You are on page 1of 2

Form No.

: FM/22A

ELECTRO METER CORPORATION


P- 5, C.I.T ROAD
KOLKATA - 700014 CALIBRATION CERTIFICATE OF:

COATING THICKNESS GAUGE

CALIBRATION CERTIFICATE NO.: EMC2022-KOL-00000283 Page: 1/2


1.0 ISSUED TO: ARMA ELECTRIC COMPANY
ALIPUR, VOBERCHAR ( DHAKA CHITTAGONG HIGHWAY )
GAZARIA, MUNSHIGANJ, BANGLADESH

1.1 Service Request / Order No.: -


1.2 Description identification of item to be calibrated:
i Name: COATING THICKNESS GAUGE ii Code No.: -
iii Sl. No.: 1901EC295590 iv Make: -
v Model / Type No.: DFT-222 vi Range: As Per Calibration Result
vii Sensor: -- viii Resolution: 0.1 mm
ix Location: Office Room
1.3 Applicable specification of item to be calibrated: Accuracy / permissible limit: Not Specified
1.4 Date of receipt of item: 16-Jan-2022 1.5 Date of calibration: 16-Jan-2022
Recommended date for
1.6 next calibration 16-Jan-2023 1.7 Frequency of calibration: 12 months

1.8a) Environmental condition Temperature: 25.2°C


during calibration Humidity: 55% RH
1.8b) Recommended Temperature: 20°C ± 2°C
Environmental condition Humidity: 50% RH ± 10% RH
S.O.P Number -
1.9 Basis of calibration: Standard Followed --
Description of Method Using Slip Gauge Set by Comparison Method
2.0 Traceability : Standards used for Calibration are Traceable to National/International Standards through ISO/IEC: 17025
Accredited Laboratory.

Name of the standard Recommended


Traceable to/
Sl. No. instrument used Range Resolution date for next
Certificate Nos.
(Sl. No./Code Nos.) calibration

Slip Gauge Set, Grade ‘0’, Make:


Aditya, Sl. No. 628, Code No.: Mikronix /
1. 0.5 mm to 100 mm - 29-07-2023
EMC/SG/494 SG/21/1631

"Digital Temperature & Humidity "19°C to 22°C (for


"0.1°C (for EMC (Kolkata) /
Meter, Make: Metravi, Model No. HT- Temperature) &
2 Temperature) & 20-03-2022 EMC2021-KOL-
3005, Sl. No. 141017137, Code No. 40% RH to 80% RH
0.1% RH (for Humidity)" 00017277
EMC/DTHM/840 (for Humidity)"
Form No.: FM/22A

ELECTRO METER CORPORATION


P- 5, C.I.T ROAD,
KOLKATA - 700014 CALIBRATION CERTIFICATE OF:

COATING THICKNESS GAUGE

CALIBRATION CERTIFICATE NO.: EMC2022-KOL-00000283 Page: 2/2


CALIBRATION RESULT

Permissible Expanded
Observation Reading on DUC Error
Sl No. Standard Reading in mm Error in µm Uncertainty k Factor
in mm in mm
(±) (±) in µm

1 5.0 5.000 0 6 2.00

2 10.0 10.000 0 6 2.00

3 20.0 20.000 0 6 2.00

4 25.1 25.000 0.1 -- 6 2.00

5 50.1 50.000 0.1 6 2.00

6 75.1 75.000 0.1 6 2.00

7 100.0 100.000 0.0 6 2.00

Note: i) Average of 10 readings had been taken in DUC/STD.


ii) The calibration certificate is issued by Electro Meter Corporation. This certificate refers only for the particular item submitted to calibrate.

iii) The certificate if produced for any purpose commercial or otherwise, should be reproduced in full. Partial production of the certificate or an abstract
thereof has to be got specifically approved from the technical and quality manager of this lab.
iv) The result in the calibration are valid at the time of an under the stated conditions on measurement.
v) The calibration result is only to be used for scientific & industrial purpose. This result can not be used for commercial purpose.
Measurement uncertainty at approx 95 % of confidence level & at a coverage factor : As Per Calibration Result.
Remarks: Statement of conformity not mentioned.
DUC- Device Under Calibration F.S.-Full Scale Physical status of the DUC: OK
Calibrated by: Saykat Mallik Checked & Approved by: Saptarshi Das
Date: 16-01-2022 Date: 17-01-2022
Report Issue Date: 17-01-2022

END OF CERTIFICATE