ORIGINAL
CERTIFICATE NUMBER: APL-C7-22-109

CALIBRATION CERTIFICATE
A. CUSTOMER
COMPANY NAME : QUALITY INSPECTORS LIMITED
ADDRESS : P.O BOX 18358-00401 NAKURU
LOCATION : WESTLANDS, NAIROBI.
A. DEVICE INFORMATION
Name : Ultrasonic Flaw Detector
Manufacturer : Sonatest
Serial number : I009422
Model : D-I0+

B. CERTIFICATE INFORMATION
Cal sticker number : APL.C10383
Cal Date : 11/10/2021
Cal Location : APL lab

C. Calibration procedure: This item whose details appear above was calibrated in
accordance with APL-TP-DM-001 Procedure for the calibration of
dimensional and length measuring equipment.
D. Reference standard: ASME B89.1.10M linear measurements.
E. Calibration Standard and Traceability: Set of Angle Gauge Blocks: S/No:
080154-3, traceable to international standards, certificate number
080154/17/15373.
F. Environmental conditions: Calibration was carried out under controlled
environmental conditions as indicated below:
Temperature: 20.0°C ±1°C
Relative 53 % ±10%
Humidity:
G. Results validation: This Certificate is valid until 10th October 2023.
Calibration by : J. JUMA
Signature :
Checked by :D. ONDITI
Signature :
Authorized/Validated By : C. ORINA
Signature : Date :12/10/2022
MANAGING DIRECTOR

Calibration certificate without signature and seal is not valid. This certificate has been issued without any alteration
and may not be reproduced other than in full except with the approval of the Managing Director, Avioteq Precision Ltd.
This Certificate only relates to the calibration status of the item/equipment whose details appear above.
Page 2 of 2
CERTIFICATE NUMBER: APL-C7-22-109
H. CALIBRATION DATA

a. Calibration readings:
Std reading Reading
Error
(mm) (mm)
2.5 2.51 -0.01
5.1 5.12 -0.02
7.7 7.71 -0.01
10.3 10.30 0.00
15.0 15.00 0.00
20.2 20.21 -0.01
25.0 25.10 -0.10

I. UNCERTAINTY STATEMENT:
The Measurement of Uncertainty of the calibration results for the given range
of measurement is ±0.05. The reported expanded uncertainty is stated as
expanded uncertainty of the measurement multiplied by a coverage factor
K=2, providing a confidence level of approximately 95%.

J. REMARKS AND OBSERVATIONS:


The results were correct during the time of calibration, and will remain
valid depending on care and handling exercised during equipment use.
No any adjustment was done to the instrument.

Issuance of this certificate indicates that the equipment calibrated


complies with specified requirements. When the equipment does not
comply with requirement, no certificate shall be issued instead a
calibration report will be issued.
………………………………………………………. END ……………………………………