You are on page 1of 1

¿What is it?

Cleaning
Proper handling, cleaning and Clean the exterior surfaces of the
maintenance of surgical Century Console Base after each use.
instruments subjected to the Clean with a damp cloth or sponge.
pediatric heart/lung machine is
Acceptable cleaning solutions include
essential to prolong its useful life,
water, mild detergents, alcohol
which will help us prevent and
avoid possible complications with solutions, and dilute sodium
end users who are patients. hypochlorite solutions.

HEART/LUNG ¿To do? Preventive Maintenance


PEDIATRIC Blood and organic elements
should be removed as soon as
Preventive maintenance procedures and
inspections should be performed annually or
possible after use.
MACHINE Do not clean with saline solution
every 1,000 hours of use, whichever comes
first. These inspections should only be
Use distilled water or sterile
water to remove bioburden
performed by a qualified technician.
Instruments with a lumen must
be vacuumed and rinsed with
distilled/sterile water, thus
preventing them from drying out.
Other maintenance measures
Use disinfectants and sterilizing solutions
that guarantee adequate disinfection and
sterilization, but that do not cause
corrosion to the instruments submitted
to the heart/lung machine.

Bibliographies: https://equipos-biomedicos.com.mx/cuidado-y-mantenimiento-del-material-quirurgico/