Professional Documents
Culture Documents
ABSTRACT OF PAPER
A gap method of measuring high voltage is desirable in
certain tests because a gap measures the maximum point of the
voltage wave.
The needle gap is unreliable, and is subject to large cor-
rections due to humidity, etc. The sphere gap is free from
complicated corrections and the curve is subject to calculation.
Wherever voltage is measured by a gap, correction must be
made for air density-that is, altitude or barometric pressure,
and temperature. The laws for this correction have been
deduced for the sphere gap, from tests, over an air density range
equivalent to that from sea level to 18,000 ft. elevation. An
equation is given for calculating sphere spark-over curves for
various spacings, radii, air density, etc.
Standard measured curves for convenient sizes of spheres at
sea level are given for practical use, as well as a table for applying
these curves to any altitude. The altitude correction may be
made very accurately to the standard curve, either by applying
the calculated correction constant or by applying the correction
factor given in the tables.
The complete standard curves should not be calculated when
reliable measured curves are available.
The effects of high frequency and impulse voltages are dis-
cussed. In making measurements certain precautions are neces-
sary for accuracy, as well as for protection of apparatus. These
precautions are discussed and test results are given.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
924 PEEK: THE SPHERE GAP [June 24
replaced after each discharge; the gap becomes very long at high
voltages, and the calibration varies somewhat with the sharpness
of the needle.
The effect of humidity is shown in Fig. 1, where it can be seen
that a higher voltage is required to spark over a needle gap when
the humidity is high than when it is low. (Curves of Fig. 1
are intended only to illustrate this effect and not for use in
making accurate measurements). It is probable that the
corona streamers in humid air cause a "fog ", and then agglom-
erate the water particles,
which, in effect, increase the t1o,PERCENT
size of the electrodes. 200
All spark curves of what- 8 _
ever form of gap must be cor- _ - _
rected for air density; that is, t60 _ i wio i
altitude and temperature. - - - _ 1
For low voltages the spark- 140 7
over voltage of the needle gap cI i / -
decreases approximately as _ _1 T
the air density. At higher I o - 1 / _ -
voltages the effect becomes I / _ _ __
erratic, probably due to t
humidity.2 -
THE SPHERE ~GAP --'- --
The voltage required to 4_
spark over a given gap be- 20-
tween spheres increases with -
the diameter of the sphere. o__
0 10 II20 30 40 50 to
Corona cannot form on SPAC1iG,Cm.
spheres, or rather, the spark- FIG. 1-APPROXIMATE NEEDLE GAP
over point and corona point CURVES FOR DIFFERENT RELATIVE
* if
are coincident * the .
spacing ~~~H UMIDITI ES
is not greater than the diameter of the sphere. Practically
a spacing as great as three times the radius may be used
without appreciable corona. The voltage limit of a given
sphere in high-voltage measurements is thus reached when a
gap setting greater than three times the radius is required. A
larger sphere should then be used.* With this space limit, the
first evidence of stress is complete spark-over; corona can never
form, and all of the undesirable effects and variables due to brush
*More accurate and consistent results are obtained, however, for
spacings less than 2 R.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GA P 925
MOO -- - .
'a I
4 a IZ 16 -~~~~~~~I
201001 24 28 52 3
0 40 lz 16 20 24 28 3'2 58
SPACING. Cm.
FIG. 2-SPARK-OVER CURVES
25 cm. diam. spheres
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
926 PEEK: THE SPHERE GAP (June 24
TABLE I
SPHERE-GAP SPARK-OVER VOLTAGES
6.25-cm. DIAMETER SPHERES.
Spacing Kilovolts effective
0.5 0.197 12 12
1. 0.394 22.5 22.5
1.5 0.591 31.5 31.5
Each point is the average of five readings. The average variation between maximum
and minimum for a given setting is less than 0.5 per cent.
TABLE II
SPHERE-GAP SPARK-OVER VOLTAGES.
12.5-cm. DIAMETER SPHERES.
Spacing Kilovolts effective
I 4 1.575 76 75
5 1.969 91 89
6 2.362 105. 102
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GAP 927
Then e=
9$-x f kv. per cm.
2. 0.787 42 42
2.5 0.983 52 52
3 1.181 61 61
4. 1.575 78 78
5 1.919 96 94
6 2.362 112 110
7.5 2.953 135 132
10. 3.937 171 166
12.5 4.92 203 196
15. 5.91 230 220
17.5 6.88 255 238
20. 7.87 278 254
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
928 PEEK: THE SPHERE GAP [June 24
TABLE IV
SPHERE-GAP SPARK-OVER VOLTAGES.
50-cm. DIAMETER SrHERES
Spacing Kilovolts effective
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GA P 929
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
930 PEEK: THE SPHERE GAP [June 24
TABLE VI
SPHERE-GAP GRADIENT FACTOR.
*Non-grounded: f 3 +1 + + 1 + 8)
I fo experimental values.
One sphere grounded: = X + + 4)
fl should not be used-see note above. fo should be used for the grounded case.
The experitnental values were determined with the grounded sphere four to five
diameters above ground.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GA P 931
/ 0.54\
gs = 19.3 (1 + )kv. per cm. effective
3.92 b
5a = 273+t
=
/ 0.54 0.5
(4)
=ea, =ek4
0
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
TABLE VII
1,500 71.77 28.25 0.948 0.946. 0.945 0.944 0.9493 0.942 0.942
2,000 70.42 27.72 0.932 0.929 0.927 0.926 0.925 0.924 0.924
2,500 69.09 27.20 0.916 0.913 0.911 0.909 0.908 0.907 0.907
3,000 67.74 26.67 0.902 0.899 0.897 0.895 0.893 0.892 0.891 j
3,500 66.51 26.18 0.887 0.881 0.882 0.880 0.878 0.876 0.875 U
4,000 65.25 25.69 0.873 0.870 0.867 0.865 0.863 0.861 0.860
4,500 64.01 25.23 0.859 0.855 0.852 0.850 ,0.848 0.846 0.845 ;
5,000 62.79 24.72 0.845 0.841 0.838 0.835 0.833 0.831 0.830
6,000 60.45 23.80 0.817 0.812 0.808 0.805 0.803 0.801 0.800
7,000 58.22 22.93 0.791 0.786 0.782 0.779 0.776 0.774 0.772
8,000 56.03 22.05 0.765 0.759 0.754 0.750 0.748 0.746 0.744
9,000 53.84 21.20 0.739 0.733 0.728 0.724 0.721 0.719 0.717
10,000 51.85 20.41 0.716 0.709 0.703 0.698 0.694 0.692 0.690 _
12,000 48.09 18.93 0.669 0.661 0.656 0.651 0.647 0.644 0.642 =
15,000 42.88 16.88 0.606 0.596 0.589 0.585 0.580 0.577 0.575 r
Authorized licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restrictions apply.
1914] PE-EK: THE SPHERE GA P 933
age strikes across a given gap, its value, el, may be found by tak-
ing the voltage e, corresponding to the gap, from the standard
curve and multiplying by the correction factor k, or a curve may
be plotted corresponding to a given barometric pressure, thus
e = ek.
TABLE VIII
Calculated values of k for different values of a
+0.541
k= a5 _ 8R !
+
0.54
I%/R j
a R + 0.54
o/R+ 0 54
.~~~~~~~~~~~~~~
Relative Values of k.
air
density Diameter of standard spheres in cm.
Table VIII gives the correction factor for various values of &
The value of 6 may be calculated for the given temperature and
barometric pressure, and a correction factor then found from the
table. Fig. 3 gives the standard curve for the 12.5-cm. sphere
(non-grounded, 25 deg. cent., 76 cm. bar. pressure) and curves
calculated therefrom for 25 deg. cent. and various barometric
pressures.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
934 PEEK: THE SPHERE GAP [June 24
G G - __ I_ a 45.76C om=.tOG
SPAC-ING,
S8CM.E
- - V ___664Cm. A8
~~ I5C - - -
=m09
~~~~~~ 53.2 Cl dQ. 7
-
80Cm.d-Q05
0 200 15 2
PAIGCm.
FIG. 3-SPHERE GAP SPARK-OVER VOLTAGE, AT VARIOUS
AIR PRESSURES
Temperature 25 deg.cent. 12.5-cm. spheres. Non-grounded. Drawn curves calculated.
76 cm. is standard curve. Points are from curves through measured values.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GA P 935
TABLE IX
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS
2.54-cm. DIAMETER SPHERES-NON-GROUNDED.
TABLE X
SPHERE-GAP SPAR K-OVER VOLTAGES AND GRADIENTS
2.54-cm. DIAMETER SPHERES-NON-GROUNDED.
Relative g
Spacing air kv. eff. max.
cdo. in. density eff. mrax. measured
ss
if
0.875 45.7 64.6 23.5 36.0
0.834 45.2 64.0 25.2 35.7
0.798 42.8 60.6 23.8 33.8
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
936 PEEK: T/HE SPIIERE GA P [June 24
TABLE XI
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS
5.08-cm. DIAMETER SPHERES-NON-GROUNDED
TABLE XII
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS
12.5-cm. DIAMETER SPHERES- NON-GROUNDED.
Relative gs
Spacing air kv. eff. max.
cm. in. density eff. max. measured.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
19143 PEEK: THE SPHERE GA P 937
TABLE XIII
SPHER E-GAP SPARK-OVER VOLTAGES AND GRADIENTS.
12.5-cm. DIAMETER SPHERES-NON-GROUNDED.
Relative gs
Spacing air kv. eff. max.
cm. in. density eff. max. measured
TABLE XIV
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS.
12.5-cm. DIAMETER SPHERES-NON-GROUNDED.
Relative gs
Spacing air kv. eff. max.
cm. in. density eff. max. measured
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
938 PEEK: THE SPHERE GAP [June 24
TABLE XV
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS.
12.5-cm. DIAMETER SPHERE-NON-GROUNDED.
Relative gs
Spacing air kv. eff. max.
cm. in. density eff. max. measured
TABLE XVI
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS
12.5-cm. DIA-METER SPHERES-NON-GROUNDED.
Relative gs
Spacing air kv. eff. max.
cm. in. density eff. max. measured
20.32 8 0.938 198.5 281 23.0 32.6
0.912 197 279 22.9 32.4
0.874 189.5 268 22.0 31.1
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GA P 939
'I'ABLE XVII
SPHERE-GAP SPARK-OVER VOLTAGLES AND GRADIENTS
12.5-cm. DIAMETER SPHERE-GROUNDED.
Relative gs
Spacing air kv. eff. max.
cm. in. density eff. max. measured
TABLE XVIII
SPHERE-GAP SPARK-OVER VOLTAGES AND GRADIENTS.
25-cm. DIAMETER SPHERE-NON-GROUNDED.
Relative gs
Spacing air kv. eff. max.
crn. in. density eff. max. measured
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
940 PEEK: THE SPHERE GAP [June 24
40
o. o
30~~~~~~~~
Tphc
equation for the correction factor was deduced fromn
measulred values as follows:
From a former investigation7 it was found that at 6 - 1 the
average gradient
g5 = go(i ± _-) (1)
gS= galv R
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
19141 PEEK: THE SPHERE GA P 941
4--
17.78
cm.1( Ccm.
7.62
cm. (3 in.) Cin
0 0.5
RTLATIVE AR DENSITYD
1VRN-SPARK-OVER
IG
5C
VOLIAGE WITH VARYING AIR DENSITIES
2032cm (mi.
15/16
: 2.5-cm. (4 6 in.) spheresr Non-grounded
254 cm. (1n.)
5.08 cm. (2 in.)
7.62 cm. (3 in.)
Spacing 1o0.16 cm. (4 in.)
12.7 cm. (5 in.)
17.78 cm. (7 in.)
(20.32 cm. (8 in.)
50 O~~50
o o. - .o o - .
-.
~~~~~~~~~~~I 5C
f1 -
12jCms
7.62 Ce
50
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
942 PEEK: THE SPHERE GAP [June 24
TABLE XIX
AVERAGE EF;FECTIVE RUPTURING-GRADIENTS FOR SPHERES OF SEVERAL DIAMETERS
AND VARYING AIR DENSITIES.
Diameter of Spheres, cm.
2.54 5.08 12.5 25
6
Surface gradients
Meas. Calc. Meas. Calc. Meas. Calc. Meas. Calc.
0.54
Columns marked Calc. " are from gs - 19.3 6 (1 +
30 -
1~~~~~~~~2.54
Cm. Dio
_50 i _-
0 0.5 1.0
RELATIVE AIR OEMSITY&
FIG. 9-SURFACE GRADIENT AT VARYING AIR DENSITIES FOR SEVERAL
SIZES OF SPHERES
Drawn curves calculated. Points measured values:
F-~~~ ~ ~ ~~~~~~0
Si 20 09z
d'= 0. 5
0 5 lO .15 20 25
DIAMETER OF SPHERES Cm.
FIG. 10-SURFACE GRADIENT FOR VARYING SIZES OF SPHERES AT SEV-
ERAL AIR DENSITIES.
0.54X
Drawn lines calculated from formula gs = 19.36 1 + R)
Points measured values.
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
19141 PEEK: THE SPHERE GA P 943
20 t0208
Ui Z0
0~~~~~~~~~~~~~1
02 0. 0j6 - -
S
04 N6 C.( L
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
944 PEEK: TIHE SPHERE GAP [June 24
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GA P 945
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
946 PEEK: THE SPHERE GAP [June 24
Gen.
Tesf Trans former
F19
CGs.
feasuring "ap g pr o
r-1C. /Cm.-_;
9Q2oAsprV/
02OM
k---105 Cm_-
ll
fote
pont *rs132
Cmo8n
ih-enin
.5 m- ->j "-.025-
te fre b
inig OH/MS fher Vo/O troansfre ne
1 50'Cm.--- - '(,C- Verl/ Smnall
n te s ar di.s a OHf per ovolt
FIG. 1 3
former under test until the " measuringF gap sparks over.
4, are brougit out at equidistant
Insulated taps, 1, 2, 3, r5,
points from the high-tension winding 'of the transformer under
test. Auxiliary needle gaps are placed between I anld 2, 2 and
3, and 1 and 3, to measure the voltage which appears across these
sections of the winding when the main measuring gap discharges.
The numbers on the figure between 1 -2, 2 -3 and 1 -3, rep-
resent the sparking distances of the local voltages caused
by a discharge of the measuring gap. Four cases are given
with different values of resistance w in the main gap. When
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
19141 PEEK: THE SPHERE GA P 947
w equals one ohm per volt the local voltages are completely
damped out.
With small resistance in the main gap, a 19-cm. spark-over
causes a voltage to build up between coils one and three which
sparks over a 150-cm. gap, although the total applied voltage
across the transformer is only equivalent to a 19-cm. gap. The
apparatus may thus be subjected to strains far beyond reason,
and either broken down or very much weakened. Water-tube
resistance is the most reliable. A metallic resistance, if non-
inductive and of small capacity, may be used. Carbon or
graphite rods should be avoided, because, although they may
measure up to a very high resistance at low voltage, the re-
sistance may become very low at high voltage by " coherer "
action. Where possible the gap should be used only to measure
the equivalent transformer ratio and not during test. When the
tested apparatus is of such nature that there is considerable
incipient arcing before spark-over, it is better to use the spheres
to determine the " equivalent ratio " of the transformer below
the voltage at which this arcing occurs. It must be remembered
that resistances do not dampen out low-frequency surges resultinrg
from a short circuit, etc.
TABLE XX.
POSSIBLE PER CENT ERROR DUE TO PLATES ON
6.25-cM. SPHERE SHANKS.
Non-grounded Grounded
Spacing 5 cm. diameter plate 5 cm. diameter plate
cm. 6.25 cm. back of 6.25 cm. back of
both one insulated grounded both
spheres sphere sphere sphere spheres
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
948 PEEK: THE SPHERE GAP [June 24
MISCELLANEOUS PRECAUTIONS
In making tests it is desirable to observe the following pre-
cautions:
The shanks should not be greater in diameter than 1/5 the
sphere diameter. Metal collars, etc., through which shanks
extend, shouldbe as small as practicable, and should not come
closer to the sphere than the gap length. For the non-grounded
case the effect of such objects is a minimum when symmetrical on
either side. For the grounded case the effect is considerable.
The effect of a large plate or plates on the shanks is given in
Table XX. This should show the maximum error in all cases,
as the maximum gap setting, and the extreme diameter of
collar that is likely to be used in practise, is taken. The
sphere diameter should not vary more than 0.1 per cenit, and
the curvature, measured by a spherometer, should not vary
more than I per cent from that of a true sphere of the re-
quired diameter. The spheres should be twice the gap setting
from surroundings.
In Table XX the spacing Xis given in terms of radius R, to make
the correction applicable to any size of sphere. The distance of the
grounded sphere above ground is, for the same reason, given in
terms of the sphere diameter. The (+) sign means that a higher
voltage is required to arc over the gap than that giveln by the
standard curve. The (-) sign indicates that the spheres read
too low. The standard curv,es were made with the grounded
spheres from four to five diameters above ground. In practise
it is desirable to work between four and teln diameters; never
ulnder three. Above ten the variation in per cent error remains
about the same.
When both spheres are insulated, with the transformer neutral
at the mid point, there is practically no variation in voltage for
different distances above ground.
Great precautions are necessary at very high voltages to pre-
vent leakage over stands, supports, etc., and to prevent corona
and brush discharges. Care must also be taken to so place the
spheres that external fields are not superposed above the sphere
gap. This is likely to result, especially in the non-grounded
case, from a large mass of resistance units or connecting leads,
etc., in back of and in electrical connection with either sphere.
The 'error may be either plus or minus, as indicated for the small
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction
1914] PEEK: THE SPHERE GAP 949
zed licensed use limited to: BFH Bern. Downloaded on February 02,2022 at 14:38:01 UTC from IEEE Xplore. Restriction