Abator will be used, skip to the
paragraph below on p. 9 titled "Abator Connection".
Connect the blue Exhaust Blower hose to the cabinet exhaust outlet. Attach the flange with four
screws to the to top cover.
9
EOGas™ Series Plus Owner’s Manual
Mount the Uninterruptible Power Supply (UPS) to the back of the cabinet.
Connect the power cord of the Exhaust Blower to the output receptacle on the UPS.
Abator Connection
If the EOGas sterilizer is provided with an Abator, the hose from the cabinet to the Abator
and the Abator Exhaust must be connected. The Abator exhaust blower must be operating
before proceeding to "Setting Sterilization Parameters and Operating Modes" (p. 20)
under "EOGas Sterilizer Operation."
Abators require space for both the Abator unit and Exhaust Blower. The induction-type
Exhaust Blower should be mounted at the end of the exhaust ducting, and care should be
taken to eliminate all leaks in the ductwork. The Abator and Exhaust Blower may be
mounted on the roof or any location that will allow access for maintenance and changing the
catalyst. The Abator weighs 385 pounds (175 kilos) and the Exhaust Blower 65 pounds (30
kilos). The Abator unit does not require a power connection. The Exhaust Blower requires
an independent power supply, preferably one that will not be subject to power outage.
10
EOGas™ Series Plus Owner’s Manual
Power Requirements
The EOGas Cabinet is fitted with a 5 ft. (1.5m) grounded three prong power cord. The
power requirements are 120V/60Hz 15 Amp (U.S.), 220V/50Hz 8Amp (Europe), or
220V/60Hz 10 Amp (Korea).
A dedicated circuit, separately fused and connected to a standby power supply is required.
Clearance of at least 28" (71cm) is necessary in front of the cabinet door. A clearance of 2
feet (61cm) above the cabinet is recommended.
Power Connection
Connect the EOGas Sterilizer power cord to the power outlet. A dedicated power
outlet that has standby power is required. Incoming power fuses are installed in the
cabinet power receptacle. Spare fuses for 110V models are enclosed in the recap
receptacle and with the Owners Manual (see p. 65 for information on "Fuse
Replacement").
Power Failure
The EOGas Series 3 Plus computer has a standby power supply to retain package
numbers and sterilization cycle information in microprocessor memory.
During a power outage the microprocessor prevents access to the cabinet. (See
page 68 "Manual Door Unlock" for emergency unlocking). During a power failure
the cabinet heaters and the circulating fan do not operate. The ventilation unit will
continue to ventilate the cabinet from the standby power supply.
When power is restored, the time of day the power outage occurred is printed.
Sterilization cycle time is extended to take into account the time the power was
disconnected. The display will return to normal, showing "Warming Up" and the
current temperature. The low temperature alarm will not sound after power
restoration while the temperature in the cabinet returns to operating temperature
(50ºC).
11
EOGas™ Series Plus Owner’s Manual
Three-eared
screw
12
EOGas™ Series Plus Owner’s Manual
PROCESSING REVIEW
Material to be sterilized by ethylene oxide (EOGas) must be meticulously cleaned and dried.
Coatings of dry protein like dry pus, blood, or feces, protect microorganisms and slow the
sterilization process. You must always be sure to take the following precautions before
sterilizing with EOGas:
Disassembly
EOGas diffuses readily through many materials. However, occlusive caps, plugs or stylets
must be removed from instruments so that the gas can penetrate freely. Syringes must be
disassembled for sterilization. Batteries must be removed from electrical devices and
sterilized separately to prevent electrical sparks.
Cleaning
Disassemble and scrub all instruments in detergent and water to the most critical standard
of cleanliness possible.
13
EOGas™ Series Plus Owner’s Manual
Humidifying
If cleaning in water will damage the items to be sterilized, pretreat for 4 hours in a
sterilization bag having saturated humidity. To humidify the contents of the sterilization bag,
wrap the items individually in paper or cloth in the usual manner. Place them in a
sterilization bag along with a Humiditube™, Humidichip®, Dosimeter® and an EOGas
cartridge (Remove the cartridge trigger guard but DO NOT ACTIVATE CARTRIDGE).
Complete the provided Dosimeter card before placing it in the sterilization bag. Use Color
Change Indicators, if desired, for further visual assurance at the end of a cycle of exposure
to Ethylene oxide.
After the Preconditioning Cycle, remove the bag from the sterilizer. It may then be placed in
the sterilizer cabinet for a normal-length, 16 hour sterilization cycle.
AN1072 Humiditube The AN1072 Humiditube provides a barrier from any surrounding
product, while allowing adequate and full release of AN1071 Humidichip moisture. The
tube does not absorb Ethylene oxide or affect the sterilization and aeration process.
AN1071 Humidichips ensure that humidification requirements for Ethylene oxide steriliza-
tion are met. The AN1071 Humidichip measures 2" x 2" (51mm x 51 mm) and contains at
least 4 gms of water. One Humidichip per sterilization bag per cycle is sufficient for stan-
dard humidification needs. Increase the number of Humidichips to two if the load contains
primarily dry cotton or paper.
14
EOGas™ Series Plus Owner’s Manual
AN1087 Dosimeters are designed to change color from yellow to blue when exposed to
Ethylene oxide. They integrate the effects of time, temperature and the concentration of
Ethylene oxide in contact with the crystals in the capillary tube. For a load to be considered
sterile, the color change from yellow to blue must extend past the triangular mark on the
label. No laboratory testing is required. The information is available immediately at the end
of a sterilization cycle.
AN85 Color Change Indicators indicate exposure to Ethylene oxide gas by turning color
from yellow to blue. They show that there has been exposure to Ethylene oxide but they do
not indicate the actual dose. The convenient self-stick backing adheres to conventional
paper or cloth wrapping, or to Seal and Peel® transparent plastic packaging.
Drying
Be sure that items to be sterilized are physically dry before wrapping and processing.
Towel drying or drain drying is sufficient. Do not use hot air drying. Water on instruments
at the time of exposure to EOGas may react with the gas and reduce its effectiveness.
Hollow bore needles and plastic or rubber tubing must be open and free from stylets and
plugs. Syringes must be packaged disassembled with their plungers out of the barrel.
Wrapping
All items to be sterilized must be wrapped in cloth or paper in the manner conventional for
steam sterilization or in ANPRO Seal and Peel® Packaging. ANPRO Seal and Peel
packaging provides a transparent, peel-open, extended shelf life package, compatible with
the EOGas sterilization system. No other plastic film packaging material may be used with
EOGas unless it has been tested and adequate ethylene oxide gas permeability has been
demonstrated. Polyamide (Nylon) and polyester (Mylar) films are known to be impermeable
to EOGas. Do not pack the sterilization bag so tightly with cloth or gauze that gas diffusion
is slowed.
15
EOGas™ Series Plus Owner’s Manual
Remove a Dosimeter card from the dispenser box. Write the date and the time sterilization
and aeration will be complete on the card. Take a sterilization bag from the EOGas
dispenser box and place the wrapped items into the sterilization bag, along with one
AN1071 Humidichip and the Dosimeter indicator card. Place the Humidichip so that it will
be near the metal shelf.
Choose an EOGas cartridge from the dispenser box. Confirm that the number printed on
the EOGas cartridge corresponds with the number printed on the sterilization bag you have
chosen. Remove the safety tab on the EOGas cartridge (covered with tape) and place the
EOGas cartridge into the sterilization bag, but do not activate it at this time.
Remove excess air with a vacuum sealer or press the excess air out of the sterilization bag
and heat seal it.
16
EOGas™ Series Plus Owner’s Manual
When the cabinet has been purged, "Door unlocked" will be displayed and the door will
unlock. Place the sterilization bag on a shelf in the sterilizer.
Without opening the sterilization bag, grasp the cartridge of EOGas through the wall of the
sterilization bag and press the plunger to activate the cartridge.
Leave the sterilization bag undisturbed in the sterilizer cabinet for 16 hours. After the
sterilization cycle is complete, the sterile contents of the sterilization bag may be removed
and used. After the sterile material has been removed from the sterilization bag, the empty
gas cartridge and the sterilization bag may be disposed of in an ordinary rubbish container.
The garbage container should be in a well-ventilated area.
17
EOGas™ Series Plus Owner’s Manual
Activating the EOGas Cartridge
Remove cartridge
1 trigger guard.
No 5
5
No 5
18
EOGas™ Series Plus Owner’s Manual
CAUTION
It is incumbent on the user to confirm that gas absorbent items sterilized in EOGas
have been adequately aired before they contact living tissue. This is particularly
important for plastic or rubber items that will contact tissue culture preparations,
microbial cell cultures, spermatocytes, oocytes, embryonic tissue, etc. Failure to
adequately air Ethylene oxide-absorbent materials may lead to contact chemical
burns.
The 16 hour sterilization cycle time at 50ºC called for in these instructions includes an
adequate amount of time to air most tubes, masks and the like. Large gas absorbent items
like breast implants and plastic extracorporeal blood filters may require additional aeration
before they can be safely implanted or used. Aeration will progress more rapidly outside of
the sterilization bag. Customers may obtain assistance in determining aeration times by
contacting the Manufacturer or Distributor as listed on page 2 of this manual.
Place the unopened Steritest with the items to be sterilized. At the end of the cycle, remove
the Steritest and look at the Dosimeter. A color change from yellow to blue that extends to
the triangular mark on the Dosimeter label indicates that a dose of Ethylene oxide sufficient
for sterilization has been delivered. Without opening the Steritest pouch, manipulate the
ampoule of culture broth inside of its break shield so that the neck of the ampoule is broken.
Gently shake the broth down to cover the spore disk. Incubate the Steritest at 37.5ºC for 72
hours. A change in the color of the broth from blue to orange indicates growth of bacteria
and therefore an unsterile load.
An incubator for Steritest is available from Andersen Products, catalog number AN810.
19
EOGas™ Series Plus Owner’s Manual
EOGas
Sterilizer
Operation
20
EOGas™ Series Plus Owner’s Manual
Pressing the SETUP button initiates the screens for setting machine parameters.
The SUPERVISOR CODE is needed to enter Setup: Press 14789 on the Control Panel
Keypad and then ENTER.
Calibrating the cabinet temperature requires the TECHNICIAN CODE: Press 5344 on the
Control Panel Keypad and then ENTER.
Pressing the ENTER button moves through the items in order. Items flash on and off when
selected.
Use the UP/DOWN arrows to cycle through the choices for items.
SUPERVISOR CODE
The Setup section may be entered from "Standby" if Standard Mode or Batch Mode are
selected or at any time from Industrial Mode. This restriction assures that the sterilizer will
not lose track of any bags that are already in process in Standard or Batch Mode.
Once SETUP is pressed the Supervisor Code (14789) or the Technician Code (5344) must
be entered on the Control Panel Keypad, followed by the ENTER button. The codes assure
that only authorized personnel will change operating parameters or calibrate the
temperature in the cabinet.
21
EOGas™ Series Plus Owner’s Manual
SUPERVISOR CODE
The two Setup screens have a total of ten parameters that may be changed, seven on the
first screen and three on the second. Parameters that may be changed flash on and off. To
move between the parameters and between the Setup screens press the ENTER key.
language mode
Sterilizer number
Sterilizer Number
An integer 1 to 9. This allows multiple machines to be used simultaneously. Each
sterilizer is assigned a unique number.
Language
Cycle through choices using the UP/DOWN arrows. Default is ENG (English).
Mode
Standard (STNDRD), Batch (BATCH) or Industrial (INDUST) - use UP/DOWN
arrows to toggle selections.
Standard Mode is the only mode that prints labels and tracks the bags
within the sterilizer. Each bag is given a unique label with a number. The
number has one digit for the sterilizer number, two digits for the bag number
(assigned by the computer) and six digits for the date in the format day/
month/year. For example, bag number 2 08 220698 would indicate
Sterilizer number 2, bag number 8 loaded on June 22, 1998.
Batch Mode is used if all bags in a particular sterilizer are loaded and
unloaded at the same time. Sterilization is done in "lots" that fill the sterilizer.
No bags may be added to the sterilizer or removed from the sterilizer during
the cycle.
Industrial Mode is for commercial use. The operator must keep track of
bag loading and unloading as well as cycle length. This Mode may be used
if appropriate, validated cycle parameters have been determined for the
22
EOGas™ Series Plus Owner’s Manual
temperature
Cycle Length
In hours (H). The default cycle length is 16 hours (16H). Cycle length may be changed
from a minimum of 16 hours to a maximum of 24 hours in increment of 1 hour using the
UP/DOWN arrows. Cycle length must not be changed except as part of a validated
sterilization protocol.
Date
In format day/month/year. Enter the date using the keypad.
Time
24-hour clock. Enter the time using the keypad. operator code
print interval
Print Interval
CE, 1M, 15M, 30M, 1HR, 2HR, 4HR, 6HR. Move through the choices using the UP/DOWN
arrows. CE stands for “Cycle End”. In this case, cycle information is printed by the panel-
mounted printer only at the end of the cycle (i.e., whenever the UNLOAD button is pressed).
1M is one minute intervals, 15M is 15 minute intervals, 30M is 30 minute intervals, 1HR is 1
hour intervals, 2HR is 2 hour intervals, 4HR is 4 hour intervals, and 6 HR is 6 hour intervals.
Code
Yes or No. Toggle the choice with the UP/DOWN arrows. The Operator Code is a 3 digit
code that must be set by the customer (unlike the Supervisor Code and Technician Code).
It is designed to avoid unauthorized entry into the sterilizer. Only one code may be
assigned for each machine, but it may be changed at any time. If CODE is set YES then
the operator is given the option of changing the code.
23
EOGas™ Series Plus Owner’s Manual
Change Code
YES or NO. Toggle the choice with the UP/DOWN arrows. Default is set to NO. To
change the OPERATOR CODE, toggle to YES and press ENTER. The following screen will
appear:
Enter a new 3 digit code using the keypad. Press ENTER then repeat the same 3 digit
code and press ENTER again. If the same 3 digit code is used both times the code will be
set to the new code.
Thermocouple
To calibrate temperature, the machine must be in INDUSTRIAL mode and the cabinet must
be at a steady-state temperature (wait 30 minutes with the door closed to reach steady-
state temperature).
To begin calibration, press SETUP, then enter the technician code (5344) and then press
ENTER.
SUPERVISOR CODE
****
24
EOGas™ Series Plus Owner’s Manual
The first screen will show the lot number of the sterilizer and the version number of the
software:
EOGAS 3+
LOT 160705 VER 1.30
PREVIOUS CALIBRATION
46.5ºC TO 50.6ºC
After the actual temperature is entered using the keypad, press ENTER. The change will
be displayed on the screen:
PREVIOUS CALIBRATION
49.8ºC TO 51.1ºC
Press SETUP to return to INDUSTRIAL mode screen or press ENTER to set the
temperature again.
25
EOGas™ Series Plus Owner’s Manual
PRECONDITIONING CYCLE
The Preconditioning Cycle is a 4 hour cycle which humidifies a load prior to sterilization.
The machine will warm up and maintain a temperature of 50ºC for 2 hours, then cool down
to room temperature for 2 hours.
Humidification is an important part of the sterilization cycle (see page 12, "Humidifying").
Items that may be immersed in water and towel or drip-dried will not need preconditioning.
Items that may be damaged by immersion in water or a load that contains a large amount of
material that will absorb water (dry paper and cloth) may need humidification. The items
should be wrapped and prepared in the usual way (see page 12, "Processing Review")
with an EOGas Cartridge, AN1071 Humidichip, AN85 Color Change Indicator and/or
AN1087 Dosimeter. The bag should be sealed but the EOGas Cartridge should not be
activated.
This cycle may be used only when the machine is in STANDBY mode (no bags in process).
Press the zero key twice (00). The following screen will appear:
ENTER key
26
EOGas™ Series Plus Owner’s Manual
Close the door after loading the bags. If the door is not closed by the time the timer gets to
00:00, the following screen will be shown and the alarm will sound until the door is closed:
Once the door is closed and the timer has finished counting down the machine will enter the
Preconditioning Cycle:
PRECONDITIONING 25ºC
WARMUP 1:59:59
The temperature will warm up to 50ºC and after 2 hours the machine will cool down to room
temperature. The display will read:
27
EOGas™ Series Plus Owner’s Manual
After the 2 hour cool down period, the display will read:
PRECONDITIONING 27ºC
DONE
If necessary, bags may be removed from the machine after the first 2 hours and allowed to
cool outside the sterilizer for 2 hours. This has the advantage of not tying up the sterilizer
for another 2 hours. However, the person operating the sterilizer will then have to keep
track of the "cool down" time. If the bags are left in the sterilizer the machine keeps track of
the time.
Once the door has been opened the display will read:
Once the bags have been removed, close the door and the sterilizer will go back to
STANDBY mode. At that point, a sterilization cycle may be started as described later in the
manual.
28
EOGas™ Series Plus Owner’s Manual
In STANDARD Mode, the microprocessor keeps track of the bags in process in the
sterilizer, and the operator may load new bags and unload finished bags while others are in
process. When a bag is loaded, two labels are printed. One must be placed on the outside
of the bag, the other should be placed in a sterilization log book. Each bag is given a
unique label. The machine notifies the operator when bags are ready to be removed from
the machine. If there are no bags in the sterilizer, the machine will enter STANDBY to save
energy. The heaters are turned off in STANDBY but the fans continue to operate.
LOAD key
If bags are already in process, LOAD may be pressed to add additional bags to the
machine.
If an OPERATOR CODE is required (3 digit code), the machine will prompt the operator for
the code, which should be entered using the keypad followed by ENTER:
OPERATOR CODE
***
ENTER key
29
EOGas™ Series Plus Owner’s Manual
The sterilizer will then prompt the operator to enter the number of bags to load:
The number of bags may be set using the keypad or the UP and DOWN arrows
After the number of bags has been set, press the ENTER key to confirm.
The sterilizer will then enter a WARMUP cycle (if there were no bags already in process) or
a PURGE cycle
(if bags are in process)
WARMUP 28.7ºC
DOOR UNLOCK AT 48ºC
The door will unlock allowing bags to be loaded after the temperature reaches 2ºC below
the operating temperature (in this example 48ºC which is 2º less then 50ºC).
30
EOGas™ Series Plus Owner’s Manual
PURGING 05:00
06 BAGS 02 DONE
Purging is high volume ventilation to remove any residual Ethylene oxide from the cabinet.
The door will unlock after the timer counts down to 00:00
At this point the door will unlock to allow bags to be loaded. A "beep" will sound every 15
seconds to notify the operator that the cabinet is ready for loading. The screen will read:
Bags may now be loaded. Once the door is opened to load the bags the screen will read:
countdown timer
The bag numbers for the bags to be loaded may be viewed on the screen, if desired. Cycle
through the bags to be loaded using the UP and DOWN arrows. The door may be opened
and closed as many times as desired while the counter is counting down to 00:00. After
loading the bags, close the door to the sterilizer. It will lock automatically after the
countdown timer reaches 00:00.
31
EOGas™ Series Plus Owner’s Manual
If the door is still open when the counter reaches 00:00, an alarm will sound to notify the
operator and the screen will display:
Once the door is closed, it locks automatically and the STANDARD Mode normal operating
screen is displayed:
STERILIZING 49.8C
08 BAGS 09:18
32
EOGas™ Series Plus Owner’s Manual
The STANDARD Mode normal operating screen displays the total number of bags in
process, and the number of bags that are done. Before the sterilization cycle is complete
only the supervisor may unload bags (see "Emergency Unloading" on p. 43).
The screen will show the following information during a cycle before any bags are done.
STERILIZING 50.0ºC
06 BAGS 15:52
Once one or more bags are done the display changes. The following example shows there
are 6 bags in the machine, 2 of which are done.
STERILIZING 50.0ºC
06 BAGS 02 DONE
6 bags in process
2 bags done
To unload the bags that are done, start by pressing the UNLOAD key:
STERILIZING 50.0ºC
06 BAGS 02 DONE
UNLOAD key
33
EOGas™ Series Plus Owner’s Manual
If an OPERATOR CODE is required (3 digit code), the machine will prompt the operator for
the code, which should be entered using the keypad, followed by ENTER:
OPERATOR CODE
***
The machine will begin purging (high volume ventilation to remove any residual Ethylene
oxide from the cabinet) and the screen will display:
PURGING 05:00
06 BAGS 02 DONE
The counter will count down to 00:00 during the purge, then the machine will unlock the
door to allow bag removal:
A "beep" will sound every 15 seconds to notify the operator that the door may be opened to
remove bags. Once the door is opened the screen will read:
countdown clock
bag numbers of finished bags - view the other finished bags using the arrow keys 34
EOGas™ Series Plus Owner’s Manual
The bag numbers for the bags to be unloaded may be viewed on the screen, if desired.
Cycle through the bags to be unloaded using the UP and DOWN arrows. The door may be
opened and closed as many times as desired while the counter is counting down to 00:00.
After unloading the bags, close the door to the sterilizer. It will lock automatically after the
countdown timer reaches 00:00. The sterilizer will assume that all bags that have
completed the sterilization cycle have been removed and they will no longer be listed as
bags in process.
If the door is still open when the counter reaches 00:00, an alarm will sound to notify the
operator and the screen will display:
Once the door is closed, it locks automatically and the STANDARD Mode normal operating
screen is displayed:
STERILIZING 50.1ºC
04 BAGS 12:07
In this example there are 4 bags remaining in the sterilizer. If the bags removed were the
last bags in the machine, the sterilizer will enter STANDBY and the heaters and fans will
shut down. The screen will display:
35
EOGas™ Series Plus Owner’s Manual
When unloading multiple bags or loads which absorb a large amount of gas, some
operators like to cut open the end of the sterilization bag while it is still in the machine, then
shut the door and repeat the PURGE cycle. This minimizes the chance of exposure to the
residual Ethylene oxide within the bag.
To repeat the purge when unloading, open the door when it is unlocked and cut open the
end of the sterilization bags that will be removed. The display will read:
Close the door and press "1" on the keypad to perform 1 hour additional Aeration, press "2"
to perform 2 hours of additional aeration or "0" to perform 5 min additional aeration. The
display will read:
“1” key
After the counter reaches 00:00 the door will unlock again and the bags may be removed.
The display will read:
After the door has been closed, the machine will return to STERILIZING or STANDBY
depending on whether or not there are bags remaining in the machine.
36
EOGas™ Series Plus Owner’s Manual
LOAD key
If an OPERATOR CODE is required (3 digit code), the machine will prompt the operator for
the code, which should be entered using the keypad, followed by the ENTER key.
OPERATOR CODE
***
ENTER key
The machine will enter a warmup cycle as the cabinet heats up to operating temperature.
WARMUP 34.1ºC
DOOR UNLOCK AT 48ºC
37
EOGas™ Series Plus Owner’s Manual
The door will unlock allowing bags to be loaded after the temperature reaches 2ºC below
the operating temperature. The screen will read:
Once the door is opened to load bags the screen will read:
The door may be opened and closed as many times as desired while the counter is
counting down to 00:00. After loading the bags, close the door to the sterilizer. It will lock
automatically after the countdown timer reaches 00:00.
If the door is still open when the counter reaches 00:00, an alarm will sound to notify the
operator and the screen will display:
Once the door is closed, it locks automatically and the BATCH Mode normal operating
screen is displayed:
Cabinet temperature
STERILIZING 49.4ºC
15:48 FINISH 07:48
To unload the machine, press the UNLOAD key on the front panel.
If an OPERATOR CODE is required (3 digit code), the machine will prompt the operator for
the code, which should be entered using the keypad:
OPERATOR CODE
***
PURGING 05:00
After purging, the door will unlock, allowing bags to be unloaded. The screen will read:
39
EOGas™ Series Plus Owner’s Manual
A "beep" will sound every 15 seconds to notify the operator that the cabinet is ready for
bags to be unloaded. Once the door is opened to unload the bags the screen will read:
The door may be opened and closed as many times as desired while the counter is
counting down to 00:00. After unloading the bags, close the door to the sterilizer. It will
lock automatically after the countdown timer reaches 00:00.
If the door is still open when the counter reaches 00:00, an alarm will sound to notify the
operator and the screen will display:
Once the door is closed, it locks automatically and the BATCH Mode STANDBY screen is
displayed:
40
EOGas™ Series Plus Owner’s Manual
When unloading multiple bags or loads which absorb a large amount of gas, some
operators like to cut open the end of the sterilization bag while it is still in the machine, then
shut the door and repeat the purge cycle. This minimizes the chance of exposure to the
residual ethylene oxide within the bag.
To repeat the purge when unloading, open the door when it is unlocked. Cut open the end
of the sterilization bag that will be removed, with the open end of the bag facing away from
the operator. The display will read:
Close the door and press "1" on the keypad to perform 1 hour additional Aeration, press "2"
to perform 2 hours of additional aeration or "0" to perform 5 min additional aeration. The
display will read:
EXTENDED AERATION
“1” key
After the counter reaches 00:00 the door will unlock again and the bags may be removed.
The display will read:
After the door has been closed, the machine will return to STANDBY.
41
EOGas™ Series Plus Owner’s Manual
The Industrial Mode is designed for customers who do not want the sterilizer to track the
bags or the cycle length. The customer must prepare Standard Operating Procedures
(SOPs) and must validate cycle parameters to sterilize in Industrial Mode. Andersen has a
sterilization cycle validation service to assist in creating the SOPs and cycle parameters if
the customer desires.
In INDUSTRIAL Mode there is no STANDBY function. The heaters and ventilation are on at
all times. The only functions used are LOAD and UNLOAD.
The following is an example of the display for INDUSTRIAL Mode: internal cabinet temperature
INDUSTRIAL 50.1ºC
09:37 29/06/05
24 - hour cycle
Date in day/month/year format
If an OPERATOR CODE is required (3 digit code), the machine will prompt the operator for
the code, which should be entered using the keypad:
OPERATOR CODE
***
PURGING 05:00
42
EOGas™ Series Plus Owner’s Manual
After purging, the door will unlock allowing bags to be loaded or unloaded. The screen will
read:
A "beep" will sound every 15 seconds to notify the operator that the cabinet is ready for
bags to be loaded or unloaded. Once the door is opened to load or unload bags the screen
will read:
The door may be opened and closed as many times as desired while the counter is
counting down to 00:00. After loading or unloading bags, close the door to the sterilizer. It
will lock automatically after the countdown timer reaches 00:00. If the door is still open
when the counter reaches 00:00, an alarm will sound to notify the operator and the screen
will display:
Once the door is closed, it locks automatically and the INDUSTRIAL Mode operating screen
is displayed:
INDUSTRIAL 50.1ºC
09:37 29/06/05
43
EOGas™ Series Plus Owner’s Manual
EMERGENCY UNLOADING
Enter the 5 digit Supervisor Code (14789) then press the ENTER key. The sterilizer will
start purging:
PURGING 05:00
SUPERVISOR UNLOAD
After the counter reaches 00:00, the door will unlock to allow bags to be unloaded:
44
EOGas™ Series Plus Owner’s Manual
when the door is opened, the Supervisor must select the bags to be removed. The number
of the bag will flash on the screen. If the ENTER key is pressed the bag number will stop
flashing, indicating that it has been selected for removal. Cycle through the bags using the
UP and DOWN arrow keys to choose the ones that will be removed. Select each of the
bags by pressing ENTER. If a bag is accidentally chosen, it may be deselected by pressing
ENTER a second time. The bag number will start flashing again indicating it has been de-
selected.
The "01" in the bag number will be flashing. Press the UP arrow twice to get to bag
number 3:
Hit the ENTER key and the "03" will stop flashing. Press the UP arrow two more times to
get to bag 5. Press ENTER to select bag 5.
At this point the bags may be removed from the sterilizer. The bag numbers for the
selected bags will be removed from computer memory, and the cycle will continue normally.
45
EOGas™ Series Plus Owner’s Manual
Enter the 5 digit Supervisor Code (14789) then press the ENTER key. The sterilizer will
start purging:
PURGING 05:00
SUPERVISOR UNLOAD
After the counter reaches 00:00, the door will unlock to allow bags to be unloaded:
The bags may be removed from the sterilizer. The sterilizer will assume that all bags have
been removed from the sterilizer and will return to STANDBY after the door open counter
reaches 00:00.
46
EOGas™ Series Plus Owner’s Manual
Power Outage
During a power outage the door can be opened mechanically. Remove the top cover panel
Once the door is opened the lock will remain lowered and will have to be raised by hand to
close again. (See also p. 68 - "Manual Door Unlock").
47
EOGas™ Series Plus Owner’s Manual
LABEL PRINTER
The label printer is only used for STANDARD Mode operation of the EOGas Series 3 Plus
sterilizer. The printer is attached to the left side of the sterilizer (when facing the front door
of the machine). There are two connections between the sterilizer and label printer, one for
power (3 pin locking ring connector), and the other for data transfer (25 pin connector). Both
of these must be securely connected for proper operation of the label printer.
Once the power is switched on, the light will notify the operator about the status of the
printer:
The label printer will print two copies of each bag identification label after the LOAD key is
pressed and the operator keys in the number of bags that will be loaded. For example, if
four bags will be loaded, four pairs of labels will be printed. One of each pair of labels must
be placed on the outside of the sterilization bag and the other should be placed in a
sterilization logbook.
48
EOGas™ Series Plus Owner’s Manual
49
EOGas™ Series Plus Owner’s Manual
50
EOGas™ Series Plus Owner’s Manual
PANEL-MOUNTED PRINTER
The Panel-Mounted Printer is located on the top front right of the Sterilizer Cabinet, next to
the Control Panel. Printing is automatic every time the UNLOAD key is pressed at the end
of a cycle. Pushing PRINT STATUS prints the same information at any time during the cy-
cle. Regular printouts during the cycle may be obtained by adjusting the printing interval
(PRINT) in the SETUP menu.
BAGS DONE: 04 tells number of bags done then lists the bag numbers
2 01 110205
2 02 110205
2 03 110205
2 04 110205
51
EOGas™ Series Plus Owner’s Manual
UNIT#2 10:43 12/02/05 sterilizer number, current time, current date (day/month/
year format)
START: 14:11 11/02/05 cycle start time, start date
CYCLE LENGTH: 16 hours length of sterilization cycle
PROBLEMS: NONE
possible problems that will be printed if they occur are:
HIGH TEMP ##:## - ##:## start and end times for high temp (up to 2 per cycle)
LOW TEMP ##:## - ##:## start and end times for low temp (up to 2 per cycle)
POWER FAIL ##:## - ##:## start and end times for power failure (up to 3 per cycle)
VENT FAIL ##:## - ##:## start and end times for ventilation failure (up to 2 per cycle)
UNIT#2 12:31 18/07/05 sterilizer number, current time, current date (day/month/
year format)
START 14:11 18/07/05 time and date the door was most recently opened
PROBLEMS: NONE
possible problems that will be printed if they occur are:
HIGH TEMP ##:## - ##:## start and end times for high temp (up to 2 per cycle)
LOW TEMP ##:## - ##:## start and end times for low temp (up to 2 per cycle)
POWER FAIL ##:## - ##:## start and end times for power failure (up to 3 per cycle)
VENT FAIL ##:## - ##:## start and end times for ventilation failure (up to 2 per cycle)
52
EOGas™ Series Plus Owner’s Manual
When the paper is nearing the end of the roll, a red dye will be seen on the edge. The
paper should be changed:
Press the PRINT FEED key on the front panel keypad to expel remaining paper from the
printer. Follow instructions on the printer for removal of the roll, left-hand side first.
53
EOGas™ Series Plus Owner’s Manual
Rotate the new cartridge dial as shown by the arrow to tighten up the ribbon. With the
ribbon facing up, feed the paper between the cartridge and the ribbon. Replace the
cartridge into the printer. Press the FEED button on the front panel keypad until 1 inch (2
cm) of paper is exposed and feed that through the outer cover. Replace the outer cover.
To confirm that the new ribbon is working properly, press the PRINT STATUS button on the
front panel keypad.
54
EOGas™ Series Plus Owner’s Manual
Maintenance
and
Troubleshooting
55
EOGas™ Series Plus Owner’s Manual
ALARMS
Ventilation Alarms
A vacuum transducer mounted in the component module controls the ventilation alarm. A
tube is connected to the main sterilizing cabinet to sense cabinet vacuum. Another tube is
open to room air.
Extraction of exhaust gas ensures a negative pressure in the main cabinet of 0.13 to 0.15
inches of H20 with respect to room pressure.
The vacuum transducer is calibrated by the manufacturer and should not require
readjustment.
56
EOGas™ Series Plus Owner’s Manual
End of Cycle in STNDRD or BATCH Mode: 5 beeps at the end of the cycle
In STANDARD and BATCH Modes, 5 beeps will be emitted at the end of the cycle to alert
the operator that one or more bags are ready to be unloaded.
Power Failure
The machine will keep track of the first 3 power failures during each cycle and will print the
time the power went off and the time it came back on. This information is also printed as
soon as the power comes back on, so if there are more than 3 power failures they may be
identified on the printout even if they are not listed under "PROBLEMS".
57
EOGas™ Series Plus Owner’s Manual
Twice per year confirm the sensitivity of the ventilation alarm and the integrity of the circuit.
Test by disconnecting the power supply to the exhaust extractor from the cabinet
receptacle. (Units with an abator, switch off the exhaust blower). The alarm should operate
within 15 seconds. Reconnect the power supply and the alarm should cease.
All painted and stainless steel finishes may be washed with any available detergent. The
polycarbonate finish of the membrane switch and front panel assembly may be cleaned with
any cleaning agent and is resistant to abrasion and chlorine based cleaners.
and replacing the cover. If there is substantial debris in the filter holder, it may be
necessary to clean it with a vacuum cleaner.
58
EOGas™ Series Plus Owner’s Manual
Maintenance
Maintenance of the Abator is restricted to inspecting the inlet and exhaust ducting, and the
replacement of the catalytic absorber. To replace the absorber material, remove the top
cover panel of the Abator catalyst, remove and replace the absorber material and refit the
cover plate.
The time interval between replacements of the catalyst depends on the amount of Ethylene
oxide used. A period of two years is usually considered appropriate. Sampling the exhaust
air as it leaves the Exhaust Blower should show zero emissions. Any level detected above
zero indicates a need for catalyst replacement.
Exhaust Blower
The Exhaust Blower bearings require oiling with a light machine oil at 6-month intervals,
accompanied by an inspection to confirm the mountings, bearings, etc., are in good working
order.
59
EOGas™ Series Plus Owner’s Manual
FEATURES
No pneumatic cylinders
Self-contained
Easy to maintain
Easy to use
AN5026
Power Usage: 110 V@60Hz or 220V@50Hz Shipping Wt: 392lbs/178kg Dimensions: 32"x29"x45" / 81.28x73.66x114.3cm
60
EOGas™ Series Plus Owner’s Manual
CIRCUIT BREAKERS
The circuit breakers are DIN rail mount type units and are located as shown on Figure 7 of
the Appendix. The circuit breakers are reset with the traditional toggle switch. Access to
reset the circuit breakers is by removing the top panel cover (see Maintenance and
Troubleshooting: Component Module Replacement on p. 69).
61
EOGas™ Series Plus Owner’s Manual
DOOR ADJUSTMENT
Door Gasket Replacement
62
EOGas™ Series Plus Owner’s Manual
BATTERY REPLACEMENT
There is a 3.0 Volt battery that acts as backup power for the computer. When the battery is
low (less than 2.0 Volts), a message will appear on the screen reading:
LOW BATTERY
REPLACE BATTERY
This will be seen for 5 seconds every time "LOAD" is pressed. The machine will still func-
tion normally, and there will be a period of time, probably several months, before the battery
will actually fail. When the message is seen, the operator may continue using the machine,
but as soon as it is convenient the battery should be replaced.
To replace the battery, gently lift the retaining clip on the battery as shown and slide the
battery out of the holder on the circuit board. Reverse the process to insert the new battery.
63
EOGas™ Series Plus Owner’s Manual
FUSE REPLACEMENT
The fuse compartment contains two fuses. On 110/120V models there is one line fuse and
one spare fuse. For 220/240V units both fuses are used as line fuses. To change the fuse,
switch off power. Remove the incoming power cord and pry out the fuseholder cover (see
photos on next page).
A list of Fuses and Circuit Breakers is attached to the back of the sterilizer
Schematic
Device Reference 220.240v Units
CIRCUIT
BREAKERS
Label Printer K9 2A
Exhaust Extractor K10 0.75A
Circulating Fan K11 0.50A
Label Printer K12 0.30A
FUSES
Incoming Power F1
Incoming Power F1/F2 8.0A
64
EOGas™ Series Plus Owner’s Manual
The fuse compartment contains two fuses - both used as line fuses.
To replace, turn off the unit and remove the incoming power cord.
Fuse compartment
Fuse unit after removal.
Replace fuses and ensure
fuseholder is returned to
the fuse compartment with
the arrow marks aligned.
65
EOGas™ Series Plus Owner’s Manual
66
EOGas™ Series Plus Owner’s Manual
67
EOGas™ Series Plus Owner’s Manual
1
2
4
3
68
EOGas™ Series Plus Owner’s Manual
Reinstall in reverse order, test the new component module and pack the replaced module
for shipping.
69
EOGas™ Series Plus Owner’s Manual
Ethylene
Oxide
Safety
70
EOGas™ Series Plus Owner’s Manual
OSHA Regulations
Ethylene oxide has been used as a sterilizing gas for more than sixty years. Throughout
that time there has been a continuous effort by all involved in its use to limit and reduce
the exposure of workers to Ethylene oxide. Almost every country has health and safety
regulations limiting exposure to Ethylene oxide. For many years, the U. S. was the most
prominent user of Ethylene oxide and as a result has clearly defined regulations placed
into legislation by the U. S. Occupational Safety and Health Administration (OSHA).
The major concern for exposure occurs after Ethylene oxide has converted from a liquid to
a gaseous state. As a gas, it is considered a respiratory hazard and is also classified as
poisonous. As with all potentially noxious substances, e.g., chlorine, ammonia, and
alcohol exposure should be kept to the absolute minimum. However, these chemicals are
useful, so methods of measuring exposure and acceptable exposure levels contribute to
safe use. Internationally there are two defined circumstances that are measured: The
Short Term Exposure Limit (STEL) for a fifteen minute period, and the Permissible
Exposure Limit (PEL) over a longer eight hour exposure period.
For either period, the amount of exposure is accumulated and averaged for the time of
exposure. This is defined as the Time-Weighted Average (TWA) and measured in units
of 1 part Ethylene oxide per 1,000,000 parts of air (ppm).
In the U.S., OSHA has established that for the short time STEL the exposure level should
not exceed 5.0 ppm, and for an 8 hour exposure PEL the level should be below 0.5 ppm.
If the exposure levels exceed these, the users have more stringent protection
requirements to meet.
Workplace Testing
Using specifically designed personnel monitor units is universally accepted as a reliable
method of measuring worker exposure to Ethylene oxide. These are designed to be worn
by an individual worker attached like a "badge" close to the person's breathing zone.
These economical devices make it easy to assess any worker's level of exposure to
ethylene oxide. Andersen Products offers a simple to use "badge" under the "AirScan"
label for measuring both the short period STEL and longer monitoring period PEL.
Information regarding their use is included in the Diagrams and Drawings section.
The information given above is a summary. Refer to the relevant health and safety
authority in the country of use.
Precautions
The EOGas cartridge contains liquid and gas under pressure. Do not use near fire, heated
surface or flame. Do not smoke near the sterilizer while loading and unloading it. Sterilize
batteries wrapped individually and separately from their electrical instruments.
71
EOGas™ Series Plus Owner’s Manual
Avoid breathing EOGas vapor. Breathing EOGas vapor is harmful. If you can smell
EOGas, you are breathing toxic amounts. In concentrated amounts EOGas sterilizing gas
is as irritating to the lungs and mucous membranes as is ammonia gas. As with other
chemical vapors, there is a chance of an occasional allergic response to EOGas in a sensi-
tive individual. Such individuals should not handle EOGas, and should neither breathe its
vapors nor allow materials sterilized in it to come into contact with their skin or mucous
membranes.
All users must avoid contact with the skin, eyes, and clothing. If contact with liquid EOGas
occurs, users must immediately remove all contaminated clothing, including shoes. Flush
skin or eyes with plenty of water for at least 15 minutes. If liquid EOGas has gotten into
your eyes, immediately see a physician for further treatment.
Failure to adequately air Ethylene oxide absorbing materials may lead to contact chemical
burns. The 16 hour sterilization cycle time at 50¡C called for in these instructions includes
an adequate amount of time to air most tubes, masks and the like. Large gas absorbent
items like silicone breast implants and plastic extracorporeal blood filters may require
additional aeration before they can be safely implanted or used. Aeration will progress
more rapidly outside of the sterilization bag. Andersen Products is prepared to determine
aeration times for our customers. Call customer service at 336-376-3000 for more
information.
It is incumbent on the user to confirm that a gas absorbent item, sterilized in EOGas has
been adequately aired before it contacts living tissue. This is a particularly important
warning to those using plastic or rubber items that directly contact tissue culture
preparations, microbial cell cultures, spermatocytes, oocytes, embryonic tissue, or the like.
EOGas is a potent polymerizing agent. It must not be used for sterilizing foodstuffs or
drugs.
The user MUST NOT deviate from the Operating Instructions. EOGas cartridges are
designed to be used only in the EOGas sterilization bag provided in the refill kit. An EOGas
sterilizer bag may not be reused. The use of non-genuine EOGas cartridges and sterilizing
bags may compromise sterilizing performance and safety and in that event the
manufacturer will be freed from all liability.
72
EOGas™ Series Plus Owner’s Manual
APPENDIX
73
EOGas™ Series Plus Owner’s Manual
EOGas Sterilizer Operation Reference Chart
For complete instructions, see instruction brochure packaged with the cartridge
WARM - UP PREHUMIDIFICATION
1. Wrap items to be prehumidified
Make sure sterilizer is connected to power. individually as for sterilization.
Sterilizer should come up to temperature in about
15 min (50ºC).
STERILIZER 2. Place wrapped items in medium
(#5), or large (#6) sterilization bag.
WARMUP 3. Mark Dosimeter card with time/
date of sterilization and place in
the core of the load. Include bio-
logical control if appropriate.
4. Match EOGas cartridge (#5, #6)
with number on sterilization bag.
Remove trigger safety and place
EOGas cartridge in sterilization
bag.
5. Heat-seal sterilization bag. DO
PREPARATION OF ITEMS NOT ACTIVATE EOGas CAR-
FOR STERILIZATION TRIDGE.
6. Place sterilization bag in sterilizer
1. Disassemble and wash with detergent and cabinet.
water. 7. Humidify sterilization bag contents
2.
3.
Air-dry; do not oven-dry.
Prehumidify items that cannot be washed.
PREPARATION for 2 hours at 50ºC.
8. Remove sterilization
4. Wrap in paper, cloth, or EO permeable film. bag from sterilizer and
humidify for an additional 2
PRE hours at room temperature
HUMIDIFICATION before sterilization.
2.
large (#6) sterilization bag.
Mark Dosimeter card with time/date of
LOAD BIOLOGICAL ule recommended by your
governing body or when-
CONTROLS
sterilization and place in the sterilization
bag. Include biological control if
LINER BAG ever changing packaging
materials or techniques.
appropriate. Use a Steritest or other
3. Place fresh Humidichip in sterilization bag. appropriate biological
4. Match EOGas cartridge size (#5, #6) with control.
number on sterilization bag. Remove Position control in core of
trigger safety and place EOGas cartridge in load.
sterilization bag.
5. Heat-seal sterilization bag.
AERATION
The 16-hour EOGas
STERILIZATION AND AERATION cycle at 50ºC includes an
AERATION aeration period adequate
1. Leave sterilization bag undisturbed in the for most EO absorbing
sterilizer for 16 hours. STERILIZATION materials. Large gas
2. Sterilization and aeration proceed simulta- absorbent items,
neously during the 16-hour cycle. & AERATION especially implants, or
3. After 16 hours, remove the sterilization bag devices that will contact
from the cabinet. Check the Dosimeter; blood or living tissue,
make sure the blue line has progressed require additional, post
beyond the triangular mark. sterilization aeration.
4. Removed the sterile items from the sterili-
zation bag. Discard the sterilization bag
and used cartridge.
Figure 1 74
EOGas™ Series Plus Owner’s Manual
Series 3 Plus Dimensions
B
29"
10 cu. ft. (280 liters) 39" (737mm)
Inner - 24.5”W(622mm) x 23.5”D(597mm) x 29”H(737mm) (991mm)
Outer - 28.5”W(745mm) x 25.75”D(654mm) x 39”H(991mm)
24.5"
(622mm)
Base
Cabinet
a: Without printer
28.5"
(724mm)
19"
B (483mm)
24.5"
29" (622mm) 23.5"
(737mm) (597mm)
a: Without printer
28.5"
25.75"
(724mm)
(654mm)
b: With printer 34.75" (883mm)
Figure 2
75
Heat Sealer EOGas™ Series Plus Owner’s Manual
Replaceable
Teflon heatstrip
and wire element
30”
41”
20”
Power Usage: 110 VA @ 60Hz 7.5 Amps Wt: 17lbs/7.7kg Dimensions: 30"x20"x41" / 75x50x102.5cmcm
Figure 3
76
EOGas™ Series Plus Owner’s Manual
Vacuum Sealer
Figure 4 77
EOGas™ Series Plus Owner’s Manual
EOGas Installation
Discharge should be
vertical to assist dilution
and dispersion
Ventilation unit
3 inch plastic
pipe to discharge
point
Rain trap
Exterior wall
Plywood panel
mounted to wall
with masonry
screws
Figure 5
78
EOGas™ Series Plus Owner’s Manual
EOGas Installation
Venting to the outside within an interior wall Surface mounted discharge piping
Discharge should
be vertical to assist
dilution and Height to conform
dispersion with local
regulations -3 foot
Mount on sheet minimum above
rock wall using 4 3 inch plastic
roof line
molly type bolts pipe to discharge
point
Rain trap
Ventilation unit
3 inch PVC 90
Clamp and brace pipe as
3 inch PVC 90 necessary for mechanical
degree elbow security and to conform
with local codes
Surface mounting
using U-shaped
Installation is shown inside a typical channel spacer
commercial building with a flat roof,
using surface mounting techniques.
Alternately the vent piping may be
installed within a hollow wall (ie.,
sheetrock). 3 inch PVC plastic
drain pipe is suitable for up to 70
equivalent feet of pipe.
Figure 6
79
EOGas™ Series Plus Owner’s Manual
Figure 7
80
EOGas™ Series Plus Owner’s Manual
EOGas Sterilizer Parts (top)
1 2 3 4 27 6 7 8 9 10
5 12 V 12 V
Light Light
Gear
Motor
28
BELOW! 15 BELOW!
13 BELOW!
13
25 18
24
19
30
11 32
12
Control Circuit 16 17
26 Board
14
29
31
23 21 22 20
Figure 8
EOGas™ Series Plus Owner’s Manual
EOGas Extractor Unit (cover removed)
Figure 9
82
EOGas™ Series Plus Owner’s Manual
Gas Disposal System
Outside
Air
Abator
Blower
EOGas
Sterilizer
Figure 10
83
EOGas™ Series Plus Owner’s Manual
Equipment Ratings Summary
Equipment Rating:
Environmental Conditions:
The EOGas Sterilizers are designed to function in an environment with these conditions:
Temperature.....................................10ºC to 40ºC
Use...................................................Indoor
Relative Humidity............................. 80% to 30%
Main Supply Voltage........................ Not to exceed +/- 10% of nominal voltage
Transient Overvoltages..................... As IEC
Installation Category.........................II
Pollution Degree...............................2
Altitude.............................................Operate up to 2,000 m above sea level
Connections:
Input: Power cord 3 x 16AWG 13Amp to 320 with cabinet power entry EMI Filter
10A 115/250v 50/60Hz
Sound Level:
Operating Condition: (2 feet from unit front panel) 55 dB
Figure 11
84
EOGas™ Series Plus Owner’s Manual
Sterilizer Controls Layout
Label Printer
Power Supply
Cable
Label Printer
Data Cable
Rear
On/Off Switch
Input Power
Receptacle &
Fuseholder
Gas Extractor
Power Receptacle
Figure 12
85
EOGas™ Series Plus Owner’s Manual
Vacuum Transducer Board
Figure 13
86
1 2 3 4 5 6
UPS
Schematic
D D
SW1
LAMP 1 LAMP 2
115VAC
TSW2
D1
CIRCULATION FAN
C1 FUSE1 FUSE2
TSW1
10uF/100V 12A 12A
DC
D2 MOTOR HEATER1 HEATER2
(2 X 600W)
SW2
POWER SUPPLY
RECEPTACLE
+20VDC
POWER SWITCH
LIGHT 1 T1 7
C C
ADAPTOR VACUUM 2 8
BOARD SENSOR 5 11
VERSION BOARD
+12VAC
EPAUA HEATER, PPC
6 12
TEMPERATURE
SENSOR
FAN
K9
K12
K10
K11
2.5A
1.6A
0.5A
0.75A
CONTROLLER
VERSION EPCUBN POWER SUPPLY
(100-240VAC)
LABEL +5VDC
PRINTER +12VDC
ALARM
EOGAS 3 PLUS
CABINET TOP SECTION
A A
SOLENOID DOOR Title
LOCK SENSOR
EOGAS 3 PLUS SYSTEM (115V)
KEYPAD VFD TAPE PRINTER Size Number Revision
B EPSUA
Date: 22-Feb-2006 Sheet 1of 1
File: C:\Hardware\EPSUA.DDB Drawn By: L. Lee
1 2 3 4 5 6
120V Figure 14
87
EOGas™ Series Plus Owner’s Manual
1 2 3 4 5 6
Schematic
UPS
TSW = THERMO SWITCH
BLOWER PLUG
PPC = PROPORTIONAL POWER CONTROLLER
D D
SW1
LAMP 1 LAMP 2
230VAC
TSW2
D1
CIRCULATION FAN
C1 FUSE1 FUSE2
TSW1
10uF/100V 6.3A 6.3A
DC
D2 MOTOR HEATER1 HEATER2
(2 X 600W)
SW2
POWER SUPPLY
RECEPTACLE
+20VDC
POWER SWITCH
LIGHT 1 T1 7
C C
ADAPTOR VACUUM 2 8
BOARD SENSOR 5 11
VERSION
+12VAC
BOARD
EPAUA HEATER, PPC
6 12
TEMPERATURE
SENSOR
FAN
K9
K12
K10
K11
0.5A
2.5A
0.75A
0.3A
CONTROLLER
VERSION EPCUBN POWER SUPPLY
(100-240VAC)
LABEL +5VDC
PRINTER +12VDC
ALARM
EOGAS 3 PLUS
CABINET TOP SECTION
A A
SOLENOID DOOR Title
LOCK SENSOR
EOGAS 3 PLUS/AN333 SYSTEM (230V)
KEYPAD VFD TAPE PRINTER Size Number Revision
B EPSEA
Date: 22-Feb-2006 Sheet 1of 1
File: C:\Hardware\EPSUA.DDB Drawn By: L. Lee
1 2 3 4 5 6
240V Figure 15
88
EOGas™ Series Plus Owner’s Manual
EOGas™ Series Plus Owner’s Manual
Figure 16
89
EOGas™ Series Plus Owner’s Manual
4
~ IEC 617, No. 02-02-06 Three-phase Alternating Current
3
5 IEC 417, No. 5017 Earth (ground) TERMINAL
12 Background color - yellow; ISO 3864, No. B.3.6 CAUTION, risk of electric shock
Symbol and outline - black
Figure 17
90
EOGas™ Series Plus Owner’s Manual
Mechanical:
Electrical:
1. Check the operation of the circulating fan for bearing wear or binding
2. Check the proximity switch stops the circulating fan with the door open
3. Check the ventilation alarm circuit by switching off the exhaust fan
4. Isolate each circuit by opening the circuit breakers then restore
5. Disconnect power to the UPS, confirm back up capacity time
6. Check calibration setting of cabinet temperature at shelf point
Replace:
Calibrate:
1. Temperature of Cabinet (see page 23)
Figure 18
91
EOGas™ Series Plus Owner’s Manual
Appendix A
Material Safety Data Sheets—Ethylene Oxide
AUTOIGNITION TEMPERATURE:
804 F (429C); burns in the absence of air ENGINEERING CONTROLS: Ethylene oxide, a major
fire hazard, can burn in the absence of oxygen. All
EXTINGUISHING MEDIA: Carbon dioxide, dry chemi- electrical devices used in areas processing or handling
cal or water spray for small fires. Water spray, polymer ethylene oxide must be engineered and designed to the
or alcohol resistant foams for large fires. Dilution of applicable local electrical/fire codes. Safeguards can
liquid ethylene oxide with 23 volumes of water should include designing electrical devices as explosion proof
render it non-flammable. Dilution with 100 parts water and/or intrinsically safe.
to one part of ethylene oxide vapor may be required to
control build up of flammable vapors in closed systems. ATTENTION: Ethylene oxide vapors are colorless and
Water spray can be used to reduce intensity of flames odorless above OSHA’S permissible exposure level. An
to cool fire-exposed containers and to dilute spills to air monitoring system and/or AirScan® badges are rec-
render non-flammable. ommended to determine airborne exposure levels.
94
EOGas™ Series Plus Owner’s Manual
95
EOGas™ Series Plus Owner’s Manual
96
EOGas™ Series Plus Owner’s Manual
Section 14 NESHAPS - National Emission Standards for Hazardous
Regulatory Information Air Pollutants
NFPA - National Fire Protection Association
U.S. REGULATIONS: NIOSH - National Institute of Occupational Safety and
Health
TSCA status: Listed NTP - National Toxicology Program
CERCLA Section 103 (40 CFR 302.4): listed OSHA - Occupational Safety and Health Administration
10 lb. Reportable Quantity p/p - parts per part
SARA Section 304 (40 CFR 356.40): Listed PEL - Permissible exposure Limit
1 lb Reportable Quantity PVC - Polyvinyl chloride
SARA Section 311/312 (40 CFR 370.21) Hazard catego ppm - Parts per million
ries met: p.s.i.g - Pounds per square inch (gauge pressure)
Acute, Chronic, Fire, Reactive, Sudden Release RCRA - Resource, Conservation and Recovery Act
SARA Section 313 (40 CFR 372.65); Listed SARA - Superfund Amendment and Reauthorization Act
OSHA (29 CFR 1910. 1200): Meets criteria as a hazard- of 1990
ous material STEL- Short-term exposure Limit
OSHA (29 CFR 1910. 1047): Ethylene Oxide Standard TDG -Transportation of Dangerous Goods
EPA list of Hazardous Air Contaminants: Listed TLV - Threshold Limit Value
EPA Organic Hazardous Air Pollutant (HAP) list: Listed TSCA - Toxic Substance Control Act
EPA list of Pesticide Chemicals (40 CFR 180.151): TWA - Time Weighted Average
Listed VOC - Volatile organic compound
EPA NESHAPS (40 CFR 63.360) WHMIS -Workplace Hazardous Material Information Sys-
VOC Rule: 100% VOC tem
Andersen Products, Inc. (“Andersen”) warrants that this EOGas® Sterilizer (“Product”) is free from defects in ma-
terial and workmanship that result in Product failure during normal usage, according to the following terms and
conditions:
1. The limited warranty for the Product extends for ONE (1) year beginning on the date of the purchase of the
Product.
2. The limited warranty extends to the original purchaser of the Product (“Consumer”) and is not assignable or
transferable to any subsequent purchaser/end-user.
3. The limited warranty extends only to Consumers who purchase the Product in the United States, Canada and
Japan, Mexico, Central and South America and the Caribbean.
4. During the limited warranty period, Andersen will repair, or replace, at Andersen's option, any defective parts, or
any parts that will not properly operate for their intended use with new or factory rebuilt replacement items if such
repair or replacement is needed because of product malfunction or failure during normal usage. No charge will be
made to the Consumer for any such parts. Andersen will also pay for the labor charges incurred by Andersen in
repairing or replacing the defective parts. The limited warranty does not cover defects in appearance that are cos-
metic or decorative. Andersen's limit of liability under the limited warranty shall be the actual cash value of the
Product at the time the Consumer returns the Product for repair, determined by the price paid by the Consumer for
the Product less a reasonable amount for usage. Andersen shall not be liable for any other losses or damages.
These remedies are the Consumer’s exclusive remedies for breach of warranty.
5. Upon request from Andersen, the Consumer must prove the date of the original purchase of the Product by a
dated bill of sale or dated itemized receipt.
6. The Consumer shall bear the cost of shipping the Product, or any part thereof, to Andersen in Haw River,
North Carolina. Andersen shall bear the cost of shipping the Product, or any part thereof, back to the Consumer
after the completion of service under this limited warranty.
7. The Consumer shall have no coverage or benefits under this limited warranty if any of the following conditions
are applicable:
a) The Product has been subject to abnormal use, abnormal conditions, improper storage, exposure to moisture
or dampness, unauthorized modifications, unauthorized connections, un-authorized repair, misuse, neglect,
abuse, accident, alteration, improper installation, or other acts which are not the fault of Andersen, including
damage caused by shipping.
b) The Product has been damaged from external causes such as collision with an object, or from fire, flooding,
sand, dirt, windstorm, lightning, earthquake or damage from exposure to weather conditions, an Act of God, or
theft, blown fuse, or improper use of any electrical source, or damage caused by the connection to other prod-
ucts not recommended for interconnection by Andersen.
c) Andersen was not advised by the Consumer in writing of the alleged defect or malfunction of the Product
within fourteen (14) days after the expiration of the applicable limited warranty period.
d) The Product serial number plate has been removed, defaced or altered.
8. If a problem develops during the limited warranty period, the Consumer shall take the following step-by-step
procedure:
a)The Consumer shall contact the place of purchase to describe the alleged defect or malfunction of the Product
and obtain a Return Authorization Number (“RMA”)
b) The Consumer shall return the Product to the place of purchase for repair or replacement processing.
c) If “b” is not convenient because of distance (more than 100 miles) or for other good cause, the Consumer
shall ship the Product prepaid and insured to:
Andersen, Products, Inc.
Attn: Repair Department
3202 Caroline Drive
Health Science Park
Haw River, NC 27258 USA
d) The Consumer shall include a return address, daytime phone number and/or fax number, complete descrip-
tion of the problem, proof of purchase and service agreement (if applicable). Expenses related to removing the
Product from an installation are not covered under this limited warranty.
e) The Consumer will be billed for any parts or labor charges not covered by this limited warranty. The Con-
sumer will be responsible for any expenses related to reinstallation of the Product.
f) Andersen will repair or authorize the repair of the Product under the limited warranty within 15 days after re-
ceipt of the Product by Andersen or an Andersen authorized service center. If Andersen cannot perform repair
covered under this limited warranty within 15 days, or after a reasonable number of attempts to repair the same
defect, Andersen at its option, will provide a replacement Product or refund the purchase price of the Product
less a reasonable amount for usage.
98
EOGas™ Series Plus Owner’s Manual
g) If the Product is returned to Andersen during the limited warranty period, but the problem with the Product is not
covered under the terms and conditions of this limited warranty, the Consumer will be notified and given an esti-
mate of the charges the Consumer must pay to have the Product repaired, with all shipping charges billed to the
Consumer. If the estimate is refused, the Product will be returned freight collect. If the Product is returned to An-
dersen after the expiration of the limited warranty period, Andersen's normal service policies shall apply and the
Consumer will be responsible for all shipping charges.
9. The Product consists of newly assembled equipment that may contain used components that have been re-
processed to allow machine compliance with Product performance and reliability specifications.
10. ANY IMPLIED WARRANTY OF MERCHANTABILITY, OR FITNESS FOR A PARTICULAR PURPOSE OR
USE, SHALL BE LIMITED TO THE DURATION OF THE FOREGOING LIMITED WRITTEN WARRANTY. OTH-
ERWISE, THE FOREGOING LIMITED WARRANTY IS THE CONSUMER'S SOLE AND EXCLUSIVE REMEDY
AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. ANDERSEN SHALL NOT BE LIABLE
FOR
SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING BUT NOT LIMITED TO, LOSS OF AN-
TICIPATED BENEFITS OR PROFITS, LOSS OF SAVINGS OR REVENUE, PUNITIVE DAMAGES, LOSS OF
USE OF THE PRODUCT OR ANY ASSOCIATED EQUIPMENT, COST OF CAPITAL, COST OF ANY SUBSTI-
TUTE EQUIPMENT OR FACILITIES, DOWN-TIME, THE CLAIMS OF ANY THIRD PARTIES, INCLUDING
CUSTOMERS, AND INJURY TO PROPERTY, RESULTING FROM THE PURCHASE OR USE OF THE PROD-
UCT OR ARISING FROM BREACH OF THE WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, STRICT
TORT, OR ANY OTHER LEGAL OR EQUITABLE THEORY, EVEN IF ANDERSEN KNEW OF THE LIKELIHOOD
OF SUCH DAMAGES. ANDERSEN SHALL NOT BE LIABLE FOR DELAY IN RENDERING SERVICE UNDER
THE LIMITED WARRANTY, OR LOSS OF USE DURING THE PERIOD THAT THE PRODUCT IS BEING RE-
PAIRED.
11. Some countries do not allow limitation of how long an implied warranty lasts, so the above one-year warranty
limitation may not apply to you (the Consumer). Some countries do not allow the exclusion or limitation of inciden-
tal and consequential damages, so certain of the above limitations or exclusions may not apply to you (the Con-
sumer). This limited warranty gives the Consumer specific legal rights and the Consumer may also have other
rights which vary from country to country.
12. Andersen neither assumes nor authorizes any authorized service center or any other person or entity to as-
sume for it any other obligation or liability beyond that which is expressly provided for in this limited warranty in-
cluding the provider or seller of any extended warranty or service agreement.
13. This is the entire warranty between Andersen and the Consumer, and supersedes all prior and contemporane-
ous agreements or understandings, oral or written, and all communications relating to the Product, and no repre-
sentation, promise or condition not contained herein shall modify these terms.
14. This limited warranty allocates the risk of failure of the Product between the Consumer and Andersen. The
allocation is recognized by the Consumer and is reflected in the purchase price of the Product.
15. Any action or lawsuit for breach of warranty must be commenced within eighteen (18) months following deliv-
ery of the Product to the Consumer.
16. Questions concerning this limited warranty may be directed to:
Andersen Customer Care Center
Andersen Products, Inc.
3202 Caroline Drive
Health Science Park
Haw River, NC 27258
Telephone: 1-800-523-1276
Facsimile: (336) 376-8153
E-mail: customerservice@anpro.com
17. The limited warranty period for Andersen supplied attachments and accessories is specifically defined within
their own warranty cards and packaging.
99