You are on page 1of 10

Analytica Chimica Acta 442 (2001) 327–336

Simultaneous spectrophotometric determination of chromium(VI)


and iron(III) with chromogenic mixed reagents by H-point
standard addition method and partial least squares regression
H. Abdollahi∗
Department of Chemistry, Institute for Advanced Studies in Basic Science, Zanjan 45195-159, Iran
Received 12 March 2001; accepted 30 May 2001
Dedicated to my best teacher professor Afsaneh Safavi.

Abstract
Simultaneous determination of Cr(VI) and Fe(III) by partial least squares (PLS) calibration and H-point standard addition
method (HPSAM) is described. Mixed reagents of diphenylcarbazide and 1,10-phenanthroline in a non-ionic micellar solution
of Triton X-100 was used as a selective chromogenic system for determination of Cr(VI) and Fe(III). The presence of a micellar
system allows one to eliminate the previous solvent extraction step that is necessary for the determination of slightly soluble
metal complexes in the absence of micelles. This reduces the cost and toxicity of the method. The total relative standard
error for applying the PLS method to 15 synthetic samples in the ranges 0.20–6.00 ␮g ml−1 Cr(VI) and 0.20–8.00 ␮g ml−1
was 1.5%. The results of applying the H-point standard addition method showed that Cr(VI) and Fe(III) can be determined
simultaneously with weight concentration ratios of Cr(VI) to Fe(III) varying from 15:1 to 1:30 in the mixed sample. Both
proposed methods (PLS and HPSAM) were also successfully applied to the determination of Cr and Fe in several synthetic
alloy solutions. © 2001 Published by Elsevier Science B.V.
Keywords: Mixed chromogenic reagents; Partial least squares; H-point standard addition method

1. Introduction present in them. Although, the determination of


chromium and iron has been studied in different sam-
The need for chromium and iron analysis in envi- ples and by using different techniques [11,12], some
ronmental and biomedical material has increased after of these methods require physical separation, pre-
reports on the different biological roles of Cr(III), liminary treatment or non-universal instrumentation.
Cr(VI), Fe(III) and Fe(II) species in plants, ani- Therefore, sensitive and selective methods for simul-
mals and humans. Several methods can be used for taneous determination of Cr(VI) and Fe(III) are very
chromium [1–8] and iron [9,10] determination. More- desirable.
over, chromium and iron are metals, which appear Mixed organic reagents have been proposed [13–16]
together in very real samples, both natural and arti- for spectrophotometric multicomponent analysis of
ficial. In most instances, the characterization of these metal ions. The limitations on the use of such systems
samples includes the determination of metal ions usually arise from the necessity for finding a com-
promise condition with respect to the different metal
∗ Tel.: +98-241-449-212; fax: +98-241-449-023. complexes. In spite of this, the use of mixed reagent
E-mail address: abd@iasbs.ac.ir (H. Abdollahi). systems can be considered as an alternative means to

0003-2670/01/$ – see front matter © 2001 Published by Elsevier Science B.V.


PII: S 0 0 0 3 - 2 6 7 0 ( 0 1 ) 0 1 1 7 4 - 6
328 H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336

extend the applicability of spectrophotometric multi- signal due to the one of the species (X) is constant
component determination. and for another one (Y) to be as be different as possi-
The organized molecular assemblies such as mi- ble. By plotting the analytical signal versus the added
celles are used in spectroscopic measurements due to Y concentration, two straight lines are obtained that
their possible effects on the systems of interest. In the have a common point with coordinates H(–CH , AH ),
field of metal ion complex, at a concentration above where –CH is the unknown Y concentration and AH
the critical micelle concentration (cmc) micelles form the analytical signal due to the X species. Absorbency
a ternary complex with advantageous properties, such increments as analytical signals were used when only
as hyperchromic and bathochromic displacements, the Y concentration was required [28]. The method
that can modify the sensitivity of the method by af- has been applied to eliminate the blank bias error due
fecting the interferences and matrix effects [17,18]. to the use of absorbent blank [29,30], in liquid chro-
The ability of micellar systems to solublize slightly matography [31] and to the analysis of kinetic data
soluble or even insoluble complexes and/or ligands [32,33], with time as an additional variable.
has been used to enhance the analytical merit of given This paper reports the simultaneous spectrophoto-
methods [19,20]. Most organic solvents that are used metric determination of chromium(VI) and iron(III)
as extraction solvents can be classified as toxic and with the use of mixed organic reagents in a Triton
environmental pollutants [21], and some have been X-100 micellar medium. Partial least squares regres-
listed as carcinogenic by the US Environmental Pro- sion and the H-point standard addition method were
tection Agency (EPA) [22]. Application of micellar employed for the resolution of binary mixtures of
systems avoids the previous step of solvent extraction Cr(VI) and Fe(III). The suggested methods were ap-
necessary following the formation of slightly soluble plied for the analysis of synthetic mixtures and artifi-
complexes in the absence of the micelles. cial mixture solutions of several alloys.
Partial least squares (PLS) modeling is a powerful
multivariate statistical tool that has been successfully
applied to the quantitative analysis of spectroscopic 2. Experimental
[23,24] and electrochemical [25] data. PLS is capable
of being a full-spectrum method and therefore enjoys 2.1. Reagents
the signal averaging advantages of other full-spectrum
methods such as principal component regression A 100 ml stock solution of mixtures of 1,10-phena-
(PCR) and classical least squares (CLS). PLS also nthrolinium chloride and diphenylcarbazide was
has characteristics and advantages of the inverse least prepared by dissolving 0.998 and 0.575 g of these
squares (ILS) method, which is limited in the number reagents, respectively, in 5.0 ml of acetonitrile, adding
of spectral frequencies that can be included in the 4 g of Triton X-100 (Merck) and adjusting the volume
analysis. [24] to 100 ml with triply distilled water. All standards
The H-point standard addition method (HPSAM) and prediction samples were 0.025 M with respect
is a modification of the standard addition method that to 1,10-phenanthrolinium chloride and diphenylcar-
transforms the incorrigible error resulting from the bazide. An excess concentration of reagents has been
presence of a direct interference in the determination chosen to ensure quantitative oxidation and subse-
of an analyte into a constant systematic error. This er- quent formation of the complexes in the whole the cal-
ror, can then be evaluated and eliminated. This method ibration region for Cr(VI) and Fe(III). A 500 ␮g ml−1
also permits both proportional and constant errors solution of chromium and iron ions was prepared from
produced by the matrix of the sample to be corrected their appropriate salts, K2 Cr2 O7 and Fe(NO3 )3 ·9H2 O
directly. The basis of the method was established (Merck). All solutions were prepared fresh daily.
previously [26,27]. This new technique was based on
the principle of dual wavelength spectrophotometry 2.2. Apparatus
and the standard addition method. The requirements
for the application of the method is that if necessary Digitized UV–VIS absorbency spectra were col-
to work only at two wavelengths where the analytical lected using a CARY 5 spectrophotometer, with 1 cm
H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336 329

quartz cells, a scan rate of 900 nm min−1 and a slit for adjusting the final pH to about 2.0 and appropriate
width of 2 mm. All spectral measurements were per- amounts of desired metal ion solution were added to
formed using the blank solution as a reference. The 10 ml volumetric flask and made up to the mark with
recorded spectra were digitized with two data points triply distilled water. For each measurement, about
per nanometer. Measurements of pH were made with a 2 ml of the above solution was transferred to a spec-
Metrohm 713 pH-meter using a combined glass elec- trophotometer cell and absorbencies were measured at
trode. The computations were made with a Pentium II appropriate wavelengths. Synthetic samples contain-
computer. All the programs in the computing process ing different concentration ratios of Cr(VI) and Fe(III)
were written in MATLAB for Windows. were prepared, and standard additions of Cr(VI) (up
to 6.00 ␮g ml−1 ) were made. Simultaneous determina-
2.3. Procedures tion of Cr(VI) and Fe(III) with HPSAM was performed
by measuring the absorbencies at 490 and 510 nm
2.3.1. Individual calibration for each sample solution. The concentration ranges of
In two 25 ml volumetric flasks, a 150 ␮g ml−1 so- Cr(VI) and Fe(III) for construction of the HPSA cali-
lution each of chromium and iron was prepared from bration graph were 0.20–6.00 and 0.20–8.00 ␮g ml−1 ,
their stock solutions. For preparation of each standard respectively.
sample solution, 5 ml of micellar mixed reagents, 1 ml
of 0.05 M HCl for adjusting the final pH to about
2.0 and appropriate amounts of considered metal ion 3. Results and discussion
solution were added to a 10 ml volumetric flask and
made up to the mark with triply distilled water. The Diphenylcarbazide and 1,10-phenanthroline are
final concentrations of chromium and iron must be be- commonly used for spectrophotometric determina-
tween 0.20–6.00 and 0.2–8.00 ␮g ml−1 , respectively. tion of Cr(VI) and Fe(II), respectively. In a solu-
The absorbencies were measured at 538 and 510 nm, tion containing a mixture of these two reagents, the
against a reagent blank, for chromium and iron ions, Cr(VI) and Fe(III) reactions occurring seems to be
respectively. the simultaneous oxidation of diphenylcarbazide to
diphenylcarbazone, reduction of Cr(VI) and Fe(III) to
2.3.2. PLS calibration Cr(III) and Fe(II) and chelate formation of diphenyl-
Each standard, prediction and synthetic mixture was carbazone with Cr(III) and 1,10-phenanthroline with
prepared as follows: 5 ml of micellar mixed reagents Fe(II) [34]. Therefore, the system seems to be appro-
solution, 1 ml of 0.05 M HCl to adjust the pH to about priate for simultaneous indirect spectrophotometric
2.0 and appropriate amounts of each metal (from the determination of Cr(VI) and Fe(III).
150 ␮g ml−1 solution of metal ion) were added to The influence of pH values on the spectrum of each
a 10 ml volumetric flask and made up to the mark complex at a constant concentration was investigated
with triply distilled water. The concentration range separately. Although, no significant changes were ob-
in these mixtures was 0.20–6.00 ␮g ml−1 for Cr(VI) served in the spectrum of the Fe–1,10-phenanthroline
and 0.20–8.00 ␮g ml−1 for Fe(III). The concentration complex in the pH range 2–6, the spectrum of the
ranges were chosen so that the absorbencies obtained chromium complex shows a decrease in maximum
for all standard samples were > / 1.5. The composi- absorbencies at pH > 3, so, for achieving higher
tions of the samples were randomly designed in order sensitivity and stability, pH 2 was selected as op-
to obtain maximum information on each ion from the timum for simultaneous determination of Cr(VI)
calibration procedure. The artificial sample solutions and Fe(III).
were prepared according to the reported composition
of some alloys. 3.1. Individual calibrations

2.3.3. H-point standard addition analysis To verify the governing Beer’s law, four calibra-
For preparation of each standard sample solution, tion graphs were constructed over the whole range of
5 ml of micellar mixed reagents, 1 ml of 0.05 M HCl wavelengths (in 10 nm increments). These calibration
330 H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336

Fig. 1. Individual calibration graphs for Fe at 510 nm (r = 0.9989) and Cr at 538 nm (r = 0.9996).

graphs were prepared for Cr(VI) and Fe(II), separately selection of appropriate wavelengths for applying
and in the presence of a fixed amount of each other. HPSAM, the following principles were followed. At
The appropriate correlation coefficients obtained (in the two selected wavelengths, the signal of X species
the range 0.9988–0.9999) indicate that the interaction must remain equal, even if the concentrations of Y
between the two complexing systems either does not species are changed, and the analytical signals of the
exist or at least does not affect the linear correlation mixture composed from the X and the Y should be
prevailing between absorbancy and concentration of equal to the sum of the individual signals of the two
each ion. Thus, chemometric methods based on factor species. In addition, the slope difference of the two
analysis such as PLS seem to be suitable for use in straight lines obtained at ␭1 and ␭2 must be as large
this system. as possible in order to get good accuracy.
Individual calibration graphs were constructed with When Cr(VI) is selected as the analyte (Y species),
several points (Fig. 1), as absorbancy versus metal ion it is possible to select several pairs of wavelengths
concentration in the range 0.20–6.00 for Cr(VI) and where they present the same absorbancy for the Fe
0.20–8.00 ␮g ml−1 for Fe(III) and evaluated by linear complex. After this, the selected wavelength pairs
regression. The intercepts on the ordinates were negli- were sorted following the criteria to give high values
gible in the two graphs. Limits of detection, calculated for the difference of the calibration slopes. The equa-
as 3s0 per slope where s0 is the standard deviation of tion that is used for calculation of CH can be seen as
the intercept on the ordinate were Cr (0.08 ␮g ml−1 ) a ratio between an absorbancy increment (A) and a
and Fe (0.05 ␮g ml−1 ). slope increment (M). M depends on the analyte
absorption characteristics, and also A depends on
3.2. H-point standard addition method (HPSAM) the sample analyte concentration. As shown previ-
ously by Compins-Falco et al. [35], the higher the
Fig. 2 shows the absorption spectra for the indi- value for the slope increment the smaller the error
vidual metal complexes and a mixture of them. For for the analyte concentration. So, the wavelength
H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336 331

Fig. 2. Absorption spectra of: (a) 1.00 ␮g ml−1 Cr(VI); (b) 2.00 ␮g ml−1 Fe(III); and (c) 1.00 ␮g ml−1 Cr(VI) + 2.00 ␮g ml−1 Fe(III) with
the mixed reagents at pH 2.

pair value that gives the greatest slope increment show the effect of change in concentration of Cr(VI)
was selected, for this reason the wavelength pair and Fe(III) on the position of the H-point.
490–520 nm were employed for obtaining the best Several synthetic mixed samples with different
accuracy. Also, according to the characteristics of concentration ratios of Cr(VI) and Fe(III) were ana-
HPSAM at the H-point, CH (concentration of Cr(IV)) lyzed by the suggested method. The concentration of
is independent of the concentration of another consid- Fe(III) was calculated in each test solution by employ-
ered species (Fe(III)) and so AH (absorbance corre- ing the calibration line obtained from ordinate values
sponding to Fe(III) concentration) also is independent of H-point for standard binary solutions. As can be
of the Cr(VI) concentration. Figs. 3 and 4 clearly seen from Table 1, the accuracy of each result is

Table 1
Results of several experiments for the analysis of Cr(VI)–Fe(III) mixtures in different concentration ratios
A–C equation r Sample (␮g ml−1 ) Found (␮g ml−1 )
Cr(VI) Fe(III) Cr(VI) Fe(III)

A520 = 0.240 Ci + 0.110 0.9997 0.25 0.50 0.22 0.48


A490 = 0.111 Ci + 0.082 0.9998
A520 = 0.230 Ci + 0.537 0.9996 1.45 1.00 1.47 1.00
A490 = 0.113 Ci + 0.365 0.9982
A520 = 0.231 Ci + 1.029 0.9995 0.24 4.88 0.21 4.85
A540 = 0.111 Ci + 1.003 0.9993
A520 = 0.235 Ci + 1.320 0.9996 0.72 5.83 0.68 5.86
A490 = 0.111 Ci + 1.236 0.9995
A520 = 0.236 Ci + 0.758 0.9997 3.00 0.20 3.05 0.18
A490 = 0.114 Ci + 0.386 0.9997
A520 = 0.232 Ci + 0.429 0.9996 1.00 1.00 0.99 1.00
A490 = 0.109 Ci + 0.307 0.9995
332 H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336

satisfactory when the concentration ratio of Cr(VI) and which are the samples not participating in the con-
Fe(III) varies from 15:1 to 1:30. struction of the model. This error was expressed as
the prediction residual error sum of squares (PRESS).
3.3. Partial least squares (PLS) method PRESS was calculated for the first variable, which
built the PLS modeling in the calibration step. After
The first step in the simultaneous determination of that, another latent variable was added for the model
different metal ions by PLS methodology involves building and the PRESS was calculated again. These
constructing the calibration matrix for the binary calculations were repeated for one to 15 latent vari-
mixture Cr–Fe. A number of 20 binary mixtures were ables, which were used in the PLS modeling. This
selected as the calibration set. Their composition was procedure was repeated for each element. Fig. 5 shows
randomly designed for obtaining more information a plot of the PRESS against the number of factors for
from the calibration procedure. Under these con- each individual component. For finding the smallest
ditions, the calibration models were obtained. The model (fewest number of factors), the F-statistic was
obtained model was validated with a 15 synthetic used to carry out the significance determination [24].
mixture set containing the considered metal ions in The PLS modeling for each element had a different
different proportions that were randomly selected. To number of factors. The optimal number of factors for
select the number of factors in the PLS algorithm, a Cr and Fe was obtained as three and four, respec-
cross-validation method, leaving out one sample at a tively. The results obtained are given in Table 2. The
time, was employed [24]. The prediction error was prediction error of a single component in the mixture
calculated for each metal ion for the prediction set, was calculated as the relative standard error (R.S.E.)

Fig. 3. Plots of H-point standard addition method for fixed Fe(III) concentration (5.83 ␮g ml−1 ) and different concentrations of Cr(VI).
H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336 333

Fig. 4. Plots of H-point standard addition method for fixed Cr(VI) concentration (0.24 ␮g ml−1 ) and different concentrations of Fe(III).

of the prediction concentration [36,37]: 3.4. Interference analysis


 N 2
1/2
j =1 (Ĉj − Cj ) For analyzing interference, the influence of several
R.S.E. (%) = N × 100 (1) ions was tested, including those that most frequently
2
j =1 (Cj )
accompany Cr(VI) and Fe(III) in real samples. The
where N is the number of samples, Cj the concentra- effect of interfering ions at different concentrations
tion of the component in the jth mixture and Ĉj the on the absorbancy of a solution containing 1 ␮g ml−1
estimated concentration. The total prediction error of of each analyte was studied. An ion was considered
N samples is calculated as follows: to interfere when its presence produced a variation in
the absorbency of the sample >5%. This increment of
 M N 2
1/2 absorbency was evaluated at two wavelengths, corre-
i=1 j =1 (Ĉij − Cij )
R.S.E.t (%) = M N sponding to the maximum absorption of Cr and Fe
2
i= j =1 (Cij ) complexes, to establish the different effects of the in-
×100 (2) terfering ions on each analyte. Among the interfer-
ing ion tested CO3 2− , CN− , F− , I− , Br− , CH3 COO− ,
where Cij is the concentration of the ith component NO3 − , Ba2+ , Mg2+ , Ca2+ , Li+ , Na+ , Zn2+ , Bi3+ ,
in the jth sample and Ĉij its estimation. Table 2 also Hg2+ , In3+ , Ni2+ , Co2+ , Cu2+ , Cd2+ , Cr3+ , Se(IV),
shows reasonable single and total relative errors for Sn(II) and Al3+ did not interfere of concentrations
such a system. Fig. 6 shows the recovery results ob- 1000 times higher than those of the analytes. However,
tained by the application of the PLS model on the V(IV) and Mo(VI) showed interferences at nearly 30
spectral data of the prediction set. times concentration level.
334 H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336

Fig. 5. Plot of PRESS against the number of factors for Cr(VI) and Fe(III).

Table 2
Composition of synthetic samples, their predictions by PLS model and statistical parameters for the system
Synthetic (␮g ml−1 ) Prediction (␮g ml−1 ) Recovery (%)

Cr(VI) Fe(III) Cr(VI) Fe(III) Cr(VI) Fe(III)


4.80 0.90 4.85 0.98 101.0 108.9
2.06 3.54 2.05 3.45 99.5 97.5
5.91 4.69 5.95 4.78 100.7 101.9
2.95 5.14 2.98 5.27 101.0 102.5
2.42 5.34 2.42 5.25 100.0 98.3
5.48 0.31 5.41 0.39 98.7 125.8
5.61 4.82 5.55 4.73 98.9 98.1
3.70 2.77 3.75 2.70 101.4 97.5
2.47 9.78 2.55 9.86 103.2 100.8
4.70 1.01 4.66 0.96 99.1 95.0
0.26 5.75 0.23 5.68 88.5 98.8
1.61 5.38 1.63 5.30 101.2 98.5
4.42 1.77 4.44 1.71 100.5 96.6
4.97 7.88 5.05 7.93 101.6 100.6
0.72 4.32 0.77 4.26 106.9 98.6
Mean recovery 100.2 101.3
R.S.E. (%)a single 1.2 1.6
R.S.E. (%)b total 1.5 1.5
a Calculated by Eq. (1).
b Calculated by Eq. (2).
H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336 335

Fig. 6. Diagrammatic representation of the percentage recovery found in the analysis of the 15 mixtures of Cr(VI) and Fe(III) by PLS model.

Table 3
Actual composition and calculated concentration of iron and chromium in synthetic mixtures of some alloys
Alloy Composition of synthetic mixture (␮g ml−1 ) Founda (␮g ml−1 )

PLS HPSAM
Fe Cr Fe Cr

Nichrome wire Fe (1.20), Cr (1.10), Ni (7.50), Mn (0.20) 1.24 ± 0.04 1.06 ± 0.04 1.18 ± 0.03 1.08 ± 0.05
Nichrome Fe (2.40), Cr (1.60), Ni (6.00) 2.31 ± 0.06 1.54 ± 0.04 2.45 ± 0.05 1.54 ± 0.04
Chromax castings Fe (5.00), Cr (1.50), Ni (3.50) 4.94 ± 0.03 1.52 ± 0.04 5.05 ± 0.06 1.52 ± 0.03
Cristite Fe (6.60), Cr (1.00), W (1.70), C (0.35), Mo (0.25) 6.68 ± 0.04 0.96 ± 0.05 6.54 ± 0.05 1.05 ± 0.04
a Mean ± S.D. (four replicates).

3.5. Analysis of synthetic mixtures results, shown in Table 3 also indicate that there are
not significant difference between the outputs of the
Both proposed methods (PLS and HPSAM) were PLS and H-point standard addition methods.
applied to the determination of Cr(VI) and Fe(III) in
several synthetic samples that had compositions pre-
pared according to some alloys [38,39]. The results are 4. Conclusions
shown in Table 3. The good agreement between these
results and known values indicates the successful ap- The most important characteristics of this work are:
plicability of the proposed methods for simultaneous (i) the possibility of simultaneous determination of
determination of Cr and Fe in complex samples. The Cr(VI) and Fe(III) in an aqueous solution owing to
336 H. Abdollahi / Analytica Chimica Acta 442 (2001) 327–336

the presence of non-ionic micelles of Triton X-100; [13] M. Blanco, J. Gene, S. Iturriga, S. Maspoch, J. Riba, Talanta
(ii) NO extraction step is required and hence the use 34 (1987) 987.
[14] M. Otto, W. Wegscheider, Anal. Chem. 61 (1989) 1847.
of organic solvents is avoided; (iii) the proposed PLS
[15] A. Safavi, H. Abdollahi, Microchem. J. 63 (2) (1999) 211.
model and HPSAM are very suitable for simultane- [16] A. Safavi, H. Abdollahi, Talanta 54 (2001) 727.
ous determination of chromium and iron and can be [17] G.L. Mclntire, Cri. Rev. Anal. Chem. 21 (1990) 257.
employed to analyze complex synthetic samples; (iv) [18] G. Jin, Y. Zhu, W. Jiang, B. Xie, B. Cheng, Analyst 122
determination of both Cr(VI) and Fe(III) are based (1997) 63.
[19] J. Hernandez, B. Moreno, J.L. Perez, J. Cerda, Inorg. Chim.
on indirect spectrophotometric responses; (v) selectiv-
Acta 140 (1987) 245.
ity is achieved mainly by choosing a suitable mixed [20] H.B. Singh, N.K. Agnihotri, V.K. Singh, Talanta 48 (1999)
reagent system which leads to the selective determi- 623.
nation of Cr(VI) and Fe(III) with sensitive spectral [21] The Merck Index, 10th Edition, Merck and Co. Inc., Rahway,
characteristics; (vi) two powerful chemometric meth- New York, 1983.
[22] Second Annual Report on Carcinogens, Environmental
ods (PLS and HPSAM) are compared for analysis of
Protection Agency, NTP 81-43, December 1981, pp. 73–75,
binary mixtures with strong signal overlapping. 78–80.
[23] A. Lorber, L. Wangen, B.R. Kowalski, J. Cemom. 1 (1986)
19.
Acknowledgements [24] D.M. Haaland, E.V. Thomas, Anal. Chem. 60 (1988) 1193.
[25] S.D. Brown, R.S. Bear, Cri. Rev. Anal. Chem. 24 (1993)
The author wish to express their gratitude to the 99.
[26] F. Bosch-Reig, P. Campains-Falco, Analyst 113 (1988) 1011.
Institute for Advanced Studies in Basic Science for [27] F. Bosch-Reig, P. Campina, P. Campins-Falco, Analyst 115
the support of this work. (1990) 111.
[28] P. Campins-Falco, F. Bosch-Reig, J.V. Andres, Talanta 39
(1992) 1.
References [29] P. Campins-Falco, F. Bosch-Reig, J.V Andres, Anal. Chim.
Acta 270 (1992) 253.
[1] D.T. Burns, C.D.F. Dangolle, Anal. Chim. Acta 356 (1997) [30] P. Campins-Falco, F. Bosch-Reig, J.V. Andres, C.M. Legua,
145. Talanta 41 (1994) 39.
[2] A.C. Harzderf, Int. J. Environ. Anal. Chem. 29 (1987) 249. [31] F. Blasco-Gomez, P. Campins-Falco, F. Bosch-Reig, C.M.
[3] M. Boussemart, C.M.G. van der Berg, Anal. Proc. 28 (1991) Legua, R.H. Hernandez, Anal. Chem. 72 (2000) 2559.
68. [32] P. Campins-Falco, F. Bosch-Reig, R.H. Hernandez, A.S.
[4] W. Frenzel, Fresenius’ J. Anal. Chem. 361 (1998) 936. Cabeza, C.M. Legua, Anal. Chem. 63 (1991) 2424.
[5] S. Nielsen, E.H. Hansen, Anal. Chim. Acta 366 (1998) 163. [33] A. Safavi, H. Abdollahi, F. Sedaghatpour, S. Zeinali, Anal.
[6] F.A. Byrdy, L.K. Olson, N.P. Vela, J.A. Caruso, J. Chromatogr. Chim. Acta 409 (2000) 275.
A 712 (1995) 311. [34] CRC Handbook of Organic Reagents in Analytical Chemistry,
[7] B. Baraj, M. Martinez, A. Sastre, M. Aguilar, J. Chromatogr. CRC Press, Boca Raton, 1982, pp. 277–283.
A 695 (1995) 103. [35] P. Compins-Falco, J. Verdu-Andres, F. Bosch-Reig, Anal.
[8] M. Korolczuk, Anal. Chim. Acta 414 (2000) 165. Chim. Acta 315 (1995) 267.
[9] S. Pehkunen, Analyst 120 (1995) 2655. [36] M. Otto, W. Wegscheider, Anal. Chem. 57 (1985) 63.
[10] K. Saitoh, T. Hasebe, N. Teshima, M. Kurihara, T. [37] M. Blanco, J. Coello, H. Ituriaga, S. Maspoch, M. Redon,
Kawashima, Anal. Chim. Acta 376 (1998) 247. Appl. Spectrosc. 48 (1994) 37.
[11] N. Teshima, K. Ayakowa, T. Kawashima, Talanta 43 (1996) [38] F.D. Snell, L.S. Ettre, Encyclopedia of Industrial Chemical
1755. Analysis, Vol. 5, Wiley, New York, 1972.
[12] H. Kutsumata, N. Teshima, T. Kawashima, Bull. Chem. Soc. [39] Composition and Physical Properties of Alloys, http://
Jpn. 70 (1997) 2151. chemistry.csudh.edu/oliver/chemdata/alloys.htm.

You might also like