CALIBRATION CERTIFICATE

Centre for Ca[i6ration


NAG MAN CALIBRATION SERVICES LLP
No. 168fl, Chennai· Bangalore National Highway,
CFC/FT·01/7.8/R3 Chembarambakkam, Chennai· 600 123. Tamilnadu. INDIA CC-2548
E·Mail: cfcchennai@nagman.com.calibration@nagman.com
Site: www.nagman.com

ULR-CC254821000005980F Certificate Number. 2020-21 / CFC / 2581/252


Date of Issue: 16.03.2021
Mis. Bharat Heavy Electricals Limited,
Customer Name and Address Power Sector Southern Region, EVR Periyar Building,
690, Anna Salai, Nandanam, Chennai - 600 035.
Customer Reference Instrument Received on: 02.03.2021
Details of the Instrument
Description RTD Sensor
Make Tempsens
Type Pt.100
Serial number 2104311486
Identification Number -
Range o to 350°C
Operating Range -
Accuracy ±0.7°C
Equipment received on 02.03.2021
Condition of the equipment on receipt Good
Calibration procedure number CFC / WI-T02A
Method of Calibration Comparison Method
Date of calibration 10.03.2021
Date of next calibration suggested 10.03.2022
Calibration environments
Temperature 25°C ± 4°C
Relative humidity 30% to 75%RH
Master Instrument Details Master 1 Master 2
Description RTD Digital Multimeter (6 112Digit)
Make Ametek Agilent
Type/Model STSI00A500 34401A
Serial number 542694-03 MY47018587
Report Number 2020-211CFCIMASTERl052 2020-211CFCIMASTERl085
Calibration valid till 24/07/2021 11109/2021
The Standards used are traceable to National/International Standards

Checked by Authorised Signatory

,1),\A~- ,
D. Yuvarani
r'~
P. Gunavathi
Team Member Team Leader
Page 01 of 02
CALIBRATION CERTIFICATE
Centre for Ca[i6ration
NAGMAN CALIBRATION SERVICES LLP
No. 168f1, Chennai· Bangalore National Highway,
CFC/FT·01/7.8/R3 Chembarambakkam, cnennei- 600 123. Tamilnadu. INDIA
CC·2548
E·Mail: cfcchennai@nagman.com.calibration@nagman.com
Site: www.nagman.com

ULR-CC254821000005980F Certificate Number. 2020-21/ CFC /2581/252

CALIBRATION DATA
Thermal - Temperature

Standard Test Reading Deviation Deviation Expanded Coverage


Eq. Reading (UUe) Allowed Uncertainty Factor
(Mean) (±) (±) k
°C Q Eq.oC °C °C °C
-0.23 99.879 -0.31 -0.08 0.70 0.13 2.00
50.21 119.411 50.04 -0.17 0.70 0.13 2.00
100.18 138.476 99.93 -0.25 0.70 0.13 2.00
150.26 157.291 149.92 -0.34 0.70 0.13 2.00
200.35 175.835 199.96 -0.39 0.70 0.13 2.00
250.31 194.044 249.87 -0.44 0.70 0.13 2.00
300.45 212.014 299.92 -0.53 0.70 0.13 2.00
350.49 229.674 349.91 -0.58 0.70 0.14 2.00

Remarks:-

I. The above UUC was calibrated at Lab.

2. UUC is defined as Unit Under Calibration.

3. The measurement confidence level is 95%.

4. The Calibration Certificate shall not bereproduced except in full without written approval of CFC.

5. The above Expanded Uncertainty includes Repeatability, Bath stability, Resolution & Uncertainty on Measurements.

6. Temperature Scale: International Temperature Scale 1990 (ITS 90).

7, The decision rule will be applied based on the statement of conformity requested by the customer.

8. The results are provided as observed without any adjustments.

9. The results are found compliance to the manufacturer specifications.

10. The above results relates only to the item calibrated.

~Checked
,'V by
"-t Auth~ri~e~at~? \
D. ~vaiani fGUnavathi
Team Member. Team Leader
Page 02 of 02
--------------- End of the Report ---------------