Bahra Cables Company ‫شركة بحرة المتطورة لصناعة للكابالت‬

BCC

CALIBRATION CERTIFICATE

Test Object : MILI OHM METER

Manufacturer : Schuetz Messtechnik Capacity : 1-10-100 mΩ


1-10-100 Ω
Model No. : MR 270 C-A 1-10 kΩ

Serial No. : 1706-6523 Accuracy : ± 0.03% from MV @ 1 mΩ


± 0.02% from MV for remaining

Calibration Date : 18/6/2023 Calibration Due Date : 18/9/2023

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (45 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by:

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge S.VP Quality & Technology

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة للكابالت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Standard Resistance
No. Serial No. Calibration Date Calibration due Date
Description Nominal Value (Ω)

Calibration
1 1m 0906-5233 10/10/2022 10/10/2024
Resistor
Calibration
2 10 m 0906-5234 10/10/2022 10/10/2024
Resistor
Calibration
3 100 m 0906-5235 11/10/2022 11/10/2024
Resistor
Calibration
4 1 0906-5236 11/10/2022 11/10/2024
Resistor
Calibration
5 10 0906-5237 11/10/2022 11/10/2024
Resistor

Resistance Measurements :

Standard
Standard
Calibration
Calibration Indicated
Resistance Error (Ω) ± Acceptable Tolerance (Ω) ± Uncertainty (Ω)
Resistance measurement (Ω)
Nominal Value
Value (Ω)
(Ω)

1m 1.00019 m 1.0002 m 0.0000 m 0.0003 m 0.0005943 m

10 m 10.0026 m 10.002 m -0.001 m 0.002 m 0.004448 m

100 m 100.014 m 100.01 m 0.00 m 0.02 m 0.038619 m

1 0.99928 0.9992 -0.0001 0.0002 0.001336

10 10.0016 10.002 0.000 0.002 0.00217

Error =Indicated Measurement - Standard Calibration Resistance Value

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage probability of
approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬
BCC

CALIBRATION CERTIFICATE

Test Object : Digital Caliper

Manufacturer : MITUTOYO Capacity : 0-150 mm

Serial No. : B16100220 Accuracy : ±0.02 mm

Calibration Date : 18/6/2023 Calibration Due Date : 18/10/2023

Temperature : Ambient temperature (23 °C)

Humidity : Ambient humidity (46 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge S.VP Quality & Technology

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Set of Slip Gauges 2899 ±0.001 20/7/2020 20/7/2023

Digital Caliper Measurements :

Standard Gauge Block Indicated measurement Acceptable Tolerance ± Max Uncertainty


Error (mm)
thickness value (mm) (mm) (mm) (mm)

1.08 1.07 -0.01

1.90 1.91 0.01

5 4.99 0

20 19.99 0 ± 0.02 0.0074

50 50.01 0

100 99.99 0

150 150.01 0

Error = Indicated Measurement-Standard gauge block thickness

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : Digital Micrometer

Manufacturer : Mitutoyo Capacity : 0-25 mm

Serial No. : 75234449 Accuracy : ±0.002 mm

Calibration Date : 9/8/2023 Calibration Due Date : 9/12/2023

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (45 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the
corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge Quality manager

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Gauge Block Set 142500 ±0.001 6/7/2023 6/7/2024

Digital Micrometer Measurements :

Standard Gauge Block Indicated measurement ± Acceptable


Error (mm) ± Max Uncertainty (µm)
thickness value (mm) (mm) Tolerance (µm)

1.008 1.007 -0.001

1.09 1.090 0.00

5 5.001 0 0.82449 2

10 9.998 0

25 25.001 0

Error = Indicated Measurement-Standard gauge block thickness

The acceptable tolerance is the maximum allowable error.

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬
BCC

CALIBRATION CERTIFICATE

Test Object : DC DIELECTRIC TEST SET

Manufacturer : Phenix Capacity : 0-40 KVdc

Serial No. : 14-8770 Accuracy : ±( 0.01 for 20 KV range, 0.1


for 40 KV rage)

Calibration Date : 18/10/2022 Calibration Due Date : 18/10/2023

Temperature : Ambient temperature (29 °C)

Humidity : Ambient humidity (48 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar

Calibration In Charge Quality manager

Calibration Certificate is electronically signed.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (Kv)
Date Date

1 spark tester calibrator 2300 ±0.01 20/10/2020 20/10/2022

DC Dielectric Test Set Measurements :

spark tester calibrator Acceptable


Indicated Voltage (Kv) Error (%) ± Max Uncertainty (Kv)
Reading (Kv) Tolerance

20 KVdc Range

3.00 3.01 -0.33

5.00 5.02 -0.40

10.00 9.96 0.40 ±3% 0.0152

12.00 12.05 -0.41

18.00 18.06 -0.33

40 KVdc Range

20.0 20.03 -0.15

22.0 22.09 -0.41 ±3% 0.0673

24.0 24.12 -0.50

Error (%) = [(Indicated Reading - Spark tester Calibrator)/Spark tester Calibrator] * 100 %

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the Maximun standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬
BCC

CALIBRATION CERTIFICATE

Test Object : Digital Caliper

Manufacturer : MITUTOYO Capacity : 0-300 mm

Serial No. : 1068014 Accuracy : ±0.02 mm

Calibration Date : 18/6/2023 Calibration Due Date : 18/10/2023

Temperature : Ambient temperature (23 °C)

Humidity : Ambient humidity (46 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge S.VP Quality & Technology

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Set of Slip Gauges 2899 ±0.001 20/7/2020 20/7/2023

Digital Caliper Measurements :

Standard Gauge Block Indicated measurement Acceptable Tolerance ± Max Uncertainty


Error (mm)
thickness value (mm) (mm) (mm) (mm)

1.08 1.09 0.01

1.90 1.90 0.00

5 4.99 0

20 20.01 0

50 50.01 0

100 99.99 0 ± 0.02 0.0755

150 150.01 0

190 189.99 0

240 240.01 0

260 260.01 0

265 265.02 0

Error = Indicated Measurement-Standard gauge block thickness

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : Laboratory Drying Oven

Manufacturer : Thermo Fisher Capacity : ambient to 250 °C

Serial No. : 41168882 Accuracy : ± 0.1 °C (ambient to 99.9 °C )


± 1 °C ((100 to 250) °C )

Calibration Date : 4/5/2023 Calibration Due Date : 4/5/2024

Temperature : Ambient temperature (21 °C)

Humidity : Ambient humidity (47 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration Manager`` S.VP Quality & Technology

Calibration Certificate is electronically signed

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (C°)
Date Date
Digital Portable
1 I103580 ± 0.1 8/5/2022 8/5/2023
Thermometer

Laboratory Drying Oven Measurements :

Standard drywell Indicated measurement Acceptable Tolerance


Error (°C) ± Uncertainty (°C)
calibrator (°C) (°C) (°C)

70.0 70.3 0.3

150.0 150.5 1
±2 1.33
200.0 200.4 0

250.0 250.7 1

The acceptable tolerance is the maximum allowable error


Error = Indicated Reading-Standard Reading

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬
BCC

Calibration Certificate

Test Object : dumbell test piece dimension

Manufacturer : NA Capacity : Specific dimensions

Serial No. : 8705-2 Accuracy : N.A

Calibration Date : 1/5/2023 Calibration Due Date : 1/5/2024

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (47 % RH)

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked By Approved by

Mousa Sharafi Eng. Asaad M.Yasin Eng. Mohammad Asker


Calibration In Charge Calibration manager S.VP Technology & QC

Calibration Certificate is electronically signed.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬

Calibration Certificate

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Digital Caliper 0070840 ± 0.02 2/5/2022 2/5/2023

dumbell test piece dimension Measurements :

Acceptable Tolerance Max. Uncertainity


Dimension (mm) Measured (mm) Error (mm)
(mm) (mm)

A 75.02 0

B 4.01 0
± 0.02 ± 0.0624
C 12.49 0.0

D 12.50 0.0

Error = Indicated dimension (on the figure) - Measured dimension

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : Digital Micrometer

Manufacturer : Mitutoyo Capacity : 0-25 mm

Serial No. : 75234449 Accuracy : ±0.002 mm

Calibration Date : 9/8/2023 Calibration Due Date : 9/12/2023

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (45 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the
corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge Quality manager

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Gauge Block Set 142500 ±0.001 6/7/2023 6/7/2024

Digital Micrometer Measurements :

Standard Gauge Block Indicated measurement ± Acceptable


Error (mm) ± Max Uncertainty (µm)
thickness value (mm) (mm) Tolerance (µm)

1.008 1.007 -0.001

1.09 1.090 0.00

5 5.001 0 0.82449 2

10 9.998 0

25 25.001 0

Error = Indicated Measurement-Standard gauge block thickness

The acceptable tolerance is the maximum allowable error.

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة للكابﻼت‬
BCC

CALIBRATION CERTIFICATE

Test Object : MILI OHM METER

Manufacturer : Chauvin Arnoux

Serial No. : P-25106940-12

Calibration Date : 7/8/2023 Calibration Due Date : 7/11/2023

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (44 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge S.VP Technology & QC

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة للكابﻼت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Standard Resistance
No. Description Nominal Value (Ω)
Serial No. Calibration Date Calibration due Date

Calibration
1 Resistor
1m 0906-5233
10/10/2022 10/10/2024
Calibration
2 Resistor
10 m 0906-5234

Calibration
3 Resistor
100 m 0906-5235

Calibration
4 Resistor
1 0906-5236 11/10/2022 11/10/2024

Calibration
5 Resistor
10 0906-5237

Resistance Measurements :

Standard
Standard
Calibration
Calibration Indicated
Resistance Error (Ω) ± Acceptable Tolerance (Ω) ± Uncertainty (Ω)
Resistance measurement (Ω)
Nominal Value
Value (Ω)
(Ω)

1m 1.00020 m 1.0001 m -0.0001 m 0.0003 m 0.000158 m

10 m 10.0041 m 10.003 m -0.001 m 0.002 m 0.000735 m

100 m 100.014 m 100.02 m 0.01 m 0.02 m 0.00736 m

1 0.99924 0.999 m 0.000 m 0.0002 0.000037

10 10.0015 10.002 0.001 0.002 0.00058

Error = Indicated Measurement - Standard Calibration Resistance Value

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage probability of
approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : WEIGHING BALANCE

Capacity : 5100 g

Manufacturer : Mettler Toledo

Serial No. : 1129250345 Accuracy : ± 0.1 g

Calibration Date : 7/8/2023 Calibration Due Date : 7/2/2024

Temperature : Ambient temperature (23 °C)

Humidity : Ambient humidity (45 % RH)

We certify that above object was calibrated for specific calibration points using certified
standards traceable to international standard.

Calibration results, data and special conditions on the use of object are stated in the
corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.yasin Eng. Mohammad Askar


Calibration In Charge Quality manager

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Test Object : WEIGHING BALANCE

Manufacturer : Mettler Toledo

Serial No. : 1129250345

Measurement Method :

Measuerment method was used, the Balance was calibrated using the
Following Laboratory Refrance Standard:

Refrance Standard
NO Description
Serial No. Calibration Date Calibration Due Date

1 10 g 158427

2 200 g 158467

3 1000 g 158697 26/4/2023 26/4/2026

4 2000 g 158707

5 5 Kg 158717

Result :

Test for Errors of Indications

Acceptable
Refrance Load ( g ) Indicated Value ( g ) Error ( g ) ± Max Uncertainty ( g )
Tolerance ( g )
10.00000 10.0 0.0

200.0000 200.0 0.0

1000.000 1000.0 0.0 ±1 0.07955

2000.000 2000.1 0.1

5000.000 5000.1 0.1

Error = Indicated value - Reference Load


The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported Max expanded uncertainty of measurement is stated as the standard uncertainty
of measurement multiplied by the coverage factor K=2, which for a normal distribution
corresponds to a coverage probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Corner Load Test with Load of : (2) kg

Loading Point No. Deviation from the Corner ( g )

1 0.0

2 0.0

3 0.1

4 0.1 Fig.1

5 0.0

Deviation from the corner = measured weight at each point from Fig.1 - measured weight at the center

Repeatability Test (repeated 5 times) :

The standard deviation was calculated at ( 2 ) kg load


and found to be 0.1 g

Note :

1) The calibration was held on lab

2) The Balance Calibration was performed on a ground level with zero sloop and
no ambient noise such as air or/and sound

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬
BCC

CALIBRATION CERTIFICATE

Test Object : Digital Caliper

Manufacturer : MITUTOYO Capacity : 0-300 mm

Serial No. : 1068014 Accuracy : ±0.02 mm

Calibration Date : 18/6/2023 Calibration Due Date : 18/10/2023

Temperature : Ambient temperature (23 °C)

Humidity : Ambient humidity (46 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge S.VP Quality & Technology

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Set of Slip Gauges 2899 ±0.001 20/7/2020 20/7/2023

Digital Caliper Measurements :

Standard Gauge Block Indicated measurement Acceptable Tolerance ± Max Uncertainty


Error (mm)
thickness value (mm) (mm) (mm) (mm)

1.08 1.09 0.01

1.90 1.90 0.00

5 4.99 0

20 20.01 0

50 50.01 0

100 99.99 0 ± 0.02 0.0755

150 150.01 0

190 189.99 0

240 240.01 0

260 260.01 0

265 265.02 0

Error = Indicated Measurement-Standard gauge block thickness

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬
BCC

Calibration Certificate

Test Object : dumbell test piece dimension

Manufacturer : NA Capacity : Specific dimensions

Serial No. : 8705-2 Accuracy : N.A

Calibration Date : 1/5/2023 Calibration Due Date : 1/5/2024

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (47 % RH)

Calibration results, data and special conditions on the use of the object are stated in the corresponding
page(s).

Calibrated By Checked By Approved by

Mousa Sharafi Eng. Asaad M.Yasin Eng. Mohammad Asker


Calibration In Charge Calibration manager S.VP Technology & QC

Calibration Certificate is electronically signed.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫ﺷﺭﻛﺔ ﺑﺣﺭﺓ ﺍﻟﻣﺗﻁﻭﺭﺓ ﻟﺻﻧﺎﻋﺔ ﺍﻟﻛﺎﺑﻼﺕ‬

Calibration Certificate

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Digital Caliper 0070840 ± 0.02 2/5/2022 2/5/2023

dumbell test piece dimension Measurements :

Acceptable Tolerance Max. Uncertainity


Dimension (mm) Measured (mm) Error (mm)
(mm) (mm)

A 75.02 0

B 4.01 0
± 0.02 ± 0.0624
C 12.49 0.0

D 12.50 0.0

Error = Indicated dimension (on the figure) - Measured dimension

The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : 3 Phase High Voltage Transformer

Manufacturer : CEFELEC Capacity : 0-3800 vac L-N

Serial No. : N1280-1/10597 Accuracy : ± 1%

Calibration Date : 18/10/2022 Calibration Due Date : 18/10/2023

Temperature : Ambient temperature (33 °C)

Humidity : Ambient humidity (46 % RH)

We certify that the above object was calibrated for specific calibration points using
certified standards traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in
the corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge Quality manager

Calibration Certificate is electronically signed

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Accuracy Calibration Calibration


No. Standard Description Serial No.
(Kvac) Date due Date

1 spark tester calibrator 2300 ± 1% 20/10/2020 20/10/2022

High Voltage Measurements :

V A-N V B-N
V L-N (Set) ± Maximum ± Maximum
Vstandard Error Error
(Kv) Vpanel (v) Uncertainty Vpanel (v) Vstandard Uncertainty
(Kv) (%) (Kv) (%)
(Kv) (Kv)

0.5 500 0.50 0.00 500 0.50 0.00

1 1000 1.00 0.00 1000 0.99 1.01

2 2000 2.01 -0.50 2000 2 0.00


0.0415 0.0409
3 3000 3.01 -0.33 3000 3.01 -0.332

3.5 3500 3.52 -0.57 3500 3.51 -0.28

3.8 3800 3.82 -0.52 3800 3.82 -0.524

V C-N
V L-N (Set) ± Maximum
Vstandard Error
(Kv) Vpanel (v) Uncertainty
(Kv) (%)
(Kv)
0.5 500 0.5 0.00
1 1000 0.99 1.01
2 2000 2.02 -0.99
0.0416
3 3000 3.02 -0.66
3.5 3500 3.52 -0.57
3.8 3800 3.83 -0.78

Error = 100 x [ (Vpanel - Vstandard)/Vstandard]


Acceptable Tolerance = ± 3%
The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported expanded Maximum uncertainty of measurement is stated as the standard
uncertainty of measurement multiplied by the coverage factor K=2, which for a normal
distribution corresponds to a coverage probability of approximately 95%.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : Digital Micrometer

Manufacturer : Mitutoyo Capacity : 0-25 mm

Serial No. : 75234449 Accuracy : ±0.002 mm

Calibration Date : 9/8/2023 Calibration Due Date : 9/12/2023

Temperature : Ambient temperature (22 °C)

Humidity : Ambient humidity (45 % RH)

We certify that the above object was calibrated for specific calibration points using certified standards
traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in the
corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Calibration In Charge Quality manager

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 1 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Calibration Calibration due


No. Standard Description Serial No. Accuracy (mm)
Date Date

1 Gauge Block Set 142500 ±0.001 6/7/2023 6/7/2024

Digital Micrometer Measurements :

Standard Gauge Block Indicated measurement ± Acceptable


Error (mm) ± Max Uncertainty (µm)
thickness value (mm) (mm) Tolerance (µm)

1.008 1.007 -0.001

1.09 1.090 0.00

5 5.001 0 0.82449 2

10 9.998 0

25 25.001 0

Error = Indicated Measurement-Standard gauge block thickness

The acceptable tolerance is the maximum allowable error.

Measurement Uncertainty:
The reported expanded uncertainty of measurement is stated as the maximum standard uncertainty of
measurement multiplied by the coverage factor K=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013 Page 2 of 2
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابﻼت‬
BCC

CALIBRATION CERTIFICATE

Test Object : Single Phase High Voltage Transformer

Manufacturer : HIPOTRONICS Capacity : 0-15 Kvac

Serial No. : 13533271 Accuracy : ± 0.1 Kvac

Calibration Date : 18/10/2022 Calibration Due Date : 18/10/2023

Temperature : Ambient temperature (32 °C)

Humidity : Ambient humidity (46 % RH)

We certify that the above object was calibrated for specific calibration points using
certified standards traceable to international standard.

Calibration results, data and special conditions on the use of the object are stated in
the corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.Yasin Eng. Mohammad Askar


Deputy Lab Director lab Director

Calibration Certificate is electronically signed.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابﻼت‬

CALIBRATION CERTIFICATE

Reference Equipment(s) Used :

Accuracy Calibration Calibration


No. Standard Description Serial No.
(Kv) Date due Date

1 spark tester calibrator 2300 ±0.01 20/10/2020 20/10/2022

High Voltage Measurements :

± Max
Set Voltage Standard Measured
Indicated Voltage (Kv) Error (%) Uncertainty
(Kv) Voltage (Kv)
(Kvac)

1 1.0 0.99 1.0

2 2.0 2.01 -0.5

3 3.0 3.01 -0.3

4 4.0 3.99 0.3

5 5.0 5.02 -0.4

6 6.0 6.04 -0.7


0.1830
7 7.0 7.02 -0.3

8 8.0 8.01 -0.1

10 10.0 10.09 -0.9

12 12.0 12.10 -0.8

14 14.0 14.11 -0.8

15 15.0 15.17 -1.1

Error = Indicated - Standard

Acceptable Tolerance = ± 3%
The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported Maximun expanded uncertainty of measurement is stated as the standard
uncertainty of measurement multiplied by the coverage factor K=2, which for a normal
distribution corresponds to a coverage probability of approximately 95%.

Form No. : WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬
BCC

CALIBRATION CERTIFICATE

Test Object : WEIGHING BALANCE

Capacity : 5100 g

Manufacturer : Mettler Toledo

Serial No. : 1129250345 Accuracy : ± 0.1 g

Calibration Date : 7/8/2023 Calibration Due Date : 7/2/2024

Temperature : Ambient temperature (23 °C)

Humidity : Ambient humidity (45 % RH)

We certify that above object was calibrated for specific calibration points using certified
standards traceable to international standard.

Calibration results, data and special conditions on the use of object are stated in the
corresponding page(s).

Calibrated By Checked & Approved by

Eng. Asaad M.yasin Eng. Mohammad Askar


Calibration In Charge Quality manager

Calibration Certificate is electronically signed.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Test Object : WEIGHING BALANCE

Manufacturer : Mettler Toledo

Serial No. : 1129250345

Measurement Method :

Measuerment method was used, the Balance was calibrated using the
Following Laboratory Refrance Standard:

Refrance Standard
NO Description
Serial No. Calibration Date Calibration Due Date

1 10 g 158427

2 200 g 158467

3 1000 g 158697 26/4/2023 26/4/2026

4 2000 g 158707

5 5 Kg 158717

Result :

Test for Errors of Indications

Acceptable
Refrance Load ( g ) Indicated Value ( g ) Error ( g ) ± Max Uncertainty ( g )
Tolerance ( g )
10.00000 10.0 0.0

200.0000 200.0 0.0

1000.000 1000.0 0.0 ±1 0.07955

2000.000 2000.1 0.1

5000.000 5000.1 0.1

Error = Indicated value - Reference Load


The acceptable tolerance is the maximum allowable error

Measurement Uncertainty:
The reported Max expanded uncertainty of measurement is stated as the standard uncertainty
of measurement multiplied by the coverage factor K=2, which for a normal distribution
corresponds to a coverage probability of approximately 95%.

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013
Bahra Cables Company ‫شركة بحرة المتطورة لصناعة الكابالت‬

CALIBRATION CERTIFICATE

Corner Load Test with Load of : (2) kg

Loading Point No. Deviation from the Corner ( g )

1 0.0

2 0.0

3 0.1

4 0.1 Fig.1

5 0.0

Deviation from the corner = measured weight at each point from Fig.1 - measured weight at the center

Repeatability Test (repeated 5 times) :

The standard deviation was calculated at ( 2 ) kg load


and found to be 0.1 g

Note :

1) The calibration was held on lab

2) The Balance Calibration was performed on a ground level with zero sloop and
no ambient noise such as air or/and sound

Form No.: WD-QMS-011.2


Issue Date : Oct.10,2013