Uncertainty
Nominal Lower Limit As Found As Left Upper Limit Result
(±)
0.00 m/s -0.200 m/s 0.000 m/s 0.000 m/s 0.200 m/s PASS 0.006 m/s
2.50 m/s 2.225 m/s 2.534 m/s 2.534 m/s 2.775 m/s PASS 0.038 m/s
5.00 m/s 4.650 m/s 4.942 m/s 4.942 m/s 5.350 m/s PASS 0.071 m/s
10.00 m/s 9.500 m/s 9.938 m/s 9.938 m/s 10.500 m/s PASS 0.136 m/s
15.00 m/s 14.350 m/s 14.907 m/s 14.907 m/s 15.650 m/s PASS 0.201 m/s
The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution
corresponds to a coverage probability of approximately 95% , unless otherwise stated.
This calibration report complies with ISO/IEC 17025:2017 and ANSI/NCSL Z540.3 Method 6-Guard Bands based on Test Uncertainty Ratio.