CONTENTS
Required equipment:
1 Phillips screwdriver, PH 2.
1 Hexagonal spanner no.13.
1 Sample weight, 5 Kg.
2 Sample weights, 2 Kg.
1 Sample weight, 1 Kg.
Or, alternatively:
1 Sample weight, min. 10 Kg.
1 Known weight, 9 Kg.
After removing the front panel, remove the M8 screw that fastens the container (34) to the
scale plate (18). Remove the container (34).
Note: it is not necessary to disconnect the hoses from the internal container (34).
With the display showing –End–, press the S Key to exit. The display shows the Stand-by
mode.
End of procedure.
Press the F key repeatedly to select the following operation parameters in sequence:
CODE VALUE
P-t02 60
P-t03 180
P-t04 60
P-t05 45
P-t06 04
P-t07 05
P-t08 02
P-t09 60
P-t10 00
P-t11 09
P-t12 60
P-t13 05
P-t14 240
P-t15 05
P-udL 0.700
P-udH 1.400
P-ud1 0.120
P--Fd 1.000
rES
P--Lb 10.000
t--AL 3600
When the display shows u---F, the pressure transducer can be calibrated.
Disconnect the quick connectors from the high and low pressure hoses.
Open the LP-HP valves on the control panel to connect the pressure transducer to the
atmospheric pressure.
With the display showing u---F, confirm with the S key.
The display shows the theorical weight measured.
Digit the value 0.960. Confirm with the S key.
P---M 19.500
rES
P-CFd 00.700
-End- OUT
Use the F key to select the desired parameter and confirm by pressing the S key.
Use the UP-DOWN keys to set the value shown in the table.
Confirm with the S key.
Select the parameter -End- and confirm with the S key to exit.
The display shows the net weight available (standby).
End of procedure.
43 41 40 OPTIONAL
11
1
10
20 3
48 7
21 45
46
47
LED W
LED T
LED P
10
42
49 2
11
39 25
31
34
53
33
12
22 28
29
25
13/14
30
15
39
65
27 34
16
18
17
13
50 CHECK VALVE
42
61A
60
12
61B
62A 62B 63 64
14
71
81
72
80 73
74
79
75
78
76
77
15
83 82
85
84
16
17
18
19
CONTROL PANEL
UP S
DOWN F
20