Technical Manual
Release 31 October 2002 (Rev. 2)
TECHNICAL MANUAL
Revision history
Revision history
Contents
1. INTRODUCTION 1-1
1.1 Icons appearing in the manual.............................................................1-1
3. DESCRIPTION 3-1
3.1 Identification labels..............................................................................3-1
3.2 Configuration of the MOVIPLAN 800 ....................................................3-2
5. PRE-INSTALLATION 5-1
5.1 Electrical wiring set-up ........................................................................5-2
5.2 Packaging ............................................................................................5-3
5.3 Space requirements .............................................................................5-3
6. INSTALLATION 6-1
6.1 Unpacking and components inspection................................................6-1
6.2 Rails assembly .....................................................................................6-2
6.3 Table positioning and fastening............................................................6-3
6.4 Column assembly ................................................................................6-5
6.5 Tubehead, Control Panel and Collimator assembly ..............................6-8
6.6 Cables connection.............................................................................. 6-10
6.6.1 Connection from column (only for TA and TF versions) ....................6-10
6.6.2 Connection MOVIPLAN 800 table – Cabinet (only for TA and
BA versions) ...................................................................................6-10
6.6.3 Internal connections MOVIPLAN 800 table (only for TF and
BF versions) ...................................................................................6-10
6.6.4 Other connections (only for TA and TF versions) ..............................6-11
6.7 Table top assembly ............................................................................ 6-12
6.8 Pedals connections............................................................................. 6-12
6.9 Final balancing and adjustment......................................................... 6-13
6.10 Electrical connections ........................................................................ 6-14
6.11 Bucky start voltage setting................................................................. 6-15
1. INTRODUCTION
NOTE:
* The present manual is updated for the product it is sold with in order to
grant an adequate reference in performing diagnostics and repair
operation normally carried out by the service engineer.
The manual may not reflect changes to the product not impacting service
operations or use mode.
The scope of this Manual is to supply user instructions for a safe and
efficient operation of the device described.
This icon indicates a WARNING: the items marked by this icon refer to
, safety aspects of the patient and/or of the operator.
2. SAFETY ASPECTS
WARNING:
, All the information provided in this chapter must be read carefully, and,
where so required, must be implemented to prevent damage to the
equipment and injury to people.
VILLA SISTEMI MEDICALI designs and builds its devices complying with
the related safety requirements; furthermore it supplies all information
necessary for a correct use.
• use of the MOVIPLAN 800 device for other than its intended use,
• damage to the device or injury to the operator or patient due to
incorrect installation and maintenance performed other than in
compliance with the procedures defined in the Technical Manual
provided with the device or due to incorrect operating techniques,
• mechanical and/or electrical changes made during and after
installation.
Only the authorised personnel can remove the covers and/or have
access to the components under voltage.
2.1 Warnings
The MOVIPLAN 800 is suitable for use in a hospital environment and
cannot be connected to a "domestic" electrical system.
The device has not been designed to be used in the presence of steam,
anaesthetic mixtures which are flammable in air, oxygen or nitrous
oxide.
The operator MUST ALWAYS PAY THE UTMOST CARE when the
motorised mechanisms are activated such as the vertical movement of
the table top or of the column. The red EMERGENCY button on the front
of the device can be pressed in case of danger. This button immediately
stops all movements and all functions and provides the operator with an
important safety tool in addition to the other intrinsic safety devices of
the device.
Before activating the movement of the column, check that the patient is
placed correctly on the table top with the legs and arms inside the
support area.
Although the MOVIPLAN 800 table was designed with a high degree of
protection against electromagnetic interference, it is necessary to install
it a certain distance away from electric energy transformers, continuity
static groups, pack units (walkie-talkies) and cellular phones. Use of the
latter is only permitted at a distance from the device as provided in the
table below.
Electromagnetic
Conformity EMC Environment
Emissions
All parts that come into contact with the patient must be cleaned,
and where necessary, disinfected.
Before activating the X-rays from the generator table, the operator shall
always check the alignment of the tubehead with respect to the image
receptor and should set the maximum aperture of the collimator in
accordance both with the format of the radiography cassette used and
with the distance of the cassette from the focal point of the tubehead.
Use the farthest focal point-skin distance compatible with the X-ray
examination to be performed.
WARNING:
, Before dismantling the column, lock the column's tube-carrier arm. Then
simultaneously remove the radiogenic-collimator tube and the counter-
weights of the sliding mechanism inserted in the column.
Villa Sistemi Medicali shall not be responsible for the disposal of the
device disposed of by the user nor for the relative costs.
Symbol Description
∼ Alternating current
Protection grounding
Operation grounding
Dangerous voltage
3. DESCRIPTION
2
3
NOTE:
* In addition, on all the above versions, the MOVIPLAN 800 table can be
ordered with a motorised elevator to allow the table top to be raised and
lowered smoothly and quickly for easier patient handling. The versions
with motorised elevator includes a control cabinet.
Grids
• Standard (Ratio 10, focal length 120 cm, 34 pl/cm) (code
4695102600).
• High performance (Ratio 12, focal length 120 cm, 36 pl/cm) and low
absorption (code 4695101600).
Accessories
The following accessories are available for all models on request:
• Single leg rest (code 7173142000).
• Band stretcher with compression band (code 7173129000).
• External side cassette carrier (code 7173115000).
9771100503 Moviplan 800 TA: with elevator, Potter Bucky 35x43, 230 V~
column and manual collimator
9771100603 Moviplan 800 TF: fixed with Potter Bucky 35x43, column 230 V~
and manual collimator
9771100703 Moviplan 800 BA: with elevator, Potter Bucky 35x43 230 V~
9771100803 Moviplan 800 BF: fixed with Potter Bucky 35x43 230 V~
4. TECHNICAL DATA
General characteristics
Type of device MOVIPLAN 800
Manufacturer Villa Sistemi Medicali S.p.A.
20090 Buccinasco (MI) Italy
Type of device and classification in
Class I with type B applied parts
accordance with Standard IEC 60601-1
Type of device and classification Class I – CE 93/42
Degree of protection in accordance with IP00
standard IEC 60529
Operating mode Continuous
Table mechanical characteristics BF BA TF TA
Weight 150 kg 180 kg 150 kg 180 kg
Maximum height 790 mm
Total length with table top centered 2180 mm
Total width with table top centered 720 mm
Column stand weight no 290 kg
Column height no 2400 mm
Column width (base) no 1070 mm
Column depth (with arm extended) no 1400 mm
Electrical characteristics BF BA TF TA
Voltage 230 Vac
Frequency 50/60 Hz
Maximum line current draw 0.25 A 2.1 A 0.9 A 2.8 A
Absorbed power 58 VA 480 VA 208 VA 645 VA
Device protection fuse 2AT 3AT 2AT 3AT
Potter protection fuse 0.2 A T
Line impedance < 1.0 Ω
Movements BF BA TF TA
Manual longitudinal movement of the yes
potter with electromagnetic braking
Manual longitudinal movement of the yes
table top with electromagnetic braking
Manual transversal movement of the yes
table top with electromagnetic braking
Motorised vertical movement of the table no yes no yes
top (elevator) with pedal command
Manual vertical movement of the no yes
column tube carrier arm with
electromagnetic braking
Manual transversal movement of the no yes
column tube carrier arm with
electromagnetic braking and central
mechanical block
Manual rotation movement of the no yes
column with mechanical stops
Manual rotation movement of the no yes
column tube carrier arm with
electromagnetic braking
Movements distance BF BA TF TA
Potter longitudinal movement 540 mm
Table top longitudinal movement 415 mm head end / 415 mm feet end
Table top transversal movement 115 mm internal / 115 mm external
Table top vertical movement (elevator) n.a. 460 mm n.a. 460 mm
min. distance Table top-Floor
Table top vertical movement (elevator) n.a. 790 mm n.a. 790 mm
max. distance Table top-Floor
Column longitudinal movement n.a. 1710 mm
Tube carrier arm vertical movement n.a. 1470 mm
Minimum focal point-floor distance (in n.a. 430 mm
function of tube used)
Maximum focal point-floor distance (in n.a. 1900 mm
function of tube used)
Maximum focal point-film distance (in n.a. 1200 1530
function of tube used) mm mm
Tube carrier arm transversal movement n.a. 190 mm
Column rotation movement n.a. 180° (± 90°)
Tube carrier arm rotation movement n.a. 360° (± 180°)
Features BF BA TF TA
Table top brake release with photocell at floor level
DFF reading n.a. centimeter scale on
column
Controls positions on control panel
Indication of brakes on n.a. luminous switches
on control panel
Indicator of degrees of inclination of the n.a. with goniometer
tube carrier arm
X-rays band - cassette centering n.a. with luminous spot
on collimator
Projections n.a. perpendicular on
potter,
oblique on potter,
side (with optional
accessories),
off-table,
on CHEST STAND
(optional)
Accessories BF BA TF TA
Handles 1
Leg rests for gynaecology examinations Optional
Compression band with band stretcher Optional
Cassette support for side projections Optional
Environmental conditions BF BA TF TA
Operating environmental conditions Temperature: from +10° to +40° Celsius
Humidity: from 30% to 75 %
Pressure: from 700 hPa to 1060 hPa
Environmental conditions for Temperature: from -20° to +70° Celsius
transportation and storage Humidity: ≤ 95% non condensing
Pressure: > 630 hPa
4.1 Dimensions
• General safety:
IEC 601-1
IEC 601-2-32
• Electromagnetic compliance:
IEC 601-1-2
5. PRE-INSTALLATION
The following pages provide instructions for the correct installation of the
MOVIPLAN 800.
The supplier can provide the necessary assistance and technical
servicing as from the pre-installation phase.
The room must be prepared for the installation of the MOVIPLAN 800 as
indicated in the Figure 5-1.
The standard installation of the MOVIPLAN 800 requires the cables to be
enclosed in channels. Please note the maximum distance allowed
between distribution box "B" and distribution box "C" (centre of the
electric cabinet which is available only in the version with elevator).
When installation is required with distances greater than those shown in
the figure, electric connecting cables of suitable lengths must be
requested when issuing the order.
Requirements for a correct installation are:
• Room minimum height of 3 mt.
• For transportation of the machine inside the room, the door sizes
must be as follows:
– 700 mm for the column
– 1200 mm for the packaged materials.
• For models with column (TA and TF), to obtain a proper cable setting
of cables coming from the column, the table must be positioned close
to a wall.
A plastic duct (section 100x100) has to be fixed on the wall aligned
with the center of the rail: its lower end must be corresponding to the
"B" collector on the floor; its upper end must be high enough so that
the cables do not interfere along the entire range of motion of the
column along the rail.
HT cables to
Cable entry underfloor tube Ø4" (Ø100)
Generator
min 1.6ft
6x6" (150x150)
(0.5m)
Cabinet
40.5x21.6x20.5"
(1030x550x520)
Cable entry
6x6" (150x150)
14" (355)
37" (940)
min 6.5ft
117.3" (2980)
(2m)
Figure 5-1
(Rev. 2) 5-1 MOVIPLAN 800
TECHNICAL MANUAL
Pre-installation
WARNING:
, The general ground system must comply with the laws in force.
Insufficient grounding of the device can be extremely dangerous for
the operator and the patient and can cause malfunctions in the
electrical equipment connected to the MOVIPLAN 800.
The device's power cable must either be connected to the X-rays
generator in series connection or to a differential thermal magnetic
switch with the following characteristics:
250V - 6A - 0.03A.
A connecting line for the emergency switch of the MOVIPLAN 800
table with the radiology room's general mains switch must also be
installed.
See chapter 4 “Technical data” for current and absorbed power data.
5.2 Packaging
The MOVIPLAN 800 is packed and shipped in one single case.
The dimensions, net weight, gross weight and contents of this case are
listed below:
Weight
Case type Contents
Net Gross
- Table base + Potter Bucky
- Table top
- Front floor rail
Moviplan 800 - Rear floor rail
TA and TF - Tube column + tube arm 600kg 850kg
330x110x110 - Collimator
- Push button control panel
- Cables
- Cabinet (*)
- Table base + Potter Bucky
Moviplan 800
- Table top
BA and BF 320kg 400kg
- Cables
248x93x110
- Cabinet (*)
NOTE:
(*) cabinet supplied with elevating table only
6. INSTALLATION
NOTE:
* The system is shipped in pre-assembled groups. The mechanical
assembly at installation consists only of final assembly of these groups.
This means that the adjustment of bearing, locks and tightening of
fixings are carried out in the factory prior to shipping and any
modification of the original settings of these particular components may
cause faulty or poor operation of the system.
NOTE:
* Actions described in this chapter have to be carried out only for versions
of MOVIPLAN 800 equipped with column (models TA and TF).
Position the front rail in the room as shown in the figure, taking into
account the size of the equipment and the cable incoming distribution
box as indicated in the pre-installation (chapter 5).
Fasten the rail to the floor with the expanding plugs provided.
Check that the rail is level and if not, level with the spacers provided.
Fasten the back rail to the front rail, level and fasten to the floor using
the ø8 plugs provided.
Rear rail
Fixing point
between rails
Front rail
Bolts Ø8
580 (22.8") 300 (11.8") 300 (11.8") 580 (22.8") (Ø.31")
= =
Figure 6-1
Front rail
Rail covers
Table
Cable exit
Cover plate
Figure 6-2
NOTE:
* The table is positioned asymmetrically with respect to the front rail to
allow the column to travel outside the space of the table on one side. It is
possible to choose the side towards which the column can exit the table
dimension, by positioning the table towards one or the other side.
Set the table in place lining it up with the holes in the rear rail. Fix the
table to the floor using the bolts and the rails using the two M8x25
screws provided.
Table
Front rail
Fixing points
Table-Rail
Figure 6-3
Check that the table is properly level and correct if necessary with the
metal shims provided.
Verify the alignment of all parts, paying especially attention to:
• the alignment of the front rail with the rear rail
• the alignment of the front rail with the table
• the levelling of both rails and the table on the floor.
WARNING:
, Remove the two supports, that are necessary during the shipment,
positioned on the lateral parts of MOVIPLAN 800 base. Do not re-install
the relevant screws.
NOTE:
* Actions described in this chapter have to be carried out only for versions
of MOVIPLAN 800 equipped with column (models TA and TF).
1. Before fitting the tube column onto its rail it is necessary to:
• remove an end-stop from one end of the rear rail in order to allow
the column to slide on
End stop
Figure 6-4
• remove the external cover plate from the rear of the column
• set two counter-weights in position in the column and check the
balancing of the tube cross-arm movement up and down the
column
• lift and secure the cross-arm in its highest position with a clamp
so that the counter-weights are at their lowest position inside the
tube column. In this way, there is less danger of the counter-
weights falling when you position the column.
Cover
Counterweight
Figure 6-5
2. Slide the column onto the rear rail making sure that the bearings
(two for each side of the column) locate on the guide and the column
wheel slides smoothly on the front guide. There is no need to make
any adjustment to the bearings as they are factory tested before
shipping to ensure correct operation.
NOTE:
* If further regulation is needed, a special hexagonal wrench, 17 mm, with
thickness lower than 4.5 mm must be used.
Column
bearing
Side
bearings
Figure 6-6
End stop
Security pin
Figure 6-7
End run up
Figure 6-8
NOTE:
* Actions described in this chapter have to be carried out only for versions
of MOVIPLAN 800 equipped with column (models TA and TF).
WARNING:
, On the column it is possible to use X-ray tubes with a maximum weight
of 25kg.
1. To fit the tube and housing on the column cross-arm, first take out
the locking pin from the tube assembly. Next insert the tube
assembly on the cross-arm pin and re-fit the locking pin temporarily.
Cross-arm
pin
Tube
assembly
Locking
pin
Figure 6-9
Base +
Push button panel
Fixing cone
Cable
passage
Figure 6-10
Collimator
Locking
points
Cross-arm Tube
assembly
Figure 6-11
5. Adjust the position of the tube assembly locking pin and adjusting
screw. When correctly adjusted, the tube assembly should balance in
any desired position when rotated.
NOTE:
* The current position of the collimator may not grant perfect alignment of
the anode focus and the collimator itself.
A regulation may be needed as described in paragraph 7.4 or 7.5.
• Cable X16 from Control Panel to X16 connector on the top of the
column and to the main components plate (A4 board – connector
X16)
• cable X31 and ground cable from the tube arm to PRL9 terminal
board on the top of the column
• ground cables to the ground connections on the main components
plate.
• Cable X14 from MOVIPLAN 800 table (A5 board – connector X14) to
cabinet (A4 board – connector X14)
• cable X17 from MOVIPLAN 800 table (A5 board – connector X17) to
cabinet (A4 board – connector X17)
• cables X0 from MOVIPLAN 800 table (A5 board) to cabinet (X0
terminal board)
• ground cables to the ground connections in the cabinet.
• HT cables, rotating anode cable and ground cable from X-ray tube to
generator
• collimator cable between 9 pins connector and X30 on control panel
(see instruction note 39119001).
NOTE:
* Make reference to the System Integration Manual for electrical
connection to the generators and the other components.
Column
connections
HT cables
Control Panel
X16
Control Panel
cable
X31
Cabinet
(TA and BA versions)
Main components
plate Main components
plate
A4 board
A4 board
X0 X0
A5 board
HT cables to Generator
Figure 6-12
Table top
Locking end
End stop stop screws
Locking ratchet
Figure 6-13
Table base
Fixing points
Pedal control
Figure 6-14
Cover
Counterweight
Figure 6-15
Rail
.078-.12" (2-3mm)
Adjusting screws
Figure 6-16
WARNING:
, Before electrically connecting the MOVIPLAN 800, check that all the
instructions and warnings given in paragraph 5.1 have been complied
with.
Disconnect from the mains before performing any operations having
to do with the electrical wiring of the system.
Table 1
After having connected the power cables, they should be tied together to
prevent accidental contact in case of wire breakage.
Figure 6-17
NOTE:
* In case the voltage selected is 24Vdc the positive must be connected to
X0-5 and negative to X0-6.
WARNING:
, Calibration of the R2 must be made exclusively with the grid in the still
position.
Measure the size of the light image starting from about halfway into the
penumbra. The sizes must be equal to the set values, ± 8 mm.
Check the size of the light field in the various formats indicated by the
knobs for a distance of 100 cm.
WARNING:
, Only in the case of light bulb replacement: adjust the position of the
collimator lamp until these measurements are obtained.
<5mm
NOTE:
* Before aligning the X-ray beam with the current procedure it is necessary
to make sure that the column is perfectly vertical and the X-ray tube is
horizontal; these checks can be easily done using a bubble level.
Tube focus
Collimator center
Correct centering
Figure 7-1
7.5.2 Alignment of the X-ray beam with the light field of the
collimator
NOTE:
* The light beam of the collimator is calibrated in the factory so that once
the alignment between the tube and the collimator is achieved, the
alignment of the X-ray beam and the light field should be granted.
1. In order to check the alignment of the X-ray beam with the light field
of the collimator turn on the collimator light and position 4 radio
opaque objects on the borders of the light field.
2. Make an exposure on a large film positioned on the table top and
check the position of the objects; the max deviation must be less
than 3% of the in each direction and the sum of the deviation along
the two directions must be less than 4%.
3. If the alignment is not achieved within the specified limits (i.e. the
X-ray focus distance from the image plan is different from the
distance of the slot of the lamp) it is possible to act on the lamp of the
collimator, remembering that moving the lamp closer to the mirror
the light field gets smaller, while moving the lamp further from the
mirror the light field gets larger.
Figure 7-2
4. Once the optimal alignment between light field and X-ray beam has
been achieved it is possible to complete the alignment between the
X-ray beam and the image receptor (Potter Bucky – see paragraph
7.6).
Figure 7-3
Tube assembly
Cross-arm
Pushbutton panel
Collimator
Column
Table top
Table
Figure 8-1
Table Emergency
push button
Bucky
tray
Potter Release
push button
Cassette
Centre
notch Magnetic lock
Locked
Jaws
Lever
Figure 8-2
Figure 8-3
Angle meter
8.4 Column
The column can make the following movements:
• Longitudinal
Pushing the longitudinal lock release button on the tube assembly
control panel allows manual movement of the column along the floor
rail.
Figure 8-4
Figure 8-5
Figure 8-6
ANGLE METER
Figure 8-7
Unlocking
pedal
Figure 8-8
Figure 8-9
WARNING:
, Before making any type of examination ensure that:
• a film loaded cassette is present in the Potter Bucky or in any of the
auxiliary accessories (chest stand, lateral cassette holder, etc.)
• the collimator has been properly adjusted according to the used
cassette format and to the set SID. Use the ruler attached to the
collimator and the light spot on purpose
• the X-ray beam is properly aligned and centered with the cassette
• no object lays between the X-ray source and the patient.
WARNING:
, The duty cycle of the collimator lamp must not be longer than 2
consecutive activation cycle separated by 180 seconds of de-activation
(lamp off).
This duty cycle assure that the parts which may come in touch with the
patient/user do not reach a too high temperature due to the heating
effect of the halogen lamp.
When the table is being used with the Potter Bucky, the X-ray beam
must be aligned with the Bucky and the Source Image Distance adjusted
accordingly as required by the operator for the type of radiographic
projection being made.
With elevating table versions, the table-top height can be adjusted by
using the foot pedals, in order to have the most comfortable working
position.
WARNING:
, The elevator pedals commands directly the elevator motor. Take care to
avoid unwanted movement of the table top caused by accidental pressure
of the elevator pedals.
WARNING:
, During the operation of the elevator make sure that no obstacle is
present in the range of the elevator movement.
Figure 8-10
3. Centre the X-ray beam to the film cassette by unlocking and moving
the Bucky tray longitudinal using the switch.
Bucky tray
Figure 8-11
4. Unlock and move the tube column on its floor rail by pushing the
lock release button on the tube assembly control panel.
Figure 8-12
Figure 8-13
NOTE:
* The distance between the table top and the cassette inserted into the
Potter Bucky is 8cm. As alternative, the distance between the focal spot-
film can be calculated by summing up this 8cm and the distance read on
the collimator measuring tape. For example, to set a focal spot-film
distance of 1meter, set the X-ray tube position at 92cm from the table
top.
WARNING:
, The scale is graduated for the following focal spot-film distances: 80, 100,
110, 180 and 200cm. For distances different from the one previously
listed, the extrapolated formats may not be reliable.
Then, once the area to be irradiated has been collimated by acting on the
dedicated knobs of the collimator, the unit is ready for the exposure.
Figure 8-14
NOTE:
* Moving the column towards the outside of the table, it is possible to lower
the tube below the level of the table top and take exposures of the legs
(knee and ankle) under load with the patient standing close to the Chest
Stand, usually associated with the Bucky table.
WARNING:
, Before cleaning the unit disconnect it from the mains.
Do not let water or liquids enter the unit, as could cause corrosion or
short-circuit.
Clean the painted surfaces, the connection cables and the accessory by
using a damp cloth and a mild detergent. Then wipe with a dry cloth. Do
not use solvents (alcohol, benzine, trichloro-ethylene) corrosive or
abrasive solutions.
Disinfect the table top, the handle of Potter Bucky, the handle of column
pushbutton and the accessory that the patient and the operator used
with a 2% solution of Glutharaldeid.
10. MAINTENANCE
As with any electrical appliance this equipment not only requires proper
operation but also regular maintenance and servicing.
These precautions serve to maintain the system in good working
conditions and operational reliability.
WARNING:
, Metallic friction noises issued from the column may be a symptom of
wear of the counterweight support cables. If this should occur, stop
using the device immediately and contact the Villa Sistemi Medicali
Technical Service Department.
NOTE:
* All component which are replaced must be only original spare parts from
Villa Sistemi Medicali.
The maintenance intervention and replacement of defective parts must be
made exclusively by Villa Sistemi Medicali authorised personnel.
MOVIPLAN 800
Components plate layout 1
TA/BA versions
Code 72711383 - Rev. 2 Page 1 of 1
MOVIPLAN 800
Electrical diagram 2
TA version
Code 39719122 - Rev. 2 Page 1 of 2
MOVIPLAN 800
Electrical diagram 2
TA version
Code 39719122 - Rev. 2 Page 2 of 2
MOVIPLAN 800
Electrical diagram 3
BA version
Code 39719128 - Rev. 2 Page 1 of 1
MOVIPLAN 800
Components plate layout 4
TF/BF versions
Code 72711366 - Rev. 1 Page 1 of 1
MOVIPLAN 800
Electrical diagram 5
TF version
Code 39719123 - Rev. 1 Page 1 of 1
MOVIPLAN 800
Electrical diagram 6
BF version
Code 39719124 - Rev. 1 Page 1 of 1
MOVIPLAN 800
Potter Bucky 7
components layout
Code GR115 - Rev. 1 Page 1 of 1
MOVIPLAN 800
Potter Bucky 8
electrical diagram
Code 39209043 - Rev. 0 Page 1 of 1
MOVIPLAN 800
A4 Board 9
components layout
Code 58714020 - Rev. 1 Page 1 of 1
MOVIPLAN 800
A4 Board 10
schematic
Code 39719125 - Rev. 1 Page 1 of 1
MOVIPLAN 800
A5 Board 11
components layout
Code 58711344 - Rev. 0 Page 1 of 1
MOVIPLAN 800
A5 Board 12
schematic
Code 39719120 - Rev. 0 Page 1 of 1
MOVIPLAN 800
Potter Bucky board 14
schematic
Code 39209044 - Rev. 0 Page 1 of 1
MOVIPLAN 800
A9 Board 15
layout and schematic
Code 58112192 - Rev. 0 Page 1 of 1
MOVIPLAN 800
Overvoltage board 16
layout and schematic
Code 58711348 - Rev. 0 Page 1 of 1
1 – TABLE
2 – CABINET
3 – COLUMN
4 – POTTER BUCKY
1 - TABLE
D 4233811100 Microswitch X X
F 4271126100 Reflector X X X X
E A B D
F C G
T1 4492820100 Transformer T1 X X
V3 4165121100 Rectifier V3 X X
2 - CABINET
C 4890306300 Wheel X X
R1 4192108200 Resistor R1 X X
T1 4492818300 Transformer T1 X X
V1 ÷ V3 4165121100 Rectifier V1 ÷ V3 X X
3 - COLUMN
2 2374862900 Lamp X X
7 6174862100 Inclinometer X X
15 6374850500 Knob X X
4 – POTTER BUCKY
E 4291305400 Microswitch S1 X X X X
F 4291305500 Microswitch S2 X X X X
A B
D C E F H G