Calibration Report of Dwyer AVUL-3DA1-LCD Air Velocity Transmitter
MPC Control #: 036R7G3DA1 Serial Number: 036R7G3DA1
Asset ID: N/A Calibration Date: October 31, 2020
Measurement Accuracy Test
Lower Upper Uncertainty Parameter Nominal As Found As Left Result Limit Limit (±) 0.20 m/s -0.010 m/s 0.183 m/s 0.183 m/s 0.410 m/s PASS 0.0026 m/s 2.50 m/s 2.225 m/s 2.476 m/s 2.476 m/s 2.775 m/s PASS 0.033 m/s Air Velocity 5.00 m/s 4.650 m/s 4.986 m/s 4.986 m/s 5.350 m/s PASS 0.065 m/s 10.00 m/s 9.500 m/s 9.977 m/s 9.977 m/s 10.500 m/s PASS 0.13 m/s 15.00 m/s 14.350 m/s 14.848 m/s 14.848 m/s 15.650 m/s PASS 0.20 m/s
- End of Calibration Report -