10151

Seite 2 von 5 Kalibrierschein vom 2021-08-19 D-K-


Page Calibration Certificate from 19153-01-00

2021-05

1 Calibration object
The calibration object (CO) is an measuring instrument for the measurement of testing
alternating voltage and testing direct voltage according to IEC 60060-2 and IEC 61083-3,
Draft.
Type: MU 18 DP
Serial no: 103 191
Input: X1
Input impedance: 10 MΩ || 50 pF
Software/Firmware: A1 version 4.3
The alternating voltages or direct voltages to be measured were led to the input X1 of the
measuring instrument.
The input impedance of the input X1 amounts 10 MΩ in parallel to 50 pF.
According to the technical specification of the measuring instrument the limit of the
uncertainty of the measurement amounts to 0.5 % and the maximum admissible systematic
measuring deviation amounts to  0.5 % for the measuring quantities average for direct
voltages and 𝑈⁄√2 and 𝑈 . . . (true r.m.s. value) for alternating voltages.

2 Calibration procedure
The measuring instrument was calibrated according to IEC 60060-2 by comparison
measurement against a Reference Multi-Product-Calibrator Wavetek 9100 (calibration mark
7314/D-K-18673-01-00/2020-12). The calibration object was connected using a coaxial cable
with a length of 1 m, a wave resistance of 75 Ω and a capacitance of 72 pF/m.

Calculation of the relative measuring deviations:

𝛿𝑈 ∙ 100 % (1)

𝑈 value measured by the calibration object


𝑈 measuring value of the standard

3 Place of calibration
The calibration was carried out in the calibration laboratory D-K-19153-01-00 of HIGHVOLT
Prüftechnik Dresden GmbH in Dresden.

Calibration Laboratory of HIGHVOLT Prüftechnik Dresden GmbH, Marie-Curie-Str. 10, 01139 Dresden, Germany
Phone +49 351 8425 225, e-mail: calibration@highvolt.com, website: http://calibration.highvolt.com
10151
Seite 3 von 5 Kalibrierschein vom 2021-08-19 D-K-
Page Calibration Certificate from 19153-01-00

2021-05

4 Measuring conditions
Direct voltage:
 Check of the voltage dependent non-linearity of the CO for the measuring quantity 𝑈
(average) in the voltage range between 10 V and 995 V
Alternating testing voltages:
 Check of the voltage dependent non-linearity of the CO for the measuring quantities
𝑈/√2, 𝑈/√2, 𝑈 . . . (true r.m.s. value) in the voltage range between 10 V and 700
V at a frequency of 50 Hz
 Check of the amplitude-frequency response of the CO for the measuring quantities
𝑈/√2, 𝑈/√2, 𝑈 . . . (true r.m.s. value) in the frequency range between 10 Hz and
500 Hz

5 Environmental conditions
Ambient temperature: (24  2) °C
Relative humidity: (43  20) %

6 Measuring results
6.1 Calibration at direct voltage
Table 1: Comparison measurements at DC, check of the voltage dependent non-linearity
𝑈 𝑈
𝑈 /V 𝑈/V 𝛿𝑈 / % 𝑈/V 𝛿𝑈 / %
10 10.00 0.00 9.988 -0.12
20 19.99 -0.05 19.98 -0.10
30 29.99 -0.03 29.98 -0.07
50 50.00 0.00 49.97 -0.06
70 70.00 0.00 69.97 -0.04
100 100.00 0.00 99.98 -0.02
200 200.00 0.00 200.00 0.00
300 300.10 0.03 300.00 0.00
500 500.10 0.02 500.00 0.00
700 700.30 0.04 700.20 0.03
995 995.60 0.06 995.50 0.05
Evaluation: The voltage dependent non-linearity is within 0.12 %.

Calibration Laboratory of HIGHVOLT Prüftechnik Dresden GmbH, Marie-Curie-Str. 10, 01139 Dresden, Germany
Phone +49 351 8425 225, e-mail: calibration@highvolt.com, website: http://calibration.highvolt.com
10151
Seite 4 von 5 Kalibrierschein vom 2021-08-19 D-K-
Page Calibration Certificate from 19153-01-00

2021-05
6.2 Calibration at alternating voltage
Table 2: Comparison measurements at AC 50 Hz,
check of the voltage dependent non-linearity
𝑈/√2 𝑈/√2 𝑈. . .

𝑈 /V 𝑈/V 𝛿𝑈 / % 𝑈/V 𝛿𝑈 / % 𝑈/V 𝛿𝑈 / %


10 9.996 -0.04 9.998 -0.02 10.005 0.05
20 19.99 -0.05 19.99 -0.08 20.02 0.08
30 29.99 -0.03 29.98 -0.08 30.04 0.12
50 50.00 0.00 49.97 -0.06 50.05 0.10
70 70.01 0.01 69.98 -0.04 70.09 0.13
100 100.00 0.00 99.94 -0.06 100.10 0.10
200 200.10 0.05 200.00 0.00 200.30 0.15
300 300.10 0.03 300.10 0.03 300.30 0.10
500 500.20 0.04 500.00 0.00 500.70 0.14
700 700.40 0.06 700.25 0.04 700.90 0.13
Evaluation: The voltage dependent non-linearity is within 0.15 %.

Table 3: Comparison measurements at 100 V AC,


check of the frequency dependent non-linearity

𝑈/√2 𝑈/√2 𝑈. . .

𝑓 / Hz 𝑈 /V 𝑈/V 𝛿𝑈 / % 𝑈/V 𝛿𝑈 / % 𝑈/V 𝛿𝑈 / %


10 100.00 0.00 99.91 -0.09 100.10 0.10
20 100.00 0.00 99.91 -0.09 100.10 0.10
40 100.10 0.10 99.92 -0.08 100.10 0.10
50 100.00 0.00 99.93 -0.07 100.10 0.10
70 100 100.00 0.00 99.97 -0.03 100.10 0.10
100 100.00 0.00 99.99 -0.01 100.00 0.00
200 100.00 0.00 100.00 0.00 100.00 0.00
300 99.95 -0.05 99.95 -0.05 99.95 -0.05
500 100.00 0.00 99.90 -0.10 99.89 -0.11
Evaluation: The frequency dependent non-linearity is within 0.11 %.

Calibration Laboratory of HIGHVOLT Prüftechnik Dresden GmbH, Marie-Curie-Str. 10, 01139 Dresden, Germany
Phone +49 351 8425 225, e-mail: calibration@highvolt.com, website: http://calibration.highvolt.com
10151
Seite 5 von 5 Kalibrierschein vom 2021-08-19 D-K-
Page Calibration Certificate from 19153-01-00

2021-05

7 Measurement uncertainty
The relative measurement uncertainty (related to the measuring value) amounts
for direct voltage 0.32 %
for alternating voltage 0.42 %
The uncertainty stated is the expanded uncertainty of measurement obtained by multiplying
the standard uncertainty by the coverage factor k = 2. It has been determined in accordance
with EA-4/02 M:2013 and in accordance with the ISO/BIPM guide GUM (“Guide to the
Expression of Uncertainty in Measurement”) considering contributions to the uncertainty of
measurement resulting from the Standard, the resolution of the instrument, the expected drift
of the instrument, the voltage dependent non-linearity, the frequency dependent non-linearity
and the influence of the temperature in the range 24 °C ± 10 K. Normally, the value of the
measurement lies within the assigned range of values with a probability of 95 %.
The uncertainty stated is related to the moment of the calibration and does not include any
portion for the long-term instability of the object calibrated.

8 Declaration of conformity
The calibration object fulfils the demands of 60060-2 to a measuring instrument for the
measurement of alternating testing voltages and direct testing voltages for the documented
range of calibration and conditions.

9 Hint concerning recognition of the DAkkS-Certificates abroad


Germany’s Accreditation Body DAkkS is signatory to the multilateral agreements of the
European co-operation for Accreditation (EA) and of the International Laboratory
Accreditation Cooperation (ILAC) for the mutual recognition of calibration certificates. The
further signatories inside and outside Europe are listed on the websites of EA
(www.european-accreditation.org) and ILAC (www.ilac.org).

--------------------------------------------------------- End of this Document ---------------------------------------------------------

Calibration Laboratory of HIGHVOLT Prüftechnik Dresden GmbH, Marie-Curie-Str. 10, 01139 Dresden, Germany
Phone +49 351 8425 225, e-mail: calibration@highvolt.com, website: http://calibration.highvolt.com