Calibration Certificate for Instruments

This document is a calibration certificate for a laboratory stirrer. It provides details of the calibration including the customer information, calibration date and validity, equipment used, observations made during calibration testing at different speeds, and approval. The calibration showed deviations within specified accuracy thresholds.

Uploaded by

Sales GoldClass
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (2 votes)
7K views1 page

Calibration Certificate for Instruments

This document is a calibration certificate for a laboratory stirrer. It provides details of the calibration including the customer information, calibration date and validity, equipment used, observations made during calibration testing at different speeds, and approval. The calibration showed deviations within specified accuracy thresholds.

Uploaded by

Sales GoldClass
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
  • Equipment Used: Lists the equipment and instruments used in the calibration process, including their identification numbers and traceability.
  • Observations: Records the observations noted during the calibration procedure, including any deviations or specific remarks.
  • Certificate Details: Provides details about the unit description, model number, and serial number of the equipment being calibrated.
  • Calibration Dates: Specifies the date of calibration and the next due date for recalibration.

REMI ELEKTROTECHNIK LTD. (INSTRUMENTS DIV.

)
VALIV, VASAI (EAST), PALGHAR-401208

Calibration Certificate

UNIT DESCRIPTION : LABORATORY STIRRER MODEL NO.: RQT – 134 H/D

SR. NO.: ZHFL-22044 CERTIFICATE NO: RMTC – 1183/21

NAME OF CUSTOMER:

ADDRESS OF CUSTOMER:

CALIBRATION DATE: 28.06.2021 CAL. VALIDITY: 27.06.2022

DETAILS OF EQUIPMENT USED FOR CALIBRATION


NOTE: Tick  mark where as applicable.

Digital Lab Thermometer LCD Tachometer Digital Stop Watch

Auto Ranging R.P.M. Meter  Digital Strobe Weight Set

Digital Temperature Indicator

Sr. No. Description Certified By Instrument Id. Certificate No. Cal. Date Validity
Auto Ranging
01 T & C Lab. IRM 01 TCL/0854/2020 09.11.2020 08.11.2021
RPM Meter

OBSERVATIONS:

Instrument Speed Speed Reading on Strobe Diff. In Actual and Strobe


Sr. No. Remarks
Displayed (RPM) Meter (RPM) Meter RPM
01 5000 5030 30 OK
02 2500 2520 20 OK
03 500 505 05 OK

NOTE: 1. Calibration Traceable to national standards.


2. Overall deviation is within specified accuracy.
3. Result is related only to the item calibrated.
4. Maximum tolerance permissible is ± 1°C & ± 5°C whichever is applicable and RPM ±2% FSD
and ± 5gm to set weight.

REMARKS: ACCEPTED

CALIBRATED BY: MANOHAR SAWANT APPROVED BY: MANAGER Q. A.