CALIBRATION CERT! FICATE

Machine : Brinell Hardness Tester Certificate No TMSC/23lM-1918s

Machine Model No KB-3000J ULR No cc24982300002s723F


Serial No 2006/1,137 Certificate issue Date 22-09-2023
Make: Krystal Tech Date of Calibration 22-09-2023
Capacity: 3000 Kgf Suggested Calibration due On 2L-09-2024
l.D No MTB/BHT/01 Job Code No: TMSC/SEPl22l02
Room Temp: 27"C Customer's Details
General Condition o.K M/s. Mangalam Testing Bureau.
Basis of Calibration: ASTM-E 10:2018 Add: P-83, Benaras Road, Howrah - 711108.
Method of verification lndirect
Calibration Location AT Site
Dlscipline & Group Mechanical-Hardness Testing Machine
STANDARDS USED FOR CALIBRATION ARE TRACEABLE TO NATIONAL OR INTERNATIONAL STANDARDS THROUGH
NABL ACCREDITED LABORATORY. MASTER (S) USED FOR CALIBRATION:-
Scale: 5 /750 HBW

DETAILS OF HARDNESS Observed Reading Mean Error Repeatabilty


STANDARDISED BLOCK 1 2 3 4 5

Scale: 5 /750 HBW Diagonal (D)mm

Std. Value: 199.8 2.L20 2.125 2.130 2.740 2.120 2.727


0.8 0.02
Certificate No: FRI/03/2UH01811 Hardness Value

Next Cali: due: 18.03.2024 202 20 1 200 198 202 200.6

Uncertainty Of Measurements: !2% (At 95% Confidence level, k=2)

DETAILS OF HARDNESS Observed Reading Mean Error Repeatabilty


STANDARDISED BLOCK 7 2 3 4 5

Scale: 5 /750 HBW Diagonal(D)mm

Std. Value: 6tO.4 7.240 L.24s 1.240 L.240 7.240 1.24L -0.4 0.01
Certificate No: FRI/03/2UH01839 Hardness Value

Next Cali: due: 26.03.2024 6tt 606 6tt 611 611. 610.0

Uncertainty Of Measurements: !2% (At 95% Confidence level, k=2) Page No: 1 of 2
Note: (1)This Certificate refers only to the particular Machine for site cal ibration.
(2) This certificate shall not be reproduced, except in full, unless prrior written permission from
CEO/JMSC. Th is certificate is invalid without sign atu re.

4'1 o a7,.1v@-
Calibrated By Rpproved {V
(Calibration Engineer) (Technical Manager)
TAPAS DEY
d]
6l
iJAL,iffRATEI] (u ffiffi BHASKAR CHATTERJEE

* *
F0RM No. : TMSC/[,4SD/FM-08

ACCREDITED BY NABL

cc-2498

CALI BRAT!ON CERTI FICATE

Machine : Brinell Hardness Tester Certificate No: TMSC/23lM-191,8s

Machine lVlodel No KB-3000J ULR No: cc249823000025723F


Serial No 200611,137 Certificate issue Date: 22-09-2023
Make Krystal Tech Date of Calibration 22-09-2023
Capacity 3000 Kgf Suggested Calibration due On 2t-09-2024
l.D No: MTB/BHT/o1 Job Code No IMSC/SEPl22l02
Room Temp 27"C Customer's Details
General Condition o.K M/s. Mangalam Testing Bureau.
Basis of Calibration: ASTM-E 10:2018 Add: P-83, Benaras Road, Howrah - 711108.
Method of verification: I ndi rect
Calibration Location AT Site
Discipline & Group Mechanical-Hardness Testing Machine
STANDARDS USED FOR CALIBRATION ARE TRACEABLE TO NATIONAL OR INTERNATIONAL STANDARDS THROUGH
NABL ACCREDITED LABORATORY. MASTER (S) USED FOR CALTBRATTON:-
e: HBW
DETAILS OF HARDNESS Observed Reading Mean Error Repeatabilty
STANDARDISED BLOCK L 2 3 4 5

Scale: 10/3000 HBW Diagonal(D)mm


Std. Value: 207.6 4.250 4.240 4.220 4.260 4.240 4.242 o.2 0.04
Certificate No: FRI/03/2 t/ H}t8tz Hardness Value

Next Cali: due: 18.03.2024 201. 202 204 200 202 20L.8
Uncertaihty Of Measurements !2% (At 95% Confidence level, k=2)

DETAILS OF HARDNESS Observed Reading Mean Error Repeatabilty


STANDARDISED BLOCK 7 2 3 4 5

Scale: 10/3000 HBW Diagonal(D)mm

Std. Value: 615.7 2.470 2.460 2.470 2.480 2.480 2.472 -0.7 o.02
Certificate No: FRI/03/2UH01838 Hardness Value

Next Cali: due: 25.03.2024 616 621 616 677 6tt 615.0
Uncertainty Of Measurements: !2% (At 95% Confidence level, k=2) Page No: 2 of 2
Note: (1)This Certificate refers only to the particular Machine for site calibration.
(2) This certificate shall not be reproduced, except in full, unless priorwritten permission
from
CEO/TMSC. This certificate is invalid without signature.
'*End Of certificate****

t*
Calibrated By
(Calibration Engineer)
TAPAS DEY
,,r)
o)
CALIBRATED
ffi
trf,ffig#
,th/\.V,a
Approved By
(Technical Manager)
BHASKAR CHATTERJEE
lt *
F0RN4 Nc. : T[,1SCi[/SD/FNl-08

ACCREDITED BY NABL

cc-2498

CALI BRATION CERTI FICATE

Name: Rockwell Hardness Tester Certificate No TMSC/23lM-19184

Machine Model No KAS ULR No cc249823000025722F


Serial No 200611203 Certificate issue Date: 22-09-2023
Make Krystal Tech Date of Calibration 22-09-2023
Ca pacity: 150 Kgf Suggested Calibration due On 27-09-2024
LD No: MTB/RHT/01 Job Code No TMSCISEP/22102
Room Temp 27"C Customer's Details
General Condition o.K
M/s. Mangalam Testing Bureau.
Basis of Calibration: ASTM-E L8t2O22
Method of verification:lndirect Add: P-83, Benaras Road, Howrah - 711108.
Location At Site
Discipline & Group Mechanical-Hardness Testing Machine
Scale: HRA
SIANDARDS USED FOR CALIBRATION ARE TRACEABLE TO NATIONAL OR INTERNATIONAL STANDARDS THROUGH NABL ACCREDITED LABORATORY, MASTER (5)
USED FOR CALIBRATION:-

DETAILS OF HARDNESS Observed Reading Mean Deviation Repeatability


STANDARDISED BLOCK 7 2 3 4 5 Allowed Actual Allowed Actual

Scale: HRA
Std. Value: 35.7 36.34
35.8 36.2 36.4 36.4 36.9 12.0 -0.64 1.3 1.10
Certificate No FRII 03 I 2t/ H01842
Next Cali: due: 18.03.2024
Uncertainty Of Measurements: t 1.5 (At 95% Confidence level, k=2)
DETAILS OF HARDNESS Observed Reading Mean Deviation Repeatability
STANDARDISED BLOCK L 2 3 4 5 Allowed Actual Allowed Actual

Scale: HRA
Std. Value: 62.5 62.74
62.6 62.6 62.8 62.8 62.9 11.5 -0.24 0.75 0.30
Certificate No: FRI/03/21lH01814
Next Cali: due: 18.03.2024
Uncertainty Of Measurements: + 1.5 (At 95%i Confidence level, k=2)
DETAILS OF HARDNESS Observed Reading Mean Deviation Repeatability
STANDARDISED BLOCK .L 2 3 4 5 Allowed Actual Allowed Actual

Scale: HRA
Std. Value: 82.6 82.68
82.7 82.s 82.8 82.6 82.8 r1.5 -0.08 0.35 0.30
Certificate No: FRI/03/2UH01813
Next Cali: due: 18.03.2024
Uncertainty Of Measurements: t 1.5 (At 95% Confidence level, k=2)
Page: 1 Of 3

s, B&ffi.rlEl
trffi
ilqbt ?TI'LE!,
Calibrated By By
(Calibration Engineer) (Technical Manager)
TAPAS DEY Hf,ffi*r# BHASKAR CHATTERJEE
FORM No : TMSC/IVSD/Ftr4-08

ACCREDITED BY NABL

cc-2498

CAL!BRATION CERTIFICATE

Name: Rockwell Hardness Tester Certificate No TMSC/23/M-79784


Machine Model No: KAS ULR No: cc249823000025722F
Serial No 2006/1203 Certificate issue Date: 22-09-2023
Make: Krystal Tech Date of Calibration 22-09-2023
Capacity 150 Kgf Suggested Calibration due On
21-09-2024
l.D No: MTB/RHT/01 Job Code No TMSC/SEPl22/02
Room Temp 27"C Customer's Details
General Condition OK
Basis of Calibration: ASTM-E Lg;2O22 M/s. Mangalam Testing Bureau.
Method of verification:lnd irect
Add: P-83, Benaras Road, Howrah _ 711l0g.
Location At Site
Disci p ine & G rou Mec h a n ical- Hardness Testing Mach ine
Scale: HRBW
ANDA RDS USED FOR CALIBRA TI N ARE TRACEABLE TO NATIONAL OR INTERNATIONAL STANDAR DS
THROUGH NABL ACCREDITED IABORA .TORY
D FOR MASTER (s)

DETAILS OF HARDNESS Observed Mean Deviation Repeatability


STANDARDISED BLOCK
t 2 3 4 5 Allowed Actual Allowed Actual
Scale: HRBW
Value 45.2
45.7 46.66
rtif icate No: FRt/ 03 / 21./H01816 45.9 46.7 46.7 48.3 14.0 1.46 3.3 2.60
Next Cali: due: 18.03.2024
U ncerta of M eas rements 1 .5 (At 95% dence
DETAILS OF HARDNESS Observed
n Deviation Repeatability
NDARDISED BLOCK
t 2 3 4 5 Allowed Actual
Scale: HRBW
Allowed Actual

Std. Value: 67.7


67.9 67.8 68.34
Certif icate No: FRt/ 03 / 2L/H01815 68.3 68.8 68.9 13.0 0.64 2.5 1.10
Next Cali: due: 18.03.2024
U ncertai Of Measure r 1.5 9s%
DETAILS OF HARDNESS
Observed Reading Deviation
Mean Repeatability
STANDARDISED BLOCK
1, 2 3 4 5 Allowed Actual Allowed Actual
le: HRBW
. Value: 99.0
99.4 99.7 99.72
certificate No: 99.8 99.8 99.9 12.0 0.72
F Rt / 03 / 27/H01840 1".21 0.s0
Next Cali: due: 25.03.2024
Unce Of Measurements: + 1.5 Al95% Confidence k=2
Page: 2 Of 3

$,
Calibrated By ,r/]t212/y1 *

ffi Approved Ay
(Calibration Engineer) CATIBRATED
d)
ol (Technical Manager)
TAPAS DEY
* i BHASKAR CHATTERJEE
F0Rl\4 No : T[4SCi]\4SD/FL4-08

ACCREDITED BY NABL

cc-2498

CALI BRATION CERTT FICATE

Name: Rockwell Hardness Tester Certificate No TMSC/23lM-19184


Machine Model No: KAS ULR No: cc249823000025722F
Serial No: 200617203 Certificate issue Date 22-09-2023
Make: Krystal Tech Date of Calibration: 22-09-2023
Ca pacity: 150 Kgf Suggested Calibration due On 2t-09-2024
LD No: MTB/RHT/o1 Job Code No TMSC/SEPl22/02
Room Temp 27'C Customer's Details
General Condition OK
Basis of Calibration: ASTM-E L8:2022
M/s. Mangalam Testing Bureau.
Method of verification:lndirect Add: P-83, Benaras Road, Howrah -7LLLO8.
Location At Site
Discipline & Group Mechanical-Hardness Testing Machine
Scale: HRC
STANDARDS USED FOR CALIBRATION ARE TRACEABLE TO NATIONAL OR INTERNATIONAL THROUGH NABL ACCREDITED LABORATORY. MASTER (S)
USED FOR CALIBRATION:-

DETAILS OF HARDNESS Observed Reading Mean Deviation Repeatability


STANDARDISED BLOCK 7 2 3 4 5 Allowed Actual Allowed Actual
Scale: HRC
Std. Value: 27.3 28.36
28.2 28.3 28.4 28.4 28.5 11.5 1.06 7.4 0.30
Certifi cate No: FRI/03/21lH01819
Next Cali: due: 18.03.2024
Uncertainty Of Measurements: + 1.08 Confidence level, k=2)
DETAILS OF HARDNESS i Mean Deviation Repeatabllty
STANDARDISED BLOCK 1 2 3 4 5 Allowed Actual Allowed Actual
Scale: HRC
Std. Value: 48.3 48.52
48.3 48.4 48.5 48.6 48.8 r1.5 0.22 1.03 0.50
Certificate No: FRI / 03 / 27 / H01818
Next Cali: due: 18.03.2024
Uncertainty Of Measurements: + 0.76 (At 95% Confidence level, k=2)
DETAILS OF HARDNESS Observted :ading Mean Dev on
Rr Repeatabilty
STANDARDISED BLOCK
Scale: HRC
1
-1 3 4 5 Allowed Actual Allowed Actual

Std. Value: 62.2 62.62


62.4 62.5 62.6 62.8 62.8 r1.5 0.42 0.75 0.40
Certificate No: FRI/03/21/ H01817
Next Cali: due: 18.03.2024
Uncertainty Of Measurements: + 0.74 (At95% Confidence level, k=2)
Page: 3 Of 3
Note:(1)This certificate refers only to the particular Machine for site calibration.
(2) This certificate shall not be reproduced, excep in full, unless prior written permission
from
CEO/TMSC. This certificate is invalid without signature.
**** END OF CERTIFICATE***

b, /-7/.'\,/V:Z--
Calibrated By Approved Ay
(Calibration Engineer) (Technical Manager)
TAPAS DEY
BHASKAR CHATTERJEE
FORM No. : TMSo/tulSD/FM"0B

ACCREDITED BY NABL

-:.ibration
cc-2498

CALI BRATION CERTI FICATE

IVlachine : Universal Testing


Certificate No: TMSC/23IM-19r83
Machine (Digital)
Machine Model No: UTES 1OO ULR No. cc249823000025721F
Serial No: 2/2070-4278 Certiflcate issue Date: 22-09-2023
Make: FIE Date of Calibration: 22-09-2023
Ca pacity 1000 KN Suggested Calibration Due On: 2t-09-2024
l.D No: 2l2OLO-4278 Job Code No. TMSC/SEP/22i02
Room Temp 27 "C Customer's Details
General Condition o.K
Basis of Calibration: ASTM E4- 2021
M/s. Mangalam Testing Bureau.
Mode Of Calibration: Compression Add: P-83, Benaras Road, Howrah -71,1,1.08.
Calibration Location: At Site
Discipline & Group M echa nica I-UTM, T ensio n Cre e p a n d To r5 on Testi ng Mach ne
I I

Scale: 1000 kN Least Count: 0.05 kN


STANDARDS USED FOR CALIBRATION ARE TRACEABLE TO NATIONAL OR INTERNATIONAL STANDARDS THROUGH NABL ACCREDITED
LABORATORY. MASTER (S) USED FOR CALIBRATION:-

Certificate No Valid Upto Calibrated By


Details Of Force lnstruments
22021.488/ D 1.0s I C- 492 23-03-2024 NPL
Used - 50/L000 kN
22021488/ D1,.0s/C-490 23-03-2024 NPL
Obs: Readings o
>
P cv E f
o
Standard Standard ooc
>EE
Force ln kN Reading At Reading At
Run 1 Run 2 Run 3 Mean _o^
6\o ElqJ g!
Error (%) !; F
0C 0C (00) (120o) (240o) (Div) 6- oEi fc
24 27 Eo^ o
o EY
Div Div Div e.
U
13 52504 52504 52706 52766 52760 52744 0.46 0.11 0.38 13.06
40 161_407 L61407 162211 162048 152052 162104 0.43 0.10 0.13 40.L7
120 23976 23976 24066 24060 24055 24060 0.3s 0.05 0.04 120.42
400 80080 80080 803ss 80362 80375 80364 0.35 0.02 0.0r. 401,.42
530 106181 106L81 106498 106510 L06522 1065L0 0.31 0.02 0.01 531.64
700 140355 140355 1,40776 740765 140782 1,40774 0.30 0.01 0.0L 702.09
930 186605 186605 1871"65 t87t7t 187185 187174 0.30 0.01 0.01 932.83
1000 200677 200677 201255 201,260 201335 201283 0.30 0.04 0.01 1003.02
Cbserved Readings on Removal of Force (Fio) = 0.002 0.000 0.000 Estimated Mean Relative Error E = (0.481s0) %
Relative Zero Error (fo) % 0.0002 0.0 0.0
Uncertainty of Measurement at 95% Confidence Level, k = 2 0.50 %
Statement of Conformity: Class-A
Note (1)This Certificate refers only to the particular Machine for site calibration
(2) This certificate shall not be reproduced, except in full, unless prior written permission from
CEO/TMSC. This certificate is invalid without signature.
**** End Of Certificate****