ETECTRO METER CORPORATION


Form No.: FM/22
...,....
P- 5, C.I.T ROAD,
CATIBRATION CERTIT'ICATE O}':
KOLKATA -7OO014,
Certificate No.:
cc - 274V DIGITAL MICRO OIIM MITI]II
CALIBMTION CERTIFICATE NO.: EMC2 0 i.9-KOL-000085 1 0 ULRN0.: CC2147790O0O7ZB7F
lpaee:t/2
1
ISSUED TO: sQ wrRE & cAtlLE CO" LTD.
124, KEODHAI,A, PIOT#1 97 9.I9BO,
MADANPUR, NARAYANGANJ

t.1 Service Request / Order No.:


t.2 Description identification of item to be calibrated:
i lName: lorctrel MICR0 oHM METER ll Code No.: DMM-1
lu Sl. No.:
lv Make: Agronic
v Model / Type No.: 53C vl Range: As Per Calibration Result
vll Sensor; vlll Resolution: As Per Calibration Result
lx Location: Testing Lab
1.3 Applicable specification of item to be calibrated: Accuracy / permissible limit: Not Specified
L.4 Date of receipt of item: 23-Apr-2019 1.5 Date of calibration: 23-Apr-2079
Recommended date for
l.t) 23-Apr-2020 1.7 Frequency of calibration:
next calibration: 72 months
Environmental condition Iemperature: 24,8'C
1.8
during calibration Humidity: 52.60/aRH
L.9 Basis of calibration: s.o.P/Er/27
2.0 )aDrrrly : )tanqaros IJseo tor calibration are Traceable to National/lnternational
standarc s through
so/ EC:17O25 Accredite rd Laboratory.

Name of the starrdard

d
Recommendetl
instrument Range Resolution 1'raccable to /
:
date for
used (Sl. No./Codr: Nos.) Certificate nos.
next calibration

Micro ohm Calibrator (Ductor


CalJ, Make: Time Electronics, Absolute
1 50pO to 2O
Sl. No. 1085C17, Model: 5070, 02-05-79 Calibration, /
Code No. : EMC/DUCTOR/7 63, 042L604, /

Iligh Precision Decade


Resistance Bcx, Make: Vaiseshika
!. Vaiseshika, Model: 7400, Serial 0.01 O to 100 kft Electron, /
2L-08-1.9
No: 6473, Code No. vED/tlABt,/17-
EMC/DRB/47s, tB(Er)/ 46,
EI ELECTRO METER CORPORATION
Form No.: FM/22

P- 5, C.I.T ROAD,

ffiffi
CALIBRATION CERTIFICATE OF:
KOLKATA.TOOO1.4. DIGITAL MICIIO OIIM MUTIR
Certificate No.:
cc - 2147
LALIBRA'|ION C RTIFICATE NO.: EMC 2019-KOL-00008510@ 900007787F Page: 2 Of-2
CALIBRATION RESULT
Parameter/ Nominal Value
Renop on IIIIC Std Value Errd? urM (r) k factor
I llesistance pO rnO pO pO
I rooopn 495 0.5 -4 2.9508 2.00
!
i
l.c ,rpo 999 1.0 -I1 5.8059 2.00
t49L r.:l -9 8.6810 2.00
rn0 mf,I mO mO
| ,r.rno,* ,1.95 5.0 -0.05 0.0129 2.00
i t .c ,o.or *o s.92 10.0 -0.08 0.0238 2.00
14.90 15.0 -0.10 0.0352 2.00
199.9mO 49.3 50.0 -0.7 0.0818 2.00
L.C 0.1mO 99.4 100.0 -0.6 0.1500 2.00
L99.2 200.0 -0.8 0.3000 2.00
1999mO 496 0.50 4 0.8180 2.00
L.C:1mO 993 1.00 7 1.2949 2.00
L49l 1.50 9 1.8319 2.00
o o () o
79.990" 4.95 5.00 -0.05 0.0500 2.00
L.C;0.01O 9.94 10.00 -0.06 0.1000 2.00
18.90 19.00 -0.10 0,1900 2.00
199.9A 49.9 50.00 0.1 0.5000 2.00
L.C :0.1O 99.7 100.00 0.3 1.0000 2.00
189.0 190.00 1.0 1.9000 2.00
ko
1999A 498 0.50 -L 0.6467 2.00
L.C :1O 997 1.00 -J 0.8204 2.00
1895 1.90 -5 1.2489 2.00
ko ko ko
19.99kO 5.03 5.00 0.03 0.0500 2.00
L.C :0.01kf} 9.98 10.00 -0.02 0.1000 2.00
18.90 19.00 -0.10 0,1900 2.00
Note: U Average of5 readings had been taken in STD/DUC.
iiJTheCalibrationcertificateisissuedbyElectroM"t".Co.
to calibrate.
rlu t ne Lertlrlcate lt produced tor any purpose commercial or otherwise, should
be reproduced ln fuLi.rr.tirtl-au.tf , oi tn"
omthetechnicalandthequalitymanagerofthislab.
lYltnereSu|tslntnecallbratloncertiiicateareVnditionsonmeasurement.
R.*".kr, P..fo.*rr." .h".k hrr b UC was not specified U"inffi
DUC: Device Under Calibration UIM: Uncertainty in Measurement F.S : Full Scale Physical status ofthe DUC; OI(
Checked & Approved by
Datet 23-04-2019 , / Date:27-04-2019
Report lssue Dare: 2z-04-ZOLg n / /
4&"
1...1r in-qhrrrc
I '
1.' :.''.-, '.. l'-'* i,-:'poretiorl Et F Clqoi\t[ rER CORp0RATIC)H