Customer name M/S: KAP FASTNERS Certificate No.

CC223921000020298F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Digital Vernier Caliper Sr. No. ------
Make Mitutoyo Location Workshop
Range / Size 0 to 150 mm Accuracy OK
Least Count 0.01 mm Visual Inspection OK
I.D. No. KAP/DVC/01

Detail of reference standards & Major equipments used


Equipment Name Caliper checker (Steel) Caliper checker (Steel)
Make KCP KCP
Model / SR No. 1083 1083
Certificate No. TI/A/CC/005/21 TI/A/CC/005/21
Calibration Validity 29/012023 29/012023
Calibration by Tanson Instrument Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial Nominal Value In Error In (micron) Uncertainty At 95% C.L.
No. (mm) OD ID Depth (coverage factor k=2)

01. 0 Set 0 0 0
02. 1 0 --- ---
03. 2 0 --- ---
04. 5 0 --- ---
05. 10 0 0 0
06. 20 0 0 0 ±8.0 µm
07. 50 0 0 0
08. 100 0 0 0
09. 150 1 1 0
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by
coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020299F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Vernier Caliper Sr. No. ------
Make Mitutoyo Location Workshop
Range / Size 0 to 150 mm Accuracy OK
Least Count 0.02 mm Visual Inspection OK
I.D. No. KAP/VC/01

Detail of reference standards & Major equipments used


Equipment Name Caliper checker (Steel) Caliper checker (Steel)
Make KCP KCP
Model / SR No. 1083 1083
Certificate No. TI/A/CC/005/21 TI/A/CC/005/21
Calibration Validity 29/012023 29/012023
Calibration by Tanson Instrument Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial Nominal Value In Error In (micron) Uncertainty At 95% C.L.
No. (mm) OD ID Depth (coverage factor k=2)

01. 0 Set 0 0 0
02. 1 0 --- ---
03. 2 0 --- ---
04. 5 0 --- ---
05. 10 0 0 0
06. 20 0 0 0 ±8.0 µm
07. 50 0 0 0
08. 100 0 0 0
09. 150 1 1 0
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by
coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020300F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Vernier Caliper Sr. No. ------
Make Mitutoyo Location Workshop
Range / Size 0 to 150 mm Accuracy OK
Least Count 0.02 mm Visual Inspection OK
I.D. No. KAP/DVC/02

Detail of reference standards & Major equipments used


Equipment Name Caliper checker (Steel) Caliper checker (Steel)
Make KCP KCP
Model / SR No. 1083 1083
Certificate No. TI/A/CC/005/21 TI/A/CC/005/21
Calibration Validity 29/012023 29/012023
Calibration by Tanson Instrument Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial Nominal Value In Error In (micron) Uncertainty At 95% C.L.
No. (mm) OD ID Depth (coverage factor k=2)

01. 0 Set 0 0 0
02. 1 0 --- ---
03. 2 0 --- ---
04. 5 0 --- ---
05. 10 0 0 0
06. 20 0 0 0 ±8.0 µm
07. 50 0 0 0
08. 100 0 0 0
09. 150 1 1 0
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by
coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020301F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Vernier Caliper Sr. No. ------
Make Mitutoyo Location Tool Room
Range / Size 0 to 150 mm Accuracy OK
Least Count 0.02 mm Visual Inspection OK
I.D. No. KAP/VC/03

Detail of reference standards & Major equipments used


Equipment Name Caliper checker (Steel) Caliper checker (Steel)
Make KCP KCP
Model / SR No. 1083 1083
Certificate No. TI/A/CC/005/21 TI/A/CC/005/21
Calibration Validity 29/012023 29/012023
Calibration by Tanson Instrument Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial Nominal Value In Error In (micron) Uncertainty At 95% C.L.
No. (mm) OD ID Depth (coverage factor k=2)

01. 0 Set 0 0 0
02. 1 0 --- ---
03. 2 0 --- ---
04. 5 0 --- ---
05. 10 0 0 0
06. 20 0 0 0 ±8.0 µm
07. 50 0 0 0
08. 100 0 0 0
09. 150 1 1 0
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by
coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020302F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Micrometer Sr. No. ------
Make Mitutoyo Location Workshop
Range / Size 0 to 25 mm Accuracy OK
Least Count 0.01 mm Visual Inspection OK
I.D. No. KAP/-MM/01

Detail of reference standards & Major equipments used


Equipment Name Slip Gauge Block Set
Make KCP
Model / SR No. 10373
Certificate No. TI/K/SGS/35/21
Calibration Validity 20/11/2023
Calibration by Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial No. Nominal Value In Error In Uncertainty At 95% C.L.
(mm) (micron) (coverage factor k=2)

01. 0 Set 0
02. 2.5 0
03. 5.1 0
04. 7.7 0
05. 10.3 0
06. 12.9 0 ±1.0 µm
07. 15.0 0
08. 17.6 0
09. 20.2 0
10. 22.8 1
11. 25.0 1
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020303F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Micrometer Sr. No. ------
Make Baker Location Workshop
Range / Size 0 to 25 mm Accuracy OK
Least Count 0.01 mm Visual Inspection OK
I.D. No. KAP/-MM/02

Detail of reference standards & Major equipments used


Equipment Name Slip Gauge Block Set
Make KCP
Model / SR No. 10373
Certificate No. TI/K/SGS/35/21
Calibration Validity 20/11/2023
Calibration by Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial No. Nominal Value In Error In Uncertainty At 95% C.L.
(mm) (micron) (coverage factor k=2)

01. 0 Set 0
02. 2.5 0
03. 5.1 0
04. 7.7 0
05. 10.3 0
06. 12.9 0 ±1.0 µm
07. 15.0 0
08. 17.6 0
09. 20.2 0
10. 22.8 0
11. 25.0 1
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020304F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Micrometer Sr. No. ------
Make Mitutoyo Location STD Room
Range / Size 50 to 75 mm Accuracy OK
Least Count 0.01 mm Visual Inspection OK
I.D. No. KAP/MM/03

Detail of reference standards & Major equipments used


Equipment Name Slip Gauge Block Set
Make KCP
Model / SR No. 10373
Certificate No. TI/K/SGS/35/21
Calibration Validity 20/11/2023
Calibration by Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial No. Nominal Value In Error In Uncertainty At 95% C.L.
(mm) (micron) (coverage factor k=2)

01. 0 Set 0
02. 52.5 0
03. 55.1 0
04. 57.7 0
05. 60.3 0
06. 62.9 0 ±1.0 µm
07. 65.0 0
08. 67.6 0
09. 70.2 0
10. 72.8 1
11. 75.0 2
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020305F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Micrometer Sr. No. ------
Make Mitutoyo Location STD Room
Range / Size 75 to 100 mm Accuracy OK
Least Count 0.01 mm Visual Inspection OK
I.D. No. KAP/MM/04

Detail of reference standards & Major equipments used


Equipment Name Slip Gauge Block Set
Make KCP
Model / SR No. 10373
Certificate No. TI/K/SGS/35/21
Calibration Validity 20/11/2023
Calibration by Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial No. Nominal Value In Error In Uncertainty At 95% C.L.
(mm) (micron) (coverage factor k=2)

01. 0 Set 0
02. 75.5 0
03. 76.1 0
04. 77.7 0
05. 80.3 0
06. 82.9 0 ±1.0 µm
07. 85.0 0
08. 87.6 0
09. 90.2 1
10. 92.8 1
11. 98.0 2
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020306F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Micrometer Sr. No. ------
Make Mitutoyo Location Workshop
Range / Size 25 to 50 mm Accuracy OK
Least Count 0.01 mm Visual Inspection OK
I.D. No. KAP/-MM/05

Detail of reference standards & Major equipments used


Equipment Name Slip Gauge Block Set
Make KCP
Model / SR No. 10373
Certificate No. TI/K/SGS/35/21
Calibration Validity 20/11/2023
Calibration by Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:2967


Condition Relative Humidity 50±10 % Calibration Procedure CP-02
Calibration Results
Serial No. Nominal Value In Error In Uncertainty At 95% C.L.
(mm) (micron) (coverage factor k=2)

01. 0 Set 0
02. 25.5 0
03. 27.1 0
04. 29.7 0
05. 30.3 0
06. 32.9 0 ±1.0 µm
07. 35.0 0
08. 37.6 0
09. 40.2 0
10. 45.8 0
11. 50.0 1
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020307F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Digital Hardness Tester Sr. No. N/A
Make ------ Location -----
Range / Size HRB/HRC Accuracy/Acceptance N/A
Least Count As per Range Visual Inspection Ok
I.D. No. 8261F

Detail of reference standards & Major equipments used


Equipment Name S.S. Weights
Make Tycon Engineering
Model / SR No. ----
Certificate No. 21.05.001
Calibration Validity 10/01/2023
Calibration By Reliable Technical Services

Environmental Temperature 23±3 ºC Calibration Reference OIML R 111


Condition Relative Humidity 50± 20% Calibration Procedure CP-17

Calibration Results (HRB)


Sr. No. Denomination in Actual Measured Value in Uncertainty At 95% C.L.
(N) Value in (N) (N) (coverage factor k=2)

01. 10 10.0 10.1


02. 30 30.0 30.0 ±6.0g
03. 45 45.0 45.2
Calibration Results (HRC)
Sr. No. Denomination in Actual Measured Value in Uncertainty At 95% C.L.
(N) Value in (N) (N) (coverage factor k=2)

01. 10 10.0 10.0


02. 30 30.0 30.0 ±6.0g
03. 45 45.0 45.0
Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020308F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Surface Plate Model No. ------
Make Bagson Location STD Room
Range / Size 1000 x 1000 mm Accuracy ------
Least Count N/A Visual Inspection OK
I.D. No. 4364

Detail of reference standards & Major equipments used


Equipment Name Electronic level
Make Qingdao
Model / SR No. 001920
URL No. TI/H/EL/002/21
Calibration Validity 08/08/2022
Calibrated by Tanson Instrument

Environmental Temperature 20±2 ºC Calibration Reference IS:12937,7327


Condition Relative Humidity 50±10% Calibration Procedure CP-24

Calibration Results
Serial Parameter Observed value in Uncertainty At 95% C.L.
No. (micron) (coverage factor k=2)

FLATNESS OF THE
1. SURFACE PLATE ±1 L+W µm
7.23 125
AS PER IS:7327
(GRADE- ‘0’)

Remarks:

 These results of calibration refer to only to the particular item submitted for calibration.
 The above results are valid at the time of and under the stated conditions measurement.
 Calibration certificate issued for weight & measure parameters like mass, weighting balance, volumetric
equipment, measuring scales/tapes etc. are scientific purpose only and should not used for commercial.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Form No.-5.10 F-01 Page No. 1 of 1
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020309F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Profile Projector Sr. No. -------
Make Meera Metzer Location Lab
Range / Size As Per Instrument Accuracy OK
Least Count As Per Range Visual Inspection OK
I.D. No. PP-01

Detail of reference standards & Major equipments used


Equipment Name Slip Gauge Block Set Angle Gauge Set
Make KCP -----
Model / SR No. 10373 AT/MECH/08
Certificate No. TI/K/SGS/041/19 TI/A/AG/001/21
Calibration Validity 20/11/2021 29/01/2023
Calibration by Tanson Instrument Tanson Instrument

Environmental Temperature 20±2 ̊ C Calibration Reference IS:----


Condition Relative Humidity 50±10 % Calibration Procedure CP-22
Calibration Results
Serial Nominal Value Error Uncertainty At 95% C.L.
No. (Deg) (Deg) (coverage factor k=2)
01. 15˚ 0
02. 30˚ 0
03. 45˚ 0 ±1.0
04. 60˚ -1
05 90˚ -1

Remarks:

 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No 1 Of 2
Customer name M/S: KAP FASTNERS Certificate No. CC223921000020309F
And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Calibration Results
Linear (X & Y AXIS)
Serial Nominal Value In Observed Value Nominal Value In Observed Uncertainty At 95%
No. (mm) In (mm) (mm) Value In (mm) C.L.
(coverage factor k=2)
X- Axis Y- Axis
01. 10.00 9.998 10.00 10.001
02. 20.00 19.998 20.00 20.000
03. 50.00 49.997 50.00 50.000 ±4.8
04. 100.00 99.997 100.00 100.001
05 150.00 149.996
06 200.00 199.998

Calibration Results
(Magnification)
Serial Nominal Value In Observed Value In Uncertainty At 95%
No. (mm) (mm) C.L.
(coverage factor k=2)
01. 10X5 5.00 50.02 ±1.8%
02. 10X10 10.00 100.02

Remarks:

 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No 2 Of 2

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020310F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 6 x 1-6g Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRG-01

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 5.322 ± 0.007 5.320

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020311F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 6 x 1-6g No Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGN-01

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. NoGo 5.205 ± 0.007 5.206

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020312F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 8 x 1.25-6g Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRG-02

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 7.158 ± 0.007 7.159

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020313F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022
Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 8 x 1.25 -6g No Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGN/02

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. No Go 7.305 ± 0.007 7.307

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020314F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022
Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 10 x 1.25-6g Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRG/03

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 9.158 ± 0.009 9.157

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020315F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 10 x 1.25-6g NoGo Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGN/03

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. No Go 9.305 ± 0.009 9.307

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020316F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 10 x 1.5-6g Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRG/04

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 8.986 ± 0.009 8.989

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020317F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 10 x 1.5-6g No Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGN/04

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. No Go 8.853 ± 0.009 8.856

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020318F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 16 x 2-6g Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRG/05

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 14.655 ± 0.009 14.657

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020319F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 16 x 2-6g NoGo Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGN/05

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. No Go 14.494 ± 0.009 14.497

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020320F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 12 x 1.75-6g Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRG/09
Detail of reference standards & Major equipments used
Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 10.821 ± 0.004 10.823

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020321F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 12 x 1.75-6g NoGo Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGN/09
Detail of reference standards & Major equipments used
Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. No Go 10.760 ± 0.009 10.763

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020322F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 6 x 1-6h Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGH/06

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 5.348 ± 0.007 5.350

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020323F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 8 x 1.25-6h Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGH/07

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 7.186 ± 0.007 7.189

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020324F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 10 x 1.5-6h Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGH/08

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 9.108 ± 0.009 9.110

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1

Customer name M/S: KAP FASTNERS Certificate No. CC223921000020325F


And Plot No. 82, Old IDC, Kutana Hissar Service request No & date A/02/03/03/2022
Address Road, Rohtak – 124001 Haryana, India Date of calibration 03/03/2022
Valid up to 02/03/2023
Date of issue 04/03/2022

Instrument Details
Instrument name Thread Ring Gauge Sr. No. -------
Make Baker Location STD Room
Range / Size M 12 x 1.75-6h Go Accuracy OK
Least Count N/A Visual Inspection OK
I.D. No. KAP/TRGH/10

Detail of reference standards & Major equipments used


Equipment Name Dial calibration tester
Make Elemech
Model / SR No. ----
Certificate No. CC233920000000969
Calibration Validity 03/11/22
Calibration by Asian Technology

Environmental Temperature 20±2 ̊ C Calibration Reference ------


Condition Relative Humidity 50±10 % Calibration Procedure ------
Calibration Results
Serial Stand. Specification Observed value
No.

01. Go 10.855 ± 0.009 10.853

Remarks:
 (1) Standard equipment use for calibration are traceable to national/ international standards.
 (2)The reported expended uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by coverage factor k = 2 such that the coverage probability corresponds to approximately 95%.
 (3)The above results are valid at the time of and under the stated conditions measurement.
 (4)This certificate is refers only to the particular item submitted for calibration.
 (5) Next calibration due date given as requested by the customer.

Calibrated By Approved By
(Calibration Engg./TM) (QM/TM)
(MAHIPAL) (NEERAJ TYAGI)
Format No.- 5.10 F-01 Page No. 1 of 1