3.0 RESULTS
Pre-adjustment results
Value of the standard (g) Balance reading (g) Deviation (g)
100.0 99.8 -0.2
500.0 499.0 -1.0
1 000.0 998.1 -1.9
Calibration Results
Linearity
Value of the standard (g) Balance reading (g) Deviation (g)
10.0 10.0 0.0
50.0 50.0 0.0
100.0 100.0 0.0
200.0 200.0 0.0
500.0 500.0 0.0
1 000.0 1 000.0 0.0
Repeatability test Eccentricity test
Repeatability Error = 0.0 g Eccentricity Error = 0.1 g
3.1 Remarks
The expanded uncertainty of the instrument based on a standard uncertainty multiplied by a coverage factor of k = 2, at a level of
confidence of approximately 95% is ± 0.1 g. This was calculated and expressed in accordance with the BIPM, JCGM 100: 2008, and
“Evaluation of measurement data - Guide to the expression of uncertainty in measurement”.
The values in this certificate are correct at the time of calibration. Subsequently the accuracy will depend on such factors as the care
exercised in handling and use of the instrument and the frequency of use. Re-calibration should be performed after a period, which has
been chosen to ensure that the instrument's accuracy remains within the desired limits.
NOTE:
This certificate of calibration 01220745 is valid for the balance with Identification: MWT-GU-BAL03 only; and calibration performed
on-site as per stated location above.
4.0 TRACEABILITY
To Reference Standards of Serial No. 80125228 and 80125227 with Calibration Certificate Number MV\M-5465.
-------------------------------------------------------END OF CERTIFICATE---------------------------------------------------