You are on page 1of 2

Form No.

: FM/22
ELECTRO METER CORPORATION
P- 5, C.I.T ROAD, CALIBRATION CERTIFICATE OF:
KOLKATA - 700014
Certificate No.: VERNIER CALIPER

CC - 2147
CALIBRATION CERTIFICATE NO.: EMC2022-KOL-00000284 ULR NO.: CC214722000000227F Page: 1/2
1.0 ISSUED TO: ARMA ELECTRIC COMPANY
ALIPUR, VOBERCHAR ( DHAKA CHITTAGONG HIGHWAY )
GAZARIA, MUNSHIGANJ, BANGLADESH

1.1 Service Request / Order No.: --


1.2 Description identification of item to be calibrated:
i Name: VERNIER CALIPER ii Code No.: --
iii Sl. No.: J1820929 iv Make: INSIZE
v Model / Type No.: -- vi Range: 0 mm to 200 mm
vii Sensor: - viii Resolution: 0.05 mm
ix Location: Office Room
1.3 Applicable specification of item to be calibrated: Accuracy / permissible limit: IS 3651 (Part 1) 1982
1.4 Date of receipt of item: 16-Jan-2022 1.5 Date of calibration: 16-Jan-2022
Recommended date for
1.6 16-Jan-2023 1.7 Frequency of calibration: 12 months
next calibration
Environmental condition Temperature: 20.6 °C
1.8(a)
during calibration Humidity: 52.9 % RH
Recommended Temperature: 20°C ± 2°C
1.8(b)
Environmental condition Humidity: 50 % RH ± 10 % RH
S.O.P Number S.O.P/Dimension/23
1.9 Basis of calibration: Standard Followed IS: 3651 (PART -1): 1982 IS: 3651 (PART -2): 1985
Description of Method Using Gauge Block & Long Slip Gauge Set by Comparison Method.
2.0 Traceability : Standards Used for Calibration are Traceable to National/International Standards through ISO/IEC: 17025
Accredited Laboratory.

Name of the standard Recommended


Traceable to /
Sl. No. instrument Range Resolution date for
Certificate Nos.
used (Sl. No./Code Nos.) next calibration

Slip Gauge Set, Grade ‘0’,


1 Make: Aditya, Sl. No. 628, 0.5 mm to 100 mm - 29-07-2023 Mikronix / SG/21/1631
Code No.: EMC/SG/494

Digimatic Outside
Micrometer, Make: Mahr, Sl. EMC (Kolkata) / EMC2021-
2 0 mm to 25 mm 1 µm 16-01-2022
No.: 10698, Code No: KOL-00000640
EMC/D/45

Slip Gauge Accessories Set,


Make: FWP, Range: 0 mm to 0 mm to
3 - 18-10-2023 Alcalab / 01/1282/21-22
250 mm (Set of 22 Pieces), 250 mm
Code No: EMC/SGA/130
"Digital Temperature &
"19°C to 22°C (for
Humidity Meter, Make:
Temperature) & "0.1°C (for Temperature) & EMC (Kolkata) / EMC2021-
4 Metravi, Model No. HT-3005, 20-03-2022
40% RH to 80% RH 0.1% RH (for Humidity)" KOL-00017277
Sl. No. 141017137, Code No.
(for Humidity)"
EMC/DTHM/840
Form No.: FM/22
ELECTRO METER CORPORATION
P- 5, C.I.T ROAD, CALIBRATION CERTIFICATE OF:
KOLKATA - 700014
Certificate No.:
VERNIER CALIPER
CC - 2147
CALIBRATION CERTIFICATE NO.: ULR NO.: CC214722000000227F Page: 2/2
CALIBRATION RESULT
External Measurement
Expanded
DUC Reading STD Reading Error Permissible Error
Sl.No. Uncertainty k factor
(mm) (mm) (µm) (±) (µm)
(±) in µm
1 0.00 0.000 0 50 28.87 2.00
2 10.05 10.050 0 50 28.87 2.00
3 20.10 20.100 0 50 28.87 2.00
4 51.05 51.050 0 60 28.87 2.00
5 75.05 75.000 50 60 28.87 2.00
6 100.05 100.000 50 60 28.87 2.00
7 150.05 150.000 50 70 28.88 2.00
8 200.05 200.000 50 70 28.91 2.00
Internal Measurement
DUC Reading STD Reading Error Permissible Error
Sl No.
(mm) (mm) (µm) (±) (µm)
1 10.05 10.000 50 50
Depth Measurement
DUC Reading STD Reading Error Permissible Error
Sl No.
(mm) (mm) (µm) (±) (µm)
1 50.05 50.000 50 60
Parallelism of Measuring Face
Observed Error Permissible Error
(µm) (±) (µm)
External Measuring
10 40
Faces
Internal Measuring Faces 6 10
Note: i) Average of 10 readings had been taken in DUC/STD.
ii) The calibration certificate is issued by Electro Meter Corporation. This certificate refers only for the particular item submitted to
calibrate.
iii) The certificate if produced for any purpose commercial or otherwise, should be reproduced in full. Partial production of the
certificate or an abstract thereof has to be got specifically approved from the technical and the quality manager of this lab.
iv) The results in the calibration certificate are valid at the time of and under the stated conditions on measurement.
v) The calibration result is only to be used for scientific & industrial purpose. This result can not be used for commercial purpose.
Measurement uncertainty at approx 95 % of confidence level & at a Coverage Factor : As Per Calibration Result
Remarks: Statement of conformity not mentioned.
DUC- Device Under Calibration F.S.-Full Scale Physical status of the DUC: OK
Calibrated by: Saykat Mallik Checked & Approved by: Saptarshi Das
Date: 16-01-2022 Date: 17-01-2022
Report Issue Date: 17-01-2022

END OF CERTIFICATE