Professional Documents
Culture Documents
1007989 O 95 (322)
Service Manual – Philips MicroDose L30, Software 8.4
When?
Once a week, or if Quick calibration fails.
When?
A Quick calibration should be performed when a line artefact occurs. See example in section
18.2.1.
When?
When four hours have passed since the last exposure was performed, the AWS will recommend
warming up the X-ray tube using this procedure.
1007989 O 96 (322)
Service Manual – Philips MicroDose L30, Software 8.4
When?
If Full calibration or Quick calibration fails because of previous problems which have now been
resolved, this step may be required. The channels which are masked because of the history
function are shown in the Channel mask viewer as Check history, see section 11.8.
1007989 O 97 (322)
Service Manual – Philips MicroDose L30, Software 8.4
Asic calibration trims the amplifier settings for the ASICs in the detector. The calibration is
performed without X-ray exposure.
Handling irregularities
See section 20.3 for information regarding handling failed calibrations.
1007989 O 98 (322)
Service Manual – Philips MicroDose L30, Software 8.4
Procedure
1. Press Emergency stop and
lower the collimator by hand.
2. Mount the lower part of the
ACA mechanics, see Figure
61. The lower part is attached
to the pre collimator assembly
by two Allen screws.
3. Loosen (do not remove) the
three screws holding the
collimator to its base, see
Figure 62. In what follows,
we refer to these as the Figure 61 The upper ACA mechanics is mounted on the so
alignment screws. You should called interface position. The lower ACA
now be able to move the mechanics is attached to the pre collimator
collimator in all directions. assembly. (The X-ray prevent cover is not
4. Tighten the alignment screws visible in this illustration.)
to 1.2 Nm or by hand until
there is no vertical play
when you pull the
collimator front up and
down. The collimator
should however still be
possible to move by
hand from left to right.
5. Mount the upper part(s)
of the ACA mechanics.
You may need to move
the collimator left/right
in order for the parts to
fit in each other. Apply
the springs (see “A” in
Figure 61) and check
that the collimator
position can be adjusted Figure 62 The three alignment screws of the collimator.
by turning the knobs.
6. Adjust both screw micrometers to the starting position:
10 mm for blue ACA (revision C or later).
13 mm for silver ACA (revision B or earlier).
7. Connect the power cable and the computer cable.
1007989 O 99 (322)
Service Manual – Philips MicroDose L30, Software 8.4
8. Release the emergency stop and perform a reset. If the X-ray prevent cover is mounted,
make sure that it is attached to the collimator interface during the reset.
Make sure that no part of the ACA mechanics collide with anything during
Notice the Reset. Otherwise the homing of the collimator movement will be
incorrect and the alignment useless.
9. Click on System Configuration (see Figure 59) in the Service Tool, under the tab Service.
Check that the Detector S/N is correct. If it is not correct, enter the correct one and click
OK.
10. Click Collimator Alignment (see Figure 59) in the Service Tool, under the tab Service to
start ACA.
11. Click OK. The ACA motor starts the homing procedure, and the collimator can be seen to
move slowly in the left/right direction.
As the screws are tightened, you must not move the motor! Therefore, at
Notice
this stage, be sure not to check Screws are made loose.
16. If the tube has two filaments and it is desired to align the pre-collimator with respect to
both filaments (optional), perform the instruction in section 11.3.3. Otherwise, continue
with step 17.
17. When the alignment is good enough (all indicators are green), click Done.
18. Disconnect and remove the ACA mechanics.
19. Perform a Geometry calibration. See section 11.3.5.
Handling irregularities
See section 20.3.1 for an alignment trouble shooting guide.
Figure 65 Verify alignment. If the indicators are green, the alignment is good.
Figure 66 Verify Alignment dialog will appear if full collimator alignment was performed with
L30 SW 8.1 or earlier.
2. Verify that the Collimator height um is set to a positive number (it might be for example
235000). Click Exposure and expose.
3. Reduce the Collimator height um by 7000 um (following the example above, we should
set the height to 228000). Click Exposure and expose.
4. The green field (asics that are still counting at the same rate) should be properly centred
inside the blue fields (asics that count less). See example of correct alignment in Figure 67
(left). If the collimator is not aligned, see an example in Figure 67 (right), a full collimator
alignment has to be performed.
Handling irregularities
See section 20.3 for information regarding handling failed calibrations.
Procedure
1. Click on Geometry calibration in the Service Tool,
under the Service tab.
2. Place the Geometry and image field phantom on the
patient support. Compress. Press Compression Done
and hold down the exposure button.
3. A progress bar is shown and the current step is
indicated. Figure 68 Geometry phantom.
(Part no. 1012092)
4. The user is informed whether the calibration was
successful or not.
5. The status field is updated.
6. It is advisable to check the channel mask after this
calibration, especially if the calibration was unsuccessful.
Handling irregularities
If the Geometry calibration fails:
1. Verify that the phantom is correctly mounted on the patient support. The front guiding
edge of the phantom should touch the front end of the patient support.
2. Run Verify alignment.
3. If the alignment is correct, check if the detector touches the patient support. This can be
performed
a. by checking the Scan speed in Measurements in the Service tool. If the scan speed
diagram shows unexpected peaks, the detector might touch the patient support.
b. or by removing the side covers of the C-arm. A visual check can performed
verifying that there is free space between the detector and patient support.
If no problems are found performing the above described actions, contact product support. See
section 20.3 for more information regarding failed calibrations.
Skew calibration measures and compensates for the deflection of the detector as an effect of
gravity.
Procedure
1. Click on Skew calibration in the Service Tool, under the Service tab.
2. Rotate the C-arm to -90 degrees. Place the Geometry and image field phantom on the
patient support. Compress gently, press Compression Done and hold down the exposure
button.
3. A progress bar is shown and the current step is indicated.
4. The user is informed whether the calibration was successful or not.
Handling irregularities
Contact product support if the calibration fails,
Procedure
1. Click AEC calibration in the Service Tool, under the Service tab.
2. The user is informed whether the calibration was successful or not.
Handling irregularities
AEC calibration is a part of Full Calibration that is performed on regular basis. However if an
AEC calibration would fail, probable explanations might be that the X-ray tube is new and has to
be “burned in” by performing a number of exposures before output is ok. A second explanation
might be that the X-ray tube is old and need to be replaced.
See section 20.3 for information regarding handling failed calibrations.
Procedure
1. Click Compression paddles in the Service Tool, under the Service tab.
2. All configured compression paddles are shown to the upper left. The text “(calibrated)” is
added after the names of the paddles that are already calibrated. See Figure 80.
3. Mark the compression paddle you want to calibrate.
4. Place the compression height calibration phantom on the patient support.
5. Enter the True height of the phantom.
6. Compress to a certain force between 50 and 200 N. Click Measure. The measured values
are displayed in the plot.
7. Compress to a different force between 50 N and 200 N. Click Measure. Repeat this until
there are at least five points in the plot.
8. Click Accept. The text “(calibrated)” is added after the name of the compression paddle.
9. Mark another compression paddle to calibrate or close the window.
It is also possible to test a calibrated compression paddle, by checking the Test box. Then the
height should equal the real height of the object, regardless of the compression force. A calibrated
compression paddle can be rejected (i.e. marked as not calibrated) by pressing Reject.
Outgoing control
After the compression paddles have been calibrated:
Compress all the way to the patient support. Check that the display shows 0 mm to 2 mm
and no compression force.
Handling irregularities
1. Check if the compression force measurement is correct.
2. The calibration might fail if the values are too spread. Repeat the procedure until it is
successful.
See section 20.3 for more information regarding handling failed calibrations.