CALIBRATION CERTIFICATE

Format No.: UCCS-QF-7.8.1 Certificate No.: UCCSAO20231121-001


Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. AOC Date of Receipt: 21-11-2023
Air Force Station, Agra Service Request No.: UCCSAOSRN20231121-002
Certificate Issue Date: 23-11-2023
Date of Calibration: 22-11-2023
Next Due Date (Suggested by Customer): 21-11-2024
Equipment Description (UUC) & Identification
Name of the Instrument: DC Volt Meter Make: CAT
Model: ------
Range: 0 to 32 V Sr. No.: ------
Least Count: As per Results I.D. No.: VM-01
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Compact Electrical As per Manual & Testncal Laboratory TNC/2023/2/18 01-02-2024
Safety Analyzer UCCS/ES/01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH UCCS 03 SOP MIS 008 --------
CALIBRATION RESULTS
DC Voltage.
Sr. No. Ref. Standard Value (V) UUC Value (V) Error (V) Expanded Uncertainty (±)
1 9.00 8.99 -0.01 0.13 % of rdg.
2 18.00 17.98 -0.02 0.13 % of rdg.
3 24.00 23.8 -0.2 0.13 % of rdg.
4 30.00 29.6 -0.4 0.13 % of rdg.
5 32.00 31.6 -0.4 0.13 % of rdg.
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: UCCS-QF-7.8.1 Certificate No.: UCCSAO20231121-002
Group: Temperature Discipline: Thermal
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. AOC Date of Receipt: 21-11-2023
Air Force Station, Agra Service Request No.: UCCSAOSRN20231121-002
Certificate Issue Date: 23-11-2023
Date of Calibration: 22-11-2023
Next Due Date (Suggested by Customer): 21-11-2024
Equipment Description (UUC) & Identification
Name of the Instrument: Temperature Gauge Make: CAT
Model: ------
Range: -20 ⁰C to 120 ⁰C Sr. No.: ------
Least Count: ------- I.D. No.: TG-01
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Thermometer -100 ⁰C to 600 ⁰C & Bansal Calibration BCS/N/23070018-001 28-07-2024
With Four Wire RTD UCCS/CAL/022A Services
Sensor & UCCS/CAL/022B
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 20 ± 15 ⁰C, Humidity: 30 to 75 %RH UCCS 03 SOP 026 DKD-R 5-7
CALIBRATION RESULTS
Sr. No. UUC Value Ref. Standard Value Error Expanded Uncertainty
(⁰C) (⁰C) (⁰C) (±) ⁰C
1 30 30.1 -0.1 2.2
2 40 40.2 -0.2 2.2
3 60 60.5 -0.5 2.2
4 80 80.8 -0.8 2.2
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: UCCS-QF-7.8.1 Certificate No.: UCCSAO20231121-003
Group: Acceleration & Speed Discipline: Mechanical
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. AOC Date of Receipt: 21-11-2023
Air Force Station, Agra Service Request No.: UCCSAOSRN20231121-002
Certificate Issue Date: 23-11-2023
Date of Calibration: 22-11-2023
Next Due Date (Suggested by Customer): 21-11-2024
Equipment Description (UUC) & Identification
Name of the Instrument: RPM Indicator of RPM Merter Make: CAT
Model: ------
Range: Up to 40 RPM Sr. No.: ------
Least Count: As per results I.D. No.: RPM-01
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Tachometer 2.5 To 99999 RPM & Asian Technology AT22000000095 06-03-2023
UCCS/CAL/014
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH UCCS 03 SOP 033 ----
CALIBRATION RESULTS
Sr. No. Set Knob Position Observed Ref. Standard Value Expanded Uncertainty (±)
(UUC) (RPM)
1 1 6
2 2 12 4.3% of rdg
3 3 24
4 4 37
5 5 49
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: UCCS-QF-7.8.1 Certificate No.: UCCSCC20231121-003 ULR No.: CC328723000011053F
Group: Pressure Indicating Devices Discipline: Mechanical
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. AOC Date of Receipt: 21-11-2023
Air Force Station, Agra Service Request No.: UCCSSRN20231121-002
Certificate Issue Date: 23-11-2023
Date of Calibration: 22-11-2023
Next Due Date (Suggested by Customer): 21-11-2024
Equipment Description (UUC) & Identification
Name of the Instrument: Pressure Guage Make: CAT
Model: -----
Range: 0 to 125 psi Sr. No.: -----
Least Count: 5 psi I.D. No.: PG-01
Condition of Instrument: Ok Location: Fuel
Master Equipment Used
Sr. No. Instrument Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Pressure 0 To 70 bar & National Center For NCQC-M/240123/002 23-01-2024
Gauge UCCS/CAL/073 Quality Calibration
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 23 ± 1.5 ⁰C, Humidity: 30 To 80 % RH UCCS 03 SOP 007 DKD-R 6-1
CALIBRATION RESULTS
Sr. No. Set UUC Value (bar) Average Value of Error Expanded Uncertainty
(psi) (bar) Ref. Standard (bar) (bar) (± % of rdg.)
1 0 0.000 0.000 0.000 ------
2 30 2.068 2.049 0.019 14.7
3 60 4.137 4.111 0.026 8.0
4 90 6.205 6.141 0.064 5.3
5 110 7.584 7.514 0.070 4.4
6 125 8.618 8.525 0.093 3.8
***End of Certificate***

Calibrated By Reviewed/Approved By
Jata Shankar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
9. 1 Pascal = 10-5 bar.
10. Scan this QR code to download our NABL Accreditation Scope & Certificate.
CALIBRATION CERTIFICATE
Format No.: UCCS-QF-7.8.1 Certificate No.: UCCSCC20231121-004 ULR No.: CC328723000011054F
Group: Pressure Indicating Devices Discipline: Mechanical
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. AOC Date of Receipt: 21-11-2023
Air Force Station, Agra Service Request No.: UCCSSRN20231121-002
Certificate Issue Date: 23-11-2023
Date of Calibration: 22-11-2023
Next Due Date (Suggested by Customer): 21-11-2024
Equipment Description (UUC) & Identification
Name of the Instrument: Pressure Guage Make: CAT
Model: -----
Range: 0 to 100 psi Sr. No.: -----
Least Count: 5 psi I.D. No.: PG-02
Condition of Instrument: Ok Location: Oil
Master Equipment Used
Sr. No. Instrument Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Pressure 0 To 70 bar & National Center For NCQC-M/240123/002 23-01-2024
Gauge UCCS/CAL/073 Quality Calibration
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 23 ± 1.5 ⁰C, Humidity: 30 To 80 % RH UCCS 03 SOP 007 DKD-R 6-1
CALIBRATION RESULTS
Sr. No. Set UUC Value (bar) Average Value of Error Expanded Uncertainty
(psi) (bar) Ref. Standard (bar) (bar) (± % of rdg.)
1 0 0.000 0.000 0.000 ------
2 20 1.378 1.362 0.016 22.1
3 40 2.757 2.729 0.028 12.0
4 60 4.136 4.095 0.041 8.0
5 80 5.515 5.448 0.067 6.0
6 100 6.894 6.811 0.083 4.8
***End of Certificate***

Calibrated By Reviewed/Approved By
Jata Shankar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
9. 1 Pascal = 10-5 bar.
10. Scan this QR code to download our NABL Accreditation Scope & Certificate.

You might also like